(161 days)
Not Found
No
The document describes automated functions and algorithms for gas delivery and alarms, but does not mention AI or ML.
Yes
The device is designed to deliver nitric oxide therapy to patients, specifically neonates with hypoxic respiratory failure, which is a therapeutic intervention.
No
Explanation: The device is a therapy system that delivers nitric oxide and monitors gas concentrations, but it does not provide a medical diagnosis.
No
The device description explicitly details hardware components such as a base unit, mobile cart, bedside rail holder, gas regulators, and patient kits, in addition to the software interface and control system.
Based on the provided information, the Ulspira TS Nitric Oxide Therapy System is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use clearly states that the device is for the "delivery of nitric oxide (NO) and the monitoring of inspired NO, NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy." This describes a therapeutic and monitoring device used directly on a patient, not a device used to examine specimens taken from the body.
- Device Description: The description focuses on the system's ability to deliver gas, monitor gas concentrations in the respiratory circuit, and provide a user interface and alarms. These are all functions related to patient care and therapy delivery, not laboratory testing of biological samples.
- Lack of IVD Characteristics: There is no mention of analyzing biological specimens (blood, urine, tissue, etc.), detecting specific analytes in those specimens, or providing diagnostic information based on such analysis.
Therefore, the Ulspira TS Nitric Oxide Therapy System is a medical device used for therapy and monitoring in a clinical setting, not an IVD device.
N/A
Intended Use / Indications for Use
The Ulspira TS Nitric Oxide Therapy System is intended for use by healthcare professionals for the delivery of nitric oxide (NO) and the monitoring of inspired NO, NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy.
The Ulspira TS must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use.
The Ulspira TS primary NO therapy system delivers NO gas in the 0-80 ppm range while in the Constant Rate or flow sensing modes. This includes:
· Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system.
· A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb.
• Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO cylinders, and a low O2 pressure alarm when using an oxygen cylinder.
· An automated emergency dosing algorithm activated by certain high-risk alarms, which impact patient dosing, to avoid sudden cessation of therapy.
· Compatibility with a wide inspiratory flow rate range of 0.25-120 //min, utilizing an automatically detected low or high flow sensor.
· An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access.
The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment during transit within hospitals.
Product codes
MRN, MRO, MRP, MRQ, CCL
Device Description
Ulspira TS Nitric Oxide Therapy System delivers physician-prescribed NO therapy gas and monitors inspired Nitric Oxide, Nitrogen Dioxide, and Oxygen gas in combination with a respiratory device.
The main device functionalities of the Ulspira TS Nitric Oxide Therapy System include:
-
A primary delivery system to administer NO gas into a respiratory device circuit.
-
Monitoring of NO, NO2, and O2 gas concentrations close to the patient interface.
-
System includes a user interface that contains all controls used to set the NO delivery and monitoring parameters. All set parameters as well as other information are shown on the user interface screen.
-
The system will produce visual and audible alarms if vital parameters vary beyond preset or default limits.
-
The system includes an integrated pneumatic back-up system for manual hand bagging in order to deliver NO therapy in the event of a failure of the primary delivery system and during manual ventilation.
The Ulspira TS system consists of the base unit, the mobile cart and bedside rail holder, and various components and accessories, including the gas regulators and patient kits for use with validated respiratory devices.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
term and near-term (>34weeks gestation) neonates
Intended User / Care Setting
healthcare professionals; Neonatal Intensive Care Unit (NICU), transport of neonates (within hospitals)
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Study Type: Nonclinical Tests
Sample Size: Not specified, but tested with seven different ventilators.
Key Results: The testing concluded that ULSPIRA TS Nitric Oxide Therapy System is compatible with the selected ventilators. The ULSPIRA TS Nitric Oxide Therapy System performed within specifications when used with each ventilator.
Evaluated conditions: Connection of the systems, Accuracy of applied NO, O2 dilution, NO2 formation, Alarm functionality, Influence on ventilator performance.
Key Metrics
NO delivery accuracy: ±20% or 2 ppm, whichever is the greatest.
Backup NO delivery accuracy: Within ±20% of set value or ±2 ppm, whichever is the greatest.
NO measurement accuracy: +- (0.5 ppm +20 % of actual concentration) in the range 0-20 ppm; +- (0.5 ppm +10 % of actual concentration) in the range 20-120 ppm
NO2 measurement accuracy: ±(20% or 0.5 ppm), whichever is the greatest.
O2 measurement accuracy: ±(2.5 % volume fraction + 2.5 % of gas concentration)
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 868.5165 Nitric oxide administration apparatus.
(a)
Identification. The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”
0
January 17, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Airgas Therapeutics LLC % Mikel Alberdi President Drug Device Consulting 13014 North Dale Mabry Highway, #326 Tampa, Florida 33618
Re: K242374
Trade/Device Name: ULSPIRA TSTM Nitric Oxide Therapy System Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: Class II Product Code: MRN, MRO, MRP, MRQ, CCL Dated: December 20, 2024 Received: December 20, 2024
Dear Mikel Alberdi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
2
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
3
Indications for Use
Submission Number (if known)
Device Name
ULSPIRA TS™ Nitric Oxide Therapy System
Indications for Use (Describe)
The Ulspira TS Nitric Oxide Therapy System is intended for use by healthcare professionals for the delivery of nitric oxide (NO) and the monitoring of inspired NO. NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy.
The Ulspira TS must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use.
The Ulspira TS primary NO therapy system delivers NO gas in the 0-80 ppm range while in the Constant Rate or flow sensing modes. This includes:
· Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system.
· A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb.
• Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO cylinders, and a low O2 pressure alarm when using an oxygen cylinder.
· An automated emergency dosing algorithm activated by certain high-risk alarms, which impact patient dosing, to avoid sudden cessation of therapy.
· Compatibility with a wide inspiratory flow rate range of 0.25-120 //min, utilizing an automatically detected low or high flow sensor.
· An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access.
The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment during transit within hospitals.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
4
510(k) SUMMARY
The following summary is provided in accordance with 21 CFR 807.92:
A. DATE
January 17, 2025
B. APPLICANT
Airgas Therapeutics 6141 Easton Rd. P.O. Box 310 Plumsteadville PA 18949 United States
- Mr. Timothy Warren Contact Person: Airgas Therapeutics Senior Director of Quality and Regulatory Affairs (215) 766-7407 x3034 tim.warren@airgas.com
C. CORRESPONDENT
Drug Device Consulting 13014 North Dale Mabry Highway, #326 Tampa FL 33618 United States
Contact Person: Mr. Mikel Alberdi Drug Device Consulting (813) 708-8303 malberdi@drugdeviceconsulting.com
D. DEVICE
Device Trade Name: | UISPIRA TS™ Nitric Oxide Therapy System |
---|---|
Common Name: | Nitric Oxide Administration Apparatus |
Classification: | Class II - 21 CFR 868.5165 |
Classification Name: | Apparatus, Nitric Oxide Delivery |
Product Codes: | MRN (Primary), MRO, MRP, MRQ, CCL |
E. PREDICATE DEVICE
K212409 - Ulspira TS Nitric Oxide Therapy System
5
ﺖ DEVICE DESCRIPTION
Ulspira TS Nitric Oxide Therapy System delivers physician-prescribed NO therapy gas and monitors inspired Nitric Oxide, Nitrogen Dioxide, and Oxygen gas in combination with a respiratory device.
The main device functionalities of the Ulspira TS Nitric Oxide Therapy System include:
-
A primary delivery system to administer NO gas into a respiratory device circuit.
-
Monitoring of NO, NO2, and O2 gas concentrations close to the patient interface.
-
System includes a user interface that contains all controls used to set the NO delivery and monitoring parameters. All set parameters as well as other information are shown on the user interface screen.
-
The system will produce visual and audible alarms if vital parameters vary beyond preset or default limits.
-
The system includes an integrated pneumatic back-up system for manual hand bagging in order to deliver NO therapy in the event of a failure of the primary delivery system and during manual ventilation.
The Ulspira TS system consists of the base unit, the mobile cart and bedside rail holder, and various components and accessories, including the gas regulators and patient kits for use with validated respiratory devices.
G. INDICATIONS FOR USE
The Ulspira TS Nitric Oxide Therapy System is intended for use by healthcare professionals for the delivery of nitric oxide (NO) and the monitoring of inspired NO, NO2 and O2 concentrations for a patient undergoing inhaled Nitric Oxide (iNO) therapy.
The Ulspira TS must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Refer to this material prior to use.
The Ulspira TS primary NO therapy system delivers NO gas in the 0-80 ppm range while in the Constant Rate or flow sensing modes. This includes:
• Continuous integrated monitoring for inspired NO, NO2 and a comprehensive alarm system.
• A touch-screen user interface with a waveform display of the ventilation device flow as measured in the inspiratory limb.
• Cylinder handling facilitated by manual or an automatic cylinder switch which is reactive to the detected gas supply state of NO cylinders, and a low O2 pressure alarm when using an oxygen cylinder.
6
- An automated emergency dosing algorithm activated by certain high-risk alarms, which impact patient dosing, to avoid sudden cessation of therapy.
• Compatibility with a wide inspiratory flow rate range of 0.25-120 //min, utilizing an automatically detected low or high flow sensor.
• An internal battery which provides at least two hours of uninterrupted therapy and a 12V DC inlet for additional external battery access.
The integrated Ulspira TS pneumatic backup NO therapy system provides backup NO delivery capability that is intended to deliver a continuous flow of NO mixed with O2, for iNO therapy which allows continuous treatment during transit within hospitals.
H. TECHNOLOGICAL COMPARISON
Ulspira TS Nitric Oxide Therapy System modifications include a modification of the labeling to include six additional respiratory devices. The Ulspira TS Nitric Oxide Therapy System with the additional respiratory devices has the identical intended use and uses the same device hardware as the cleared Ulspira TS Nitric Oxide Therapy System predicate device. All features are identical except for those described in the table below.
| Device name | Ulspira TS
(K212409) | Ulspira TS
(Proposed Device) | Comparison |
|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Comparison of general attributes, indication, patient population, operating environment, etc. | | | |
| Product code(s) | MRN | MRN | Identical |
| Regulation number | 21 CFR 868.5165 | 21 CFR 868.5165 | Identical |
| Regulation Description | - Nitric oxide administration apparatus (Nitric
oxide delivery/backup delivery)
- Nitric oxide analyzer
- Nitrogen dioxide analyzer | - Nitric oxide administration apparatus (Nitric
oxide delivery/backup delivery) - Nitric oxide analyzer
- Nitrogen dioxide analyzer | Identical |
| Indications for Use | The Ulspira TS Nitric Oxide Therapy System is
intended for use by healthcare professionals
for the delivery of nitric oxide (NO) and the
monitoring of inspired NO, NO2 and O2
concentrations for a patient undergoing
inhaled Nitric Oxide (iNO) therapy and must
only be used in accordance with the
indications, contraindications, warnings and
precautions described in the nitric oxide
drug packaging inserts and labeling
(currently neonates). Refer to this material
prior to use.
The Ulspira TS primary NO therapy system
delivers NO gas in the 0-80 ppm range while
in the Constant Rate or flow sensing modes.
This includes:
• Continuous integrated monitoring for
inspired NO, NO2 and O2 and a
comprehensive alarm system.
• A touch-screen user interface with a
waveform display of the ventilation device
flow as measured in the inspiratory limb.
• Cylinder handling facilitated by manual or
an automatic cylinder switch which is
reactive to the detected gas supply state of
NO cylinders, and a low O2 pressure alarm
when using an oxygen cylinder. | The Ulspira TS Nitric Oxide Therapy System is
intended for use by healthcare professionals
for the delivery of nitric oxide (NO) and the
monitoring of inspired NO, NO2 and O2
concentrations for a patient undergoing
inhaled Nitric Oxide (iNO) therapy and must
only be used in accordance with the
indications, contraindications, warnings and
precautions described in the nitric oxide
drug packaging inserts and labeling
(currently neonates). Refer to this material
prior to use.
The Ulspira TS primary NO therapy system
delivers NO gas in the 0-80 ppm range while
in the Constant Rate or flow sensing modes.
This includes:
• Continuous integrated monitoring for
inspired NO, NO2 and O2 and a
comprehensive alarm system.
• A touch-screen user interface with a
waveform display of the ventilation device
flow as measured in the inspiratory limb.
• Cylinder handling facilitated by manual or
an automatic cylinder switch which is
reactive to the detected gas supply state of
NO cylinders, and a low O2 pressure alarm
when using an oxygen cylinder. | Identical |
7
K242374 – 510(k) SUMMARY
| Device name | Ulspira TS
(K212409) | Ulspira TS
(Proposed Device) | Comparison | | | | |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-------------------------------------------------------------------|-----------------------------|-----------|--|
| | • An automated emergency dosing algorithm
activated by certain high-risk alarms, which
impact patient dosing, to avoid sudden
cessation of therapy.
• Compatibility with a wide inspiratory flow
rate range of 0.25-120 l/min, utilizing an
automatically detected low or high flow
sensor.
• An internal battery which provides at least
two hours of uninterrupted therapy and a
12V DC inlet for additional external battery
access.
The integrated Ulspira TS pneumatic backup
NO therapy system provides backup NO
delivery capability that is intended to deliver
a continuous flow of NO mixed with O2, for
iNO therapy which allows continuous
treatment during transit within hospitals. | • An automated emergency dosing algorithm
activated by certain high-risk alarms, which
impact patient dosing, to avoid sudden
cessation of therapy.
• Compatibility with a wide inspiratory flow
rate range of 0.25-120 l/min, utilizing an
automatically detected low or high flow
sensor.
• An internal battery which provides at least
two hours of uninterrupted therapy and a
12V DC inlet for additional external battery
access.
The integrated Ulspira TS pneumatic backup
NO therapy system provides backup NO
delivery capability that is intended to deliver
a continuous flow of NO mixed with O2, for
iNO therapy which allows continuous
treatment during transit within hospitals. | | | | | |
| Physical dimensions and
weight (excl. carrier/cart) | Weight: 7.0 kg
Width: 320 mm
Depth: 150 mm
Height: 300 mm | Weight: 7.0 kg
Width: 320 mm
Depth: 150 mm
Height: 300 mm | Identical | | | | |
| NO gas connectors | CGA 626 | CGA 626 | Identical | | | | |
| NO Injection location | NO injection connected between ventilator
and humidifier. | NO injection connected between ventilator
and humidifier. | Identical | | | | |
| Battery backup | Yes | Yes | Identical | | | | |
| Battery backup time | 2h | 2h | Identical | | | | |
| Power Supply - Main | Voltage: 100-240V, 50 - 60Hz | Voltage: 100-240V, 50 - 60Hz | Identical | | | | |
| Automated Pre-Use
check | Yes | Yes | Identical | | | | |
| | Alarms | | | | | | |
| NO Delivery/Flow Sensor
Alarms | Yes | Yes | Identical | | | | |
| Power Supply/Battery
Alarms | Yes | Yes | Identical | | | | |
| NO, NO2, O2 Monitoring
Alarms | Yes | Yes | Identical | | | | |
| | Primary NO administration system | | | | | | |
| NO administration
principle | NO delivery into the inspiratory limb of a
ventilation device's patient circuit. | NO delivery into the inspiratory limb of a
ventilation device's patient circuit. | Identical | | | | |
| Range of NO gas
concentration delivered | 0-80 ppm | 0-80 ppm | Identical | | | | |
| NO delivery accuracy | ±20% or 2 ppm, whichever is the greatest. | ±20% or 2 ppm, whichever is the greatest. | Identical | | | | |
| Operating Modes | The subject device incorporates two modes
(Autosense and Jet Sense) that provides a
user set dose of NO into the inspiratory limb
of the respiratory device circuit, based on
the measured respiratory device flow.
The subject device also includes a third
mode, Constant Rate, which provides flow
of NO (0.5 - 60 l/min) into the inspiratory
limb of the respiratory device circuit. | The subject device incorporates two modes
(Autosense and Jet Sense) that provides a
user set dose of NO into the inspiratory limb
of the respiratory device circuit, based on
the measured respiratory device flow.
The subject device also includes a third
mode, Constant Rate, which provides flow
of NO (0.5 - 60 l/min) into the inspiratory
limb of the respiratory device circuit. | Identical | | | | |
| Backup NO administration system | | | | | | | |
| Backup power source | Pneumatic system | Pneumatic system | Identical | | | | |
| Backup NO
Administration | Backup system delivering a user set
concentration of NO, via adjustment of O2
flow, to the patient via a resuscitator. | Backup system delivering a user set
concentration of NO, via adjustment of O2
flow, to the patient via a resuscitator. | Identical | | | | |
| Backup NO delivery
accuracy | accuracy | Within ±20% of set value or ±2 ppm,
whichever is the greatest. | Which ever is the greatest. | Within ±20% of set value or ±2 ppm,
whichever is the greatest. | Which ever is the greatest. | Identical | |
8
| Device name | Ulspira TS
(K212409) | Ulspira TS
(Proposed Device) | Comparison |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Gas analysis (NO, NO2, O2) - General characteristics | | | |
| Breathing circuit sample
source location | On the inspiratory limb of the breathing
circuit, after the humidifier. | On the inspiratory limb of the breathing
circuit, after the humidifier. | Identical |
| Sample flow rate | 150 ml/min | 150 ml/min | Identical |
| NO gas analysis | | | |
| Integrated NO Gas
Analyzer | Yes | Yes | Identical |
| NO measurement
accuracy | +- (0.5 ppm +20 % of actual concentration) in
the range 0-20 ppm
+- (0.5 ppm +10 % of actual concentration) in
the range 20-120 ppm | +- (0.5 ppm +20 % of actual concentration) in
the range 0-20 ppm
+- (0.5 ppm +10 % of actual concentration) in
the range 20-120 ppm | Identical |
| NO Measurement range | 0 - 120 ppm | 0 - 120 ppm | Identical |
| NO2 gas analysis | | | |
| Integrated NO2 Gas
Analyzer | Yes | Yes | Identical |
| NO2 measurement
accuracy | ±(20% or 0.5 ppm), whichever is the
greatest. | ±(20% or 0.5 ppm), whichever is the
greatest. | Identical |
| NO2 measurement range | 0 - 30 ppm | 0 - 30 ppm | Identical |
| O2 gas analysis | | | |
| Integrated O2 Gas
Analyzer | Yes | Yes | Identical |
| O2 measurement
accuracy | ±(2.5 % volume fraction + 2.5 % of gas
concentration) | ±(2.5 % volume fraction + 2.5 % of gas
concentration) | Identical |
| O2 measurement range | 18 - 100 % | 18 - 100 % | Identical |
| Software | | | |
| Software version | Version 1.5 | Version 1.6
Software Changes:
- Improvements in onscreen appearance
- correct minor items noted during software
testing | Substantially Equivalent:
Minor updates/
differences
The changes in the
software are minor and
would not require a
510(k) submission to
justify the changes in
accordance with FDA
Guidance Deciding
When to Submit a
510(k) for a Software
Change to an Existing
Device. |
| User Manual | | | |
| User Manual | Original | Revision 5
Updated to add new respiratory devices | Substantially Equivalent:
Minor updates/
differences |
9
K242374 – 510(k) SUMMARY
NON-CLINICAL TESTS SUMMARY & CONCLUSION l.
Summary of Nonclinical Tests:
The testing concluded that ULSPIRA TS Nitric Oxide Therapy System is compatible with the selected ventilators.
The seven ventilators, and the ULSPIRA TS Nitric Oxide Therapy System, were set up and calibrated according to the manufacturer's recommendations, and tested using the settings established for each ventilator.
Multiple levels of nitric oxide delivery were evaluated [0 (baseline), 1, 5, 20, 40, and 80 ppm].
The following conditions were evaluated for the ULSPIRA TS Nitric Oxide Therapy System when used in conjunction with the ventilators:
- Connection of the systems
- Accuracy of applied NO
- 02 dilution
- NO2 formation
- Alarm functionality
- Influence on ventilator performance
Nonclinical Testing Conclusion:
The ULSPIRA TS Nitric Oxide Therapy System performed within specifications when used with each ventilator.
CONCLUSIONS J.
The above described nonclinical data support the substantial equivalence of the device with the predicate device. The supporting hardware, and the software verification and validation and usability testing demonstrate that the Ulspira TS Nitric Oxide Therapy System performs as intended in the specified use conditions. Risk assessments and completed testing did not raise different questions of safety and effectiveness. Airgas Therapeutics concludes that the performance data for the subject device shows that it is substantially equivalent to the cleared predicate device.