K Number
K242261
Manufacturer
Date Cleared
2024-11-06

(98 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Accelx Implant System implants are indicated for use in partially or fully edentulous patients to support maxillary and mandibular single-unit, multiple-unit, and overdenture dental restorations. Accelx Implant System implants are indicated for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Implant diameters larger than 6.0 mm are dedicated for the molar region and are indicated for delayed loading.

Accelx Implant System implants are compatible with Accelx Implant System abutments.

Accelx Implant System implants are compatible with MegaGen AnyRidge Internal Implant System titanium abutments as listed below.

Implant System CompatibilityPlatform Diameter, mmMaximum Angulation
MegaGen AnyRidge Internal Implant System3.5, 4.0, 4.25, 5.025°

Accelx Implant System cover screws are compatible with MegaGen AnyRidge Internal Implant System components as listed below.

Implant System CompatibilityImplant Body Diameter, mmImplant Platform, mm
MegaGen AnyRidge Internal Implant System4.03.5
4.43.5
4.93.5, 4.0
5.43.5, 4.25
5.93.5, 4.25
6.45.0
6.95.0
7.45.0
7.95.0
8.45.0
Device Description

The Accelx Implant System comprises Accelx dental implants, Accelx abutments previously cleared in K203240. The Accelx Implant System is compatible with the MegaGen AnyRidge Internal Implant System, regardless of the implant platform; this compatibility is due to the internal taper connection.

Accelx dental implants have an internal taper abutment interface connection and an internal hexagonal feature for abutment anti-rotation and instrument. The internal threaded section is for mating to the corresponding Accelx cover screws, abutments and screws previously cleared in K203240, and MegaGen AnyRidge Internal Implant System restorative devices.

Accelx dental implants are provided in ten body diameters ranging from 4.0 mm to 8.4 mm, and in total lengths ranging from 7.7 mm to 14.2 mm. Implants with a body diameter of 4.0 mm and 4.4 mm have a platform diameter of 3.5 mm. Implants with a body diameter of 4.9 mm have a platform diameter of 3.5 mm or 4.0 mm. Implants with a body diameter of 5.4 mm to 5.9 mm have a platform diameter of 3.5 mm. Implants with a body diameter of 6.4 mm to 8.4 mm have a platform diameter of 5.0 mm. The implants are manufactured from titanium alloy conforming to ASTM F136. The endosseous surface of the implant is aluminum oxide-blasted arom the implant collar to the apex. All Accelx dental implants are intended to be placed 0.5-1.0 mm sub crestal.

Accelx cover screws are provided with a coronal diameter of 3.5 mm in overall lengths of 5.7 mm, 6.5 mm and gingival heights (cuff heights) of 0.8 mm, 1.6 mm, respectively. The cover screws are manufactured from titanium alloy confirming to ASTM F136 and anodized. The cover screws are single use only. The smallest cover screw is provided sterile within the Subject device implant packaging. They are also provided individually as replacement or alternative parts in the same non-sterile packaging configuration as cleared in K203240.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Accelx Implant System. It primarily focuses on demonstrating substantial equivalence to predicate devices based on non-clinical performance data and a comparison of technological characteristics. It does not contain information about clinical studies, AI/algorithm performance, multi-reader multi-case (MRMC) studies, or any form of human-in-the-loop performance evaluation.

Therefore, I cannot provide a table of acceptance criteria or details about a study proving the device meets those criteria in the context of an AI/ML-driven device, as the submission is for a medical device (dental implant system) that does not incorporate AI/ML.

However, I can extract the non-clinical acceptance criteria and reported device performance from the document.

Non-Clinical Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criterion / Test Method (as recommended in FDA guidance)Reported Device Performance
Modified SurfacesEvaluation by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) following media blasting and cleaning."Implant surfaces were evaluated after media blasting and cleaning by scanning electron microscope (SEM) and energy dispersive X-ray spectroscopy (EDS) characterization meeting acceptance criteria."
BiocompatibilityCytotoxicity testing according to ISO 10993-5.
Bacterial endotoxin testing to ANSI/AAMI ST72, USP and USP .
Material mediated pyrogenicity per USP Pyrogen Test."Cytotoxicity testing according to ISO 10993-5, bacterial endotoxin to ANSI/AAMI ST72, USP and USP and material mediated pyrogenicity per USP Pyrogen Test was performed on representative devices." (Implies successful completion as per general context of 510(k) submission for substantial equivalence).
SterilizationGamma sterilization validation for implants according to ISO 13004.
Moist heat (steam) sterilization for non-sterile cover screws according to ISO 17665-2."Gamma sterilization validation was performed according to ISO 13004 was performed for implants; moist heat (steam) sterilization according to ISO 17665-2 for non-sterile cover screws." (Implies successful validation as per general context of 510(k) submission for substantial equivalence).
Magnetic Resonance Imaging (MRI) SafetyNon-clinical worst-case MRI review performed using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795).
Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque."Non-clinical worst-case MRI review was performed to evaluate the subject device components in the MRI environment using scientific rationale and published literature... rationale addressed parameters per the FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment, including magnetically induced displacement force and torque." (Implies favorable assessment supporting MRI safety, likely "MR Conditional" labeling although not explicitly stated here for all components). The document also states the results demonstrate the device is suitable for intended use.
Reverse EngineeringTo demonstrate physical compatibility with MegaGen AnyRidge Internal Implant connection. Testing performed on OEM implant bodies, OEM abutments, and OEM abutment screws."Reverse engineering of the MegaGen AnyRidge Internal Implant connection was performed to demonstrate physical compatibility. This testing was performed on OEM implant bodies, OEM abutments, and OEM abutment screws." (Implies successful demonstration of physical compatibility).
Mechanical PerformanceMechanical performance testing according to ISO 14801:2016, "Dentistry – Implants – Dynamic loading test for Endosseous Dental Implants."
Worst-case scenario chosen based on FDA guidance document "Root Form Endosseous Dental Implants and Endosseous Dental Implant Abutments.""The fatigue limit data demonstrated that constructs of the previously cleared Accelx abutments (K203240) and previously cleared MegaGen AnyRidge Internal Implant System OEM abutments in combination with Accelx implants have sufficient strength for their intended use."

Since the device described is not an AI/ML device, the following points are not applicable and thus, the document does not contain information on them:

  1. Sample sizes used for the test set and data provenance (e.g., country of origin of the data, retrospective or prospective): No clinical test set data from patients is referenced. The "test set" here refers to the physical units tested in non-clinical lab settings.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not applicable, as no ground truth for image interpretation or diagnosis is relevant here. The ground truth for mechanical testing is established by engineering standards and measurements.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Ground truth is based on engineering specifications, material science standards (ASTM, ISO), and validated test methods for physical properties.
  7. The sample size for the training set: Not applicable, as there is no AI/ML model for training.
  8. How the ground truth for the training set was established: Not applicable.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.