K Number
K242488
Device Name
Soteria E-View
Date Cleared
2025-01-06

(138 days)

Product Code
Regulation Number
892.1650
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The System is intended for use in Radiographic / Fluoroscopic applications including general radiographic / fluoroscopic diagnostic, and interventional X-Ray imaging for General and Pediatric Populations.

Device Description

The Soteria E-View system is classified as an interventional fluoroscopic X-ray system. The fundamental performance characteristics of the Soteria E-View interventional fluoroscopic X-ray system consists of:

  • The patient table and C- arm with X-ray source on one side and the flat panel detector on the opposite side. The C-arm can be angulated in both planes, and the flat panel detector can be lifted vertically. The tabletop can be shifted from side to side and move forward/backward by an operator.
  • Real-time image visualization of patient anatomy during procedures.
  • Imaging techniques and tools to assist interventional procedures.
  • . Post-processing functions after interventional procedures.
  • Storage of reference/control images for patient records.
  • Compatibility to images of other modalities via DICOM.
  • . Built-in radiation safety controls-with the already FDA cleared CA-100S /FluoroShield (K182834).
    This array of functions provides the physician with the imaging information required to achieve minimally invasive interventional procedures.
    The Soteria E-View system is available as a Model GI-100 configuration. It is similar to the currently marketed predicate Soteria.Al consisting of an X-ray generator, image processor, collimator, X-ray Tube, Positioner, and patient table with CA-100S / FluoroShield Accessory, (K212336).
AI/ML Overview

This document describes the Omega Medical Imaging, LLC Soteria E-View (K242488) system, which is a fluoroscopic/radiographic X-ray system. The submission aims to demonstrate substantial equivalence to its predicate device, the Soteria.AI (K212336).

The provided text details the device's technical specifications and comparisons to the predicate, as well as adherence to various safety standards and guidance documents. However, it does not contain information regarding specific acceptance criteria, corresponding device performance data, or detailed study results (including sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone performance studies).

The document primarily focuses on demonstrating substantial equivalence through:

  • Indications for Use: The Soteria E-View has similar indications, but with cardiac and vascular applications removed and gastrointestinal/ERCP applications added.
  • Technological Characteristics: Changes include an updated flat panel detector (Varex Azure 3131Z CXP Detector with IGZO technology) with improved specifications like resolution and MTF, a different X-ray generator, positioner configuration, power supplies, control buttons, actuators, table size, and power unit.
  • Non-Clinical Performance: States that non-clinical and "Sample clinical images" were used to validate image performance and demonstrate conformance to intended use, claims, user, and service needs. It also mentions software verification testing for functional requirements, performance, reliability, and safety.
  • Safety and Effectiveness: Asserts that the differences do not raise new safety or effectiveness concerns, and the device complies with relevant 21 CFR regulations and international safety standards.

Therefore, many of the requested details about acceptance criteria and the study proving adherence cannot be extracted from this document.

Based on the provided text, here is what can be inferred or explicitly stated:


1. Table of Acceptance Criteria and Reported Device Performance

Specific, quantifiable "acceptance criteria" and direct "reported device performance" against those criteria are not explicitly laid out in a table format in the provided text. The document focuses on demonstrating substantial equivalence to a predicate device, meaning the new device performs at least as well as the predicate and does not raise new safety or effectiveness concerns.

However, some performance characteristics are compared to the predicate, implying they meet or exceed the predicate's performance, which can be seen as an indirect form of demonstrating acceptance.

Feature / CharacteristicAcceptance Criteria (Implied by Predicate performance or general standards)Reported Device Performance (Soteria E-View)
Flat Panel Detector
Detector NameTeledyne Dalsa Xineos-3030HS (Predicate)Varex Azure 3131Z CXP Detector
Receptor TypeCMOS (Predicate)IGZO (indium gallium zinc oxide)
Conversion ScreenCesium Iodide (Predicate & Device)Cesium Iodide
Pixel Area - Active29.6 cm x 29.6 cm (Predicate)30.7 cm x 30.7 cm
Pixel Matrix - Active1952 x 1952 (Predicate)2048 x 2048
Pixel Pitch151.8 μm (Predicate)150 μm
Limiting Resolution3.15 lp/mm (Predicate)3.3 lp/mm (Improved)
MTF, X-ray58% @ (1.0 lp/mm at RQA5 1x1) (Predicate)60% @ (1.0 lp/mm) (Improved)
Energy Range40 - 125 kV (Predicate & Device)40-125 kV
Fill Factor83% (Predicate)66% (Difference noted, but not stated as an issue for equivalence)
Dynamic Range77 dB (Predicate)65 dB (Difference noted, but not stated as an issue for equivalence)
A/D Conversion16-bits (Predicate & Device)16-bits
X-Ray Specifications
Half Value LayerMinimum 3.6 mm Al equiv. @ 100 kV; Maximum 4.0 mm Al Equiv. @ 100 kV (Predicate)Minimum 3.6 mm Al equiv. @ 100 kV; Maximum 4.0 mm Al Equiv. @ 100 kV
Tabletop Aluminum EquivalenceRejection Limit: 2.00 mm Al max. (Predicate)Rejection Limit: 1.6 mm Al max. (Improved)
Image Quality: ResolutionUp to 3.15 lp/mm (Predicate)Up to 3.15 lp/mm

Note: The table above extracts direct comparisons from Section "PRODUCT OVERVIEW" within the 510(k) summary. For other aspects like software performance, safety, and functional requirements, the document only states that "Results demonstrated that the executed verification test was passed" without providing specific criteria or data.


2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document states: "Non-clinical and Sample clinical images were used to validate image performance." It also mentions: "Non- clinical and Sample clinical images were used for validation testing of the Soteria E-View system to demonstrate conformance to the intended use, claims, user, and service needs..."
However, no specific sample size for the test set (number of images, number of cases, number of patients) is provided.
The data provenance (country of origin, retrospective/prospective) is not mentioned.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not mention an MRMC study or any AI components for human reader assistance. The device is an X-ray system, and while the predicate device is named "Soteria.AI", the specific AI functionality (automatic region of interest to reduce exposure) is a built-in feature of the system and not framed as an AI-assistance tool for human readers in the context of comparative effectiveness for diagnosis. The focus is on the device's technical performance and safety.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not provided in the document, as the device is a complete X-ray system, not a standalone algorithm. The "FluoroShield / CA-100S device to provide an automated Region of interest that reduces exposure to the patient and operator" is described as an integrated hardware/software component of the system.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

This information is not provided in the document. The text only mentions "Sample clinical images" were used for validation testing, but how their "ground truth" was established is not detailed.


8. The sample size for the training set

The document primarily discusses validation and substantial equivalence, not the development of a specific algorithm requiring a "training set." Therefore, no information on a training set size is provided.


9. How the ground truth for the training set was established

As no training set is described, this information is not applicable/provided.

{0}------------------------------------------------

January 6, 2025

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The seal is a stylized depiction of an eagle and a caduceus, while the FDA acronym and name are written in blue, with the word "Administration" appearing below the acronym and name.

Omega Medical Imaging, LLC % Matthew Anderson Director of RA/OA 3400 St. Johns Parkway, Suite 1020 SANFORD, FL 32771

Re: K242488

Trade/Device Name: Soteria E-View Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: JAA, OWB Dated: August 15, 2024 Received: December 11, 2024

Dear Matthew Anderson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Submission Number (if known)

K242488

Device Name

Soteria E-View

Indications for Use (Describe)

The System is intended for use in Radiographic / Fluoroscopic applications including general radiographic / fluoroscopic diagnostic, and interventional X-Ray imaging for General and Pediatric Populations.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Omega Medical Imaging. The logo consists of a blue stylized omega symbol on the left, followed by the text "OMEGA" in a bold, blue sans-serif font. Below the word "OMEGA" is the text "Medical Imaging" in a smaller, light gray sans-serif font. The overall design is clean and professional, suggesting a company in the medical field.

Traditional 510(k) SUMMARY K242488

Company Name:Omega Medical Imaging, LLC
Address:3400 St. Johns Parkway, Suite 1020, Sanford, FI 32771
Telephone No:407-323-9400
Registration No:1052701
Contact person:Matthew Anderson, Director RA/QA
Date Prepared:06/04/2024
Device (trade) name:Soteria E-View
Common/usual name:Fluoroscopic/Radiographic X-ray system
Classification Name:Image-intensified Fluoroscopic X-ray system
Classification Panel:Radiology
CFR section:892.1650
Device Class:Class II
Primary Product code:JAA
Secondary product code:OWB

Predicate Device K212336

Company Name:Omega Medical Imaging, LLC
Address:3400 St. Johns Parkway, Suite 1020, Sanford, FI 3277
Telephone No:407-323-9400
Registration No:1052701
Contact person:John Newman
Date Prepared:08/19/2021
Device (trade) name:Soteria.AI
Common/usual name:Fluoroscopic/Radiographic X-ray system
Classification Name:Image-intensified Fluoroscopic X-ray system
Classification Panel:Radiology
CFR section:892.1650
Device Class:Class II
Primary Product code:JAA
Secondary Product code:OWB

{5}------------------------------------------------

Indications for use:

The System is intended for use in Radiographic/fluoroscopic applications including general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging for General and Pediatric Populations.

Device Description:

The Soteria E-View system is classified as an interventional fluoroscopic X-ray system. The fundamental performance characteristics of the Soteria E-View interventional fluoroscopic X-ray system consists of:

  • The patient table and C- arm with X-ray source on one side and the flat panel detector on the opposite side. The C-arm can be angulated in both planes, and the flat panel detector can be lifted vertically. The tabletop can be shifted from side to side and move forward/backward by an operator.
  • Real-time image visualization of patient anatomy during procedures.
  • Imaging techniques and tools to assist interventional procedures.
  • . Post-processing functions after interventional procedures.
  • Storage of reference/control images for patient records.
  • Compatibility to images of other modalities via DICOM.
  • . Built-in radiation safety controls-with the already FDA cleared CA-100S /FluoroShield (K182834).

This array of functions provides the physician with the imaging information required to achieve minimally invasive interventional procedures.

The Soteria E-View system is available as a Model GI-100 configuration. It is similar to the currently marketed predicate Soteria.Al consisting of an X-ray generator, image processor, collimator, X-ray Tube, Positioner, and patient table with CA-100S / FluoroShield Accessory, (K212336).

Patient Population:

General and Pediatric Population, special concerns must be taken for pediatric use.

Based on the information provided above, the Soteria E-View system is considered substantially equivalent to the current marketed predicate device Soteria. Al (K212336). Both share Indications for use.

(Note: Cardiac and Vascular application removed from the Soteria E-View indications for use)

{6}------------------------------------------------

Technological characteristics and Summary of Critical Improvements:

The Soteria E-View system has similar technological characteristics compared to the predicate device with some changes to it. Below is a summary of the changes between the new Soteria E-View system and the predicate device. FluoroShield was integrated with the Varex Azure 3131Z CXP Flat Panel Detector.

  • Replace 80KV high voltage generator with GI Epsilon Generator (E-view), X-ray generator is certified . separately.
  • . The positioner changed from L/C to I/C configuration.
  • Updated 12 and 24V Power supplies in electronics cabinet with newer more available UL compliant units.
  • Added fluoro, spot and save buttons to positioner controls.
  • Upgrade actuators to support Linak instead of SKF.
  • . Shorter GI patient table that supports an 800-pound patient, as this is a more suitable table configuration for Gastrointestinal interventional procedures.
  • . The Nyquist image processor is now configured with a Varex 3131x CXP detector, as Indium Gallium Zinc Oxide technology has a high quantum efficiency and lower dose than traditional CMOS detectors.
  • . Replaced Powertronix IsoStation Transformer and Ametek UPS combination with a single Astrodyne TDI UL certified PowerBridge which combines the isolation transformer and UPS into a single unit.

The difference between the Soteria E-View system and the predicate device does not raise any new safety or effectiveness. Based on the information provided in this 510K submission, Soteria E-View is considered substantially equivalent to the current marketed predicate Soteria. Al (K212336)

Clinically:

  • . Legacy E-view (K062647) and Soteria E-view (GI-100) are virtually identical in terms of the Positioner and Table movements and capabilities.
  • Same Tube ●
  • Same Generator
  • Same Table Size

Summary of Non- Clinical Performance

Non-clinical and Sample clinical images were used to validate image performance. This testing has been performed on the Soteria E-View system to demonstrate substantial equivalence to the predicate device.

    1. IEC 62304 Medical device software Software life cycle processes. FDA/CDRH recognition number 13-79.
    1. ISO 14971 Medical devices Application of risk management to medical devices
    1. IEC 60601-2-54 Particular requirements for basic safety and essential performance of X-ray Safety.

{7}------------------------------------------------

    1. IEC60601-2-43 - Particular requirements for the basic safety and essential performance of x-ray equipment for interventional procedures.
    1. Guidance for Industry and FDA Staff Guidance for the Content of premarket Submissions for Device Software function. June 2023
  • "Guidance for Industry and FDA Staff The 510(k) Program: Evaluating Substantial Equivalence in 6. Premarket Notifications [510(k)]", July 28, 2014 (document number 1766).
    1. Guidance for industry and FDA staff Electronic Submission Template for Medical Device 510(k) submissions. Oct. 2023

Software verification testing of the functional requirements, as well as performance, reliability, and safety, have been performed to verify that all conditions of System requirements Specifications, as well as the safety risk control measures from the detailed Risk management and security requirements, have been implemented. Results demonstrated that the executed verification test was passed.

Non- clinical and Sample clinical images were used for validation testing of the Soteria E-View system to demonstrate conformance to the intended use, claims, user, and service needs, effectively satisfying measures and instruction for use.

The Soteria E-View was demonstrated with the following attributes:

  • Indication for use.
  • Technological characteristics.
  • Non-clinical and Sample clinical image performance testing.
  • Safety and effectiveness.

Substantial equivalence Conclusion:

The Soteria E-View system is substantially equivalent to the currently marketed predicate device Soteria.Al regarding indications for use, technological characteristics, safety, and effectiveness.

Therefore, Soteria E-View is as safe and effective as its predicate device and does not raise any new safety and effectiveness concerns.

{8}------------------------------------------------

Comparison with Predicate Devices:

Indications for use comparison:
510(k)Number andDevice NameK212336 (Predicate Device)Soteria.Al, Model Al-100K242488 (This Submission)Soteria E-View, Model GI-100
Intended UseThe Omega Medical Imaging, LLC AI-100,Soteria.AI (SSXI) systems with FluoroShield / CA-100S device to provide an automated Region ofinterest that reduces exposure to the patient andoperator. The System is intended for use inRadiographic/fluoroscopic applications, includingcardiac, vascular, generalradiographic/fluoroscopic diagnostic, andinterventional x-ray imaging for General andPediatric Populations.The Omega Medical Imaging, LLC GI-100,Soteria E-View (SSXI) systems withFluoroShield / CA-100S device to provide anautomated Region of interest that reducesexposure to the patient and operator. TheSystem is intended for use inRadiographic/fluoroscopic applications, includinggastrointestinal, ERCP, generalradiographic/fluoroscopic diagnostic, andinterventional x-ray imaging for General andPediatric Populations.
ClassificationName:Image-intensified Fluoroscopic X-ray systemImage-intensified Fluoroscopic X-ray system
CFRRegulation #:892.1650892.1650
Device ClassClass IIClass II
ClassificationProduct Code:JAA, OWBJAA, OWB

PRODUCT OVERVIEW

Substantial Equivalence:

1. Flat Panel Detectors
510(k) Cleared (PredicateDevice) K212336This Submission Soteria E-View
Detector Name:Teledyne Dalsa Xineos-3030HSVarex Azure 3131Z CXP Detector
1.1 Receptor TypeCMOSIGZO (indium gallium zinc oxide)
1.2 Conversion ScreenCesium IodideCesium Iodide
1.3 Pixel Area - Active29.6 cm x 29.6 cm30.7 cm x 30.7 cm
1.4 Pixel Matrix - Active1952 x 19522048 x 2048
1.5 Pixel Pitch151.8 μm150 μm
1.6 Limiting Resolution3.15 lp/mm3.3 lp/mm
1.7 MTF, X-ray58% @ (1.0 lp/mm at RQA5 1x1)60% @ (1.0 lp/mm)
1.8 Energy Range40 - 125 kV40-125 kV
1.9 Fill Factor83%66%
1.10 Dynamic Range77 dB65 dB
1.11 A/D Conversion16-bits16-bits

{9}------------------------------------------------

2. Acquisition is the same for all Flat Panel Detectors
2.1 ModesPulsed Fluoro, Spot, Cine, Roadmaping (Subtration Fluoroscopy) and DSAPulsed Fluoro and Spot
2.2 Pulsed Fluoro Rates3.75, 7.5, and 14 pps3, 6, and 12.5 pps
2.3 Spot Rates1, 2, 3,4, and 6 fps1, 2, 3,4, and 6 fps
2.4 Cine Rates7.5, 15, and 30 fpsNo Cine
2.5 PolarityPositive & NegativePositive & Negative
2.6 Edge Enhancement5 levels5 levels
2.7 Noise Reduction5 levels5 levels
2.8 Last Image HoldYesYes
2.9 Fluoro LoopsYesYes
2.10 Horizontal &Vertical FlipYesYes
3. Acquisition Workstation
3.1 Workstation ModelNyquistNyquist
3.2 Monitor, size2 x 21.3 inch, Medical GradeColor, High Contrast2 x 21.3 inch, Medical GradeColor, High Contrast
3.3 Grey Levels10-bit, Display Port10-bit, Display Port
3.4 Review (FluoroLoops/Cine)Forward, Reverse, Pause, andPlayForward, Reverse, Pause, andPlay
3.5 Window/Level AdjustYesYes
3.6 AnnotationYesYes
3.7 Image InvertYesYes
3.8 Horizontal & VerticalFlipYesYes
3.9 Zoom & RoamYes (2x)Yes (2x)
3.10 Image StorageMinimum: 225,000 imagesMinimum: 225,000 images
3.11 NetworkingCapabilitiesGigabit ethernet10/100/1000Gigabit ethernet 10/100/1000
3.12 DICOM Worklist
3.13 DICOM StorageYesYes
3.14 DICOM Storage to DVD/CDWith Embedded ViewerWith Embedded Viewer
4. Reference Monitor Identifiers [Review Station]
4.1 Patient NameYesYes
4.2 Patient I.D.YesYes
4.3 DateNoNo
4.4 TimeNoNo
4.5 Window/LevelNoNo
4.6 Hospital NameNoNo
4.7 Physician NameYesYes
4.8 Study I.D.YesYes
4.9 Series/ImageYesYes
4.10 Acquisition ModeYesYes
4.11 Image OrientationYesYes
4.12 RadiationNoNo
14.13 Image I.D.YesYes
5. Live Monitor Identifiers [Image Area]
5.1 Patient NameYesYes
5.2 Patient I.D.YesYes
5.3 DateYesYes
5.4 TimeYesYes
5.5 Window/LevelNoNo
5.6 Hospital NameYesYes
5.7 Physician NameNoNo
5.8 Study I.D.YesYes
5.9 Series/ImageOnly Image NumberOnly Image Number
5.10 Image OrientationYesYes
5.11 Radiation SymbolRadiation symbol during LiveRadiation symbol during Live
5.12 Image I.D.e.g., FL LIH, FL Loops, SpotReplay, or Cine Replaye.g., FL LIH, FL Loops, SpotReplay, or Cine Replay
5.13 Detector StatusReady/Not ReadyReady/Not Ready
5.14 Magnification ModeNormal, MAG 1, MAG 2Normal, MAG 1, MAG 2
5.15 Loop ReplayPlay, Forward, Reverse, Play,& PausePlay, Forward, Reverse, Play, &Pause
GUI [Graphic User Interface on Reference Monitor]
Patient/Exam Entry
5.16 DirectoryOpens 'Patient Directory'screenOpens 'Patient Directory'screen
5.17 NewOpens 'Patient Data Entry'screenOpens 'Patient Data Entry'screen
5.18 CloseCloses active patientCloses active patient
Acquisition
5.19 NR Hi / NR LoSelects Noise Reduction LevelSelects Noise Reduction Level
5.20 L -> RLive to Reference (monitor)transferLive to Reference (monitor)transfer
5.21 Fluoro Pulse Rates:Pulse Rates: [14 f/s default,7.5, & 3.75 f/s]Pulse Rates: [12.5 f/s default,6, & 3 f/s]
Note: Image Orientation available for Acquisition Mode [from Review]
Review
5.22 Edge[enhancement]Increments from 1 - 5Increments from 1 - 5
5.23 Win./Lev.Window/Level adjustmentWindow/Level adjustment
5.24 SplitInvokes 4 on 1 image display,NoInvokes 4 on 1 image display,No
5.25 Reference - LiveToggles image between theReference and Live monitorsToggles image between theReference and Live monitors
5.26 ZoomUp to 2X zoom with panningUp to 2X zoom with panning
5.27 PolarityVideo white on black or blackon whiteVideo white on black or blackon white
5.29 R [imageorientation]Toggles thru 1 of 4 positionsToggles thru 1 of 4 positions
Additional Functions
5.30 SendOpens 'DICOM server nodeselection'Opens 'DICOM server nodeselection'
5.31 PrintSends image to a Windowscompatible printer (No)Sends image to a Windowscompatible printer (No)
5.32 SaveSaves current image to HardDriveSaves current image to HardDrive
5.33TextInvokes the Annotation andMeasurement functionInvokes the Annotation andMeasurement function
C-Arm & Operator Controls
C-ARM Specifications:510(k) Cleared (PredicateDevice) K212336This Submission Soteria E-View.AI
C-AngulationHead $\pm$ 110° $\pm$ 2°,Side $\pm$ 45° $\pm$ 2°,$\pm$ 90°, $\pm$ 2°
C-RollHead $\pm$ 45° $\pm$ 2°,Side $\pm$ 90° $\pm$ 2°,+ 90°, $\pm$ 2°, -75°, $\pm$ 2°
Iso-Rotation$\pm$ 90°, $\pm$ 2°No
Collimator - LongitudeOpens to full field, closes fullyOpens to full field, closes fully
Collimator -LatitudeOpens to full field, closes fullyOpens to full field, closes fully
Cardio Filter -Open CloseOpens to full field, closes fully(vascular), closes to mid field(lung)No Filter
Cardio Filter -RotateRotates cardio filter(s) 360°No Filter
Variable SID13", $\pm$ 1"13", $\pm$ 1"
Operator Controls
Fluoro EnableEnables/disables Fluoro functionEnables/disables Fluoro function
Fluoro Timer ResetResets 5-minute alarm on highvoltage generatorResets 5-minute alarm on highvoltage generator
Emergency StopDisables motion actuators andhigh voltage generatorDisables motion actuators andhigh voltage generator
Magnification modeSelects field sizes: Normal, Mag1, and Mag 2Selects field sizes: Normal, Mag 1,and Mag 2
Controls selectRequests active control is 2controllers are installedRequests active control is 2controllers are installed
5th Axis Rotate± 90°, ± 2°No
Cardio filter selectLung or vascularNo
DAP meter resetResets DAP meter outputs to 0Resets DAP meter outputs to 0
Pediatric Filter SelectSelects none, 0.1, 0.2, or 0.3 mmCu filtersNo
Parking ControlControls parking actuatorNo
FluoroEnables X-ray fluoroEnables X-ray fluoro
Fluoro loopsEnables X-ray fluoro loopsEnables X-ray fluoro loops
CineEnables X-ray cine modeNo
Miscellaneous Functions:
System Digital DisplayMiscellaneous System StatusMiscellaneous System Status
Collision Sensing on FPDBump stop.Bump stop.
Collimator Functions:
Collimator: Maximum Sizeand TrackingYesYes
Collimator: Minimum SizeYesYes
X-Ray Specifications
Half Value LayerMinimum value = 3.6 mm Alequiv. @ 100 kVMaximum value = 4.0 mm AlEquiv. @ 100 kVMinimum value = 3.6 mm Alequiv. @ 100 kVMaximum value = 4.0 mm AlEquiv. @ 100 kV
Tabletop AluminumEquivalenceRejection Limit: 2.00 mm Al max.Rejection Limit: 1.6 mm Al max.
Dose Calibration: MaximumInput Dose Rate80 to 85 mGy/minute80 to 85 mGy/minute
Dose Calibration: Max InputDose Rate2.3 to 2.9 µR/frame2.3 to 2.9 µR/frame
Dose Calibration: Flat PanelInput Dose rate0.8 µGy/sec.0.8 µGy/sec.
Generator Function: PulsedFluoro FunctionYesYes
Generator Function: ABCFunctionYesYes
Image Quality: ResolutionUp to 3.15 lp/mmUp to 3.15 lp/mm
Last Image HoldYesYes
Table Specifications:510(k) Cleared (PredicateDevice) K212336This Submission Soteria E-View
Table Lock Manual DisableControlsUnlocks table rotate, latitudeand longitude.No
Table Trendelenburg± 15°, ± 1°± 12°, ± 1°
Table height33" to 45"Same
Table cradle (LAO/RAO)± 15°, ± 1°± 12°, ± 1°
Table Rotation0° to 90°No
X-RAY TUBE Specifications:
X-RAY TubeSpecifications:510(k) Cleared (PredicateDevice) K212336This Submission Soteria E-View
Tube Model:G-2090G-1092
Heat ExchangerHE-280HE-100
Focal Spot sizes0.6 - 1.0 mm0.6 – 1.2mm
Heat Unit rating2.0 MHU1.0 MHU
Housing Model:B-240HB-160
High Voltage Generator
H.V. GeneratorSpecifications:510(k) Cleared (PredicateDevice) K212336This Submission Soteria E-View
Model:EMD Technologies EPS HighVoltage Generator seriesEMD Technologies EPS HighVoltage Generator series
Mains Input Power:480 Vac +/- 10% 50/60 Hz (3phases)480 Vac +/- 10% 50/60 Hz (3phases)
kV range:40 kV - 125 kV40 kV - 125 kV
mA range:Min: 5mA to Max: 1000mAMin: 5mA to Max: 1000mA
CA-100S / FluoroShield
ROI: 'ON' ,'OFF','ERROR', or'BYPASS'YesYes
'AUTO' or'MANUAL' ROIYesYes
'HIGH' or 'LOW'reductionYesYes

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

Safety information:

  • The Omega Soteria E-View systems comply with the applicable requirements of 21 CFR 1020.30, 21 CFR 1020.31, and 21 CFR 1020.32.
  • The Omega Soteria E-View systems comply with the international safety standards EN 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-4, IEC 60601-2-54, IEC 60601-2-54, IEC 60601-2-43, EN ISO 15223-1, and EN ISO 14971.
  • The Omega Soteria E-View systems comply with UL 60601-1 and CAN/USA C22.2 No.601.1-M90.
  • The device is designed and manufactured under the Quality System Regulations outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with applicable parts of the IEC60601-1 standards and its collateral standards. All Federal Diagnostic Equipment Standard requirements, as outlined in 21 CFR § 1020, that apply to this device will be met and reported in this initial report.
  • . This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard].

Safety is assured through a risk management process, and manufacturing complies with the Quality System Regulations.

{16}------------------------------------------------

Referenced Guidance Documents:

  • · Guidance for this submission of 510/k) for Indications of use as provided in Pediatric Information for X-ray Imaging Device Premarket Notifications (Document issued on November 28, 2017) Guidance for Industry and Food and Drug Administration Staff.
  • · Guidance for the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. (Document issued September 2023)
  • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices . (Document issued June 2023)
  • · Guidance for Industry and FDA Staff: Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically powered Medical Devices. (Document issued June 2022)
  • Guidance for this submission of 510(k) for (SSXID) Solid State X-ray Imaging Devices issued on: September 1, 2016, was used to establish substantial equivalence.
  • · Guidance for industry and FDA Staff User Fees and Refunds for Premarket Notification Submissions 510(k)s, (Document issued October 2022)
  • Guidance for industry and FDA Staff Refuse to Accept Policy for 510(k) (Document issued April 2022)
  • Guidance for industry and FDA Staff -Format for Traditional and Abbreviated 510(k)s (Document issued on September 2019)
  • · Guidance for industry and FDA Staff Deciding when to submit a 510(k) for a change to an existing device. (Document issued on October 25, 2017)
  • Guidance for industry and FDA Staff The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] (Document Issued on July 28, 2014)
  • · Guidance for industry and FDA Staff Guidance for Off-The-Shelf Software Use in Medical Devices (Document issued August 2023)
  • · Guidance for industry and FDA Staff Guidance for the Content of Premarket Submission for Software in Medical Devices. (Document issued June 2023)
  • Guidance for industry and Food and Drug Administration Staff Policy Clarification for Certain . Fluoroscopic Equipment Requirements (Document issued on May 8, 2019)
  • · Guidance for Industry and FDA Staff Medical X-Ray Imaging Devices Conformance with IEC Standards. (Document issued February 2023)
  • Guidance for Industry and FDA Staff Clarification of Radiation Control Regulations for Manufacturers of Diagnostic X-Ray Equipment. (Document issued on December 17, 2018)
  • · Guidance for Industry and FDA Administration Staff Pediatric Information for X-ray Imaging Device Premarket Notifications (Document issued on November 28, 2017)

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.