K Number
K242783

Validate with FDA (Live)

Date Cleared
2024-12-17

(92 days)

Product Code
Regulation Number
864.3700
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Roche Digital Pathology Dx is an automated digital slide creation, viewing and management system. Roche Digital Pathology Dx is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of scanned pathology slides prepared from formalin-fixed paraffin-embedded (FFPE) tissue. Roche Digital Pathology Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. Roche Digital Pathology Dx is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy.

Roche Digital Pathology Dx is composed of VENTANA DP 200 slide scanner, VENTANA DP 600 slide scanner, Roche uPath enterprise software, and ASUS PA248QV display. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using Roche Digital Pathology Dx.

Device Description

Roche Digital Pathology Dx (hereinafter referred to as RDPD), is a whole slide imaging (WSI) system. It is an automated digital slide creation, viewing, and management system intended to aid pathologists in generating, reviewing, and interpreting digital images of surgical pathology slides that would otherwise be appropriate for manual visualization by conventional light microscopy. RDPD system is composed of the following components:

  • · VENTANA DP 200 slide scanner,
  • · VENTANA DP 600 slide scanner,
  • · Roche uPath enterprise software, and
  • · ASUS PA248QV display.
    VENTANA DP 600 slide scanner has a total capacity of 240 slides through 40 trays with 6 slides each. The VENTANA DP 600 slide scanner and VENTANA DP 200 slide scanner use the same Image Acquisition Unit.

Both VENTANA DP 200 and DP 600 slide scanners are bright-field digital pathology scanners that accommodate loading and scanning of 6 and 240 standard glass microscope slides, respectively. The scanners each have a high-numerical aperture Plan Apochromat 20x objective and are capable of scanning at both 20x and 40x magnifications. The scanners feature automatic detection of the tissue specimen on the glass slide, automated 1D and 2D barcode reading, and selectable volume scanning (3 to 15 focus layers). The International Color Consortium (ICC) color profile is embedded in each scanned slide image for color management. The scanned slide images are generated in a proprietary file format, Biolmagene Image File (BIF), that can be uploaded to the uPath Image Management System (IMS), provided with the Roche uPath enterprise software.

Roche uPath enterprise software (uPath), a component of Roche Digital Pathology Dx system, is a web-based image management and workflow software application. uPath enterprise software can be accessed on a Windows workstation using the Google Chrome or Microsoft Edge web browser. The user interface of uPath software enables laboratories to manage their workflow from the time the whole slide image is produced and acquired by VENTANA DP 200 and/or DP 600 slide scanners through the subsequent processes, such as review of the digital image on the monitor screen and reporting of results. The uPath software incorporates specific functions for pathologists, laboratory histology staff, workflow coordinators, and laboratory administrators.

AI/ML Overview

The provided document is a 510(k) summary for the "Roche Digital Pathology Dx" system (K242783), which is a modification of a previously cleared device (K232879). This modification primarily involves adding a new slide scanner model, VENTANA DP 600, to the existing system. The document asserts that due to the identical Image Acquisition Unit (IAU) between the new DP 600 scanner and the previously cleared DP 200 scanner, the technical performance assessment from the predicate device is fully applicable. Therefore, the information provided below will primarily refer to the studies and acceptance criteria from the predicate device that are deemed applicable to the current submission due to substantial equivalence.

Here's an analysis based on the provided text, focusing on the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a discrete "acceptance criteria" table with corresponding numerical performance metrics for the current submission (K242783). Instead, it states that the performance characteristics data collected on the VENTANA DP 200 slide scanner (the predicate device) are representative of the VENTANA DP 600 slide scanner performance because both scanners use the same Image Acquisition Unit (IAU). The table below lists the "Technical Performance Assessment (TPA) Sections" as presented in the document (Table 3), which serve as categories of performance criteria that were evaluated for the predicate device and are considered applicable to the current device. The reported performance for these sections is summarized as "Information was provided in K232879" for the DP 600, indicating that the past performance data is considered valid.

TPA Section (Acceptance Criteria Category)Reported Device Performance (for DP 600 scanner)
Components: Slide FeederNo double wide slide tray compatibility, new FMEA provided in K242783. (Predicate: Information on configuration, user interaction, FMEA)
Components: Light SourceInformation was provided in K232879. (Predicate: Descriptive info on lamp/condenser, spectral distribution verified)
Components: Imaging OpticsInformation was provided in K232879. (Predicate: Optical schematic, descriptive info, testing for irradiance, distortions, aberrations)
Components: Focusing SystemInformation was provided in K232879. (Predicate: Schematic, description, optical system, cameras, algorithm)
Components: Mechanical Scanner MovementSame except no double wide slide tray compatibility, replaced references & new FMEA items provided in K242783. (Predicate: Information/specs on stage, movement, FMEA, repeatability)
Components: Digital Imaging SensorInformation was provided in K232879. (Predicate: Information/specs on sensor type, pixels, responsivity, noise, data, testing)
Components: Image Processing SoftwareInformation was provided in K232879. (Predicate: Information/specs on exposure, white balance, color correction, subsampling, pixel correction)
Components: Image CompositionInformation was provided in K232879. (Predicate: Information/specs on scanning method, speed, Z-axis planes, analysis of image composition)
Components: Image File FormatsInformation was provided in K232879. (Predicate: Information/specs on compression, ratio, file format, organization)
Image Review Manipulation SoftwareInformation was provided in K232879. (Predicate: Information/specs on panning, zooming, Z-axis displacement, comparison, image enhancement, annotation, bookmarks)
Computer EnvironmentSelect upgrades of sub-components & specifications. (Predicate: Information/specs on hardware, OS, graphics, color management, display interface)
DisplayInformation was provided in K232879. (Predicate: Information/specs on pixel density, aspect ratio, display surface, and other display characteristics; performance testing for user controls, spatial resolution, pixel defects, artifacts, temporal response, luminance, uniformity, gray tracking, color scale, color gamut)
System-level Assessments: Color ReproducibilityInformation was provided in K232879. (Predicate: Test data for color reproducibility)
System-level Assessments: Spatial ResolutionInformation was provided in K232879. (Predicate: Test data for composite optical performance)
System-level Assessments: Focusing TestInformation was provided in K232879. (Predicate: Test data for technical focus quality)
System-level Assessments: Whole Slide Tissue CoverageInformation was provided in K232879. (Predicate: Test data for tissue detection algorithms and inclusion of tissue in digital image file)
System-level Assessments: Stitching ErrorInformation was provided in K232879. (Predicate: Test data for stitching errors and artifacts)
System-level Assessments: Turnaround TimeInformation was provided in K232879. (Predicate: Test data for turnaround time)
User InterfaceIdentical workflow, replacement of new scanner component depiction. (Predicate: Information on user interaction, human factors/usability validation)
LabelingSame content, replaced references. (Predicate: Compliance with 21 CFR Parts 801 and 809, special controls)
Quality ControlSame content, replaced references. (Predicate: QC activities by user, lab technician, pathologist prior to/after scanning)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states that the technical performance assessment data was "collected on the VENTANA DP 200 slide scanner" (the predicate device). However, the specific sample sizes for these technical studies (e.g., number of slides used for focusing tests, stitching error analysis, etc.) are not detailed in this summary. The data provenance (country of origin, retrospective/prospective) for these underlying technical studies from K232879 is also not provided in this document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable or not provided by the document. The studies mentioned are primarily technical performance assessments related to image quality, system components, and usability, rather than diagnostic accuracy studies requiring expert pathologist interpretation for ground truth. For the predicate device, it mentions "a qualified pathologist" is responsible for interpretation, but this refers to the end-user clinical use, not the establishment of ground truth for device validation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable or not provided by the document. As noted above, the document details technical performance studies rather than diagnostic performance studies that would typically involve multiple readers and adjudication methods for diagnostic discrepancies.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study is mentioned in the provided text. The device, "Roche Digital Pathology Dx," is described as a "digital slide creation, viewing and management system" intended "as an aid to the pathologist to review and interpret digital images." It is a Whole Slide Imaging (WSI) system, and the submission is focused on demonstrating the technical equivalence of a new scanner component, not on evaluating AI assistance or its impact on human reader performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No standalone algorithm performance study is mentioned. The device is a WSI system for "aid to the pathologist to review and interpret digital images," implying a human-in-the-loop system. The document does not describe any specific algorithms intended for automated diagnostic interpretation or analysis in a standalone capacity.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the technical performance studies referenced from the predicate device (K232879), the "ground truth" would be established by physical measurements and engineering specifications for aspects like spatial resolution, color reproducibility, focusing quality, tissue coverage, and stitching errors, rather than clinical outcomes or diagnostic pathology. For instance, color reproducibility would be assessed against a known color standard, and spatial resolution against resolution targets. The document does not explicitly state the exact types of ground truth used for each technical assessment but refers to "test data to evaluate" these characteristics.

8. The sample size for the training set

Not applicable/not provided. The document describes a WSI system, not an AI/ML-based diagnostic algorithm that would typically require a training set. The specific "Image Acquisition Unit" components (hardware and software for pixel pipeline) are stated to be "functionally identical" to the predicate, implying established design rather than iterative machine learning.

9. How the ground truth for the training set was established

Not applicable/not provided. As there is no mention of a training set for an AI/ML algorithm, the method for establishing its ground truth is not discussed.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 17, 2024

Ventana Medical Systems, Inc. Cameron Smith Regulatory Affairs Manager (Ventana, VMSI; Roche Tissue Diagnostics, RTD) 1910 E. Innovation Park Drive Tucson, Arizona 85755

Re: K242783

Trade/Device Name: Roche Digital Pathology Dx Regulation Number: 21 CFR 864.3700 Regulation Name: Whole slide imaging system Regulatory Class: Class II Product Code: PSY Dated: September 13, 2024 Received: September 16, 2024

Dear Cameron Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Shyam Kalavar -S

Shyam Kalavar Deputy Branch Chief Division of Molecular Genetics and Pathology OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242783

Device Name Roche Digital Pathology Dx

Indications for Use (Describe)

Roche Digital Pathology Dx is an automated digital slide creation, viewing and management system. Roche Digital Pathology Dx is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of scanned pathology slides prepared from formalin-fixed paraffin-embedded (FFPE) tissue. Roche Digital Pathology Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. Roche Digital Pathology Dx is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy.

Roche Digital Pathology Dx is composed of VENTANA DP 200 slide scanner, VENTANA DP 600 slide scanner, Roche uPath enterprise software, and ASUS PA248QV display. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using Roche Digital Pathology Dx.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary Roche Digital Pathology Dx

Date Prepared: December 12, 2024

SUBMITTER

Ventana Medical Systems, Inc.

CONTACT PERSON

Ventana Medical Systems, Inc. Cameron Smith Regulatory Affairs Manager (Ventana, VMSI; Roche Tissue Diagnostics, RTD) 1910 E. Innovation Park Drive Tucson, Arizona 85755

DEVICE INFORMATION

Subject Device

Proprietary Name:Roche Digital Pathology Dx
Common Name:Roche Digital Pathology Dx
Classification Name:Whole Slide Imaging System
Regulation Section:21 CFR 864.3700
Classification:Class II
Product Code:PSY
Review Panel:88 – Pathology
510(k) Number:K242783

Predicate Device

Proprietary Name:Roche Digital Pathology Dx (VENTANA DP 200)
Submission Number:K232879

{5}------------------------------------------------

A new model of the slide scanner component, VENTANA DP 600 slide scanner, is being added to the Roche Digital Pathology Dx [previously known as Roche Digital Pathology Dx (VENTANA DP 200)] system. As a part of the system modification, the system name is being changed to "Roche Digital Pathology Dx".

INTENDED USE I.

Roche Digital Pathology Dx is an automated digital slide creation, viewing and management system. Roche Digital Pathology Dx is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of scanned pathology slides prepared from formalin-fixed paraffin-embedded (FFPE) tissue. Roche Digital Pathology Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. Roche Digital Pathology Dx is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy.

Roche Digital Pathology Dx is composed of VENTANA DP 200 slide scanner, VENTANA DP 600 slide scanner, Roche uPath enterprise software, and ASUS PA248QV display. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using Roche Digital Pathology Dx.

II. DEVICE DESCRIPTION

Roche Digital Pathology Dx (hereinafter referred to as RDPD), is a whole slide imaging (WSI) system. It is an automated digital slide creation, viewing, and management system intended to aid pathologists in generating, reviewing, and interpreting digital images of surgical pathology slides that would otherwise be appropriate for manual visualization by conventional light microscopy. RDPD system is composed of the following components:

  • · VENTANA DP 200 slide scanner,
  • · VENTANA DP 600 slide scanner,
  • · Roche uPath enterprise software, and

{6}------------------------------------------------

  • · ASUS PA248QV display.
    VENTANA DP 600 slide scanner has a total capacity of 240 slides through 40 trays with 6 slides each. The VENTANA DP 600 slide scanner and VENTANA DP 200 slide scanner use the same Image Acquisition Unit.

Both VENTANA DP 200 and DP 600 slide scanners are bright-field digital pathology scanners that accommodate loading and scanning of 6 and 240 standard glass microscope slides, respectively. The scanners each have a high-numerical aperture Plan Apochromat 20x objective and are capable of scanning at both 20x and 40x magnifications. The scanners feature automatic detection of the tissue specimen on the glass slide, automated 1D and 2D barcode reading, and selectable volume scanning (3 to 15 focus layers). The International Color Consortium (ICC) color profile is embedded in each scanned slide image for color management. The scanned slide images are generated in a proprietary file format, Biolmagene Image File (BIF), that can be uploaded to the uPath Image Management System (IMS), provided with the Roche uPath enterprise software.

Roche uPath enterprise software (uPath), a component of Roche Digital Pathology Dx system, is a web-based image management and workflow software application. uPath enterprise software can be accessed on a Windows workstation using the Google Chrome or Microsoft Edge web browser. The user interface of uPath software enables laboratories to manage their workflow from the time the whole slide image is produced and acquired by VENTANA DP 200 and/or DP 600 slide scanners through the subsequent processes, such as review of the digital image on the monitor screen and reporting of results. The uPath software incorporates specific functions for pathologists, laboratory histology staff, workflow coordinators, and laboratory administrators.

{7}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE III.

The candidate device, Roche Digital Pathology Dx, is substantially equivalent to the predicate device, Roche Digital Pathology Dx (VENTANA DP 200), which was cleared on June 14, 2024 through K232879. Below is a table that provides a comparison between the two devices:

Table 1: Comparison of Predicate Device and Candidate Device

(differences underlined)

AspectPredicate Device:Roche Digital Pathology Dx (VENTANA DP 200)Candidate Device:Roche Digital Pathology Dx (RDPD)
510(k) NumberK232879K242783
ManufacturerVentana Medical Systems, Inc.same
Product CodePSY - Whole Slide Imaging Systemsame
RegulationSection21 CFR 864.3700same
Panel(88) Pathologysame
Intended UseRoche Digital Pathology Dx (VENTANA DP 200) is an automated digital slidecreation, viewing and management system. Roche Digital Pathology Dx (VENTANADP 200) is intended for in vitro diagnostic use as an aid to the pathologist to reviewand interpret digital images of scanned pathology slides prepared from formalin-fixedparaffin-embedded (FFPE) tissue. Roche Digital Pathology Dx (VENTANA DP 200)is not intended for use with frozen section, cytology, or non-FFPE hematopathologyspecimens. Roche Digital Pathology Dx (VENTANA DP 200) is for creation andviewing of digital images of scanned glass slides that would otherwise be appropriatefor manual visualization by conventional light microscopy.Roche Digital Pathology Dx (VENTANA DP 200) is composed of VENTANA DPRoche Digital Pathology Dx is an automated digital slide creation, viewing andmanagement system. Roche Digital Pathology Dx is intended for in vitrodiagnostic use as an aid to the pathologist to review and interpret digital images ofscanned pathology slides prepared from formalin-fixed paraffin-embedded(FFPE) tissue. Roche Digital Pathology Dx is not intended for use with frozensection, cytology, or non-FFPE hematopathology specimens. Roche DigitalPathology Dx is for creation and viewing of digital images of scanned glass slidesthat would otherwise be appropriate for manual visualization by conventionallight microscopy.Roche Digital Pathology Dx is composed of VENTANA DP 200 slide scanner,
AspectPredicate Device:Roche Digital Pathology Dx (VENTANA DP 200)Candidate Device:Roche Digital Pathology Dx (RDPD)
200 slide scanner. Roche uPath enterprise software, and ASUS PA248QV display. Itis the responsibility of a qualified pathologist to employ appropriate procedures andsafeguards to assure the validity of the interpretation of images obtained using RocheDigital Pathology Dx (VENTANA DP 200).VENTANA DP 600 slide scanner, Roche uPath enterprise software, and ASUSPA248QV display. It is the responsibility of a qualified pathologist to employappropriate procedures and safeguards to assure the validity of the interpretationof images obtained using Roche Digital Pathology Dx.
Principle ofOperationAfter conducting Quality Control (QC) on the glass slides per laboratory standards(e.g., staining, coverslipping, barcode placement, etc.), the technician loads the slidesinto the VENTANA DP 200 slide scanner. The scanner scans the slides and generatesa whole slide image for each slide. The technician performs QC on scanned WSIimages by checking image data and image quality. The operator has the option toreject slide scans and initiate a re-scan. The acquired WSI images are stored in an enduser provided image storage attached to the local network. During review, thepathologist opens WSI images from the image storage in Roche uPath enterprisesoftware, performs further QC to ensure image quality, and reads the WSI images ofthe slides to make a diagnosis.After conducting Quality Control (QC) on the glass slides per laboratorystandards (e.g., staining, coverslipping, barcode placement, etc.), the technicianloads the slides into the VENTANA DP 200 or DP 600 slide scanner. The scannerscans the slides and generates a whole slide image for each slide. The technicianperforms QC on scanned WSI images by checking image data and image quality.The operator has the option to reject slide scans and initiate a re-scan. Theacquired WSI images are stored in an end user provided image storage attached tothe local network. During review, the pathologist opens WSI images from theimage storage in Roche uPath enterprise software, performs further QC to ensureimage quality, and reads the WSI images of the slides to make a diagnosis.
DeviceComponentsWSI scanner (VENTANA DP 200 slide scanner),Image Management System (Roche uPath enterprise software), andcolor monitor display (ASUS PA248QV)WSI scanners (VENTANA DP 200 slide scanner and VENTANA DP 600 slidescanner),Image Management System (Roche uPath enterprise software), andcolor monitor display (ASUS PA248QV)
Slide ScannerCapacityVENTANA DP 200 slide scanner: 1 tray with 6 slides or 1 tray with 3 double-wideslidesVENTANA DP 200: 1 tray with 6 slides or 1 tray with 3 double-wide slidesVENTANA DP 600 slide scanner: 40 trays with 6 slides each; 240 slides total
Specimen TypeSurgical pathology slides prepared from FFPE tissuesame
ScanningMagnification20x and 40xsame
Scan OutputImage FormatBIFsame

{8}------------------------------------------------

{9}------------------------------------------------

PERFORMANCE DATA IV.

1. Design Changes

A new model of the slide scanner component, VENTANA DP 600 slide scanner, is being added, while the image management software and display monitor have not changed. The VENTANA DP 200 slide scanner and VENTANA DP 600 slide scanner differ mainly in slide capacity. They share the same Image Acquisition Unit (IAU) for creating the scanned slide image. The IAU includes all hardware components, subassemblies and software modules whose form, fit and function have a direct effect on the pixel values of the digital image of the scanned slide. The hardware and software components comprising the Image Acquisition Unit implement the pixel pipeline, that is, the process of imaging the tissue specimen and creating the digital image. The IAU matches the 'Image Acquisition' subsystem defined in the FDA's device-specific guidance "Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices (April 20, 2016)" (FDA TPA guidance).

All hardware and software components not part of the IAU implement Workflow and User Support Functions (WUSF). These components and software modules do not have a direct effect on the pixel values of the digital tissue image. See below Table 2 which breaks the main scanner components into either IAU or WUSF. Hardware components and software modules comprising the WUSF may or may not be equivalent.

Similarities between the predicate and subject devices were analyzed by comparing their IAU hardware and software components. The IAU hardware components, including the superstructure, illuminator, imaging optics (objective, tube lens, and imaging sensor), focus assembly, focus imaging system, and focusing subsystem, and stage assembly are identical. The IAU software components, including the pixel shaping and image composition modules, are functionally identical. The IAU support software components, including the Focus Tracking Algorithm, Service Tool, and Calibration and Diagnostic Utility, are functionally identical.

{10}------------------------------------------------

Differences between the predicate and subject devices were analyzed by comparing their

WUSF components, including slide loading mechanism, tray storage rack, frame and cover.

FunctionAssembly
IAUSuperstructure
Optical Path
Illuminator Assembly
Focus Assembly
Focusing Imaging Subsystem
Stage Assembly
WUSFFrames and Covers
Slide Loading Mechanism
Tray Storage Rack
Workflow Management Software

Table 2: Assemblies by their Function of Image Acquisition Unit (IAU) or Workflow and User Support Functions (WUSF)

Design controls were implemented throughout the design process of the VENTANA DP 600 slide scanner to maintain the entire IAU of the predicate device scanner (VENTANA DP 200 slide scanner) to be unaltered in form, fit and function. The design controls and design changes are documented in the Quality Management System (QMS) of the Predicate Device scanner (VENTANA DP 200 slide scanner) and the VENTANA DP 600 slide scanner, specifically the Customer Requirement and Product Requirement documents, which are part of the Design History File. As part of the design controls, all requirements passed the required verification and validation process.

Since the VENTANA DP 600 slide scanner IAU is identical to that of the Predicate Device scanner by design, the Technical Performance Assessment provided in conjunction with the Predicate Device is fully applicable to the VENTANA DP 600 slide scanner as shown in Table 3.

{11}------------------------------------------------

2. Technical Studies

Multiple studies were conducted on the VENTANA DP 200 slide scanner to evaluate the performance assessment data assocrated with the technical evaluation of Roche Digital Pathology Dx per the aforementioned FDA TPA guidance. All performance characteristics data were collected on the VENTANA DP 200 slide scanner but by virtue of identical imaging acquisition units by design, its results are also representative of the VENTANA DP 600 slide scanner performance. See below Table 3 which compares the predicate device to the subject device for these aspects.

TPA SectionDP 200 slide scanner (K232879)DP 600 slide scanner (K242783)
Components
Slide FeederInformation was provided on the configuration of the slide feed mechanism, including a physical description ofthe slide, the number of slides in queue (carrier), and the class of automation. Information was provided on theuser interaction with the slide feeder, including hardware, software, feedback mechanisms, and Failure Mode andEffects Analysis (FMEA).no double wide slide traycompatibility, new FMEA providedin K242783
Light SourceDescriptive information associated with the lamp and the condenser was provided. Testing information wasprovided to verify the spectral distribution of the light source as part of the color reproduction capability ofVENTANA DP 200 scanner.Information was provided inK232879
Imaging OpticsInformation including an optical schematic with all optical elements identified from slide (object plane) to digitalimage sensor (image plane) was provided. Descriptive information regarding the microscope objective, theauxiliary lenses, and the magnification of imaging optics was provided. Testing information regarding therelative irradiance, optical distortions, and lateral chromatic aberrations was provided.Information was provided inK232879
Focusing SystemSchematic diagrams and a description of operation of the focus system, which includes the optical system, 2imaging cameras, and a focus tracking algorithm, was provided.Information was provided inK232879
Mechanical Scanner MovementInformation and specifications on the configuration of the stage, method of movement, control of movement ofthe stage, and FMEA was provided. Test data to verify the repeatability of the stage movement mechanismstaying within limit during operation was provided.same except no double wide slidetray compatibility, replacedreferences & new FMEA itemsprovided in K242783
Digital Imaging SensorInformation and specifications on the sensor type, pixel information, responsivity specifications, noisespecifications, readout rate, and digital output format was provided. Test data to determine the correctfunctioning of the digital image sensor that converts optical signals of the slide to digital signals which consist ofa set of numerical values corresponding to the brightness and color at each point in the optical image wasprovided.Information was provided inK232879
TPA SectionDP 200 slide scanner (K232879)DP 600 slide scanner (K242783)
Image Processing SoftwareInformation and specifications on exposure control, white balance, color correction, subsampling, pixel-offsetcorrection, pixel-gain or flat-field correction, and pixel-defect correction were provided.Information was provided inK232879
Image CompositionInformation and specifications on the scanning method, the scanning speed, and the number of planes at the Z-axis to be digitized were provided. Test data to analyze the image composition performance was provided.Information was provided inK232879
Image File FormatsInformation and specifications on the compression method, compression ratio, file format, and file organizationwere provided.Information was provided inK232879
Image Review ManipulationSoftwareInformation and specifications on continuous panning and pre-fetching, continuous zooming, discrete Z-axisdisplacement, the ability to compare multiple slides simultaneously on multiple windows, image enhancementand sharpening functions, color manipulation, annotation tools, and digital bookmarks were provided.Information was provided inK232879
Computer EnvironmentInformation and specifications on the computer hardware, operating system, graphics card, graphics card driver,color management settings, color profile, and display interface were provided.select upgrades of sub-components& specifications
DisplayInformation and specifications on the technological characteristics of the display such as pixel density, aspectratio, display viewing area, display surface, backlight type, panel type, viewing angle, pixel pitch, resolution,color space, max brightness, contrast ratio, response time, refresh rate (max), color accuracy, color adjustment,gamma adjustment, adjustments, display interface, and certificate were provided. Test data to verify theperformance of the display for user controls, spatial resolution, pixel defects (count and map), artifacts, temporalresponse, maximum and minimum luminance (achievable and recommended), grayscale, luminance uniformity,bidirectional reflection distribution function, gray tracking, color scale, and color gamut volume was provided.Information was provided inK232879
System-level Assessments
Color ReproducibilityTest data to evaluate the color reproducibility of the system was provided.Information was provided inK232879
Spatial ResolutionTest data to evaluate the composite optical performance of all components in the image acquisition phase wasprovided.Information was provided inK232879
Focusing TestTest data to evaluate the technical focus quality of the system was provided.Information was provided inK232879
Whole Slide Tissue CoverageTest data to demonstrate that the entire tissue specimen on the glass slide is detected by the tissue detectionalgorithms and that all the tissue specimens are included in the digital image file was provided.Information was provided inK232879
Stitching ErrorTest data to evaluate the stitching errors and artifacts in the reconstructed image was provided.Information was provided inK232879
Turnaround TimeTest data to evaluate the turnaround time of the system was provided.Information was provided inK232879
User InterfaceInformation on the parts of the system that users interact with, along with human factors/usability validationtesting performed to demonstrate that representative users of the WSI system can perform essential tasks andthose critical to safety under simulated use conditions was provided.identical workflow, replacement ofnew scanner component depiction
LabelingThe labeling is sufficient, and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable, and thespecial controls for this device type under 21 CFR 864.3700.same content, replaced references
TPA SectionDP 200 slide scanner (K232879)DP 600 slide scanner (K242783)
Quality ControlQuality control (QC) activities are performed by the user per the laboratory standards and professional guidelines (e.g., staining, cover-slipping, barcode placement) prior to loading the slides into the VENTANA DP 200. After completing a scan, the lab technician checks image data and image quality as per the instructions for use. Before review, the pathologist ensures the quality of the WSI images before reviewing the image for diagnostic purposes.same content, replaced references

Table 3: Comparison of technical characteristics

{12}------------------------------------------------

{13}------------------------------------------------

3. User Interface/Human Factors Validation

An analysis was performed to determine if additional human factors / usability evaluations were necessary to supplement the study done for the predicate device for this modification. The analysis spanned usability heuristics, comparative analysis of the workflows, risks, and related tasks for whole slide imaging. While minor differences in the User Interface (UI) exist and are notable (such as differences in some terminology, the appearance of the UI, the difference in slide capacity, and the consolidation of certain UI elements), it was determined that there are no high-level workflow differences in user input and the identified differences in the UI have no impact on users' ability to safely complete the whole slide imaging workflow. Importantly, the users' mental model of successfully completing a scan workflow does not change, regardless of the volume of trays / slides being loaded. In fact, the updated UI improved the information architecture structure, encouraging users to be more intentional about the tasks they perform during the scan workflow by explicitly confirming through the UI which tasks they are trying to accomplish. Additionally, a review and analysis of the critical tasks from the predicate device's usability evaluation & risk analysis revealed no new risks as they are virtually identical, and, upon review of DP 600 slide scanner's risk documentation and summative usability results, no significant risks related to the safe completion of the workflow were found. In summary, no additional usability study is required for this modification and the usability validation efforts completed for the predicate to the modified device.

{14}------------------------------------------------

4. Electromagnetic Compatibility (EMC) Testing

VENTANA DP 600 slider scamer EMC testing was performed by external vendor testing services using the following test standards: EN 61326-1:2013, IEC 61010-1, IEC 62368-1, and EN 61000-4. According to the test report, the emission testing was performed in accordance with EN 55011:2016 and VENTANA DP 600 slide scanner met the testing requirements classified for Group 1, Class A equipment. VENTANA DP 600 slide scanner also passed the various testing categories based on the corresponding test standards.

5. Clinical Testing

The modification to the system of increased slide scanner capacity does not require additional clinical performance data, so the clinical performance of the modified device is substantially equivalent to the original, predicate device cleared in K232879. As displayed above, the design, components, and relevant performance data of the scanners are identical with a few exceptions that do not impact the Image Acquisition Unit's produced image quality, the final output of the device, or its ability to intended use as a whole slide imaging system. Therefore it is concluded that the scanners are equivalent in these aspects.

V. CONCLUSION

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 864.3700 Whole slide imaging system.

(a)
Identification. The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
e.g., main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.