(270 days)
The PPTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation.
When the patient is in the seated position using the chair, the System is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the head, neck and thorax.
The P-CURE Proton Therapy System (PPTS) is comprised of four main subsystems that function in tandem to generate the desired dose level and distribution at the target site:
-
Beam production system (Synchrotron based accelerator)
- Injector – produces and delivers protons to the synchrotron
- Synchrotron ring – accelerates the proton beam in circular orbit (within the ring) to the desired energy level
- Extraction system - extracts the beam from the ring to the beam delivery subsystem
-
Beam delivery system for a single fixed beam treatment room. Steers and monitors the extracted proton pencil beam from the synchrotron to the desired treatment location (Nozzle).
-
Patient Positioning System (P-ARTIS). Mechanically orients the patient (seated or on supine); provides independent means of patient registration using CT (3D) and X-ray (2D)
- CT system (P-ARTIS CT)
- Robotic arm and chair/couch (6 Degree of freedom Couch) (P-ARTIS PPS)
- X-ray system (P-ARTIS XR)
- Positioning Software (P-ART)
-
Control and Safety Systems
- Control Subsystem (TSM). Synchronizes the various subsystem actions and connects with hospital oncology information systems and PACS.
- Safety Subsystem. Includes hardware and software means to ensure safe system operation for patient and personnel. It includes subsystem interlocks, treatment beam parameters monitoring, and others.
The provided FDA 510(k) clearance letter for the P-Cure Proton Therapy System (PPTS) does not contain specific acceptance criteria or a detailed study description with performance metrics that would allow for a comprehensive table and answer to all the questions. This document is a clearance letter, which summarizes the outcome of a review, rather than providing the full technical details of the submission.
However, based on the information provided, here's what can be extracted and inferred:
1. Table of Acceptance Criteria and Reported Device Performance
The clearance letter does not list specific numerical acceptance criteria (e.g., minimum accuracy percentages, maximum error values) or direct quantitative performance results in a table format. It states that:
- "In all instances, the PTTS functioned as intended and met its specifications."
- "Testing demonstrated substantial equivalence in terms of performance and safety to the predicate."
To construct a table, we would need the actual specifications and the measured performance against those specifications, which are typically found in the full 510(k) submission, not the clearance letter.
Inferred Performance Claims (from "Performance Data" section):
| Acceptance Criteria Category (Inferred) | Reported Device Performance (Summary from Letter) |
|---|---|
| Mechanical Performance | Verified performance of the positioning system. |
| Beam Performance | Evaluated beam dose shape, beam dose, dose rate, dose monitoring, and spot positioning. (Implied: met specifications) |
| Safety Interface Performance | Verified collision sensors, mechanical interlocks. (Implied: functioned as intended) |
| Integration with Oncology Info Systems | Verification testing for integration. (Implied: functioned as intended) |
| Integration with Positioning & Treatment Planning Systems | Validation testing for integration. (Implied: functioned as intended) |
| Repeatability/Reproducibility | Testing to support repeatability and reproducibility of patient positioning and immobilization. (Implied: met specifications) |
| Electrical Safety & Essential Performance | Conducted based on IEC 60601-1, IEC 60601-1-2, IEC TR 60601-4-2, EN 606601-2-44, IEC 60601-1-3, IEC 60601-1-8, IEC 60601-2-54, IEC 60601-1-64, IEC 60601-2-68, IEC 62667, and AAPM TG-224. (Implied: device complies with these standards) |
| Software Documentation & Validation | Documented and validated per FDA Guidance Document "Content of Premarket Submissions for Device Software Functions," and per IEC. (Implied: software functions as intended and safely) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not provide details on the "sample size" in terms of patient data or case numbers for the performance testing. The testing described is primarily technical and engineering verification and validation of the system's components and functions (mechanical, beam, safety, software integration, repeatability). It does not mention clinical studies with patient data.
- Sample Size for Test Set: Not specified for clinical cases. The testing appears to be system-level verification and validation, not patient-based clinical performance data.
- Data Provenance: Not applicable as no patient data (e.g., country of origin, retrospective/prospective) is referenced for the performance testing cited. The submitter, P-Cure, Ltd., is located in Israel, but this pertains to the company, not necessarily the origin of any clinical data.
3. Number of Experts Used to Establish Ground Truth and Qualifications
The document does not mention any "ground truth" established by experts in the context of clinical performance, as it focuses on the technical verification and validation of the device's physical and software functions. Therefore, this question is not applicable based on the provided text.
4. Adjudication Method for the Test Set
As no expert review or clinical case evaluation is mentioned, there is no adjudication method described. This question is not applicable based on the provided text.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No MRMC study is mentioned. The clearance letter details technical and engineering performance testing, not studies comparing human reader performance with or without AI assistance. This question is not applicable based on the provided text.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
The device described is a physical medical device (Proton Therapy System), not an AI algorithm to be used standalone or with human-in-the-loop for diagnostic or prognostic purposes. While the system has software, its "performance" refers to the entire system's ability to produce and deliver a proton beam accurately and safely. This question is not applicable in the typical sense of AI standalone performance.
7. The Type of Ground Truth Used
The "ground truth" for the performance testing appears to be based on:
- Known engineering specifications and physical laws: For beam performance, mechanical movements, dose delivery accuracy, etc.
- Safety standards: Compliance with IEC and AAPM standards.
- Software requirements: Validation against specified software functions.
There is no mention of expert consensus, pathology, or outcomes data as "ground truth" for the reported performance testing.
8. The Sample Size for the Training Set
The concept of a "training set" is usually applicable to machine learning algorithms. While the system involves software, the document describes traditional software validation and verification for system control and safety functions, not the development of a machine learning model that would require a distinct training set. Therefore, this question is not applicable based on the provided text.
9. How the Ground Truth for the Training Set Was Established
As no "training set" (in the context of machine learning) is mentioned, the method for establishing its ground truth is also not applicable.
FDA 510(k) Clearance Letter - P-Cure Proton Therapy System
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
May 12, 2025
P-Cure, Ltd.
℅ Lina Kontos
Regulatory Counsel
Hogan Lovells US LLP
555 Thirteenth Street NW
Washington, District of Columbia 20004
Re: K242418
Trade/Device Name: P-Cure Proton Therapy System (PPTS)
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: LHN
Dated: September 19, 2024
Received: September 19, 2024
Dear Lina Kontos:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K242419 - Lina Kontos
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
Page 3
K242419 - Lina Kontos
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Indications for Use
Submission Number (if known)
K242418
Device Name
P-Cure Proton Therapy System (PPTS)
Indications for Use (Describe)
The PPTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation.
When the patient is in the seated position using the chair, the System is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the head, neck and thorax.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
Page 5
K242418 - 510(k) SUMMARY
P-Cure, Ltd.'s P-Cure Proton Therapy System (PTTS)
Submitter:
P-Cure, Ltd.
14 Carmel St.
Shilat Industrial Zone, 7318800, Israel
Phone: +972 54 488 1399
Contact Person: Ori Lubin
Date Prepared: May 9, 2025
Name of Device: P-Cure Proton Therapy System (PTTS)
Common or Usual Name: Proton beam therapy systems
Classification Name: Medical Charged-Particle Radiation Therapy System, 21 CFR 892.5050
Regulatory Class: Class II
Product Code: LHN
Predicate Device:
P-Cure Proton Beam Therapy System (PPBTS) (K221996)
Reference Device:
ProTom Radiance 330 (K191521)
Intended Use / Indications for Use:
The PPTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation.
When the patient is in the seated position using the chair, the System is indicated for treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the head, neck and thorax.
Technological Characteristics:
The P-CURE Proton Therapy System (PPTS) is comprised of four main subsystems that function in tandem to generate the desired dose level and distribution at the target site:
- Beam production system (Synchrotron based accelerator)
- Injector – produces and delivers protons to the synchrotron
Page 6
K242418
Page 2 of 5
-
Synchrotron ring – accelerates the proton beam in circular orbit (within the ring) to the desired energy level
-
Extraction system - extracts the beam from the ring to the beam delivery subsystem
-
Beam delivery system for a single fixed beam treatment room. Steers and monitors the extracted proton pencil beam from the synchrotron to the desired treatment location (Nozzle).
-
Patient Positioning System (P-ARTIS). Mechanically orients the patient (seated or on supine); provides independent means of patient registration using CT (3D) and X-ray (2D)
- CT system (P-ARTIS CT)
- Robotic arm and chair/couch (6 Degree of freedom Couch) (P-ARTIS PPS)
- X-ray system (P-ARTIS XR)
- Positioning Software (P-ART)
-
Control and Safety Systems
- Control Subsystem (TSM). Synchronizes the various subsystem actions and connects with hospital oncology information systems and PACS.
- Safety Subsystem. Includes hardware and software means to ensure safe system operation for patient and personnel. It includes subsystem interlocks, treatment beam parameters monitoring, and others.
Summary of Technological Characteristics and Comparison to Predicate Device:
The PPTS, the predicate device, and the reference device all share the same fundamental principles of operation, namely that they accelerate a proton beam to the same energy range by using a synchrotron and identically deliver the prescribed treatment dose to the patient by a pencil beam scanning technique. The PPTS uses the same image guided and robotic patient positioning system (P-ARTIS) to move the patient ensuring delivery of the beam to the target area, as prescribed.
While the previously cleared PPBTS positioned patients in a seated position alone, the PPTS can additionally position the patient in a supine position for treatment using an off-the-shelf couch, similar to the identified reference device, ProTom Radiance 330. The PPTS, PPBTS, and ProTom Radiance 330 all use the same robotic arm for mounting the patient support.
In all three systems, the simulation CT dataset are acquired in the treatment patient orientation. For a "seated" planning, the CT dataset is acquired by using in room vertical CT (performed identically for planning for PPBTS and PPTS). For a "supine" planning, the CT simulation data is generated at standard simulation settings of radiation oncology departments (performed identically for planning for ProTom Radiance 330 and PPTS). In all three systems, orthogonal x-ray imaging is used for image guided position verification at isocenter.
A table comparing the key features of the subject, predicate, and reference devices is provided below.
Page 7
K242418
Page 3 of 5
| Device Characteristic | P-Cure Proton Therapy System (PPTS) | P-Cure Proton Beam Therapy System (PPBTS) (Predicate Device) | Radiance 330 Proton Beam Therapy System (Reference Device) |
|---|---|---|---|
| 510(k) Number | K242418 | K221996 | K191521 |
| Intended Use/ Indications for Use | The PPTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation. When using the chair, the System is indicated to treatment of patients with localized tumors and other conditions susceptible to treatment by radiation in the head, neck and thorax. | The P-CURE system is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation in the head, neck and thorax. | The ProTom Radiance 330 is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other condition susceptible to treatment by radiation |
| Accelerator | Synchrotron | Synchrotron | Synchrotron |
| Particle | Protons | Protons | Protons |
| Variable | Yes; 70-250 MeV | Yes; 70-250 MeV | Yes; 70-250 MeV |
| Cycle Time | Variable | Variable | Variable |
| Beam Transport | Bending and focusing magnets | Bending and focusing magnets | Bending and focusing magnets |
| Nozzles | Pencil Beam Scanning | Pencil Beam Scanning | Pencil Beam Scanning |
| Beam Range in Patient (Tissue Depth) | 3 cm to 38 cm clinical use | 3 cm to 38 cm clinical use | 3 cm to 38 cm clinical use |
| Range Verifier | Optional | Optional | Optional |
| Control and Safety System | Software controls accelerator, beam transport and delivery, sets operational parameters, monitors systems and provides alerts regarding error conditions. | Software controls accelerator, beam transport and delivery, sets operational parameters, monitors systems and provides alerts regarding error conditions. | Software controls accelerator, beam transport and delivery, sets operational parameters, monitors systems and provides alerts regarding error conditions. |
| Mechanical Beam Stops | Yes | Yes | Yes |
| Beam Intensities | Continuously variable over range | Continuously variable over range | Continuously variable over range |
Page 8
K242418
Page 4 of 5
| Device Characteristic | P-Cure Proton Therapy System (PPTS) | P-Cure Proton Beam Therapy System (PPBTS) (Predicate Device) | Radiance 330 Proton Beam Therapy System (Reference Device) |
|---|---|---|---|
| Shielding | Steel and Concrete | Steel and Concrete | Steel and Concrete |
| Gantry Rotating Angle | N/A. Fixed beam | N/A. Fixed beam | Rotating gantry and Fixed beam |
| Patient Positioner | Yes – couch or chair mounted on a robotic patient positioning device | Yes – chair mounted on a robotic patient positioning device | Yes – couch mounted on a robotic patient positioning device |
| Patient Positioner | 6 degree of freedom robotic positioning device. | 6 degree of freedom robotic positioning device. | 6 degree of freedom robotic positioning device. |
| Maximum Load | Chair: 180 kg Couch: 235 kg | Chair: 180 kg | Unknown |
| Imaging | Yes – Diagnostic quality CT (seated configuration) and orthogonal x-ray system | Yes – Diagnostic quality CT and orthogonal x-ray system | Yes – couch mounted CBCT and x-ray system |
| Treatment Room | Horizontal fixed gantry with a robotic chair or a couch | Horizontal fixed gantry with a robotic chair | Rotating gantry with a robotic couch Horizontal fixed gantry with a robotic couch |
Performance Data:
The Company performed testing as follows:
- Mechanical testing to verify performance of the positioning system
- Beam performance testing to evaluate beam dose shape, beam dose, dose rate, dose monitoring, and spot positioning
- Safety interface testing to verify collision sensors, mechanical interlocks
- Simulation and validation testing to verify integration with oncology information systems
- Validation testing to verify integration with positioning system and treatment planning system
- Testing to support repeatability and reproducibility of patient positioning and immobilization
Product safety and essential performance electrical testing was conducted based upon the following consensus standards: IEC 60601-1, IEC 60601-1-2, IEC TR 60601-4-2, EN 606601-2-44, IEC 60601-1-3, IEC 60601-1-8, IEC 60601-2-54, IEC 60601-1-64, IEC 60601-2-68, IEC 62667, and AAPM TG-224.
Software was documented and validated per FDA's Guidance Document "Content of Premarket Submissions for Device Software Functions," and per IEC
In all instances, the PTTS functioned as intended and met its specifications. Testing demonstrated substantial equivalence in terms of performance and safety to the predicate.
Page 9
K242418
Page 5 of 5
Substantial Equivalence Conclusion:
The PPTS has the same intended use and substantially similar indications for use and technological characteristics as the predicate, PPBTS (K221996). The minor differences in indications for use statements and technological characteristics between the PPTS and the predicate device do not raise new or different questions of safety or effectiveness. The completed verification and validation data demonstrate that the device meets its specifications with the commercially available patient positioning configuration, which are comparable to those of the reference device. Thus, the PPTS system can be found substantially equivalent to the predicate.
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.