K Number
K241731
Manufacturer
Date Cleared
2025-08-11

(420 days)

Product Code
Regulation Number
874.4760
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Outlook Surgical Versa One System is intended for use in examination and visualization in otolaryngology and head and neck procedures.

Device Description

The Outlook Surgical Versa One System is a video nasopharyngoscopy system that provides illumination and live video stream of a patient's internal anatomy during otolaryngology and head and neck examinations and procedures. The device is composed of an endoscope, a connection cable available in 6ft or 12ft lengths, a wireless enabled display unit, and a wireless HDMI receiver. Users attach accessory instruments to the semi-flexible shaft of the endoscope enabling the user to visualize with the instrument attached.
The device will be used by clinical practitioners in a sterile field in an operating room, healthcare clinic, or medical office. The Outlook Surgical Versa One System is a reusable device initially supplied as non-sterile to the user and requiring the user to process (i.e., clean and sterilize) the device for initial use, as well as to reprocess the device after each use.
The Versa One Endoscope has a semi-flexible shaft that enables the user to attach instruments. The semi-flexible shaft is rigid near the adapter end and semi-flexible at the distal end to conform the shape of the attached instrument. When an instrument is attached to the scope, the flexible portion becomes rigid and the scope functions as a rigid endoscope. The Versa One endoscope is compatible with Versa One straight and curved sheaths.
The Outlook Surgical Versa One System live stream video can be viewed on the display unit and an off-the-shelf monitor connected to the display unit via an HDMI cable or wirelessly. The off-the shelf monitor is not provided with the device. The device operates off rechargeable battery power.
The Versa One Endoscope uses an integrated LED light source to illuminate glass fiber bundles in the shaft and a CMOS sensor at the distal tip to send the signal to a custom PCB in the endoscope handle that converts the signal to live video.

AI/ML Overview

Based on the provided FDA 510(k) Clearance Letter, the "Outlook Surgical Versa One System" is a medical device for examination and visualization in otolaryngology and head and neck procedures, and not an AI/ML powered device. Therefore, the information typically required for an AI/ML device's acceptance criteria, such as a table of performance metrics (sensitivity, specificity, AUROC), sample size for test/training sets, expert qualifications, and comparison to human readers, is not present in this document.

The document discusses the substantial equivalence of the "Outlook Surgical Versa One System" to a predicate device (Stryker Precision S 4K Sinuscope) based on physical and functional characteristics, and a range of non-AI/ML specific performance tests.

Here's an analysis of what information is provided regarding the device's performance and acceptance, and why the requested AI/ML specific details are absent:

Acceptance Criteria and Device Performance (as per the document):

The acceptance criteria for this device are demonstrated through various performance tests, with the reported device performance consistently being "Passed. All acceptance criteria were met."

TestTest Method SummaryResults (Device Performance / Acceptance)
Performance Testing: Simulated Use/CyclingVersa One endoscope tested through multiple simulated use procedures and repeated instrument attachment/detachment cycles.Passed. All acceptance criteria were met. Demonstrated intended performance as a rigid scope and safe use throughout its minimum expected life.
User ValidationMultiple users followed IFU to set up and use the system with an anatomical model and confirmed results in a cadaver model, following FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices" and IEC 62366-1.Passed. All users were able to choose appropriate attachment, attach instrument, and perform simulated procedure as intended.
Performance Testing: Optical PerformanceTested according to ISO 8600-X, ISO 12233, and ISO 15739 for optical performance. Evaluated signal to noise ratio, dynamic range, image quality (geometric distortion, MTF, FOV), image intensity uniformity, color accuracy, and endoscope image latency.Passed. All devices met performance criteria.
Performance Testing: Simulated Use (Specifics)Tested automatic exposure control, wireless link distance, Display power up, battery charge, battery status indicator, image rotation, HDMI port activity. Evaluated shaft durability and bend radius using an automated process.Passed. All systems met performance criteria. Demonstrated reliable performance and wireless link reliability.
Performance Testing: Electrical Requirements (non 60601-1)Inspected Versa One Display and Wireless HD Receiver for electrical requirements not included in IEC 60601-1 or IEC 60601-2-18 testing. Inspected Device charger, Display MOPP, isolation barriers, and HDMI connectors.Passed. All devices met criteria. Demonstrated safe and reliable electrical performance.
Performance Testing: Dimension, Material, EnvironmentalTested patient-contacting surface temperature, water splash, ambient temperature, dimensional and material requirements, cable force measurement, and leak testing of the distal end.Passed.
Electrical Safety, electromagnetic compatibility (EMC), and RF SafetyTested in compliance with IEC 60601-1, IEC 60601-2-18 (electrical safety), IEC 60601-1-2 (EMC), IEC 63195-1, and ANSI/IEEE C95.1-1 (RF Safety).Passed.
Photobiological SafetyTested in accordance with IEC 62471.Passed.
Cleaning ValidationTested per FDA Guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," AAMI TIR12 and AAMI ST98.Passed.
Biocompatibility TestingTested per FDA Guidance, "Use of International Standard ISO 10993-1," ISO 10993-5, ISO 10993-10, ISO 10993-23.Passed. Non-cytotoxic, non-sensitizing, non-irritating, non-systemically toxic, and non-pyrogenic.
Sterilization ValidationFor scope: ISO 22441:2022 (low temp vaporized hydrogen peroxide). For sheaths: ISO 11135:2014 (ethylene oxide). Overkill test method used for scope.Passed.

Regarding the specific AI/ML and clinical study requirements you listed, the document does not provide the following information:

  1. Sample size used for the test set and the data provenance: Not applicable as this is not an AI/ML device with a test set of data. The "test sets" here refer to physical devices undergoing various engineering and biological tests.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an imaging device like this is primarily about validating its physical performance against standards and its ability to capture high-quality images, not expert interpretation of AI outputs. "User Validation" involved multiple users, but their qualifications are not specified beyond being "clinical practitioners" in the "use environment."
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For this device, "ground truth" is established through engineering specifications, international standards (ISO, IEC, ANSI/IEEE), and regulatory guidance documents (FDA). For instance, optical performance is validated against ISO standards, and biocompatibility against ISO 10993. The "User Validation" involved using anatomical models and cadaver models, which serve as a form of ground truth for assessing usability and visualization capability.
  7. The sample size for the training set: Not applicable. This is not an AI/ML device that requires a training set.
  8. How the ground truth for the training set was established: Not applicable.

In summary, the provided 510(k) clearance letter details the substantial equivalence of a medical imaging device (a nasopharyngoscope) by demonstrating its physical, electrical, and performance characteristics meet established safety and effectiveness standards, and are comparable to a predicate device. It is not an AI/ML device, and therefore the specific criteria, studies, and data provenance related to AI/ML device performance are not present in this document.

FDA 510(k) Clearance Letter - Outlook Surgical Versa One System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 11, 2025

ResnENT, LLC
℅ Billi-Jo Pfalzgraf
Regulatory Consultant
MedEdge Consulting
7005 Evergreen Pl
Roswell, Georgia 30076

Re: K241731
Trade/Device Name: Outlook Surgical Versa One System (8900139)
Regulation Number: 21 CFR 874.4760
Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOB
Dated: July 29, 2025
Received: July 29, 2025

Dear Billi-Jo Pfalzgraf:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.


August 11, 2025

ResnENT, LLC
℅ Billi-Jo Pfalzgraf
Regulatory Consultant
MedEdge Consulting
7005 Evergreen Pl
Roswell, Georgia 30076

Re: K241731
Trade/Device Name: Outlook Surgical Versa One System (8900139)
Regulation Number: 21 CFR 874.4760
Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOB
Dated: July 29, 2025
Received: July 29, 2025

Dear Billi-Jo Pfalzgraf:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 2

K241731 - Billi-Jo Pfalzgraf Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K241731 - Billi-Jo Pfalzgraf Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joyce C. Lin -S

for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K241731

Device Name
Outlook Surgical Versa One System (8900139)

Indications for Use (Describe)
The Outlook Surgical Versa One System is intended for use in examination and visualization in otolaryngology and head and neck procedures.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Outlook Surgical Versa One System 510(k)

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date: August 8, 2025

1.0 SUBMITTER:

ResnENT, LLC
3210 Mulligan Drive
Bloomington, IL 61704

1.1 PRIMARY CONTACT PERSON:

Billi-Jo Pfalzgraf
Regulatory Consultant
MedEdge Consulting
(704) 840-8453
bjpfalzgraf@mededge.io

1.2 SECONDARY CONTACT PERSON:

Tiffini Wittwer
Regulatory Consultant and Owner
MedEdge Consulting
(707) 799-6732
twittwer@mededge.io

2.0 DEVICE:

Trade Name: Outlook Surgical Versa One System
Common/Usual Name: Flexible Nasopharyngoscope System
Classification Name: Nasopharyngoscope (flexible or rigid) (21 CFR 874.4760)
Regulatory Class: II
Product Code: EOB

3.0 PREDICATE DEVICE:

The subject device is equivalent to the following devices:

  • K191102 Stryker Precision S 4K Sinuscope

4.0 DEVICE DESCRIPTION:

The Outlook Surgical Versa One System is a video nasopharyngoscopy system that provides illumination and live video stream of a patient's internal anatomy during otolaryngology and head and neck examinations and procedures. The device is composed of an endoscope, a connection cable available in 6ft or 12ft lengths, a wireless enabled display unit, and a wireless HDMI receiver. Users attach accessory instruments to the semi-flexible shaft of the endoscope enabling the user to visualize with the instrument attached.

The device will be used by clinical practitioners in a sterile field in an operating room, healthcare clinic, or medical office. The Outlook Surgical Versa One System is a reusable device initially

Page 6

supplied as non-sterile to the user and requiring the user to process (i.e., clean and sterilize) the device for initial use, as well as to reprocess the device after each use.

The Versa One Endoscope has a semi-flexible shaft that enables the user to attach instruments. The semi-flexible shaft is rigid near the adapter end and semi-flexible at the distal end to conform the shape of the attached instrument. When an instrument is attached to the scope, the flexible portion becomes rigid and the scope functions as a rigid endoscope. The Versa One endoscope is compatible with Versa One straight and curved sheaths.

The Outlook Surgical Versa One System live stream video can be viewed on the display unit and an off-the-shelf monitor connected to the display unit via an HDMI cable or wirelessly. The off-the shelf monitor is not provided with the device. The device operates off rechargeable battery power.

The Versa One Endoscope uses an integrated LED light source to illuminate glass fiber bundles in the shaft and a CMOS sensor at the distal tip to send the signal to a custom PCB in the endoscope handle that converts the signal to live video.

5.0 INDICATIONS FOR USE:

The Outlook Surgical Versa One System is intended for use in examination and visualization in otolaryngology and head and neck procedures.

6.0 SUBSTANTIAL EQUIVALENCE:

The Outlook Surgical Versa One System is similar to the predicate in several ways. Both devices have the same intended use, indication for use, and the same use environment. Both devices illuminate and provide visualization of the patient's anatomy. The subject and predicate devices perform as rigid endoscopes with the same direction of view and lack of articulation control and internal channel. Both devices are reusable and use similar methods of reprocessing. Performance and safety testing has demonstrated that technical differences raise no new questions of safety and effectiveness. Thus, the subject Outlook Surgical Versa One System is substantially equivalent to the predicate Stryker Precision S 4K Sinuscope.

The table below compares the intended use and technological characteristics of the subject and predicate device.

FeatureOutlook Surgical Versa One System (Subject Device)Stryker Precision S 4K Sinuscope (Predicate Device)Similarities and Differences
510(k)K241731K191102
Common NameNasopharyngoscope SystemSinuscope
Classification NameNasopharyngoscope (flexible or rigid) and AccessoriesNasopharyngoscope (flexible or rigid) and AccessoriesSame
Device ClassClass IIClass IISame
Regulation21 CFR 874.476021 CFR 874.4760Same
Product CodeEOBEOBSame
Indication for UseThe Outlook Surgical Versa One System is intended for use in examination andThe Precision S 4K Sinuscope is intended for use in otolaryngology andSimilar. Narrower indication for use does not

Page 7

FeatureOutlook Surgical Versa One System (Subject Device)Stryker Precision S 4K Sinuscope (Predicate Device)Similarities and Differences
visualization in otolaryngology and head and neck procedures.Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.raise new questions of safety or effectiveness.
Use EnvironmentMedical facilitiesMedical facilitiesSame
Principle of OperationAttach proximal end to provided power source, attach scope shaft to instrument to create rigid scope configuration and insert into nose or mouthAttach proximal end to light source and camera and insert rigid scope into nose or mouthPerformance testing and User Validation demonstrate this difference does not raise new questions of safety and effectiveness.
Image transmissionCMOS sensor at distal tip transfers images to integrated handheld monitor with video processor to display live stream images.Rod lenses with camera/monitor connected to sinuscope eyepieceOptical performance testing demonstrates this difference does not raise new questions of safety and effectiveness.
Wireless video transmissionYes. Optional wireless data transfer between handheld monitor and wireless receiver.NoSafety testing in accordance with IEC 60601-1, IEC 60601-2-18, IEC 63195-1, and ANSI/IEEE C95.1 demonstrates this difference does not raise new questions of safety and effectiveness.
Internal working channelNoNoSame
Shaft typeRigid when in use (attached to instrument)Rigid shaftSame
Required AccessoriesSheath required for useNoUser Validation and Performance testing demonstrate this difference does not raise new questions of safety and effectiveness.
Shaft Outer Diameter3.0mm, 3.8mm3.1mm, 4.0mmPerformance and User Validation demonstrates this difference does not raise new questions of safety or effectiveness.
Shaft Working Length102mm - 105mm125mm - 180mmPerformance and User Validation testing demonstrate this difference does not raise new questions of safety and effectiveness.
Articulation ControlNoNoSame

Page 8

FeatureOutlook Surgical Versa One System (Subject Device)Stryker Precision S 4K Sinuscope (Predicate Device)Similarities and Differences
Scope Direction of View0° - 70°Same
Scope Field of View120°80° - 105°User Validation and Performance testing demonstrate this difference does not raise new questions of safety and effectiveness.
Scope Power SourceRechargeable Li-ion battery or AC power from control box through cable connected to scope.Light source powered externally and coupled to scope.Electrical safety testing in accordance with IEC 60601-1 and 60601-2-18 demonstrate this difference does not raise new questions of safety and effectiveness.
Illumination OutputIllumination fibers; commercially available sourceIllumination fibers, commercially available sourceSame
Light SourceIntegrated LEDExternal light source coupled to scopeElectrical safety testing in accordance with IEC 60601-1 and 60601-2-18 and Photobiological testing in accordance with IEC 62471 demonstrates this difference does not raise new questions of safety and effectiveness.
Key Patient-Contacting Materials304 stainless steel, glass, borosilicate, crosslinked polyvinylidene fluoride, epoxyStainless Steel, optical glass, glass fibersBiocompatibility testing in accordance with ISO 10993 and performance testing demonstrates this difference does not raise new questions of safety and effectiveness.
Single Use or ReusableReusable – scope Single Use – accessoriesReusableSterilization validation in accordance with ISO 11135 and ISO 10993-7 demonstrates this difference does not raise new questions of safety and effectiveness.
Scope ReprocessingManual cleaning followed by APS STERRAD SterilizationManual and automated cleaning, manual and automated disinfection, autoclave, Steris VPRO, and APS STERRAD sterilizationSimilar

Page 9

7.0 PERFORMANCE DATA:

The table below lists all of the testing that has been performed on the Outlook Surgical Versa One System.

TestTest Method SummaryResults
Performance Testing: Simulated Use/CyclingThe Versa One endoscope was tested to show the device performs as intended through multiple simulated use procedures and repeated instrument attachment/detachment cycles.Passed. All acceptance criteria were met. The simulated use testing is relevant to determining substantial equivalent because it demonstrates the subject device performs as intended as a rigid scope when attached to an instrument. The cycling testing demonstrates that the subject device performs as expected and can be used safely throughout its minimum expected use life.
User ValidationFDA Guidance document "Applying Human Factors and Usability Engineering to Medical Devices", IEC 62366-1. Multiple users were presented with all device components and followed the IFU to set up and use the system using an anatomical model and to confirm results in a cadaver model.Passed. All users were able to choose the appropriate attachment for a procedure, attach the instrument to the scope, and perform the simulated procedure with the device. This test is relevant to determining substantial equivalence because it demonstrates users can use the device as intended in the expected use environment.
Performance Testing: Optical PerformanceTesting was conducted according to ISO 8600-X, ISO 12233, and ISO 15739 for optical performance. Image capture and analysis software was used to evaluate signal to noise ratio and dynamic range, image quality (geometric distortion, MTF, and FOV), image intensity uniformity, and color accuracy. Endoscope image latency was also measured.Passed. All devices met performance criteria. This testing is relevant in determining substantial equivalence because the subject device and predicate are both intended to illuminate and visualize patient anatomy.
Performance Testing: Simulated UseThe Outlook Surgical Versa One System was tested for automatic exposure control, wireless link distance, Display power up, battery charge, battery status indicator, image rotation, and HDMI port activity. Versa One Endoscopes were evaluated for shaft durability and bend radius using an automated process.Passed. All systems met performance criteria. This testing is relevant in determining substantial equivalence because it demonstrates the subject device reliably performs as intended. The reliability of the wireless link distance supports substantial equivalence of the subject device to the predicate device with wireless transmission of images.

Page 10

TestTest Method SummaryResults
Performance Testing: Electrical Requirements (non 60601-1)The Versa One Display and Wireless HD Receiver were inspected for electrical requirements not included in IEC 60601-1 or IEC 60601-2-18 testing. Device charger and Display MOPP, isolation barriers, and HDMI connectors were inspected.Passed. All devices met criteria. This test is relevant in determining substantial equivalence because is demonstrates the subject device safely and reliably performs as intended.
Performance Testing: Dimension, Material, EnvironmentalVersa One Endoscopes were tested for patient-contacting surface temperature, water splash testing, ambient temperature testing, dimensional and material requirements, cable force measurement and leak testing of the distal end.Passed. This is relevant to determining substantial equivalence because the subject device and predicate used to illuminate and visualize patient anatomy and have the same use environment.
Electrical Safety, electromagnetic compatibility (EMC), and RF SafetyTesting was conducted in compliance with the IEC 60601-1 and IEC 60601-2-18 for electrical safety, IEC 60601-1-2 for EMC, IEC 63195-1, and ANSI/IEEE C95.1-1 for RF SafetyPassed. This test is relevant in determining substantial equivalence because the it demonstrates the electrical safety, electromagnetic compatibility, and RF safety of the subject device.
Photobiological SafetyTesting was conducted in accordance with IEC 62471 for photobiological safety.Passed. This test is relevant in determining substantial equivalence because it demonstrates the photobiological safety of the subject device.
Cleaning ValidationFDA Guidance document Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, AAMI TIR12 and AAMI ST98.Passed. This test is relevant in determining substantial equivalence because the subject device and predicate device are both cleaned manually.
Biocompatibility TestingFDA Guidance, "Use of International Standard ISO 10993-1, 'Biological evaluation and testing within a risk management Process' and 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. (ISO 10993-5, ISO 10993-10, ISO 10993-23)Passed. Non-cytotoxic, non-sensitizing, non-irritating, non-systemically toxic, and non-pyrogenic. This test is relevant in determining substantial equivalence because it demonstrates any material differences between the subject device and the predicate do not raise new questions of safety.
Sterilization ValidationFor scope: ISO 22441:2022 Sterilization of health care products, low temperature vaporized hydrogen peroxide – Requirements for the development, validation and routine control of sterilization process for medical devices. The Overkill test method was used. For sheaths: ISO 11135:2014 Sterilization of health care products –Passed. This test is relevant in determining substantial equivalence because the subject device and predicate device are both reprocessed with APS STERRAD sterilization. Sterilization validation for the sheaths is relevant in determining substantial equivalence because it demonstrates the sterile, single-use accessories do not raise

Page 11

| | Ethylene oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices. | new questions of safety or effectiveness. |

8.0 CONCLUSION:

The results of the performance testing described above demonstrate that the Outlook Surgical Versa One System is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 874.4760 Nasopharyngoscope (flexible or rigid) and accessories.

(a)
Identification. A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.