K Number
K241355
Date Cleared
2024-10-09

(148 days)

Product Code
Regulation Number
878.4430
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SER Pen Carain MicroSystem is intended for use as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years and older and as a treatment to improve the appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years or older.

Device Description

The SER Pen Carain MicroSystem (MP1209SP) is a handheld device that creates channels as well as microinjuries into the skin, by virtue of a DC motor drive system that rapidly reciprocates an array of 12 needles and 36 needles that are no longer than 2.5mm. The device consists of a power source, a motor body with depth adjustment, and a disposable, single use cartridge containing an array of needles. The power source consists of a rechargeable lithium-ion battery that delivers no more than 5 volts DC and 1 amp of current to power the motor. A power cord connects the wall adaptor to the device via a USB connector and a standard 1/8" headphone plug on the device side. The motor body is comprised of anodized aluminum with a dial mechanism that controls the depth of penetration of the microneedles from 0.0 mm to a maximum of 2.5mm. The SER Pen Carain MicroSystem's disposable cartridge is designed in two configurations, a 12-needle array and a 36-needle array. The needle array is housed in a specially designed cartridge housing with a silicon spring that prevents liquids from entering the motor body via the inside lumen of the cartridge. The SER Pen Carain MicroSystem biolock sleeve is a one end elastic opening PE bag to protect the whole system from cross contamination from the outside during the operation procedure. The SER Pen Carain MicroSystem microneedles are composed of 304 18/8 surgical steel tested and certified as biocompatible under GLP testing conditions that conform to ISO 10993-5, 10993-10, and 10993-11. A metallurgical analysis under GLP testing conditions that conform to ASTM E1019-11(Method A)(C/S Analyzer) was also completed and is attached. Each lot of cartridges are individually packaged and then gamma ray sterilized by TUV Sud accredited facility under procedure code P1501314 with a minimum dose of 25kGy. Certifications on file with the Sponsor and available for review.

AI/ML Overview

This document describes the SER Pen Carain MicroSystem (MP1209SP), a microneedling device cleared via a 510(k) submission (K241355) by demonstrating substantial equivalence to predicate devices, rather than through a direct study proving specified acceptance criteria.

The submission specifically states: "The Sponsor intends to rely on the same performance data submitted in Collagen P.I.N. (K222199) and SkinStylus (K200044 and K231073) devices and not include any clinical data." This means that acceptance criteria and performance data are inferred from the predicate devices through a comparison of technological characteristics, rather than direct testing of the SER Pen Carain MicroSystem against pre-defined acceptance criteria with a dedicated study.

Therefore, the following information is based on the provided document and the nature of a 510(k) submission relying on predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) submission relying on predicate devices, explicit "acceptance criteria" and "reported device performance" in the context of a de novo clinical study are not directly presented for the SER Pen Carain MicroSystem. Instead, the "acceptance criteria" are implied by meeting the same performance and safety standards demonstrated by the predicate devices, and "device performance" is established by showing identical or highly similar technological characteristics and performance to those predicates.

Based on the document, the "performance" of the SER Pen Carain MicroSystem is derived from its substantial equivalence to the predicate devices across various technological characteristics.

Feature/TestAcceptance Criteria (Implied by Predicate Devices)SER Pen Carain MicroSystem Performance (Reported as "Identical" or similar to Predicates)
Indications for UseImprove appearance of facial acne scars (Fitzpatrick Skin Types I-III, 22+ years); Improve appearance of surgical/traumatic hypertrophic scars on abdomen (22+ years).Intended for use to improve appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, and to improve appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years and older. (Matches combined indications of predicates)
Mode of ActionMicroneedling (mechanically puncturing skin for aesthetic use)Identical (Microneedling)
Power Source5 Volt DC/1 amp rechargeable lithium-ion batterySame: 5Volt DC/ 1 amp rechargeable lithium-ion battery
Length/Width/shape/weightIdentical to predicate devicesIdentical to predicate device(s)
Range of Needle Length0.0-3.0mm (Collagen P.I.N.) or 0.0-2.5mm (SkinStylus SteriLock)0.0-2.5mm
Maximum Penetration3.0mm (Collagen P.I.N.) or 2.5mm (SkinStylus SteriLock)2.5mm
Needle Geometry36 solid needles (Collagen P.I.N.) or 12 solid needles (SkinStylus SteriLock)Identical, 12 or 36 solid needles
Speed7000RPM-9000RPM (Collagen P.I.N.) or 6200RPM-9000RPM (SkinStylus SteriLock)Identical, 6200RPM-9000RPM
Sterility & CleaningDisposable cartridge, gamma ray or ethylene oxide sterilized prior to packagingSame. Disposable cartridge Gamma Ray Sterilized prior to packaging
Cross Contamination FeatureCartridge design with sealing spring or other features to prevent cross-contaminationIdentical: Cartridge design with a sealing spring inside the cartridge housing, and a biolock sleeve to cover from the cartridge outshell and the whole handpiece to stop cross contamination.
BiocompatibilityConforms to ISO 10993-5, 10993-10, 10993-11 standardsMet (through reliance on predicate data and direct testing of microneedles per ISO 10993-5, -10, -11, and ASTM E1019-11 for metallurgical analysis, as listed in "Summary of Non-Clinical Test Reports" for the needles themselves).
Sterilization ValidationConforms to ISO 17665-1, ISO 11737-1, ISO 11737-2, USP 71Met (through reliance on predicate data and direct testing per ISO 17665-1, ISO 11737-1, ISO 11737-2, and USP 71).
Reprocessing ValidationConforms to ISO 11737-1, AAMI TIR-30, AAMI TIR-12Met (through reliance on predicate data and direct testing per ISO 11737-1, AAMI TIR-30, AAMI TIR-12).
Fluid Ingress ValidationDemonstrated to prevent fluid ingress into motor/housingMet (through reliance on predicate data and similarity of design).
Shelf-life TestingConforms to ASTM F88/F88M-15 and ASTM F1929-15 for seal strength and leak detectionMet (through reliance on predicate data and direct testing per ASTM F88/F88M-15, ASTM F1929-15).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: This information is not provided as the submission relies on existing data from predicate devices. There was no specific "test set" for the SER Pen Carain MicroSystem in the context of a new clinical study.
  • Data Provenance: The document explicitly states the "Sponsor intends to rely on the same performance data submitted in Collagen P.I.N. (K222199) and SkinStylus (K200044 and K231073) devices and not include any clinical data." Therefore, the data provenance would be from the studies conducted for these predicate devices, which are not detailed in this document. Information on country of origin or whether the data was retrospective or prospective is not available in this document.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • This information is not applicable as no new clinical study with a "test set" requiring expert ground truth was conducted for the SER Pen Carain MicroSystem. The data used for establishing substantial equivalence comes from pre-existing regulatory submissions for the predicate devices.

4. Adjudication Method for the Test Set

  • This information is not applicable as no new clinical study with a "test set" requiring adjudication was conducted for the SER Pen Carain MicroSystem.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done. The submission explicitly states "not include any clinical data," indicating a reliance on non-clinical data and comparison to predicate devices for substantial equivalence.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • No, this is a microneedling device, not an AI algorithm. "Standalone performance" in the context of an algorithm is not applicable.

7. The Type of Ground Truth Used

  • As no new clinical studies were conducted for the SER Pen Carain MicroSystem, the concept of "ground truth" in this context is not directly applicable. The basis for clearance is established through demonstrating similar technological characteristics and performance to predicate devices that were previously cleared by the FDA based on their own evidence (which may have included clinical data, expert consensus, etc., but that information is not present here). For the non-clinical tests (biocompatibility, sterilization, etc.), the "ground truth" is adherence to established international and national standards (ISO, ASTM, USP).

8. The Sample Size for the Training Set

  • This information is not applicable as there was no AI algorithm or "training set" for this device.

9. How the Ground Truth for the Training Set was Established

  • This information is not applicable as there was no AI algorithm or "training set" for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 9, 2024

Su-Ko Technologies LLC % Marc Sanchez, Esq. Regulatory Counsel Contract In-House Counsel and Consultants, LLC d/b/a FDA Atty 1717 Pennsylvania Ave NW Suite 1025 Washington, District of Columbia 20006

Re: K241355

Trade/Device Name: SER Pen Carain MicroSystem (MP1209SP) Regulation Number: 21 CFR 878.4430 Regulation Name: Microneedling Device For Aesthetic Use Regulatory Class: Class II Product Code: QAI Dated: September 11, 2024 Received: September 11, 2024

Dear Marc Sanchez, Esq.:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Wilmarie Flores -S

Wilmarie Flores, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241355

Device Name SER Pen Carain MicroSystem (MP1209SP)

Indications for Use (Describe)

The SER Pen Carain MicroSystem is intended for use as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, aged 22 years and older and as a treatment to improve the appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years or older.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary

The following information is provided as required by 21 CFR 807.92 for the Carain MicroSystem 510(k) premarket notification.

Sponsor:Su-Ko Technologies, LLCAttention: Yuan Yuan Korrodi700 Smith Street, #61070Houston, TX 77002
Manufacturer:GUANGZHOU CARAIN BEAUTY EQUIPMENT501, No. 3 of Xin Liu Mu Road, ZhongCun StreetGuangzhou Guangdong, CHINA 511495Establishment Registration: 3011568699
Contact:Marc C. Sanchez, Esq.Contract In-House Counsel and Consultants, LLC (d/b/a FDA Atty) 1717Pennsylvania Ave NW Suite 1025 Washington D.C. 20006Ph: 202.765.4491E-mail: msanchez@fdaatty.com
Date of Submission:October 8, 2024
Proprietary Name:SER Pen Carain MicroSystem (MP1209SP)
Common Name:Powered Microneedle Device
Regulation Number:21 CFR 882.4810
Regulatory Class:Class II

Product Code: QAI

Predicate Device: Collagen P.I.N. (Percutaneous Induction Microneedling) System (K222199).

Secondary Predicate Device: SkinStylus SteriLock® MicroSystem (K200044).

Device Description:

The SER Pen Carain MicroSystem (MP1209SP) is a handheld device that creates channels as well as microinjuries into the skin, by virtue of a DC motor drive system that rapidly reciprocates an

{5}------------------------------------------------

SuKo Technologies, LLC

array of 12 needles and 36 needles that are no longer than 2.5mm. The device consists of a power source, a motor body with depth adjustment, and a disposable, single use cartridge containing an array of needles.

The power source consists of a rechargeable lithium-ion battery that delivers no more than 5 volts DC and 1 amp of current to power the motor. A power cord connects the wall adaptor to the device via a USB connector and a standard 1/8" headphone plug on the device side.

The motor body is comprised of anodized aluminum with a dial mechanism that controls the depth of penetration of the microneedles from 0.0 mm to a maximum of 2.5mm.

The SER Pen Carain MicroSystem's disposable cartridge is designed in two configurations, a 12-needle array and a 36-needle array. The needle array is housed in a specially designed cartridge housing with a silicon spring that prevents liquids from entering the motor body via the inside lumen of the cartridge.

The SER Pen Carain MicroSystem biolock sleeve is a one end elastic opening PE bag to protect the whole system from cross contamination from the outside during the operation procedure.

The SER Pen Carain MicroSystem microneedles are composed of 304 18/8 surgical steel tested and certified as biocompatible under GLP testing conditions that conform to ISO 10993-5, 10993-10, and 10993-11. A metallurgical analysis under GLP testing conditions that conform to ASTM E1019-11(Method A)(C/S Analyzer) was also completed and is attached.

Each lot of cartridges are individually packaged and then gamma ray sterilized by TUV Sud accredited facility under procedure code P1501314 with a minimum dose of 25kGy. Certifications on file with the Sponsor and available for review.

Indications for Use:

The SER Pen Carain MicroSystem is intended to be used as a treatment to improve the appearance of facial acne scars in adults with Fitzpatrick Skin Types I - III, and to improve the appearance of surgical or traumatic hypertrophic scars on the abdomen in adults aged 22 years and older.

{6}------------------------------------------------

Traditional 510k Submission Carain MicroSystem

Summary of Non-Clinical Test Reports

The Sponsor intends to rely on the same performance data submitted in Collagen P.I.N. (K222199) and SkinStylus (K200044 and K231073) devices and not include any clinical data. Specifically, the Sponsor intends to rely on the following performance data reports.

Test CompletedStandard
BiocompatibilityCytotoxity - ISO 10993-5:2009
Sensitization -ISO 10993-10:2010
Irritation - Intracutaneous Injection Test GLP -ISO 10993-10:2010
Systemic Toxicity – Systemic Injection GLP -ISO 10993-11:2017
Pyrogenicity ISO 10993
Metallurgical Analysis – GLP - ASTM E1019-11(Method A)(C/S Analyzer)
Sterilization ValidationSterilization of Health Care Products – Moistheat ISO 17665-1:2006Sterilization of Medical Devices ISO 11737-1:2006 Sterilization of Medical Devices ISO11737-2: 2009 Sterility and Bacteriostasis/Fungistasis Tests ISO11737-2: 2009 and USP 71
Reprocessing ValidationISO 11737-1: 2018 (AAMI TIR-30; 2011;AAMI TIR-12:2004)
Fluid Ingress Validation
Shelf-life Testing
Standard Test Method for Seal Strength of FlexibleBarrier Materials ASTM F88/F88M-15
Standard Test Method for Detecting Seal Leaks inPorous Medical Packaging by Dye PenetrationASTM F1929-15

{7}------------------------------------------------

Summary of Substantial Equivalence

The SER Pen Carain MicroSystem and the predicate are for similar uses and rely on the same mode of action. Both devices include disposable needle cartridges with design features to mitigate the likelihood of cross-contamination between patients and to prevent needle depth greater than 2.5mm. Both devices also include a redundant safety feature to ensure no fluid enters the motor or housing.

{8}------------------------------------------------

Technological Characteristics
Device Name/ModelSER Pen CarainMicroSystem(MP1209SP)Collagen P.I.N.(PercutaneousInductionMicroneedling)SystemSkinStylusSteriLock®MicroSystem
510(k) Numbert/b/dK222199K200044
Indication for UseThe SER Pen CarainMicroSystem isintended to be usedas a treatment toimprove theappearance of facialacne scars in adultswith Fitzpatrick SkinTypes I - III, and toimprove theappearance ofsurgical or traumatichypertrophic scarson the abdomen inadults aged 22 yearsand older.for use as atreatment toimprove theappearance of facialacne scars in adultswith FitzpatrickSkin Types I - III,aged 22 years andolder.The SkinStylusSteriLock®MicroSystem isintended to be usedas a treatment toimprove theappearance ofsurgical ortraumatichypertrophic scarson the abdomen inadults aged 22years or older.
Mode of ActionIdenticalMicroneedling (usingone or more needlesto mechanicallypuncture and injureskin tissue foraesthetic use)Microneedling (usingone or more needlesto mechanicallypuncture and injureskin tissue foraesthetic use)
Power SourceSame 5VoltDC/ 1 amprechargeablelithium-ionbattery5 Volt DC/ 1 amprechargeable lithium-ion battery5 Volt DC/1 amprechargeable lithium-ion battery
Length/Width/shape/weight of bothhandpiece andcartridgeIdenticalIdenticalIdentical
Range of NeedleLength0.0-2.5mm0.0mm-3.0mm0.0-2.5mm
MaximumPenetration2.5mm3.0mm2.5mm
Needle GeometryIdentical, 12 or 36solid needles36 solid needles12 solid needle
SpeedIdentical ,6200RPM-9000 RPM7000RPM-9000RPM6200RPM-9000RPM
Sterility andCleaningSame. Disposablecartridge GammaRay Sterilized priorto packagingDisposablecartridge EthyleneOxideDisposablecartridge GammaRay Sterilized priorto packaging
Cross ContaminationFeatureIdentical:Cartridgedesign with asealing springinside the cartridgehousing, and abiolock sleeve tocover from thecartridge outshelland the wholehandpiece to stopcrosscontamination.SameCartridgedesign andintermediatedisinfectedhandpiece andnosecone withoptionalreprocessed(autoclavesterilized)nosecone;

Table 2 Technological Characteristics

{9}------------------------------------------------

{10}------------------------------------------------

Traditional 510k Submission Carain MicroSystem

Conclusion:

Therefore, taking into consideration Table 1 for Substantial Equivalence, an analysis of safety, indications, intended uses, performance, and technological properties, the SER Pen Carain MicroSystem raises no new issues of safety or effectiveness and has been found to be substantially equivalent to the predicate device.

§ 878.4430 Microneedling device for aesthetic use.

(a)
Identification. A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.