K Number
K242745

Validate with FDA (Live)

Date Cleared
2025-03-27

(197 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AI-Rad Companion Organs RT is a post-processing software intended to automatically contour DICOM CT and MR pre-defined structures using deep-learning-based algorithms.

Contours that are generated by AI-Rad Companion Organs RT may be used as input for clinical workflows including external beam radiation therapy treatment planning. AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by AI-Rad Companion Organs RT.

The outputs of AI-Rad Companion Organs RT are intended to be used by trained medical professionals.

The software is not intended to automatically detect or contour lesions.

Device Description

AI-Rad Companion Organs RT provides automatic segmentation of pre-defined structures such as Organs-at-risk (OAR) from CT or MR medical series, prior to dosimetry planning in radiation therapy. AI-Rad Companion Organs RT is not intended to be used as a standalone diagnostic device and is not a clinical decision-making software.

CT or MR series of images serve as input for AI-Rad Companion Organs RT and are acquired as part of a typical scanner acquisition. Once processed by the AI algorithms, generated contours in DICOMRTSTRUCT format are reviewed in a confirmation window, allowing clinical user to confirm or reject the contours before sending to the target system. Optionally, the user may select to directly transfer the contours to a configurable DICOM node (e.g., the Treatment Planning System (TPS), which is the standard location for the planning of radiation therapy).

AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept the automatically generated contours. Then the output of AI-Rad Companion Organs RT must be reviewed and, where necessary, edited with appropriate software before accepting generated contours as input to treatment planning steps. The output of AI-Rad Companion Organs RT is intended to be used by qualified medical professionals, who can perform a complementary manual editing of the contours or add any new contours in the TPS (or any other interactive contouring application supporting DICOM-RT objects) as part of the routine clinical workflow.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Device Performance Study for AI-Rad Companion Organs RT

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the AI-Rad Companion Organs RT device, particularly for the enhanced CT contouring algorithm, are based on comparing its performance to the predicate device and relevant literature/cleared devices. The primary metrics used are Dice coefficient and Absolute Symmetric Surface Distance (ASSD).

Table 3: Acceptance Criteria of AIRC Organs RT VA50

Validation Testing SubjectAcceptance CriteriaReported Device Performance (Summary)
Organs in Predicate DeviceAll organs segmented in the predicate device are also segmented in the subject device.Confirmed. The device continued to segment all organs previously handled by the predicate.
The average (AVG) Dice score difference between the subject and predicate device is < 3%.Confirmed. "For existing organs, the average (AVG) Dice score difference between the subject device and predicate device is smaller than 3%."
New Organs for Subject DeviceThe subject device in the selected reference metric has a higher value than the defined baseline value.Confirmed. "The performance results of the subject device for the new CT organs are comparable to the reference literature & cleared devices. Here equivalence for the new organs is defined such that the selected reference metric has a higher value than the defined baseline."

Table 3: Performance Summary of the Subject Device CT Contouring (Overall Average Dice Coefficients)

Anatomic RegionAvg Dice (%)Std Dice (%)95% CI
Head & Neck76.114.3[75.1, 77.2]
Head & Neck lymph nodes69.313.9[68.7, 70.0]
Thorax76.915.8[76.2, 77.6]
Abdomen87.310.1[86.3, 88.2]
Pelvis85.79.6[85.0, 86.5]
Cardiac75.615.1[74.1, 77.1]

Table 4: Detailed Performance Evaluation of the New Organs in the Subject Device (Selected Examples)

Organ NameNo.AVG Dice (%)STD Dice (%)MED Dice (%)95%CI DiceAVG ASSD (mm)STD ASSD (mm)MED ASSD (mm)95%CI ASSD
Left Breast3090.43.891[89, 91.8]2.42.21.8[1.5, 3.2]
Right Breast3090.23.790.8[88.8, 91.5]1.90.71.8[1.7, 2.2]
Bowel Bag33953.696.5[93.7, 96.3]1.91.51.4[1.4, 2.5]
Pituitary3075.87.477[73.1, 78.6]0.70.30.6[0.5, 0.8]
Brainstem3088.42.588.8[87.5, 89.3]10.30.9[0.9, 1.1]
Esophagus3085.64.286[84, 87.2]0.60.30.6[0.5, 0.7]
MEDIASTINAL LN 9L3138.321.142.9[30.6, 46.1]5.34.43.7[3.7, 6.9]

(Note: The full Table 4 from the document provides detailed performance for all 37 new organs. This table includes a selection for illustrative purposes.)

2. Sample Sizes and Data Provenance

  • Test Set Sample Size:
    • CT Contouring Algorithm: N = 579 cases
    • MR Contouring Algorithm: The MR algorithm is unchanged from the predicate, so its performance is unchanged. The predicate was validated using 66 cases.
  • Data Provenance (CT Contouring Algorithm Test Set):
    • Geographic Origin (Overall N=579): Data from multiple clinical sites across North American, South American, Asia, Australia, and Europe.
    • Example Cohorts (Table 5: Validation Testing Data Information based on Cohort):
      • Cohort A.1 (N=73): Germany (14), Brazil (59)
      • Cohort A.2 (N=40): Canada (40)
      • Cohort A.3 (N=301): South/North America (184), EU (44), Asia (33), Australia (28), Unknown (12)
      • Cohort B (N=165): South/North America (100), EU (51), Asia (6), Australia (3), Unknown (5)
    • Retrospective/Prospective: "retrospective performance study on CT data previously acquired for RT treatment planning."

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Number of Experts for Ground Truth: "a team of experienced annotators mentored by radiologists or radiation oncologists" for initial manual annotation. "a board-certified radiation oncologist" performed a quality assessment including review and correction of each annotation. The document does not specify an exact number of individuals for these teams, but describes the roles and qualifications.
  • Qualifications of Experts:
    • "experienced annotators"
    • "radiologists or radiation oncologists" (mentors for annotators)
    • "board-certified radiation oncologist" (for quality assessment/review)

4. Adjudication Method for the Test Set

The document describes the ground truth establishment process as: "manual annotation" by experienced annotators mentored by radiologists/radiation oncologists, followed by a "quality assessment including review and correction of each annotation was done by a board-certified radiation oncologist." This indicates a hierarchical review/correction process rather than a multi-reader consensus adjudication between equally-weighted readers (e.g., 2+1 or 3+1). The final accepted contour after the board-certified radiation oncologist's review served as the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was described. The study focused on the standalone performance of the AI algorithm against established ground truth and comparison with a predicate device and literature. The document does not mention an effect size of how much human readers improve with AI vs. without AI assistance. The intended use specifies that the AI-generated contours must be reviewed, edited, and accepted by trained medical professionals, implying a human-in-the-loop workflow, but the validation study presented focuses on the AI's autonomous segmentation accuracy.

6. Standalone (Algorithm Only) Performance Study

Yes, a standalone performance study was done. The performance metrics (Dice coefficient, ASSD) and the comparison to an expert-established ground truth demonstrate the algorithm's autonomous segmentation capability. The study validated the "autocontouring algorithms" and their performance.

7. Type of Ground Truth Used

The ground truth used for the test set was expert consensus / manual annotation based on clinical guidelines. Specifically: "Ground truth annotations were established following RTOG and clinical guidelines using manual annotation." This was further reviewed and corrected by a board-certified radiation oncologist.

8. Sample Size for the Training Set

The document provides the sample sizes for the training set for new organs introduced:

  • Table 6: Training Dataset Characteristics (Examples):
    • Lacrimal Glands Left/Right: 247
    • Pituitary Gland: 247
    • Humeral Head Left/Right: 207
    • Bowel Bag: 544
    • Pelvic Bone Left/Right: 160
    • Sacrum: 160
    • Mediastinal LN (various): 136
    • Femoral Head Left/Right: 160
    • Brainstem: 247
    • Esophagus: 247
    • Breast Left/Right: 172
    • Supraglottic Larynx: 247
    • Glottis: 247

The total training set size for all organs is not explicitly summed, but these numbers indicate the scale of the training data used for the specific new organs.

9. How the Ground Truth for the Training Set Was Established

"In both the annotation process for the training and validation testing data, the annotation protocols for the OAR were defined following the applicable guidelines. The ground truth annotations were drawn manually by a team of experienced annotators mentored by radiologists or radiation oncologists using an internal annotation tool. Additionally, a quality assessment including review and correction of each annotation was done by a board-certified radiation oncologist using validated medical image annotation tools."

This indicates the same rigorous process of expert manual annotation and review was applied to establish ground truth for the training set as for the test set. The validation testing and training data were explicitly stated to be independent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 27, 2025

Siemens Healthcare GmbH Kira Morales Regulatory Affairs Manager Henkestrasse 127 Erlangen, 91052 Germany

Re: K242745

Trade/Device Name: AI-Rad Companion Organs RT Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: OKB Dated: September 9, 2024 Received: September 11, 2024

Dear Kira Morales:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Locon Weidner

Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242745

Device Name

Al-Rad Companion Organs RT

Indications for Use (Describe)

Al-Rad Companion Organs RT is a post-processing software intended to automatically contour DICOM CT and MR pre-defined structures using deep-learning-based algorithms.

Contours that are generated by Al-Rad Companion Organs RT may be used as input for clinical workflows including external beam radiation therapy treatment planning. Al-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated bv Al-Rad Companion Organs RT.

The outputs of Al-Rad Companion Orqans RT are intended to be used by trained medical professionals.

The software is not intended to automatically detect or contour lesions.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots.

510(k) SUMMARY FOR AI-Rad Companion Organs RT K242745

Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Date Prepared: September 9, 2024

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990 and 21 CFR §807.92.

1. Submitter

Importer/DistributorSiemens Medical Solutions USA, Inc.40 Liberty BoulevardMalvern, PA 19355Registration Number: 2240869
Manufacturing SiteSiemens Healthcare GmbHHenkestrasse 127Erlangen, Germany 91052Registration Number: 3002808157

2. Contact Person

Kira Morales Senior Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19335 Phone: +1 (484) 901 - 9471 Email: kira.morales(@siemens-healthineers.com

3. Device Name and Classification

Product Name:AI-Rad Companion Organs RT
Common Name:Medical Imaging Software
Classification Name:Medical Image Management and Processing System

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots.

Classification Panel: CFR Section: Device Class: Product Code:

4. Predicate Device

Product Name: Common Name: 510(k) Number: Clearance Date: Classification Name: Classification Panel: CFR Section: Device Class: Primary Product Code: Recall Information:

5. Reference Devices

Product Name: Common Name: 510(k) Number: Clearance Date: Classification Name: Classification Panel: CFR Section: Device Class: Primary Product Code:

Product Name:

Common Name: 510(k) Number: Clearance Date: Classification Name: Classification Panel: CFR Section: Device Class: Primary Product Code:

Product Name:

Common Name: 510(k) Number: Clearance Date:

Radiology 21 CFR §892.2050 Class II OKB

AI-Rad Companion Organs RT Medical Imaging Software K232899 April 3, 2024 Medical Image Management and Processing System Radiology 21 CFR §892.2050 Class II QKB N/A

Contour Protégé AI

Medical Imaging Software K231765 November 8, 2023 Medical image management and processing system Radiology 21 CFR §892.2050 Class II QKB

Contour Protégé AI

Medical Imaging Software K223774 April 6, 2023 Medical image management and processing system Radiology 21 CFR §892.2050 Class II QKB

AI Segmentation

Medical Image Segmentation Software K211881 September 2, 2021

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" are several orange dots.

Classification Name: Classification Panel: CFR Section: Device Class: Primary Product Code:

Medical charged-particle radiation therapy system Radiology 21 CFR §892.2050 Class II MUJ

6. Indications for Use

AI-Rad Companion Organs RT is a post-processing software intended to automatically contour DICOM CT and MR pre-defined structures using deep-learning-based algorithms.

Contours that are generated by AI-Rad Companion Organs RT may be used as input for clinical workflows including external beam radiation therapy treatment planning. AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by AI-Rad Companion Organs RT.

The outputs of AI-Rad Companion Organs RT are intended to be used by trained medical professionals.

The software is not intended to automatically detect or contour lesions.

7. Device Description

AI-Rad Companion Organs RT provides automatic segmentation of pre-defined structures such as Organs-at-risk (OAR) from CT or MR medical series, prior to dosimetry planning in radiation therapy. AI-Rad Companion Organs RT is not intended to be used as a standalone diagnostic device and is not a clinical decision-making software.

CT or MR series of images serve as input for AI-Rad Companion Organs RT and are acquired as part of a typical scanner acquisition. Once processed by the AI algorithms, generated contours in DICOMRTSTRUCT format are reviewed in a confirmation window, allowing clinical user to confirm or reject the contours before sending to the target system. Optionally, the user may select to directly transfer the contours to a configurable DICOM node (e.g., the Treatment Planning System (TPS), which is the standard location for the planning of radiation therapy).

AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept the automatically generated contours. Then the output of AI-Rad Companion Organs RT must be reviewed and, where necessary, edited with appropriate software before accepting generated contours as input to treatment planning steps. The output of AI-Rad Companion Organs RT is intended to be used by qualified medical professionals, who can perform a complementary manual editing of the contours or add any new contours in the TPS (or any other interactive contouring application supporting DICOM-RT objects) as part of the routine clinical workflow.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

8. Substantially Equivalent (SE) and Technological Characteristics

The intended use of the subject device is unchanged from the predicate device. The following modifications have been made in the subject device, compared to the predicate, AI-Rad Companion Organs RT (K232899).

  • Enhanced CT contouring algorithm include 37 new organs & structures
  • Multi Guideline Support and corresponding update to UI ●

AI-Rad Companion Organs RT VA60 and AI-Rad Companion Organs RT VA50 both use a deep learning algorithm to support their AI claims. Additionally, they both process CT and MR data in DICOM format, making them vendor agnostic and create outputs which can be used by any TPS system. The deep learning CT algorithm within AI-Rad Companion Organs RT VA60 has been enhanced from the algorithm in AI-Rad Companion Organs RT VA50 (K232899). All models contained within AI-Rad Companion Organs RT VA60 and AI-Rad Companion Organs RT VA50 (K232899) are locked and cannot be modified by the user.

The subject device, AI-Rad Companion Organs RT, is substantially equivalent with regards to the software features, functionalities, and core algorithms. The performance of the enhanced CT contouring algorithm has been validated against FDA/CE cleared devices or from literature.

Subject DevicePredicate Device
DeviceManufacturerSiemensSiemens
Device NameAI-Rad Companion Organs RTAI-Rad Companion Organs RT
510(k) NumberK242745K232899
Indications forUseAI-Rad Companion Organs RT isa post-processing softwareintended to automatically contourDICOM CT and MR pre-definedstructures using deep-learning-based algorithms.Contours that are generated by AI-Rad Companion Organs RT maybe used as input for clinicalworkflows including externalbeam radiation therapy treatmentAI-Rad Companion Organs RT isa post-processing softwareintended to automatically contourDICOM CT and MR pre-definedstructures using deep-learning-based algorithms.Contours that are generated by AI-Rad Companion Organs RT maybe used as input for clinicalworkflows including externalbeam radiation therapy treatment
planning. AI-Rad Companionplanning. AI-Rad Companion
Organs RT must be used inOrgans RT must be used in
conjunction with appropriateconjunction with appropriate
software such as Treatmentsoftware such as Treatment
Planning Systems and InteractivePlanning Systems and Interactive
Contouring applications, toContouring applications, to
review, edit, and accept contoursreview, edit, and accept contours
generated by AI-Rad Companiongenerated by AI-Rad Companion
Organs RT.Organs RT.
The outputs of AI-Rad CompanionThe outputs of AI-Rad Companion
Organs RT are intended to be usedOrgans RT are intended to be used
by trained medical professionals.by trained medical professionals.
The software is not intended toThe software is not intended to
automatically detect or contourautomatically detect or contour
lesionslesions
AlgorithmDeep LearningDeep Learning
Segmentation ofCT: Head & Neck, Thorax,CT: Head & Neck, Thorax,
Organ at Risk in theAbdomen & PelvisAbdomen & Pelvis
Anatomic RegionsHead & Neck lymph nodesHead & Neck lymph nodes
(203 OAR)(166 OAR)
MR: Pelvis (9 OAR)MR: Pelvis (9 OAR)
CompatibleModalityCT & MR ImagesCT & MR Images
CompatibleScanner ModelsNo Limitation on scanner modelfor CT. Siemens Healthineers'data only for MR. DICOMcompliance required.No Limitation on scanner modelfor CT. Siemens Healthineers'data only for MR. DICOMcompliance required.
CompatibleTreatment PlanningSystemNo Limitation on TPS model,DICOM compliance required.No Limitation on TPS model,DICOM compliance required.
ContraindicationsAdult use onlyAdult use only
Target PopulationAI-Rad Companion Organs RT isdesigned for use only in adultpopulations.AI-Rad Companion Organs RT isdesigned for any patient for whomrelevant modality scans areavailable.AI-Rad Companion Organs RT isdesigned for use only in adultpopulations.AI-Rad Companion Organs RT isdesigned for any patient for whomrelevant modality scans areavailable.
Clinical conditionthe device isintended todiagnose, treat ormanageLimited to patients previouslyselected for Radiation Therapy.Limited to patients previouslyselected for Radiation Therapy.
SoftwareArchitectureAI-Rad Companion (Engine)architecture enabling thedeployment of AI Rad CompanionOrgans RT using Edge and in theCloud. The UI is provided using aweb-based interface.AI-Rad Companion (Engine)architecture enabling thedeployment of AI Rad CompanionOrgans RT using Edge and in theCloud. The UI is provided using aweb-based interface.
DeploymentFeatureEdge & Cloud DeploymentEdge & Cloud Deployment
Organ TemplatesCreating, editing and deletion oforgan templates. Customizepredefined structure database withmapping to internationalnomenclature schemes.Creating, editing and deletion oforgan templates. Customizepredefined structure database withmapping to internationalnomenclature schemes.
AutomatedworkflowAI-Rad Companion Organs RTautomatically processes inputimage data and sends the results asDICOM-RT Structure Sets to auser-configurable target node.AI-Rad Companion Organs RTautomatically processes inputimage data and sends the results asDICOM-RT Structure Sets to auser-configurable target node.
Contourvisualization andediting featureAI-Rad Companion Organs RTprovides basic result preview ofautomatic segmentation results,and no editing feature of theautomatic segmented contour.AI-Rad Companion Organs RTprovides basic result preview ofautomatic segmentation results,and no editing feature of theautomatic segmented contour.
SegmentationPerformanceMR: The algorithm is unchangedfrom the predicate and is separatefrom the CT algorithm thereforethe performance is unchangedfrom the predicate.CT: The target performance wasMR: The target performance wasvalidated using 66 cases tovalidate the overall performance ofthe MR contouring.CT: The target performance wasvalidated using 414 cases
validated using 579 casesdistributed to four cohorts.distributed to three cohorts.
Both: To objectively evaluate thetarget performance, the DICEcoefficient, the absolute symmetricsurface distance (ASSD) and thefail rate was evaluated. TheBoth: To objectively evaluate thetarget performance, the DICEcoefficient, the absolute symmetricsurface distance (ASSD) and thefail rate was evaluated. Thesegmentation performance of the
segmentation performance of thesubject was equivalent to theoverall performance compared tothe predicate, reference device andcomparable literature & devices.subject was equivalent to theoverall performance compared tothe predicate, reference device andcomparable literature & devices.
User Interface -Results Preview(Confirmation)Basic visualization functionality oforiginal data and generatedcontoursBasic visualization functionality oforiginal data and generatedcontours
User InterfaceConfigurationConfiguration UIConfiguration UI
AutomatedWorkflow to TPSResults send to Confirmation UI &Optional bypassing ofConfirmation UI to TPSResults send to Confirmation UI &Optional bypassing ofConfirmation UI to TPS
Human FactorsDesign to be used by trainedclinicians.Design to be used by trainedclinicians.

The risk analysis and non-clinical data support that both devices perform equivalently and do not raise different questions of the safety and effectiveness.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a cluster of orange dots arranged in a circular pattern.

Table 1: Indications for Use and Segmentation Feature Comparison

The conclusions from all verification and validation data suggests that these enhancements are equivalent with respect to safety and effectiveness of the predicate device. These modifications do not change the intended use of the product. Siemens is of opinion that AI-Rad Companion Organs RT VA60 is substantially equivalent to the currently marketed device, AI-Rad Companion Organs RT (K232899).

9. Nonclinical Tests

Non-clinical tests were conducted to test the functionality of AI-Rad Companion Organs RT. Software validation and bench testing have been conducted to assess the performance claims as well as the claim of substantial equivalence to the predicate device.

AI-Rad Companion has been tested to meet the requirements of conformity to multiple industry standards. Non-clinical performance testing demonstrates that AI-Rad Companion Organs RT complies with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Device Software Functions" (June 2023) as well as with the following voluntary FDA recognized Consensus Standards listed in Table 2.

RecognitionNumberProductAreaTitle of StandardReferenceNumber andDateStandardsDevelopmentOrganization
5-129GeneralMedical Devices –Application of usabilityengineering to medicaldevices62366-1 Ed1.1 2020-06CVIEC
5-125GeneralMedical Devices –application of risk14971:2019-12ISO

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.

management to medicaldevices
13-79Software/InformaticsMedical device software –software life cycleprocesses [IncludingAmendment 1 (2016)]62304 Ed 1.12015-06 CVAAMIANSIIEC
12-352RadiologyDigital Imaging andCommunications inMedicine (DICOM) SetPS 3.1 – 3.202023eNEMA
5-134GeneralMedical devices – symbolsto be used with informationto be supplied by themanufacturer – Part 1:General Requirements15223-1Fourth edition2021-07ISOIEC
13-97Software/InformaticsHealth software – Part 1:General requirements forproduct safety82304-1Edition 1.02016-10IEC
13-122Software/InformaticsHealth software and HealthIT system safetyeffectiveness and security81001-5-1Edition 1.02021-12IEC
5-135GeneralMedical devices –Information to be suppliedby the manufacturer20417 Firstedition 2021-04ISO

Table 2: List of recognized standards

Verification and Validation

Software documentation level, per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on June 14, 2023, is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the subject device during product development.

Software bench testing in the form of Unit, System and Integration tests were performed to evaluate the performance and functionality of the new features and software updates. All testable requirements in the Requirement Specifications and the Risk Analysis have been successfully verified and traced in accordance with the Siemens Healthineers DH product development process. Human factor usability validation is addressed in system testing and usability validation test records. Software verification and regression testing have been performed successfully to meet their previously determined acceptance criteria as stated in the test plans.

Siemens Healthineers adheres to the cybersecurity recommendations as defined the FDA Guidance "Cybersecurity in Medical Devices: Ouality System Considerations and Content of Premarket Submissions" (September 2023) by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.

10. Performance Software Validation

To validate the AI-Rad Companion Organs RT software from clinical perspective, the autocontouring algorithms underwent a scientific evaluation. The results of clinical data-based software validation for the subject device AI-Rad Companion Organs RT (SW VA60A) demonstrated equivalent performance in comparison to the predicate device (SW VA50A. K232899). The performance of the enhanced CT organ contouring algorithm is comparable to the predicate device and comparable reference literature and cleared devices. A complete scientific evaluation report is provided in support of the device modifications.

The MR Contouring algorithm contained within AI-Rad Companion Organs RT VA60 is identical to the predicate device AI-Rad Companion Organs RT VA50 (K232899).

CT Contouring Algorithm Performance

The performance of the AI-Rad Companion Organs RT has been validated in a retrospective performance study on CT data previously acquired for RT treatment planning (N= 579, data from multiple clinical sites across the North American, South American, Asia, Australia and Europe). Ground truth annotations were established following RTOG and clinical guidelines using manual annotation. The mean and standard deviation Dice coefficients, along with the lower 95th percentile confidence bound, were calculated for each organ in the subject device. The results of subject device were equivalent or had better performance than the predicate device. To encountered for different datasets, variation, we first calculate the average of multiple references or the average of anatomical region for the specific organ or anatomical region. We then define the baseline value by subtracting the reference value using 5% error margin in case of Dice and 0.1 mm in case of ASSD.

The performance results of the subject device for the new CT organs are comparable to the reference literature & cleared devices. Here equivalence for the new organs is defined such that the selected reference metric has a higher value than the defined baseline. For existing organs, the average (AVG) Dice score difference between the subject device and predicate device is smaller than 3%.

Validation Testing SubjectAcceptance Criteria
Organs in Predicate DeviceAll the organs segmented in the predicate device are also segmented in the subject device
The average (AVG) Dice score difference between the subject and predicate device is smaller than 3%

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.

New Organs for Subject Device• The subject device in the selected referencemetric has a higher value than the definedbaseline value
-----------------------------------------------------------------------------------------------------------------------------------------------

Table 3: Acceptance Criteria of AIRC Organs RT VA50

Dice (%)
AvgStd95% CI
Head & Neck76.114.3[75.1, 77.2]
Head & Neck lymphnodes69.313.9[68.7, 70.0]
Thorax76.915.8[76.2, 77.6]
Abdomen87.310.1[86.3, 88.2]
Pelvis85.79.6[85.0, 86.5]
Cardiac75.615.1[74.1, 77.1]

Table 3: Performance summary of the subject device CT contouring

Organ NameNO.Dice (%)ASSD (mm)
AVGSTDMED95%CIAVGSTDMED95%CI
Left Breast3090.43.891[89, 91.8]2.42.21.8[1.5, 3.2]
Right Breast3090.23.790.8[88.8, 91.5]1.90.71.8[1.7, 2.2]
Bowel Bag33953.696.5[93.7, 96.3]1.91.51.4[1.4, 2.5]
Left Pelvic Bone3093.41.693.8[92.8, 94]0.60.20.5[0.5, 0.6]
Right Pelvic Bone3093.91.394[93.4, 94.4]0.50.10.5[0.4, 0.6]
Pituitary3075.87.477[73.1, 78.6]0.70.30.6[0.5, 0.8]
Sacrum3090.7491.9[89.2, 92.2]0.80.60.6[0.5, 1]
Brainstem3088.42.588.8[87.5, 89.3]10.30.9[0.9, 1.1]
Glottis30688.770.9[64.8, 71.3]1.20.41.1[1.1, 1.4]
Supraglottic Larynx30805.981[77.8, 82.20]0.80.30.7[0.7, 0.9]
Esophagus3085.64.286[84, 87.2]0.60.30.6[0.5, 0.7]
Left Lacrimal Gland3072.17.371.3[69.4, 74.9]0.80.30.8[0.7, 0.9]
Right Lacrimal Gland3069.912.473.6[65.3, 74.6]0.90.70.8[0.7, 1.2]
Left Femur Head3095.21.195.2[94.8, 95.6]0.50.20.5[0.5, 0.6]

{14}------------------------------------------------

Hea

Right Femur Head3094.91.395.1[94.5, 95.4]0.60.20.5[0.5, 0.6]
Left Humeral Head3094.22.194.8[93.4, 94.9]0.70.30.6[0.6, 0.8]
Right Humeral Head3094.63.295.4[93.4, 95.8]0.70.50.5[0.5, 0.9]
MEDIASTINAL LN 10L3152.99.755.1[49.4, 56.5]1.20.61.1[1, 1.5]
MEDIASTINAL LN 10R3150.29.550.6[46.7, 53.7]1.40.61.2[1.1, 1.6]
MEDIASTINAL LN 1L3170.71073[67, 74.4]2.41.21.9[1.9, 2.8]
MEDIASTINAL LN 1R316612.867.9[61.3, 70.7]2.40.92.3[2.1, 2.8]
MEDIASTINAL LN 2L3167.67.668.6[64.8, 70.5]1.40.41.3[1.2, 1.5]
MEDIASTINAL LN 2R3155.513.656.9[50.6, 60.5]2.31.52[1.7, 2.8]
MEDIASTINAL LN 3A3175.66.377.6[73.3, 77.9]1.60.41.5[1.5, 1.8]
MEDIASTINAL LN 3P3165.9666.5[63.7, 68.1]1.60.61.5[1.4, 1.8]
MEDIASTINAL LN 4L3165.78.467.2[62.6, 68.8]1.40.71.1[1.1, 1.7]
MEDIASTINAL LN 4R3172.8874.6[69.9, 75.7]1.70.71.6[1.4, 1.9]
MEDIASTINAL LN 5316114.865.7[55.5, 66.4]1.80.81.6[1.5, 2.1]
MEDIASTINAL LN 63158.814.862.5[53.4, 64.2]21.11.6[1.6, 2.4]
MEDIASTINAL LN 731597.661.3[56.2, 61.8]1.90.81.7[1.6, 2.2]
MEDIASTINAL LN 83165.8867.9[62.9, 68.8]20.91.7[1.7, 2.4]
MEDIASTINAL LN 9L3138.321.142.9[30.6, 46.1]5.34.43.7[3.7, 6.9]
MEDIASTINAL LN 9R2938.815.638.5[32.9, 44.7]3.723.5[2.9, 4.4]

Table 4: Detailed Performance evaluation of the new organs in the subject device

Test Cohort
# Patients244
(Data origin)South/North America: 163
EU: 70
Asia: 6
Australia: 3
Unknown: 2
GenderF: 111
M: 100
Unknown: 33
ManufacturerGE: 29
Philips: 33
Siemens: 158
Other/Unknown: 24
Slice Thickness<= 1: 10

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.

(1, 2]: 124
(2, 3]: 106
> 3: 4

Table 5: Validation Testing Data Information for new OARs

CohortAB
A.1A.2A.3
# Patients7340301165
# Of clinicalsites(Data origin)3412+20+
Germany: 14Brazil: 59Canada: 40South/North America: 184EU: 44Asia: 33Australia: 28Unknown: 12South/North America: 100EU: 51Asia: 6Australia: 3Unknown: 5
Body RegionHead & Neck: 24Thorax & Abdomen: 20Pelvis: 29Head & Neck: 40Head & Neck: 50Thorax: 81Abdomen: 115Pelvis: 55Head & Neck: 40Thorax: 69Abdomen: 25Pelvis: 31

Table 5: Validation Testing Data Information based on Cohort

Organ GroupNo. of TrainingNo. of Validation
Lacrimal Glands Left24762
Lacrimal Glands Right
Pituitary Gland24762
Humeral Head Left20752
Humeral Head Right
Bowel Bag54425
Pelvic Bone Left16040
Pelvic Bone Right
Sacrum16040
Mediastinal LN I Left
Mediastinal LN 1 Right13634
Mediastinal LN II Left

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the words is a graphic of orange dots.

Mediastinal LN II Right
Mediastinal LN III Anterior
Mediastinal LN III Posterior
Mediastinal LN IV Left
Mediastinal LN IV Right
Mediastinal LN V
Mediastinal LN VI
Mediastinal LN VII
Mediastinal LN VIII
Mediastinal LN IX Left
Mediastinal LN IX Right
Mediastinal LN X Left
Mediastinal LN X Right
Femoral Head Left16040
Femoral Head Right
Brainstem24762
Esophagus24762
Breast Left17244
Breast Right
Supraglottic Larynx24762
Glottis

Table 6: Training Dataset Characteristics

Standard Annotation Process

In both the annotation process for the training and validation testing data, the annotation protocols for the OAR were defined following the applicable guidelines. The ground truth annotations were drawn manually by a team of experienced annotators mentored by radiologists or radiation oncologists using an internal annotation tool. Additionally, a quality assessment including review and correction of each annotation was done by a board-certified radiation oncologist using validated medical image annotation tools.

Validation Testing & Training Data Independence

The training data used for the training of the algorithm is independent of the data used to test the algorithm.

11. Clinical Tests

No clinical tests were conducted to test the performance and functionality of the modifications introduced within AI-Rad Companion Organs RT. Verification and validation of the enhancements and improvements have been performed and these modifications have been validated for their intended use. The data from these activities were used to support the subject device and the substantial equivalence argument. No animal testing has been performed on the subject device.

12. Safety and Effectiveness

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.

The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device as compared to the predicate.

Risk management is ensured via ISO 14971:2019 compliance to identify and provide mitigation of potential hazards in a risk analysis early in the design phase and continuously throughout the development of the product. These risks are controlled via measures realized during software development, testing and product labeling.

13. Conclusion

Based on the discussion and validation testing and performance data above, the proposed device is determined to be as safe and effective as its predicate device, AI-Rad Companion Organs RT VA50 (K232899).

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).