(40 days)
Pocguide Multi-Drug Test Panel OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:
| Drug ( Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 1000 ng/mL or 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Benzoylecognine (COC) | 300 ng/mL or 150 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
| Methamphetamine (MET) | 1000 ng/mL or 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (OPI2000/MOP300) | 2000 ng/mL or 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
The single or multi-test panels can consist of up to the above listed analytes in any combination. The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
For over-the-counter use. For in vitro diagnostic use only
Pocguide™ Multi-Drug Test Panel and Pocguide™ Multi-Drug Test Panel OTC are immunochromatographic assays that use a lateral flow system for the qualitative detection of single drugs in human urine at or above the cut-off levels as indicated. The products are single use in vitro diagnostic devices.
This device is a dipcard format in which the test strips are integrated into the plastic dipcard. After removing the cap of the dipcard, the absorbent end of the test strips is exposed and can be in direct contact with the urine sample. The device is in a ready-to-use format and no longer requires assembly before use.
The provided document describes the Pocguide Multi-Drug Test Panel and Pocguide Multi-Drug Test Panel OTC, which are in vitro diagnostic devices for qualitative and simultaneous detection of various drugs in human urine.
Here's an analysis of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this type of qualitative diagnostic device are typically related to its analytical performance, specifically precision (reproducibility) around the cutoff concentration, and its ability to correctly identify positive and negative samples when compared to a confirmed method (method comparison).
Acceptance Criteria (Implied based on study design and regulatory context for qualitative drug tests):
- Precision/Reproducibility: A high percentage of agreement (e.g., typically >80%) for samples near the cutoff (e.g., +/- 25% cutoff, cutoff itself) over multiple lots and runs, and 100% agreement for samples far from the cutoff (e.g., +/- 100% cutoff).
- Method Comparison: High overall agreement (sensitivity and specificity) with a gold standard confirmatory method (LC/MS or GC/MS) for clinical samples, especially for true positive and true negative samples. Acceptable performance for samples near the cutoff where some discordance is expected due to the nature of qualitative assays.
- Analytical Specificity (Cross-Reactivity): No significant cross-reactivity with common substances or structurally similar compounds to avoid false positives.
- Interference: No interference from common physiological substances found in urine.
- Lay-User Study (for OTC devices): High agreement with trained professionals and ease of use for the intended lay user.
Reported Device Performance (from "Precision/Reproducibility" and "Method Comparison" sections):
| Test Characteristic | Drug (Cutoff) Examples | Reported Device Performance |
|---|---|---|
| Precision/Reproducibility | AMP 500 ng/mL, BUP 10 ng/mL, etc. | +100%, +75%, +50%, +25% Cutoff: 100.0% Positive (Across all tested drugs and cutoffs) -100%, -75%, -50%, -25% Cutoff: 100.0% Negative (Across all tested drugs and cutoffs) Cutoff: * AMP 500: 82.0% Positive, 18.0% Negative * BUP 10: 84.0% Positive, 16.0% Negative * BAR 300: 82.7% Positive, 17.3% Negative * And similar ranges for other drugs listed in Table 2. Each drug showed similar performance around the cutoff. |
| Method Comparison | AMP 1000 ng/mL, AMP 500 ng/mL, BAR 300 ng/mL, etc. | Excellent agreement for Drug-Free, Low Negative, and High Positive Samples: Typically 100% correct classification by the device for these categories (e.g., "Viewer A Positive" for High Positive by LC/MS and "Viewer A Negative" for Drug-Free by LC/MS are 100% for almost all drugs). Expected Discordance Near Cutoff: As anticipated for qualitative tests, some samples near the cutoff (especially -25% and +25%) show mixed results (discordance) between the device and LC/MS, as detailed in Table 5 and Table 6 (Discordant results). These are typically within acceptable ranges for qualitative tests, acknowledging the inherent variation around a precise cutoff. |
| Analytical Specificity | AMP, BUP, BAR, etc. | Tested numerous substances. Most showed no cross-reactivity or very low percentages at very high concentrations, indicating good specificity. Specific cross-reactivity percentages are provided in Table 3. |
| Interference | N/A - broadly tested | No interference observed for a wide range of common substances and physiological conditions (urine specific gravity 1.000-1.035, pH 4-9) as listed in Tables 4. |
| Lay-User Study | All Configuration 1 & 2 Drugs | Agreement (%): * -100%, -75%, -50% Cutoff: 100% negative calls. * +25%, +50%, +75% Cutoff: Mostly 100% positive calls, some 95% for +25% cutoff. * -25% Cutoff: 95% negative calls for most drugs. * Raw numbers show 19/20 or 20/20 correct calls for most categories. Ease of Use: All participants indicated instructions were easy to understand and follow (Flesch-Kincaid Grade Level 7). |
2. Sample Sizes and Data Provenance
- Test Set Sample Sizes:
- Precision/Reproducibility: For each drug and each cutoff, 50 samples were tested at each concentration level (-100%, -75%, -50%, -25%, Cutoff, +25%, +50%, +75%, +100%). This was done across 3 lots, so $50 \text{ samples/level} \times 9 \text{ levels} \times 3 \text{ lots} = 1350$ tests per drug. (For AMP alone, this would be $1350 \times 2 \text{ cutoffs} = 2700$ tests).
- Method Comparison: 100 unaltered clinical samples were used for each target drug (40 negative, 40 positive, and an additional 20 samples around the cutoff as seen in the breakdown of results). So, for 13 drug analytes, this would be $13 \times 100 = 1300$ clinical samples.
- Cross-Reactivity / Interference: Specific numbers for each substance are not given, but samples were spiked at various concentrations and tested using three lots of each device.
- Lay-User Study: For Configuration 1, 140 participants (58 male, 82 female). For Configuration 2, 140 participants (56 male, 84 female). Each participant tested 1 blind-labeled sample. For each drug within each configuration, 20 samples were prepared per concentration level (-100%, -75%, -50%, -25%, +25%, +50%, +75%).
- Data Provenance: The document does not explicitly state the country of origin for the data. The consulting firm is in Shanghai, China, and the applicant's address is Irvine, CA, USA. Given the FDA 510(k) submission, it's implied that the data is intended to be representative and valid for the US market. The studies are described as retrospective as they involve samples prepared at specific concentrations or existing clinical samples compared to a gold standard.
3. Number of Experts and their Qualifications
- For Precision/Reproducibility, Cross-Reactivity, Interference, and Method Comparison: The document does not explicitly state the number of "experts" used to establish ground truth or interpret results. These are quantitative/analytical laboratory tests where the ground truth (concentration by LC/MS or GC/MS) is established by analytical instrumentation. The "Viewers" (A, B, C) mentioned in the Method Comparison section appear to be individuals performing the visual interpretation of the device results, not necessarily independent experts establishing ground truth. Their qualifications are not specified but are implied to be trained laboratory personnel.
- For Lay-User Study: No "experts" were used to establish ground truth for the lay-user study. The ground truth for the samples used in this study was established by LC-MS/MS confirming the spiked drug concentrations.
4. Adjudication Method for the Test Set
- For Precision/Reproducibility, Cross-Reactivity, Interference: No adjudication method is described. The results are reported as counts of positive/negative readings against a known (spiked) concentration.
- For Method Comparison: No explicit "adjudication" among multiple readers is described. Results for the candidate device were observed by "Viewer A, B, C." The comparison is directly between the "Candidate Device Result" (presumably individual Viewer results, though aggregated in Table 5) and the LC/MS reference method. The discordant results in Table 6 specify which viewers made the discordant call (e.g., "Viewer A, B," "Viewer C"). This suggests independent readings by three viewers, but no formal adjudication process to resolve disagreements among them is mentioned; the individual viewer calls are presented relative to the ground truth.
- For Lay-User Study: No adjudication method is mentioned. The tables report the number of positive/negative results per concentration.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done. The study design is an analytical performance study and a comparison study against a laboratory reference method, along with a lay-user study for OTC claims. It assesses device performance in a standalone or simulated user setting, not direct human reader improvement with AI assistance. The device is a lateral flow immunoassay, not an AI-powered diagnostic.
6. Standalone (Algorithm Only) Performance
- This question is not applicable as the device is a lateral flow immunochromatographic assay, not an algorithm or software-based diagnostic. Its performance is inherent to the chemical reactions on the test strip and visual interpretation, not an algorithm.
7. Type of Ground Truth Used
- The primary ground truth used for performance evaluation (Precision/Reproducibility, Method Comparison, Lay-User Study) is analytical confirmation by Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions (LC-MS/MS), which are stated as the "preferred confirmatory methods." This is a highly accurate and quantitative method for determining drug concentrations.
8. Sample Size for the Training Set
- This question is not applicable. The device is a qualitative diagnostic test based on immunoassay principles, not a machine learning or AI-based device that requires a "training set" in the computational sense. The development of the immunoassay itself relies on antigen-antibody binding characteristics and optimization, not a deep learning model.
9. How the Ground Truth for the Training Set Was Established
- This question is not applicable for the same reasons as #8. The "ground truth" for developing the test and optimizing its performance would be established through standard immunoassay R&D processes, involving controlled experiments with known concentrations of analytes and cross-reactants, guided by established analytical chemistry principles.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The logo is simple and professional, reflecting the agency's role in protecting public health.
Aicheck Biotech, Inc. % Dylan Wu IVD Senior Consultant Shanghai Sungo Management Consulting Co., Ltd. 14th Floor, 1500# Century Avenue Shanghai, 200122, China
Re: K242498
Trade/Device Name: Pocguide Multi-Drug Test Panel, Pocguide Multi-Drug Test Panel OTC Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: NFT, NGL, PTH, NFV, NFY, PTG, NGG, NGM, QAW, NFW Dated: August 22, 2024 Received: August 22, 2024 Dear Dylan Wu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely. Digitally signed by Joseph A. Joseph A. Kotarek -S Kotarek -S Date: 2024.10.0 10:15:07 -04'00' Joseph Kotarek Branch Chief for Toxicology Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K242498
Device Name
Pocguide Multi-Drug Test Panel OTC Pocguide Multi-Drug Test Panel
Indications for Use (Describe)
Pocquide Multi-Drug Test Panel OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:
| Drug ( Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 1000 ng/mL or 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Benzoylecognine (COC) | 300 ng/mL or 150 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
| Methamphetamine (MET) | 1000 ng/mL or 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (OPI2000/MOP300) | 2000 ng/mL or 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
The single or multi-test panels can consist of up to the above listed analytes in any combination. The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
For over-the-counter use. For in vitro diagnostic use only
Pocguide Multi-Drug Test Panel is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:
| Drug ( Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 1000 ng/mL or 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Benzoylecognine (COC) | 300 ng/mL or 150 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
| Methamphetamine (MET) | 1000 ng/mL or 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (OPI2000/MOP300) | 2000 ng/mL or 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
{3}------------------------------------------------
| Phencyclidine (PCP) | 25 ng/mL |
|---|---|
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
| The single or multi-test panels can consist of up to the above listed analytes in any combination. | |
| The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended | |
| to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should | |
| be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. | |
| The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a | |
| confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass | |
| Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful | |
| consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating | |
| preliminary positive results. | |
| For in vitro diagnostic use only |
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(K) Summary
K242498
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement by 21 CFR 807.92
Date prepared: 15th, Aug 2024
1 Submitter's Information
Name: Aicheck Biotech, Inc. Address: 17701 Cowan Ste 230, Irvine, CA, USA, 92614 Contact Person: Lisa Liu Title: Manager of the quality department Tel: +86-571-89055299 Email: ra.rep@aichek.com
2 Official Contact Person Information
Primary contact: Mr. Dylan Wu Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-18616797137 Email: haiyang.wu@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com
3 Subject Device
3.1 Trade Name and Regulatory Information
Pocguide™ Multi-Drug Test Panel Pocguide™ Multi-Drug Test Panel OTC
3.2 Regulatory Information
| Drug (Identifier) | Code (s) | Classification | Regulation Section | Panel |
|---|---|---|---|---|
| Amphetamine (AMP) | NFT | Class II | 21 CFR § 862.3100Amphetamine Test System | ClinicalToxicology |
| Buprenorphine (BUP) | NGL | Class II | 21 CFR § 862.3650Opiate Test System | ClinicalToxicology |
| Secobarbital (BAR) | PTH | Class II | 21 CFR § 862.3150Barbiturate test system | ClinicalToxicology |
| Oxazepam (BZO) | NFV | Class II | 21 CFR § 862.3170Benzodiazepine test system | ClinicalToxicology |
| Benzoylecognine (COC) | NFY | Class II | 21 CFR § 862.3250Cocaine and Cocaine Metabolites Test System | ClinicalToxicology |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | PTG | Class II | 21CFR § 862.3620Methadone Test System | ClinicalToxicology |
| Methamphetamine (MET) | NGG | Class II | 21 CFR § 862.3610Methamphetamine Test System | ClinicalToxicology |
{5}------------------------------------------------
| Drug (Identifier) | Code (s) | Classification | Regulation Section | Panel |
|---|---|---|---|---|
| Methylenedioxymethamphetamine (MDMA) | NGG | Class II | 21 CFR § 862.3610Methamphetamine Test System | ClinicalToxicology |
| Morphine (OPI/MOP) | NGL | Class II | 21 CFR § 862.3650Opiate Test System | ClinicalToxicology |
| Methadone (MTD) | PTG | Class II | 21CFR § 862.3620Methadone Test System | ClinicalToxicology |
| Oxycodone (OXY) | NGL | Class II | 21 CFR § 862.3650Opiate Test System | ClinicalToxicology |
| Phencyclidine (PCP) | NGM | Unclassified | ||
| Nortriptyline (TCA) | QAW | Class II | 21 CFR § 862.3910Tricyclic antidepressant drugs test system | ClinicalToxicology |
| Marijuana (THC) | NFW | Class II | 21 CFR § 862.3870Cannabinoid test system | ClinicalToxicology |
CA IICA 92614
4 Predicate device
K202567
Wondfo T-Dip® Multi-Drug Urine Test Panel Wondfo T-Dip® Multi-Drug Urine Test Panel Rx
5 Indications for use/Intended use
Pocguide™ Multi-Drug Test Panel is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Mothadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 1000 ng/mL or 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Benzoylecognine (COC) | 300 ng/mL or 150 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
| Methamphetamine (MET) | 1000 ng/mL or 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (OPI2000/MOP300) | 2000 ng/mL or 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
The single or multi-test panels can consist of up to the above listed analytes in any combination.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
For in vitro diagnostic use only.
{6}------------------------------------------------
Pocguide™ Multi-Drug Test Panel OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencycline and Marijuana in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 1000 ng/mL or 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Benzoylecognine (COC) | 300 ng/mL or 150 ng/mL |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 ng/mL |
| Methamphetamine (MET) | 1000 ng/mL or 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (OPI2000/MOP300) | 2000 ng/mL or 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
The single or multi-test panels can consist of up to the above listed analytes in any combination.
The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
For over-the-counter (OTC) use. For in vitro diagnostic use only.
б Device Description
Pocguide™ Multi-Drug Test Panel and Pocguide™ Multi-Drug Test Panel OTC are immunochromatographic assays that use a lateral flow system for the qualitative detection of single drugs in human urine at or above the cut-off levels as indicated. The products are single use in vitro diagnostic devices.
This device is a dipcard format in which the test strips are integrated into the plastic dipcard. After removing the cap of the dipcard, the absorbent end of the test strips is exposed and can be in direct contact with the urine sample. The device is in a ready-to-use format and no longer requires assembly before use.
7 Principle of Operation
Pocguide™ Multi-Drug Test Panel is a competitive immunoassay that is used to screen for the presence of drugs of abuse in urine. It is chromatographic absorbent device in which drugs or drug metabolites in a sample competitively combined to a limited number of antibody-dye conjugate binding sites. When the absorbent end of the test device is immersed into the urine sample, the urine is absorbed into the device by capillary action, mixes with the antibody
{7}------------------------------------------------
dye conjugate, and flows across the pre-coated membrane.
When sample drug levels are at or above the target cutoff (the detection sensitivity of the drug in the sample binds to the antibody-dye conjugate preventing the antibody-dye conjugate from binding to the drug-protein pre-coated in the test region (T). This prevents the development of a distinct colored band in the test region indicating a potentially positive result.
When sample drug levels are zero or below the target cutoff, antibody-dye conjugate binds to the drug-protein precoated in the test region (T) of the device. This produces a colored test line that, regardless of its intensity, indicates a negative result.
To serve as a procedural control, a colored line will always appear at the control region (C) indicating that proper volume of specimen has been added and membrane wicking has occurred.
{8}------------------------------------------------
Substantial Equivalence Information 8
8.1 Technological characteristics
| Table 1 Comparison of Technological characteristics | ||
|---|---|---|
| Device | Proposed Device | Predicate Device |
|---|---|---|
| 510K number | K242498 | K202567 |
| Device name | Pocguide™ Multi-Drug Test PanelPocguide™ Multi-Drug Test Panel OTC | Wondfo T-Dip® Multi-Drug Urine Test PanelWondfo T-Dip® Multi-Drug Urine Test Panel Rx |
| Indication (s) for use | For the qualitative determination of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Marijuana in human urine. | Same |
| Methodology | Competitive binding, lateral flow immunochromatographic assay based on antigen-antibody reaction | Same |
| Type of Test | Qualitative | Same |
| Specimen Type | Human Urine | Same |
| Target Drugand CutoffValue | Target Drug | Cutoff (ng/mL) |
| Amphetamine (AMP) | 1000 or 500 | |
| Buprenorphine (BUP) | 10 | |
| Secobarbital (BAR) | 300 | |
| Oxazepam (BZO) | 300 | |
| Benzoylecognine (COC) | 300 or 150 | |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 | |
| Methamphetamine (MET) | 1000 or 500 | |
| Methylenedioxymethamphetamine (MDMA) | 500 | |
| Morphine (OPI2000/MOP300) | 2000 or 300 | |
| Methadone (MTD) | 300 | |
| Oxycodone (OXY) | 100 | |
| Phencyclidine (PCP) | 25 | |
| Nortriptyline (TCA) | 1000 | |
| Marijuana (THC) | 50 | The number of drugs detected are different, the predicate device also includes Propoxyphene (PPX) 300 ng/mL |
| Configurations | Test Panel | Same |
{9}------------------------------------------------
д Summary of Non-Clinical Testing
9.1 Analytical performance
Precision / Reproducibility a
Precision studies were carried out for samples with contract, - 75% atoff, -25% cutoff, -25% cutoff, -15% cutoff, +50% cutoff +15% cutoff +15% cutoff and +100% cutoff. Samples with concentration of -100% cutoff were prepared by spiking target drug in drug-free urines samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were tested in a blinded fashion, tests were performed in 2 runs per day for 25 days using 3 lots of test panels. The results obtained are summarized in Table 2.
| Drug testedClaimed Cutoff | %cutoff | Lot 1 | Lot 2 | Lot 3 | Total Result | %Positive | %Negative | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| + | - | + | - | + | - | + | - | |||||
| AMP500 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| AMP500 | cutoff | 41 | 9 | 42 | 8 | 40 | 10 | 123 | 27 | 82.0% | 18.0% | |
| AMP500 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| AMP1000 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| AMP1000 | cutoff | 42 | 8 | 43 | 7 | 43 | 7 | 128 | 22 | 85.3% | 14.7% | |
| AMP1000 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| BUP10 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| BUP10 | cutoff | 43 | 7 | 41 | 9 | 42 | 8 | 126 | 24 | 84.0% | 16.0% | |
| BUP10 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| Negative | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| BAR300 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| BAR300 | cutoff | 42 | 8 | 41 | 9 | 41 | 9 | 124 | 26 | 82.7% | 17.3% | |
| BAR300 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| BZO300 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| BZO300 | cutoff | 43 | 7 | 41 | 9 | 42 | 8 | 126 | 24 | 84.0% | 16.0% | |
| BZO300 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| COC300 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | ||
| COC300 | cutoff | 43 | 7 | 42 | 8 | 41 | 9 | 126 | 24 | 84.0% | 16.0% | |
| COC300 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 0.0% | 100.0% |
| Table 2 Precision / Reproducibility results of Pocquide™ Multi-Drug Test Panel | |||
|---|---|---|---|
| -- | -- | -- | -------------------------------------------------------------------------------- |
{10}------------------------------------------------
Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614
| Drug tested | Lot 1Lot 2Lot 3 | Total Result | %Positive | %Negative | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Claimed Cutoff | %cutoff | + | - | + | - | + | - | + | |||
| +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| COC | cutoff | 41 | 9 | 41 | 9 | 42 | 8 | 124 | 26 | 82.7% | 17.3% |
| 150 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| EDDP | cutoff | 42 | 8 | 43 | 7 | 43 | 7 | 128 | 22 | 85.3% | 14.7% |
| 300 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% |
| 50 | 50 | 0 | 50 | 0 | |||||||
| -50% cutoff | 0 | 0 | 150 | 0.0% | 100.0% | ||||||
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| MET | +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% |
| 1000 | cutoff | 40 | 10 | 41 | 9 | 41 | 9 | 122 | 28 | 81.3% | 18.7% |
| -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| MET | cutoff | 43 | 7 | 41 | 9 | 42 | 8 | 126 | 24 | 84.0% | 16.0% |
| 500 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| Negative | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| MDMA | cutoff | 41 | 9 | 43 | 7 | 42 | 8 | 126 | 24 | 84.0% | 16.0% |
| 500 | -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | |||
| 100.0% | 0.0% | ||||||||||
| OPI | +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% |
| 2000 | cutoff | 41 | 9 | 43 | 7 | 42 | 8 | 126 | 24 | 84.0% | 16.0% |
| -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| MOP300 | +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| cutoff | 42 | 8 | 42 | 8 | 40 | 10 | 124 | 26 | 82.7% | 17.3% | |
| -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| Drug tested | %cutoff | Lot 1 | Lot 2 | Lot 3 | Total Result | %Positive | %Negative | ||||
| Claimed Cutoff | + | - | + | - | + | - | + | - | |||
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| MTD300 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| cutoff | 42 | 8 | 42 | 8 | 43 | 7 | 127 | 23 | 84.7% | 15.3% | |
| -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| OXY100 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| cutoff | 43 | 7 | 44 | 6 | 43 | 7 | 130 | 20 | 86.7% | 13.3% | |
| -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| PCP25 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| cutoff | 42 | 8 | 43 | 7 | 42 | 8 | 127 | 23 | 84.7% | 15.3% | |
| -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| TCA1000 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| cutoff | 43 | 7 | 41 | 9 | 42 | 8 | 126 | 24 | 84.0% | 16.0% | |
| -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -100% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| THC50 | +100% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% |
| +75% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +50% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| +25% cutoff | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0 | 100.0% | 0.0% | |
| cutoff | 42 | 8 | 43 | 7 | 43 | 7 | 128 | 22 | 85.3% | 14.7% | |
| -25% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -50% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% | |
| -75% cutoff | 0 | 50 | 0 | 50 | 0 | 50 | 0 | 150 | 0.0% | 100.0% |
{11}------------------------------------------------
Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614
Linearity/assay reportable range b
Linearity is not applicable since this is a qualitative test.
Analytical specificity/Interference C
c.1 Cross-Reactivity
To test cross-reactivity, drug netabolites and other compounds that are likely to cross-react in urine samples were spiked into drug-free unine sample and were tested using three lots of each device. The lowed a positive result for each compound is listed below along with the corresponding % crossreactivity. The highest concentration tested is shown if no cross reactivity was observed.
| Drug (Cutoff) | Substances | Minimum concentration requiredto obtain a positive result (ng/mL) | % Cross-Reactivity |
|---|---|---|---|
| AMP 500 | d-Amphetamine | 500 | 100% |
| l-Amphetamine | 15000 | 3.30% | |
| dl- Amphetamine | 1500 | 33.30% | |
| (+/-) 3,4-Methylenedioxyamphetamine (MDA) | 5000 | 10% |
Table 3 Cross-Reactivity results of Pocguide™ Multi-Drug Test Panel
{12}------------------------------------------------
| Drug (Cutoff) | Substances | Minimum concentration requiredto obtain a positive result (ng/mL) | % Cross-Reactivity |
|---|---|---|---|
| Phentermine | 1500 | 33.30% | |
| Hydroxyamphetamine | 8000 | 6.25% | |
| d-Methamphetamine | >100000 | -- | |
| l-Methamphetamine | >100000 | -- | |
| (+/-) 3,4-Methylenedioxyethylamphetamine (MDE) | >100000 | -- | |
| (+/-) 3,4-Methylenedioxymethamphetamine (MDMA) | >100000 | -- | |
| Ephedrine | >100000 | -- | |
| β-Phenylethylamine | 100000 | 0.50% | |
| Tyramine | 100000 | 0.50% | |
| p-Hydroxynorephedrine | 100000 | 0.50% | |
| Phenylpropanolamine | >100000 | -- | |
| (±) Phenylpropanolamine | >100000 | -- | |
| p-Hydroxyamphetamine | 100000 | 0.50% | |
| d/l-Norephedrine | 100000 | 0.50% | |
| Benzphetamine | >100000 | -- | |
| l-Ephedrine | >100000 | -- | |
| l-Epinephrine | >100000 | -- | |
| d/l-Epinephrine | >100000 | -- | |
| d-Amphetamine | 1000 | 100% | |
| AMP 1000 | l-Amphetamine | 30000 | 3.30% |
| dl- Amphetamine | 3000 | 33.30% | |
| (+/-)3,4-Methylenedioxyamphetamine (MDA) | 5000 | 20% | |
| Phentermine | 3000 | 33.30% | |
| Hydroxyamphetamine | 8000 | 12.50% | |
| d-Methamphetamine | > 100000 | -- | |
| l-Methamphetamine | > 100000 | -- | |
| (+/-)3,4-Methylenedioxyethylamphetamine (MDEA) | > 100000 | -- | |
| (+/-)3,4-Methylenedioxymethamphetamine (MDMA) | > 100000 | -- | |
| Ephedrine | >100000 | -- | |
| β-Phenylethylamine | 100000 | 1% | |
| Tyramine | 100000 | 1% | |
| p-Hydroxynorephedrine | 100000 | 1% | |
| Phenylpropanolamine | > 100000 | -- | |
| (±) Phenylpropanolamine | > 100000 | -- | |
| p-Hydroxyamphetamine | 100000 | 1% | |
| d/l-Norephedrine | 100000 | 1% | |
| Benzphetamine | > 100000 | -- | |
| l-Ephedrine | > 100000 | -- | |
| l-Epinephrine | > 100000 | -- | |
| d/l-Epinephrine | > 100000 | -- | |
| BUP 10 | Buprenorphine | 10 | |
| Buprenorphine -3-D-Glucuronide | 15 | 66.67% | |
| Norbuprenorphine | 20 | 50% | |
| Norbuprenorphine-3-D-Glucuronide | 100 | 10% | |
| Morphine | > 100000 | - | |
| Oxymorphone | > 100000 | - | |
| Hydromorphone | > 100000 | - | |
| Secobarbital | 300 | 100% | |
| Amobarbital | 5000 | 6% | |
| BAR 300 | Alphenol | 150 | 200% |
| Aprobarbital | 200 | 150% | |
| Butabarbital | 150 | 200% | |
| Butethal | 50 | 600% | |
| Butalbital | 2500 | 12% | |
| Cyclopentobarbital | 600 | 50% | |
| Pentobarbital | 2000 | 15% | |
| Phenobarbital | 5000 | 6% | |
| BZO 300 | Oxazepam | 300 | 100% |
| Alprazolam | 200 | 150% | |
| a-Hydroxyalprazolam | 1000 | 30% | |
| Bromazepam | 500 | 60% | |
| Chlordiazepoxide | 1500 | 20% |
Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614
{13}------------------------------------------------
| Drug (Cutoff) | Substances | Minimum concentration required to obtain a positive result (ng/mL) | % Cross-Reactivity | ||||
|---|---|---|---|---|---|---|---|
| Clobazam | 100 | 300% | |||||
| Clonazepam | 3000 | 10% | |||||
| Clorazepate dipotassium | 200 | 150% | |||||
| Delorazepam | 1500 | 20% | |||||
| Desalkylflurazepam | 400 | 75% | |||||
| Diazepam | 200 | 150% | |||||
| Estazolam | 500 | 60% | |||||
| Flunitrazepam | 1000 | 30% | |||||
| Midazolam | 5000 | 6% | |||||
| Nitrazepam | 1000 | 30% | |||||
| Norchlordiazepoxide | 200 | 150% | |||||
| Nordiazepam | 500 | 60% | |||||
| Temazepam | 250 | 120% | |||||
| Triazolam | 1200 | 25% | |||||
| Demoxepam | 2000 | 15% | |||||
| Flurazepam | 500 | 60% | |||||
| D, L-Lorazepam | 1500 | 20% | |||||
| COC 300 | Benzoylecgonine | 300 | 100% | ||||
| Cocaine | 1000 | 30% | |||||
| Cocaethylene | 15000 | 2% | |||||
| Ecgonine | 30000 | 1% | |||||
| Ecgonine methyl ester | >100000 | -- | |||||
| Norcocaine | >100000 | -- | |||||
| COC 150 | Benzoylecgonine | 150 | 100% | ||||
| Cocaine | 500 | 30% | |||||
| Cocaethylene | 6250 | 2.40% | |||||
| Ecgonine | 16000 | 0.94% | |||||
| Ecgonine methyl ester | > 100000 | -- | |||||
| Norcocaine | > 100000 | -- | |||||
| EDDP 300 | 2-ethylidene-1,5-dimethyl-3,3- diphenylpyrrolidine | 300 | 100% | ||||
| Methadone | 200000 | 0.15% | |||||
| EMDP | 300000 | 0.10% | |||||
| Doxylamine | > 100000 | -- | |||||
| Disopyramide | > 100000 | -- | |||||
| LAAM (Levo-alpha-acetylmethadol) HCl | > 100000 | -- | |||||
| Alpha Methadol | > 100000 | -- | |||||
| D (+)-Methamphetamine | 1000 | 100% | |||||
| (+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA) | 2000 | 50% | |||||
| D/L-Methamphetamine | 1000 | 100% | |||||
| p-Hydroxymethamphetamine | 30000 | 3.30% | |||||
| D-Amphetamine | > 100000 | -- | |||||
| L-Amphetamine | 100000 | 1% | |||||
| Chloroquine | 20000 | 5% | |||||
| (+/-) -Ephedrine | 50000 | 2% | |||||
| MET 1000 | (-) -Methamphetamine | 25000 | 4% | ||||
| (+/-)3,4-Methylenedioxyamphetamine (MDA) | 2500 | 40% | |||||
| (+/-)3,4-Methylenedioxymethamphetamine (MDMA) | 4000 | 25% | |||||
| ẞ-Phenylethylamine | 50000 | 2% | |||||
| Trimethobenzamide | 10000 | 10% | |||||
| d,l-Amphetamine | 100000 | 1% | |||||
| Mephetermine | 50000 | 2% | |||||
| (1R,2S)-(-)-Ephedrine | > 100000 | -- | |||||
| 1-phenylephrine | > 100000 | -- | |||||
| L-Methamphetamine | 20000 | 5% | |||||
| MET 500 | D (+) -Methamphetamine | 500 | 100% | ||||
| (+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA) | 2000 | 25% | |||||
| D/L-Methamphetamine | 500 | 100% | |||||
| MET 500 | p-Hydroxymethamphetamine | (+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA) | 15000 | 2000 | 3.33% | 25% | |
| D/L-Methamphetamine | 500 | 100% | |||||
| p-Hydroxymethamphetamine | 15000 | 3.33% | |||||
| D-Amphetamine | 50000 | 1% | |||||
| L-Amphetamine | 100000 | 0.50% | |||||
| Chloroquine | 10000 | 5% | |||||
| (+/-)-Ephedrine | 25000 | 2% |
Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614
{14}------------------------------------------------
| Drug (Cutoff) | Substances | Minimum concentration requiredto obtain a positive result (ng/mL) | % Cross-Reactivity |
|---|---|---|---|
| (-)-Methamphetamine | 12500 | 4% | |
| (+/-)3,4-Methylenedioxyamphetamine (MDA) | 2000 | 25% | |
| (+/-)3,4-Methylenedioxymethamphetamine (MDMA) | 2000 | 25% | |
| β-Phenylethylamine | 25000 | 2% | |
| Trimethobenzamide | 5000 | 10% | |
| d,l-Amphetamine | 75000 | 0.67% | |
| Mephetermine | 25000 | 2% | |
| (1R,2S)- (-)-Ephedrine | 50000 | 1% | |
| l-phenylephrine | 100000 | 0.50% | |
| L-Methamphetamine | 10000 | 5% | |
| 3,4-Methylenedioxymethamphetamine (MDMA) | 500 | 100% | |
| 3,4-Methylenedioxyamphetamine HCl (MDA) | 3000 | 16.67% | |
| 3,4-Methylenedioxyethylamphetamine (MDEA) | 500 | 100% | |
| MDMA 500 | d-methamphetamine | > 100000 | -- |
| d-amphetamine | > 100000 | -- | |
| l-methamphetamine | 50000 | 1% | |
| l-amphetamine | > 100000 | -- | |
| Morphine | 2000 | 100% | |
| Normorphine | 50000 | 4% | |
| Codeine | 2000 | 100% | |
| s-Monoacetylmorphine | 2000 | 100% | |
| Ethyl Morphine | 1500 | 133.30% | |
| OPI 2000 | Heroin | 2000 | 100% |
| Hydrocodone | 12500 | 16% | |
| Hydromorphone | 3500 | 57.10% | |
| Morphinie-3-β-d-glucuronide | 2000 | 100% | |
| Oxycodone | 25000 | 8% | |
| Oxymorphone | 25000 | 8% | |
| Thebaine | 5000 | 40% | |
| Levorphanol | 75000 | 2.70% | |
| 6-Monoacetylmorphine (6-MAM) | 1500 | 133.30% | |
| Norcodeine | 12500 | 16% | |
| Procaine | 150000 | 1.30% | |
| MOP 300 | Morphine | 300 | 100% |
| Normorphine | 250 | 120% | |
| Codeine | 300 | 100% | |
| s-Monoacetylmorphine | 300 | 100% | |
| Ethyl Morphine | 100 | 300% | |
| Heroin | 300 | 100% | |
| Hydrocodone | 5000 | 6% | |
| Hydromorphone | 2000 | 15% | |
| Morphinie-3-β-d-glucuronide | 1000 | 30% | |
| Oxycodone | > 100000 | -- | |
| Oxymorphone | 100000 | 0.30% | |
| Thebaine | 3000 | 10% | |
| Levorphanol | 5000 | 6% | |
| 6-Monoacetylmorphine (6-MAM) | 150 | 200% | |
| Norcodeine | 6500 | 4.60% | |
| Procaine | 100000 | 0.30% | |
| MTD 300 | Methadone | 300 | 100% |
| Doxylamine | 3000 | 10% | |
| EDDP | > 100000 | -- | |
| EMDP | > 100000 | -- | |
| LAAM | > 100000 | -- | |
| Alpha Methadol | > 100000 | -- | |
| OXY 100 | Oxycodone | 100 | 100% |
| Dihydrocodeine | 20000 | 0.50% | |
| Hydrocodone | 10000 | 1% | |
| Oxymorphone | 1000 | 10% | |
| Codeine | 100000 | 0.10% | |
| Hydromorphone | 32000 | 0.30% | |
| Morphine | >100000 | -- |
Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614
{15}------------------------------------------------
| Drug (Cutoff) | Substances | Minimum concentration requiredto obtain a positive result (ng/mL) | % Cross-Reactivity |
|---|---|---|---|
| Acetylmorphine | >100000 | -- | |
| Buprenorphine | >100000 | -- | |
| Ethylmorphine | >100000 | -- | |
| Thebaine | >100000 | -- | |
| PCP 25 | Phencyclidine | 25 | 100% |
| 4-Hydroxyphencyclidine | 12500 | 0.20% | |
| Nortriptyline | 1000 | 100% | |
| Nordoxepine | 1000 | 100% | |
| Trimipramine | 2000 | 50% | |
| Amitriptyline | 1500 | 66.70% | |
| Promazine | 1500 | 66.70% | |
| Desipramine | 400 | 250% | |
| TCA 1000 | Imipramine | 400 | 250% |
| Clomipramine | 12500 | 8% | |
| Doxepin | 1000 | 100% | |
| Maprotiline | 2000 | 50% | |
| Promethazine | 25000 | 4% | |
| Cyclobenzaprine | 1500 | 66.70% | |
| Norclomipramine | 10000 | 10% | |
| THC 50 | 11-nor-Δ9-THC-9-COOH | 50 | 100% |
| (-)-11-nor-9-carboxy-Δ9-THC | 50 | 100% | |
| 11-nor-Δ8-THC-9-COOH | 30 | 166.67% | |
| 11-nor-Δ9-THC-carboxy glucuronide | 100 | 50% | |
| Cannabinol | 20000 | 0.25% | |
| Cannabidiol | 100000 | 0.05% | |
| Δ8- Tetrahydrocannabinol | 1300 | 3.80% | |
| Δ9- Tetrahydrocannabinol | 5000 | 1% | |
| 511-hydroxy-Δ9-Tetrahydrocannabinol | 5000 | 1% |
Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614
c.2 Interference Study
Potential interfering substances found in human urine of physiological conditions were added to drug-free urine and urine samples with concentrations of the target drugs at 25% below and 25% above Cutoff levels. These urine three lots of each device. The compounds were tested at 100 µg/ml (unless otherwise noted in the table). No interference was observed for anyounds tested, and the list of compounds is provided in the following table.
Table 4 Interfering substances
| Acetaminophen | Ciprofloxacin Hydrochloride | 1% Ethanol | Lidocaine Hydrochloride | Ondansetran |
|---|---|---|---|---|
| Acetophenetidin | Citalopram | Fenofibrate | Lisinopril | Paliperidone |
| Acetylsalicylic Acid | Clarithromycin | Fenoprofen | Lithium Carbonate | Pantoprazole |
| Acyclovir | Clonidine | Fentanyl Citrate | Liverite | Papaverine |
| Afrin | Clopidogrel Hydrogen Sulphate | Fluoxetine Hydrochloride | Loperamide | Paroxetine Hydrochloride |
| Albumin(100mg/dL) | Clozapine | Fluvoxamine | Loratadine | Penfluridol |
| Aminophylline | Conjugated Estrogens | Furosemide | Magnesium | PenicillinV Potassium |
| Aminopyrine | Cortisone | Gabapentin | Meperidine | Penicillin-G |
| Amiodarone Hydrochloride | Creatinine | Gentisic Acid | Meprobamate | Phenelzine |
| Amlodipine Mesylate | (-) Cotinine | Glibenclamide | Metoprolol Tartrate | Pioglitazone Hydrochloride |
| Amoxicillin | chlorpheniramine | Gliclazide | Mifepristone | Piracetam |
| Ampicillin | D, L-Octopamine | Glipizide | Mirtazapine | Pravastatin Sodium |
| Apomorphine | D, L-Propranolol | Glucose | Montelukast Sodium | Prednisone |
| Aripiprazole | D, L-Tyrosine | Haloperidol | Mosapride Citrate | Propylthiouracil |
| Aspartame | Deoxy- corticosterone | Hemoglobin | Minocycline | Quetiapine Fumarate |
| Atomoxetine | Dextromethorphan | Hydrochlorothiazide | Nalidixic Acid | Quinine |
| Atorvastatin Calcium | Diclofenac | Hydrocortisone | Naproxen | Ranitidine |
| Atropine | Diflunisal | 3-Hydroxytyramine | Niacinamide | Rifampicin |
| Benzilic Acid | Digoxin | Isosorbide Dinitrate | Nifedipine | Risperidone |
| Benzoic Acid | Diphenhydramine | Isoxsuprine | Nikethamide | Salicylic Acid |
| Bilirubin | Dirithromycin | Ibuprofen | Nimodipine | Serotonin |
| Bupropion | Domperidone | Ketoconazole | Nitroglycerin | Sertraline Hydrochloride |
| Captopril | D-Pseudoephedrine | Ketoprofen | Norethindrone | Sildenafil Citrate |
| Carbamazepine | Duloxetine | Ketamine | N-Acetylprocain-amide | Simvastatin |
| Cefradine | Dicyclomine | Kratom powder | 0-Hydroxyhippuric Acid | Sodium Valproate |
| Cephalexin | Effexor | Labetalol | Olanzapine | Spironolactone |
| Chloral Hydrate | Enalapril Maleate | Lamotrigine | Omeprazole | Sulfamethazine |
| Chloramphenicol | Erythromycin | Levofloxacin Hydrochloride | Oxalic Acid | Sulindac |
{16}------------------------------------------------
| Chlorothiazide | Esomeprazole Magnesium | Levonorgestrel | Oxolinic Acid | Tetracycline |
|---|---|---|---|---|
| Cholesterol | ß-Estradiol | Levothyroxine Sodium | Oxymetazoline | Tetrahydrocortisone 3 -acetate |
| Tetrahydrocortisone 3-(ß-D glucuronide) | Thioridazine | Trazodone Hydrochloride | Trimethoprim | Verapamil |
| Tetrahydrozoline | Topiramate | Triamterene | Uric Acid | Vitamin B2 |
| Thiamine | Tramadol Hydrochloride | Trifluoperazine | Valproate | Vitamin C |
| Chloroquine | Ecgonine Methyl Ester | Doxylamine | Epinephrine Hydrochloride | Maprotiline |
| Promethazine |
c.3 Effect of urine specific gravity and pH
To investigate the effect of urine specific gravity and urine samples, with specific gravity ranging from 1.000 to 1.035 or urine samples with pH ranging from 4 to 9 were spiked with the target drugs to concentrations at -25% Cutoff levels. These samples were tested using three lots of each device. Results vere all postive for samples at +25% Cutoff and all negative for samples demonstrated that urine specific gravity levels of 1000 to 1.035 and urine pH levels of 4 to 9 do not affect the results of the assays.
Traceability, Stability, Expected values (controls, calibrators, or methods): d
d.1 Traceability
The device is traceable to commercially available materials.
d.2 Stability
The stability data supports that the products have the shelf life of 24 months when stored at 4-30°C.
Real time Stability
A 27-month real time stability test is planned to vertify of the device as 24 morths. Three batches for each configuration in sealed foil pouch are currently stable for 21 months at 4°C and 30°C, and the real time stability study is still on-going.
Accelerated Stability
Stable at 4-30°C for 24 months based on the accelerated stability study at 55°C for 31 days and real time stability determination at both 4°C and 30°C.
e Detection limit
Not Applicable.
Assay Cut-off f
Refer to section 9.1.a.
9.2 Comparison Studies
Method comparison studies a
Method comparison studies were performed by the 100 unaltered clinical samples (40 negative and 40 positive) were run for each target drug. The samples were tested in a blinded fashion and compared to LC/MS results are presented in Table 5 below, and the discordant results are summarized in below.
| Drugcutoff | Candidate Device Result | Drug-Free | Low Negative by LC/MS(less than - 50%) | Near Cutoff Negative by LC/MS(Between -50% and cutoff) | Near Cutoff Positive by LC/MS(Between cutoff and +50%) | High Positive by LC/MS(greater than +50%) | |
|---|---|---|---|---|---|---|---|
| AMP 1000 | Viewer A | Positive | 0 | 0 | 1 | 14 | 25 |
| Negative | 7 | 15 | 17 | 1 | 0 | ||
| Viewer B | Positive | 0 | 0 | 1 | 15 | 25 | |
| Negative | 7 | 15 | 17 | 0 | 0 | ||
| Viewer C | Positive | 0 | 0 | 1 | 14 | 25 | |
| Negative | 7 | 15 | 17 | 1 | 0 | ||
| AMP 500 | Viewer A | Positive | 0 | 0 | 2 | 15 | 23 |
| Negative | 13 | 7 | 18 | 2 | 0 | ||
| Viewer B | Positive | 0 | 0 | 2 | 16 | 23 | |
| Negative | 13 | 7 | 18 | 1 | 0 | ||
| Viewer C | Positive | 0 | 0 | 1 | 15 | 23 | |
| Negative | 13 | 7 | 19 | 2 | 0 | ||
| BAR 300 | Viewer A | Positive | 0 | 0 | 1 | 19 | 19 |
| Negative | 10 | 15 | 14 | 2 | 0 | ||
| Viewer B | Positive | 0 | 0 | 1 | 19 | 19 | |
| Negative | 10 | 15 | 14 | 2 | 0 | ||
| Viewer C | Positive | 0 | 0 | 1 | 19 | 19 | |
| Negative | 10 | 15 | 14 | 2 | 0 | ||
| BUP 10 | Viewer A | Positive | 0 | 0 | 3 | 20 | 18 |
| Negative | 10 | 15 | 12 | 2 | 0 | ||
| Viewer B | Positive | 0 | 0 | 3 | 20 | 18 | |
| Negative | 10 | 15 | 10 | 2 | 0 |
Table 5 The Results of Method comparison with predicate device
| Viewer C | Positive | 0 | 0 | 2 | 20 | 18 | |
|---|---|---|---|---|---|---|---|
| Negative | 10 | 15 | 13 | 2 | 0 | ||
| BZO 300 | Viewer A | Positive | 0 | 0 | 2 | 18 | 20 |
| Negative | 10 | 15 | 13 | 2 | 0 | ||
| Viewer B | Positive | 0 | 0 | 2 | 18 | 20 | |
| Negative | 10 | 15 | 13 | 2 | 0 | ||
| Viewer C | Positive | 0 | 0 | 3 | 18 | 20 | |
| Negative | 10 | 15 | 12 | 2 | 0 | ||
| COC 300 | Viewer A | Positive | 0 | 0 | 0 | 18 | 21 |
| Negative | 9 | 16 | 15 | 1 | 0 | ||
| Viewer B | Positive | 0 | 0 | 0 | 17 | 21 | |
| Negative | 9 | 16 | 15 | 2 | 0 | ||
| Viewer C | Positive | 0 | 0 | 1 | 19 | 21 | |
| Negative | 9 | 16 | 14 | 0 | 0 | ||
| COC 150 | Viewer A | Positive | 0 | 0 | 0 | 21 | 15 |
| Negative | 10 | 15 | 15 | 4 | 0 | ||
| Viewer B | Positive | 0 | 0 | 1 | 23 | 15 | |
| Negative | 10 | 15 | 14 | 2 | 0 |
12
2
าก
0
1Q
15
Negative
Docitive
10
0
{17}------------------------------------------------
Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614
| Drug cutoff | Candidate Device Result | Drug-Free | Low Negative by LC/MS (less than - 50%) | Near Cutoff Negative by LC/MS (Between -50% and cutoff) | Near Cutoff Positive by LC/MS (Between cutoff and +50%) | High Positive by LC/MS (greater than +50%) | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| EDDP 300 | Viewer A Positive | 0 | 0 | 0 | 21 | 15 | ||||||||
| Viewer A Negative | 10 | 14 | 15 | 4 | 0 | |||||||||
| Viewer A Positive | 0 | 0 | 2 | 19 | 20 | |||||||||
| Viewer A Negative | 10 | 14 | 14 | 1 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 3 | 19 | 20 | |||||||||
| Viewer B Negative | 10 | 14 | 13 | 1 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 2 | 20 | 20 | |||||||||
| Viewer C Negative | 10 | 14 | 14 | 0 | 0 | |||||||||
| MET 1000 | Viewer A Positive | 0 | 0 | 1 | 16 | 22 | ||||||||
| Viewer A Negative | 8 | 13 | 18 | 2 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 2 | 17 | 22 | |||||||||
| Viewer B Negative | 8 | 13 | 17 | 1 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 2 | 17 | 22 | |||||||||
| Viewer C Negative | 8 | 13 | 17 | 1 | 0 | |||||||||
| MET 500 | Viewer A Positive | 0 | 0 | 1 | 21 | 16 | ||||||||
| Viewer A Negative | 8 | 20 | 11 | 3 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 2 | 23 | 16 | |||||||||
| Viewer B Negative | 8 | 20 | 10 | 1 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 0 | 21 | 16 | |||||||||
| Viewer C Negative | 8 | 20 | 12 | 3 | 0 | |||||||||
| OPI 2000 | Viewer A Positive | 0 | 0 | 1 | 22 | 18 | ||||||||
| Viewer A Negative | 10 | 15 | 14 | 0 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 1 | 20 | 18 | |||||||||
| Viewer B Negative | 10 | 15 | 14 | 2 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 2 | 20 | 18 | |||||||||
| Viewer C Negative | 10 | 15 | 13 | 2 | 0 | |||||||||
| MOP 300 | Viewer A Positive | 0 | 0 | 1 | 23 | 15 | ||||||||
| Viewer A Negative | 10 | 14 | 15 | 2 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 2 | 23 | 15 | |||||||||
| Viewer B Negative | 10 | 14 | 14 | 2 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 1 | 23 | 15 | |||||||||
| Viewer C Negative | 10 | 14 | 15 | 2 | 0 | |||||||||
| MTD 300 | Viewer A Positive | 0 | 0 | 1 | 14 | 24 | ||||||||
| Viewer A Negative | 10 | 18 | 11 | 2 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 1 | 15 | 24 | |||||||||
| Viewer B Negative | 10 | 18 | 11 | 1 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 0 | 14 | 24 | |||||||||
| Viewer C Negative | 10 | 18 | 12 | 2 | 0 | |||||||||
| OXY 100 | Viewer A Positive | 0 | 0 | 2 | 23 | 15 | ||||||||
| Viewer A Negative | 10 | 15 | 13 | 2 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 2 | 23 | 15 | |||||||||
| Viewer B Negative | 10 | 15 | 13 | 2 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 1 | 22 | 15 | |||||||||
| Viewer C Negative | 10 | 15 | 14 | 3 | 0 | |||||||||
| PCP 25 | Viewer A Positive | 0 | 0 | 1 | 20 | 18 | ||||||||
| Viewer A Negative | 9 | 16 | 14 | 2 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 1 | 20 | 18 | |||||||||
| Viewer B Negative | 9 | 16 | 14 | 2 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 0 | 20 | 18 | |||||||||
| Viewer C Negative | 9 | 16 | 15 | 2 | 0 | |||||||||
| TCA 1000 | Viewer A Positive | 0 | 0 | 2 | 20 | 18 | ||||||||
| Viewer A Negative | 10 | 15 | 13 | 2 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 1 | 19 | 18 | |||||||||
| Viewer B Negative | 10 | 15 | 14 | 3 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 2 | 20 | 18 | |||||||||
| Viewer C Negative | 10 | 15 | 13 | 2 | 0 | |||||||||
| THC 50 | Viewer A Positive | 0 | 0 | 1 | 18 | 20 | ||||||||
| Viewer A Negative | 10 | 15 | 14 | 2 | 0 | |||||||||
| Viewer B Positive | 0 | 0 | 2 | 19 | 20 | |||||||||
| Viewer B Negative | 10 | 15 | 13 | 1 | 0 | |||||||||
| Viewer C Positive | 0 | 0 | 1 | 17 | 20 | |||||||||
| Viewer C Negative | 10 | 15 | 14 | 3 | 0 | |||||||||
| Viewer A Positive | Viewer A | 0 | Positive | 0 | 0 | 1 | 0 | 20 | 1 | 18 | 18 | 20 | ||
| Negative | 10 | 15 | 14 | 2 | 0 | |||||||||
| THC 50 | Viewer B | Positive | 0 | 0 | 2 | 19 | 20 | |||||||
| Negative | 10 | 15 | 13 | 1 | 0 | |||||||||
| Viewer C | Positive | 0 | 0 | 1 | 17 | 20 | ||||||||
| Negative | 10 | 15 | 14 | 3 | 0 | |||||||||
| Viewer A | Positive | 0 | 0 | 1 | 20 | 18 | ||||||||
| Negative | 10 | 15 | 14 | 2 | 0 |
| MDMA 500 | |||||||
|---|---|---|---|---|---|---|---|
| Viewer B | Positive | 0 | 0 | 1 | 21 | 18 | |
| Negative | 10 | 15 | 14 | 1 | 0 | ||
| Viewer C | Positive | 0 | 0 | 1 | 20 | 18 | |
| Negative | 10 | 15 | 14 | 2 | 0 |
{18}------------------------------------------------
| Table 6 Discordant results | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Drug cutoff | Operator | Sample ID | LC/MS Results(ng/mL) | Test Result | Drug cutoff | Operator | Sample ID | LC/MS Results(ng/mL) | Test Result |
| AMP 1000 | Viewer C | AH621 | 968.6 | Positive | MET 500 | Viewer B | ML217 | 457.3 | Positive |
| Viewer A, B | AH618 | 970.1 | Positive | Viewer A, B | ML177 | 491 | Positive | ||
| Viewer A, C | AH562 | 1101 | Negative | Viewer A, C | ML224 | 506 | Negative | ||
| Viewer B | AL554 | 472.6 | Positive | Viewer B, C | ML216 | 512 | Negative | ||
| Viewer A | AL486 | 475.7 | Positive | Viewer A, C | ML168 | 518 | Negative | ||
| AMP 500 | Viewer A, B, C | AL546 | 489 | Positive | Viewer A | ML233 | 523 | Negative | |
| Viewer A, C | AL491 | 501.2 | Negative | Viewer A, C | OP094 | 1869.3 | Positive | ||
| Viewer B, C | AL551 | 508.9 | Negative | OPI 2000 | Viewer B, C | OP084 | 1983 | Positive | |
| Viewer A | AL483 | 536.2 | Negative | Viewer B, C | OP149 | 2045 | Negative | ||
| Viewer A, B, C | BA436 | 296 | Positive | Viewer B, C | OP087 | 2080 | Negative | ||
| BAR 300 | Viewer A, C | BA445 | 302.1 | Negative | Viewer B | M0033 | 287.5 | Positive | |
| Viewer A, B, C | BA474 | 307.2 | Negative | Viewer A, B, C | M0039 | 298 | Positive | ||
| Viewer B | BA403 | 319.83 | Negative | MOP 300 | Viewer A, B, C | M0028 | 301 | Negative | |
| Viewer A, B | BU001 | 8.36 | Positive | Viewer A, C | MO009 | 301.6 | Negative | ||
| Viewer A, B, C | BU052 | 9.21 | Positive | Viewer B | M0071 | 319.5 | Negative | ||
| BUP 10 | Viewer A, B, C | BU053 | 9.42 | Positive | Viewer C | OX236 | 87 | Positive | |
| Viewer A, B, C | BU007 | 10.2 | Negative | Viewer B | OX228 | 92.3 | Positive | ||
| Viewer A, B, C | BU003 | 10.5 | Negative | Viewer A, B | OX204 | 95 | Positive | ||
| Viewer B, C | BZ100 | 282 | Positive | OXY 100 | Viewer A | OX185 | 96 | Positive | |
| Viewer A, C | BZ112 | 291.2 | Positive | Viewer A, B, C | OX199 | 100 | |||
| Viewer A, B, C | BZ149 | 298 | Positive | Viewer A, C | OX209 | 102 | |||
| BZO 300 | Viewer C | BZ103 | 301 | Negative | Viewer B, C | OX195 | 103 | Negative | |
| Viewer A, B, C | BZ096 | 305 | Negative | Viewer A | PC252 | 23.16 | Positive | ||
| Viewer B | BZ108 | 309 | Negative | Viewer B | PC260 | 23.97 | Positive | ||
| Viewer A | BZ155 | 317.6 | Negative | PCP 25 | Viewer A, B, C | PC286 | 25.39 | Negative | |
| Viewer C | CH945 | 294.8 | Positive | Viewer A, B, C | PC297 | 25.9 | |||
| COC 300 | Viewer A, B | CH948 | 328.5 | Negative | Viewer C | TA752 | 849 | Positive | |
| Viewer B | CH893 | 338.5 | Negative | Viewer A, C | TA776 | 892 | Positive | ||
| Viewer B | CL801 | 145 | Positive | Viewer A, B | TA724 | 934 | Positive | ||
| Viewer B | CL827 | 154 | Negative | TCA 1000 | Viewer A | TA774 | 1025 | Negative | |
| Viewer A, C | CL806 | 156 | Negative | Viewer B, C | TA761 | 1058 | |||
| COC 150 | Viewer A, B, C | CL839 | 159 | Negative | Viewer B, C | TA767 | 1069 | Negative | |
| Viewer A, C | CL830 | 160.1 | Negative | Viewer A, B | TA799 | 1079 | Negative | ||
| Viewer A, C | CL858 | 163.5 | Negative | Viewer B, C | TH334 | 48 | Positive | ||
| Viewer B | ED366 | 280.8 | Positive | Viewer A, B | TH375 | 48.8 | Positive | ||
| Viewer A, B, C | ED350 | 281 | Positive | THC 50 | Viewer A, C | TH374 | 50.35 | Negative | |
| EDDP 300 | Viewer A, B, C | ED335 | 284.1 | Positive | Viewer A, C | TH339 | 53.46 | ||
| Viewer A | ED331 | 325 | Negative | Viewer B, C | TH398 | 54.45 | Negative | ||
| Viewer B | ED358 | 335.7 | Negative | Viewer B | MD642 | 476.1 | Positive | ||
| Viewer A, B, C | MH307 | 981.1 | Positive | MDMA 500 | Viewer A, C | MD708 | 482.34 | Positive | |
| MET 1000 | Viewer B, C | MH291 | 995.8 | Positive | Viewer A, C | MD706 | 502 | ||
| Viewer A, B | MH246 | 1039.7 | Negative | Viewer B, C | MD681 | 527.89 | Negative | ||
| Viewer A, C | MH302 | 1098 | Negative | Viewer A | MD710 | 532 | Negative | ||
| Viewer A, B | MT464 | 278.8 | Positive | ||||||
| MTD 300 | Viewer A, B, C | MT424 | 326.5 | Negative | |||||
| Viewer A, C | MT475 | 329.8 | Negative |
b Matrix comparison
Not applicable. This device is for testing with human urine only.
10 Clinical studies
Clinical Sensitivity a
Not applicable
Clinical specificity b
Not applicable
Other clinical supportive data (when a. and b. are not applicable) C
c.1 Lay-User Study
The lay-user study was performed at three intensed by and 82 female tested Porguite™ Multi-Drug Test Panel Configuration 1 (including AMP 500, MET 500, NET 500, NET 500, NET MOP 300, COC 150); 56 male and 84 female tested Poguide™ Multi-Drug Test Panel Configuration 2 (including AMP 1000, MET 1000). Ney had
direrse educational and occupational backgrounds and the > 50. Urine samples were prepared at the following concentrations: -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug free-pooled urine specimens. The concentrations of the sample were confirmed by LC-MS/MS Each sample was aliquoted into individual containers and blind-label. Each participant was provided with the package insert, 1 blind label sample and a device. The results are summarized in below.
| Drug | Cutoff(ng/mL) | Results | Concentration | ||||||
|---|---|---|---|---|---|---|---|---|---|
| AMP | 500 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| AMP | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |
| AMP | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |
| AMP | Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | |
| BAR | 300 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| BAR | Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 |
Table 7 Result of Pocquide™ Multi-Drug Test Panel Configuration 1 (including AMP 500, MET 500, MOP 300, COC 150)
{19}------------------------------------------------
| Cutoff | Concentration | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Drug | (ng/mL) | Results | -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff | |
| BZO | 300 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% | |||
| BUP | 10 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | |||
| COC | 150 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% | |||
| EDDP | 300 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% | |||
| MDMA | 500 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 100% | 95.00% | 100% | 100% | |||
| MET | 500 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | |||
| MOP | 300 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% | |||
| MTD | 300 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 100% | 95.00% | 100% | 100% | |||
| OXY | 100 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | |||
| PCP | 25 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% | |||
| TCA | 1000 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% | |||
| THC | 50 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% |
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% |
|---|---|---|---|---|---|---|---|
| --------------- | ------ | ------ | ------ | -------- | ------ | ------ | ------ |
| Drug | Cutoff(ng/mL) | Results | Concentration | ||||||
|---|---|---|---|---|---|---|---|---|---|
| -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff | |||
| AMP | 1000 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100.00% | 100% | 100% | ||
| BAR | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | ||
| Drug | Cutoff(ng/mL) | Results | Concentration | ||||||
| -100% cutoff | -75% cutoff | -50% cutoff | -25% cutoff | +25% cutoff | +50% cutoff | +75% cutoff | |||
| BZO | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | ||
| BUP | 10 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100.00% | 95.00% | 100% | 100% | ||
| COC | 300 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100.00% | 95.00% | 100% | 100% | ||
| EDDP | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | ||
| MDMA | 500 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95.00% | 100% | 100% | ||
| MET | 1000 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100.00% | 95.00% | 100% | 100% | ||
| OPI | 2000 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | ||
| MTD | 300 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95.00% | 100% | 100% | ||
| OXY | 100 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100.00% | 95.00% | 100% | 100% | ||
| PCP | 25 | Negative | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100.00% | 95.00% | 100% | 100% | ||
| TCA | 1000 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 95.00% | 100% | 100% | ||
| THC | 50 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.00% | 100% | 100% | 100% |
Table 8 Result of Pocguide™ Multi-Drug Test Panel Configuration 2 (including AMP 1000, MET 1000, MOP 2000 (OPI), COC 300)
{20}------------------------------------------------
Participants were given surveys on the ease of under on to use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.
Clinical Cut-off d
Not applicable
11 Conclusion
Based on the test principle and performance charactive including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that Pocguide™ Multi-Drug Test Panel OTC are substantially equivalent to the predicate devices. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
§ 862.3100 Amphetamine test system.
(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).