K Number
K241713
Device Name
HiRise (1040-230)
Manufacturer
Date Cleared
2024-07-12

(28 days)

Product Code
Regulation Number
892.1750
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The HiRise is intended to be used for 3-D imaging of the upper and the lower extremities and pelvis of adult and pediatric patients weighing from 40 to 450 lbs. The device is to be operated in a professional healthcare environment by qualified health care professionals only.
Device Description
The HiRise is a Cone Beam Computed Tomography Imaging Device that acquires 360-degree rotational projection sequences which are reconstructed into 3D volumetric images of the examined anatomical region. The device uses a gantry assembly, which is comprised of an X- ray source, image detector, and a motorized gantry. The gantry facilitates the acquisition of a full X-ray projection sequence by the acquisition software. For non-weight bearing scans of the lower extremity, a patient positioner accessory allows the patient to sit into a position where he/she can comfortably place his/her anatomy into the imaging bore. The gantry assembly is mounted on vertical actuators and can travel vertically to capture weight-bearing anatomy at various heights ranging from the feet to the pelvis regions. The HiRise provides total vertical travel of 37 inches to accommodate patients of various sizes. Images produced by the HiRise can be sent electronically to a DICOM complaint image viewing software.
More Information

Not Found

No
The summary describes a standard Cone Beam Computed Tomography (CBCT) device using a traditional filtered back projection reconstruction algorithm. There is no mention of AI, ML, or related concepts in the device description, intended use, or performance studies.

No.
The device is described as an imaging device for diagnosis, not for treating any condition.

Yes

Explanation: The device is intended for 3-D imaging of anatomical regions, and the images it produces are described as being of "adequate diagnostic quality" and "clinically acceptable," indicating their use in diagnosis by qualified healthcare professionals.

No

The device description explicitly states it is a "Cone Beam Computed Tomography Imaging Device" that includes hardware components such as an X-ray source, image detector, motorized gantry, and vertical actuators.

Based on the provided information, the HiRise device is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • IVD devices are used to examine specimens derived from the human body. The HiRise device is an imaging system that directly images the human body (upper and lower extremities and pelvis) using X-rays. It does not analyze biological samples like blood, urine, or tissue.
  • The intended use clearly describes imaging of anatomical regions. The purpose is to create 3D images of the patient's body for diagnostic purposes, not to perform tests on samples taken from the body.
  • The device description details an imaging system. It describes an X-ray source, detector, and gantry for acquiring images, which is characteristic of an in-vivo imaging device, not an in-vitro diagnostic device.

Therefore, the HiRise device falls under the category of medical imaging devices rather than in vitro diagnostic devices.

N/A

Intended Use / Indications for Use

The HiRise is intended to be used for 3-D imaging of the upper and the lower extremities and pelvis of adult and pediatric patients weighing from 40 to 450 lbs.
The device is to be operated in a professional healthcare environment by qualified health care professionals only.

Product codes (comma separated list FDA assigned to the subject device)

JAK

Device Description

The HiRise is a Computed Tomography X-ray System. The HiRise is a Cone Beam Computed Tomography Imaging Device that acquires 360-degree rotational projection sequences which are reconstructed into 3D volumetric images of the examined anatomical region. The device uses a gantry assembly, which is comprised of an X- ray source, image detector, and a motorized gantry. The gantry facilitates the acquisition of a full X-ray projection sequence by the acquisition software. For non-weight bearing scans of the lower extremity, a patient positioner accessory allows the patient to sit into a position where he/she can comfortably place his/her anatomy into the imaging bore.

The gantry assembly is mounted on vertical actuators and can travel vertically to capture weight-bearing anatomy at various heights ranging from the feet to the pelvis regions. The HiRise provides total vertical travel of 37 inches to accommodate patients of various sizes. Images produced by the HiRise can be sent electronically to a DICOM complaint image viewing software.

Key Device Components:
Embedded Controller
Flat Panel Detector
Gantry Assembly
X-Ray Source
X-Ray Power Supply
Patient Platform and Positioners
Patient non-weight bearing chair
Operator Control Box
External Server

Mentions image processing

Acquires 360-degree rotational projection sequences which are reconstructed into 3D volumetric images.
Images produced by the HiRise can be sent electronically to a DICOM complaint image viewing software.

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Computed Tomography X-ray

Anatomical Site

Upper and lower extremities, including the Elbow, Hand, Wrist, Pelvis, Hip, Femur, Knee, shin (lower leg or tib/fib), Ankle, and Foot.

Indicated Patient Age Range

Adult and pediatric patients

Intended User / Care Setting

Professional healthcare environment by qualified health care professionals only.
The HiRise is to be used in medical facilities including hospitals, and orthopedic clinics. Revolution CT is to be operated by a trained technologist.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Study Type: Verification/Validation
Sample Size: Not specified for image review, but testing included internal verification, external third-party testing, and validation by a fellowship-trained MSK radiologist.
Key Results:

  • All applicable test plans passed with no anomalies increasing risk.
  • Completed 3rd party testing with no outstanding failures.
  • Images of new imaging protocols were validated by a fellowship-trained MSK radiologist to be clinically acceptable and of clinical adequate diagnostic quality.
  • Comparison of characteristics and performance data with the predicate device demonstrated substantial equivalence.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K203187

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

July 12, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

CurveBeam LLC % Ryan Conlon Director of Quality and Regulatory Affairs 2800 Bronze Drive Suite #110 Suite 110 HATFIELD, PA 19440

Re: K241713

Trade/Device Name: HiRise Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: June 11, 2024 Received: June 14, 2024

Dear Ryan Conlon:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

1

2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

2

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Submission Number (if known)K241713
Device NameHiRise

Indications for Use (Describe)
The HiRise is intended to be used for 3-D imaging of the upper and the lower extremities and pelvis
of adult and pediatric patients weighing from 40 to 450 lbs.
The device is to be operated in a professional healthcare environment by qualified health care
professionals only.
Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Number: K241713 510(k) Summary

Contents

Table 5-1 Device Information2
Predicate Device:2
High-level change description:2
Device Description:3
Device Characteristics and Performance3
Key Device Components3
Environment of Use3
Patient Contacting Materials4
Changes that resulted in a new 510(k)4
Substantial equivalence comparison table:5
Design Control Activities Summary:11
Substantial Equivalence Discussion:12
Safety and Effectiveness Information:12
Conformity12
FDA Guidance13
Conclusion:13

4

Table 5-1 Device Information

Manufacturer Name:Curvebeam LLC
Manufacturer Address:2800 Bronze Drive, Suite 110, Hatfield, PA
Registration Number/FEI number:3009543051
Contact Person:Ryan Conlon
Submission DateJuly 11, 2024
Device Name:HiRise
Classification:Class II
Product code:JAK
Regulation Number:21 CFR 892.1750
Regulation Name:Computed Tomography X-Ray System
Device Description:The HiRise is a Computed Tomography X-ray
Intended Use:The HiRise is intended to be used for 3-D

Predicate Device:

| Company | Device name | Product
Code | 510(k) | Regulation Number | Device
Classification |
|-------------------|-------------|-----------------|---------|-------------------|--------------------------|
| CurveBeam,
LLC | HiRise | JAK | K203187 | 892.1750 | Class II |

High-level change description:

Device Name:HiRise
Indications for Use:The HiRise is intended to be used for 3-D imaging of the upper and the lower
extremities and pelvis of adult and pediatric patients weighing from 40 to 450 lbs.
The device is to be operated in a professional healthcare environment by qualified
health care professionals only.
Device class:Class II (21 CFR 892.1750)
Product code:JAK
510k Number/Type:K203187 / Traditional 510k
Basic UDI-DI8631520003CBCTDevicesCL
Date of change:TBA
Description of Change:Changing the maximum mA output from 6.5mA to 20mA
Reason for Change:The HiRise is indicated for imaging of upper and lower extremities. Current imaging
protocols for the HiRise have been deemed clinically acceptable, but the image

CurveBeam, LLC – HiRise 510(k) Application

5

| characteristics on certain patients, based on anatomy, could be improved for

enhanced contrast and lower noise.
When it comes to imaging the Hip, there is naturally a significantly higher volume of
adipose tissue for the X-Rays to penetrate. Therefore, a higher energy imaging
protocol is proposed to enhance the image characteristics of hip scans (especially for
larger patients, based on anatomy).
The HiRise is a versatile device, as prior to each scan the imaging protocol is selected
based on anatomy being scanned, purpose of the scan and size of patient. Due to
such complexities and variabilities the HiRise is also indicated to be operated by
trained professionals only.

Device Description:

Device Characteristics and Performance

The HiRise is a Cone Beam Computed Tomography Imaging Device that acquires 360-degree rotational projection sequences which are reconstructed into 3D volumetric images of the examined anatomical region. The device uses a gantry assembly, which is comprised of an X- ray source, image detector, and a motorized gantry. The gantry facilitates the acquisition of a full X-ray projection sequence by the acquisition software. For non-weight bearing scans of the lower extremity, a patient positioner accessory allows the patient to sit into a position where he/she can comfortably place his/her anatomy into the imaging bore.

The gantry assembly is mounted on vertical actuators and can travel vertically to capture weight-bearing anatomy at various heights ranging from the feet to the pelvis regions. The HiRise provides total vertical travel of 37 inches to accommodate patients of various sizes. Images produced by the HiRise can be sent electronically to a DICOM complaint image viewing software.

Key Device Components

Embedded Controller Flat Panel Detector Gantry Assembly X-Ray Source X-Ray Power Supply Patient Platform and Positioners Patient non-weight bearing chair Operator Control Box External Server

Environment of Use

The HiRise is to be used in medical facilities including hospitals, and orthopedic clinics. The HiRise should be installed and operated according to state and federal regulations regarding radiationemitting products and the device room layout should be approved by a certified medical physicist prior to CurveBeam, LLC – HiRise 510(k) Application Page 3

6

installation.

Patient Contacting Materials

The materials that could contact the patient are listed below.

  • (1) Polyester patient platform lining
  • (2) Vinyl Transporter cushion
  • (3) Carbon fiber positioners
  • (5) Powder coated Aluminum handlebars

Changes that resulted in a new 510(k).

The proposed higher energy protocols (performance specifications) and the underlying component changes required to support the proposed protocols were evaluated. In summary, the conclusion is that this results in a performance change and there is an increased risk when it comes to patient radiation dose.

The most significant changes to the device include a redesign of the x-ray generator electronics. This includes the x-ray source (tube housing) and the x-ray power supply (high voltage power supply) along with the supporting cabling and safety electronics (isolation toroid and circuit breaker). All the above components exist in the current design but have been modified to support and enable higher energy imaging protocols to enhance the image characteristics of hip scans (especially for larger patients, based on anatomy).

The HiRise is a versatile device, as prior to each scan the imaging protocol is selected based on anatomy being scanned, purpose of the scan and size of patient. Due to such complexities and variabilities both the current and higher energy protocol versions of the HiRise are indicated to be operated by trained professionals only.

7

Substantial equivalence comparison table:

SUBSTANTIAL EQUIVALENCE TABLE
Feature/
ComponentHiRise V1HiRise V2Variance/Rationale
510(k) numberK203187
Product codeJAKNo changes
Regulation
number21 CFR 892.1750No changes
Indications for
useThe HiRise is intended to be used
for 3-D imaging of the upper and
the lower extremities and pelvis
of adult and pediatric patients
weighing from 40 to 450 lbs.
The device is to be operated in a
professional healthcare
environment by qualified health
care professionals only.No changes
Principal of
operationComputed Tomography X-ray
SystemNo changes
Scan axisHorizontal and verticalNo changes
Mechanical
LayoutA horizontal doughnut, with the
x-ray source and flat panel
detector mounted at each end of
a rotating gantry. The gantry
assembly lifts for scanning of the
lower extremities. Rotation of the
gantry to vertical imaging mode
allows for the positioning and
imaging of upper extremities. A
non-weight bearing chair allows
for non-weight bearing imaging of
the foot, ankle, and knee.No changes
ControllerFirmware Exposure ControllerNo changes
TubeStationary anode, glass envelope
x-ray tube.
Max KV: 130kV
Focal Spot: 0.5mm nominal
Target Angle: 15° degreesRotating Anode, glass
envelope x-ray tube.
Max KV: 130kV
Focal Spot: 0.3mm &
0.6mm nominal
Target Angle: 16°
degreesThe overall changes utilize a more robust
tube. Change of target angle and does
not impact the image sequence output
quality. Both the focal spot size and
target angle have been assessed in the
data sets sent to board-certified
radiologist and found to be of diagnostic
quality.
Tube HousingSame except for shape of
apertureEnlarged to account
for the rotating
anode tubeThe updated Tube housing was designed
with the same principles as the previous
version but is designed for the updated
Imaging requirements.
While the tube housing has been made
larger, the collimation ensures that the
output is kept as low as possible and only
exposes the active imaging area of the
panel.
All electrical changes have been verified
by third party to meet IEC 60601-1
testing, including all particular and
collateral standards (specifically 60601-1-
2, 60601-1-3, and 60601-2-44).
High Voltage
Power SupplyHigh frequency generatorHigh frequency
generator updated to
allow for the
increased output
requirementsThe updated High Voltage Power supply
utilizes the same design principles as the
previous version but is designed for the
updated Imaging requirements.
All electrical changes have been verified
by third party IEC 60601-1 testing,
including all particular and collateral
standards (specifically 60601-1-2, 60601-
1-3, and 60601-2-44).
Tube voltage100-130 kVP for CT scans100-120 kVP for CT
scansPrevious imaging protocols were limited
by the system requirements.
Bench testing determined optimal X-Ray
tube voltage (imaging characteristics) for
each anatomy and patient size.
Imaging characteristics have been
optimized for clinical diagnostic quality.
(especially for larger patients, based on
anatomy).
Labeling has been updated in alignment
with the new tube voltage range.
Tube current5.5 or 6.5 mA12 or 20 mAPrevious imaging protocols were limited
by the system requirements.
Bench testing determined optimal tube
current (imaging characteristics) for each
anatomy and patient size.
Imaging characteristics have been
optimized for clinical diagnostic quality.
Scan time26 sec for CT26 to 60 sec(especially for larger patients, based on anatomy).
Labeling has been updated in alignment with the new tube current range.
Bench testing determined optimal imaging characteristics for each anatomy and patient size.
Based on device design, maintaining an appropriate duty cycle is critical.
Increases in mAs require longer scan times to maintain device duty cycle.
Labeling has been updated in alignment with the new scan times.
Max exposure time15 sec26 secBench testing determined optimal imaging characteristics for each anatomy and patient size. Higher energy (exposure time) imaging protocol is proposed to enhance the image characteristics of hip scans (especially for larger patients, based on anatomy).
Labeling has been updated in alignment with the new max exposure times.
Image detectorAmorphous Silicon flat panelNo changes
Reconstruction AlgorithmFiltered back projection with non-linear filtering for 3D Cone-Beam CT reconstructionNo changes
Gray scale16 bitNo changes
3D Imaging VolumeLarge FOV: 7.7" (19.5cm) height x 15.79" (40.1cm) diameter
Medium FOV: 7.7" (19.5cm) height x 9.9" (25.2cm) diameterLarge FOV: 7.67" (19.5cm) height x 16.53" (42.0 cm) diameter*
Medium FOV: 7.67" (19.5cm) height x 9.84" (25.0 cm) diameter*
*Reconstructed Imaged VolumeSlight changes to the imaging volume are based on the new Tube housing design (see Tube housing comparison).
The imaging volumes were reviewed by a board-certified radiologist and found to be of diagnostic quality.
Labeling has been updated in alignment with the new Imaging volumes.
Typical resolutionLFOV: 0.3mm, MFOV: 0.25mmTypical resolution defined by anatomy imaged
0.25 mm, 0.3 mm, 0.5 mmboard-certified radiologist and found to be of diagnostic quality.
Labeling has been updated in alignment with the changes in protocol resolution. Minor specific protocol changes from the predicate device and are covered under the intended use (lower extremity) of both the predicate and updated HiRise. Removal of humerus, forearm (radius/ulna) particular protocols were deemed not critical as imaging of these partial anatomy protocols has not been utilized in real-world use. The specific humerus and forearm protocols configurations have been removed in the updated HiRise. Labeling has been updated in alignment with body parts imaged.
Body part scannedUpper and lower extremities, including the Humerus, Elbow, Forearm (radius/ulna), Hand, Wrist, Hip, Femur, Knee, Shin (lower leg or tib/fib) and Ankle, and Foot.Upper and lower extremities, including the Elbow, Hand, Wrist, Pelvis, Hip, Femur, Knee, shin (lower leg or tib/fib), Ankle, and Foot.
Size, inches h x d x w57"x58"x73" (145cm x 147cm x 185cm)No changes
Weight, lbsScanner 850 lbs (385.554 kg), Transporter 250 lb (113.398 kg)Weight changed to 900lbsIncreased weight of tube housing plus Added counterweight. Both Devices were designed to be as small as possible and to be installed in a professional medical environment with minimal modifications. Labeling has been updated in alignment with the new scanner weight.
Power Requirements920VA4800 VAThe power consumption has increased from the predicate in alignment with the updated imaging characteristics. All electrical changes have been verified by third party 60601-1 testing, including all particular and collateral standards (specifically 60601-1-2, 60601-1-3, and 60601-2-44). Labing has been updated in alignment with the new power requirements.
Tissue Density Range0 to 2000 HU's (Hounsfield Units)No changes
Scatter Radiation Range0.0003 to 4.83mR0.2 to 12.9mRWhen comparing the new scatter measurement numbers to the predicate,
expected as higher energy imaging
protocols are proposed to enhance the
image characteristics of hip scans
(especially for larger patients, based on
anatomy).
The scatter radiation is in alignment with
the increased imaging protocols and has
not increased outside of the expected
amount.
Labeling has been updated in alignment
with the new scatter numbers.
Patient Support
StructureFlat plastic platform and
handlebars for weight-bearing
imaging sequences, positioner
plate for knee, transporter
accessory for non-weight-bearing
imaging sequencesNo changes
Detector
PositionSource to Imager Distance Fixed
distance from detector to x-ray
beam center, parallel to axis of
rotation and orthogonal to x-ray
beam.No changes
X-Ray Beam
PositionBeam size matches the 40cm x
30cm flat panel (with a narrow
unexposed margin), there are two
vertical beam offsets from panel
center: beam offset below the
panel center for foot scans, and
beam offset above the panel
center for other extremity scansBeam size matches
the 40cm x 30cm flat
panel (with a narrow
unexposed margin),
there are two
vertical beam offsets
from panel center:
Beam offset below
the panel center is
utilized for foot scans
or any protocols that
are continuous and
combine with the
foot. Beam offsets
above the panel
center for other
extremity scansThe slight change in the definition of
beam position was included in the
datasets sent to the board-certified
radiologist and found to be of diagnostic
quality.
Software
Capture ToolVirtualized Windows environment
based applicationNo changesThere is new version of the software to
include the updated imaging protocols.
The application is still a virtualized
windows environment-based application.
DisplayComputer with mouse and
keyboardNo changes
Projection
GeometryBeam collimated to a square
shape, Source to Imager Distance:Beam collimated to a
square shape, Source
to Imager Distance:Slight changes to the projection
geometry are based on the new Tube
767 mm, Source to Axis of
rotation Distance: 503 mm730.55 mm, Source
to Axis of rotation
Distance: 466.95 mmhousing design (see Tube housing
comparison).
Performance testing demonstrated that
new geometry does not harm image
quality.
The imaging volumes were reviewed by a
board-certified radiologist and found to
be of diagnostic quality.
Patient
Positioning
GuideFoot: Circular markings on
platform
Knee/Hips: Accumeasure controls
height of the Gantry
Upper Extremity: Marks on upper
extremity platform insertNo changes
Patient
Contacting
MaterialPowder coated Aluminum
(Handlebar and Device cover),
Polyester Vinyl, Carbon fibersPowder coated
Aluminum
(Handlebar and
Device cover),
Polyester Vinyl,
Carbon fibers,
Nylon (fastener for
hand platform &
Shielding curtain)
and PolycarbonateMinor material changes, biocompatibility
assessed in alignment with 10993 and do
not impact the safety of the device.
The instructions for use regarding contact
with the machine on the predicate and
new version have not changed. The User
Manual on both devices recommends
avoiding direct contact with the scanner
using readily available materials.
Reconstruction
MethodFiltered back projection with
non-linear filtering for 3D
Cone-Beam CT reconstruction
Filtered back projection with nonlinear
filtering for 3D Cone-BeamNo changes

8

9

10

CurveBeam, LLC – HiRise 510(k) Application

11

CurveBeam, LLC – HiRise 510(k) Application

12

13

Design Control Activities Summary:
Device ChangeRisksVerification/Validation methodsAcceptance CriteriaSummary of Results
Change of X-Ray generating
circuitry and supporting
electronics
• Tube
• Tube housing.
◦ Projection
geometry
◦ 3D imaging
volume
• HVPS
• Tube current
• Increased power
requirements
• Additional cooling
features
• Additional bow tie
filter mechanismIncreased Dose
to the patientInternal Verification in
alignment with all internal
testing of previous HiRise.

External third-party testing
by ASCA testing facility to
60601-1 and all applicable
collateral and particular
standards.

*Note: Bowtie filter not
used in this update.
Future use will be enabled
under the same design
controls employed by the
current QMS. | Passes all internal
verification with no
anomalies causing
increased risk.

Passing all third
party testing to all
applicable
requirements. | All applicable test
plans passed with
no anomalies
increasing risk.

Completed 3rd party
testing with no
outstanding
failures. |
| Increased energy output
protocols
• Scan time
• Max Exposure time
• Increased mA
capabilities | | Validation by fellowship-
trained MSK radiologist. | Updated labeling

A letter stating
Images of new
imaging protocols
are clinically
acceptable and
adequate
diagnostic quality. | Labeling updated.

Letter from
fellowship-trained
MSK radiologist
validating images to
be clinically
acceptable and of
clinical adequate
diagnostic quality. |
| Increased weight of the
system | Negligible
increase
(approx. 50
pounds) | No verification required | Updated labeling | Labeling updated. |

Design Control Activities Summary:

14

Substantial Equivalence Discussion:

The overall technology, key components and intended use of the HiRise Computed Tomography x-ray system and the predicate device are substantially similar, with certain inconsequential differences as described.

Both the predicate HiRise and updated HiRise devices image extremities of similar bone density ranges and utilize a similar gantry to capture the desired anatomy. Images of the difference between the two devices have been reviewed by a radiologist and have been found to be of adequate diagnostic quality.

The combination of performance testing internally, safety and functional testing by a certified thirdparty testing body, and clinical review of images by a radiologist indicate that HiRise is safe and effective when used as labeled.

Safety and Effectiveness Information:

The HiRise Computed Tomography X-ray system is a Class II medical device.

The HiRise Computed Tomography X-ray system complicable FDA and international standards pertaining to electrical, mechanical, software, EMC, and radiation safety of medical devices.

Conformity

The HiRise device has been tested to the following standards to ensure safety, effectiveness, and compliance with industry norms:

IEC 60601-1:2005, Edition 3.2, 08/2020 IEC 60601-1-3:2008, Edition 2.2, 01/2021 IEC 60601-1-6, Edition 3.2, 07/2022 IEC 60601-2-44, Edition 3.2, 03/2016 (12-302) IEC 62366-1, Edition 1.1, 06/2020 IEC 60825-1, Edition 3, 05/2014 IEC 62304:2006 Ed.1 +A1, 06/2015 IEC 60601-1-2, Edition 4, 02/2014 IEC 61223-3-5, Edition 2.0, 09/2019 (12-328) NEMA PS 3.1-3.20, 2023e (12-352) NEMA PS 3.1-3.20, 2023d (12-349) 61223-2-6 Second Edition 2006-11 (12-266)

15

FDA Guidance

The following FDA Guidance Documents were referenced to the extent they were applicable in the preparation of this 510(k) Submission

  • . Provision for Alternate Measure of the Computed Tomography Dose Index (CTDI) to Assure Compliance with the Dose Information Requirements of the Federal Performance Standard for Computed Tomography
  • . Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices
  • Guidance for the Content of Premarket Submissions for Device Software Functions
  • Applying Human Factors and Usability Engineering to Medical Devices
  • Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered ● Medical Devices
  • Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions
  • Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices
  • Pediatric Information for X-ray Device Premarket Notifications Imaging

Conclusion:

CurveBeam, LLC has demonstrated through its comparison of characteristics with the predicate device and comparison of performance data with the predicate device that the HiRise Computed Tomography X-ray System is substantially equivalent to the predicate device.