(88 days)
ACUSON Sequoia and Sequoia Select
The ACUSON Sequoia and Sequoia Select ultrasound imaging systems are intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Neonatal Cephalic, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Cardiac, Transesophageal, Pelvic, Vascular, Adult Cephalic, Musculoskeletal and Peripheral Vascular applications.
The system supports the Ultrasonically-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult and pediatric patients with hepatic steatosis.
The system also provides the ability to measure anatomical structures for fetal, abdominal, pediatric, small organ, cardiac, transvaginal, peripheral vessel, musculoskeletal and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
ACUSON Origin and Origin ICE
The ACUSON Origin and Origin ICE ultrasound imaging systems are intended to provide images of. or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric. OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Cardiac, Transesophageal, Intracardiac, Vascular, Adult Cephalic, and Peripheral Vascular applications.
The catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
The system also provides the ability to measure anatomical structures for fetal, abdominal, pediatric, cardiac, peripheral vessel, and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The ACUSON Sequoia, Sequoia Select, Origin and Origin ICE Diagnostic Ultrasound Systems are multi-purpose, mobile, software-controlled, diagnostic ultrasound systems with an on-screen display of thermal and mechanical indices related to potential bioeffect mechanisms. The ultrasound system function is to transmit and receive ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes, Panoramic Imaging, Contrast agent Imaging, Virtual Touch Strain Imaging (except Origin), Virtual Touch - pSWE Imaging, Virtual Touch - SWE Imaging, Custom Tissue Imaging, 3D/4D Volume lmaging or Harmonic Imaging on a Display and provide cardiac anatomical and quantitative function software applications.
This document describes the acceptance criteria and study that proves the device called ACUSON Sequoia Diagnostic Ultrasound System from Siemens Medical Solutions USA, Inc. meets the acceptance criteria.
1. Table of Acceptance Criteria and Reported Device Performance
The submission details two AI features: "AI Abdomen" and "Trace AI."
AI Abdomen
| Feature/Metric | Acceptance Criterion | Reported Device Performance |
|---|---|---|
| View Classification | Not explicitly stated as a numerical criterion, but the overall expectation is a high accuracy and usefulness. The summary indicates a focus on specific view types. | Achieved a success rate of 77.8% for view classification when aggregated across all 17 view types. When eliminating CBD and pancreas sagittal from the aggregated results, the accuracy increased to 82.5%. |
| Semi-automated Measurements | Individual Equivalence Coefficient (IEC) below a prespecified success criterion of 0.25. (This indicates that the algorithm's measurements are very close to the ground truth, likely meaning a low difference/high agreement). | The IEC fell below the prespecified success criterion of 0.25 for all 12 measurements, with and without users editing the landmark locations. |
Trace AI
| Feature/Metric | Acceptance Criterion | Reported Device Performance |
|---|---|---|
| DICE Coefficient for Orifice-type Structures | The 90th percentile shall have a minimum DICE coefficient of at least 80% with 90% confidence. (DICE coefficient is a measure of similarity between two images or segmentations, with 1 being perfect overlap and 0 being no overlap). | The lower 90% confidence bound for the 90th percentile is 0.95, which is greater than the requirement of 0.8, therefore, the test passes. |
2. Sample Size Used for the Test Set and Data Provenance
AI Abdomen:
- Sample Size: 105 exams from individual patients, resulting in 1,785 images extracted (15 B-Mode images per view and per transducer).
- Data Provenance: Retrospective, collected from 3 institutions in the United States.
Trace AI:
- Sample Size: 10 different adult patients, from which 24 volumes were extracted.
- Data Provenance: Retrospective, collected from three institutions in the US, Mexico, and Germany.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
AI Abdomen:
- Number of Experts: Three clinical users.
- Qualifications: All registered Clinical Sonographers with ARDMS accreditation (or equivalent), each with at least 3 years of clinical experience.
Trace AI:
- Number of Experts: Three clinical users.
- Qualifications: All registered Clinical Sonographers with ARDMS accreditation (or equivalent), each with at least 3 years of clinical experience.
4. Adjudication Method for the Test Set
AI Abdomen:
- For view classification, the ground truth was associated with each image at the time of imaging using a system protocol, implying a single determination at the point of acquisition.
- For semi-automated measurements, ground truth measurements were provided by three clinical users. The document does not explicitly describe an adjudication method (like 2+1 or 3+1 consensus) for these measurements, but the acceptance criterion of Individual Equivalence Coefficient (IEC) suggests a comparison of the algorithm's results against these expert measurements.
Trace AI:
- Ground truth measurements were provided by three clinical users. Each sonographer independently refined the initial circle provided by Trace AI to annotate the underlying orifice structure. They could only see their own annotations. The contour created by sonographers (Ground Truth) was then compared with the contour detected by Trace AI (Detection Truth). This implies a comparison against each expert's delineation, potentially implying an aggregation or statistical analysis of agreement rather than a formal consensus adjudication before comparison with the AI.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was explicitly mentioned in the provided text, and therefore, no effect size of human readers' improvement with AI vs. without AI assistance can be determined from this information. The studies described are performance evaluations of the AI algorithms themselves against ground truth.
6. Standalone (Algorithm Only) Performance Study
- Yes, the studies described for both "AI Abdomen" and "Trace AI" are standalone performance evaluations of the algorithms. They assess the algorithms' accuracy in view classification, semi-automated measurements, and segmentation (DICE coefficient) against established ground truth without a human-in-the-loop component for the performance measurement. The "AI Abdomen" notes testing "with and without users editing the landmark locations" for measurements, implying that the algorithm's initial measurement (standalone) was evaluated. "Trace AI" describes experts refining initial circles given by Trace AI, and then comparing the expert-refined contour to the contour detected by Trace AI algorithm, further confirming a standalone evaluation of the algorithm's output.
7. Type of Ground Truth Used
AI Abdomen:
- View Classification: Ground truth view labels were associated with each image at the time of imaging using a system protocol. This suggests a form of expert labeling or pre-categorization at the point of data capture.
- Semi-automated Measurements: Expert consensus (or at least independent expert measurements) provided by three registered Clinical Sonographers.
Trace AI:
- Expert consensus/independent expert annotations: Ground truth measurements (delineations of orifice structures) were provided by three registered Clinical Sonographers independently.
8. Sample Size for the Training Set
- The document does not explicitly state the sample size for the training set for either AI Abdomen or Trace AI. It only focuses on the test set and ensures its independence from the training data.
9. How the Ground Truth for the Training Set Was Established
- The document does not explicitly state how the ground truth for the training set was established. It only mentions that the testing dataset was from different clinical sites than those used for training, to ensure independence.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
November 19, 2024
Siemens Medical Solutions USA, Inc. Sulgue Choi Regulatory Affairs Manager 22010 South East 51st Street Issaquah, Washington 98029
Re: K242523
Trade/Device Name: ACUSON Sequoia Diagnostic Ultrasound System: ACUSON Sequoia Select Diagnostic Ultrasound System; ACUSON Origin Diagnostic Ultrasound System; ACUSON Origin ICE Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX, OIJ, QIH, OBJ Dated: August 29, 2024 Received: October 28, 2024
Dear Sulgue Choi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YANNA S. KANG -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
Submission Number (if known)
Device Name
ACUSON Sequoia Diagnostic Ultrasound System;
ACUSON Sequoia Select Diagnostic Ultrasound System;
ACUSON Origin Diagnostic Ultrasound System;
ACUSON Origin ICE Diagnostic Ultrasound System
Indications for Use (Describe)
ACUSON Sequoia and Sequoia Select
The ACUSON Sequoia and Sequoia Select ultrasound imaging systems are intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Neonatal Cephalic, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Cardiac, Transesophageal, Pelvic, Vascular, Adult Cephalic, Musculoskeletal and Peripheral Vascular applications.
The system supports the Ultrasonically-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult and pediatric patients with hepatic steatosis.
The system also provides the ability to measure anatomical structures for fetal, abdominal, pediatric, small organ, cardiac, transvaginal, peripheral vessel, musculoskeletal and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Operating Modes
2D-mode
- · 2D-mode
- · 2D-mode with Harmonics Imaging
- · 2D-mode with Harmonics Imaging for Contrast Agent Imaging
- Color flow Doppler
- · Color (velocity)
- · Power (energy)
Doppler
- · Pulsed Wave Doppler
- · Pulsed Wave Doppler Tissue Imaging
- · High Pulsed Repetition Frequency Pulsed Wave Doppler
- · Steerable Continuous Wave Doppler for imaging transducers
- · Continuous Wave Doppler for non-imaging transducers M-mode
- · M-mode with Harmonics Imaging
- · Anatomical M-Mode
- Elastography
- · Strain Imaging
- · Shear Wave Elastography
3D/4D Volume Imaging
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
{4}------------------------------------------------
Combined Modes 2D-mode with color 2D-mode with Doppler 2D-mode with color and Doppler 2D-mode with M-mode 2D-mode with M-mode and Color 2D-mode with Elastography 3D/4D Volume Imaging with color
ACUSON Origin and Origin ICE
The ACUSON Origin and Origin ICE ultrasound imaging systems are intended to provide images of. or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Cardiac, Transesophageal, Intracardiac, Vascular, Adult Cephalic, and Peripheral Vascular applications.
The catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
The system also provides the ability to measure anatomical structures for fetal, abdominal, pediatric, cardiac, peripheral vessel, and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Operating Modes
2D-mode
· 2D-mode
- · 2D-mode with Harmonics Imaging
- · 2D-mode with Harmonics Imaging for Contrast Agent Imaging
- Color flow Doppler
- · Color (velocity)
- · Power (energy)
Doppler
- · Pulsed Wave Doppler
- · Pulsed Wave Doppler Tissue Imaging
- · High Pulsed Repetition Frequency Pulsed Wave Doppler
- · Steerable Continuous Wave Doppler for imaging transducers
- · Continuous Wave Doppler for non-imaging transducers
- M-mode
- · M-mode with Harmonics Imaging
- · Anatomical M-Mode
3D/4D Volume Imaging
Combined Modes
2D-mode with color
2D-mode with Doppler
2D-mode with color and Doppler
{5}------------------------------------------------
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
510(k) Summary
K242523
| Date: | August 21, 2024 | ||
|---|---|---|---|
| 1. Sponsor: | Siemens Medical Solutions USA, Inc.Ultrasound Division22010 South East 51st StreetIssaquah, Washington 98029 | ||
| Contact Person: | Sulgue ChoiTel: (425) 281-9898 | ||
| 2. Device Name: | ACUSON Sequoia Diagnostic Ultrasound SystemACUSON Sequoia Select Diagnostic Ultrasound SystemACUSON Origin Diagnostic Ultrasound SystemACUSON Origin ICE Diagnostic Ultrasound System | ||
| Common Name: | Diagnostic Ultrasound System with Accessories | ||
| Classification: | Regulatory Class:IIClassification Panel:Radiology | ||
| Classification Name | 21 CFRSection | ProductCode | |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN | |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO | |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX | |
| Biopsy Needle Guide Kit | 892.1560 | OIJ | |
| Automated Radiological Image ProcessingSoftware | 892.2050 | QIH | |
| Ultrasound Intravascular Catheter | 892.1200 | OBJ | |
| ManufacturingSite: | Siemens Medical Solutions USA, Inc.22010 South East 51st Street,Issaquah, Washington 98029, UNITED STATES |
{7}------------------------------------------------
Siemens Healthcare s.r.o.
Panattoni Park Kosice Airport ul. Andreja Kvasa 5 040 17. Kosice-Barca, Slovakia
3. Legally Marketed Predicate Devices and Reference Device
The ACUSON Sequoia, Sequoia Select, Origin and Origin ICE Diagnostic Ultrasound Systems are multi-purpose, diagnostic ultrasound systems with accessories and proprietary software, and are substantially equivalent to the company's own productsthe ACUSON Sequoia, Sequoia Select, Origin and Origin ICE (K240704) which is the primary predicate device and the ACUSON Juniper Select (K230207) as the reference device.
4. Device Description
The ACUSON Sequoia, Sequoia Select, Origin and Origin ICE Diagnostic Ultrasound Systems are multi-purpose, mobile, software-controlled, diagnostic ultrasound systems with an on-screen display of thermal and mechanical indices related to potential bioeffect mechanisms. The ultrasound system function is to transmit and receive ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes, Panoramic Imaging, Contrast agent Imaging, Virtual Touch Strain Imaging (except Origin), Virtual Touch - pSWE Imaging, Virtual Touch - SWE Imaging, Custom Tissue Imaging, 3D/4D Volume lmaging or Harmonic Imaging on a Display and provide cardiac anatomical and quantitative function software applications.
5. Intended Use/Indications for Use
ACUSON Sequoia and ACUSON Sequoia Select
The ACUSON Sequoia and Sequoia Select ultrasound imaging systems are intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Neonatal Cephalic, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Cardiac, Pelvic, Vascular, Adult Cephalic, Musculoskeletal and Peripheral Vascular applications.
The system supports the Ultrasonically-Derived Fat Fraction (UDFF) measurement tool to report an index that can be useful as an aid to a physician managing adult and pediatric patients with hepatic steatosis.
The system also provides the ability to measure anatomical structures for fetal, abdominal, pediatric, small organ, cardiac, transvaginal, peripheral vessel,
{8}------------------------------------------------
musculoskeletal and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Operating Modes
2D-mode
- . 2D-mode
- . 2D-mode with Harmonics Imaging
- . 2D-mode with Harmonics Imaging for Contrast Agent Imaging Color flow Doppler
- Color (velocity) ●
- . Power (energy)
Doppler
- . Pulsed Wave Doppler
- . Pulsed Wave Doppler Tissue Imaging
- . High Pulsed Repetition Frequency Pulsed Wave Doppler
- . Steerable Continuous Wave Doppler for imaging transducers
- . Continuous Wave Doppler for non-imaging transducers
M-mode
- M-mode with Harmonics Imaging .
- . Anatomical M-Mode
Elastography
-
- Strain Imaging
-
- Shear Wave Elastography
3D/4D Volume Imaging
Combined Modes 2D-mode with color 2D-mode with Doppler 2D-mode with color and Doppler 2D-mode with M-mode 2D-mode with M-mode and Color 2D-mode with Elastography 3D/4D Volume Imaging with color
ACUSON Origin and Origin ICE
The ACUSON Origin and Origin ICE ultrasound imaging systems are intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal. Pediatric. OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Cardiac, Transesophageal, Intracardiac, Vascular, Adult Cephalic, and Peripheral Vascular applications.
{9}------------------------------------------------
The catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
The system also provides the ability to measure anatomical structures for fetal, abdominal, pediatric, cardiac, peripheral vessel, and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Operating Modes
2D-mode
- . 2D-mode
- . 2D-mode with Harmonics Imaging
- 2D-mode with Harmonics Imaging for Contrast Agent Imaging .
Color flow Doppler
- Color (velocity) ●
- Power (energy) .
Doppler
- Pulsed Wave Doppler .
- Pulsed Wave Doppler Tissue Imaging .
- . High Pulsed Repetition Frequency Pulsed Wave Doppler
- Steerable Continuous Wave Doppler for imaging transducers .
- . Continuous Wave Doppler for non-imaging transducers
M-mode
- M-mode with Harmonics Imaging ●
- Anatomical M-Mode .
3D/4D Volume Imaging
Combined Modes 2D-mode with color 2D-mode with Doppler 2D-mode with color and Doppler 2D-mode with M-mode 2D-mode with M-mode and Color 4D Volume Imaging with color
6. Substantially Equivalent Devices and Summary of Technological Characteristics
The modified ACUSON Sequoia, Sequoia Select, Origin and Origin ICE Ultrasound Systems are based on the same system configuration as the company's own currently cleared ACUSON Sequoia, Sequoia Select, Origin ICE (K240704) with regard to the intended use and technological characteristics. The ACUSON Sequoia, Sequoia Select, Origin and Origin ICE will have different software and transducer
ACUSON Sequoia, Sequoia Select, Origin ICE Diagnostic Ultrasound System 510(k) Submission Page 4 of 10
{10}------------------------------------------------
options for different markets. The ACUSON Sequoia, Sequoia Select, Origin and Origin ICE will offer an optional cardiac imaging and analysis package with new transducers and software applications have been previously cleared by FDA under K240704 and not modified in the current submission. The modified ultrasound systems under this review and the predicate ultrasound systems function in the same manner as all diagnostic ultrasound systems and transducers.
The submission devices include following changes from predicate devices:
ACUSON Sequoia and Sequoia Select
- The addition of the 11M2 transducer which is substantially equivalent to the . 11M3 transducer previously cleared on ACUSON Sequoia, Sequoia Select, Origin, and Origin ICE (K240704).
- . The addition of the HLX transducer which is substantially equivalent to the 15L4 and 18L6 transducer previously cleared on ACUSON Sequoia, Sequoia Select, Origin, and Origin ICE (K240704).
- The addition of needle quide kit, HLX, which is substantially equivalent to the . needle guide Kit, 16L4 previously cleared on the ACUSON Juniper (K230207)
- . The modified ACUSON Sequoia and Sequoia Select Ultrasound System includes the expansion of the 5Z1 transduce, 4D Heart Al and Al Assist which was already cleared on ACUSON Sequoia, Sequoia Select, Origin, and Origin ICE (K240704).
- . Also this Traditional 510(k) includes Al Abdomen which leverages Al Measure and Al Assist features released on the predicate ACUSON Origin (K240704), Trace Al which is an improvement of the generic eclipse measurement tool released on predicate ACUSON Origin (K240704), Needle Enhancement and Freehand 3D Volume Imaging which were cleared on the predicate device, ACUSON Sequoia, Sequoia Select, Origin, and Origin ICE (K240704) and ACUSON Juniper and Juniper Select (K230207) to enable an improved customer experience.
ACUSON Origin and Origin ICE
- . The addition of the HLX transducer which is substantially equivalent to the 15L4 and 18L6 transducer previously cleared on ACUSON Sequoia, Sequoia Select, Origin, and Origin ICE (K240704).
- . The addition of needle quide kit, HLX, which is substantially equivalent to the needle guide Kit, 16L4 previously cleared on the ACUSON Juniper (K230207)
- . Also this Traditional 510(k) includes Al Abdomen which leverages Al Measure and Al Assist features released on the predicate ACUSON Origin (K240704), Trace Al which is an improvement of the generic eclipse measurement tool released on predicate ACUSON Origin (K240704) to enable an improved customer experience.
All other hardware and software features of the ACUSON Sequoia, Sequoia Select,
ACUSON Sequoia, Sequoia Select, Origin ICE Diagnostic Ultrasound System 510(k) Submission Page 5 of 10
{11}------------------------------------------------
Origin and Origin IEC Diagnostic Ultrasound devices remain unchanged. The foundation of the ACUSON Sequoia, Sequoia Select, Origin and Origin ICE (this submission) is the ACUSON Sequoia, Sequoia Select, Origin, and Origin ICE (K240704) with features and transducers integrated with the ACUSON Sequoia, Sequoia Select, Origin, and Origin ICE (K240704) hardware and the ACUSON Sequoia, Sequoia Select, Origin and Origin ICE (this submission) reuse software developed for Sequoia, Sequoia Select and Origin (K240704) mainly.
7. Nonclinical Data
The subject devices have been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and have been found to conform to applicable medical device safety standards. The systems comply with the following voluntary standards:
- . IEC 62359: Edition 2.1 2017-09, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields / Combines IEC 62359 (2010-10) and AMD 1 (2017-09)
- AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005)
- IEC 60601-1:2005/A1(2012), Medical electric equipment Part 1: General . requirements for basic safety and essential performance / This document and its separate amendments continue to be valid together with the consolidated version
- 트 IEC 60601-1-2 Edition 4.0 2014-02, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- 트 IEC 60601-2-18: Edition 3.0 2009-08. Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment Part 2-37: .. Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
{12}------------------------------------------------
- . ISO 10993-1 Fifth edition 2018-08, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- IEC 60601-1-6:2010+A1:2013+A2:2020 Medical Electrical Equipment Part 1-6, General Requirements for Basic Safety and Essential Performance- Collateral standard: Usability
- ANSI AAMI ISO 14971: Medical devices- Applications of risk management to medical devices, 2019
- .. FDA Ultrasound Guidance document, "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers," issued in February 2023 (https://www.fda.gov/media/71100/download) for determining the measurement accuracy
Al Summary of Testing
Al Abdomen
- Summary of test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance
- Al Abdomen covers a set of semi-automated Abdomen view classifications and । measurements for B-Mode acquisitions.
- -The algorithm achieved a success rate of 77.8% for view classification when aggregated across all 17 view types. When eliminating CBD and pancreas sagittal from the aggregated results, the accuracy increased to 82.5%.
- -For semi-automated measurements, the Individual Equivalence Coefficient (IEC) fell below the prespecified success criterion of 0.25 for all 12 measurements, with and without users editing the landmark locations.
- The number of individual patients' images were collected from
- The dataset was consisted of 105 exams from 3 institutions collected on the following transducers: 4V1, 5C1, 9C2, 10L4, 11M2, 11M3, and DAX.
- The number of samples, if different from above, and the relationship between the two
- A total of 1,785 images were extracted from the 105 exams, corresponding to the 17 abdomen views.
- Demographic distribution including Gender, Age, Ethnicity
{13}------------------------------------------------
- Age: 10 74 years -
- Gender: Female; 57%, Male; 41%, Unknown; 2% -
- Ethnicity/Country: US -
- BMI: 16.3 47.3 -52% Normal (BMI <= 25 kg/m2) 48% Overweight + Obese (BMI > 25 kg/m2)
- Information about clinical subgroups and confounders present in the dataset
- During testing of the Al algorithm, we have included images from subjects with । Normal and High BMI. The test datasets consist of 15 B-Mode images per view and per transducer.
- Information about equipment and protocols used to collect images
- The dataset consists of exams from three institutions, seven probes (4V1, 5C1. -9C2, 10L4, 11M2, 11M3, DAX) and different Sequoia Systems. A system protocol was distributed to each data collection site to ensure standardization of the data collection process.
- Information about how the reference standard was derived from the dataset (i.e., the "truthing" process)
- For view classification, the ground truth view label was associated with each image at the time of imaging using a system protocol.
- -Ground truth measurements were provided by three clinical users, all registered Clinical Sonographers with ARDMS accreditation (or equivalent), each with at least 3 years of clinical experience.
- Description of how independence of test data from training data was ensured
- -To ensure that the testing dataset is not mixed with the training data, we used datasets from different clinical sites for testing as compared to the clinical sites for training.
Trace Al
- Summary of test statistics or other test results including acceptance criteria or other information supporting the appropriateness of the characterized performance
{14}------------------------------------------------
- -Trace AI is a semi-automated measurement tool that provides an ability to measure orifice-tvpe anatomic structures on MPRs generated from volumetric ICE. TEE and TTE transducers.
- The accuracy goal is the 90th percentile shall have a minimum DICE coefficient of at least 80% with 90% confidence.
- The lower 90% confidence bound for the 90th percentile is 0.95, which is greater than the requirement of 0.8, therefore, the test passes.
- The number of individual patients' images were collected from
- The testing data for our internal verification is from 10 different adult patients ।
- The number of samples, if different from above, and the relationship between the two
- -24 volumes were extracted from 10 exams.
- Demographic distribution including Gender, Age, Ethnicity ■
- Age: 18 79 years -
- Gender: Female; 25%, Male; 62%, Unknown; 13% -
- Ethnicity/Country: US, Germany, Mexico -
- BMI: 18.1 44.2 -
20% Normal (BMI <= 25 kg/m2) 20% Overweight + Obese (BMI > 25 ka/m2) 60% Unknown
- Information about clinical subgroups and confounders present in the dataset:
- -The data used for testing captures a wide range of cardiac structures from different imaging transducers including 3D TEE, 3D TTE and 3D ICE. A total number of 24 samples were created during internal examinations, consisting of a variety of structures: left ventricle (LV), right ventricle (RV), left ventricle outflow tract (LVOT), left atrium (LA), ascending aorta (Asc Ao), mitral valve annulus (MV Annulus), ascending aorta sinotubular junction (Ao ST jnct.), left atrium appendage opening (LA App. Opening), sinus of valsalva (AO Sin. Vals.), tricuspid annulus (TV Ann). In summary, the test data are representative of the intended use population for TraceAl.
- Information about equipment and protocols used to collect images:
- The datasets were collected from three institutions in the US, Mexico and -Germany. The testing data was collected using the Origin system at external
{15}------------------------------------------------
clinical sites. Transducers used include 5Z1 (3D TTE), Z6T (3D TEE) and Lumos (3D ICE).
- . Information about how the reference standard was derived from the dataset (i.e., the "truthing" process)
- Ground truth measurements were provided by three clinical users, all registered -Clinical Sonographers with ARDMS accreditation (or equivalent), each with at least 3 years of clinical experience.
- Each sonographer was given the same volume images with the same -MPRs, the target orifice structures to delineate and which MPR the structure is located in (which avoids confusion when multiple orifice structures appear in the same MPR). Then each of the sonographers is asked to independently refine the initial circle given by Trace41 to annotate the underlying orifice structure. During the process, sonographers can only see their own annotations and cannot see annotation results from other sonographers or a contour detected by TraceAl algorithm. The contour created by sonographers (Ground Truth) is compared with the contour detected by TraceAl (Detection Truth).
- Description of how independence of test data from training data was ensured
- To ensure that the testing dataset is not mixed with the algorithm training data. we used datasets from different clinical sites for verification testing as compared to the clinical sites for algorithm training.
8. Clinical Data
Since the ACUSON Sequoia, Sequoia Select, Origin and Origin ICE Diagnostic Ultrasound Systems use the same technology and principles as existing devices, clinical studies were not required to support substantial equivalence.
9. Summary
Intended uses and other key features are consistent with traditional clinical practice and FDA quidelines. The design and development process of the manufacturer conforms to 21 CFR 820 Quality System Requlation and ISO 13485:2016 quality system standards. The product is designed to conform to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound system has accumulated a long history of safe and effective performance. Therefore, it is the opinion of Siemens Medical Solutions USA. Inc. that the ACUSON Sequoia, Sequoia Select, Origin ICE systems are substantially equivalent with respect to safety and effectiveness of the primary predicate device.
ACUSON Sequoia, Sequoia Select, Origin ICE Diagnostic Ultrasound System 510(k) Submission Page 10 of 10
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.