K Number
K242018
Manufacturer
Date Cleared
2024-11-12

(125 days)

Product Code
Regulation Number
870.2910
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UbiqVue 2A Multiparameter System is a wireless remote patient monitoring system intended for use by healthcare professionals for continuous collection of physiological data at home and in healthcare settings. This shall include electrocardiography (2-Channel ECG), heart Rate, SpO2%, respiration rate, pulse Rate, photoplethysmography waveform, skin temperature, body posture, body motion, R-R Interval, heart Rate variability (HRV) and blood Pressure (optional). Data is transmitted wirelessly near real time from UbigVue 2A Wearable Biosensor and 3rd party device (for Blood Pressure only) to remote central server for display, storage, and analysis.

The UbiqVue 2A Multiparameter System is intended for non-critical, adult population.

The UbiqVue 2A Multiparameter System shall include the ability to notify healthcare professionals through alerts when physiological parameters fall outside the set limits, manual trigger by the patient and to display multiple patient's physiological data for remote monitoring at home and with visual alarm for active monitoring at hospitals and out-of-hospital patient care settings (such as clinics, outpatient surgery facilities, long-term care facilities and physician offices) in which care is administered by healthcare professionals.

Device Description

The UbiqVue™ 2A Multiparameter System is similar to LifeSignals Multiparameter Remote Monitoring Platform that is cleared under K202868 but upgraded with monitoring of additional physiological parameters, minor performance improvements and additional User Interface.

The UbiqVue™ 2A Multiparameter System consists of following components:

    1. UbiqVue™ 2A Wearable Biosensor
    1. UbigVue™ Single Patient Relay
    1. UbiqVue™ Multi-Patient Relay
    1. UbiqVue™ Central Server
    1. UbiqVue™ Active Monitoring Portal
    1. UbiqVue™ Relay Bridge Software (optional)
    1. UbiqVue™ Bluetooth Gateway Software (optional)
    1. Third-party Blood Pressure device (optional)

The UbigVue™ 2A Wearable Biosensor when attached to body acquires two channels of ECG signals, TTI respiration signals (one of the inputs for deriving Respiration Rate), resistance variation of a Thermistor attached to body (used for deriving Skin Temperature) & accelerometer data (input for deriving Respiration Rate & Posture), pre-processes them and wirelessly transmits to a paired Relay device, as cleared under K202868.

Following are additional features proposed in this submission:

  • Ambient temperature sensor for Body temperature derivation.
  • Acquisition of PPG (Photoplethysmography) signals using the integrated Optical sensor array.
  • Derivation of SpO2% and Pulse Rate within the Biosensor, based on PPG signals acquired through the Optical sensor array.

Biosensor uses standard WLAN (802.11b) secured (AES) communication protocol for wireless data transmission to the Relay Device, as cleared under K202868. The Biosensor is also additionally integrated with BLE Radio for communication with Single Patient Relay device.

As cleared under K202868, when Relay device is available within the wireless range, the acquired data is continuously transmitted to the Relay device immediately. If Relay device is not available or if there is any interruption in the communication between Relay device and Biosensor, data shall be temporarily buffered locally in Biosensor till the wireless connection is re-established.

The UbigVue™ Single Patient Relay Device, a standard Mobile device or custom device that is installed with a UbigVue Single Patient Relay Application software, that receives the physiological data transmitted from any paired Biosensor. It shall transmit the data received from the Biosensor, immediately without any day to Central Server that is configured to receive, after necessary data encryption. The Single Patient Relay Device shall receive data from Biosensor either through WLAN or BLE and transmit it to Central Server using its mobile data network (3G, 4G, LTE, 5G, etc.)
Single Patient Relay Device function is identical to the one cleared under K202868, except it is upgraded to support additional parameters & also improvements in GUI as below:

  • To handle additional data packets of Biosensor such as Ambient Temperature, PPG signals. SpO2% and Pulse Rate.
  • Interactive GUI that provides feedback to Users about the signal quality of PPG signals to identify the optimal body location for affixing the Biosensor.
  • Shall pair with a 3rd party Blood Pressure device using BLE and transfer the data to the Central Server.
  • Allows User to enter Oral Temperature reading (for calibration) and/or manual Blood pressure data (optional if no 3td Party device is connected).

The UbiqVue™ Multi-Patient Relay Device (MPR) is an UbiqVue™ Multi-Patient Relay Application Software installed in a standard Linux platform (Physically or in cloud), that has ability to pair & receives data from multiple Biosensors, through wireless access points (WLAN) connected in its network. It shall transmit the received data from Biosensors to Central Server that is configured to receive through internet or private network (VPN), after necessary encryption & authentication. Multi-Patient Relay Device also receives data from multiple third-party Blood Pressure devices worn by the patients & connected to its network (through BLE gateway).

If the Central Server is not available or there is no connectivity, the data shall be temporarily buffered securely in the MPR device itself. There is no GUI component for this software application.

The UbigVue™ Central Server is UbiqVue™ Central Server Application software installed in a compatible Linux Server hardware Platform. The Central Server Application functions are also identical to Secure Server Application in LifeSignals Multi-parameter Remote Monitoring Platform (K202868), other than being upgraded to handle additional parameters. Also, the "Sensor Processing Library", the algorithm that derives various vital parameters is upgraded for performance improvements. The following are additional features:

  • Extending the minimum skin temperature measurement capability from 32.0℃ (89.6 °F) to 15.0°C (59.0°F)
  • Body Temperature estimation using Skin Temperature, Ambient temperature, Heart Rate (based on clinical condition) and Activity.
  • Uses the Oral Temperature entered manually for initial / periodic calibration of derived Body temperature.
  • Additional Posture classification & Body motion detection capabilities.
  • Improvements in Respiration Rate & Heart Rate derivation algorithm.
  • Integration with 3rd Party beat analysis & classification software for deriving Sinus Heart Rate, R-R Interval & Heart Rate Variability.
  • The PPG Signal Quality Index (SQI) algorithm module for displaying the SQI on the Active Monitoring dashboard & to access the quality of SpO2 value derived by the Biosensor.
  • Support for Early Warning Score (NEWS2) calculation.

Central Server shall have ability to send alert notifications to any configured one or more Users through E-mail, SMS or WhatsApp, for any Clinical, Technical or Manual Alert conditions, as cleared under K202868. However, this alert engine is upgraded to support additional parameters and improved configuration capabilities such as Acknowledgement, priority & condition delay time.

Active Monitoring Portal in UbiqVue™ 2A Multiparameter System is identical to Web UI in LifeSignals Multi-parameter Remote Monitoring Platform cleared under K202868. It is a browser-based User Interface Application that enables Clinical personnel to login to the Central server remotely and access the patient physiological data (Biosensor & derived data) and view/Acknowledge the Alert(s). The Clinical personnel, depending on the roles (normal or supervisory) can access data of multiple patients assigned to them and search them based on the recent alert status. This includes patients that are active (wearing Biosensor) and monitoring procedures completed.
It is in-built with a Monitoring Dashboard that continuously displays physiological parameters, waveforms & alert status of any assigned patients to any authenticated Clinical personnel for (quasi) real time Active Monitoring. It has the option for the User to select multiple patient tile view or Single Patient Zoom View/Hybrid view of any patient based on multiple available filter settings.

Following are the few changes in this 510(k):

  • Additional parameters (SpO2%. PR, EWS, SYS, DIA, MAP, PPG, SQI, EWS).
  • Visual alarm display for any Clinical, Technical, or manual alert/alarm.
  • Hybrid tile view (Zoom View + Tile view) and Group tile view.
  • Multi-level User Management, roles & privileges.

Note: Any Third-Party Application / Server may communicate with UbiqVue™ Central Server using a defined set of Application Programming Interface (API) for entering patient information, configuring & generating customized report or any other requirement that does not alter the intended use of the system (e.g., EMR/EHR, HIMS) or to meet the additional intended use claimed by the third-party application. UbigVue™ Active Monitoring Portal shall continue to be used for the claimed Active Monitoring intended use. However, UbiqVue™ Active Monitoring portal may be optional for integration with any other regulatory approved monitoring dashboard or different intended use claims.

Relay Bridge Software in UbiqVue™ 2A Multiparameter System when installed in a standard OpenWrt based routers, shall provide communication between Biosensor(s) & Multi-Patient Relay. Relay Bridge is an alternate for a standard access point and is intended when Multi-Patient Relay Software is hosted in the cloud. Similar to standard access point, the Relay Bridge communicates with multiple Biosensor and transfers the data to the Multi-Patient Relay without buffering the data. Since the Relay bridge transfers the data to the designated Multi-Patient Relay bridge in the cloud, it further encrypts the data for securely transmitting through the internet via ethernet, Wi-Fi or cellular network.

Bluetooth Gateway Software in UbigVue™ 2A Multiparameter System when installed in a standard OpenWrt based BLE gateway or BLE supported routers, shall provide communication between 3rd party BLE devices (e.g. BP device) & Multi-Patient Relay. Bluetooth Gateway communicates with multiple 3rd party devices and transfers the data to the Multi-Patient Relay without buffering.

AI/ML Overview

LifeSignals, Inc. has developed the UbiqVue™ 2A Multiparameter System (UX2550), a wireless remote patient monitoring system. The device collects continuous physiological data including 2-Channel ECG, heart rate, SpO2%, respiration rate, pulse rate, photoplethysmography waveform, skin temperature, body temperature, body posture, body motion, R-R Interval, heart rate variability (HRV), and optionally blood pressure. The system transmits data wirelessly to a remote central server for display, storage, and analysis, and can notify healthcare professionals through alerts. It is intended for non-critical, adult populations in home and healthcare settings.

The following information describes the acceptance criteria and the study conducted to prove the device meets these criteria:

1. Acceptance Criteria and Reported Device Performance

ParameterAcceptance CriteriaReported Device Performance
Heart RateRange: 30 – 250 BPM Accuracy: ± 3 BPM or 10% whichever is greater (Predicate-I)Range: 30 – 250 BPM Accuracy: ± 3 BPM or 5% whichever is greater
Respiration RateRange: 6-60 Breaths per Minute (Predicate-I); 4-40 Breaths per Minute (Predicate-II) Accuracy: ≤ 1 Breath per minute (MAE simulation), ≤ 3 Breaths per minute (MAE clinical) (Predicate-I); ± 3 Breaths per Minute (Predicate-II)Range: 5-60 Breaths per Minute Accuracy: ≤ 1 Breath per minute (MAE simulation), ≤ 3 Breaths per minute (MAE clinical)
Skin TemperatureRange: 32.0°C – 43.0°C (89.6°F – 109.4°F) Accuracy: As per ASTM E1112-00 (Predicate-I)Range: 15.0°C – 43.0°C (59.0°F – 109.4°F) Accuracy: As per ASTM E1112-00
Body TemperatureRange: 32.0°C – 43.0°C (89.6°F – 107.6°F) (Predicate-II) Clinical Accuracy: ±0.3°C (Predicate-II)Range: 32.0°C - 43.0°C (89.6°F - 109.4°F) Clinical Accuracy: -0.27 °C (Clinical Bias), 1.01°C (Limits of agreement)
SpO2%Range: 40 to 100% (Predicate-II) Accuracy: ±2% (Predicate-II)Range: 0 to 100% Accuracy: ± 3 % (100 to 70%), Less than 70% unspecified
Pulse RateRange: 40 to 250 BPM (Predicate-II) Accuracy: ± 3% (Predicate-II)Range: 30 to 250 BPM Accuracy: ± 3 or 5% whichever is greater
Wear Life120 hours120 hours
Alerts/AlarmsCompliance with IEC 60601-1-8 (Ed. 2.2) (validated for visual alarm, alerts via email/SMS for primary visual alarm indication)Compliance with IEC 60601-1-8 (Ed. 2.2)
BiocompatibilityAccording to ISO 10993-1: 2018Testing conducted according to ISO 10993-1:2018 (In-vitro cytotoxicity, skin irritation, skin sensitization)
Electrical Safety & EMCCompliance with IEC 60601-1(Ed.3.2), IEC 60601-1-2 (Ed. 4.1), IEC 60601-1-11(Ed. 2.1)Compliance achieved
Wireless PerformanceConformance to ANSI/IEEE C63.27:2017Conformance achieved for Single Patient Relay System & Multi-Patient Relay System
SoftwareIEC 62304 & USFDA Guidance (Class B, Moderate-Enhanced documentation), IEC 81001-5-1Designed, documented, verified & validated per specified standards
Shelf-life & ReliabilityAs per acceptance criteriaTesting conducted and verified
PackagingAs per ASTM D7386-16Testing conducted
UsabilityCompliance with IEC 60601-1-6 (Ed. 3.2) and "USFDA Guidance- Applying Human Factors and Usability Engineering to Medical Devices."Study conducted

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the exact sample sizes for all test sets. However, it does mention specific studies:

  • SpO2% Accuracy: Clinical validation was performed to determine the accuracy of SpO2% with respect to SaO2 values from simultaneously drawn arterial blood as per ISO 80601-2-61 (Ed. 2.0). The specific sample size for this clinical validation is not provided in the summary.
  • Body Temperature Accuracy: Validated using "non-randomized, self-control comparative on-body comparative performance with oral digital thermometer on multiple subject populations" in accordance with ISO 80601-2-56 (Ed.2.0). The specific number of subjects or populations is not provided.
  • Wear-life Performance: Validated using "non-randomized, self-control comparative on-body comparative performance study for the wear period of 120 hours." The sample size for this study is not provided.
  • Respiration Algorithm Improvement: Validated by comparing against "clinician manually scored end tidal CO2 (EtCO2) capnography, under spontaneous breathing & metronome breathing during normal activity conditions as part of wearlife study." The sample size is part of the wear-life study, but not explicitly stated.

Data Provenance: The document does not explicitly state the country of origin for the clinical study data or whether studies were retrospective or prospective. Given the FDA 510(k) submission, it is implied that the data is acceptable for the US market.

3. Number of Experts Used to Establish the Ground Truth and Qualifications

The document indicates that for the respiration algorithm validation, the comparison was made against "clinician manually scored end tidal CO2 (EtCO2) capnography." This implies that clinicians (experts) were involved in establishing the ground truth. However, the exact number of experts and their specific qualifications (e.g., years of experience, specialty) are not provided.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1, none) for the test sets. For the "clinician manually scored" data mentioned for respiration, it's not specified if multiple clinicians scored and how discrepancies were resolved.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any study evaluating the effect size of how much human readers (or healthcare professionals using the device) improve with AI vs without AI assistance. The device is a monitoring system that provides data and alerts, rather than an AI-driven diagnostic interpretation system typically evaluated with MRMC studies.

6. Standalone (Algorithm Only) Performance

The document describes specific performance tests for various parameters (e.g., heart rate, respiration rate, SpO2%, skin temperature, body temperature) that would inherently reflect the algorithm's performance in deriving these parameters from sensor data. For example:

  • SpO2% accuracy was proven through "clinical validation to determine the accuracy of SpO2% with respect to SaO2 values from simultaneously drawn arterial blood." This implies a comparison of the device's output (algorithm only) to a gold standard.
  • Body temperature accuracy was validated against an "oral digital thermometer."
  • Respiration algorithm validation was against "clinician manually scored end tidal CO2 (EtCO2) capnography."

These tests demonstrate standalone performance metrics against established reference standards.

7. Type of Ground Truth Used

The types of ground truth used include:

  • Physiological measurements from reference devices/standards:
    • SaO2 values from simultaneously drawn arterial blood (for SpO2% accuracy).
    • Oral digital thermometer (for Body Temperature accuracy).
    • Bench testing standards (e.g., ASTM E1112-00 for skin temperature, ISO 80601-2-61 for Pulse Rate).
  • Clinician-established ground truth:
    • Clinician manually scored end tidal CO2 (EtCO2) capnography (for Respiration Rate algorithm).

8. Sample Size for the Training Set

The document does not explicitly state the sample size used for training the algorithms within the UbiqVue™ 2A Multiparameter System. It discusses validation of improved algorithms and additional features, but provides no detail on the development or training phase data.

9. How the Ground Truth for the Training Set Was Established

As the document does not provide details on the training set, it does not describe how the ground truth was established for the training set. It primarily focuses on the verification and validation (testing) of the final device/algorithms.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol representing the Department of Health & Human Services on the left, and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in a sans-serif font, with "FDA" in a blue square and the rest of the text in blue as well.

November 12, 2024

LifeSignals, Inc. Saravanan Balasubramanian Vice President - Global, Medical Technology & Regulatory Operations 426 S Hillview Dr. Milpitas, California 95035

Re: K242018

Trade/Device Name: UbiqVue™ 2A Multi-parameter System (UX2550) Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter And Receiver Regulatory Class: Class II Product Code: DRG, MHX, FLL, DQA Dated: October 10, 2024 Received: October 10, 2024

Dear Saravanan Balasubramanian:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer W. Shih -S

Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242018

Device Name

UbiqVue™ 2A Multiparameter System (UX2550)

Indications for Use (Describe)

The UbiqVue 2A Multiparameter System is a wireless remote patient monitoring system intended for use by healthcare professionals for continuous collection of physiological data at home and in healthcare settings. This shall include electrocardiography (2-Channel ECG), heart Rate, SpO2%, respiration rate, pulse Rate, photoplethysmography waveform, skin temperature, body posture, body motion, R-R Interval, heart Rate variability (HRV) and blood Pressure (optional). Data is transmitted wirelessly near real time from UbigVue 2A Wearable Biosensor and 3rd party device (for Blood Pressure only) to remote central server for display, storage, and analysis.

The UbiqVue 2A Multiparameter System is intended for non-critical, adult population.

The UbiqVue 2A Multiparameter System shall include the ability to notify healthcare professionals through alerts when physiological parameters fall outside the set limits, manual trigger by the patient and to display multiple patient's physiological data for remote monitoring at home and with visual alarm for active monitoring at hospitals and out-of-hospital patient care settings (such as clinics, outpatient surgery facilities, long-term care facilities and physician offices) in which care is administered by healthcare professionals.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform inside, followed by the text "LifeSignals" in blue. The waveform graphic is on the left side of the logo, and the text is on the right.

510(k) Summary for UbiqVue™ 2A Multiparameter System

1 Company name : LifeSignals, Inc.

  • Address :426 S Hillview Drive Milpitas, CA 95035. USA
    Contact : Saravanan Balasubramanian Vice President - Global, Medical Technology & Regulatory Operations Email: saravanan@lifesignals.com Tel: 510.770.6412 Ext. 4

  • 2 Date prepared : October 07, 2024

3 Device

Trade Name: UbiqVue™ 2A Multiparameter System
Model Name: 2A Multiparameter System
Common Name: Transmitters and Receivers, Physiological Signal, Radio Frequency

4 Classification Product code

RegulationClassificationProductCodeDescription
Cardiovascular21 CFR 870.2910Class IIDRGTransmitters and Receivers,Physiological Signal, Radiofrequency

5 Subsequent Product code

RegulationClassificationProductCodeDescription
Cardiovascular21 CFR 870.1025Class IIMHXMonitor, Physiological,Patient (With ArrhythmiaDetection Or Alarms)
General Hospital21 CFR 880.2910Class IIFLLThermometer, Electronic,Clinical

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for LifeSignals. The logo consists of a blue square on the left side, which contains a white waveform graphic. To the right of the square is the text "LifeSignals" in a blue sans-serif font. The logo is simple and modern, and the waveform graphic suggests that the company is involved in healthcare or technology.

RegulationClassificationProductCodeDescription
Anesthesiology21 CFR 870.2700Class IIDQAOximeter

6 Predicate & Reference Devices

Primary Predicate Device:
510(k) Number: K202868
Model: LifeSignals Multiparameter Remote Monitoring Platform
Manufacturer: LifeSignals, Inc. USA
Secondary Predicate (for SpO2 , Pusle Rate & Body temperature intended use)
510(k) Number: K212153
Model: Biobeat Platform, BB-613WP Patch
Manufacturer: Biobeat Technologies Ltd. Israel
Reference Device-I (for Body temperature, HRV, RR Interval, EWS Intended Use)
510(k) Number: K183078
Model: Vitalconnect Platform, VitalPatch Biosensor
Manufacturer: Vital Connect, Inc. USA
Reference Device -II (for SpO2, Pulse Rate, PPG & SQI technology):
510(k) Number: K200160
Model: 740 SafeSAT
Manufacturer: Zoe Medical, Inc. USA
Reference Device-III (for Visual Alarm & Active Monitoring Intended Use):
510(k) Number: K160951
Model: Philips Efficia CMS200 Central Monitoring System
Manufacturer: Philips Medical Systems, USA
Reference Device-IV (for Body Temperature accuracy claim):
510(k) Number: K203215
Model: Radius T Wearable Thermometer
Manufacturer: Masimo Corporation, USA

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform inside, followed by the text "LifeSignals" in blue. The waveform is a series of vertical lines that vary in height, resembling a signal. The text is in a sans-serif font and is aligned to the right of the square.

7 Device description

The UbiqVue™ 2A Multiparameter System is similar to LifeSignals Multiparameter Remote Monitoring Platform that is cleared under K202868 but upgraded with monitoring of additional physiological parameters, minor performance improvements and additional User Interface.

The UbiqVue™ 2A Multiparameter System consists of following components:

    1. UbiqVue™ 2A Wearable Biosensor
    1. UbigVue™ Single Patient Relay
    1. UbiqVue™ Multi-Patient Relay
    1. UbiqVue™ Central Server
    1. UbiqVue™ Active Monitoring Portal
    1. UbiqVue™ Relay Bridge Software (optional)
    1. UbiqVue™ Bluetooth Gateway Software (optional)
    1. Third-party Blood Pressure device (optional)
  • The UbigVue™ 2A Wearable Biosensor when attached to body acquires two channels of ● ECG signals, TTI respiration signals (one of the inputs for deriving Respiration Rate), resistance variation of a Thermistor attached to body (used for deriving Skin Temperature) & accelerometer data (input for deriving Respiration Rate & Posture), pre-processes them and wirelessly transmits to a paired Relay device, as cleared under K202868.

Following are additional features proposed in this submission:

  • Ambient temperature sensor for Body temperature derivation. O
  • Acquisition of PPG (Photoplethysmography) signals using the integrated Optical o sensor array.
  • Derivation of SpO2% and Pulse Rate within the Biosensor, based on PPG signals o acquired through the Optical sensor array.

Biosensor uses standard WLAN (802.11b) secured (AES) communication protocol for wireless data transmission to the Relay Device, as cleared under K202868. The Biosensor is also additionally integrated with BLE Radio for communication with Single Patient Relay device.

As cleared under K202868, when Relay device is available within the wireless range, the acquired data is continuously transmitted to the Relay device immediately. If Relay device is not available or if there is any interruption in the communication between Relay device and Biosensor, data shall be temporarily buffered locally in Biosensor till the wireless connection is re-established.

{7}------------------------------------------------

"In LifeSignals

  • The UbigVue™ Single Patient Relay Device, a standard Mobile device or custom device . that is installed with a UbigVue Single Patient Relay Application software, that receives the physiological data transmitted from any paired Biosensor. It shall transmit the data received from the Biosensor, immediately without any day to Central Server that is configured to receive, after necessary data encryption. The Single Patient Relay Device shall receive data from Biosensor either through WLAN or BLE and transmit it to Central Server using its mobile data network (3G, 4G, LTE, 5G, etc.)
    Single Patient Relay Device function is identical to the one cleared under K202868, except it is upgraded to support additional parameters & also improvements in GUI as below:

  • To handle additional data packets of Biosensor such as Ambient Temperature, PPG O signals. SpO2% and Pulse Rate.

  • Interactive GUI that provides feedback to Users about the signal quality of PPG O signals to identify the optimal body location for affixing the Biosensor.

  • Shall pair with a 3rd party Blood Pressure device using BLE and transfer the data о to the Central Server.

  • Allows User to enter Oral Temperature reading (for calibration) and/or manual o Blood pressure data (optional if no 3td Party device is connected).

  • The UbiqVue™ Multi-Patient Relay Device (MPR) is an UbiqVue™ Multi-Patient Relay ● Application Software installed in a standard Linux platform (Physically or in cloud), that has ability to pair & receives data from multiple Biosensors, through wireless access points (WLAN) connected in its network. It shall transmit the received data from Biosensors to Central Server that is configured to receive through internet or private network (VPN), after necessary encryption & authentication. Multi-Patient Relay Device also receives data from multiple third-party Blood Pressure devices worn by the patients & connected to its network (through BLE gateway).

If the Central Server is not available or there is no connectivity, the data shall be temporarily buffered securely in the MPR device itself. There is no GUI component for this software application.

  • The UbigVue™ Central Server is UbiqVue™ Central Server Application software ● installed in a compatible Linux Server hardware Platform. The Central Server Application functions are also identical to Secure Server Application in LifeSignals Multi-parameter Remote Monitoring Platform (K202868), other than being upgraded to handle additional parameters. Also, the "Sensor Processing Library", the algorithm that derives various vital parameters is upgraded for performance improvements. The following are additional features:
    • Extending the minimum skin temperature measurement capability from 32.0℃ o (89.6 °F) to 15.0°C (59.0°F)

{8}------------------------------------------------

"It" LifeSignals

  • Body Temperature estimation using Skin Temperature, Ambient temperature, о Heart Rate (based on clinical condition) and Activity.
  • Uses the Oral Temperature entered manually for initial / periodic calibration of O derived Body temperature.
  • Additional Posture classification & Body motion detection capabilities. O
  • Improvements in Respiration Rate & Heart Rate derivation algorithm. O
  • Integration with 3rd Party beat analysis & classification software for deriving Sinus O Heart Rate, R-R Interval & Heart Rate Variability.
  • The PPG Signal Quality Index (SQI) algorithm module for displaying the SQI on O the Active Monitoring dashboard & to access the quality of SpO2 value derived by the Biosensor.
  • Support for Early Warning Score (NEWS2) calculation. o

Central Server shall have ability to send alert notifications to any configured one or more Users through E-mail, SMS or WhatsApp, for any Clinical, Technical or Manual Alert conditions, as cleared under K202868. However, this alert engine is upgraded to support additional parameters and improved configuration capabilities such as Acknowledgement, priority & condition delay time.

  • Active Monitoring Portal in UbiqVue™ 2A Multiparameter System is identical to Web . UI in LifeSignals Multi-parameter Remote Monitoring Platform cleared under K202868. It is a browser-based User Interface Application that enables Clinical personnel to login to the Central server remotely and access the patient physiological data (Biosensor & derived data) and view/Acknowledge the Alert(s). The Clinical personnel, depending on the roles (normal or supervisory) can access data of multiple patients assigned to them and search them based on the recent alert status. This includes patients that are active (wearing Biosensor) and monitoring procedures completed.
    It is in-built with a Monitoring Dashboard that continuously displays physiological parameters, waveforms & alert status of any assigned patients to any authenticated Clinical personnel for (quasi) real time Active Monitoring. It has the option for the User to select multiple patient tile view or Single Patient Zoom View/Hybrid view of any patient based on multiple available filter settings.

Following are the few changes in this 510(k):

  • 0 Additional parameters (SpO2%. PR, EWS, SYS, DIA, MAP, PPG, SQI, EWS).
  • Visual alarm display for any Clinical, Technical, or manual alert/alarm. O
  • Hybrid tile view (Zoom View + Tile view) and Group tile view. O
  • o Multi-level User Management, roles & privileges.

Note: Any Third-Party Application / Server may communicate with UbiqVue™ Central Server using a defined set of Application Programming Interface (API) for entering patient information, configuring & generating customized report or any other requirement that

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform inside, followed by the text "LifeSignals" in blue. The waveform graphic is on the left side of the logo, and the text is on the right.

does not alter the intended use of the system (e.g., EMR/EHR, HIMS) or to meet the additional intended use claimed by the third-party application. UbigVue™ Active Monitoring Portal shall continue to be used for the claimed Active Monitoring intended use. However, UbiqVue™ Active Monitoring portal may be optional for integration with any other regulatory approved monitoring dashboard or different intended use claims.

  • Relay Bridge Software in UbiqVue™ 2A Multiparameter System when installed in a . standard OpenWrt based routers, shall provide communication between Biosensor(s) & Multi-Patient Relay. Relay Bridge is an alternate for a standard access point and is intended when Multi-Patient Relay Software is hosted in the cloud. Similar to standard access point, the Relay Bridge communicates with multiple Biosensor and transfers the data to the Multi-Patient Relay without buffering the data. Since the Relay bridge transfers the data to the designated Multi-Patient Relay bridge in the cloud, it further encrypts the data for securely transmitting through the internet via ethernet, Wi-Fi or cellular network.
  • Bluetooth Gateway Software in UbigVue™ 2A Multiparameter System when installed in ● a standard OpenWrt based BLE gateway or BLE supported routers, shall provide communication between 3rd party BLE devices (e.g. BP device) & Multi-Patient Relay. Bluetooth Gateway communicates with multiple 3rd party devices and transfers the data to the Multi-Patient Relay without buffering.

8 Indications for Use

  • . The UbiqVue 2A Multiparameter System is a wireless remote patient monitoring system intended for use by healthcare professionals for continuous collection of physiological data at home and in healthcare settings. This shall include electrocardiography (2-Channel ECG), heart rate, SpO2%, respiration rate, pulse rate, photoplethysmography waveform, skin temperature, body temperature, body posture, body motion, R-R Interval, heart Rate Variability (HRV) and Blood Pressure (optional). Data is transmitted wirelessly near real time from Ubiq Vue 2A Wearable Biosensor and 3rd party device (for Blood Pressure only) to remote central server for display, storage, and analysis.
  • The UbiqVue 2A Multiparameter System is intended for non-critical, adult population.
  • The UbiqVue 2A Multiparameter System shall include the ability to notify healthcare professionals through alerts when physiological parameters fall outside the set limits, manual trigger by the patient and to display multiple patient's physiological data for remote monitoring at home and with visual alarm for active monitoring at hospitals and out-of-hospital patient care settings (such as clinics, outpatient surgery facilities, long-term care facilities and physician offices) in which care is administered by healthcare professionals.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the LifeSignals logo. The logo consists of a blue square with white sound wave-like lines on the left side. To the right of the square is the word "LifeSignals" in blue, with the "L" being slightly larger than the rest of the letters. The logo is simple and modern, and the sound wave-like lines suggest that the company is involved in some kind of technology or communication.

Predicate-I (K202868) Predicate-II (K212153) Subject Device Comparison Manufacturer Biobeat Technologies Ltd. LifeSignals, Inc., USA LifeSignals, Inc., USA DRG, MHX, FLL, DQA Product Codes DRG, MHX, FLL DQA, DRG, BZQ, DXG, FLL Regulation Classification 21 CFR 870.2910 21 CFR 870.2700 21 CFR 870.2910 (Primary) Class II Class II Class II The LifeSignals Multi-Parameter The Biobeat Platform is a wireless The UbiqVue™ 2A Multiparameter noninvasive remote monitoring System is a wireless remote patient Remote Monitoring Platform is a remote monitoring system intended for use by monitoring system intended for use by wireless system intended for use healthcare professionals for spotbv healthcare professionals for professionals continuous collection of physiological of healthcare for checking collection continuous collection of physiological data in home and data at home and in healthcare physiological data at home and in healthcare settings. This shall include This can settings. healthcare settings. This shall functional electrocardiography, heart include. oxygen rate. include Electrocardiography (2 saturation of arterial hemoglobin SpO2%, respiration rate, pulse rate, channel ECG), Heart Rate, (%SpO2), pulse rate, blood photoplethysmography waveform, Respiration Rate, Skin pressure, respiration rate (RRp), skin temperature, body temperature, Temperature & Posture. Data is and body temperature. body posture, body motion, R-R Intended use / Interval, heart rate variability transmitted wirelessly from The Biobeat Platform tracks Indications for use LifeSignals Biosensor to Remote changes in blood pressure based on (HRV) and Blood Pressure (optional). secure server for display, storage Pulse Wave Transit Time (PWTT) Data is transmitted wirelessly near & analysis. The LifeSignals which is obtained utilizing real time from Ubiq Vue 2A Wearable Multi-Parameter Remote pulsemeasurements from Biosensor and 3rd party device (for Monitoring Platform is intended Blood Pressure only) to Remote the integrated SpO2 sensor. for non-critical, adult population. following a calibration process Central server for display, storage, The LifeSignals Multi-Parameter using an FDA-cleared and analysis. The UbiqVue™ 2A Multiparameter Remote Monitoring Platform can oscillometric blood pressure include the ability to notify System is intended for non-critical, monitor. healthcare professionals when The Biobeat Platform is intended adult population. for spot- checking and tracking The UbiqVue™ 2A Multiparameter physiological parameters fall outside the set limits and to changes of adult patients in System shall include the ability to

ல் Substantial Equivalence comparison (Subject device & Predicate Device)

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform pattern on the left, followed by the word "LifeSignals" in a bold, blue sans-serif font. The waveform pattern is stylized and represents the company's focus on biosensors and remote monitoring.

K242018 510(k) Summary

ComparisonPredicate-I (K202868)Predicate-II (K212153)Subject Device
multipledisplaypatientphysiological data for remotemonitoring.hospitals, clinics, long-term care,and at home . The data from theBiobeat Platform are intended foruse by healthcare professionals asan aid to diagnosis and treatment.The device is not intended for useon critical care patients.healthcareprofessionalsnotifythrough alerts when Physiologicalparameters fall outside the set limits,manual trigger by patient andtomultipledisplaypatient'sphysiological data for remotemonitoring at home and with visualalarm for active monitoring athospitals and out of hospital patientsettings (such as clinics,careoutpatient surgery facilities, long-term care facilities and physicianwhichcareoffices)inisadministeredhealthcarebyprofessionals.
Intended PopulationNon-critical, adult population,18 years or olderNon-critical, adult population,18 years or olderNon-critical, adult population, 18years or older
Intended UseEnvironmentHome & Healthcare settingsHome & Healthcare settingsHome & Healthcare settings
Monitored Parameters
ECG (Two Channel)V==============================================================================================================================================================================×V
Heart RateV.×
Heart Rate Variability×
R-R Interval×
Respiration Rate------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------VV
Skin TemperatureV×
Body Temperature×V
Body PostureV×V
Body Motion××
SpO2%×VV
ComparisonPredicate-I (K202868)Predicate-II (K212153)Subject Device
Pulse Ratex
PPG Waveformx
Blood Pressurex✓ (using 3rd party device)
Continuous Physiologicaldata monitoringx(Spot check only)
Heart RateRange30 – 250 BPMx30 – 250 BPM
Accuracy± 3 BPM or 10% whicheveris greaterx± 3 BPM or 5% whichever isgreater
Respiration RateRange6-60 Breaths per Minute4-40 Breaths per Minute5-60 Breaths per Minute
Accuracy≤ 1 Breath per minute (Mean Absolute Error based onSimulation study)≤ 3 Breaths per minute (Mean Absolute Error based on Clinicalstudy)± 3 Breaths per Minute≤ 1 Breath per minute (Mean Absolute Error based on Simulation study)≤ 3 Breaths per minute (Mean Absolute Error based on Clinicalstudy)
MethodTrans-thoracic Impedance (TTI),ECG Derived Respiration (EDR)based on RS Amplitude andAccelerometer data.Derived fromphotoplethysmogramTrans-thoracic Impedance (TTI),ECG Derived Respiration (EDR)based on RS Amplitude &Respiratory Sinus arrhythmia(RSA) and Accelerometer data.
Skin TemperatureRange32.0°C – 43.0°C89.6°F – 109.4°Fx15.0°C – 43.0°C59.0°F – 109.4°F
AccuracyAs per ASTM E1112-00xAs per ASTM E1112-00
Body TemperatureRangex32.0°C – 43.0°C89.6°F – 107.6°F32.0°C - 43.0°C89.6°F - 109.4°F
ComparisonPredicate-I (K202868)Predicate-II (K212153)Subject Device
Clinical Accuracy×±0.3°C-0.27 °C (Clinical Bias)1.01°C (Limits of agreement)
Method×Adjusted mode (with SkinTemperature & AmbientTemperature)Adjusted mode (with SkinTemperature, AmbientTemperature, Heart Rate &Activity)
Posture DetectionLying down, Upright orInclined×Supine, Prone, Lying- Left/Right,Reclining, Reclining-Left/Right,Leaning Forward, Upright, Upsidedown
SpO2%
Range×40 to 100%0 to 100%
Accuracy×±2%± 3 % (100 to 70%),Less than 70% unspecified
PPG×
Signal Strength Indicator××
Measurement Site×ChestChest
Optical Sensor×880nm (IR), 650nm (Red)Spectral Sensor
Pulse Rate
Range×40 to 250 BPM30 to 250 BPM
Accuracy×± 3%± 3 or 5% whichever is greater
R-R Interval××
Beat Classification××
Sinus Heart Rate××
Heart Rate Variability(HRV)
SDNN××
SDANN××
ComparisonPredicate-I (K202868)Predicate-II (K212153)Subject Device
ASDNN××
RMSSD××
VLF Band××
LF Band××
HF Band××
Body Motion××No Body Motion, Mild BodyMotion, Moderate Body Motion,Severe Body Motion
Sensor
Single Use
Wear Life120 hours120 hours120 hours
Dimension105 x 94 x 12 mmMonitor enclosure - 38 x 38 x 15 mm.Adhesive unit - 85 x 85 x 12 mm116 x 91 x17 mm
Weight28 g14g35 g
ECG Electrode distance
ECG-A90.0 mm×90.0 mm
ECG-B57.9 mm73.0 mm
Location of BodyUpper Left ChestChest-patch attached to the skin& Wrist deviceUpper Center Chest orUpper Left Chest
Data can be transferred &temporarily stored
BatteryLi-MnO2 (Primary)Li-MnO2 (Primary)Li-MnO2 (Primary)
Applied part categoryType CFType BFType CF
Defibrillation Proof×
HF Surgical equipmentcompatibility×
Communication protocolWLAN (802.11b)BLE 4.2WLAN (802.11b), BLE 5.2
Wireless Radio Frequency2.4 - 2.4835 GHz2.402-2.480 GHz2.4 - 2.4835 GHz
ComparisonPredicate-I (K202868)Predicate-II (K212153)Subject Device
Communication SecurityWPA2-PSK/CCMPAES-CCM 128WPA2-PSK /CCMPAES-CCM 128
Patient Relay Device
Authentication from Server-
Data transferred to Server-
Data buffered if there is noconnection with Server
GUI for Patient admission(Only for Single Patient Relay)-
GUI for PPG signal quality(Only for Single Patient Relay)×-
Secure ServerData is stored for access byActive Monitoring Portal &any 3rd party server (throughAPI)-Data is stored for access byActive Monitoring Portal & any3rd party server (through API)
Programmable Alert
Notification & Configurable✓ (Clinical, Technical, Manual)-✓ (Clinical, Technical, Manual)
Early Warning Score(NEWS2)×-
Active Monitoring Portal
Multi-patient Tile view-
Single Patient Zoom view-
Visual Alarm Indication×-
Early Warning score×-

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform graphic inside, followed by the text "LifeSignals" in a bold, blue font. The waveform graphic appears to represent a life signal or vital sign, aligning with the company's name.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for LifeSignals. The logo consists of a blue square with a white waveform inside, followed by the word "LifeSignals" in blue, sans-serif font. The waveform graphic is on the left side of the logo, and the text is on the right.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains the logo for LifeSignals. The logo consists of a blue square on the left side, which contains a waveform graphic. To the right of the square is the company name, "LifeSignals", in a bold, blue font.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for LifeSignals. The logo consists of a blue square with a white waveform inside, followed by the word "LifeSignals" in blue, sans-serif font. The waveform graphic is on the left side of the logo, and the text is on the right.

Note: ✔ Feature is identical

× Feature not available - Feature Unknown

{16}------------------------------------------------

Image /page/16/Picture/1 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform graphic on the left, followed by the word "LifeSignals" in a bold, blue font. The waveform graphic appears to represent a signal or wave pattern, possibly related to health or technology. The overall design is clean and modern.

Differences and Risks associated with that:

  • The Subject device has additional physiological data monitoring capabilities of, R-R interval, Heart Rate Variability & Body . motion in the system, compared the Predicate device(s). However, performance of these parameters is validated & the technology to measure these parameters is equivalent to other 510(k) cleared device which is included as Reference Device-I under Section #10 below (K183978). Hence, these additional parameters do not affect the intended use of the Proposed device.
  • The Subject device has an Early Warning Score display in the Active Monitoring dashboard as an added feature, compared to ● the Predicate Device(s). However, this does not add any risk, as Early Warning Score is based on well-known NEWS2 (National Early Warning Score by Royal College of Physicians, UK) adopted by other 510(k) cleared device which is included as Reference Device-I section #10 (K183078) and the integration of which is validated.
  • The Subject device has improved algorithms to derive Respiration Respiratory Sinus arrhythmia (RSA) . based EDR, derive Heart Rate from either of the two channels dynamically & extended Skin temperature compared to the Predicate device. However, performance of improved Respiration Rate, Skin temperature algorithm is validated and hence does not add any risk.
  • The Subject device has a Visual Alarm indication and intended for Active monitoring at hospitals patient care ● settings (such as clinics, outpatient surgery facilities and physician offices) in which care is administered by healthcare professionals, using the updated Active Monitoring portal. However, this does not add any risk, as the Visual Alarm indication is in compliance with IEC 60601-1-8 and Intended use is equivalent to other 510(k) cleared device which is included as Reference Device-III under Section #10 (K160951)
  • The Subject Device uses IR/Red/White LED and Spectral sensor for continuous SpO2% & Pulse rate derivation compared to . the standard IR/Red LED/sensor used in the Predicate device. However, performance of these parameters is validated & the technology to measure these parameters is equivalent to other 510(k) cleared device which is included as Reference Device-II under Section #10 (K200160).
  • The Body temperature accuracy claim of the predicate device is different from the subject device. However, the acuracy claims ● of the Reference Device-IV (K203215) that is included under Section #10 is same as the subject device.

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform graphic inside, followed by the company name "LifeSignals" in blue text. The waveform graphic is a series of vertical lines of varying heights, resembling a signal or sound wave.

10 Comparison with Reference devices for additional physiological monitoring parameters & Technology.

ComparisonReference Device -I (K183078)Reference Device- II (K200160)Reference Device-III (K160951)Reference Device-IV (K203215)Subject Device
ManufacturerVital Connect, Inc.Zoe Medical, Inc.Philips Medical SystemsMasimo CorporationLifeSignals, Inc., USA
Product CodesDRG, MHX, DSIMWI, DQA, DXNMSXFLLDRG, MHX, FLL, DQA
Regulation21 CFR 870.291021 CFR 870.230021 CFR 870.230021 CFR 880.291021 CFR 870.2910
Classification (Primary)Class IIClass IIClass IIClass IIClass II
Intended use /Indications for useThe VitalConnectPlatform is a wirelessremote monitoringsystem intended foruse by healthcareprofessionals forcontinuous collectionof physiological datain home andhealthcare settings.This can include heartrate,electrocardiography(ECG), heart ratevariability, R-Rinterval, respiratoryrate, bodytemperature, skintemperature, activityThe 740 SafeSAT isindicated for use asa bedside, portabledevice for use byhealth careprofessionals,clinicians andmedically qualifiedpersonnel for spotchecking,continuousmonitoring andrecording of adultand pediatricpatients,excluding neonates.The Monitorfeatures technologyto facilitate theThe Efficia CMS200central monitoringsystem is intended foruse by healthcareprofessionals forcentral viewing ofphysiologic waves,parameters, and trendsfrom other networkedmedical devices(patient monitors andvital signsmonitors) for multiplepatients. It providessecondary operatornotification of alarmsfrom other networkedmedical devices. Itprovides for theRadius T wearablethermometer isintended for single-use, continuousnoninvasivemeasurement of bodytemperature on theupper chest viawirelesscommunication to asmart deviceapplication orcompatible patientmonitor (i.e., MasimoRoot, Masimo Rad-97).The Radius T isindicated for single-use, continuous bodyThe UbiqVue™ 2AMultiparameter System is awireless remote monitoringsystem intended for use byhealthcare professionals forcontinuous collection ofphysiological data at homeand in healthcare settings.This shall includeelectrocardiography, heartrate, SpO2%, respiration rate,pulse rate,photoplethysmographywaveform, skin temperature,body temperature, bodyposture, body motion, R-RInterval, heart rate Variability(HRV) and Blood Pressure(optional). Data is transmitted
ComparisonReference Device -I (K183078)Reference Device- II (K200160)Reference Device-III (K160951)Reference Device-IV (K203215)Subject Device
(includingstepcount), and posture(bodypositionrelative to gravityincluding fall). Dataaretransmitted wirelesslyfromtheVitalConnect Sensorforstorageandanalysis. The VitalConnect Platform caninclude the ability tonotifyhealthcareprofessionalswhenphysiological data falloutsideselectedparameters. The datafromtheVitalConnectPlatform are intendedfor use by healthcareprofessionals as anaid to diagnosis andtreatment.monitoring of non-invasivebloodpressure, pulse rateandfunctionalarterialoxygensaturation (SpO2) ina variety of clinicalenvironments.retrospective review ofalarmconditions,physiologic waves andparametersfrommultiple patients. Theintended use of theprinter, when present,is to provide hardcopytext, graphics,andwave data. The EfficiaCMS200 may providefor connection andinformation exchangeto external systems.The Efficia CMS200 isintended for use inhospitals and out-of-hospital patient caresettings(suchasclinics,outpatientsurgeryfacilities,long-termcarefacilitiesandphysician offices) inwhichcareisadministeredbyhealthcareProfessionals.temperaturemeasurementsofpersons 5 years of ageorolder in hospitals,hospital-typefacilities, and homeenvironments.wirelessly near real time fromUbiqVue2AWearableBiosensor and 3rd partydevice (for Blood Pressureonly) to remote central serverfor display, storage, andanalysis. The UbiqVue2AMultiparameter System isintended for non-criticaladult population.TheUbiqVue2AMultiparameter System shallinclude the ability to notifyhealthcare professionalsthroughalertswhenphysiological parameters falloutside the set limits, manualtrigger by the patient and todisplay multiple patient'sphysiological data for Remotemonitoring at home and withvisual alarm for activemonitoring at hospitals andout of- hospital patient caresettings (such as clinics,outpatient surgery facilities,long-term care facilities andphysician offices) in whichcare is administered byhealthcare professionals.
ComparisonReference Device -I (K183078)Reference Device- II (K200160)Reference Device-III (K160951)Reference Device-IV (K203215)Subject Device
Intended PopulationGeneral carepatients who are 18years or olderGeneral & AcutecareAdult andPediatric PatientsGeneral patients5 years or olderNon-critical, adultpopulation, 18 years orolder
Intended UseEnvironmentHome & HealthcaresettingsHealthcareSettingsHealthcare SettingsHome &Healthcare settingsHome & Healthcaresettings
Monitored Parameters
ECG✓ (SingleChannel)××✓ (Two Channels)
Heart Rate××
Heart Rate Variability××
R-R Interval××
Respiration Rate××
Skin Temperature×××
Body Temperature×
Body Posture×××
Activity / Body Motion✓ (Activity)×××✓ (Body motion)
SpO2%×××
Pulse Rate×××
PPG Waveform×××
Blood Pressure×××✓ (using 3rd partydevice)
ContinuousPhysiological datamonitoring
ComparisonReference Device -I (K183078)Reference Device -II (K200160)Reference Device-III (K160951)Reference Device-IV (K203215)Subject Device
Body TemperatureRange32.0°C - 43.0°C89.6°F - 109.4°F×-25.0°C – 43.0°C77.0°F - 109.4°F32.0°C - 43.0°C89.6°F - 109.4°F
Clinical Accuracy≤ 1°C (Meanabsolute error)×--0.27 °C(Clinical Bias)≤1.0°C(Limits of agreement)-0.27 °C (Clinical Bias)1.01°C (Limits of agreement)
MethodAdjusted mode(with SkinTemperature &AmbientTemperature)×-Adjusted mode(surfacetemperature)Adjusted mode (withSkin Temperature,Ambient Temp, HeartRate & Activity)
SpO2%Range×0 to 100%-×0 to 100%
Accuracy×± 2 % (100 to70%),Less than 70%unspecified-×± 3 % (100 to 70%),Less than 70% unspecified
PPG×-×
Signal Strength Indicator×-×
Method×LED - Red, IR &WhiteTwo Photo sensor-×LED - Red, IR & WhiteTwo Photo sensor
Pulse RateRange×30 to 250 BPM-×30 to 250 BPM
Accuracy×± 3 or 5%whichever isgreater-×± 3 or 5% whichever isgreater
R-R Interval×-×
ComparisonReference Device -I (K183078)Reference Device- II (K200160)Reference Device-III (K160951)Reference Device-IV (K203215)Subject Device
Beat Classification×-×
Sinus Heart Rate×-×
Heart Rate Variability(HRV)×-×
Monitoring Dashboard
Multi-patient Tile view××
Single Patient Zoomview×
Visual Alarm Indication×××
Viewing the Trend××
Early Warning Score(NEWS2)×××

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a waveform graphic on the left and the word "LifeSignals" in blue on the right. The waveform graphic is made up of several vertical lines of varying heights, resembling a signal or sound wave.

K242018 510(k) Summary

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the LifeSignals logo. The logo consists of a blue square with a white waveform inside, followed by the word "LifeSignals" in blue. The waveform is a series of vertical lines of varying heights, resembling a sound wave or a heartbeat signal. The text "LifeSignals" is written in a bold, sans-serif font.

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image contains the logo for LifeSignals. The logo consists of a blue square with white signal wave patterns on the left, followed by the text "LifeSignals" in blue. The text is in a sans-serif font, with the word "Life" slightly larger than "Signals".

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform inside, followed by the word "LifeSignals" in blue. The waveform is a series of vertical lines that vary in height, resembling a signal or sound wave.

K242018 510(k) Summary

Note: ✔ Feature is identical × Feature not available - Feature Unknown

{22}------------------------------------------------

Image /page/22/Picture/1 description: The image shows the logo for LifeSignals. The logo consists of a blue square with a white waveform inside, followed by the word "LifeSignals" in blue. The waveform is a stylized representation of a life signal, such as an electrocardiogram (ECG) or electroencephalogram (EEG). The logo is simple and modern, and it conveys the company's focus on life science.

  • The function of Active Monitoring portal of the Subject device at hospitals & out-of-. hospital patient care settings (such as clinics, outpatient surgery facilities, long-term care facilities and physician offices) in which care is administered by healthcare professionals is equivalent to the Reference device-III. The Reference device-III obtains data from networked patient monitors, while the subject device obtains data from the networked Biosensors. Both Subject device & Reference device-III provide centralized viewing of waveforms, numeric data and alarms from all the connected beds. Both Reference Device-III & Subject device allow for retrospective data review and can transfer data to an electronic medical record system (EMR).
    • One main difference between the Reference Device-III & Subject device is that the O Reference Device-III provides secondary alarm notification (both audio & visual), while the Subject device provides primary visual alarm notification. The Subject device Portal does not have the claims of audio alarm. The Reference Device-III software is intended to be installed in a dedicated PC that meets the audio alarm specification, while the subject device is browser-based solution & not specific to any specific hardware (PC). However, the alert notification (e.g. Email or SMS) by the Subject device can be configured to be an audible one and further not intended for critical patients. Hence, this does not add any risk.
  • The Subject device uses the same technology of three LED & two photosensors used in ● Reference device-II. However, the location of SpO2 monitoring in Subject device is on "Chest", while it is "Finger" in the Reference device-II. However, the sensor design is suitably modified for the "Chest" based measurement and the performance of SpO2% & Pulse Rate on "Chest" is validated through hypoxia tests as per ISO80601-2-61. Further, the subject device has additional controls for Chest-based measurement, such as providing visual guidance to the User to identify suitable location on the Chest for affixing the Biosensor. All the subject device has an in-built algorithm to invalidate or discard the derived SpO2% & Pulse rate value if there is a severe body motion or heavy respiration, by accessing the PPG signal & accelerometer data. Hence, this does not add any risk.

11 Summary of Performance Testing

Verification & Validation activities were performed on UbiqVue™ 2A Multiparameter System to demonstrate substantial equivalence to the predicate device:

  • Biocompatibility testing of In-vitro cytotoxicity, skin irritation and skin sensitization ● were conducted on UbiqVue™ 2A Wearable Biosensor, according to ISO 10993-1: 2018 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process.

{23}------------------------------------------------

Image /page/23/Picture/1 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform inside of it. To the right of the square is the word "LifeSignals" in a blue sans-serif font.

  • Electrical Safety and electromagnetic compatibility testing were conducted on . UbiqVue™ 2A Multiparameter System with UbiqVue™ 2A Wearable Biosensors for compliance with IEC 60601-1(Ed.3.2), IEC 60601-1-2 (Ed. 4.1) and IEC 60601-1-11(Ed. 2.1).
  • UbiqVue™ 2A Multiparameter System with UbiqVue™ 2A Wearable Biosensors (at ● Upper Center Chest location & Upper Left Chest location) were verified & validated for compliance with ISO 80601-2-61(Ed. 2.0) standard including clinical validation to determine the accuracy of SpO2% with respect to SaO2 values from simultaneously drawn arterial blood.
  • The Pulse Rate accuracy performance of UbiqVue™ 2A Multiparameter System was . verified by using bench testing, as per ISO 80601-2-61(Ed. 2.0) and through clinical validation by a reference device.
  • Body temperature accuracy performance of UbiqVue™ 2A Multiparameter System ● was validated using non-randomized, self-control comparative on-body comparative performance with oral digital thermometer on multiple subject populations in accordance with the ISO 80601-2-56 (Ed.2.0).
  • The Skin temperature accuracy performance of Ubiq Vuelle 2A Multiparameter System ● was reverified by using bench testing, as per ASTM E1112-00 & ISO80601-2-56 (Ed. 2.0).
  • ECG & Heart Rate Performance testing was repeated on the UbiqVue™ 2A . Multiparameter System for compliance with ANSI AAMI IEC 60601-2-47:2012, IEC 60601-2-27(Ed.3.0) & IEC 60601-2-25 (Ed. 2.0).
  • . Wear-life performance of ECG waveform quality, Heart Rate, Respiration Rate, Skin Temperature, and SpO2% of UbiqVue™ 2A Multiparameter System under normal use condition was validated using non-randomized, self-control comparative on-body comparative performance study for the wear period of 120 hours.
  • Improvement in the Respiration algorithm was validated by comparing against ● clinician manually scored end tidal CO2 (EtCO2) capnography, under spontaneous breathing & metronome breathing during normal activity conditions as part of wearlife study.
  • The Alert system & Visual alarm display on UbiqVue™ 2A Multiparameter System . was validated for compliance with IEC 60601-1-8 (Ed. 2.2).
  • Usability study was conducted on UbiqVue™ 2A Multiparameter System for . compliance with IEC 60601-1-6 (Ed. 3.2) and "USFDA Guidance- Applying Human Factors and Usability Engineering to Medical Devices."

{24}------------------------------------------------

Image /page/24/Picture/1 description: The image contains the logo for LifeSignals. The logo consists of a blue square with a white waveform inside of it. To the right of the square is the word "LifeSignals" in blue, sans-serif font. The logo is simple and modern, and the waveform suggests that the company is involved in the medical or healthcare industry.

  • Wireless performance & coexistence testing conducted as per ANSI/IEEE . C63.27:2017: American National Standard for Evaluation of Wireless coexistence on Ubig Vue™ 2A Multiparameter System for both Single Patient Relay System & Multi-Patient Relay System.
  • Software in the UbiqVue™ 2A Multiparameter System was designed, documented, ● verified & validated as per the IEC 62304: Medical device Software - Software Life Cycle Process and USFDA Guidance for the content of premarket submissions for Software contained in medical device. The Software for this device is determined as Class B. "Moderate-Enhanced documentation" level of concern.
  • . Software in the UbiqVue™ 2A Multiparameter System was securely designed by conforming to IEC 81001-5-1 Health software and health IT systems safety, effectiveness and security – Activities in the product life cycle.
  • Shelf-life & Reliability and testing were conducted and verified on UbiqVue™ 2A Wearable Biosensor, as per the acceptance criteria.
  • Packaging test conducted on the packaged Ubiq Vue™ 2A Wearable Biosensors as per ● ASTM D7386-16.

12 Conclusion

The UbiqVue 2A Multiparameter System is substantially equivalent in indications for use & intended use to the legally marketed Predicate Device(s) & Reference Device(s). The Technology adopted for additional physiological parameters in UbiqVue 2A Multiparameter System is substantially equivalent to the technology adopted in the Reference Device(s). Minor differences between UbiqVue 2A Multiparameter System and the Predicate & Reference devices have no significant effect on safety or effectiveness, as established through various performance tests. Thus, the proposed device/system is substantially equivalent to the legally marketed predicate & reference devices.

§ 870.2910 Radiofrequency physiological signal transmitter and receiver.

(a)
Identification. A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.(b)
Classification. Class II (performance standards).