(184 days)
The Fingertip Pulse Oximeter are intended for measuring function of atterial hemoglobin (SpO2) and pulse rate for adults as non-invasive spot checking in professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during no-motion condition. The device is prescription only.
The subject device AVITA Pulse Oximeter with Bluetooth is non-invasive spot checking, not provided sterile, multi-use device, which can measure and display user's oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) through finger during no-motion condition. The subject device is not for lifesustaining, not for implant and does not contain drug or biological products. The device is for prescription use. The subject device consists of sensor, electronic circuits, display and plastic enclosures. It is a battery powered device and is adopted with two color OLED screen to display SpO2 and PR. It is software-driven and does not include alarms.
The provided text is a 510(k) Summary for the AViTA Pulse Oximeter (SP62B). It addresses medical device regulation, but it does not describe acceptance criteria and a study that proves the device meets specific acceptance criteria in the manner requested (e.g., in the context of an AI/ML algorithm's performance).
Instead, the document details the equivalence of the new device to a predicate device based on various non-clinical performance and safety tests, and a clinical validation of SpO2 performance.
Therefore, I cannot fully answer your request based on the provided text because the information specifically about "acceptance criteria and a study that proves the device meets the acceptance criteria" for an AI/ML device's performance, including details like sample size for test sets, data provenance, expert ground truth establishment, adjudication, MRMC studies, or standalone algorithm performance, is not present.
The document mentions "acceptance criteria" generally for biocompatibility tests, electrical safety, and EMC testing, but not for an AI/ML performance study as implied by your detailed questions.
Here's what can be extracted and inferred from the text regarding the device's overall performance validation, with limitations noted:
1. A table of acceptance criteria and the reported device performance:
The document states performance specifications rather than explicit acceptance criteria in a table format for each test, but the overall statement is that the device "met the acceptance criteria" for various tests. For SpO2 accuracy, the performance is explicitly stated.
| Test/Parameter | Acceptance Criteria (Implied/General) | Reported Device Performance |
|---|---|---|
| SpO2 Accuracy (70-100% range) | Within acceptable limits per ISO 80601-2-61 (likely ±2% or similar based on predicate) | ARMS (Accuracy Root Mean Square) = 1.89% |
| Electrical Safety & EMC | Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and ISO 80601-2-61 | Complied with all listed standards |
| Biocompatibility | Compliance with ISO 10993-1, -5, -10, -23 and FDA Biocompatibility guidance | All tested standards met the acceptance criteria |
| Software V&V | Compliance with FDA Guidance for "Moderate" level of concern software | Demonstrates functional performance and compliance |
| Cleaning Validation | Device performance not affected after multiple cleaning procedures | Performance not affected as illustrated in user manual |
2. Sample size used for the test set and the data provenance:
- Test set sample size: For the clinical performance testing, it mentions "healthy, adult volunteers." The exact number (sample size) is not specified in the provided text.
- Data provenance: "healthy, adult volunteers." The country of origin is not specified. The study was prospective as it involved clinical validation testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable in the context of a pulse oximeter's SpO2 accuracy against a reference measurement. Pulse oximeters are typically validated against arterial blood gas (ABG) measurements (co-oximetry) in a controlled desaturation study, where the ABG is the direct, objective ground truth. Expert readers are not usually involved in establishing ground truth for SpO2 values.
- The document implies the adherence to ISO 80601-2-61 for clinical performance, which outlines the methodology for such studies.
4. Adjudication method for the test set:
- Not applicable/Not specified. For SpO2 accuracy, the "ground truth" is typically the co-oximeter measurement from arterial blood, not a subjective interpretation requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI/ML device for diagnostic image interpretation. It is a pulse oximeter measuring physiological parameters.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable in the AI/ML sense. The device itself is "standalone" in that it directly measures SpO2 and PR; it's not an AI algorithm assisting a human interpreter.
7. The type of ground truth used:
- For SpO2 accuracy, the ground truth is implied to be co-oximetry measurement from arterial blood gas (ABG), as is standard for pulse oximeter validation according to ISO 80601-2-61. The document states "Clinical performance was conducted per ISO 80601-2-61. The clinical validation testing of the SpO2 performance under no motion on healthy, adult volunteers in the range of 70% to 100%."
8. The sample size for the training set:
- Not applicable. This is not an AI/ML device that requires a "training set" in the machine learning sense.
9. How the ground truth for the training set was established:
- Not applicable. (See point 8)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 19, 2025
AVITA Corporation Maggie Chao Adm. Manager 9F, No.78, Sec.1, Kwang Fu Road, Sanchong Dist., New Taipei City, 24158 Taiwan
Re: K242455
Trade/Device Name: AViTA Pulse Oximeter (SP62B) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: January 22, 2025 Received: January 22, 2025
Dear Maggie Chao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K242455
Device Name AVITA Pulse Oximeter (SP62B)
Indications for Use (Describe)
The Fingertip Pulse Oximeter are intended for measuring function of atterial hemoglobin (SpO2) and pulse rate for adults as non-invasive spot checking in professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during no-motion condition. The device is prescription only.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AViTA. The logo features the word "AViTA" in blue, with a stylized heart shape surrounding it. The top of the heart is yellow, the bottom is red, and the right side is green. There is a registered trademark symbol to the bottom right of the heart.
AVITA CORPORATION
510(K) Submission for AViTA Pulse Oximeter
510(K) Summary
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AViTA. The logo features the word "AViTA" in blue, with a stylized heart shape behind it. The heart is composed of three colored segments: yellow at the top, red at the bottom, and green on the right side. A registered trademark symbol is present near the bottom right of the heart.
APPLICANT INFORMATION 1
| Applicant: | AVITA Corporation |
|---|---|
| Address: | 9F, No.78, Sec.1, Kwang-Fu Rd., San-Chung District, New Taipei |
| City 24158, Taiwan | |
| Applicant Establishment Number: | 9617543 |
| Applicant Contact: | Maggie Chao |
| Email: | Maggie_chao@avita.com.tw |
| Phone Number: | +886-2-8512-1568 |
| Fax Number: | +886-2-8512-1347 |
| Date of Submission: | August 19, 2024 |
SUBJECT DEVICE 2
| Trade/Proprietary Name: | AViTA Pulse Oximeter |
|---|---|
| Common Name: | Pulse Oximeter |
| Review Panel: | Anesthesiology |
| Classification Product Code: | DQA |
| Regulation Number: | 870.2700 |
| Device Class: | II |
3 PREDICATE DEVICE
| 510(k) Number: | K223399 |
|---|---|
| Manufacturer: | AViTA Corporation |
DEVICE DESCRIPTION 4
The subject device AVITA Pulse Oximeter with Bluetooth is non-invasive spot checking, not provided sterile, multi-use device, which can measure and display user's oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) through finger during no-motion condition. The subject device is not for lifesustaining, not for implant and does not contain drug or biological products. The device is for prescription use. The subject device consists of sensor, electronic circuits, display and plastic enclosures. It is a battery powered device and is adopted with two color OLED screen to display SpO2 and PR. It is software-driven and does not
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the AViTA logo, which is a stylized heart shape with the letters "AViTA" to the left of the heart. The heart is made up of three different colors: yellow at the top, red on the left side, and green on the right side. Below the logo, the text "include alarms." is written in a simple, sans-serif font.
5 INDICATIONS FOR USE
The Fingertip Pulse Oximeter are intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for adults as non-invasive spot checking in professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches) and for patients during no-motion condition. The device is prescription only.
TECHNOLOGICAL CHARACTERISTICS 6
AViTA pulse oximeters work by the principles of spectrophotometry, emitting two different wavelengths of light, typically red and infrared, through a pulsating capillary bed, such as a fingertip. The sensor on the other side of the tissue detects the light that emerges from the tissues. The intensity of red and infrared light that is transmitted through the capillary bed.
Based on the differences in absorption between oxygenated and deoxygenated blood at specific wavelengths, the device can calculate the ratio of oxygenated hemoglobin (HbO) to total hemoglobin in the blood, which is known as oxygen saturation (SpO2).
It's important to keep the finger or the measurement site stationary during the reading to avoid introducing motion artifacts that could affect the accuracy of the measurement. Additionally, it is recommended to use the pulse oximeter before or after engaging in sports activities rather than during physical exercise. Do not use for continuous monitoring.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for AViTA. The logo features the word "AViTA" in blue, with a stylized heart shape surrounding it. The heart is composed of three colored sections: a yellow arc at the top, a green curve forming the right side, and a red curve at the bottom left. A registered trademark symbol is visible near the bottom right of the heart.
DEVICE COMPARISON TABLE 7
| Item | Proposed Device | Predicate Device | Predicate Device |
|---|---|---|---|
| ProductName | AViTA Pulse Oximeter | AViTA Pulse Oximeter | Beijing Choice PulseOximeter |
| Model No. | SP62B | SP61 | MD300C228 |
| 510(k) Information | |||
| RegulationNumber | 870.2700 | 870.2700 | 870.2700 |
| Classification | Class II | Class II | Class II |
| Product Code | DQA | DQA | DQA |
| K. No. | - | K223399 | K232975 |
| Indication for Use | |||
| Statement | The Fingertip Pulse Oximeterare intended for measuringfunctional oxygen saturation ofarterial hemoglobin (SpO2) andpulse rate for adults as non-invasive spot checking inprofessional caring environment.It is designed for fingersbetween 0.8cm and 2.3cm (0.3inches to 0.9 inches) and forpatients during no-motioncondition. | The Fingertip Pulse Oximeterare intended for measuringfunctional oxygen saturationof arterial hemoglobin (SpO2)and pulse rate for adults asnon-invasive spot checking inprofessional caringenvironment. It is designed forfingers between 0.8cm and2.3cm (0.3 inches to 0.9inches) and for patients duringno-motion condition. | The Pulse Oximeter is ahandheld non-invasive deviceintended forspot-checking of oxygensaturation of arterialhemoglobin (SpO2) andPulse Rate of adult,adolescent child and infantpatients in hospitals,hospital-type facilities andhomecare. The device can beused by the people whosefinger thickness is between8mm and 22mm (0.3 inchesto 0.9 inches). |
| Population | adults | adults | adult, adolescent, childand infant patients |
| Application site | finger | finger | finger |
| Performance | normal condition | normal condition | normal condition |
| Stand-alone ormodule | stand-alone | stand-alone | stand-alone |
| Single use ornot | multiple use | multiple use | multiple use |
| Useenvironment | professionalcaring environment | professionalcaring environment | hospitals, hospital-typefacilities and homecare |
| Comparison | The proposed device and the predicated device have the same intended use and classification. Allchanges in indications for use would not affect the safety andeffectiveness. The AViTA Pulse Oximeter SP62B is substantially equivalent to the predicatedevice (K223399) concerning the same intended use. | ||
| Test Principle | Traditional 510(k) | ||
| AViTA pulse oximeters | AViTA pulse oximeters | The pulse oximeter works by | |
| work by the principles of | work by the principles of | applying a sensor | |
| spectrophotometry, emitting | spectrophotometry, emitting | to a pulsating arteriolar | |
| two different wavelengths of | two different wavelengths | vascular bed. The sensor | |
| light, typically red and | of light, typically red and | contains a dual light | |
| infrared, through a pulsating | infrared, through a pulsating | source and photo detector. | |
| capillary bed, such as a | capillary bed, such as a | The one | |
| fingertip. The sensor on the | fingertip. The sensor on the | wavelength of light source is | |
| other side of the tissue | other side of the tissue | 660nm, which is red light; | |
| detects the light that | detects the light that | the other is 905nm, which is | |
| emerges from the tissues. | emerges from the tissues. | infrared-red light. | |
| The device then measures | The device then measures | Skin, bone, tissue and | |
| the intensity of red and | the intensity of red and | venous vessels | |
| infrared light that is | infrared light that is | normally absorb a | |
| transmitted through the | transmitted through the | constant | |
| capillary bed. | capillary bed. | amount of light over time. | |
| The photo detector | |||
| Based on the differences in | Based on the differences in | in finger sensor collects | |
| absorption between | absorption between | and converts the light into | |
| Principle | oxygenated and | oxygenated and | electronic signal which is |
| deoxygenated blood at | deoxygenated blood at | proportional to the light | |
| specific wavelengths, the | specific wavelengths, the | intensity. The arteriolar | |
| device can calculate the ratio | device can calculate the | bed normally pulsates and | |
| of oxygenated hemoglobin | ratio of oxygenated | absorbs variable amounts | |
| (HbO) to total hemoglobin in | hemoglobin (HbO) to total | of light during systole and | |
| the blood, which is known as | hemoglobin in the blood, | diastole, as blood volume | |
| oxygen saturation (SpO2). | which is known as oxygen | increases and decreases. | |
| saturation (SpO2). | The ratio of light absorbed | ||
| It's important to keep the | at systole and diastole is | ||
| finger or the measurement | It's important to keep the | translated | |
| site stationary during the | finger or the measurement | into an oxygen saturation | |
| reading to avoid introducing | site stationary during the | measurement. This | |
| motion artifacts that could | reading to avoid introducing | measurement is referred to | |
| affect the accuracy of the | motion artifacts that could | as SpO2. | |
| measurement. | affect the accuracy of the | ||
| Additionally, it is | measurement.Page 6 of 11Additionally, it is | ||
| Image: AVITA logo | recommended to use thepulse oximeter before orafter engaging in sportsactivities rather than duringphysical exercise. Do notuse for continuousmonitoring.The proposed devices of thissubmission do not differfrom the predicate device. | recommended to use thepulse oximeter before orafter engaging in sportsactivities rather than duringphysical exercise. Do notuse for continuousmonitoring.The proposed devices of thissubmission do not differfrom the predicate device. | Traditional 510(k) |
| MeasurementMethod | Wavelength | Wavelength | Wavelength |
| Comparison | The proposed device and the predicated device have the same principle. There is noadditional question of safety and effectiveness as compared to the predicate device.AVITA Pulse Oximeter SP62B is substantially equivalent to the predicateddevice K223399) concerning the test principle. | ||
| Energy | |||
| Type | Battery | Battery | Battery |
| Battery | AAA Alkaline battery x 1 | AAA Alkaline battery x 1 | AAA Alkaline battery x 2 |
| Comparison | There is no additional question of safety and effectiveness as compared to the predicatedevice raised by the battery. The AVITA Pulse Oximeter SP62B is substantiallyequivalent to the predicate device (K223399) concerning the energysource. | ||
| Operation Features | |||
| On/Off | Automatic turn on and off | Automatic turn on and off | Automatic turn on and off |
| Display | Two color OLED | Two color OLED | LCD |
| Input Key | A 4-directional key | A 4-directional key | A 2-directional key |
| Warning/Indicator | Visual indicator | Visual indicator | Visual indicator |
| Warning /IndicatorFunction | Reading starts to flash whendetects user SpO2 and Pulserate.Display shows "Finger Out"when nomeasurement detects. | Reading starts to flash whendetects user SpO2 and Pulserate.Display shows “FingerOut” when nomeasurement detects. | the device will power offautomatically in 8 secondswhen "Finger out" isdisplayed |
| DisplayRotation | Yes | Yes | Yes |
| Bluetooth | BLE 4.2 | N/A | BLE 5.0 |
| Comparison | The proposed device is configured the Bluetooth 4.2 module which predicate |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for AViTA. The logo has the word "AViTA" in blue font. To the right of the word is a heart shape, with the top of the heart in green, the top left in yellow, and the bottom left in red.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for AVITA. The word "AVITA" is in blue, and there is a heart-shaped design to the right of the word. The heart is colored in yellow, red, and green.
{10}------------------------------------------------
{11}------------------------------------------------
| device is configured the Bluetooth 5.0 module. The Bluetooth is the independent functionmodule, it would not affect other function module of the device. In additional 510(k)embedded software could meet the requirements of Submissions for Software Containedin Medical Devices, the related document of software could refer to Software Documentsection of this submission. Therefore, this difference does not affect substantiallyequivalence between proposed device and predicatedevice on safety and effectiveness. | ||||||
|---|---|---|---|---|---|---|
| General Specification | ||||||
| OperatingTemp | 5 °C to 40 °C ( 41 °F to104 °F) | 5 °C to 40 °C ( 41 °F to104 °F) | 0 °C to 40 °C ( 32 °F to104 °F) | |||
| StorageTemp. | -30°C to 70 °C (-22 °F to158 °F) | -30°C to 70 °C (-22 °F to158 °F) | -25 | |||
| Humidity | 10% - 90%(non-condensing) | 10% - 90%(non-condensing) | 15% to 93%, non-condensing for bothoperating and storage | |||
| AtmosphericPressure | 700 hPa - 1060 hPa for bothoperating andstorage | 700 hPa - 1060 hPa for bothoperating andstorage | 70kPa~106kpa for bothoperating and storage | |||
| WaterResistance | IP22 | IP22 | IEC 60601-1-11 | |||
| Comparison | There's no difference in usage life and atmospheric pressure. The AVITA Pulse OximeterSP62B is substantially equivalent to the predicate device (K223399)concerning the technological characteristics. | |||||
| Classification | ||||||
| Applied Part | Type BF | Type BF | Type BF | |||
| Safety | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | |||
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | |||
| HarmonizedStandard | ISO 80601-2-61 | ISO 80601-2-61 | ISO 80601-2-61 | |||
| Mode ofOperation | Spot checking | Spot checking | Spot checking | |||
| Appearance | ||||||
| Weight | weight without battery:26.5g | weight without battery:25g | weight without battery:50g | |||
| Size | L68 mm x W37.8 mmxH28.5 mm | L68 mm x W37.8 mmxH27.7 mm | L58 mm x W32mm x H34mm | |||
| Comparison | The differences of general specification and appearance will not be considered as a NSEbetween the proposed device and the predicate device. The AVITA Pulse OximeterSP62B is substantially equivalent to the predicate device (K223399)concerning the general specification and appearance.Pulse Oximetry Specification | |||||
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows a logo for AViTA. The logo features the word "AViTA" in blue, sans-serif font. Above the "A" is a curved yellow shape, and below the "A" is a curved red shape. To the right of the word is a green heart shape, with the "R" symbol inside a circle in the bottom right corner of the heart.
| Traditional 510(k | |||
|---|---|---|---|
| Range | 0% to 100% | 0% to 100% | 0% to 100% |
| Resolution unit | 1% | 1% | 1% |
| Accuracy | 70% to 100% range ± 2%,less than 70% areunspecified | 70% to 100% range ± 2%,less than70% are unspecified | 70% to 100% range ± 2%,less than0~69% no definition |
| Comparison | The AVITA Pulse OximeterSP62B is substantially equivalent to the predicatedevice (K223399) concerning the general specification and appearance. | ||
| Biocompatibility Testing | |||
| cytotoxicity | In accordance with ISO10993-1 | In accordance with ISO10993-1 | In accordance with ISO10993-1 |
| skinsensitization, | In accordance with ISO10993-1 | In accordance with ISO10993-1 | In accordance with ISO10993-1 |
| skin irritation | In accordance with ISO10993-1 | In accordance with ISO10993-1 | In accordance with ISO10993-1 |
| Material used | Enclosure material: ABSLens: PMMAMylar: PET | Enclosure material: ABSLens: PMMAMylar: PET | Battery Cover: ABSEnclosure: ABSScreen protect Film:PMMAPower Button: ABS |
| Comparison | The AVITA Pulse Oximeter SP62B is substantially equivalent to the predicatedevice (K223399) concerning the general specification and appearance. | ||
| Pulse Rate Specification | |||
| Range | 30 to 250 bpm | 30 to 250 bpm | 30 to 250 bpm |
| Resolution unit | 1 bpm (no decimal) | 1 bpm (no decimal) | 1 bpm (no decimal) |
| Accuracy | ±2 bpm or ±2%,whichever is greater | ±2 bpm or ±2%,whichever is greater | 30 bpm |
| Comparison | The AVITA Pulse Oximeter SP62B is substantially equivalent to the predicatedevice (K223399) concerning the performance specification |
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the logo for AViTA. The logo features the text "AViTA" in a bold, sans-serif font, with the "A" and "V" in blue and the "iTA" in black. Above the text is a stylized heart shape, with the top portion in yellow, the bottom left portion in red, and the right side in green. A registered trademark symbol is located to the bottom right of the heart.
8 PERFORMANCE TESTING
The following tests were conducted to evaluate the safety and effectiveness of the subject device, and the test results indicated that the subject device is safe and effective.
8.1 Performance Data
All necessary performance testing was conducted oxygenated hemoglobin characterization testing on the subject device to support a determination of substantial equivalence to the predicate device in accordance with ISO 80601-2-61.
8.2 Electrical Safety and EMC Testing
The laboratory tests of electroal safety, electromagnetic compatibility, testing were conducted and showed that the subject device complied with IEC 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility – Requirements and tests, IEC 60601-1-11 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical systems used in home healthcare environment, and ISO 80601-2-61 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
Biocompatibility testing 8.3
The biocompatibility evaluation for the subject device with surface contact and prolonged duration was in accordance with the FDA Biocompatibility guidance (Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"). The standards below are tested and met the acceptance criteria.
- Biological evaluation (ISO10993-1)
- Cytotoxicity (ISO10993-5)
- Sensitization (ISO10993-10)
- Irritation (ISO10993-23)
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image shows the logo for AViTA. The logo consists of the word "AViTA" in blue, with a stylized heart shape surrounding it. The top of the heart is yellow, the left side is red, and the right side is green. There is a registered trademark symbol in the bottom right corner of the heart.
8.4 Software Verification and Validation
Software verification and validation were provided in compliance with FDA Guidance "The Content of the Premarket Submission for Software Contained in Medical Devices". The verifications and validations demonstrate that the subject device work functionally. The software for the subject device is a "moderate" level of concern, which is identical to the predicate device. A failure or latent flaw in the software could not directly cause serious injury or death to the patient or operator, but a non-serious injury could occur. According to FDA Guidance document, the software validation documentation summarized the required for a Moderate level of concern device.
Cleaning Validation 8.5
Cleaning validation was executed in accordance with FDA Guidance "Reprocessing Medical Device in Health Care Setting: Validation Methods and Labeling" The performance of the subject device will not be affected after multiple cleaning procedures as illustrated in user manual.
Clinical Performance 8.6
Clinical performance was conducted per ISO 80601-2-61. The clinical validation testing of the SpO2 performance under no motion on healthy, adult volunteers in the range of 70% to 100%. The ARMS for SpO2 under no motion was found to be 1.89%. No adverse effects and complications happened during the clinical study.
9 SUBSTANTIAL EQUIVALENCE
Based upon equivalences in: intended use, patient population, conditions of use, operating principles, and the non-clinical performance data, the subject device have been shown to be safe and effective and to perform equivalently as compared to the legally marketed predicate device. Therefore, the subject devices are substantially equivalent to the legally marketed predicate device.
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).