K Number
K242128

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-02-27

(223 days)

Product Code
Regulation Number
876.1075
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Forsvall biopsy needle is intended to collect histological biopsy samples from soft tissue intended for clinical examination. The Forsvall biopsy needle is part of the Forsvall biopsy needle system for prostate biopsy and shall be used together with the Forsvall biopsy gun.

The Forsvall biopsy gun is intended to be used by medical professionals, together with the Forsvall biopsy needle, on male patients to obtain soft tissue samples. The Forsvall biopsy gun is part of the Forsvall biopsy needle system for prostate biopsy.

The Forsvall biopsy needle system for prostate biopsy is indicated whenever there is a need to obtain prostate biopsy samples.

Device Description

The Forsvall biopsy needle system for prostate biopsy consists of two items: the Forsvall biopsy needle and the Forsvall biopsy gun. The Forsvall biopsy needle is intended for use with the Forsvall biopsy gun. The Forsvall biopsy gun is reusable and the biopsy needle is single use. The Forsvall biopsy system uses a spring loaded gun that mediates a biopsy needle with a side notch used to obtain tissue samples. The biopsy gun is made of metal and hard plastic and the 16/18 gauge 10/15/20/25 cm biopsy needle is made of stainless medical grade steel.

By pulling a handle on the Forsvall biopsy gun, the front wagon therein is tensioned and connected to the Forsvall biopsy needle. When tensioning the load pin also moves along a guide curve. When the front wagon is in its loaded position, the load pin has moved the front needle bracket a small distance backwards in the Forsvall biopsy gun.

The handle is pulled a second time to tension the rear wagon that is connected to the needle in the Forsvall biopsy needle. A tension is created between the outer tube and the needle in order to force closing of the Forsvall biopsy needle. The purpose is to prevent an interstice from arising in order to prevent accumulation of bacteria.

The Forsvall biopsy gun is fired by pushing the trigger, which causes the rear wagon to move forward of the front wagon. This results in the inner part of the needle being forced forward, followed by forward movement of the outer tube. When the Forsvall biopsy gun is fired, the load pin moves the outer tube the final millimeter in order to achieve a soft closing mechanism. On the Forsvall biopsy needle, the outer tube thus stops against the needle head and it is forced closed.

The Forsvall biopsy operates by the outer tube stopping against the needle head in the closing motion. Additionally, the final millimeter in the closing motion is a soft close action in order to protect the needle head from too strong forces. To enable the soft closing action, the Forsvall biopsy gun has a design similar to the predicate device.

AI/ML Overview

This looks like a 510(k) summary for a medical device called the "Forsvall biopsy needle system for prostate biopsy." Unfortunately, the provided text does not contain the information necessary to answer your questions about acceptance criteria and the study proving the device meets them.

The document focuses on demonstrating substantial equivalence to a predicate device (Möller Medical Biopsy Needles and Systems) based on similar indications for use, technological characteristics, and general safety and performance.

Here's why I cannot provide the requested information from the given text:

  • Acceptance Criteria and Device Performance Table: The document lists "Non-Clinical and/or Clinical Tests Summary & Conclusions," but it does not provide specific acceptance criteria (e.g., minimum sample length, specific force measurements) or a table comparing these criteria to the device's reported performance. It only states that "All non-clinical tests related to performance as described above passed."
  • Study Details (Sample Size, Data Provenance, Experts, Adjudication, MRMC, Standalone, Ground Truth, Training Set): The document briefly mentions "User Testing" in a "simulated clinical setting" but provides no details about:
    • The sample size of this user testing.
    • The provenance of any data (e.g., country, retrospective/prospective).
    • The number or qualifications of experts.
    • Adjudication methods.
    • Whether it was an MRMC comparative effectiveness study or a standalone study.
    • The type of ground truth used.
    • Any training set information.

In summary, the provided FDA 510(k) summary (K242128) does not contain the detailed study information you are asking for regarding acceptance criteria and performance data. It's a regulatory document focused on establishing substantial equivalence, not a detailed scientific study report. Such detailed study information would typically be found in the full 510(k) submission, which is not publicly available in its entirety.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below that.

February 27, 2025

Xaga Surgical Anders Grim OA Manager Alnarpsgatan 81 Helsingborg, Scan SE-256 67 SWEDEN

K242128 Re:

Trade/Device Name: Forsvall biopsy needle system for prostate biopsy (Forsvall biopsy gun and Forsvall biopsy needle) Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: II Product Code: KNW Dated: July 19, 2024 Received: January 31, 2025

Dear Anders Grim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/uniquedevice-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific

{2}------------------------------------------------

regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino -S

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242128

Device Name

Forsvall biopsy needle system for prostate biopsy (Forsvall biopsy gun and Forsvall biopsy needle)

Indications for Use (Describe)

The Forsvall biopsy needle is intended to collect histological biopsy samples from soft tissue intended for clinical examination. The Forsvall biopsy needle is part of the Forsvall biopsy needle system for prostate biopsy and shall be used together with the Forsvall biopsy gun. The Forsvall biopsy gun is intended to be used by medical professionals, together with the Forsvall

biopsy needle, on male patients to obtain soft tissue samples. The Forsvall biopsy gun is part of the Forsvall biopsy needle system for prostate biopsy.

The Forsvall biopsy needle system for prostate biopsy is indicated whenever there is a need to obtain prostate biopsy samples.

Type of Use (Select one or both, as applicable)

scription Use (Part 21 CFR 801 Subpart D)

he-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) #:K242128K242128 Page 1 of 2
510(k) SummaryPrepared on: 2025-02-26
Contact Details21 CFR 807.92(a)(1)
Applicant NameXaga Surgical
Applicant AddressAlnarpsgatan 81 Helsingborg Scania SE-256 67 Sweden
Applicant Contact Telephone+46707109969
Applicant ContactMr. Anders Grim
Applicant Contact Emailandersgrim@xagasurgical.com
Correspondent NameStravus AB
Correspondent AddressIliongränden 320 Lund Scania 22471 Sweden
Correspondent Contact Telephone+46708960161
Correspondent ContactMr. Michael Lundh
Correspondent Contact Emailmichaellundh@xagasurgical.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameForsvall biopsy needle system for prostate biopsy (Forsvall biopsy gun and Forsvall biopsy needle)
Common NameGastroenterology-urology biopsy instrument
Classification NameInstrument, Biopsy
Regulation Number876.1075
Product Code(s)KNW, KNW (CLASS 2) - INSTRUMENT, BIOPSY
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K162588Möller Medical Biopsy Needles and SystemsKNW
Device Description Summary21 CFR 807.92(a)(4)
-------------------------------------------------

The Forsvall biopsy needle system for prosists of two items: the Forsvall biopsy needle and the Forsvall biopsy gun. The Forsvall biopsy needle is intended for use with the Forsvall biopsy gun is reusable and the biopsy needle is single use. The Forsvall biopsy system uses a spring loaded gun that mediates a biopsy needle with a side notch used to obtain tissue samples. The biopsy gun is made of metal and hard plastic and the 16/18 gauge 10/15/20/25 cm biopsy needle is made of stainless medical grade steel.

By pulling a handle on the Forsvall biopsy gun, the front wagon therein is tensioned and connected to the Forsvall biopsy needle. When tensioning the foad pin also moves along a quide curve. When the front wagon is in its loaded position, the load pin has moved the front needle bracket a small distance backwards in the Forsvall biopsy gun.

{5}------------------------------------------------

The handle is pulled a second time to tear wagon that is connected to the needle in the Forsvall biopsy needle. A tension is created between the outer tube and the needle in order to force closing of the Forsvall biopsy needle. The purpose is to prevent an interstice from arising in order to prevent accumulation of bacteria.

The Forsvall biopsy gun is fired by pushing the trigger, which causes the rear wagon to move forward of the front wagon. This results in the inner part of the Forced forward, followed by forward movement of the outer tube. When the Forsvall biopsy gun is fired, the load pin moves the outer tube the final millimeter in order to achieve a soft closing mechanism. On the Forsvall biopsy needle, the outer tube thus stops against the needle head and it is forced closed.

The Forsvall biopsy operates by to the outer tube stoping against the needle head in the closing motion. Additionally, the final millimeter in the closing motion is a soft close action in order to protect the needle head from too strong forces. To enable the soft closing action, the Forsvall biopsy gun has a design similar to the predicate device.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Forsvall biopsy needle is intended to collect histological biopsy samples from soft tissue intended for clinical examination. The Forsvall biopsy needle is part of the Forsvall biopsy needle biopsy and shall be used together with the Forsvall biopsy qun.

The Forsvall biopsy gun is intended to be used by medical professionals, together with the Forsvall biopsy needle, on male patients to obtain soft tissue samples. The Forsvall biopsy gun is part of the Forsvall biopsy needle system for prostate biopsy.

The Forsvall biopsy needle system for prostate biopsy is indicated whenever there is a need to obtain prostate biopsy samples.

Indications for Use Comparison

The Forsvall biopsy needle system for prostate biopsy is indicated for use for prostate biopsy only, whereas the predicate device (Möller) has multiple use indications where prostate biopsy is one (see Möller models: FNB, MBY, ABG, SBG + Model Möller Medical Blue RBG),

These differences are concluded to not constitute a new intended use since models (as presented above) and the Forsvall biopsy needle system for prostate biopsy are intended to be used by the same patient group and for the same medical reasons.

Technological Comparison

The Forsvall biopsy system consists of the disposable biopsy needle and the reusable biopsy gun. The needle consist of a single use 16/18 gauge 10/15/20/25 cm needle in medical grade stainless steel with a side notch 20 mm pocket for biopsy sampling just as the preducate device. The biopsy gun is spring loaded and has the same principle of action using force, the same design and materials and moving the needle in the same principal way as the predicate device.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The main non-clinical testing that was done to determine substantial equivalence:

-User Testing: Users have fired the biopsy model to confirm that it performs as expected in a simulated clinical setting. -Dimensional Verification: Confirmed that the needle dimensions are appropriate to achieve the proper sample length.

-Needle Specifications: Verified that the needle length and diameter are suitable.

-Sample Retrieval: Validated that a tissue sample can be effectively removed using the needle.

-Needle Gap Minimization: Confirmed that the gap between the needle components is minimized to limit patient exposure of in-vivo bacteria.

-Needle Tip Quality: Ensured that the needle tip is smooth and free from defects, which is crucial for patient safety and procedure efficacy.

-Shelf life and transport testing.

See bench_tests.pdf and user_tests.pdf for details.

N/A

All non-clinical tests related to performance as described above passed. Risk management was performed according to ISO 14971. Design properties and risks scenarios that differs from the predicate device have been assessed and concluded to be either state-of-theart or related to a risk control measure where the benefit is deemed greater than the risk.

The device is therefore at least as good as the predicate device with respect to safety, effectiveness and performance

21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.