K Number
K242256
Manufacturer
Date Cleared
2024-10-29

(90 days)

Product Code
Regulation Number
866.3970
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QIAstat-Dx Meningitis/Encephalitis (ME) Panel is a qualitative multiplexed nucleic acid real-time PCR based in vitro diagnostic test intended for use with the QIAstat-Dx Analyzer 1.0. The QIAstat-Dx ME Panel is capable of simultaneous detection and identification of multiple bacterial, viral, and yeast nucleic acids from cerebrospinal fluid (CSF) specimens obtained via lumbar puncture from individuals with signs and/or symptoms of meningitis and/or encephalitis.

The following organisms are identified using the OlAstat-Dx ME Panel: Enterovirus, Escherichia coli K1, Haemophilus influenzae, Listeria monocytogenes, Neisseria meningitidis (encapsulated), Streptococcus agalactiae, Streptococcus pneumoniae, Streptococcus pyogenes, and Cryptococcus neoformans/gattii*.

The QIAstat-Dx ME Panel is indicated as an aid in the diagnosis of meningitis and/or encephalitis and results must be used in conjunction with other clinical, endemiological, and laboratory data. Results from the OlAstat-Dx ME Panel are not intended to be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with organisms not included in the QIAstat-Dx ME Panel. The agents detected may not be the definite cause of the disease. Negative results do not preclude central nervous system infection.

Not all agents of central nervous system infection are detected by this test and sensitivity in clinical use may differ from that described in the instructions for use.

The QIAstat-Dx ME Panel is not intended for testing specimens collected from indwelling central nervous system medical devices.

The QIAstat-Dx ME Panel is intended to be used in conjunction with standard of care culture for organism recovery, serotyping, and antimicrobial susceptibility testing.

*Cryptococcus neoformans and Cryptococcus gattii are not differentiated.

Device Description

The QIAstat-Dx® Meningitis/Encephalitis (ME) Panel is part of the QIAstat-Dx Meningitis/Encephalitis system and works with the OIAstat-Dx Analyzer 1.0.

The QIAstat-Dx ME Panel is intended to be used with cerebrospinal fluid (CSF) specimens.

Once the cartridge has been inserted into the instrument, the test starts automatically and runs for approximately 80 minutes. When the test is finished, the cartridge is removed by the user and discarded. The OIAstat-Dx Analyzer 1.0 automatically interprets test results and displays a summary on the analyzer display screen. The results can be printed using a connected printer, if needed. The detected analytes are displayed in red. For other analytes tested, they are displayed in green if not detected or in gray if not applicable or invalid. The analyzer will report if an error occurs during processing, in which case the test will fail and no results will be provided (screen will show "FAIL").

QIAstat-Dx consists of single-test cartridges with pre-packaged reagents including both wet and dry chemistry necessary to perform the sample preparation, nucleic acid amplification and detection to be used in conjunction with the QIAstat-Dx Analyzer 1.0. All sample preparation and assay steps are performed within the cartridge, so the user does not need to manipulate any reagent during the test. This eliminates exposure of the user or the Analyzer to chemicals contained in the cartridge during the test and up to the disposal of used cartridges.

Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer the sample and fluids via the Transfer Chamber (TC) to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:

  • Resuspension of air-dried internal control and Proteinase K (ProtK) enzyme using . provided buffer and mixing with the liquid sample (IC Cavity and ProtK Cavity);
  • Cell lysis using mechanical (rotation) and chemical (chaotropic and isotonic) ● means (lysis chamber):
  • Membrane-based nucleic acid purification from Lysate by: ●
    • Mixing lysate with binding buffer and capturing on the membrane -(purification chamber);
    • First washing of membrane to remove bound proteins (purification chamber and waste chamber);
    • Second washing of membrane to leave only bound nucleic acids -(purification chamber and waste chamber);
    • Rinsing of Transfer Chamber (TC) using the rinsing buffer before introduction of the eluate (Transfer Chamber);
  • Drying of membrane with bound nucleic acids with an air flow generated by a high flow vacuum pump (purification chamber); and
    • Elution of nucleic acids with elution buffer (purification chamber and TC);
  • Mixing of the purified nucleic acid (eluate) with lyophilized "Master Mix" reagents ● (Dry chemistry container (DCC) and TC);
  • Sequential transfer of defined aliquots of mixed eluate/Master Mix from the ● Transfer Chamber to each of eight Reaction Chambers containing the specified, airdried primers and probes;
  • Within each Reaction Chamber, real-time, multiplex PCR ("rtPCR") testing is ● performed. Increase in fluorescence (indicative of detection of each target analyte) is detected directly within each Reaction Chamber; and
  • The detected signal per fluorescent marker per Reaction Chamber is then used by the system software to generate the assay result.
AI/ML Overview

The provided document is a 510(k) Summary for the QIAGEN GmbH QIAstat-Dx Meningitis/Encephalitis (ME) Panel. Here's a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal acceptance criteria (i.e., predefined thresholds for sensitivity and specificity). However, it reports sensitivity (Positive Percent Agreement - PPA) and specificity (Negative Percent Agreement - NPA) values from its clinical performance study. The reported performance is implicitly the "met acceptance criteria" as the device received 510(k) clearance.

Here's a summary of the clinical performance for each pathogen from the prospective clinical study (Table 16), which represents the primary evidence for diagnostic performance:

PathogenSample CategoryPPA (TP/TP+FN) (%)95% CI (PPA)NPA (TN/TN+FP) (%)95% CI (NPA)
Escherichia coli K1Overall50.0 (2/4)15.0-85.0100.0 (1520/1520)99.7-100.0
Haemophilus influenzaeOverall80.0 (4/5)37.6-96.499.8 (1516/1519)99.4-99.9
Listeria monocytogenesOverall80.0 (4/5)37.6-96.4100.0 (1519/1519)99.7-100.0
Neisseria meningitidis (encapsulated)Overall100.0 (1/1)20.7-100.099.9 (1522/1523)99.6-100.0
Streptococcus agalactiaeOverall100.0 (3/3)43.9-100.0100.0 (1521/1521)99.7-100.0
Streptococcus pneumoniaeOverall100.0 (8/8)67.6-100.099.6 (1360/1365)99.1-99.8
Streptococcus pyogenesOverallN/A (0/0)N/A100.0 (1291/1291)99.7-100.0
Enterovirus (EV)Overall91.7 (22/24)74.2-97.799.9 (1499/1500)99.6-100.0
Cryptococcus gattii / Cryptococcus neoformansOverall57.1 (4/7)25.0-84.299.9 (1515/1517)99.5-100.0

Additionally, for contaminants that were confirmed by culture (fungal and bacterial), in table 19 and 20:

PathogenSample CategorySensitivity (compared to culture) (%)95% CI (Sensitivity)Specificity (compared to culture) (%)95% CI (Specificity)
Bacteria
Escherichia coli KlArchived100.0 (1/1)20.7-100.090.9 (10/11)62.3-98.4
Prospective Fresh50.0 (1/2)9.5-90.5100.0 (760/760)99.5-100.0
Haemophilus influenzaeArchived100.0 (1/1)20.7-100.090.9 (10/11)62.3-98.4
Prospective Frozen100.0 (3/3)43.9-100.099.7 (339/340)98.4-99.9
Listeria monocytogenesProspective Fresh100.0 (1/1)20.7-100.0100.0 (761/761)99.5-100.0
Prospective Frozen66.7 (2/3)20.8-93.9100.0 (340/340)98.9-100.0
Neisseria meningitidis (encapsulated)Archived100.0 (2/2)34.2-100.090.0 (9/10)59.6-98.2
Streptococcus agalactiaeProspective Fresh100.0 (1/1)20.7-100.099.9 (760/761)99.3-99.9
Prospective Frozen100.0 (1/1)20.7-100.0100.0 (342/342)98.9-100.0
Streptococcus pneumoniaeProspective Frozen100.0 (3/3)43.9-100.099.7 (339/340)98.4-99.9
Fungi / Yeast
Cryptococcus gattii / Cryptococcus neoformansArchived100.0 (2/2)34.2-100.0100.0 (1/1)20.7-100.0
Prospective Fresh100.0 (1/1)20.7-100.098.5 (129/131)94.6-99.6

2. Sample Sizes Used for the Test Set and Data Provenance

  • Test Set (Clinical Performance Study):
    • Prospective Specimens: 1524 evaluable specimens.
      • 552 (36.2%) were frozen before testing.
      • 972 (63.8%) were tested fresh.
    • Archived Specimens: 41 evaluable archived specimens (from an initial 195 collected).
    • Contrived Specimens: Not specified as a "test set" in the context of clinical performance, but used to supplement for rare analytes.
      • Ranges from 79 to 96 samples per pathogen, tested at 2xLoD and 5xLoD (e.g., Cryptococcus neoformans/gattii had 79 samples). These were likely individual spiked samples.
  • Data Provenance:
    • Country of Origin: 13 geographically diverse clinical sites across 4 countries (10 U.S. sites and 3 European sites).
    • Retrospective/Prospective: The study included both prospective (March 2022 to March 2023) and retrospective (archived) specimens.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of "experts" used to establish ground truth or their specific qualifications (e.g., radiologist with X years of experience).

Instead, the ground truth for the clinical performance study was established using:

  • An FDA-cleared molecular comparator method.
  • Two validated end point PCRs followed by bidirectional sequencing (BDS) for Streptococcus pneumoniae and Streptococcus pyogenes.
  • Standard of Care (SoC) testing, which varied across sites and included bacterial culture, Laboratory Developed PCR tests (LDT), FDA-cleared molecular methods, and Cryptococcus antigen screen and culture.
  • Discrepancy investigations were conducted for discordant results, implying a review process, but details on who performed this review are not given.

4. Adjudication Method for the Test Set

The document mentions that discrepancies between the QIAstat-Dx ME Panel and the comparator methods were investigated. This implies an adjudication process was in place to determine the true positive/negative status for discordant results. However, the specific method (e.g., 2+1, 3+1, none) is not explicitly described. The footnotes in Table 16 (Clinical Performance) provide details on how some discordant cases were resolved (e.g., "no organisms were detected with resolution method PCR/BDS," "negative result was confirmed positive with SoC culture and LDT result was positive").

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. This device is an in vitro diagnostic (IVD) test for nucleic acid detection, not an imaging AI device that assists human readers.

6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

Yes, a standalone performance study was done. The entire clinical performance study (Prospective, Archived, and Contrived Specimens Testing) evaluates the performance of the QIAstat-Dx ME Panel (the algorithm/device) directly against comparator methods (other molecular tests, culture, PCR/BDS), without a human-in-the-loop component. The device generates results automatically, and its accuracy is assessed based on these outputs.

7. Type of Ground Truth Used

The ground truth for the clinical performance studies was established using a combination of:

  • FDA-cleared molecular comparator method.
  • Validated end point PCRs followed by bidirectional sequencing (BDS).
  • Standard of Care (SoC) culture (for bacterial and fungal analytes).
  • Laboratory-Developed PCR tests (LDT).
  • Discrepancy investigations where discordant results were resolved using additional testing.

For the contrived specimens, the ground truth was known by design, as the samples were intentionally spiked with quantified strains.

8. Sample Size for the Training Set

The document does not provide details about a specific "training set" sample size for the QIAstat-Dx ME Panel. As an IVD based on real-time PCR, its "training" is typically in the form of analytical validation and optimization during its development, rather than machine learning model training with a distinct training dataset. The studies described are primarily for clinical performance validation, demonstrating the device's accuracy in a real-world setting.

9. How the Ground Truth for the Training Set Was Established

Since a dedicated "training set" for a machine learning model is not described, the concept of establishing ground truth for it is not applicable here in the conventional sense of AI/ML. The analytical validation studies (Limit of Detection, Analytical Reactivity/Inclusivity, Analytical Specificity/Exclusivity, etc.) involved known concentrations and strains of pathogens (e.g., ATCC strains, commercial stocks) in artificial or negative clinical CSF. The "ground truth" for these analytical studies was based on the known composition and concentration of these prepared samples.

{0}------------------------------------------------

October 29, 2024

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

QIAGEN GmbH % Sonia Pablo Pablo Senior Manager, Regulatory Affairs STAT-Dx Life, S.L. (A QIAGEN Company) Carrer Baldiri Reixac. 4 Barcelona, 08028 Spain

Re: K242256

Trade/Device Name: QIAstat-Dx Meningitis/Encephalitis (ME) Panel Regulation Number: 21 CFR 866.3970 Regulation Name: Device To Detect And Identify Microbial Pathogen Nucleic Acids In Cerebrospinal Fluid Regulatory Class: Class II Product Code: PLO Dated: July 29, 2024 Received: July 31, 2024

Dear Sonia Pablo Pablo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely, Bryan M. Grabias -S

Digitally signed by Bryan M. Grabias -S Date: 2024.10.29 15:52:56 -04'00'

Bryan Grabias Acting Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242256

Device Name QIAstat-Dx Meningitis/Encephalitis (ME) Panel

Indications for Use (Describe)

The QIAstat-Dx Meningitis/Encephalitis (ME) Panel is a qualitative multiplexed nucleic acid real-time PCR based in vitro diagnostic test intended for use with the QIAstat-Dx Analyzer 1.0. The QIAstat-Dx ME Panel is capable of simultaneous detection and identification of multiple bacterial, viral, and yeast nucleic acids from cerebrospinal fluid (CSF) specimens obtained via lumbar puncture from individuals with signs and/or symptoms of meningitis and/or encephalitis.

The following organisms are identified using the OlAstat-Dx ME Panel: Enterovirus, Escherichia coli K1, Haemophilus influenzae, Listeria monocytogenes, Neisseria meningitidis (encapsulated), Streptococcus agalactiae, Streptococcus pneumoniae, Streptococcus pyogenes, and Cryptococcus neoformans/gattii*.

The QIAstat-Dx ME Panel is indicated as an aid in the diagnosis of meningitis and/or encephalitis and results must be used in conjunction with other clinical, endemiological, and laboratory data. Results from the OlAstat-Dx ME Panel are not intended to be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with organisms not included in the QIAstat-Dx ME Panel. The agents detected may not be the definite cause of the disease. Negative results do not preclude central nervous system infection.

Not all agents of central nervous system infection are detected by this test and sensitivity in clinical use may differ from that described in the instructions for use.

The QIAstat-Dx ME Panel is not intended for testing specimens collected from indwelling central nervous system medical devices.

The QIAstat-Dx ME Panel is intended to be used in conjunction with standard of care culture for organism recovery, serotyping, and antimicrobial susceptibility testing.

*Cryptococcus neoformans and Cryptococcus gattii are not differentiated.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

General Information

Submitted by:QIAGEN GmbHQIAGEN Strasse 1Hilden, Germany 40724
Contact Person:Sonia PabloSenior Manager, Regulatory AffairsSTAT-Dx Life, S.L. (A QIAGEN Company)Carrer Baldiri Reixac, 408028 BarcelonaSpainPhone: +34 696 85 81 85Email: sonia.pablo@qiagen.com
Date Prepared:October 28, 2024
Device Name:QIAstat-Dx® Meningitis/Encephalitis (ME) Panel
Classification:21 CFR 866.3970 - Device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid.
Product Code:PLO
Predicate Device:
ManufacturerProduct NameDe Novo/510(k) No.
BioFire Diagnostics, LLCFilmArray Meningitis/Encephalitis (ME) PanelK160462

{5}------------------------------------------------

Device Description

The QIAstat-Dx® Meningitis/Encephalitis (ME) Panel is part of the QIAstat-Dx Meningitis/Encephalitis system and works with the OIAstat-Dx Analyzer 1.0.

The QIAstat-Dx ME Panel is intended to be used with cerebrospinal fluid (CSF) specimens.

Once the cartridge has been inserted into the instrument, the test starts automatically and runs for approximately 80 minutes. When the test is finished, the cartridge is removed by the user and discarded. The OIAstat-Dx Analyzer 1.0 automatically interprets test results and displays a summary on the analyzer display screen. The results can be printed using a connected printer, if needed. The detected analytes are displayed in red. For other analytes tested, they are displayed in green if not detected or in gray if not applicable or invalid. The analyzer will report if an error occurs during processing, in which case the test will fail and no results will be provided (screen will show "FAIL").

QIAstat-Dx consists of single-test cartridges with pre-packaged reagents including both wet and dry chemistry necessary to perform the sample preparation, nucleic acid amplification and detection to be used in conjunction with the QIAstat-Dx Analyzer 1.0. All sample preparation and assay steps are performed within the cartridge, so the user does not need to manipulate any reagent during the test. This eliminates exposure of the user or the Analyzer to chemicals contained in the cartridge during the test and up to the disposal of used cartridges.

Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer the sample and fluids via the Transfer Chamber (TC) to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:

  • Resuspension of air-dried internal control and Proteinase K (ProtK) enzyme using . provided buffer and mixing with the liquid sample (IC Cavity and ProtK Cavity);
  • Cell lysis using mechanical (rotation) and chemical (chaotropic and isotonic) ● means (lysis chamber):
  • Membrane-based nucleic acid purification from Lysate by: ●
    • Mixing lysate with binding buffer and capturing on the membrane -(purification chamber);
    • First washing of membrane to remove bound proteins (purification chamber and waste chamber);
    • Second washing of membrane to leave only bound nucleic acids -(purification chamber and waste chamber);
    • Rinsing of Transfer Chamber (TC) using the rinsing buffer before introduction of the eluate (Transfer Chamber);

{6}------------------------------------------------

  • Drying of membrane with bound nucleic acids with an air flow generated by a high flow vacuum pump (purification chamber); and
    • Elution of nucleic acids with elution buffer (purification chamber and TC);
  • Mixing of the purified nucleic acid (eluate) with lyophilized "Master Mix" reagents ● (Dry chemistry container (DCC) and TC);
  • Sequential transfer of defined aliquots of mixed eluate/Master Mix from the ● Transfer Chamber to each of eight Reaction Chambers containing the specified, airdried primers and probes;
  • Within each Reaction Chamber, real-time, multiplex PCR ("rtPCR") testing is ● performed. Increase in fluorescence (indicative of detection of each target analyte) is detected directly within each Reaction Chamber; and
  • The detected signal per fluorescent marker per Reaction Chamber is then used by the system software to generate the assay result.

Intended Use

The QIAstat-Dx® Meningitis/Encephalitis (ME) Panel is a qualitative multiplexed nucleic acid real-time PCRbased in vitro diagnostic test intended for use with the QIAstat-Dx Analyzer 1.0. The QIAstat-Dx ME Panel is capable of simultaneous detection and identification of multiple bacterial, viral, and yeast nucleic acids from cerebrospinal fluid (CSF) specimens obtained via lumbar puncture from individuals with signs and/or symptoms of meningitis and/or encephalitis.

The following organisms are identified using the QIAstat-Dx ME Panel: Enterovirus, Escherichia coli K1, Haemophilus influenzae, Listeria monocvtogenes, Neisseria meningitidis (encapsulated), Streptococcus agalactiae, Streptococcus pneumoniae, Streptococcus pyogenes, and Cryptococcus neoformans/gattii*.

The OIAstat-Dx ME Panel is indicated as an aid in the diagnosis of specific agents of meningitis and/or encephalitis and results must be used in conjunction with other clinical, epidemiological, and laboratory data. Results from the QIAstat-Dx ME Panel are not intended to be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with organisms not included in the QIAstat-Dx ME Panel. The agent or agents detected may not be the definite cause of the disease. Negative results do not preclude central nervous system (CNS) infection.

Not all agents of CNS infection are detected by this test and sensitivity in clinical use may differ from that described in the instructions for use.

The QIAstat-Dx ME Panel is not intended for testing of specimens collected from indwelling CNS medical devices.

The QIAstat-Dx ME Panel is intended to be used in conjunction with standard of care culture for organism recovery, serotyping, and antimicrobial susceptibility testing.

{7}------------------------------------------------

  • Cryptococcus neoformans and Cryptococcus gattii are not differentiated.

Comparison of the QIAstat-Dx ME Panel and the Predicate Device

Similarities and differences between the QIAstat-Dx ME Panel and the predicate device are shown in Table 1.

CharacteristicDevicePredicate
NameQIAstat-Dx ME PanelFilmArrayMeningitis/Encephalitis (ME)Panel
510(k) No.K242256K160462
Regulation21 CFR 866.397021 CFR 866.3970
Product CodePLOPLO
Device ClassClass IIClass II
Similarities
Intended UseThe QIAstat-Dx®Meningitis/Encephalitis (ME)Panel is a qualitative multiplexednucleic acid real-time PCR-basedin vitro diagnostic test intendedfor use with the QIAstat-DxAnalyzer 1.0. The QIAstat-DxME Panel is capable ofsimultaneous detection andidentification of multiplebacterial, viral, and yeast nucleicacids from cerebrospinal fluid(CSF) specimens obtained vialumbar puncture from individualswith signs and/or symptoms ofmeningitis and/or encephalitis.The following organisms areidentified using the QIAstat-DxME Panel:Bacteria:• Escherichia coli K1• Haemophilus influenzae• Listeria monocytogenesThe FilmArrayMeningitis/Encephalitis (ME)Panel is a qualitativemultiplexed nucleic acid-basedin vitro diagnostic test intendedfor use with FilmArray,FilmArray 2.0, and FilmArrayTorch systems. The FilmArrayME Panel is capable ofsimultaneous detection andidentification of multiplebacterial, viral, and yeastnucleic acids directly fromcerebrospinal fluid (CSF)specimens obtained via lumbarpuncture from individuals withsigns and/or symptoms ofmeningitis and/or encephalitis.The following organisms areidentified using the FilmArrayME Panel:Bacteria:• Escherichia coli K1• Haemophilus influenzae• Listeria monocytogenes
Table 1: Comparison of the QIAstat-Dx ME Panel with the predicate device
--------------------------------------------------------------------------------------

{8}------------------------------------------------

510(k) Premarket Notification

Page 5 of 35

CharacteristicDevicePredicate
• Neisseria meningitidis(encapsulated)• Neisseria meningitidis(encapsulated)
• Streptococcus agalactiae• Streptococcus agalactiae
• Streptococcus pneumoniae• Streptococcus pneumoniae
• Streptococcus pyogenes
Viruses:
Virus:• Enterovirus
• Enterovirus• Cytomegalovirus
• Herpes simplex virus 1
Yeast:• Herpes simplex virus 2
• Cryptococcusneoformans/gattii*• Human herpesvirus 6
• Human parechovirus
• Varicella zoster virus
The QIAstat-Dx ME Panel isindicated as an aid in thediagnosis of specific agents ofmeningitis and/or encephalitis andresults must be used inconjunction with other clinical,epidemiological, and laboratorydata. Results from the QIAstat-Dx ME Panel are not intended tobe used as the sole basis fordiagnosis, treatment, or otherpatient management decisions.Positive results do not rule out co-infection with organisms notincluded in the QIAstat-Dx MEPanel. The agent or agentsdetected may not be the definitecause of the disease. Negativeresults do not preclude centralnervous system (CNS) infection.Not all agents of CNS infectionare detected by this test andsensitivity in clinical use may
Yeast:
• Cryptococcusneoformans/gattii
The FilmArray ME Panel isindicated as an aid in thediagnosis of specific agents ofmeningitis and/or encephalitisandresults are meant to be used inconjunction with other clinical,epidemiological, and laboratorydata. Results from theFilmArray ME Panel are notintended to be used as the solebasis for diagnosis, treatment, orother patient managementdecisions. Positive results do notrule out co-infection withorganisms not included in theFilmArray ME Panel. The agentdetected may not be the definitecause of the disease. Negativeresults do not preclude centralnervous system (CNS)infection. Not all agents of CNSinfection are detected by thistest and sensitivity in clinicaluse may differ from that
CharacteristicDevicePredicate
differ from that described in theinstructions for use.described in the package insert.
The QIAstat-Dx ME Panel is notintended for testing of specimenscollected from indwelling CNSmedical devices.The FilmArray ME Panel is notintended for testing ofspecimens collected fromindwelling CNS medicaldevices.
The QIAstat-Dx ME Panel isintended to be used in conjunctionwith standard of care culture fororganism recovery, serotyping, andantimicrobial susceptibility testing.The FilmArray ME Panel isintended to be used inconjunction with standard ofcare culture for organismrecovery,serotyping, and antimicrobialsusceptibility testing.
* Cryptococcus neoformans andCryptococcus gattii are notdifferentiated.
Specimen TypeCerebrospinal FluidCerebrospinal Fluid
Analyte DetectedRNA/DNARNA/DNA
OrganismsDetectedSee aboveSee above
Amplification andDetectionTechnologyPCRPCR
Differences
Assay ControlsOne internal control in eachcartridge to control for sampleprocessing that is subjected to allnucleic acid extraction andamplification steps similar topatient samples. Labeling willrecommend use of negative andpositive external controlsregularly. Use transport media asthe external negative control, andpreviously characterized positivesamples or negative samplespiked with well characterizedtarget organisms as externalpositive controls.Two controls are included ineach reagent pouch to controlfor sample processing andboth stages of PCR and meltanalysis. Labeling recommendsthe use of external positive andnegative controls regularly.Enteric transport media can beused as an external negativecontrol, and previouslycharacterized positive samplesor negative samples spiked withwell characterized organisms asexternal positive controls.
CharacteristicDevicePredicate
Assay TargetsThe QIAstat-Dx ME Panel detects:Bacteria:• Streptococcus pyogenesThe FilmArray ME Panel detects:Viruses:• Cytomegalovirus• Herpes simplex virus 1• Herpes simplex virus 2• Human herpesvirus 6• Human parechovirus• Varicella zoster virus
Nucleic AcidExtractionTechnologyExtraction of nucleic acids using silica membraneExtraction of nucleic acids using standard magnetic beads
QIAstat-Dx detection of amplified targets uses an increase in fluorescence to generate the assay results.The FilmArray ME Panel uses melting curve analysis to confirm the identity of amplified targets to generate assay results.
OperationalThe sample is loaded straight into the cartridge.The FilmArray ME Panel must rehydrate and prime the cartridge before use.
Amplification andDetectionInstrumentSystemQIAstat-Dx Analyzer 1.0FilmArray Instruments

{9}------------------------------------------------

{10}------------------------------------------------

Performance Characteristics - Non-clinical Studies

The studies presented have been performed to demonstrate the non-clinical performance of the QIAstat-Dx ME Panel.

Limit of Detection

The Limit of Detection (LoD) is defined as the lowest concentration at which >95% of samples tested generate a positive call.

The LoD for each QIAstat-Dx ME Panel target was assessed, using 26 pathogen strains, by analyzing dilutions of analytical samples prepared from stocks obtained from commercial suppliers. The LoD was determined using selected strains representing individual pathogens that are possible to detect with the QIAstat- Dx ME Panel. All sample dilutions were prepared using artificial CSF. To confirm the established LoD concentration, the required detection rate of all replicates was ≥95%. Additional testing of samples prepared using negative clinical CSF was conducted to assess equivalency.

{11}------------------------------------------------

Individual LoD concentrations for each QIAstat-Dx ME Panel target are shown in Table 2.

PathogenStrainSource /Catalog IDLoDConcentration*DetectionRate
Escherichia coli KlStrain C5 [Bort];018ac:K1:H7ATCC7009733.48E+02 CFU/mL30/30
Escherichia coli K1NCTC 9001.Serovar 01:K1:H7ATCC117757.86E+02 CFU/mL30/30
Haemophilus influenzaetype b (cap)ATCC102113.16E+02 CFU/mL32/32
Haemophilus influenzaeType e [strain AMC36-A-7ATCC81422.54E+03 CFU/mL30/30
Listeria monocytogenesType 1/2bZeptoMetrix08015341.86E+03 CFU/mL21/21
Listeria monocytogenesType 4b. Strain Li 2ATCC191156.64E+03 CFU/mL30/30
Neisseria meningitidis(encapsulated)Serotype B. M2092ATCC130908.28E-02 CFU/mL31/32
Neisseria meningitidis(encapsulated)Serotype Y. M-112[BO-6]ATCC355611.33E+01 CFU/mL30/30
Streptococcus agalactiaeZ019ZeptoMetrix08015451.75E+03 CFU/mL30/30
Streptococcus agalactiaeG19 group BATCC138133.38E+03 CFU/mL31/31
Streptococcus pneumoniae19FZeptoMetrix08014397.14E+02 CFU/mL29/30
Streptococcus pneumoniaeSerotype 1. NCTC7465ATCC334006.22E-01 CFU/mL29/29
Streptococcus pyogenesZ472; Serotype M1ZeptoMetrix08043511.80E+03 CFU/mL30/30
Streptococcus pyogenesBruno [CIP104226]ATCC196159.10E+01CFU/mL30/30
Enterovirus ACoxsackievirusA16ZeptoMetrix0810107CF3.79E+00 TCID50/mL31/31
Enterovirus AA6, species A.Strain GdulaATCCVR-18011.60E+02 TCID50/mL30/30
Enterovirus BCoxsackievirus B5ZeptoMetrix0810019CF8.91E+01 TCID50/mL30/30
Enterovirus BCoxsackievirus A9,species BZeptoMetrix0810017CF4.36E+01 TCID50/mL28/29
Enterovirus CCoxsackievirusA17, species C.Strain G-12ATCCVR-10231.58E+01 TCID50/mL30/30
Enterovirus CCoxsackievirusA24. Starin DN-19ATCCVR-5834.99E+00 TCID50/mL30/30
PathogenStrainSource /Catalog IDLoDConcentration*DetectionRate
Enterovirus DEV 70, species D,strain J670/71ATCCVR-836$4.99E+01$ TCID50/mL30/31
Enterovirus DEnterovirus D68.Strain US/MO/14-18947ATCCVR-1823$5.06E+02$ TCID50/mL30/30
Cryptococcus neoformansSerotype D strainWM629, typeVNIVATCCMYA-4567$2.21E+03$ CFU/mL31/31
Cryptococcus neoformansC. neoformans H99ATCC208821$1.64E+02$ CFU/mL31/31
Cryptococcus gattiiSerotype B strainR272, type VGIIbATCCMYA-4094$1.32E+04$ CFU/mL30/30
Cryptococcus gattiiA6MR38 [CBS11545]ATCCMYA-4877$2.60E+03$ CFU/mL29/29

Table 2: LoD concentrations for the strains tested with QIAstat-Dx ME Panel

{12}------------------------------------------------

*Highest LoD concentration from LoD/Matrix Equivalency studies is reported.

Analytical Reactivity (Inclusivity)

The Inclusivity (analytical reactivity) study extended the list of pathogen strains tested during the QIAstat-Dx ME Panel Limit of Detection (LoD) study to confirm the reactivity of the detection system in the presence of different strains of the same organisms at a concentration near or above the respective Limit of Detection.

A variety of clinically relevant strains of each target organism of the OIAstat-Dx ME Panel representing organism sub-types, strains, serotypes, and genotypes of different temporal and geographic diversity of each analyte were included in the study. Analytical Reactivity was performed in two steps (in vitro and in silico).

Based on in vitro (wet) and in silico analysis, the QIAstat-Dx ME Panel has demonstrated a broad coverage for all targets detected by the panel.

In vitro analysis

A total of 130 strains covering 10 different analytical strains for each targeted organism or relevant species were tested spiked in combined samples. Strains tested for inclusivity are shown in Table 3.

125 out of 130 pathogen strains were successfully detected by the QIAstat-Dx ME Panel when tested in vitro. Five strains were not detected by the assay (detailed in Table 4).

PathogenSource(Catalog ID)Strain/SerotypeLowest ConcentrationTested
Escherichiacoli K1ATCC (700973)aStrain C5 [Bort]; O18ac:K1:H71x LoD
ATCC (11775)aNCTC 9001. Serovar O1:K1:H71x LoD
NCTC (9007)Strain Bi 7509/41; O7:K1:H-0.03x LoD
PathogenSource(Catalog ID)Strain/SerotypeLowest ConcentrationTested
HaemophilusinfluenzaeATCC (23509)NCDC Bi 7509-41 Serotype 07:K1(L) :NM1:100 from stock
ATCC (23511)NCDC F 11119-410.03x LoD
BEI Resources (NR-17666)O-2, U9-411:100 from stock
BEI Resources (NR-17674)O-16, F1119-411:100 from stock
ZeptoMetrix (0801905)Z136 CTX-M-153x LoD
NCTC (11101)Sc15 02:K1:H63x LoD
NCTC (9045)Strain H61; O45:K1:H103x LoD
ATCC (10211) atype b (cap)1x LoD
ATCC (8142) aType e [strain AMC 36-A-7]1x LoD
ATCC (51907)Non-typeable [strain Rd [KW20]0.014x LoD
ATCC (11116)Non-typeable [strain 180-a]3x LoD
ATCC (9006)Type a [strain AMC 36-A-3]3x LoD
ATCC (31512)Type b [strain Rab]0.016x LoD
ATCC (49699)Type c [strain C 9007]0.016x LoD
ATCC (9008)Type d [strain AMC 36-A-6]0.014x LoD
ATCC (700223)Type f [strain GA-1264]0.016x LoD
ATCC (49766)L-3783x LoD
ListeriamonocytogenesZeptoMetrix (0801534) aType 1/2b1x LoD
ATCC (19115) aType 4b. Strain Li 21x LoD
ATCC (BAA-2659)Type 1/2a. Strain 2011L-26763x LoD
ATCC (19111)Type 1/2a. Strain Li 203x LoD
ZeptoMetrix (0804339)Type 4b0.03x LoD
ATCC (13932)serotype 4b. Strain 1071/53 [LMG21264, NCTC 10527]0.011x LoD
ATCC (19114)Li 23. Serotype 4a0.02x LoD
BEI Resources (NR-13237)FSL J2-0641:100 from stock
ATCC (7644)Gibson2x LoD
ATCC (BAA-679)EGDe0.026x LoD
Neisseriameningitidis(encapsulated)ATCC (13090) aSerotype B. M2092 [CIP 104218,L. Cunningham]1x LoD
ATCC (35561) aSerotype Y. M-112 [BO-6]1x LoD
ATCC (13077)Serogroup A, M1027[NCTC10025]0.03x LoD
ATCC (13102)Serogroup C, M16280.03x LoD
ATCC (13113)Serotype D. M158 [37A]3x LoD
IDT (gBlock 77859371)bsequence with variant ctrA gene1:1000 from stock
ATCC (43744)W1350.03x LoD
ATCC (BAA-335)MC580.03x LoD
StreptococcusagalactiaeATCC (23255)79 Eur. Serogroup B1:1000 from stock
ATCC (13092)Serotype B. M997 [S-3250-L]1:1000 from stock
ZeptoMetrix (0801545) aZ0191x LoD
ATCC (13813) aG19 group B1x LoD
ATCC (12403)Serotype III. Typing strainD136C(3) [3 Cole 106, CIP 82.45]1.2x LoD
ATCC (BAA-611)2603 V/R. Serotype V0.013x LoD
PathogenSource(Catalog ID)Strain/SerotypeLowest ConcentrationTested
ATCC (31475)type III-ST2833x LoD
BEI Resources (NR-43898)MNZ9291:100 from stock
ATCC (12401)Typing strain H36B - type Ib1.4x LoD
ATCC (27591)CDC SS700 [A909; 5541], type 1c0.014x LoD
ATCC (49446)3139 [CNCTC 1/82] Serotype IV0.0127x LoD
ZeptoMetrix (0801556)Z0233x LoD
ZeptoMetrix (0801439) a19F1x LoD
ATCC (33400) aSerotype 1. NCTC 74651x LoD
StreptococcuspneumoniaeATCC (BAA-334)Serotype 4. TIGR4 [JNR.7/87]0.03x LoD
ATCC (BAA-341)Serotype 5. SPN1439-106[Colombia 5-19]0.03x LoD
ATCC (10343)Serotype 11A. Type 433x LoD
ATCC (700672)Serotype 14. VH140.03x LoD
ATCC (700673)Serotype 19A. Hungary 19A-6[HUN663]0.026x LoD
ZeptoMetrix (0804016)Z319; 12F3x LoD
ATCC (6303)Diplococcus pneumoniae; Type 3.Strain [CIP 104225]3x LoD
ATCC (BAA-661)DCC1476 [Sweden 15A-25]0.03x LoD
ZeptoMetrix (0804351) aZ472; Serotype M11x LoD
ATCC (19615) aBruno [CIP 104226]1x LoD
ZeptoMetrix (0801512)Z018; Serotype M583x LoD
StreptococcuspyogenesATCC (BAA-947)Serotype M1. MGAS 50050.03x LoD
ATCC (14289)Lancefield's group A / C203 S3x LoD
ATCC (12203)NCTC 8709 (Type 6 glossy)0.03x LoD
ATCC (12353)Group a, type 12. Typing strainT12 [F. Griffith SF 42]1:1000 from stock
ATCC (12972)Group a, type 140.03x LoD
ATCC (8133)Group a, type 233x LoD
ATCC (12384)C203 -Type 30.029x LoD
ZeptoMetrix (0810107CF)Coxsackievirus A161x LoD
Enterovirus AATCC (VR-1801) aA6, species A. Strain Gdula1x LoD
ATCC (VR-168)A10. Μ.Κ. (Kowalik)3x LoD
ATCC (VR-1432)Enterovirus 71. Strain H3x LoD
ZeptoMetrix (0810236CF)Species A, Serotype EV-A71(2003 Isolate)1:100 from stock
BEI Resources (NR-471)Tainan/4643/19983x LoD
ATCC (VR-1550)A2 F1 [Fleetwood]3x LoD °
ATCC (VR-673)A7 - 275/583x LoD
ATCC (VR-170)A12 – Texas 121:100 from stock
ATCC (VR-1775)EV-A71. Strain BrCr3x LoD
Enterovirus BZeptoMetrix (0810019CF)Coxsackievirus B51x LoD
ZeptoMetrix (0810017CF) aCoxsackievirus A9, species B1x LoD
ATCC (VR-28)Species B, Serotype CV-B1, StrainConn-53x LoD
ATCC (VR-29)Species B, Serotype CV-B2. StrainOhio-13x LoD
ZeptoMetrix (0810075CF)Coxsackievirus B41:100 from stock
ZeptoMetrix (0810076CF)Echovirus 61:100 from stock
ZeptoMetrix (0810077CF)Echovirus 91:100 from stock
ZeptoMetrix (0810074CF)Coxsackievirus B31:10000 from stock
PathogenSource(Catalog ID)Strain/SerotypeLowest ConcentrationTested
NCPV (0901047v)Echovirus 183x LoD
ATCC (VR-41)Species B, Serotype E-113x LoD
Enterovirus CATCC (VR-1023)aCoxsackievirus A17, species C.Strain G-121x LoD
ATCC (VR-583)aCoxsackievirus A24. Starin DN-191x LoD
ATCC (VR-850)Coxsackievirus A21. StrainKuykendall [V-024-001-012]3x LoDc
ATCC (VR-169)A11 - Belgium-13x LoD
ATCC (VR-1488)A13 - Flores3x LoD
ATCC (VR-182)A22 - Chulman1:100 from stock
ATCC (VR-178)A20 - IH Pool 351:100 from stock
ATCC (VR-176)A18 - G-131:100 from stock
NCTC (0812075v)CV-A21. Strain H06452 4723x LoD
NCTC (0812074v)CV-A21. Strain H06418 5080.03x LoD
ATCC (VR-836)aEV 70, species D, strain J670/711x LoD
ATCC (VR-1823)aEnterovirus D68. StrainUS/MO/14-189471x LoD
ZeptoMetrix (0810237CF)Enterovirus 68. 2007 Isolate0.03x LoD
ATCC (VR-1824)Enterovirus D68. Strain US/IL/14-189523x LoD
ATCC (VR-1197)D68. Strain F02-3607 Corn3x LoD
Enterovirus DZeptoMetrix (0810302CF)Type 68 Major Group (09/2014Isolate 2)1:100 from stock
ATCC (VR-1825)Enterovirus D68. StrainUS/KY/14-189533x LoD
ATCC (VR-1826)Enterovirus D68. Strain Fermon3x LoD
BEI Resources (NR-49130)Enterovirus D68. US/MO/14-189493x LoD
BEI Resources (NR-51998)Enterovirus D68. USA/2018-230893x LoD
ATCC (MYA-4567)aSerotype D strain WM629, typeVNIV1x LoD
ATCC (208821)aH991x LoD
ATCC (32045)type strain, CBS 1323x LoD
ATCC (MYA-4564)Serotype A strain WM148, typeVNI1:1000 from stock
ATCC (13690)M20921:100 from stock
CryptococcusneoformansATCC (MYA-4566)Serotype AD strain WM628, typeVNIII1:1000 from stock
ZeptoMetrix (0801803)Serotype A3x LoD
BEI Resources (NR-50335)NIH9hi901:100 from stock
BEI Resources (NR-50332)NIH3061:100 from stock
BEI Resources (NR-48776)Var grubiiYL99α1:100 from stock
ATCC (MYA-4094)aSerotype B strain R272, typeVGIIb1x LoD
ATCC (MYA-4877)aA6MR381x LoD
CryptococcusgattiiATCC (MYA-4560)Serotype B strain WM179, typeVGI1:100 from stock
ATCC (MYA-4562)Serotype B strain WM161, typeVGIII1:1000 from stock
ATCC (MYA-4563)Serotype C strain WM779, typeVGIV1:1000 from stock
ATCC (MYA-4138)A1M R2653x LoD
PathogenSource(Catalog ID)Strain/SerotypeLowest ConcentrationTested
ATCC (14248)110 [CBS 883]1:1000 from stock
BEI Resources (NR-50184)AIR2651:100 from stock
BEI Resources (NR-50195)Alg1661:100 from stock
BEI Resources (NR-50198)Alg2541:100 from stock

Table 3: Strains tested for Inclusivity

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

a Strains tested and evaluated during the Limit of Detection studies.

b Commercial artificial dsDNA fragment (gBlock) including ctrA gene sequence was tested due to unavailability of analytical sample for this variant (GenBank Accession ID HO156899).

6 Higher concentration tested to meet 100% detection rate (30xLoD for ATCC (VR-1550, VR-850 and VR-41) and 3xLoD for ZeptoMetrix (0804339)).

PathogenStrain/Serotype
Escherichia coli K1NCDC Bi 7509-41 Serotype 07:K1(L):NM
Escherichia coli K1Z136 CTX-M-15
Enterovirus CCV-A21. Strain H06452 472
Enterovirus CCV-A21. Strain H06418 508
Streptococcus agalactiaeSerotype III. Typing strain D136C(3) [3 Cole 106, CIP 82.45]

Table 4: Inclusivity Strains not detected by the QIAstat-Dx ME Panel

In silico analysis

An in silico prediction of the reactivity of all primers-probe oligonucleotide sequences included in the panel against sequences publicly available in databases was performed. The analysis demonstrated a complete coverage for the detection of all main subtypes (species and serotypes) for every On-Panel target organism included in the QIAstat-Dx ME Panel (Table 5).

PathogensClinically relevant strains/subtypes detected
Neisseria meningitidis(encapsulated)Encapsulated serotypes (A, B, C, D, E, H, I, K, L, NG, W, W135, X,Y, Z, 29E).
Cryptococcus gattii/neoformansSerotype A ( C. neoformans var neoformans ), serotype D ( C.neoformans var grubii ), serotpyes B and C ( C. gattii including allVGI,VGII, VGIII, VGIV molecular types).
Listeria monocytogenesSerotypes 1/2a, 1/2b, 1/2c, 3a, 3b, 3c, 4a, 4b, 4c, 4d, 4e, 7.
Haemophilus influenzaeAll encapsulated serotypes (a, b, c, d, e, f) and unencapsulated strains(non-typeable, NTHi) including var. H. Aegyptus
EnterovirusCoxsackievirus A (CV-A1 through CV-A24), coxsackievirus B (CV-B1 through CV-B6), Echovirus (E-1 through E-33), Enterovirus A(EV-A71, EV-A76, EV-A89 through EV-A92, EV-A119, EV-A120),Enterovirus B (EV-B69, EV-B73 through EV-B75, EV-B79, EV-B80through EV-B88, EV-B93, EV-B97, EV-B98, EV-B100, EV-B101,EV-B106, EV-B107, EV-B111), Enterovirus C (EV-C96, EV-C99,EV-C102, EV-C104, EV-C105, EV-C109, EV-C116 through EV-C118), Enterovirus D (EV-D68, EV-D70, EV-D94), Poliovirus (PV-1through PV-3).
Escherichia coli K1K1 strains

Table 5: Organisms with predicted reactivity based on in silico analysis

{17}------------------------------------------------

PathogensClinically relevant strains/subtypes detected
Rest of On-Panel organismwith no biologicalsubclassification (S.pneumoniae, S. agalactiae, S.pyogenes)All genomic sequences available in databases detected

Analytical Specificity (Exclusivity)

The analytical specificity study was carried out by in vitro testing and in silico analysis to assess the potential cross-reactivity and exclusivity of the QIAstat-Dx ME Panel. On-Panel organisms were tested to assess the potential for intra-panel cross-reactivity and Off-Panel organisms were tested to evaluate cross-reactivity with organisms not covered by the panel content (panel exclusivity). The Off-Panel organisms were selected because they are clinically relevant (colonize the central nervous system or cause meningitis and/or encephalitis symptoms), are common skin flora or laboratory contaminants, are genetically similar to On-Panel analytes, or are microorganisms for which much of the population may have been infected.

After testing several On-Panel (13) and Off-Panel (115) strains, the QIAstat-Dx ME Panel demonstrated high specificity levels for all the rtPCR assays included in the device. Additional bioinformatic prediction analysis supported the specificity level, with few exceptions.

In vitro analysis

Samples were prepared by spiking potential cross-reactive organisms into artificial CSF matrix at 105 TCIDso/ml for viral targets, 105 CFU/mL for fungi / yeast target, and 100 CFU/mL for bacterial and fungi / yeast target, or the highest concentration possible based on the organism stock. The On-Panel and Off-Panel organisms tested are shown in Table 6.

All On-Panel pathogens resulted in specific detection, and all Off-Panel pathogens tested showed a negative result and no cross-reactivity was observed in the QIAstat-Dx ME Panel, except for the pathogens shown in Table 7.

TypePathogenStrain/Serotype
On-Panel
BacteriaEscherichia coli K1Strain C5 [Bort];O18ac:K1:H7
BacteriaHaemophilus influenzaeType e [strain AMC 36-A-7]
BacteriaListeria monocytogenesType 4b. Strain Li 2
Table 6: Off-Panel and On-Panel Analytical strains tested during Exclusivity

{18}------------------------------------------------

TypePathogenStrain/Serotype
On-Panel
BacteriaNeisseria meningitidisSerotype Y. M-112 [BO-6]
BacteriaStreptococcus pneumoniae19F
BacteriaStreptococcus agalactiaeZ019
BacteriaStreptococcus pyogenesZ472; Serotype M1
VirusEnterovirus A1A6, species A.Strain Gdula
VirusEnterovirus B1Coxsackievirus B5
VirusEnterovirus C1Coxsackievirus A17,species C. Strain G-12
VirusEnterovirus D1Enterovirus D68. StrainUS/MO/14-18947
YeastCryptococcus neoformans1WM629 [CBS 10079]
YeastCryptococcus gattii1Serotype B strain R272,type VGIIb
Off-Panel
VirusAdenovirus A12Huie
VirusAdenovirus C2Adenoid 6 (NIAID 202-001-014)
VirusAdenovirus D20A.A
VirusAdenovirus E4RI-67
VirusAdenovirus F41Tak
VirusBK polyoma virusN/A
VirusCoronavirus 229E229E
VirusCoronavirus NL63NL63 (Amsterdam I)
VirusCoronavirus OC43OC43
VirusDengue virus (Type 2)New Guinea C
VirusEpstein-Barr VirusB95-8
VirusHepatitis B virus (HBV)N/A
VirusHepatitis C virus (HCV)N/A
VirusHuman herpes virus 7SB
VirusHuman herpes virus 8N/A
VirusHuman ImmunodeficiencyVirusQuantitative SyntheticHuman immunodeficiencyvirus 1 (HIV-1) RNA
VirusHuman Rhinovirus A1b2060
VirusHuman Rhinovirus A1611757
VirusHuman Rhinovirus B3FEB
VirusHuman Rhinovirus B83Baylor 7 [V-190-001-021]
VirusInfluenza A H1N1A/Florida/3/2006
TypePathogenOff-PanelStrain/Serotype
VirusInfluenza A H1N1-2009A/California/08/2009(H1N1pdm)
VirusInfluenza A H3N2A/Port Chalmers/1/73
VirusInfluenza BB/Virginia/ATCC4/2009
VirusJC polyoma virusMAD-4
VirusMeasles VirusEdmonston
VirusMumps VirusJones
VirusWest Nile Virus1986
VirusParainfluenza virus 2Greer
VirusParainfluenza virus 4N/A
VirusParvovirus B19B19
VirusRespiratory Syncytial VirusA2
VirusRotavirusRRV (Rhesus Rotavirus)
VirusRubella VirusN/A
VirusSt. Louis EncephalitisVirus 1Parton
VirusCytomegalovirusDavis
VirusHerpes simplex virus 1Macintyre
VirusHerpes simplex virus 2HSV-2. (Strain: MS)
VirusHuman herpes virus 6HHV-6B. (Strain: Z29)
VirusHuman parechovirusSerotype 3
VirusVaricella-zoster virusEllen
Fungi/parasiteCandida glabrataCBS 138
Fungi/parasiteCandida kruseiN/A
Fungi/parasiteCandida lusitaniaeZ010
Fungi/parasiteCandida metapsilosisMCO429
Fungi/parasiteCandida orthopsilosisMCO471
Fungi/parasiteCandida viswanathiiPK 233 [NCYC 997,pK233]
Fungi/parasiteCandida parapsilosisCBS 604
Fungi/parasiteCandida tropicalisVitek #8935
Fungi/parasiteCryptococcus albidusAmMS 228
Fungi/parasiteCryptococcus amylolentusNRRY Y-7784
Fungi/parasiteCryptococcus laurentiiCBS 139
Fungi/parasiteCryptococcus uniguttulatusAmMS 234
Fungi/parasiteCryptococcus adeliensis =Cryptococcus adeliae =Naganishia adeliensisCryptococcus adeliae
Fungi/parasiteCryptococcus flavescens =Papiliotrema flavescensCryptococcus laurentii var.Flavescens (Saito)
TypePathogenStrain/Serotype
Off-Panel
Fungi/parasiteCryptococcus wingfieldii =Tsuchiyaea wingfieldiiOTU 26
Fungi/parasiteCryptococcus depauperatus= Aspergillus depauperatus= FilobasidielladepauperateK [ARSEF 2058, CBS7842]
Fungi/parasiteFilobasidium capsuligenumML-186
Fungi/parasiteNaegleria fowleriGenomic DNAfrom Naegleria fowleri
Fungi/parasiteToxoplasma gondiiHaplogroup 2
Fungi/parasiteAspergillus fumigatusZ014
Fungi/parasiteCandida albicansCBS 562
Fungi/parasiteCandida dubliniensisZ145
BacteriaBacillus cereusZ091
BacteriaCitrobacter freundii[ATCC 13316, NCTC9750]
BacteriaCorynebacterium striatumCDC F6683
BacteriaCorynebacteriumurealyticus3 [Garcia strain]
BacteriaCronobacter(Enterobacter) sakazakiiCDC 4562-70
BacteriaEnterobacter aerogenesZ052
BacteriaEnterobacter cloacaeCDC 442-68
BacteriaEscherichia coli (non-K1)2003-3055
BacteriaEscherichia fergusoniiZ302
BacteriaEscherichia hermanniiCDC 980-72
BacteriaEscherichia vulnerisCDC 875-72
BacteriaHaemophilus ducreyiDCC1476 [Sweden 15A-25]
BacteriaHaemophilus haemolyticusNCTC 10659
BacteriaHaemophilusparahaemolyticus536 [NCTC 8479]
BacteriaHaemophilusparainfluenzaeNCTC 7857
BacteriaKlebsiella pneumoniaeNCTC 9633 [NCDC 298-53,NCDC 410-68]
BacteriaListeria innocuaSLCC 3379
BacteriaListeria ivanoviiLi 1979
BacteriaMorganella morganiiAM-15
BacteriaStreptococcus salivariusC699
BacteriaStreptococcus sanguinisDSS-10
TypePathogenStrain/Serotype
Off-Panel
BacteriaStreptococcuspseudopneumoniaeCDC-SS-1757
BacteriaMycoplasma genitaliumM30
BacteriaNeisseria lactamicaNCDC A7515
BacteriaNeisseria mucosaAmMS 138
BacteriaNeisseria siccaAMC 14-D-1
BacteriaNeisseria gonorrhoeaeZ017
BacteriaPantoea agglomeransEnterobacter agglomerans
BacteriaProprionibacterium acnesNCTC 737
BacteriaProteus mirabilisLRA 08 01 73 [API SA,DSM 6674]
BacteriaPseudomonas aeruginosaPRD-10 [CIP 103467,NCIB 10421, PCI 812]
YeastSaccharomyces cerevisiaeNRRL Y-567
BacteriaSalmonella bongoriCIP 82.33
BacteriaSalmonella entericaCDC K-1891[ATCC 25928]
BacteriaSerratia marcescensPCI 1107
BacteriaShigella boydiiCDC C-123
BacteriaShigella flexneriZ046
BacteriaShigella sonneiAMC 43-GG9
BacteriaStaphylococcus aureusFDA 209
BacteriaStaphylococcus capitisPRA 360 677
BacteriaStaphylococcus epidermidisFDA strain PCI 1200
BacteriaStaphylococcushaemolyticusSM 131
BacteriaStaphylococcus hominisZ031
BacteriaStaphylococcus lugdunensisLRA 260.05.79
BacteriaStaphylococcussaprophyticusNCTC 7292
BacteriaStreptococcus anginosusNCTC 10713
BacteriaStreptococcus bovisZ167
BacteriaStreptococcus dysgalactiaeGrouping strain C74
BacteriaStreptococcus intermediusZ126
BacteriaStreptococcus oralisZ307
BacteriaStreptococcus mitis(tigurinus)Clinical Isolate
BacteriaStreptococcus mutansLRA 28.02.81

{19}------------------------------------------------

Page 16 of 35

{20}------------------------------------------------

QIAGEN QI KOLE :

QIAstat-Dx® Meningitis/Encephalitis (ME) Panel
510(k) Summary

Page 17 of 35

{21}------------------------------------------------

Page 18 of 35

1 Enterovirus A, B, C, D and Cryptococcus neoformans/gattii are detected by their specific rtPCR assay with no species discrimination.

2 N. meningitidis was tested at 1.00E+05 CFU/mL instead of 1.00E+06 CFU/mL.

{22}------------------------------------------------

QIAstat-Dx ME Panel TargetPotential cross-reactive organismCross reactiveconcentration
Haemophilus influenzaeHaemophilus haemolyticus≥1.00E+03 CFU/mL
Cryptococcus neoformans/gattiiCryptococcus wingfieldii = Tsuchiyaea wingfieldii≥1.00E+01 CFU/mL
Cryptococcus flavescens = Papiliotrema flavescens≥4.00E+03 CFU/mL
Cryptococcus neoformans/gattiiCryptococcus amylolentus≥1.00E+01 CFU/mL

Table 7: Samples showing cross-reactivity with the panel

In silico analysis

In silico analysis was performed, for the primer/probe designs included in the QIAstat-Dx ME Panel, in two steps to further characterize the sequence specificity of the rtPCR assays included in the QIAstat-Dx ME Panel to detect possible unspecific homologies and/or unspecific cross-reactions.

The result of the analysis performed for the primer/probe designs included in the QIAstat-Dx ME Panel pointed at 6 potential cross-reactions with Off-Panel targets (listed in Table 8).

Table 8: Samples showing cross-reactivity with the panel
Off-Panel organismOn-Panel signal
Streptococcus pseudopneumoniae *Streptococcus pneumoniae
Listeria innocua *Listeria monocytogenes
Haemophilus haemolyticusHaemophilus influenzae
Cryptococcus amylolentusCryptococcusneoformans/gatti
Cryptococcus depauperatus *
Cryptococcus wingfieldii
  • in silico cross-reactive risk was not confirmed by in vitro testing.

Interfering Substances

The effect of potentially interfering substances on the detectability of the QIAstat-Dx ME Panel organisms was evaluated. The substances tested in the study included endogenous as well as exogenous substances that are commonly found and/or introduced into CSF specimens during specimen collection.

The QIAstat-Dx ME Panel target organisms were tested at 3x LoD in artificial CSF matrix and testing was performed in triplicate. Potential interfering substances were spiked into

{23}------------------------------------------------

the samples at a level predicted to be above the concentration of the substance likely to be found in CSF sample.

Potentially interfering endogenous and exogenous substances have been evaluated and have been confirmed not to interfere with any of the panel target assays at concentrations potentially found in clinical samples (above the concentration of the substance likely to be found in a real CSF specimen). As an exception, bleach was shown to cause inhibition at high concentrations (1% v/v, 0.1% v/v), and showed no interference when tested at 0.01% v/v. Results are provided in Table 9.

Table 9: Final highest concentrations per potentially interfering substance without
observable interference
SubstanceConcentrationTestedResult
Endogenous substances
Human Blood10% (v/v)No interference
gDNA20 µg/mLNo interference
D(+)Glucose10 mg/mLNo interference
L-lactate (Na)2.2 mg/mLNo interference
Immunoglobulin G (human)20 mg/mLNo interference
Albumin (human)30 mg/mLNo interference
Peripheral blood mononuclear cells10,000 cells/µLNo interference
Exogenous substances
Chlorhexidine0.4% (w/v)No interference
Ethanol7.0% (v/v)No interference
Bleach1.0% (v/v)Interference
Bleach0.1% (v/v)Interference
Bleach0.01% (v/v)No interference
Acyclovir69 µg/mLNo interference
Amphotericin B5.1 µg/mLNo interference
Ampicillin210 µg/mLNo interference
Ceftriaxone840 µg/mLNo interference
Cefotaxime645 µg/mLNo interference

Microbial Interference

A microbial interference study was conducted to assess the inhibitory effects of select nontarget organisms on the ability to detect the QIAstat-Dx ME Panel target organisms. Challenging concentrations (105 units/ml for viral targets and 106 CFU/mL for bacterial targets) of non-target organisms were individually mixed with artificial CSF matrix containing spiked targeted QIAstat-Dx ME Panel organisms at 3x LoD. Testing was performed in triplicate. All QIAstat-Dx ME Panel organisms were successfully detected

{24}------------------------------------------------

(100% hit rate) when tested in combination with the potentially microbial interferents. See Table 10 for a list of the non-target organisms tested and the result summary.

SubstanceSupplier / Catalog #ConcentrationTestedResult
Epstein-Barr virusZeptoMetrix,0810008CF$1E+05$ cp/mLNo interference
Influenza A H1N1- 2009ATCC, VR-1895$1E+05$ CEID50/mLNo interference
Cutibacterium acnesATCC, 6919$1E+06$ CFU/mLNo interference
Staphylococcus epi-dermidisATCC, 14990$1E+06$ CFU/mLNo interference
Escherichia coli (non-K1)ATCC, 25922$1E+06$ CFU/mLNo interference
Staphylococcus aureusATCC, 29213$1E+06$ CFU/mLNo interference
Measles VirusATCC, VR-24$1E+05$ TCID50/mLNo interference

Table 10: Final highest concentration without observable inhibitory effect

Competitive Inhibition

Combined samples containing a mixture of two different targets spiked at low and high concentrations into artificial CSF were tested. Selection of bacteria, virus, and yeast pathogens and combinations of targets tested was based on clinical relevance.

Clinically relevant co-infection testing demonstrated that when at least two QIAstat-Dx ME Panel pathogens of different concentrations are simultaneously present in one sample, all targets can be detected by the assay. A summary of the final co-infection mixes whereby the High Positive Analyte (HPA) does not inhibit the Low Positive Analyte (LPA) is shown in Table 11.

LPAHPA
PathogenConcentrationUnitsPathogenConcentrationUnits
Escherichia coli K13.30E+02CFU/mLHaemophilus influenzae1.00E+06CFU/mL
Haemophilusinfluenzae9.48E+02CFU/mLEscherichia coli K11.00E+06CFU/mL
Haemophilusinfluenzae9.48E+02CFU/mLStreptococcus pneu-moniae1.00E+06CFU/mL
Streptococcus pneu-moniae6.78E+02CFU/mLHaemophilus influenzae1.00E+06CFU/mL
Listeriamonocytogenes5.58E+03CFU/mLStreptococcus pneu-moniae1.00E+06CFU/mL
Streptococcus pneu-moniae6.78E+02CFU/mLListeria monocytogenes1.00E+06CFU/mL
Cryptococcus neo-formans6.63E+03CFU/mLStreptococcus pneu-moniae1.00E+06CFU/mL
Streptococcus pneu-moniae6.78E+02CFU/mLCryptococcus neo-formans1.00E+05CFU/mL
Neisseria meningitidis3.99E+01CFU/mLHaemophilus influenzae1.00E+06CFU/mL

Table 11: Summary of competitive inhibition testing results

{25}------------------------------------------------

LPAHPA
PathogenConcentrationUnitsPathogenConcentrationUnits
Enterovirus4.80E+02TCID50/mLStreptococcuspyogenes1.00E+06CFU/mL
Streptococcuspyogenes1.71E+03CFU/mLEnterovirus1.00E+05TCID50/mL

Carrvover

A carryover study was performed to evaluate the potential occurrence of crosscontamination between consecutive runs when using the QIAstat-Dx ME Panel on the QIAstat-Dx Analyzer 1.0. Pathogenic CSF samples with alternating high-positive (10-10° organism/mL) and negative samples, were conducted on two OIAstat-Dx Analyzer 1.0 instruments.

No carryover between samples was observed in the OIAstat-Dx ME Panel, demonstrating that the system design and recommended sample handling and testing practices are effective in preventing unexpected results due to carryover or cross-contamination between samples.

Sample Stability

Sample stability testing demonstrated that the OlAstat-Dx ME Panel assay is capable of processing samples (clinical cerebrospinal fluid specimens) that are stored at different temperatures without affecting the performance.

Sample stability testing demonstrated that CSF specimens may be stored at the conditions listed below.

  • Room temperature up to 24 hours at 15-25℃ -
  • -Refrigerated up to 7 days at 2-8℃

Matrix Equivalency

A matrix equivalency study was conducted to compare the performance of analytical samples prepared in negative clinical CSF matrix to samples prepared in artificial CSF matrix.

A total of 26 pathogen strains (at least two different strains per panel target) were tested in combined samples spiked in clinical true-negative CSF. Clinical CSF was sourced from commercial suppliers and was tested prior to its use in mix manufacture for true-negativity with either the QIAstat-Dx ME Panel or an alternative method.

Samples were prepared by spiking pathogens in combined (multiple-spike) sample mixes containing up to 4 organisms per sample in clinical CSF at 5x LoD, 1x LoD and 0.1x LoD concentrations following the Limit of Detection in Combined Samples study. The

{26}------------------------------------------------

concentration of the pathogen strain that achieved a result around the detection limit (1x LoD) was assessed by testing a minimum of 30 replicates; 5x LoD and 0.1x LoD concentrations were assessed by testing a minimum of 10 replicates. Some pathogens required an additional round of testing to establish the dilution to achieve LoD, in addition to the initial testing described.

Testing for each concentration was executed during at least 3 different days by different operators using four different lots of QIAstat-Dx ME Panel cartridges executed on 3 or more QIAstat-Dx Analyzers.

Negative samples consisted of unspiked clinical CSF, and a minimum of 10 replicates were tested.

The LoD in clinical sample matrix using combined samples has been established for 26 pathogen strains, and in 18 of those strains the LoD concentration was found to be equivalent to the one established in artificial CSF matrix. Eight (8) pathogen strains were not equivalent and had a new LoD concentration determined in clinical CSF.

Claimed LoD concentrations represent the highest (most concentrated) titer of analyte identified from LoD determination in clinical CSF or artificial CSF.

Precision (Repeatability and Reproducibility)

For the reproducibility assessment, a multi-site scheme was followed by testing both negative and positive samples at three different study sites with varying workflow variables, such as sites, days, instruments, operators and cartridge lots that could have an impact on the precision of the system. Negative samples consisted of artificial CSF. Positive combined samples consisted of artificial CSF spiked with a representative panel of pathogens covering all types of organisms targeted by the QIAstat-Dx ME Panel (i.e., RNA virus, gram (+) bacteria, gram (-) bacteria and yeast) at the limit of detection (1x LoD) and at 3x LoD. For each site, testing was performed across 5 non-consecutive days per mix with 6 replicates per day per mix (leading to a total of 90 replicates per target, concentration, and site), a minimum of 9 different QIAstat-Dx Analyzers per site, and at least 3 operators on each testing day.

Reproducibility testing was designed to evaluate the critical variables that may impact the performance of the QIAstat-Dx ME Panel in the context of its routine and intended use.

For the repeatability study, the same sample panel was tested following a single-site scheme. Repeatability testing was designed to evaluate the precision of a OIAstat-Dx ME Panel Cartridge under similar (intra laboratory) conditions. The Repeatability study was assessed with the same samples used for Reproducibility testing using Site 1.

The Reproducibility and Repeatability studies for the OIAstat-Dx ME Panel demonstrated that no test condition has a specific impact on the test performance or pathogen calling and

{27}------------------------------------------------

confirmed its successful precision since the panel performance or pathogen calling showed to be repeatable along the testing days (results are shown in Table 12 and Table 13).

Grouping Variable(s)ProportionTwo-Sided 95%Confidence Limit
TargetConcentrationSiteFractionPercentage(%)LowerUpper
Cryptococcus neoformans/gattii1xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00
Cryptococcus neoformans/gattii3xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00
Enterovirus1xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00
Enterovirus3xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00
Escherichia coli K11xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00
Escherichia coli K13xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00
Listeria monocytogenes1xLoD129 / 3096.6782.7899.92
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All89 / 9098.8993.9699.97
Listeria monocytogenes3xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00

Table 12: Reproducibility study agreement rate (detected vs. expected) per analyte and the 2-sided 95% Confidence Interval by target for 1x LoD, 3x LoD and Negative samples

{28}------------------------------------------------

Grouping Variable(s)ProportionTwo-Sided 95%Confidence Limit
TargetConcentrationSiteFractionPercentage(%)LowerUpper
Streptococcus agalactiae1xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00
Streptococcus agalactiae3xLoD130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00
n/aNegative130 / 30100.0088.43100.00
230 / 30100.0088.43100.00
330 / 30100.0088.43100.00
All90 / 90100.0095.98100.00

Table 13: Repeatability study agreement rate (detected vs. expected) per analyte and the 2-sided 95% Confidence Interval by target for 1x LoD, 3x LoD and Negative samples

Grouping Variable(s)ProportionTwo-Sided 95%Confidence Limit
TargetConcentrationFractionPercentage(%)LowerUpper
Cryptococcus neoformans/gattii1xLoD60 / 60100.0094.04100.00
Cryptococcus neoformans/gattii3xLoD60 / 60100.0094.04100.00
Enterovirus1xLoD57 / 6095.0086.0898.96
Enterovirus3xLoD60 / 60100.0094.04100.00
Escherichia coli K11xLoD56 / 6093.3383.8098.15
Escherichia coli K13xLoD60 / 60100.0094.04100.00
Listeria monocytogenes1xLoD57 / 6095.0086.0898.96
Listeria monocytogenes3xLoD59 / 6098.3391.0699.96
Streptococcus agalactiae1xLoD60 / 60100.0094.04100.00
Streptococcus agalactiae3xLoD60 / 60100.0094.04100.00
NegativeNegative60 / 60100.0094.04100.00

Performance Characteristics - Clinical Studies

Expected values

In the prospective clinical performance evaluation of the QIAstat-Dx Meningitis / Encephalitis (ME) Panel residual cerebrospinal fluid specimens were collected. Specimens

{29}------------------------------------------------

were tested at 13 geographically diverse clinical sites across 4 countries (3 sites in Europe and 10 sites in USA) from March 2022 to March 2023. The number and percentage of positive results as determined by the QIAstat-Dx ME Panel, stratified by age group are presented in Table 14. Overall. 1527 specimens were included in the prevalence assessment and the QIAstat-Dx ME Panel detected at least one organism in a total of 65 prospective specimens analysed in the study (4.3% positivity rate).

1-1718-4445-6465-84>85
PathogenOverall<1 yearyears oldyears oldyears oldyears oldyears oldUnknown
Bacteria
Escherichia coli K12(0.1%)2(1.2%)0(0.0%)0(0.0%)0(0.0%)0(0.0%)0(0.0%)0(0.0%)
Haemophilusinfluenzae7(0.5%)0(0.0%)0(0.0%)1(0.2%)5(1.1%)1(0.3%)0(0.0%)0(0.0%)
Listeriamonocytogenes4(0.3%)0(0.0%)0(0.0%)0(0.0%)2(0.5%)1(0.3%)1(3.2%)0(0.0%)
Neisseriameningitidis(encapsulated)2(0.1%)0(0.0%)0(0.0%)1(0.2%)1(0.2%)0(0.0%)0(0.0%)0(0.0%)
Streptococcusagalactiae3(0.2%)1(0.6%)0(0.0%)0(0.0%)2(0.5%)0(0.0%)0(0.0%)0(0.0%)
Streptococcuspneumoniae17(1.1%)3(1.8%)0(0.0%)3(0.7%)7(1.6%)4(1.2%)0(0.0%)0(0.0%)
Streptococcuspyogenes1(0.1%)0(0.0%)1(0.8%)0(0.0%)0(0.0%)0(0.0%)0(0.0%)0(0.0%)
Virus
Enterovirus (EV)23(1.5%)12(7.4%)3(2.3%)5(1.2%)1(0.2%)1(0.3%)0(0.0%)1(50.0%)
Fungi / Yeast
Cryptococcus gattii /Cryptococcusneoformans(not differentiated)6(0.4%)0(0.0%)0(0.0%)1(0.2%)5(1.1%)0(0.0%)0(0.0%)0(0.0%)
Overall Panel Result
Negative1462(95.7%)145(89.0%)124(96.9%)407(97.4%)419(94.8%)336(98.0%)30(96.8%)1(50.0%)
Positive65(4.3%)18(11.0%)4(3.1%)11(2.6%)23(5.2%)7(2.0%)1(3.2%)1(50.0%)

Table 14: Table Summary of Prevalence of Target Pathogens, as identified by QIAstat-Dx ME Panel, stratified by age

N.B. Total of positive + negative specimens is 1527 due to inclusion of 3 specimens where a QIAstat-Dx result was available but no FilmArray result was available (therefore classed as non-evaluable for the performance analysis).

{30}------------------------------------------------

Clinical Performance

The performance characteristics of the QIAstat-Dx ME Panel was assessed by a multicentre, observational, prospective, and retrospective, clinical performance study, testing fresh and frozen cerebrospinal fluid (CSF) residual specimens obtained by lumbar puncture from patients with signs and symptoms of meningitis and/or encephalitis. The study was conducted at 13 geographically diverse study sites: ten (10) U.S. sites and three (3) European sites. Testing of prospectively collected residual specimens occurred between March 2022 and March 2023. All study sites were hospital-associated or independent clinical diagnostics laboratories that perform routine diagnostics.

Prospective Specimens Testing

A total of 1737 prospective residual CSF specimens were enrolled into the clinical study, of which 205 were withdrawn. The most common reason for specimen withdrawal was ineligibility. Additionally, some prospective samples could not be included in the agreement analysis due to missing data. The final dataset consisted of 1524 prospective specimens of which 552 (36.2%) were frozen before testing and 972 (63.8%) were tested fresh.

Table 15 provides a summary of demographic information for the 1524 specimens included in the prospective study.

Sample CategoryVariableSubgroupN%
Prospective FreshAge Group<1 year13614.0
1-17 years old879.0
18-44 years old28429.2
45-64 years old26627.4
65-84 years old18719.2
≥85 years old111.1
Unknown10.1
GenderFemale49851.2
Male47448.8
Prospective FrozenAge Group<1 year274.9
1-17 years old417.4
18-44 years old13324.1
45-64 years old17431.5
65-84 years old15628.3
≥85 years old203.6
Unknown10.2
GenderFemale27149.1
Male28050.7
Not available10.2
CombinedAge Group<1 year16310.7
1-17 years old1288.4
18-44 years old41727.4
45-64 years old44028.9
65-84 years old34322.5
>85 years old312.0

Table 15: Demographic data for prospective evaluated specimens

{31}------------------------------------------------

Sample CategoryVariableSubgroupN%
GenderUnknown20.1
Female76950.5
Male75449.5
Not available10.1

Residual CSF specimens were tested with the OlAstat-Dx ME Panel and two types of comparator methods (an FDA-cleared molecular comparator or two validated end point PCRs followed by bidirectional sequencing (BDS) for selected targets). All targets were compared to the FDA-cleared molecular method except Streptococcus pneumoniae and Streptococcus pvogenes which were compared against two validated end point PCRs followed by bidirectional sequencing. The standard of care (SoC) testing varied across all sites but included bacterial culture, Laboratory Developed PCR test (LDT), FDA-cleared molecular methods and Cryptococcus antigen screen and culture. Standard of care (SoC) culture results were collected to allow an assessment of clinical sensitivity and specificity and were investigated in cases of discordant result.

Clinical sensitivity or positive percent agreement (PPA) was calculated as 100% x (TP / (TP + FN)). True positive (TP) indicates that both QIAstat-Dx ME Panel and comparator method have a positive result for the specific pathogen. False negative (FN) indicates that the OIAstat-Dx result is negative while the comparator result is positive for the specific pathogen. Clinical Specificity or Negative Percent agreement (NPA) was calculated as 100% x (TN / (TN + FP)). True negative (TN) indicates that both the OIAstat- Dx Panel and the comparator method have negative results for the specific pathogen. False positive (FP) indicates that the OIAstat- Dx Panel result is positive for the specific pathogen, but the comparator result is negative. The two-sided 95% confidence intervals were calculated (Wilson Score Method).

The QIAstat-Dx ME Panel prospective data for positive percent agreement (PPA) and negative percent agreement (NPA) against the comparator methods are presented by analyte in Table 16. Discrepancies between the QIAstat-Dx ME Panel and the comparator methods were investigated and discrepancy investigations are footnoted in Table 16 below.

Positive Percent AgreementNegative Percent Agreement
PathogenSampleCategoryTP/TP+FN%95% CI*TN/TN+FP%95% CI*
Bacteria
Escherichiacoli K1aProspectiveFresh2/366.720.8-93.9969 / 969100.099.6-100.0
ProspectiveFrozen0 / 10.00.0-79.3551 / 551100.099.3-100.0
Overall2 / 450.015.0-85.01520 / 1520100.099.7-100.0

Table 16: Prospectively Collected Clinical Specimens (fresh and frozen) Performance Agreement (PPA and NPA) between QIAstat-Dx ME Panel and Comparator Method by Target

{32}------------------------------------------------

QIAGEN

QIAstat-Dx® Meningitis/Encephalitis (ME) Panel
510(k) Summary

Page 29 of 35
Positive Percent AgreementNegative Percent Agreement
PathogenSampleCategoryTP/TP+FN%95% CI*TN/TN+FP%95% CI*
Bacteria
Haemophilusinfluenzae bProspectiveFresh0 / 10.00.0-79.3970 / 97199.999.4-100.0
ProspectiveFrozen4 / 4100.051.0-100.0546 / 54899.698.7-99.9
Overall4 / 580.037.6-96.41516 / 151999.899.4-99.9
Listeriamonocytogenes cProspectiveFresh1 / 1100.020.7-100.0971 / 971100.099.6-100.0
ProspectiveFrozen3 / 475.030.1-95.4548 / 548100.099.3-100.0
Overall4 / 580.037.6-96.41519 / 1519100.099.7-100.0
Neisseriameningitidis(encapsulated)dProspectiveFresh1 / 1100.020.7-100.0971 / 971100.099.6-100.0
ProspectiveFrozen0 / 0N/AN/A551 / 55299.899.0-100.0
Overall1 / 1100.020.7-100.01522 / 152399.999.6-100.0
StreptococcusagalactiaeProspectiveFresh2 / 2100.034.2-100.0970 / 970100.099.6-100.0
ProspectiveFrozen1 / 1100.020.7-100.0551 / 551100.099.3-100.0
Overall3 / 3100.043.9-100.01521 / 1521100.099.7-100.0
Streptococcuspneumoniae eProspectiveFresh1 / 1100.020.7-100.0845 / 84899.699.0-99.9
ProspectiveFrozen7 / 7100.064.6-100.0515 / 51799.698.6-99.9
Overall8 / 8100.067.6-100.01360 / 136599.699.1-99.8
StreptococcuspyogenesProspectiveFresh0 / 0N/AN/A778 / 778100.099.5-100.0
ProspectiveFrozen0 / 0N/AN/A513 / 513100.099.3-100.0
Overall0 / 0N/AN/A1291 / 1291100.099.7-100.0
Virus
Enterovirus(EV)fProspectiveFresh18 / 2090.069.9-97.2951 / 95299.999.4-100.0
ProspectiveFrozen4 / 4100.051.0-100.0548 / 548100.099.3-100.0
Overall22 / 2491.774.2-97.71499 / 150099.999.6-100.0
Fungi / Yeast
Cryptococcusgattii /Cryptococcusneoformans(notdifferentiated)gProspectiveFresh2 / 540.011.8-76.9965 / 96799.899.2-99.9
ProspectiveFrozen2 / 2100.034.2-100.0550 / 550100.099.3-100.0
Overall4 / 757.125.0-84.21515 / 151799.999.5-100.0

{33}------------------------------------------------

ª For the prospective fresh Escherichia coli KI discordant sample, no organisms were detected with resolution method PCR/BDS. For the frozen Escherichia coli K1 discordant sample no organisms were detected with bacterial culture.

b For the prospective fresh Haemophilus influenzae discordant sample, no organisms were detected by the SoC bacterial culture and resolution testing with PCR/BDS. Of the three (3) false positive Haemophilus influenzae samples, no organisms were detected in one frozen by SoC culture and PCR/BDS resolution method was also negative. No additional testing results associated with the final frozen false positive sample were available.

6 For the frozen Listeria monocytogenes discordant sample the negative result was confirmed positive with SoC culture and LDT result was positive.

d For the frozen Neisseria meningitidis (encapsulated) sample, no organisms were detected by SoC culture and LDT resolution testing with PCR/BDS also returned a negative result for this sample.

e Of the five (5) false positive Streptococcus pneumoniae samples, no organisms were detected in four (3 fresh and 1 frozen) prospective samples with SoC culture. One prospective frozen sample had no SoC result available. However, an FDA cleared method conducted as part of the study also produced a negative result.

f For the prospective fresh Enterovirus discordant samples, no organisms were detected in one sample by two independent SoC LDT assays. The negative result for the second sample returned a negative result with PCR/BDS. For the prospective fresh false positive Enterovirus sample, a negative result was returned when tested with PCR/BDS.

8 Of the three false negative Cryptococcus neoformans (not differentiated) results, no organisms were detected in two fresh samples with fungal culture and PCR/BDS. The remaining false negative fresh sample was confirmed negative for Cryptococcus gattii / Cryptococcus neoformans (not differentiated) with PCR/BDS. Of the two false positive results, no organisms were detected for one fresh sample with PCR/BDS. No organisms were detected in the second fresh sample with SoC fungal culture.

Archived Specimens Testing

Several analytes were either not encountered or had a low prevalence in the prospective arm of the study, so an effort was made to collect and test as many archived samples (positive targeted archived specimens) as possible for all targets within the panel. A total of 195 retrospective archived specimens were enrolled into the study. One hundred and fifty-four (154) archived specimens were excluded from the analysis as positivity was not confirmed by the comparator method. A total of 41 evaluable archived specimens were used in the analysis to support the QIAstat-Dx ME Panel performance evaluation and Table 17 provides a summary of demographic information for the archived specimens.

Table 17: Demographic Summary of Evaluable Archived Specimens for QIAstat-Dx ME Panel Clinical Evaluation

Sample CategoryVariableSubgroupN%
ArchivedAge Group<1 year1126.8
1-17 years old922.0
18-44 years old1229.3
45-64 years old512.2
65-84 years old49.8
ArchivedGenderFemale1946.3
ArchivedGenderMale2253.7

The QIAstat-Dx ME Panel retrospective archived specimen data in positive percent agreement and negative percent against the comparator methods are presented by analyte in Table 18.

{34}------------------------------------------------

Positive Percent AgreementNegative Percent Agreement
PathogenTP/TP+FN%95% CITN/TN+FP%95% CI
Overall Panel41 / 41100.091.4-100.0314 / 314100.098.8-100.0
Bacteria
Escherichia coli Kl2 / 2100.034.2-100.039 / 39100.091.0-100.0
Haemophilusinfluenzae6 / 6100.061.0-100.035 / 35100.090.1-100.0
Listeriamonocytogenes0 / 0N/AN/A41 / 41100.091.4-100.0
Neisseria meningitidis(encapsulated)3 / 3100.043.9-100.038 / 38100.090.8-100.0
Streptococcusagalactiae9 / 9100.070.1-100.032 / 32100.089.3-100.0
Streptococcuspneumoniae4 / 4100.051.0-100.018 / 18100.082.4-100.0
Streptococcuspyogenes0 / 0N/AN/A23 / 23100.085.7-100.0
Virus
Enterovirus (EV)9 / 9100.070.1-100.032 / 32100.089.3-100.0
Fungi / Yeast
Cryptococcus gattii /Cryptococcusneoformans (notdifferentiated)8 / 8100.067.6-100.033 / 33100.089.6-100.0

Table 18: Archived Clinical Performance Agreement (PPA) between QlAstat-Dx ME Panel and Comparator Method by Target*

*Please note that due to volume constraints, not all targets have the same number of specimens tested.

For each analyte where archived samples were available, a PPA of 100% was achieved.

Co-Infections

No specimens with multiple detections were identified by the QIAstat-Dx ME Panel during the performance evaluation.

Clinical sensitivity and specificity determined against culture

The performance measure of sensitivity and specificity was calculated only for bacterial and fungi / yeast analytes for which the gold-standard CSF culture result was available in the standard of care for the specimen. This data was used in additional performance calculations outlined in Table 19 and Table 20.

{35}------------------------------------------------

Sensitivity(compared to culture)Specificity(compared to culture)
PathogenSampleCategoryTP/TP+FN%95% CITN/TN+FP%95% CI
Bacteria
Escherichiacoli KlArchived1/1100.020.7-100.010/1190.962.3-98.4
ProspectiveFresh1/250.09.5-90.5760/760100.099.5-100.0
ProspectiveFrozen0/0N/AN/A343/343100.098.9-100.0
HaemophilusinfluenzaeArchived1/1100.020.7-100.010/1190.962.3-98.4
ProspectiveFresh0/0N/AN/A761/76299.999.3-100.0
ProspectiveFrozen3/3100.043.9-100.0339/34099.798.4-99.9
ListeriamonocytogenesArchived0/0N/AN/A12/12100.075.8-100.0
ProspectiveFresh1/1100.020.7-100.0761/761100.099.5-100.0
ProspectiveFrozen2/366.720.8-93.9340/340100.098.9-100.0
Neisseriameningitidis(encapsulated)Archived2/2100.034.2-100.09/1090.059.6-98.2
ProspectiveFresh0/0N/AN/A761/76299.999.3-100.0
ProspectiveFrozen0/0N/AN/A342/34399.798.4-99.9
StreptococcusagalactiaeArchived0/0N/AN/A12/12100.075.8-100.0
ProspectiveFresh1/1100.020.7-100.0760/76199.999.3-100.0
ProspectiveFrozen1/1100.020.7-100.0342/342100.098.9-100.0
StreptococcuspneumoniaeArchived0/0N/AN/A11/1291.764.6-98.5
ProspectiveFresh0/0N/AN/A757/76299.398.5-99.7
ProspectiveFrozen3/3100.043.9-100.0339/34099.798.4-99.9
StreptococcuspyogenesArchived0/0N/AN/A12/12100.075.8-100.0
ProspectiveFresh0/0N/AN/A761/76299.999.3-100.0
ProspectiveFrozen0/0N/AN/A343/343100.098.9-100.0

Table 19: Bacterial Culture comparison for diagnostic sensitivity and specificity by sample category

Cryptococcus was also calculated in comparison to fungal culture performed as a standard of care diagnostic at clinical testing sites. For data that were available, QIAstat-Dx ME Panel performance is summarized in Table 20.

{36}------------------------------------------------

Positive Percent AgreementNegative Percent Agreement
PathogenSample CategoryTP/TP+FN%95% CITN/TN+FP%95% CI
Fungi / Yeast
CryptococcusgattiiArchived2 / 2100.034.2-100.01 / 1100.020.7-100.0
Prospective1 / 1100.020.7-100.0129 / 13198.594.6-99.6
Cryptococcusneoformans(notdifferentiated)Fresh
ProspectiveFrozen0 / 0N/AN/A24 / 24100.086.2-100.0

Table 20: Fungal Culture Comparison for diagnostic sensitivity and specificity by sample category

Validity of results

The failure rate for initial clinical specimens tests for prospective fresh was 2.7% (26/977), for prospective frozen was 1.3% (7/555) for prospective frozen and for archived samples was 1.7% (3/176). Those failures consisted of 5 instrument errors and 2 invalid results, and 29 were other run failures. All specimens except 5 (3 prospective fresh and 2 prospective frozen) were retested and were successful after retest, yielding a final success rate of 99.7% for prospective fresh, 99.6% for prospective frozen and 100.0% for archived samples. The failure rates breakdown due to instrument, invalid results and other run failures is summarized in Table 21. Withdrawn specimens were not included in the validity assessment; however some prospective samples which were excluded from the agreement analysis were included for the validity assessment (3 prospective fresh, 5 prospective frozen and 135 archive specimens).

Table 21: Summary of the Number of Samples with Failed Test Results (Initial and
Final)
Placeholder Column 1Placeholder Column 2
Placeholder Data 1Placeholder Data 2
Initial RunsFinal Runs(after repeats)
Sample CategoryFailure ReasonN/Total%N/Total%
Prospective FreshInvalid*0 / 9770.00 / 9770.0
Prospective FreshInstrument3 / 9770.30 / 9770.0
Prospective FreshOther**23 / 9772.43 / 9770.3
Prospective FrozenInvalid0 / 5550.00 / 5550.0
Prospective FrozenInstrument1 / 5550.20 / 5550.0
Prospective FrozenOther6 / 5551.12 / 5550.4
ArchivedInvalid2 / 1761.10 / 1760.0
ArchivedInstrument1 / 1760.60 / 1760.0
ArchivedOther0 / 1760.00 / 1760.0

*Internal Control failures with at least one analyte detected and the other analytes reported as 'invalid'

** Run failures related to workflow checkpoints.

{37}------------------------------------------------

Contrived Specimens Testing

Contrived specimens testing was conducted to support the nine (9) targets on the panel as there were insufficient positive specimens obtained from both prospective and archived collection efforts. Contrived specimens were prepared by spiking five different quantified strains representative of the genetic diversity of each pathogen. For each pathogen, the LoD concentration was manufactured at 2x (at least 50%) and 5x LoD spiked into screened individual unique samples of negative CSF. Contrived specimens were tested alongside negative specimens in a blinded manner. The results are summarized in Table 22.

Table 22: Overall Proportion of Positive Results Post Exclusions or withdrawals of Contrived Positive Samples for Each Target

Classification(genome type)PathogenConcentrationLevelFrequencyof PositiveResultsProportion(%) ofPositiveResultsLower95%ConfidenceLimitUpper95%ConfidenceLimit
BacteriaEscherichia coli Kl2xLoD48 / 48100.092.6100.0
5xLoD37 / 37100.090.6100.0
Total85 / 85100.095.7100.0
Haemophilus influenzae2xLoD57 / 57100.093.7100.0
5xLoD36 / 36100.090.4100.0
Total93 / 93100.096.0100.0
Listeria monocytogenes2xLoD47 / 4995.986.398.9
5xLoD38 / 38100.090.8100.0
Total85 / 8797.792.099.4
Neisseria meningitidis(encapsulated)2xLoD46 / 4895.886.098.8
5xLoD39 / 4097.587.199.6
Total85 / 8896.690.598.8
Streptococcusagalactiae2xLoD49 / 49100.092.7100.0
5xLoD39 / 39100.091.0100.0
Total88 / 88100.095.8100.0
Streptococcuspneumoniae2xLoD55 / 5796.588.199.0
5xLoD39 / 39100.091.0100.0
Total94 / 9697.992.799.4
Streptococcus pyogenes2xLoD47 / 4995.986.398.9
5xLoD40 / 40100.091.2100.0
Total87 / 8997.892.299.4
VirusEnterovirus (EV)2xLoD48 / 4998.089.399.6
5xLoD39 / 39100.091.0100.0
Total87 / 8898.993.899.8
Fungi / YeastCryptococcus gattii /Cryptococcusneoformans (notdifferentiated)2xLoD41 / 41100.091.4100.0
5xLoD38 / 38100.090.8100.0
Total79 / 79100.095.4100.0

{38}------------------------------------------------

Target Proportion of Positive Result ≥95% was achieved for all prepared contrived samples 2xLoD and 5xLoD in all tested analytes.

Conclusions

The QIAstat-Dx ME Panel is substantially equivalent to the to the predicate device, FilmArray Meningitis/Encephalitis (ME) Panel.

§ 866.3970 Device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid.

(a)
Identification. A device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid is a qualitative in vitro device intended for the detection and identification of microbial-associated nucleic acid sequences from patients suspected of meningitis or encephalitis. A device to detect and identify microbial pathogen nucleic acids in cerebrospinal fluid is intended to aid in the diagnosis of meningitis or encephalitis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical studies: Analytical sensitivity (limit of detection), inclusivity, reproducibility, interference, cross reactivity, and specimen stability.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted comparator methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) The Intended Use statement in the device labeling must include a statement that the device is intended to be used in conjunction with standard of care culture.
(6) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(7) The device labeling must include a limitation stating that the negative results do not preclude the possibility of central nervous system infection.
(8) The device labeling must include a limitation stating that device results are not intended to be used as the sole basis for diagnosis, treatment, or other patient management decisions.
(9) The device labeling must include a limitation stating that positive results do not mean that the organism detected is infectious or is the causative agent for clinical symptoms.
(10) As part of the risk management activities performed as part of your 21 CFR 820.30 design controls, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.