K Number
K241649
Device Name
DUO1 and DUO2
Manufacturer
Date Cleared
2024-07-05

(28 days)

Product Code
Regulation Number
872.1800
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The DUO Sensors are USB driven intraoral sensors which are intended to acquire intraoral radiographic images. The DUO Sensors shall be operated by trained healthcare professionals, who are educated and competent to perform the acquisition of intraoral radiographs. The DUO Sensors can be used either in combination with special positioning devices to facilitate positioning and alignment with the X-ray beam or they may also be positioned by hand with the assistance of the patient. The DUO sensors are intended for any dental practice that uses X-ray equipment for intraoral diagnostic purposes. DUO sensors can be used by trained dental professionals for patients receiving intraoral X-ray examinations and procedures for capturing digital X-ray images. Captured digital X-ray images can be used for examinations and diagnostic purposes with the help of optional image analysis software. The optional image analysis software is not part of this submission. DUO sensors can be used with dental positioning devices and holders to assist with aligning an X-ray source beam with the sensor and anatomy.
Device Description
The DUO sensors are USB-driven digital X-ray sensors designed for health care professionals who are already acquainted with the standard procedures for acquiring dental intraoral radiographs. Digital X-ray imaging is an aide for diagnosis and should always be confirmed by the doctor using appropriate additional diagnostic aides, professional judgment, and experience. The DUO Sensors are indirect converting X-ray detectors. A scintillating material converts the incident Xrays into visible light, this light is coupled optically to a CMOS technology light detection imager, and then converted to digital data. The design of the sensor assembly supports the automatic detection of incident X-rays to generate digital images for intraoral applications, once armed via a software command. The digital image created is immediately visible on the screen of a personal computer connected to the DUO sensor through the standard USB port. For DUO sensors to be used in a dental practice, an optional image analysis software will be necessary. Image analysis software is not part of the submission. DUO captured X-ray images are suitable for recognition of normal anatomical structures, dental pathologies, and abnormal conditions. The DUO sensors support USB2.0 connectivity to computers using a dedicated electronic assembly and a sensor software driver. Functions of the DUO sensors are controlled by software drivers and utilities support sensor activation and settings. The DUO sensors are manufactured with the same device firmware as the predicate device, Brasseler GEM.
More Information

Not Found

No
The description focuses on the sensor technology and image acquisition, explicitly stating that optional image analysis software (which might contain AI/ML) is not part of this submission.

No
The device is described as an intraoral sensor used to acquire radiographic images for diagnostic purposes, not for treating a condition or disease.

Yes

Explanation: The "Intended Use / Indications for Use" section explicitly states, "DUO sensors can be used by trained dental professionals for patients receiving intraoral X-ray examinations and procedures for capturing digital X-ray images. Captured digital X-ray images can be used for examinations and diagnostic purposes with the help of optional image analysis software." This indicates that the device produces images that are intended to be used for diagnostic purposes.

No

The device is a physical sensor that converts X-rays into digital data, which is then processed by software. It is not solely software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health.
  • DUO Sensors Function: The DUO Sensors acquire radiographic images using X-rays. They are used to visualize internal structures (teeth, bones) within the mouth. This is an in vivo (within the living body) imaging technique, not an in vitro (outside the living body) diagnostic test.
  • Intended Use: The intended use clearly states the device is for acquiring intraoral radiographic images for diagnostic purposes, which is a form of medical imaging, not an in vitro test.
  • Device Description: The description details how the sensor converts X-rays into digital images, which is consistent with an imaging device, not an IVD.

Therefore, the DUO Sensors fall under the category of medical imaging devices, not In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

The DUO Sensors are USB driven intraoral sensors which are intended to acquire intraoral radiographic images. The DUO Sensors shall be operated by trained healthcare professionals, who are educated and competent to perform the acquisition of intraoral radiographs.

The DUO Sensors can be used either in combination with special positioning devices to facilitate positioning and alignment with the X-ray beam or they may also be positioned by hand with the assistance of the patient.

Product codes (comma separated list FDA assigned to the subject device)

MUH

Device Description

The DUO sensors are USB-driven digital X-ray sensors designed for health care professionals who are already acquainted with the standard procedures for acquiring dental intraoral radiographs. Digital X-ray imaging is an aide for diagnosis and should always be confirmed by the doctor using appropriate additional diagnostic aides, professional judgment, and experience.

The DUO Sensors are indirect converting X-ray detectors. A scintillating material converts the incident Xrays into visible light, this light is coupled optically to a CMOS technology light detection imager, and then converted to digital data.

The design of the sensor assembly supports the automatic detection of incident X-rays to generate digital images for intraoral applications, once armed via a software command. The digital image created is immediately visible on the screen of a personal computer connected to the DUO sensor through the standard USB port. For DUO sensors to be used in a dental practice, an optional image analysis software will be necessary. Image analysis software is not part of the submission. DUO captured X-ray images are suitable for recognition of normal anatomical structures, dental pathologies, and abnormal conditions.

The DUO sensors support USB2.0 connectivity to computers using a dedicated electronic assembly and a sensor software driver. Functions of the DUO sensors are controlled by software drivers and utilities support sensor activation and settings.

The DUO sensors are manufactured with the same device firmware as the predicate device, Brasseler GEM.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-ray

Anatomical Site

Intraoral / dental anatomy

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Trained healthcare professionals / dental practice

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance Testing-Bench: Intertek performed extensive bench testing on DUO Sensor in accordance with the following reference standards: ANSI AMI ES 60601-1 Edition 3.2 2020-08; IEC 62304 Edition 1.1 2015-06; IEC 60601-1-2 Edition 4.0 2014-02. The results of these tests indicate that the DUO Sensor is substantially equivalent to the predicate device.

Performance Testing-Clinical: Zinnovi performed a clinical performance testing to determine that DUO performed similar or better than the predicate device, Brasseler GEM. Zinnovi has provided clinical images captured with both the devices in a performance testing report which were evaluated by US dentists. These clinical images provide evidence in addition to the bench performance data that the complete system works as intended.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K221955

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.

0

July 5, 2024

Zinnovi Inc. % Abdel Halim President & MRP Global Quality and Regulatory Services 10 Scenic Way MONROE, NJ 08831

Re: K241649

Trade/Device Name: DUO1 and DUO2 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: June 7, 2024 Received: June 7, 2024

Dear Abdel Halim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

1

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

2

Indications for Use

510(k) Number (if known) K241649

Device Name DU01 and DUO2

Indications for Use (Describe)

The DUO Sensors are USB driven intraoral sensors which are intended to acquire intraoral radiographic images. The DUO Sensors shall be operated by trained healthcare professionals, who are educated and competent to perform the acquisition of intraoral radiographs.

The DUO Sensors can be used either in combination with special positioning devices to facilitate positioning and alignment with the X-ray beam or they may also be positioned by hand with the assistance of the patient.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

K241649

510(k) Summary

Date Prepared: June 14th, 2024

Manufacturer and 510(k) Owner:

Name : Zinnovi, Inc. Address : 135 W. Central Rd, Suite 204, Schaumburg, IL 60195 Telephone: +1 (224) 801-5902 Official Contact: Swaroop Patel (CEO and Co-founder) Email : spatel@zinnovi.com

Representative/Consultant:

Name : Erica Livingston, RAC Company : Qualomics LLC Address : 5938 Priestly Dr, Suite 103, Carlsbad, CA 92008 Telephone: +1 (858) 247-6804 Email: erica@qualomics.com

Identification of New Device

Owner/Operator:Zinnovi Inc.
Premarket Notification Number:K241649
Trade Name:DUO1 and DUO2
Common Name:Dental digital X-ray sensor
Classification Name:Extraoral Source X-ray System
Product Code:MUH
Submission Type:510(k)
Classification:Class II
Panel/Medical Specialty:Radiology/Dental
Regulation Number:21 CFR 872.1800
Manufacturer:Zinnovi Inc.

Identification of Predicate Device

Owner/Operator:RealCloud Imaging Inc. doing business as RealCloud Imaging
510(k) Premarket Notification Number: K221955
Trade Name:Brasseler GEM

4

Common Name: Dental digital X-ray sensor Classification Name: Extraoral Source X-ray system Product Code: MUH Classification: Class II Panel/Medical Specialty: Radiology/Dental Regulation Number: 21 CFR 872.1800 Manufacturer: RealCloud Imaging Inc. doing business as RealCloud Imaging

Device Description

The DUO sensors are USB-driven digital X-ray sensors designed for health care professionals who are already acquainted with the standard procedures for acquiring dental intraoral radiographs. Digital X-ray imaging is an aide for diagnosis and should always be confirmed by the doctor using appropriate additional diagnostic aides, professional judgment, and experience.

The DUO Sensors are indirect converting X-ray detectors. A scintillating material converts the incident Xrays into visible light, this light is coupled optically to a CMOS technology light detection imager, and then converted to digital data.

The design of the sensor assembly supports the automatic detection of incident X-rays to generate digital images for intraoral applications, once armed via a software command. The digital image created is immediately visible on the screen of a personal computer connected to the DUO sensor through the standard USB port. For DUO sensors to be used in a dental practice, an optional image analysis software will be necessary. Image analysis software is not part of the submission. DUO captured X-ray images are suitable for recognition of normal anatomical structures, dental pathologies, and abnormal conditions.

The DUO sensors support USB2.0 connectivity to computers using a dedicated electronic assembly and a sensor software driver. Functions of the DUO sensors are controlled by software drivers and utilities support sensor activation and settings.

The DUO sensors are manufactured with the same device firmware as the predicate device, Brasseler GEM.

Indications for Use

The DUO Sensors are USB driven intraoral sensors which are intended to acquire intraoral radiographic images. The DUO Sensors shall be operated by trained healthcare professionals, who are educated and competent to perform the acquisition of intraoral radiographs.

The DUO Sensors can be used either in combination with special positioning devices to facilitate positioning and alignment with the X-ray beam or they may also be positioned by hand with the assistance of the patient.

5

Intended Use

The DUO sensors are intended for any dental practice that uses X-ray equipment for intraoral diagnostic purposes. DUO sensors can be used by trained dental professionals for patients receiving intraoral X-ray examinations and procedures for capturing digital X-ray images. Captured digital X-ray images can be used for examinations and diagnostic purposes with the help of optional image analysis software. The optional image analysis software is not part of this submission. DUO sensors can be used with dental positioning devices and holders to assist with aligning an X-ray source beam with the sensor and anatomy.

Comparison Table

The following comparison table compares Zinnovi DUO to the predicate device, Brasseler GEM, with respect to intended use, indications of use, environment of use, limitations of use, technical performance, and technological characteristics, and provides more detailed information regarding the basis for the determination of substantial equivalence.

| Descriptive

InformationZinnovi DUOBrasseler GEMVERDICT
510(k) NumberK241649K221955NA
ManufacturerZinnovi Inc.RealCloud Imaging Inc. doing
business as RealCloud
ImagingNA
Trade NameDUO1
DUO2Brasseler GEM10
Brasseler GEM15
Brasseler GEM20NA
Classification NameDigital X-ray SensorDigital X-ray SensorSame
ClassificationClass IIClass IISame
Classification Panel/
Medical SpecialtyRadiology/DentalRadiology/DentalSame
Regulation Number21 C.F.R. §872.180021 C.F.R. §872.1800Same
Product CodeMUHMUHSame
Number of Sensors23Different
Sensor Exterior Sizes36.36 mm x 24.53.mm36.36 mm x 24.53.mmSame
38.83 mm x 29.63 mmNA
41.80 mm x 30.48 mm41.80 mm x 30.48 mmSame
Sensor Imaging Sizes30.26 mm x 20.32 mm30.26 mm x 20.32 mmSame
33.15 mm x 26.25 mmNA
36.08 mm x 26.25 mm36.08 mm x 26.25 mmSame
All with four clipped cornersAll with four clipped cornersSame
Overall Imaging Areas615 mm2615 mm2Same
870 mm2NA
947.1mm2947.1mm2Same
Pixel Size19.5 μm19.5 μmSame
Image Resolution1539 x 1026 pixels
(1.70 M pixels)1539 x 1026 pixels
(1.70 M pixels)Same
1692 x 1324 pixels
(2.2 M Pixels)NA
1842 x 1324 pixels
(2.40 M pixels)1842 x 1324 pixels
(2.40 M pixels)Same
X-Ray Resolution20 visible lp/mm20+ visible lp/mmDifferent
Dynamic Range16,384:116,384:1Same
TechnologyCMOSCMOSSame
Scintillator TechnologyCesium lodideCesium lodideSame
Operating SystemMicrosoft Windows 10 and
Windows 11Microsoft Windows 7 and 10Different
Extraoral sourceX-ray systemX-ray systemSame
Interface to PCUSB 2.0, Type AUSB 2.0, Type ASame
Cable Length0.6m and 1.9m1.9 m and 2.9 mDifferent
Power Consumption0.8 Watts Max0.8 Watts MaxSame
Electrical RatingDC 5V, 350 mA maxDC 5V, 350 mA maxSame
SterilizationNot suitable for sterilizationNot suitable for sterilizationSame
HousingIPx8 Equivalent ISO 10993-1
BiocompatibleIPx8 Equivalent ISO 10993-1
BiocompatibleSame
Indications for useThe DUO Sensors are USB
driven digital intraoral sensors
which are intended to acquire
intraoral radiographic images.
The DUO Sensors shall be
operated by trained
healthcare professionals, who
are educated and competent
to perform the acquisition of
intraoral radiographs.
The DUO Sensors can be used
either in combination with
special positioning devices to
facilitate positioning and
alignment with the X-ray beam
or they may also be positioned
by hand with the assistance of
the patient.Brasseler GEM is a USB
driven digital intraoral X-ray
sensor which is intended to
acquire dental radiographic
images.
Brasseler GEM must be
operated by healthcare
professionals who have
been trained and are
competent using various
methods of acquiring
radiographic images of
dental anatomy. Brasseler
GEM can be used with
dental positioning devices
and holders to assist with
aligning an X-ray source
beam with the sensor and
anatomy. Brasseler GEM can
also be aligned by hand with
assistance of the patient.Same
Intended UseThe DUO sensors are intended
for any dental practice that
uses X-ray equipment for
intraoral diagnostic purposes.
DUO sensors can be used by
trained dental professionals
for patients receiving intraoral
X-ray examinations and
procedures for capturing
digital X-Ray images. Captured
digital X-ray images can be
used for examinations and
diagnostic purposes with theBrasseler GEM is intended
for any dental practice that
uses X-ray equipment for
intraoral diagnostic
purposes. Brasseler GEM
can be used by trained
dental professionals for
patients receiving intraoral
X-ray examinations and
produces digital images for
patients receiving intraoral
X-ray examinations for
diagnostic purposes. An
image analysis software isSame
help of optional image analysis
software.not part of this submission.
The optional image analysis
software is not part of thisWhen Brasseler GEM is to
be used in a dental practice,
submission. DUO sensors can
be used with dentalan optional software will be
necessary. Brasseler GEM is
positioning devices and
holders to assist with aligninga USB-driven digital intraoral
X-ray sensor which is
an X-ray source beam with the
sensor and anatomy.intended to acquire dental
radiographic images.
Brasseler GEM must be
operated by healthcare
professionals who have
been trained and are
competent using various
methods of acquiring
radiographic images of
dental anatomy. Brasseler
GEM can be used with
dental positioning devices
and holders to assist with
aligning an X-ray source
beam with the sensor and
anatomy.

Comparison Table

6

7

8

Comparison of MTF of Zinnovi DUO and MTF of Brasseler GEM

Zinnovi DUO uses a sensor component from contract manufacturer BAE Systems Imaging Solutions, a division of BAE Systems Inc., which is also responsible for the manufacture of Brasseler GEM which is the predicate device. The components used in Zinnovi DUO are the exact same components used in predicate device Brasseler GEM. The sensor MTF versus spatial frequency for Zinnovi DUO and the sensor MTF versus spatial frequency for the predicate device, Brasseler GEM, are identical.

Comparison of DQE for Zinnovi DUO and DQE of Brasseler GEM

On behalf of Zinnovi Inc., BAE Systems Imaging Solutions has compared the DQE for Zinnovi DUO with the DQE of Brasseler GEM and has determined that they are substantially equivalent to each other.

Meaningful Differences

Zinnovi DUO and Brasseler GEM are CMOS X-ray image sensors. Zinnovi has compared the MTF and the DQE of DUO sensor to the MTF and the DQE of Brasseler GEM and has determined they are substantially equivalent to each other. DUO is not only similar in performance to Brasseler GEM but is also safe and effective based on performance testing.

9

Number of sensors: Zinnovi offers the smallest and largest sizes equivalent to the Brasseler GEM and would therefore be effective on the same patient population.

Operating System: Zinnovi offers device drivers to work with the latest Microsoft operating system.

X-Ray Resolution: DUO and predicate Brasseler GEM both have the same theoretical maximum resolution of 25 lp/mm.

Cable Length: Cable length has no effect on the performance of the product.

Sterilization

DUO sensors are not sold as sterile. Zinnovi Inc. recommends that usage of DUO sensor requires a hygienic barrier that meets ISO 10993 requirements for biocompatibility.

Biocompatibility

Biocompatibility testing is not needed with a rationale that considers all relevant endpoints because all materials and manufacturing/processing are identical to the predicate device, Brasseler GEM. The biocompatibility of DUO Sensors are based on the biocompatibility of the predicate device, Brasseler GEM. The DUO Sensor and the predicate Brasseler GEM are intended to be used in the oral cavity by a trained clinician.

SABIC Resin for Sensor Housing of DUO

BAE Systems Imaging Solutions uses the same SABIC resin for the sensor housing of DUO sensor that BAE Systems Imaging Solutions uses for the sensor housing of the predicate device, Brasseler GEM.

Software Information

The software for the DUO sensor consists of a simple API that can be provided to developers of existing FDA cleared image capture/dental imaging software to facilitate DUO sensor.

Risk Analysis

Zinnovi Inc. used a specified procedure for evaluating the safety of a device by identifying potential hazards and estimating the associated risks as required in the Software Design and Development Procedure subsection of the Software Information section. Zinnovi Inc. performed an extensive risk analysis of DUO sensor in accordance with the following reference standards: ISO 14971, IEC 60601-1; and IEC 60601-1-2: Medical Electrical Equipment, Part 1-2 to generate a risk analysis report.

Comparison of Safety and Effectiveness of DUO to Safety and Effectiveness of Brasseler GEM

Using the above Comparison Table, DUO sensor can be compared to the predicate device, Brasseler GEM, with respect to intended use, indications of use, environment of use, limitations of technical performance and technological characteristics and provides more detailed information regarding the basis for the

10

determination of substantial equivalence. There are no differences between DUO and Brasseler GEM with respect to indications and intended use.

Performance Testing-Bench

On behalf of Zinnovi Inc, Intertek performed extensive bench testing on DUO Sensor in accordance with the following reference standards: ANSI AMI ES 60601-1 Edition 3.2 2020-08; IEC 62304 Edition 1.1 2015-06; IEC 60601-1-2 Edition 4.0 2014-02.

The results of these tests indicate that the DUO Sensor is substantially equivalent to the predicate device.

Performance Testing-Clinical

Zinnovi performed a clinical performance testing to determine that DUO performed similar or better than the predicate device, Brasseler GEM. Zinnovi has provided clinical images captured with both the devices in a performance testing report which were evaluated by US dentists. These clinical images provide evidence in addition to the bench performance data that the complete system works as intended.

Conclusion

On behalf of Zinnovi Inc., BAE Systems Imaging Solutions has compared DUO Sensor with the predicate device, Brasseler GEM, and has determined that they are substantially equivalent to each other in intended use, indications for use, safety and effectiveness, and technical characteristics.