K Number
K241847

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-08-30

(65 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Imbio PHA Software device is designed to measure the maximal diameters of the right and left ventricles of the heart, the main pulmonary artery, and the ascending aorta from a volumetric CTPA acquisition and report the RV/LV and Pa/Ao ratios. PHA analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the anatomy. The PHA software provides the user with annotated images showing measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.

Device Description

Imbio's PHA Software is a set of medical image post-processing computer algorithms that together perform automated right and left ventricle, pulmonary artery (PA), and aorta (Ao) measurements from CT pulmonary angiography (CTPA) scans, and reports the RV/LV and PA/Ao ratios. The Imbio PHA is a single command-line executable program that may be run directly from the or through scripting and thus the user interface is minimal. Imbio PHA Software is a Software as a Medical Device (SaMD) intended to provide annotated images and a PDF report that will be read most typically at a PACS workstation. Imbio PHA Software is an aid, only used to support a physician in the analysis of CTPA images. The Imbio PHA Software program reads in CTPA DICOM datasets, processes the data, then writes output DICOM files and summary reports to a specified directory. Imbio PHA Software outputs DICOM CT images overlaid with color-codings indicating where the measurements were made. Additionally, a summary PDF report is output. Imbio PHA Software does not interface directly with any CT scanner or data collection equipment; instead the software imports data previously generated by such equipment and is integrated as part of the radiological workflow, reducing the risk of use errors.

AI/ML Overview

The Imbio PHA (4.0.0) device is intended to measure the maximal diameters of the right and left ventricles of the heart, the main pulmonary artery, and the ascending aorta from a volumetric CTPA acquisition and report the RV/LV and Pa/Ao ratios. The device utilizes an artificial intelligence algorithm for identification and measurement of the anatomy, providing annotated images to the user. Its results are not intended for stand-alone clinical decision-making.

Here's a breakdown of the acceptance criteria and the study proving the device meets these criteria:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Imbio PHA (4.0.0) device were based on the Intra-class Correlation Coefficient (ICC) of measurements.

Acceptance CriteriaReported Device PerformanceConclusion
ICC of all measurements (annotators and algorithm) > 0.90 (excellent)ICC = 0.94Met
ICC of all measurements (annotators and algorithm) within 95% CI of the ICC of only annotators(Specific CI not provided, but the overall ICC of 0.94 indicates strong agreement)Inferred as Met (given "excellent agreement")

The reported device performance, with an ICC of 0.94, indicates excellent agreement between the algorithm's measurements and those performed by expert radiologists.

2. Sample Size and Data Provenance

  • Test Set Sample Size: 100 contrast-enhanced CT pulmonary angiography (CTPA) datasets.
  • Data Provenance: The datasets were sourced from multiple centers and multiple databases. While specific countries are not mentioned, the experts establishing ground truth were U.S. board-certified radiologists practicing within the United States. The data was likely retrospective, as it mentions "datasets were sourced."

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Three expert radiologists.
  • Qualifications: U.S. board-certified radiologists practicing within the United States, each with a minimum of 10 years of clinical experience.

4. Adjudication Method for the Test Set

The adjudication method involved each radiologist independently measuring the diameter of the pulmonary artery trunk at the bifurcation and the aorta diameter at the same slice. Intra-class correlation coefficients were then calculated to show equivalency between the radiologists and between the radiologists and the algorithm. This implies a comparison of individual expert measurements rather than a formal consensus adjudication where experts reconcile their differences to arrive at a single ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was specifically mentioned that directly assesses the improvement of human readers with AI assistance versus without AI assistance. The study focused on the agreement between the AI algorithm and expert radiologists' measurements.

6. Standalone Performance (Algorithm Only)

Yes, a standalone performance assessment was done. The study specifically states, "100 contrast-enhanced CT pulmonary angiography (CTPA) datasets were used for standalone performance assessment." The ICC calculation of 0.94 directly reflects the agreement between the algorithm's measurements and the expert radiologists' measurements, indicating its standalone performance in providing measurements consistent with human experts.

7. Type of Ground Truth Used

The type of ground truth used was expert consensus/agreement based on measurements. Specifically, the ground truth for the pulmonary artery and aorta diameter measurements was established by three expert radiologists, and the agreement between these experts and the algorithm was quantified using ICC.

8. Sample Size for the Training Set

The document does not specify the sample size for the training set. It only describes the test set used for performance assessment.

9. How Ground Truth for Training Set Was Established

The document does not provide information on how the ground truth for the training set was established. It only details the ground truthing process for the performance (test) datasets.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG' are written in blue, followed by 'ADMINISTRATION' in a smaller font size.

Imbio, Inc Lauren Keith Director of Engineering 1015 Glenwood Ave Floor 4 Minneapolis, Minnesota 55405

Re: K241847

August 30, 2024

Trade/Device Name: Imbio PHA (4.0.0) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: August 6, 2024 Received: August 6, 2024

Dear Lauren Keith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K241847

Device Name

Imbio PHA (4.0.0)

Indications for Use (Describe)

The Imbio PHA Software device is designed to measure the maximal diameters of the right and left ventricles of the heart, the main pulmonary artery, and the ascending aorta from a volumetric CTPA acquisition and report the RV/LV and Pa/Ao ratios. PHA analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the PHA software provides the user with annotated images showing measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) #:K241847510(k) SummaryPrepared on: 2024-08-27
Contact Details21 CFR 807.92(a)(1)
Applicant NameImbio, Inc
Applicant Address1015 Glenwood Ave Floor 4 Minneapolis MN 55405 United States
Applicant Contact Telephone608-213-7100
Applicant ContactDr. Lauren Keith
Applicant Contact Emaillaurenkeith@imbio.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameImbio PHA (4.0.0)
Common NameMedical image management and processing system
Classification NameAutomated Radiological Image Processing Software
Regulation Number892.2050
Product Code(s)QIH
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K203256Imbio RV/LVQIH
Device Description Summary21 CFR 807.92(a)(4)

{4}------------------------------------------------

lmbio's PHA Software is a set of medical image post-processing computer algorithms that together perform automated right and left ventricle, pulmonary artery (PA), and aorta (Ao) measurements from CT pulmonary angiography (CTPA) scans, and reports the RV/LV and PA/Ao ratios. The Imbio PHA is a single command-line executable program that may be run directly from the or through scripting and thus the user interface is minimal.

lmbio PHA Software is a Software as a Medical Device (SaMD) intended to provide annotated images and a PDF report that will be read most typically at a PACS workstation. Imbio PHA Software is an aid, only used to support a physician in the analysis of CTPA images.

The Imbio PHA Software program reads in CTPA DICOM datasets, processes the data, then writes output DICOM files and summary reports to a specified directory. Imbio PHA Software outputs DICOM CT images overlaid with color-codings indicating where the measurements were made. Additionally, a summary PDF report is output.

lmbio PHA Software does not interface directly with any CT scanner or data collection equipment; instead the software imports data previously generated by such equipment and is integrated as part of the radiological workflow, reducing the risk of use errors.

100 contrast-enhanced CT pulmonary angiography (CTPA) datasets were used for standalone performance assessment. The datasets were sourced from multiple centers and multiple databases and included a range of convolution kernels (vendor equivalent of GE's Soft, Detail, Standard, and Standard Iterative kernels) and sice thicknesses (0.75 mm, 1.5 mm, and 2.00 mm) from the four major CT scanner vendors. The datasets are representative of the intended patients because they are all cinically indicated for a CTPA scan, which is primarily indicated for suspicion of pulmonary embolism or pulmonary hypertension.

Three expert, U.S. board-certified radiologists practicing within the United States were enrolled to perform ground truthing of the performance datasets. Each truther had a minimum of 10 years clinical experience. The ground truthing exercise consisted of each radiologist measuring the diameter of the pulmonay artery trunk at the bifurcation and the aorta diameter at the same slice. Intra-class correlation coefficients were calculated to show equivalency. There were two acceptance criterion:

*ICC of all measurements (annotators and algorithm) > 0.90 (excellent)

*ICC of all measurements (annotators and algorithm) within 95% Cl of the ICC of only annotators

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Imbio PHA Software device is designed to maximal diameters of the right and left ventricles of the heart, the main pulmonary artery, and the ascending aorta from a volumetric CTPA acquisition and report the RV/LV and Pa/Ao ratios. PHA analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the anatomy. The PHA software provides the user with annotated images showing measurements. Its results are not intended to be used on a stand-alone basis for clinical decisionmaking or otherwise preclude clinical assessment of CTPA cases.

Indications for Use Comparison

The Indications for Use differ slightly between the proposed devices. The FUs have been reworded slightly to include the main pulmonary artery and ascending aorta measurements. The overall intended use of the devices are equivalent: to use deeplearning models to measure diameters of clinically relevant cardiovascular strom a CTPA image. The differences in IFU statements have no neqative impact on safectiveness because neither the predicate's nor the subject device's results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.

Technological Comparison

21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

{5}------------------------------------------------

Proposed Device Characteristics Name: Imbio PHA 510(k) Number: TBD Product Code: OlH Requlation Name: Picture Archiving and communications systems Regulation Number: 21 CFR 892.2050 lnput Data Requirements - General: Non-qated, CT Pulmonary Angiography images DICOM compliant: Yes LV Segmentation: Yes - Deep Learning RV Segmentation: Yes - Deep Learning mPA Segmentation: Yes - Deep Learning Ao Segmentation: Yes - Deep Learning Diameters Measurements: Yes - Automated Fully-automated: Yes Interface: Command-line Ability to QA Results: Yes - via outputs Outputs: Report, DICOM Secondary Capture series

Proposed Indications for Use: The Imbio PHA Software device is designed to measure the maximal diameters of the right and left ventricles of the heart, the main pulmonary and the ascending aorta from a volumetric CTPA acquisition and report the RV/LV and Pa/Ao ratios. PHA analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the anatomy. The PHA software provides the user with annotated images showing measurements. Its results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of CTPA cases.

Predicate Device Characteristics Name: Imbio RV/LV 510(k) Number: K203256 Product Code: QIH Regulation Name: Picture Archiving and communications systems Requlation Number: 21 CFR 892.2050 Input Data Requirements - General: Non-qated, CT Pulmonary Angiography images DICOM compliant: Yes LV Segmentation: Yes - Deep Learning RV Segmentation: Yes - Deep Learning mPA Segmentation: No Ao Segmentation: No Diameters Measurements: Yes - Automated Fully-automated: Yes Interface: Command-line Ability to QA Results: Yes - via outputs Outputs: Report, DICOM Secondary Capture series

Predicate Indications for Use: The Imbio RV/LV Software device is designed to maximal diameters of the right and left ventricles of the heart from a volumetric CTPA acquisition and report the ratio of those measurements. RV/LV analyzes cases using an artificial intelligence algorithm to identify the location and measurements of the RV/LV software provides the user with annotated images showing ventricular measurements. Its results are not intended to be used on a stand-alone basis for clinical decisionmaking or otherwise preclude clinical assessment of CTPA cases.

{6}------------------------------------------------

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

No clinical testing was conducted. The Imbio PHA Software provides quantitative metrics with well-known clinical value. Non-Clinical testing was conducted to ensure that the measurements provided by the PHA Software, specifically the PA and Ao diameters, resemble measurements of the same anatomy performed by radiologists. The ICC between three expert radiologists and the algorithm results is 0.94, indicating excellent agreement.

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective as Imbio RV/LV (K203256).

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).