K Number
K242189
Manufacturer
Date Cleared
2024-08-20

(26 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.

The tape may be used in the following sterilization cycles:

· Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX, maX 2, 60, and s2 Low Temperature Sterilization Systems.

· Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALL Clear Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear

Device Description

The CELERITY HP Indicator Tape (Indicator Tape) is a ¾'' wide crepe paper tape printed with diagonal stripes of a vaporized hydrogen peroxide (VHP) reactive ink. The ink is sealed with a varnish, the function of which is to inhibit removal of the ink via transference or by the adhesive (as the tape is dispensed). The reactive ink meets the performance specifications for a Type 1 process indicator for vaporized hydrogen peroxide as defined in ANSI/AAMI/ISO 11140-1:2014.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called "CELERITY HP Indicator Tape (PCC077)". It does not contain information about acceptance criteria or a study proving that the device meets those criteria.

The document states, "No testing was required to verify the labeling change," and clarifies that "The revised labeling is in accordance with recommendations in 'User Labeling for Devices that Contain Natural Rubber (21 CFR 801.437); Small Entity Compliance Guide; Guidance for Industry'' April 1, 2003." This implies that the current submission (K242189) is for a labeling change on an existing device, and therefore, no new performance testing was conducted for this specific submission.

The document also refers to the predicate device, K240760, and states that the proposed and predicate devices have "Identical" technological characteristics and performance limitations. This suggests that the performance criteria and related studies would have been established and demonstrated for the predicate device, and since the current device is identical in these aspects (apart from labeling), no new studies are needed.

Therefore, the requested information regarding acceptance criteria and a study proving the device meets them cannot be extracted from this document, as this submission primarily addresses a labeling change.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.

August 20, 2024

STERIS Anthony Piotrkowski Director, Regulatory Affairs 5960 Heisley Rd Mentor. Ohio 44060

Re: K242189

Trade/Device Name: CELERITY HP Indicator Tape (PCC077) Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: July 24, 2024 Received: July 25, 2024

Dear Anthony Piotrkowski:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen A.
Anisko -S

Digitally signed by Stephen A. Anisko -S Date: 2024.08.20 15:45:50 -04'00'

for: Christopher Dugard Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K242189

Device Name

CELERITY HP Indicator Tape (PCC077)

Indications for Use (Describe)

The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.

The tape may be used in the following sterilization cycles:

· Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-PRO: 1, 1 Plus, maX, maX 2, 60, and s2 Low Temperature Sterilization Systems.

· Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALL Clear Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary For CELERITY HP Indicator Tape (PCC077) K242189

Sponsor Facility

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459

Manufacturing Facility

Contact: Anthony Piotrkowski Director, Regulatory Affairs Telephone: (440) 392-7437 Fax No: (440) 357-9198 Tony piotrkowski@steris.com

Submission Date: July 24, 2024

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

1. Predicate Device

Trade Name: Common/Usual Name: Classification: Classification Name: 510(k) Submitter/Holder: 510(k) Number:

CELERITY HP Indicator Tape (PCC077) Chemical Indicator Class II Physical/chemical Sterilization Process Indicator STERIS Corporation K240760

2. Device Description

The CELERITY HP Indicator Tape (Indicator Tape) is a ¾'' wide crepe paper tape printed with diagonal stripes of a vaporized hydrogen peroxide (VHP) reactive ink. The ink is sealed with a varnish, the function of which is to inhibit removal of the ink via transference or by the adhesive (as the tape is dispensed). The reactive ink meets the performance specifications for a Type 1 process indicator for vaporized hydrogen peroxide as defined in ANSI/AAMI/ISO 11140-1:2014.

3. Indications for Use:

The Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink/magenta) to orange, yellow or lighter. The Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.

The tape may be used in the following sterilization cycles:

  • Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible, and Specialty Cycles of the V-. PRO: 1, 1 Plus, maX, maX 2, 60, and s2 Low Temperature Sterilization Systems.
  • . Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without ALLClear
  • . Standard, Flex Scope, Express and DUO Cycles of the STERRAD® 100NX Sterilizer with or without ALLClear.

4. Technological Characteristics

The proposed and predicate devices are chemical indicators in accordance with ISO 11140-1:2014 for use in monitoring Vaporized Hydrogen Peroxide sterilization cycles. The mechanism of action and endpoint are the same and when exposed to the defined processing conditions, the proposed and predicate devices exhibit a visible color change.

{5}------------------------------------------------

FeatureProposed: CELERITY HP Indicator TapePredicate: CELERITY HP Indicator Tape(K240760)Comparison
Intended Use/ Indicationsfor UseThe Indicator Tape is intended to secure itemswrapped in synthetic wrap materials until usedand to distinguish between processed andunprocessed units through a color changefrom the start color (pink/magenta) to orange,yellow or lighter. The Indicator Tape is alsointended to be used as an external processindicator when applied to synthetic wrapmaterials and/or Tyvek pouches.The tape may be used in the followingsterilization cycles:Lumen, Non Lumen, Fast Non Lumen,Fast, Flexible, and Specialty Cycles of theV-PRO: 1, 1 Plus, maX, maX 2, 60, ands2 Low Temperature SterilizationSystems.Standard and Advanced Cycles of theSTERRAD® NX Sterilizer with orwithout ALLClearStandard, Flex Scope, Express and DUOCycles of the STERRAD® 100NXSterilizer with or without ALLClearThe Indicator Tape is intended to secure itemswrapped in synthetic wrap materials until usedand to distinguish between processed andunprocessed units through a color changefrom the start color (pink/magenta) to orange,yellow or lighter. The Indicator Tape is alsointended to be used as an external processindicator when applied to synthetic wrapmaterials and/or Tyvek pouches.The tape may be used in the followingsterilization cycles:Lumen, Non Lumen, Fast Non Lumen,Fast, Flexible, and Specialty Cycles of theV-PRO: 1, 1 Plus, maX, maX 2, 60, ands2 Low Temperature SterilizationSystems.Standard and Advanced Cycles of theSTERRAD® NX Sterilizer with orwithout ALLClearStandard, Flex Scope, Express and DUOCycles of the STERRAD® 100NXSterilizer with or without ALLClearIdentical
ChemicalIndicatorAgent andPerformanceSpecificationThe indicator agent is a vaporized hydrogenperoxide reactive ink that meets therequirement specified in ANSI/AAMI/ISO11140-1:2014 for a Type 1 Process Indicatorfor a vaporized hydrogen peroxide processThe indicator agent is a vaporized hydrogenperoxide reactive ink that meets therequirement specified in ANSI/AAMI/ISO11140-1:2014 for a Type 1 Process Indicatorfor a vaporized hydrogen peroxide processIdentical
DeviceDesignTape: 3/4" wide by 60 yards long crepe(masking) tape which is wound around a 3"core. A hydrogen peroxide reactive ink is laiddown on the non-adhesive surface. The ink isprotected from transfer to the adhesive via acoating.Tape: 3/4" wide by 60 yards long crepe(masking) tape which is wound around a 3"core. A hydrogen peroxide reactive ink is laiddown on the non-adhesive surface. The ink isprotected from transfer to the adhesive via acoating.Identical
Shelf Life5 months - real-time testing is ongoing5 months - real-time testing is ongoingIdentical
PerformanceLimitationsDo not overlap the tape onto itself as this mayprevent the underlying layer from beingappropriately exposed to hydrogen peroxideresulting in a failed indicator response.Do not overlap the tape onto itself as this mayprevent the underlying layer from beingappropriately exposed to hydrogen peroxideresulting in a failed indicator response.Identical

Table 1 Comparison of CI Physical Description and Technological Properties

5. Performance Testing

No testing was required to verify the labeling change. The revised labeling is in accordance with recommendations in "User Labeling for Devices that Contain Natural Rubber (21 CFR 801.437); Small Entity Compliance Guide; Guidance for Industry'' April 1, 2003.

{6}------------------------------------------------

6. Conclusion

Based on the intended uses, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate device (K240760, Class II as per 21 CFR 880.2800, product code JOJ).

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).