K Number
K242383
Device Name
Apex Locator
Date Cleared
2025-03-07

(207 days)

Product Code
Regulation Number
N/A
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

Device Description

Apex Locator is an oral equipment used for root canal measurements. It features an LCD color display with functional buttons, enabling users to conveniently view parameters such as battery status, test wire connection status, and apex position etc. Users can easily adjust sound level, brightness, and reference settings using intuitive buttons, and perform functional checks of both the device and cable. The device is intended to be sterilized prior to use.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called "Apex Locator". It details a comparison to a predicate device and states that non-clinical tests were performed. However, it explicitly states that no clinical test data was used to support the decision of substantial equivalence. This means that a study proving the device meets acceptance criteria, as typically understood in the context of clinical trials or performance evaluations with human subjects/data, was not conducted or reported in this submission. The "acceptance criteria" and "device performance" mentioned here refer to specifications confirmed through non-clinical testing and comparison to the predicate.

Given this, I cannot provide information on:

  • The study that "proves the device meets the acceptance criteria" in a clinical sense.
  • Sample size used for the test set and data provenance (as no clinical test data was used).
  • Number of experts used to establish ground truth or their qualifications.
  • Adjudication method.
  • Multi-reader multi-case (MRMC) comparative effectiveness study.
  • Standalone (algorithm only) performance.
  • Type of ground truth (as no clinical data).
  • Sample size for the training set (as no clinical data).
  • How ground truth for the training set was established (as no clinical data).

I can, however, provide the acceptance criteria based on the non-clinical tests and the device performance as reported in the comparison table and discussion of differences.


Non-Clinical Acceptance Criteria and Reported Device Performance

The provided document describes acceptance criteria and reported "performance" based on non-clinical tests and comparison to a predicate device. The primary method for demonstrating performance and safety was through compliance with recognized standards and verification that design differences do not affect safety and effectiveness.

Key takeaway: This submission focuses on engineering, safety, and functional conformance rather than clinical performance data.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Standard / Metric)Reported Device Performance (Proposed Device)
Electrical Safety (Compliance)Compliance with IEC 60601-1
EMC (Compliance)Compliance with IEC 60601-1-2
Usability Engineering (Compliance)Compliance with IEC 62366-1
Software (Compliance)Compliance with IEC 62304
Biocompatibility (Compliance)Compliance with ISO 10993-1 for patient-contacting components (Lip hook: Stainless steel; File clip: stainless steel, plastic PBT). Also meets ISO 10993-11.
Power Supply (Type)Lithium battery (DC3.7V)
Charger Power Supply (Input)AC100V-240V, 50Hz ~ 60Hz
Charger OutputDC 5V 1A, Class II (adaptor)
ComponentsMeasuring wire, File clip, Lip hook, Touch Probe, Tester
Measurement Accuracy±0.5mm
SterilizationLip clip and file clip are user sterilized by steam sterilization.
Battery Safety (Compliance)Li-ion battery complies with IEC 62133-2
Function TestMeets requirements of pre-defined acceptance criteria and intended uses. (Specifics not detailed beyond general statement).

Explanation of "Study that proves the device meets the acceptance criteria":

The "study" referenced in the document is a series of non-clinical tests and a detailed comparison to a legally marketed predicate device. The conclusion of substantial equivalence is based on demonstrating that the proposed device:

  1. Complies with relevant recognized safety and performance standards: These include IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 62366-1 (usability), IEC 62304 (software), ISO 10993-1 and ISO 10993-11 (biocompatibility), and IEC 62133-2 (battery safety).
  2. Performs the same intended function as the predicate device.
  3. Has fundamental technological characteristics that are substantially equivalent to the predicate device.
  4. Any differences in technological characteristics (e.g., dimensions, weight, battery capacity, means of input) do not raise new questions of safety or effectiveness. For each noted difference, the submitter provided a justification that it "would not affect its safety and effectiveness" or "could not effects the performance and safety."
  5. Achieves the same specified performance metric as the predicate device where applicable (e.g., Measurement Accuracy: ±0.5mm).

The document explicitly states: "No clinical test data was used to support the decision of substantial equivalence." Therefore, there is no clinical study that "proves the device meets the acceptance criteria" in terms of patient outcomes or comparative clinical performance. The acceptance is based on engineering principles, compliance with standards, and a robust non-clinical comparison to a predicate device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 7, 2025

Shenzhen Rogin Medical Co., Ltd. % Salon Chen System Engineer IMD Medical & Drug Technology Service Institutions Room 308, Building 11, No. 23 Jinqu Road Wanjiang District. Dongguan City, Guangdong Province Guangdong, 523039 CHINA

Re: K242383

Trade/Device Name: Apex Locator Regulatory Class: Unclassified Product Code: LQY Dated: February 5, 2025 Received: February 5, 2025

Dear Salon Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242383

Device Name Apex Locator

Indications for Use (Describe)

Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K242383

510(k) Summary

1. Submitter's Identification:

  • Company Name: Shenzhen Rogin Medical Co., Ltd.

  • Address: Rm 3109, Bldg 6, Tian'an Cloud Park, Bantian Str, Longgang Dist, Shenzhen, > Guangdong, 518129, CHINA.
  • Phone: +86-755- 26758272

  • Fax: +86-755- 26758272

  • Contact Person (Title): Nature Xu (General Manager)

  • E-mail: admin@rogindental.com

  • Date of Preparation: August 24, 2024

Common Name and Classification: 2.

  • Device Classification Name: Locator, Root apex

  • Classification Product Code: LQY

  • Requlation Number: Pre-Amendment

  • ア Class: Unclassified
  • Review Panel: Dental

  • Device Name: Apex Locator

  • Model: APEX-S, APEX-R

  • Predicate Device Information: 3.
  • ア 510(k) Number: K212178
  • ア Device Classification Name: Locator, Root apex
  • Sponsor: Foshan COXO Medical Instrument Co., Ltd.

  • Classification Product Code: LQY

  • Regulation Number: Pre-Amendment

  • Class: Unclassified

  • Review Panel: Dental

{5}------------------------------------------------

  • Device name: Root Apex Locator

  • Common Name: Locator, Root Apex

Application Correspondent 4.

  • Company Name: IMD Medical & Drug technology service institutions

  • ア Phone: +86-18613190779
  • Fax: +86-755-62809168

  • Contact Person (Title): Salon Chen (System engineer)

  • E-mail: 33999439@qq.com

  • Address: Room 308, Building 11, No. 23 Jinqu Road, Wanjiang District, Dongguan City, > Guangdong Province, China

5. Device Description

Apex Locator is an oral equipment used for root canal measurements. It features an LCD color display with functional buttons, enabling users to conveniently view parameters such as battery status, test wire connection status, and apex position etc. Users can easily adjust sound level, brightness, and reference settings using intuitive buttons, and perform functional checks of both the device and cable. The device is intended to be sterilized prior to use.

6. Indications for Use

Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.

7. Comparison to the predicate device

{6}------------------------------------------------

Table 1 General Comparison

Elements ofComparisonProposed DevicePredicate DeviceJudgment
CompanyNameShenzhen Rogin Medical Co., Ltd.Foshan COXO Medical Instrument Co.,Ltd./
Device NameApex LocatorRoot Apex Locator/
ClassificationProductCodeLQYLQYSE
RegulationN/AN/ASE
ClassificationNameDentalDentalSE
ClassUnclassifiedUnclassifiedSE
Prescriptionor OTCPrescription UsePrescription UseSE
AnatomicalSitesRoot canal Softened dentinRoot canal Softened dentinSE
TargetPopulationPatients in need of root canalsurgeryPatients in need of root canal surgerySE
Intended UseApex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.The Root Apex Locator is used to detect the apex of root canal. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.SE
Safety factor &PerformanceProposed DevicePredicate DeviceJudgment
Electrical SafetyCompliance with IEC 60601-1Compliance with IEC 60601-1SE
EMCCompliance with IEC60601-1-2Compliance with IEC 60601-1-2SE
UsabilityEngineeringCompliance with IEC 62366-1Compliance with IEC 62366-1SE
SoftwareCompliance with IEC 62304Compliance with IEC 62304SE
BiocompatiblityCompliance with ISO 10993-1Compliance with ISO 10993-1SE
power SupplyLithium battery (DC3.7V)Lithium battery (DC3.7V)SE
Charger Power SupplyAC100V-240VAC100V-240VSE
Frequency of SupplyVoltage50Hz ~ 60Hz50Hz ~ 60HzSE
ComponentsMeasuring wireFile clipLip hookTouch ProbeTesterMeasuring wireFile clipLip hookTouch probeTesterSE
Safety factor &PerformanceProposed DevicePredicate DeviceJudgment
DimensionsAPEX-S:Length:106mmWidth: 75mmThickness:24mmAPEX-R:Length:110mmWidth: 67mmThickness: 52mmLength: 110mmWidth: 103 mmThickness: 21.5mmDifference(1)
Cable lengthCharger cable: 1.5mMeasuring wire: 1.5mCharger cable: 1.5mTester cable: 1.4mDifference(2)
WeightAPEX-S: 540gAPEX-R: 330g270gDifference(3)
Power supplyLi-ion Battery3.7V, 800mAhLi-ion BatteryDC 3.7V 1800mAhDifference(4)
ChargerInput: AC 100-240 V, 50/60 HzOutput: DC 5V 1AClassification of Protectionagainst Electric Shock:Class II (adaptor)Input: AC 100-240 V,50/60 HzOutput: DC 5V 1AClassification of Protectionagainst Electric Shock:Class II (adaptor)SE
Means of inputButton controlTouch screenFoldable main unitDifference(5)
Safety factor &PerformanceProposed DevicePredicate DeviceJudgment
Adjustment beforemeasurementReference setting functionenables to mark an individualpredetermined referenceposition at the requireddistance from the apex. Thisvariable apical arrow can beset between the first green barand the last green bar.The DR'S CHOICE Apical Arrowfunction enables to mark anindividual predetermined referenceposition at the required distancefrom the apex. This variable apicalarrow can be set between the firstgreen bar and the last yellow bar.Difference(6)
Measurement Accuracy±0.5mm±0.5mmSE
Patient contactingcomponents materialsLip hook: Stainless steel;File clip: stainless steel, plastic PBTLip hook: 304 Stainless steel;File clip: PI and 304Stainless steelDifference(7)
SterilizationLip clip and file clip are usersterilized by steamsterilization.Lip clip and file clip are usersterilized by steam sterilization.SE

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Review of Differences :

Difference (1)

The proposed device includes the same main specifications as compared to the predicate device but differs in dimension, which is caused by different appearance design and could not effects the performance and safety.

Difference (2)

There is a difference in the cable length, but these cables include the same function, the cable length difference would not affect its safety and effectiveness.

Difference (3)

The proposed device and the predicate device have a different weight, however, this could not effect the performance and safety.

{10}------------------------------------------------

Difference (4)

Both the proposed device, the predicate device is powered by rechargeable 3.7V Li-ion battery, there is a difference in the battery capacity. The proposed device complies with electrical safety standard IEC 60601-1, and its Li-ion battery complies with the battery safety standard IEC 62133-2, so the difference in battery capacity would not affect its safety and effectiveness.

Difference (5)

The proposed device includes the same main specifications as compared to the predicate device but Means of input in appearance, which is caused by different appearance design and could not effects the performance and safety.

Difference (6)

Differences in color at the reference position do not affect the performance of the product.

Difference (7)

Both the Patient contacting components materials of proposed device and predicate device meet the ISO10993-5 & ISO10993-10 standards, and the proposed device also meets the ISO10993-11 standard, the difference would not affect its safety and effectiveness.

8. Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence are as follows:

The proposed device is the same as the predicate device but the two devices are different. A series of safety and performance tests were conducted on the subject device:

  • Biocompatibility
  • Software Validation
  • Electromagnetic compatibility and electrical safety
  • Function test

All the test results demonstrate Apex Locator meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate device.

{11}------------------------------------------------

9. Clinical Tests Performed

No clinical test data was used to support the decision of substantial equivalence.

10. Conclusion

The subject devices have all features of the predicate devices. The few differences do not affect the safety and effectiveness of the subject devices.

Thus, the subject devices are substantially equivalent to the predicate devices.

N/A