(265 days)
The Hip Spacer with gentamicin preserves the joint space and the length of the affected limb, which results in the maintenance of the entire abductor and stabilizer apparatus of the hip. Its use is indicated for a limited period, in patients who require a Two-stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Hip Spacer is implanted after the removal of the infected implant, as a regular Hemiarthroplasty. The spacer is inserted into the femoral canal, and a handmade ring of PMMA bone cement with antibiotic may be added to the base of the spacer's neck in order to increase stability. The spacer is kept "in place" until it is replaced by the final prosthesis according to medical criteria. The Hip Spacer must not remain implanted for more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
The Knee Spacer with gentamicin maintains the articular space, the length of the affected limb, and the ligament apparatus of the knee. Its use is indicated for a limited period, in patients who require a Two stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Knee Spacer is placed like a regular arthroplasty, after the removal of the original implant. This knee spacer consists of two independent parts: a tibial plate and a femoral component. The first has a flat base where the femoral component articulates. It is recommended that both components be adapted to the bone by means of a small amount of bone cement with antibiotic. The Knee Spacer must not remain implanted more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
The Shoulder Spacer with gentamicin preserves joint space and length of the affected limb, resulting in maintenance of the entire shoulder muscle and stabilizer complex. Its use is indicated for a limited period, in patients who require a Two-stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Shoulder Spacer is placed as a hemiarthroplasty after the original implant has been removed. The spacer is inserted into the humeral canal and a ring of bone cement with antibiotic can be added to the base of the neck of the spacer for added stability. The Shoulder Spacer should not remain in place for more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
The Synicem Hip, Knee, and Shoulder Spacers are combination products made from fully formed polymethylmethacrylate (bone cement) with gentamicin. The bone cement is prepared from a powder component and a liquid component. The hip and shoulder spacers contain a stainless steel core of 316L in accordance with ASTM F138. The spacers are temporary implants utilized to maintain the joint space during two-stage revision arthroplasties. The spacer implant is placed as part of the first stage of the two-stage revision when the original prostheses are removed due to joint infection. Once the infection is cleared, the spacers are removed and replaced with a permanent prosthesis as part of the second stage of the revision process. The joint spacers are not intended to be implanted for longer than 6 months. They are a single use device and supplied sterile.
This document is a 510(k) clearance letter for the Synicem Hip, Knee, and Shoulder Spacers. These are medical devices, specifically temporary implants used in two-stage revision arthroplasties following joint infections.
The information provided outlines the substantial equivalence of the Synicem Spacers to previously cleared predicate devices. It describes the device, its intended use, technological characteristics, material composition, sizes, and various performance tests conducted.
However, a critical point to understand is that this document DOES NOT describe an AI/ML-based device. It is a clearance for a traditional medical device (joint spacers). Therefore, many of the requested criteria related to AI/ML systems (such as test set sample size, data provenance, expert ground truth, MRMC studies, standalone performance, training set details) are not applicable to this submission.
The document focuses on demonstrating that the new Synicem Spacers are "substantially equivalent" to existing, legally marketed predicate devices in terms of safety and effectiveness, based on non-clinical performance testing (mechanical, antibiotic elution), biocompatibility, and MRI safety.
Here's an analysis based on the provided document, addressing the applicable criteria and explicitly stating where information is not available due to the nature of the device:
Device: Synicem Hip Spacer, Synicem Knee Spacer, Synicem Shoulder Spacer
Type of Device: Traditional Medical Device (Non-AI/ML)
Acceptance Criteria and Reported Device Performance
Given that this is a traditional medical device (joint spacer) and not an AI/ML algorithm, the "acceptance criteria" are based on demonstrating substantial equivalence to predicate devices through various non-clinical performance tests and adherence to relevant standards. There isn't a "performance metric" in the sense of accuracy, sensitivity, or specificity as one would expect for a diagnostic AI.
The acceptance criteria are implicitly met by demonstrating that the Synicem Spacers "met performance requirements and is substantially equivalent to the predicate device" across several categories.
Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):
| Category | Acceptance Criteria (Implicit) | Reported Device Performance (Summary from Document) |
|---|---|---|
| Mechanical Performance | Equivalent or superior to predicate devices in relevant mechanical tests. | - Hip Spacers: Femoral stem and neck fatigue, and wear testing performed. Results showed equivalence to predicate. - Knee Spacers: Fatigue performance and wear testing performed. Results showed equivalence to predicate. - Shoulder Spacers: Static and dynamic fatigue testing performed. Results showed equivalence to predicate. |
| Antibiotic Elution Profile | Elution kinetics of gentamicin are equivalent to predicate devices. | Antibiotic Elution Kinetics testing conducted. Results demonstrated elution kinetics equivalent to predicate spacers. |
| Biocompatibility | Biological safety suitable for intended use, in accordance with ISO 10993 standards and FDA guidance. | - Cytotoxicity (ISO 10993-5) - Sensitization (ISO 10993-10) - Irritation (ISO 10993-23) - Acute systemic toxicity (ISO 10993-11) - Material mediated pyrogenicity (ISO 10993-11) - Genotoxicity (ISO 10993-3) - Intramuscular implantation (ISO 10993-6) - Local Effects after Bone Implantation / Subchronic Systemic Toxicity (ISO 10993-6 and ISO 10993-11) - Chemical characterization (ISO 10993-18) - Toxicological risk assessment (ISO 10993-17) Conclusion: Materials are biologically safe and suitable for intended use. |
| MRI Safety | Meet MR safety classifications as "MR Safe" or "MR Conditional" as determined by testing. | - Synicem Knee Spacer: "MR Safe" - Synicem Hip and Shoulder Spacers: "MR Conditional" (Testing conducted according to FDA guidance) |
| Sterilization & Shelf Life | Sterilization methods validated according to international standards and FDA guidance; packaging maintains functionality and safety. | Sterilized using well-established methods; validations performed following international standards and FDA guidance. Compatibility between packaging and sterilization ensured; packaging preserves functionality and safety throughout declared shelf-life. |
Study Details (Applicability to AI/ML context)
Since this is a non-AI/ML device submission, most of the requested study details pertinent to AI/ML performance validation are not applicable (N/A). The "study" here refers to the non-clinical performance testing done to establish substantial equivalence.
-
Sample size used for the test set and the data provenance:
- N/A. This is not an AI/ML device. The "test set" would refer to the physical samples of the spacers used for mechanical, biocompatibility, and elution testing. The provenance would be the manufacturing site (United Kingdom).
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- N/A. This is not an AI/ML device requiring expert-labeled ground truth data. Performance for this device is based on objective measurements from mechanical and chemical tests, and biological assays.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- N/A. Not applicable for objective non-clinical performance testing of a physical device.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a physical medical device (joint spacer), not an AI system designed to assist human readers. MRMC studies are not relevant.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- N/A. There is no "algorithm" in this device. Its performance is inherent in its physical and chemical properties.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- N/A. The "ground truth" for this device's performance is established by validated engineering and scientific test methods (e.g., ASTM standards for material properties, ISO standards for biocompatibility, gravimetric analysis for elution profiles). It's based on physical measurements and biological reactions, not subjective interpretation requiring "expert consensus" in the diagnostic sense.
-
The sample size for the training set:
- N/A. This is not an AI/ML device; there is no "training set."
-
How the ground truth for the training set was established:
- N/A. Not an AI/ML device; no "training set" or "ground truth for training."
FDA 510(k) Clearance Letter - Synicem Spacers
Page 1
June 12, 2025
Biocomposites Ltd.
Melisa Lamanna
Research and Development Manager
Keele Science Park
Keele, Staffordshire ST5 5NL
United Kingdom
Re: K242865
Trade/Device Name: Synicem Hip Spacer; Synicem Knee Spacer; Synicem Shoulder Spacer
Regulation Number: 21 CFR 888.3360
Regulation Name: Hip Joint Femoral (Hemi-Hip) Metallic Cemented Or Uncemented Prosthesis
Regulatory Class: Class II
Product Code: KWL, KWY, JWH, HSD, KWS
Dated: May 14, 2025
Received: May 15, 2025
Dear Mrs. Lamanna:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K242865 - Melisa Lamanna Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K242865 - Melisa Lamanna Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert M. Stefani -S
Digitally signed by Robert M. Stefani -S
Date: 2025.06.12 09:44:17 -04'00'
For: Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Indications for Use
Submission Number (if known)
K242865
Device Name
Synicem Hip Spacer, Knee Spacer, Shoulder Spacer
Indications for Use (Describe)
The Hip Spacer with gentamicin preserves the joint space and the length of the affected limb, which results in the maintenance of the entire abductor and stabilizer apparatus of the hip. Its use is indicated for a limited period, in patients who require a Two-stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Hip Spacer is implanted after the removal of the infected implant, as a regular Hemiarthroplasty. The spacer is inserted into the femoral canal, and a handmade ring of PMMA bone cement with antibiotic may be added to the base of the spacer's neck in order to increase stability. The spacer is kept "in place" until it is replaced by the final prosthesis according to medical criteria. The Hip Spacer must not remain implanted for more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
The Knee Spacer with gentamicin maintains the articular space, the length of the affected limb, and the ligament apparatus of the knee. Its use is indicated for a limited period, in patients who require a Two stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Knee Spacer is placed like a regular arthroplasty, after the removal of the original implant. This knee spacer consists of two independent parts: a tibial plate and a femoral component. The first has a flat base where the femoral component articulates. It is recommended that both components be adapted to the bone by means of a small amount of bone cement with antibiotic. The Knee Spacer must not remain implanted more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
The Shoulder Spacer with gentamicin preserves joint space and length of the affected limb, resulting in maintenance of the entire shoulder muscle and stabilizer complex. Its use is indicated for a limited period, in patients who require a Two-stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Shoulder Spacer is placed as a hemiarthroplasty after the original implant has been removed. The spacer is inserted into the humeral canal and a ring of bone cement with antibiotic can be added to the base of the neck of the spacer for added stability. The Shoulder Spacer should not remain in place for more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
Page 5
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 6
510(k) SUMMARY
GENERAL INFORMATION
Submitter: Biocomposites Ltd.
Date Prepared: June 12, 2025
Contact Information:
Contact Name: Melisa Lamanna
Contact Title: Research and Development Manager
Contact Number: +44 (11) 5272 4060
CONTACT INFORMATION
Contact Name: Melisa Lamanna
Contact Title: Research and Development Manager
Contact Address: Keele Science Park, Keele, Staffordshire ST5 5NL, United Kingdom
Contact Email Address: mlamanna@subiton.com
Contact Phone: +44 (11) 5272 4060
COMMON NAME: Temporary Hip, Knee, and Shoulder Spacers
PROPOSED CLASS: II
CLASSIFICATION NAME: Prosthesis hip femoral or metallic
Prosthesis hip femoral or metallic cemented or uncemented
Prosthesis knee patellofemoral, semiconstrained cemented
polymer/metal/polymer
Prosthesis shoulder hemihumeral metallic uncemented or cemented
REGULATION NUMBER: 21 CFR §888.3360, §888.3560, §888.3690
PRODUCT CODES: KWL, KWY, JWH, HSD, KWS
PREDICATE DEVICES:
- Spacer-G Temporary Hip Spacer with Gentamicin (K031841, K062273 and K101356)
- Spacer-K Temporary Knee Spacer with Gentamicin (K032522, K062274 and K101356)
- Spacer-S Temporary Shoulder Spacer with Gentamicin (K060535 and K112983)
DEVICE DESCRIPTION
The Synicem Hip, Knee, and Shoulder Spacers are combination products made from fully formed polymethylmethacrylate bone cement with gentamicin. The bone cement is prepared from a powder component and a liquid component. The hip and shoulder spacers contain a stainless steel core of 316L in accordance with ASTM F138. The spacers are temporary implants utilized to maintain the joint space during two-stage revision arthroplasties. The spacer implant is placed as part of the first stage of the two-stage revision when the original prostheses are removed due to joint infection. Once the infection is cleared, the spacers are removed and replaced with a permanent prosthesis as part of the second stage of the revision process. The joint spacers are not intended to be implanted for longer than 6 months. They are a single use device and supplied sterile.
K242865
Page 1 of 5
Page 7
510(k) Summary
Applicant information:
Company Name: Biocomposites Ltd.
Street Address: Keele Science Park
City: Keele
State / Province: Staffordshire
ZIP/Postal Code: ST5 5NL
Country: England
Contact information:
Contact name: Melisa Lamanna
Contact Title: Research and Development Manager
Phone Number: +54 (11) 5272 4060
Contact Email Address: mlamanna@subiton.com
Date Prepared: June 9, 2025
Proprietary Name: Synicem Hip, Knee, and Shoulder Spacers
Common Name: Temporary Hip, Knee, and Shoulder Spacers with Gentamicin
Proposed Class: II
Classification Name: Prosthesis, hip, hemi-, femoral, metal;
Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented;
Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer;
Prosthesis, shoulder, hemi-, humeral, metallic uncemented or cemented.
Regulation Number: 21 CFR §888.3360, §888.3560, §888.3690
Product Codes: KWL, KWY, JWH, HSD, KWS
Predicate Device(s):
- Spacer-G, Temporary Hip Spacer with Gentamicin (K031841, K062273 and K101356).
- Spacer-K, Temporary Knee Spacer with Gentamicin (K032522, K062274 and K101356).
- Spacer-S, Temporary Shoulder Spacer with Gentamicin (K060535 and K112983).
Device Description:
The Synicem Hip, Knee, and Shoulder Spacers are combination products made from fully formed polymethylmethacrylate (bone cement) with gentamicin. The bone cement is prepared from a powder component and a liquid component. The hip and shoulder spacers contain a stainless-steel core of AISI 316L in accordance with ASTM F138. The spacers are temporary implants utilized to maintain the joint space during two-stage revision arthroplasties. The spacer implant is placed as part of the first stage of the two-stage revision when the original prostheses are removed due to joint infection. Once the infection is cleared, the spacers are removed and replaced with a permanent prosthesis as part of the second stage of the revision process. The joint spacers are not intended to be implanted for longer than 6 months. They are a single use device and supplied sterile.
K242865
Page 1 of 5
Page 8
INDICATIONS FOR USE
Hip Spacers
INDICATIONS:
The Hip Spacer with gentamicin preserves the joint space and the length of the affected limb, which results in the maintenance of the entire abductor and stabilizer apparatus of the hip. Its use is indicated for a limited period, in patients who require a two-stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Hip Spacer is implanted after the removal of the infected implant, as a regular Hemiarthroplasty. The spacer is inserted into the femoral canal, and a handmade ring of PMMA bone cement with antibiotic may be added to the base of the spacer's neck in order to increase stability. The spacer is kept "in place" until it is replaced by the final prosthesis according to medical criteria. The Hip Spacer must not remain implanted for more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
Knee spacers
INDICATIONS:
The Knee Spacer with gentamicin maintains the articular space, the length of the affected limb, and the ligament apparatus of the knee. Its use is indicated for a limited period, in patients who require a Two stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Knee Spacer is placed like a regular arthroplasty, after the removal of the original implant. This knee spacer consists of two independent parts: a tibial plate and a femoral component. The first has a flat base where the femoral component articulates. It is recommended that both components be adapted to the bone by means of a small amount of bone cement with antibiotic. The Knee Spacer must not remain implanted more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
Shoulder Spacers
INDICATIONS:
The Shoulder Spacer with gentamicin preserves joint space and length of the affected limb, resulting in maintenance of the entire shoulder muscle and stabilizer complex. Its use is indicated for a limited period, in patients who require a Two-stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Shoulder Spacer is placed as a hemiarthroplasty after the original implant has been removed. The spacer is inserted into the humeral canal and a ring of bone cement with antibiotic can be added to the base of the neck of the spacer for added stability. The Shoulder Spacer should not remain in place for more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
TECHNOLOGICAL CHARACTERISTICS
Device comparison demonstrated that Synicem Hip Spacer is substantially equivalent to the previously cleared Spacer-G (K031841, K062273 and K101356), Synicem Knee Spacer is substantially equivalent to the previously cleared Spacer-K (K032522, K062274 and K101356) and Synicem Shoulder Spacer is substantially equivalent to the previously cleared Spacer-S (K060535 and K112983), regarding intended use, indications for use, technological characteristics, design features, material, sterilization, and performance as well as operating principle. The minor differences in design and size between the subject and the predicate devices do not raise any significant new risks. Through performance and mechanical testing it has been established that these differences do not present any new issues of safety or effectiveness.
K242865
Page 2 of 5
Page 9
Indications for Use:
Hip Spacers
INDICATIONS:
The Hip Spacer with gentamicin preserves the joint space and the length of the affected limb, which results in the maintenance of the entire abductor and stabilizer apparatus of the hip. Its use is indicated for a limited period, in patients who require a two-stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Hip Spacer is implanted after the removal of the infected implant, as a regular Hemiarthroplasty. The spacer is inserted into the femoral canal, and a handmade ring of PMMA bone cement with antibiotic may be added to the base of the spacer's neck in order to increase stability. The spacer is kept "in place" until it is replaced by the final prosthesis according to medical criteria. The Hip Spacer must not remain implanted for more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
Knee spacers
INDICATIONS:
The Knee Spacer with gentamicin maintains the articular space, the length of the affected limb, and the ligament apparatus of the knee. Its use is indicated for a limited period, in patients who require a Two stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Knee Spacer is placed like a regular arthroplasty, after the removal of the original implant. This knee spacer consists of two independent parts: a tibial plate and a femoral component. The first has a flat base where the femoral component articulates. It is recommended that both components be adapted to the bone by means of a small amount of bone cement with antibiotic. The Knee Spacer must not remain implanted more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
Shoulder spacers
INDICATIONS:
The Shoulder Spacer with gentamicin preserves joint space and length of the affected limb, resulting in maintenance of the entire shoulder muscle and stabilizer complex. Its use is indicated for a limited period, in patients who require a Two-stage Revision Arthroplasty. A systemic antibiotic therapy should also be prescribed while the spacer remains implanted. The Shoulder Spacer is placed as a hemiarthroplasty after the original implant has been removed. The spacer is inserted into the humeral canal and a ring of bone cement with antibiotic can be added to the base of the neck of the spacer for added stability. The Shoulder Spacer should not remain in place for more than 6 months. Once this period has elapsed, it must be explanted and a permanent device implanted or another appropriate treatment performed.
Technological Characteristics:
Device comparison demonstrated that Synicem Hip Spacer is substantially equivalent to the previously cleared Spacer-G (K031841, K062273 and K101356), Synicem Knee Spacer is substantially equivalent to the previously cleared Spacer-K (K032522, K062274 and K101356) and Synicem Shoulder Spacer is substantially equivalent to the previously cleared Spacer-S (K060535 and K112983), regarding intended use, indications for use, technological characteristics (design
K242865
Page 2 of 5
Page 10
features, material, sterilization, and performance) as well as operating principle. The minor differences in design and size between the subject and the predicate devices do not raise any significant new risks. Through performance and mechanical testing it has been established that these differences do not present any new issues of safety or effectiveness.
| Subject Device | Predicate | Predicate | Predicate |
|---|---|---|---|
| 510(k) Number | K242865 | K031841, K062273, and K101356 | K032522, K062274 and K101356 |
| Trade Name | Synicem Hip Spacer, Knee Spacer, and Shoulder Spacer | Spacer-G, Temporary Hip Spacer with Gentamicin | Spacer-K, Temporary Knee Spacer with Gentamicin |
| Regulation and Procode | 21 CFR 888.3360 (Hip; KWL, KWY) 21 CFR 888.3560 (Knee; JWH) 21 CFR 888.3690 (Shoulder; HSD) | 21 CFR 888.3360 (Hip; KWL, KWY) | 21 CFR 888.3560 (Knee; JWH) |
| Classification Name | Prosthesis, hip, hemi-, femoral, metal; Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented; Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer; Prosthesis, shoulder, hemi-, humeral, metallic uncemented or cemented. | Prosthesis, hip, hemi-, femoral, metal; Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented | Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/ polymer |
| Intended Use | Provide a temporary implant to allow for natural range of motion and partial weight bearing of the hip, knee, or shoulder. Preserve soft tissue structure | Provide a temporary implant to allow for natural range of motion and partial weight bearing of the hip. Preserve soft tissue structure | Provide a temporary implant to allow for natural range of motion and partial weight bearing of the knee. Preserve soft tissue structure |
MATERIAL COMPOSITION
| Material Composition | Gentamicin-impregnated PMMA spacer with stainless steel core | Gentamicin-impregnated PMMA spacer with stainless steel core | Gentamicin-impregnated PMMA spacer with stainless steel core | Gentamicin-impregnated PMMA spacer with stainless steel core |
SIZES
| Hip | Hip | Knee | Shoulder |
|---|---|---|---|
| Head: 40, 48, 56 mm | Head: 46, 54, 60 mm | Femoral: 54, 64, 74, 84 mm | Head: 41, 46 mm |
| Stem: 120, 127, 130, 250 mm | Stem: 96, 94, 98 mm | Tibial: 60, 70, 80, 90 mm | Stem: 99, 125 mm |
| Knee | |||
| Femoral: 58, 65, 72, 79 mm | |||
| Tibial: 65, 73, 77.5, 85 mm | |||
| Shoulder | |||
| Head: 40, 48 mm | |||
| Stem: 120 mm |
| Supplied Sterile | Yes | Yes | Yes | Yes |
Performance Data
The Synicem Spacers have been evaluated through non-clinical performance testing. The following mechanical testing was performed and results showed the Synicem Spacers were equivalent to the predicate spacers.
- Femoral stem and neck fatigue, and wear testing for the hip spacers.
- Fatigue performance and wear for the knee spacers.
- Static and dynamic fatigue testing for the shoulder spacers.
Antibiotic Elution Profile
Antibiotic Elution Kinetics testing was conducted to determine the antibiotic elution properties hip, knee, and shoulder subject's spacers compared to the predicate spacers. These results demonstrate that the elution kinetics of the antibiotic in the subject hip, knee and shoulder spacers are equivalent to the elution kinetics of the antibiotic obtained for the predicate spacers.
Biocompatibility
Biocompatibility testing of Synicem Spacers was conducted in accordance with FDA guidance document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process".
- Cytotoxicity as per ISO 10993-5.
- Sensitization as per ISO 10993-10.
- Irritation as per ISO 10993-23.
- Acute systemic toxicity as per ISO 10993-11.
- Material mediated pyrogenicity as per ISO 10993-11.
- Genotoxicity as per ISO 10993-3.
K242865
Page 4 of 5
Page 11
- Intramuscular implantation as per 10993-6.
- Local Effects after Bone Implantation / Subchronic Systemic Toxicity as per ISO 10993-6 and ISO 10993-11.
It has been demonstrated through the mentioned biological tests, the chemical characterization (ISO 10993-18) and the toxicological risk assessment (ISO 10993-17) that the materials of Synicem Spacers are biological safe, and suitable for their intended use.
MRI Safety
MR testing was conducted to FDA guidance document, Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment. The Synicem knee spacer is "MR Safe" and the Synicem Hip and Shoulder Spacers are "MR Conditional".
The testing demonstrated that the Synicem Spacers met performance requirements and is substantially equivalent to the predicate device.
Packaging, Sterilization and Shelf Life
Synicem Spacers are sterilized using well established sterilization methods and the validations have been performed following international standards and FDA guidance document "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile". The compatibility between the packaging material and the sterilization process is ensured and the packaging preserves the functionality and safety of Synicem Spacers throughout its declared shelf-life.
Substantial Equivalence:
The Synicem Spacers have the same indications for use and similar design features as compared with the predicate systems. The bench testing demonstrates that the performance characteristics of the Synicem Spacers are equivalent to those of the other legally marketed temporary spacer devices, and therefore supports a determination of Substantial Equivalence for the proposed indications for use. Any differences between the subject and predicate devices would not render the device NSE, affect the safety or effectiveness, or raise different questions of safety and effectiveness.
K242865
Page 5 of 5
§ 888.3360 Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis.
(a)
Identification. A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.(b)
Classification. Class II.