(1 days)
BracePaste Fluoride Sealant is a light-cure primer / sealant intended to be used in the bonding procedure (direct or indirect) of orthodontic bondable devices by preparing the bonding surface during orthodontic treatment to the etched enamel.
BracePaste Fluoride Sealant is indicated for the orthodontic treatment of malocclusions and craniofacial abnormalities as diagnosed by a trained dental professional or orthodontist.
American Orthodontics' BracePaste Fluoride Sealant is a light-curing primer that is intended for use in orthodontic treatment as diagnosed by a trained dental professional or orthodontist. The primer prepares the enamel during orthodontic treatment to increase adhesion to the bonding surface by wetting the bonding surface of the object to be bonded.
The provided document describes the FDA 510(k) clearance for the BracePaste Fluoride Sealant. It is a submission for a dental device, specifically a light-cure primer/sealant for orthodontic bonding. As such, it does not involve AI/ML technology or image processing. Therefore, the questions related to AI/ML device performance, such as sample size for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, and training set details, are not applicable to this submission.
The acceptance criteria and study information provided in the document are primarily for demonstrating substantial equivalence to a predicate device (Opal Seal) through non-clinical performance and biocompatibility testing.
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria with numerical targets for each performance test. Instead, it states that "Results of all conducted testing was found acceptable and does not raise any new issues of safety or effectiveness." and "Results of bench testing indicate that BracePaste Fluoride Sealant performs as well as the predicate Opal Seal." This implies that the acceptance criteria are met if the new device's performance is comparable to or performs as well as the predicate device or meets established standards (e.g., ISO 29022:2013 for Shear Bond Strength).
However, based on the non-clinical performance tests listed, here's a conceptual table:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Shear Bond Strength (ISO 29022:2013): Must be comparable to or perform as well as the predicate device. | Performed the test. Results were found acceptable and indicate performance as well as the predicate Opal Seal. (Specific values not provided) |
| Stability: Maintain performance characteristics over shelf life. | Stability Study conducted. Results found acceptable. 3-year shelf life under recommended storage conditions (3-25 °C). |
| Interaction with Accessories: Compatible with intended accessories. | Test conducted. Results found acceptable. |
| Fluoride Measurements: Demonstrate fluoride release/presence. | Test conducted. Results found acceptable. (Presence of Fluoride: Yes) |
| Depth of Cure: Achieve adequate curing depth. | Test conducted. Results found acceptable. |
| Evaluation of Reapplication: Performance after reapplication. | Test conducted. Results found acceptable. |
| Water Sorption and Water Solubility: Meet dental material standards. | Test conducted. Results found acceptable. |
| Evaluation of Air Dry, Tack Cure: Demonstrate proper handling characteristics. | Test conducted. Results found acceptable. |
| Biocompatibility (ISO 7405, ISO 10993-1): No adverse biological reactions. | Biocompatible per ISO 7405, ISO 10993-1. Specific tests conducted include Cytotoxicity (Direct Contact & Elution), Acute Systemic Toxicity, Genotoxicity (Bacterial Reverse Mutation & Mouse Lymphoma Assay), Irritation (Intracutaneous Reactivity), Sensitization (Delayed Hypersensitization), and Subacute/Subchronic Systemic Toxicity. All results found acceptable. |
2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not specify the exact sample sizes for each non-clinical test conducted (e.g., how many samples were tested for Shear Bond Strength). It only lists the types of tests.
Given the nature of the device and studies (non-clinical bench testing), data provenance typically refers to the testing being performed in an accredited lab, often in the country of the manufacturer. However, this specific detail (country of origin for the data) is not provided in the document. All testing described is prospective, as it was conducted specifically for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable. This is a non-clinical bench testing submission for a dental material, not an AI/ML device requiring expert ground truth for interpretation.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. This is a non-clinical bench testing submission for a dental material.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a non-clinical bench testing submission for a dental material, not an AI-assisted diagnostic or treatment planning device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a non-clinical bench testing submission for a dental material, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for non-clinical performance testing typically refers to:
- Established physical and chemical standards: For tests like depth of cure, water sorption, water solubility, and fluoride measurements, the ground truth is often defined by industry standards or specifications for dental materials.
- Predicate device performance: For comparative tests like shear bond strength, the "ground truth" or benchmark is the performance of the legally marketed predicate device (Opal Seal).
- Biocompatibility standards: For biocompatibility, the ground truth is defined by international standards like ISO 10993 series and ISO 7405, which specify acceptable limits for various biological responses.
8. The sample size for the training set
Not applicable. This is a non-clinical bench testing submission for a dental material; there is no "training set" as it does not involve AI/ML.
9. How the ground truth for the training set was established
Not applicable. There is no training set for this type of device submission.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 27, 2024
American Orthodontics Corp. % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K242537
Trade/Device Name: BracePaste Fluoride Sealant Regulation Number: 21 CFR 872.3750 Regulation Name: Bracket adhesive resin and tooth conditioner Regulatory Class: Class II Product Code: DYH Dated: August 24, 2024 Received: August 26, 2024
Dear Dave Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
Submission Number (if known)
Device Name
BracePaste Fluoride Sealant
Indications for Use (Describe)
BracePaste Fluoride Sealant is a light-cure primer / sealant intended to be used in the bonding procedure (direct or indirect) of orthodontic bondable devices by preparing the bonding surface during orthodontic treatment to the etched enamel.
BracePaste Fluoride Sealant is indicated for the orthodontic treatment of malocclusions and craniofacial abnormalities as diagnosed by a trained dental professional or orthodontist.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized, geometric font, with the word "AMERICAN" above the word "ORTHODONTICS" to the right of the letters. The logo is in a blue color and the background is white.
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
K242537
510(k) Summary 21CFR807.92
Preparation Date: August 5th, 2024
Company Information:
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081 Phone: 920-457-5051 Fax: 920-457-5773
Submitter Information:
Luke Hetue | Supplier Quality Engineer
Device Information:
Trade Name: BracePaste Fluoride Sealant Common Name: Orthodontic Primer Classification Name: Bracket Adhesive Resin and Tooth Conditioner 510(k) Number: unknown Product Code: DYH Regulation Number (21CFR): 872.3750
Predicate Device Information:
Product/Trade Name: Opal Seal Classification Name: Bracket Adhesive Resin and Tooth Conditioner 510(k) Number: K090355 Product Code: DYH Regulation Number (21CFR): 872.3750
Description of the Device:
American Orthodontics' BracePaste Fluoride Sealant is a light-curing primer that is intended for use in orthodontic treatment as diagnosed by a trained dental professional or orthodontist. The primer prepares the enamel during orthodontic treatment to increase adhesion to the bonding surface by wetting the bonding surface of the object to be bonded.
Indications for Use:
BracePaste Fluoride Sealant is a light-cure primer / sealant intended to be used in the bonding procedure (direct or indirect) of orthodontic bondable devices by preparing the bonding surface during orthodontic treatment to the etched enamel.
BracePaste Fluoride Sealant is indicated for the orthodontic treatment of malocclusions and craniofacial abnormalities as diagnosed by a trained dental professional or orthodontist.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, with the word "AMERICAN" above the word "ORTHODONTICS" to the right of the letters. The logo is in a blue color.
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
Substantial Equivalence Discussion:
BracePaste Fluoride Sealant has the following similarities to the legally marketed predicate Opal Seal (K090355):
- Same intended use, and .
- Same technological characteristics through delivery system, chemical characteristics, curing . method, and incorporation of similar materials.
The table below outlines the comparison of the predicate device (Opal Seal), and American Orthodontics' device (BracePaste Fluoride Sealant) to show Substantial Equivalence.
| Device Name / Manufacturer | ||
|---|---|---|
| Element | Device: Opal SealManufacturer: Ultradent Products, Inc. | Device: BracePaste Fluoride SealantManufacturer: American Orthodontics |
| 510(k) Number | K090355 | unknown |
| ClassificationCode / RegulationNumber | DYH872.3750 | DYH872.3750 |
| Indications forUse / IntendedUse | Opal Seal is a light cure primer that is used whenbonding orthodontic appliances to etched enamel. | BracePaste Fluoride Sealant is a light-cure primer /sealant intended to be used in the bondingprocedure (direct or indirect) of orthodonticbondable devices by preparing the bonding surfaceduring orthodontic treatment to the etched enamel.BracePaste Fluoride Sealant is indicated for theorthodontic treatment of malocclusions andcraniofacial abnormalities as diagnosed by a traineddental professional or orthodontist. |
| Type andDuration ofTissue Contact | Long term use (used continuously for more thanthirty days)Location - oral cavity, direct contact with toothenamelDuration - greater than thirty days | Long term use (used continuously for more thanthirty days)Location - oral cavity, direct contact with toothenamelDuration - greater than thirty days |
| Delivery System | 1.7g syringe | 6mL Dropper Bottle |
| Material Type | Methacrylate based resin | Methacrylate based resin |
| Color | Light Yellow | Light Yellow |
| Physical Form | Liquid | Liquid |
| StorageConditions | Under recommended storage conditions of 2 – 8 °C(36 - 46 °F), product has a shelf life of 18 months. | Under recommended storage conditions of 3 - 25 °C(38 - 77 °F), product has a shelf life of 3 years. |
| Curing Method | Light Activated | Light Activated |
| Presence ofFluoride | Yes | Yes |
| Sterile | Shipped Unsterile | Shipped Unsterile |
| Limits of Reuse | Tips are single use only | Do not re-use used/dispensed product |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, with the word "AMERICAN" above the word "ORTHODONTICS" to the right of the letters. The logo is in blue.
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
| Biocompatibility | Biocompatible per ISO 7405, ISO 10993-1Refer to FDA 510(k) submission K090355. | Biocompatible per ISO 7405, ISO 10993-1Refer to Section 15 – Biocompatibility |
|---|---|---|
| ------------------ | ------------------------------------------------------------------------------------ | ----------------------------------------------------------------------------------- |
Performance Testing:
Clinical Performance Testing
No clinical performance testing has been conducted.
Non-Clinical Performance Testing
The following non-clinical performance tests were conducted:
-
- Shear Bond Strength Test (ISO 29022:2013) predicate device
-
- Stability Study
-
- Interaction with Accessories
-
- Fluoride Measurements
-
- Depth of Cure
- Evaluation of Reapplication 6.
-
- Water Sorption and Water Solubility
-
- Evaluation of Air Dry, Tack Cure
Non-Clinical Biocompatibility Testing
-
- Cytotoxicity Direct Contact
-
- Cytotoxicity Elution
-
- Acute Systemic Toxicity Test
-
- Genotoxicity Bacterial Reverse Mutation Test
- Genotoxicity Mouse Lymphoma Assay 5.
-
- Irritation Intracutaneous Reactivity
-
- Sensitization Delayed Hypersensitization Test
-
- Subacute/Subchronic Systemic Toxicity Test
The combination of in-house testing and side-by-side comparison performed by the original manufacturer has demonstrated the efficacy or suitability to the intended purpose of the BracePaste Fluoride Sealant. Results of all conducted testing was found acceptable and does not raise any new issues of safety or effectiveness.
Conclusion:
The combination of in-house testing and side-by-side comparison performed by the original manufacturer has demonstrated the efficacy or suitability to the intended purpose of the BracePaste Fluoride Sealant. Results of bench testing indicate that BracePaste Fluoride Sealant performs as well as the predicate Opal Seal.
Both devices have the same intended use – to prepare the enamel and increase adhesion to the bonding surface during orthodontic treatment.
Any slight differences do not affect the original function or intended purpose of the device.
Information contained in this 510(k) does not raise new questions or safety and effectiveness and, demonstrates BracePaste Fluoride Sealant is at least as safe and effective as the predicate.
§ 872.3750 Bracket adhesive resin and tooth conditioner.
(a)
Identification. A bracket adhesive resin and tooth conditioner is a device composed of an adhesive compound, such as polymethylmethacrylate, intended to cement an orthodontic bracket to a tooth surface.(b)
Classification. Class II.