K Number
K241884

Validate with FDA (Live)

Date Cleared
2025-03-20

(265 days)

Product Code
Regulation Number
886.5925
Age Range
15 - 49
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye.

Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.

Disposable Wear:

Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.

Frequent / Planned Replacement Wear:

Eyecare practitioners may prescribe any of the above lenses for frequent / planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.

Device Description

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is a spherical lens. It is fabricated from polymacon which has been adopted by the United Stated Adopted Names Council (USAN). The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The lens material, polymacon, is a hydrophilic polymer of 2-hydroxyethyl methacrylate (HEMA) cross-linked with ethylene glycol dimethacrylate (EGDMA), initiated by 2, 2'-azobisisobutyronitrile (AIBN). The lens consists of 62% polymacon and 38% water by weight when immersed in standard saline solution.

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is available cosmetic tinted to enhance or alter the apparent color of the eye. It is tinted in an annular pattern, providing a clear optic zone, using a combination of one or more of the following 'listed' color additives: Reactive black 5, Titanium dioxide, Iron oxides, Solvent Yellow 18, Carbazole violet, Pigment Blue 36, Phthalocyanine green.

The proposed color lenses are manufactured by sandwiching the pattern layer containing color additives between two layers of lens materials (polymacon).

In the hydrated state, HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus, and it acts as a refracting media to focus light rays on the retina.

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is available in the spherical configuration with the following features and properties: Back vertex power -10.00D to 0.00D, step: 0.25D; Base curve 8.40 mm to 8.70 mm, step: 0.10 mm; Total diameter 13.80 mm to 14.50 mm, step:0.10 mm.

The physical properties of the proposed lens are: Specific gravity 1.124; Refractive index 1.437; Light Transmission 93% ± 5%; Surface character Hydrophilic; Water content 38%; Oxygen permeability 8.99x10^-11 (cm²/s) [ml O₂/(ml·mmHg)] at 35°C.

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is supplied sterile in the foil blister pack containing the borate buffer solution.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for a contact lens, the HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens. It details the device's characteristics, its comparison to a legally marketed predicate device, and the non-clinical tests performed to demonstrate its safety and effectiveness.

Based on the provided text, no information on a study proving the device meets acceptance criteria for an AI/ML powered device, specifically in the context of human-in-the-loop performance, standalone algorithm performance, or ground truth establishment by experts, is present.

The document specifically states in Section 8: "This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply." This confirms that there was no clinical study, let alone one involving AI or expert human readers, to assess and prove the device meets acceptance criteria as envisioned in the prompt's requirements.

The "acceptance criteria" discussed in this document are related to the physical, chemical, optical, and biological safety and performance of the contact lens itself, not an AI/ML algorithm.

Therefore, I cannot provide the requested information for acceptance criteria and a study proving the device meets acceptance criteria within an AI/ML context because the document is a 510(k) premarket notification for a contact lens and does not describe such a study.

However, I can extract the acceptance criteria and performance related to the contact lens itself from the document.


Acceptance Criteria and Reported Device Performance (Non-AI/ML Medical Device)

The document primarily focuses on demonstrating the substantial equivalence of the HPM38 contact lens to a predicate device (Neo Cosmo (polymacon) Soft (hydrophilic) Contact Lens, K142275) based on physical, optical, physicochemical, and biocompatibility properties. The "acceptance criteria" are implied by meeting the standards outlined in regulatory guidances and internationally recognized standards (e.g., ISO, ASTM).

1. A table of acceptance criteria and the reported device performance:

The document presents a comparison table (Section 6) and lists performance tests (Section 7). Below is a summary of the key physical and optical properties and their reported values, which implicitly serve as the device's performance against industry and regulatory expectations for these parameters. The "acceptance criteria" are generally that these parameters fall within acceptable ranges for contact lenses and demonstrate equivalence to the predicate device.

Table: Acceptance Criteria (Implied) and Reported Device Performance for HPM38 Contact Lens

ParameterAcceptance Criteria (Implied / Predicate Range)Reported Device Performance (HPM38)
Physical & Optical Properties
Back Vertex PowerMeets standard for vision correction (Predicate: -25.00D to +25.00D)-10.00D to 0.00D, step: 0.25D
Base CurveWithin typical ranges for corneal fit (Predicate: 8.3 mm to 9.0 mm)8.40 mm to 8.70 mm, step: 0.10 mm
Total DiameterWithin typical ranges for corneal coverage (Predicate: 13.5 mm to 14.5 mm)13.80 mm to 14.50 mm, step: 0.10 mm
Center ThicknessWithin typical ranges based on power (Predicate: 0.03 mm to 0.50 mm)0.050 mm to 0.090 mm
Specific GravityConsistent with polymacon material1.124
Refractive IndexConsistent with polymacon material (Predicate: 1.428)1.437
Light Transmission>90% (Predicate: >90%)93% ± 5%
Surface CharacterHydrophilicHydrophilic
Water Content38% ± 2% (Predicate: 38% ± 2%)38% ± 2%
Oxygen PermeabilitySufficient for corneal health (Predicate: 9.77x10⁻¹¹ (cm²/s) [ml O₂/(ml·mmHg)])8.99x10⁻¹¹ (cm²/s) [ml O₂/(ml·mmHg)] at 35°C
ModulusAppropriate for handling and comfort (Predicate: 0.350 MPa)0.61 MPa
Tensile StrengthAdequate for durability (Predicate: 0.425 MPa)0.62 MPa
Elongation at BreakAdequate for material flexibility (Predicate: 127%)279%
BiocompatibilityNo cytotoxicity, ocular irritation, skin sensitization, acute systemic toxicityPassed all tests for lens and packaging
SterilizationSterility Assurance Level 10⁻⁶Validated per ISO 17665-1:2006
Shelf Life5 yearsValidated via accelerated aging per ISO 11987:2012
TransportationWithstands transportation stressesPassed per ASTM D4169-22
Compatibility with Lens CareCompatible with marketed lens care regimensDemonstrated per ISO 11981:2017
Color Additive LeachabilityColor additives are stable and do not leachDemonstrated

Since the document explicitly states that no clinical study was utilized and does not mention any AI/ML components, the remaining points of your request cannot be fulfilled as they pertain to such a study.

2. Sample size used for the test set and the data provenance: Not applicable. No test set for an AI/ML model or human reader study is described. The "test set" here refers to samples of contact lenses undergoing physical, chemical, and biological testing, not patient data for an AI.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth in an AI/ML context (e.g., for image interpretation) is not established by experts in this type of submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable in the context of AI/ML. The "ground truth" for this device's performance is established through physical measurements, chemical analyses, and biological assays (e.g., cytotoxicity tests).

8. The sample size for the training set: Not applicable. There is no mention of a training set for an AI/ML model.

9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on the first line and "ADMINISTRATION" on the second line.

March 20, 2025

Hunan Haipuming Technology Co., Ltd. % Grace Liu Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Nanshan District Shenzhen, Guangdong 518000 China

Re: K241884

Trade/Device Name: HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens (HPM38) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: February 13, 2025 Received: February 13, 2025

Dear Grace Liu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketingsafety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-deviceadvice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

J Angelo Green -S

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K241884

Device Name

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens (HPM38)

Indications for Use (Describe)

HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye.

Daily wear replacement schedules may vary from patient and should be decided by evecare practitioners in consultation with their patients.

Disposable Wear:

Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.

Frequent / Planned Replacement Wear:

Eyecare practitioners may prescribe any of the above lenses for frequent / planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary - K241884

1. Contact Details


1.1 Applicant information

Applicant NameHunan Haipuming Technology Co., Ltd.
AddressThe Intersection of Chuangye Avenue and Lvting Avenue, DahetangSub-district Office, Shaodong City, Shaoyang City, Hunan Province,China
Contact personZhang Mingxuan
Phone No.+86-739-2712889+86-18107399879
E-mail164059332@qq.com
Date Prepared2025.03.14

1.2 Submission Correspondent

Image: LogoShenzhen Joyantech Consulting Co., Ltd.
1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District,
Shenzhen, Guangdong Province, China
Phone No.+86-755-86069197
Contact personGrace Liu
Contact person's e-mailgrace@cefda.com
Websitehttp://www.cefda.com

2. Device Information

Trade nameHPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens (HPM38
Common nameSoft (Hydrophilic) Contact Lens
ClassificationII
Classification nameLenses, Soft Contact, Daily Wear
Product codeLPL, MVN
Regulation No.21 CFR 886.5925

3. Legally Marketed Predicate Device

Trade NameNeo Cosmo (polymacon) Soft (hydrophilic) Contact Lens
510(k) NumberK142275
Product CodeLPL
ManufacturerNeo Vision Co., Ltd.

The predicate device has not been subject to a design-related recall.

4. Device Description

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is a spherical lens. It is fabricated from

{5}------------------------------------------------

polymacon which has been adopted by the United Stated Adopted Names Council (USAN). The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

The lens material, polymacon, is a hydrophilic polymer of 2-hydroxyethyl methacrylate (HEMA) cross-linked with ethylene glycol dimethacrylate (EGDMA), initiated by 2, 2'-azobisisobutyronitrile (AIBN). The lens consists of 62% polymacon and 38% water by weight when immersed in standard saline solution.

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is available cosmetic tinted to enhance or alter the apparent color of the eye. It is tinted in an annular pattern, providing a clear optic zone, using a combination of one or more of the following 'listed' color additives:

Color additiveListing No.
Reactive black 521 CFR 73.3121
Titanium dioxide21 CFR 73.3126
Iron oxides21 CFR 73.3125
Solvent Yellow 1821 CFR 73.3122
Carbazole violet21 CFR 73.3107
Pigment Blue 3621 CFR 73.3110a
Phthalocyanine green21 CFR 73.3124

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens contains only the amount of color additives to accomplish the intended coloring effect. The proposed color lenses are manufactured by sandwiching the pattern layer containing color additives between two layers of lens materials (polymacon). As part of the manufacturing process, the color lenses are thoroughly washed to remove the unbound reactive color additives. Because of the 'sandwich' structure, the color additives will not be removed by lens handling and cleaning procedures.

In the hydrated state, HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus, and it acts as a refracting media to focus light rays on the retina. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution compatible with the eye.

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is available in the spherical configuration with the following features and properties:

Back vertex power-10.00D to 0.00D, step: 0.25D
Base curve8.40 mm to 8.70 mm, step: 0.10 mm
Total diameter13.80 mm to 14.50 mm, step:0.10 mm

The physical properties of the proposed lens are:

Specific gravity1.124
Refractive index1.437

{6}------------------------------------------------

Light Transmission93% ± 5%
Surface characterHydrophilic
Water content38%
Oxygen permeability$8.99\times10^{-11}$ (cm²/s) [ml O₂/(ml·mmHg)] at 35°C

HPM38 (polymacon) Daily Wear Soft (hydrophilic) Contact Lens is supplied sterile in the foil blister pack containing the borate buffer solution.

5. Intended Use/Indication for Use

HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens is indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia. The lens may be worn by persons who exhibit refractive astigmatism of 0.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available tinted and used to enhance or alter the apparent color of the eye.

Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.

Disposable Wear:

Eyecare practitioners may prescribe any of the above lenses for Daily Disposable Wear. When prescribed for Daily Disposable Wear, the lenses are not to be used with disinfecting systems as they are to be discarded after a single use.

Frequent / Planned Replacement Wear:

Eyecare practitioners may prescribe any of the above lenses for frequent / planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.

ComparisonitemProposed Device(K241884)Predicate Device(K142275)Comment
ManufacturerHunan Haipuming Technology Co.,Ltd.Neo Vision Co., Ltd./
Product NameHPM38 (polymacon) Daily Wear Soft(hydrophilic) Contact LensNeo Cosmo (polymacon) Soft(hydrophilic) Contact Lens/
RegulationNumber21 CFR 886.592521 CFR 886.5925Same
ClassificationClass IIClass IISame
PrescriptionUseYesYesSame
Intended Use /Indications forUseHPM38 (polymacon) Daily Wear Soft(Hydrophilic) Contact Lens isindicated for the correction of visualThe Neo Cosmo (Polymacon) Soft(hydrophilic) Contact Lens isindicated for daily wear for theEquivalent
acuity in aphakic and not-aphakicpersons with non-diseased eyes withcorrection of visual acuity in aphakicand non-aphakic persons with
myopia. The lens may be worn bynon-diseased eyes with myopia or
persons who exhibit refractivehyperopia. The lens may be worn by
astigmatism of 0.50 diopters or lesspersons who exhibit refractive
where the astigmatism does notstigmatism of 0.50 diopters or less
interfere with visual acuity. The lens iswhere the astigmatism does not
available tinted and used to enhanceinterfere with visual acuity. The lens is
or alter the apparent color of the eye.available clear or colored and may be
Daily wear replacement schedulesmay vary from patient to patient andused to enhance or alter the apparentcolor of the eye.
should be decided by eyecareEye care practitioners may prescribe
practitioners in consultation with theirthe lens frequent replacement wear
patients.with cleaning, disinfecting and
Disposable Wear:Eyecare practitioners may prescribeschedule replacement. The lens maybe disinfected using a chemical (notheat) lens care system only.
any of the above lenses for Daily
Disposable Wear. When prescribed
for Daily Disposable Wear, the lenses
are not to be used with disinfecting
systems as they are to be discarded
after a single use.
Frequent/PlannedReplacementWear:
Eyecare practitioners may prescribe
any of the above lenses for frequent /
planned replacement wear, with
cleaning, disinfection and scheduled
replacement. When prescribed for
frequent/planned replacement wear,
the lens may be disinfected using a
chemical disinfecting system.
FunctionalityThe contact lenses act as a refractivemedium that focus light rays fromnear and distant objects on the retina.The contact lenses act as a refractivemedium that focus light rays fromnear and distant objects on the retina.Same
Lens DesignSphericalSpherical, ToricEquivalent
ManufacturingProcessLathe-cut(semi-molded)Lathe-cut(semi-molded)Same
Lens MaterialPolymaconPolymaconSame
MaterialClassificationGroup ILow water content (<50%), NonionicGroup ILow water content (<50%), NonionicSame
Color AdditivesReactive Black 5Reactive Black 5Different
(21CFR 73.3121)Titanium Dioxide(21 CFR 73.3126)Iron Oxides(21CFR 73.3125)Solvent Yellow 18(21CFR 73.3122)Carbazole Violet(21CFR 73.3107)Pigment Blue 36(21 CFR 73.3110a)Phthalocyanine Green(21 CFR 73.3124)(21CFR 73.3121)Titanium Dioxide (White)(21 CFR 73.3126)Iron Oxides(Red)(21CFR 73.3125)C.I Pigment Green 7(21CFR 73.3124)(Phthalocyaninato(2-)) Copper(21CFR 74.3045)
Storage SolutionBorate buffer solutionNaCl 0.85~0.95%(w/v) solutionDifferent
Primary PackagingBlister packBlister pack and VialEquivalent
Total Diameter13.80 mm to 14.50 mm13.5 mm to 14.5 mmEquivalent
Base Curve8.40 mm to 8.70 mm8.3 mm to 9.0 mmEquivalent
Center Thickness(varies with power)0.050 mm to 0.090 mm0.03 mm to 0.50 mmEquivalent
Power Range-10.00D to 0.00D-25.00D to +25.00DEquivalent
Water Content38% ±2 %38% ±2 %Same
Spectral Transmission93% ± 5%>90%Same
Refraction Index1.4371.428Same
Oxygen Permeability8.99×10-11 (cm²/s) [ml O₂/(ml·mmHg)]at 35°C9.77×10-11 (cm²/s) [ml O₂/(ml·mmHg)]at 35°CDifferent
Modulus0.61 MPa0.350 MPaDifferent
Tensile Strength0.62 MPa0.425 MPaDifferent
Elongation at break279%127%Different
SterilizationMoist Heat SterilizationMoist Heat SterilizationSame
BiocompatibilityBiocompatibleBiocompatibleSame

6. Substantial Equivalence Comparison

{7}------------------------------------------------

{8}------------------------------------------------

HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens has the similar indication for use as the predicate device as well as similar technical characteristics, and the differences don't raise any additional questions for safety and effectiveness, therefore, the proposed device is substantially equivalent to the predicate device.

7. Summary of Non-clinical Testing

{9}------------------------------------------------

Non-clinical tests were conducted in accordance with the FDA guidance - Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses" issued on June 27, 1994. The non-clinical performance tests had demonstrated the safety and effectiveness of HPM38 (polymacon) Daily Wear Soft (Hydrophilic) Contact Lens and the substantial equivalence to the predicate device.

> Performance Testing

The physical, optical, physicochemical performance tests were performed in accordance with ISO 18369-2:2017, ISO 18369-3:2017, ISO 18369-4:2017 and ASTM D882-18, and the test results showed that the proposed device meets the requirements.

  • Appearance
  • । Physical parameters (Total diameter, Base curve, Center thickness)
  • Optical parameter (Back vertex power) ।
  • Spectral transmission ।
  • । Refractive index
  • Water content ।
  • -Specific gravity
  • i Extractables
  • i Oxygen permeability
  • Mechanical properties i
  • pH and Osmolality of the storage solution ।

In addition, the compatibility test of the lens with the lens care regimen has been conducted as per ISO 11981:2017, and the results demonstrate the proposed device is compatible with the marketed lens care regimens. The leachability of the color additives has also been conducted, and the results also demonstrate that the color additives contained in the proposed device are stable.

> Biocompatibility Testing

The following biocompatibility tests were performed respectively on the contact lens and primary packaging according to FDA guidance - Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses" issued on June 27, 1994, and the test results demonstrated that the contact lens and primary packaging have no cytotoxicity, no ocular irritation, no skin sensitization and no acute systemic toxicity.

  • Cytotoxicity (ISO 10993-5:2009) ।
  • i Ocular Irritation (ISO 10993-23:2021)
  • i Skin Sensitization (ISO 10993-10:2021)
  • Acute System Toxicity (ISO 10993-11:2017) ।

In addition, the following biocompatibility tests were performed on the storage solution, and the test

{10}------------------------------------------------

results demonstrated that the storage solution had no cytotoxicity, no ocular irritation and no skin sensitization.

  • Cytotoxicity (ISO 10993-5:2009) ।
  • Ocular Irritation (ISO 10993-23:2021) i
  • Skin Sensitization (ISO 10993-10:2021)

In conclusion, the results of the above biocompatibility tests showed that the proposed device has no biocompatibility issues.

A Sterilization

A Sterility Assurance Level (SAL) of 10° has been validated in accordance with the requirements of ISO 17665-1:2006.

A Shelf Life

Accelerated aging testing has been performed to assure the shelf-life of 5 years in accordance with ISO 11987:2012.

A Simulated Transportation Testing

The simulated transportation testing has been conducted according to ASTM D4169-22.

8. Clinical Testing

This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply.

9. Conclusions

The results of comparing the design specifications and non-clinical testing between the proposed device and the legally marketed predicate device (K142275) show that they are Substantially Equivalent (SE).

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.