K Number
K242383
Device Name
Apex Locator
Date Cleared
2025-03-07

(207 days)

Product Code
Regulation Number
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.
Device Description
Apex Locator is an oral equipment used for root canal measurements. It features an LCD color display with functional buttons, enabling users to conveniently view parameters such as battery status, test wire connection status, and apex position etc. Users can easily adjust sound level, brightness, and reference settings using intuitive buttons, and perform functional checks of both the device and cable. The device is intended to be sterilized prior to use.
More Information

Not Found

No
The description focuses on basic electronic functions, display, and user interface for measuring root canal length. There is no mention of AI, ML, or any complex data processing that would suggest the use of these technologies. The performance studies listed are standard for medical devices and do not indicate AI/ML validation.

No
This device is for measurement and diagnostic purposes (detecting the apex of the root canal), not for treating a disease or condition.

Yes

The device is used to "detect the apex of root canal" and provides "root canal measurements" with a "Measurement Accuracy: ±0.5mm." This functional capability to detect and measure a specific anatomical point for treatment planning and execution categorizes it as a diagnostic device in the context of dentistry.

No

The device description explicitly mentions hardware components such as an LCD color display, functional buttons, and a cable, and also mentions sterilization, indicating a physical device.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
  • Apex Locator Function: The Apex Locator described is used to measure the length of a root canal within a tooth. It does not analyze biological samples taken from the body. It's a device used directly on the patient's tooth during a dental procedure.
  • Intended Use: The intended use clearly states it's for "detecting the apex of root canal" and is used "in hospital environments, clinics or dental offices by qualified dental personnel." This is a clinical procedure, not an in vitro test.

Therefore, based on the provided information, the Apex Locator is a medical device used in a clinical setting for a dental procedure, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

Product codes (comma separated list FDA assigned to the subject device)

LQY

Device Description

Apex Locator is an oral equipment used for root canal measurements. It features an LCD color display with functional buttons, enabling users to conveniently view parameters such as battery status, test wire connection status, and apex position etc. Users can easily adjust sound level, brightness, and reference settings using intuitive buttons, and perform functional checks of both the device and cable. The device is intended to be sterilized prior to use.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Root canal Softened dentin

Indicated Patient Age Range

Not Found

Intended User / Care Setting

qualified dental personnel. This device must only be used in hospital environments, clinics or dental offices.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The proposed device is the same as the predicate device but the two devices are different. A series of safety and performance tests were conducted on the subject device:

  • Biocompatibility
  • Software Validation
  • Electromagnetic compatibility and electrical safety
  • Function test

All the test results demonstrate Apex Locator meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate device.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K212178

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

N/A

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 7, 2025

Shenzhen Rogin Medical Co., Ltd. % Salon Chen System Engineer IMD Medical & Drug Technology Service Institutions Room 308, Building 11, No. 23 Jinqu Road Wanjiang District. Dongguan City, Guangdong Province Guangdong, 523039 CHINA

Re: K242383

Trade/Device Name: Apex Locator Regulatory Class: Unclassified Product Code: LQY Dated: February 5, 2025 Received: February 5, 2025

Dear Salon Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

2

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

510(k) Number (if known) K242383

Device Name Apex Locator

Indications for Use (Describe)

Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

K242383

510(k) Summary

1. Submitter's Identification:

  • Company Name: Shenzhen Rogin Medical Co., Ltd.

  • Address: Rm 3109, Bldg 6, Tian'an Cloud Park, Bantian Str, Longgang Dist, Shenzhen, > Guangdong, 518129, CHINA.
  • Phone: +86-755- 26758272

  • Fax: +86-755- 26758272

  • Contact Person (Title): Nature Xu (General Manager)

  • E-mail: admin@rogindental.com

  • Date of Preparation: August 24, 2024

Common Name and Classification: 2.

  • Device Classification Name: Locator, Root apex

  • Classification Product Code: LQY

  • Requlation Number: Pre-Amendment

  • ア Class: Unclassified
  • Review Panel: Dental

  • Device Name: Apex Locator

  • Model: APEX-S, APEX-R

  • Predicate Device Information: 3.
  • ア 510(k) Number: K212178
  • ア Device Classification Name: Locator, Root apex
  • Sponsor: Foshan COXO Medical Instrument Co., Ltd.

  • Classification Product Code: LQY

  • Regulation Number: Pre-Amendment

  • Class: Unclassified

  • Review Panel: Dental

5

  • Device name: Root Apex Locator

  • Common Name: Locator, Root Apex

Application Correspondent 4.

  • Company Name: IMD Medical & Drug technology service institutions

  • ア Phone: +86-18613190779
  • Fax: +86-755-62809168

  • Contact Person (Title): Salon Chen (System engineer)

  • E-mail: 33999439@qq.com

  • Address: Room 308, Building 11, No. 23 Jinqu Road, Wanjiang District, Dongguan City, > Guangdong Province, China

5. Device Description

Apex Locator is an oral equipment used for root canal measurements. It features an LCD color display with functional buttons, enabling users to conveniently view parameters such as battery status, test wire connection status, and apex position etc. Users can easily adjust sound level, brightness, and reference settings using intuitive buttons, and perform functional checks of both the device and cable. The device is intended to be sterilized prior to use.

6. Indications for Use

Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

7. Comparison to the predicate device

6

Table 1 General Comparison

| Elements of

ComparisonProposed DevicePredicate DeviceJudgment
Company
NameShenzhen Rogin Medical Co., Ltd.Foshan COXO Medical Instrument Co.,
Ltd./
Device NameApex LocatorRoot Apex Locator/
Classification
Product
CodeLQYLQYSE
RegulationN/AN/ASE
Classification
NameDentalDentalSE
ClassUnclassifiedUnclassifiedSE
Prescription
or OTCPrescription UsePrescription UseSE
Anatomical
SitesRoot canal Softened dentinRoot canal Softened dentinSE
Target
PopulationPatients in need of root canal
surgeryPatients in need of root canal surgerySE
Intended UseApex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.The Root Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.SE
Safety factor &
PerformanceProposed DevicePredicate DeviceJudgment
Electrical SafetyCompliance with IEC 60601-1Compliance with IEC 60601-1SE
EMCCompliance with IEC
60601-1-2Compliance with IEC 60601-1-2SE
Usability
EngineeringCompliance with IEC 62366-1Compliance with IEC 62366-1SE
SoftwareCompliance with IEC 62304Compliance with IEC 62304SE
BiocompatiblityCompliance with ISO 10993-1Compliance with ISO 10993-1SE
power SupplyLithium battery (DC3.7V)Lithium battery (DC3.7V)SE
Charger Power SupplyAC100V-240VAC100V-240VSE
Frequency of Supply
Voltage50Hz ~ 60Hz50Hz ~ 60HzSE
ComponentsMeasuring wire
File clip
Lip hook
Touch Probe
TesterMeasuring wire
File clip
Lip hook
Touch probe
TesterSE
Safety factor &
PerformanceProposed DevicePredicate DeviceJudgment
DimensionsAPEX-S:
Length:106mm
Width: 75mm
Thickness:24mm
APEX-R:
Length:110mm
Width: 67mm
Thickness: 52mmLength: 110mm
Width: 103 mm
Thickness: 21.5mmDifference
(1)
Cable lengthCharger cable: 1.5m
Measuring wire: 1.5mCharger cable: 1.5m
Tester cable: 1.4mDifference
(2)
WeightAPEX-S: 540g
APEX-R: 330g270gDifference
(3)
Power supplyLi-ion Battery
3.7V, 800mAhLi-ion Battery
DC 3.7V 1800mAhDifference
(4)
ChargerInput: AC 100-240 V, 50/60 Hz
Output: DC 5V 1A
Classification of Protection
against Electric Shock:
Class II (adaptor)Input: AC 100-240 V,
50/60 Hz
Output: DC 5V 1A
Classification of Protection
against Electric Shock:
Class II (adaptor)SE
Means of inputButton controlTouch screen
Foldable main unitDifference
(5)
Safety factor &
PerformanceProposed DevicePredicate DeviceJudgment
Adjustment before
measurementReference setting function
enables to mark an individual
predetermined reference
position at the required
distance from the apex. This
variable apical arrow can be
set between the first green bar
and the last green bar.The DR'S CHOICE Apical Arrow
function enables to mark an
individual predetermined reference
position at the required distance
from the apex. This variable apical
arrow can be set between the first
green bar and the last yellow bar.Difference
(6)
Measurement Accuracy±0.5mm±0.5mmSE
Patient contacting
components materialsLip hook: Stainless steel;
File clip: stainless steel, plastic PBTLip hook: 304 Stainless steel;
File clip: PI and 304Stainless steelDifference
(7)
SterilizationLip clip and file clip are user
sterilized by steam
sterilization.Lip clip and file clip are user
sterilized by steam sterilization.SE

7

8

9

Review of Differences :

Difference (1)

The proposed device includes the same main specifications as compared to the predicate device but differs in dimension, which is caused by different appearance design and could not effects the performance and safety.

Difference (2)

There is a difference in the cable length, but these cables include the same function, the cable length difference would not affect its safety and effectiveness.

Difference (3)

The proposed device and the predicate device have a different weight, however, this could not effect the performance and safety.

10

Difference (4)

Both the proposed device, the predicate device is powered by rechargeable 3.7V Li-ion battery, there is a difference in the battery capacity. The proposed device complies with electrical safety standard IEC 60601-1, and its Li-ion battery complies with the battery safety standard IEC 62133-2, so the difference in battery capacity would not affect its safety and effectiveness.

Difference (5)

The proposed device includes the same main specifications as compared to the predicate device but Means of input in appearance, which is caused by different appearance design and could not effects the performance and safety.

Difference (6)

Differences in color at the reference position do not affect the performance of the product.

Difference (7)

Both the Patient contacting components materials of proposed device and predicate device meet the ISO10993-5 & ISO10993-10 standards, and the proposed device also meets the ISO10993-11 standard, the difference would not affect its safety and effectiveness.

8. Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence are as follows:

The proposed device is the same as the predicate device but the two devices are different. A series of safety and performance tests were conducted on the subject device:

  • Biocompatibility
  • Software Validation
  • Electromagnetic compatibility and electrical safety
  • Function test

All the test results demonstrate Apex Locator meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate device.

11

9. Clinical Tests Performed

No clinical test data was used to support the decision of substantial equivalence.

10. Conclusion

The subject devices have all features of the predicate devices. The few differences do not affect the safety and effectiveness of the subject devices.

Thus, the subject devices are substantially equivalent to the predicate devices.