(223 days)
The Electric Breast Pump (Model LD-208L, LD-3010L, LD-3010, LD-2010) is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Electric Breast Pump is intended for a single user.
The Electric Breast Pump (LD-208L, LD-3010L, LD-3010, LD-3010, LD-2010) is an electrically powered breast pump to be used in a home environment by a single user. The device is provided non-sterile and can be used on one breast (single pumping) or on both breasts at the same time (double pumping).
The subject devices are capable of expression, stimulation, and the hybrid 'bionio' and 'variable frequency' modes. Stimulation mode is associated with six suction levels, and expression, bionic, and variable frequency modes are associated with nine suction levels. The subject devices feature various combinations of on/off buttons, level up/down buttons, mode selection buttons, suction level/time indicators, battery visual indicator display, and mode displays. The LD-208L, LD-3010L, and LD-2010L pumps are powered by internal, nonreplaceable, rechargeable lithium-ion batteries which are charged using the included AC power supply and cable. The LD-2010 and LD-3010 are powered by alkaline batteries or AC power.
The breast pumps use cyclic negative pressure (suction) to mimic the suckling patterns of a feeding infant. A DC motor drives a membrane vacuum pump to generate the suction required to stimulate and express breast milk. The timing of this pattern is dependent upon the suction/speed settings selected by the user and is pre-programmed in the devices. The devices are capable of producing peak suction levels between -40 and -290 mmHg at speeds between 20 and 120 cycles per minute. The subject device ensures backflow protection between the breast shield and the electronic components via a physical barrier (silicone diaphragm) mechanism.
All other components (i.e., motor unit) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 177.
The provided document is a 510(k) summary for an Electric Breast Pump (LD-208L, LD-3010L, LD-2010L, LD-3010, LD-2010). This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving independent effectiveness or advanced performance metrics typically associated with AI/ML devices. Therefore, much of the information requested in your prompt (e.g., sample size for test sets, expert ground truth, MRMC studies, standalone performance with effect sizes) is not applicable or generally required for a 510(k) clearance of a mechanical device like a breast pump.
Here's the information that can be extracted or inferred from the provided text:
1. A table of acceptance criteria and the reported device performance
For a breast pump, the "acceptance criteria" and "device performance" are primarily based on meeting safety standards, functional specifications, and demonstrating equivalence to a predicate. The document highlights various tests performed to ensure this.
| Acceptance Criteria (Measured Performance) | Reported Device Performance |
|---|---|
| Biocompatibility: Non-cytotoxic, non-irritating, non-sensitizing | User-contacting materials shown to be non-cytotoxic, non-irritating, and non-sensitizing. |
| Electrical Safety: Compliance with IEC 60601-1, IEC 62133-2, IEC 60086-5, IEC 60601-1-11 | Testing conducted in accordance with specified IEC standards. |
| Electromagnetic Compatibility: Compliance with IEC 60601-1-2 | Testing conducted in accordance with IEC 60601-1-2. |
| Software: Basic Documentation level as per FDA guidance | Software evaluated at the Basic Documentation level. |
| Vacuum Level Verification: Meet mode/cycle specifications | Devices meet mode/cycle specifications. |
| Backflow Protection: No liquid backflow into tubing | Liquid does not backflow into the tubing. |
| Use Life: Maintains specifications throughout proposed use life | Device maintains its specifications throughout its proposed use life. |
| Battery Performance: Battery functional during stated use-life | Battery remains functional during its stated use-life. |
| Battery Status Indicator: Functional during stated battery life | Battery status indicator remains functional during its stated battery life. |
| Peak Suction Levels: Between -40 and -290 mmHg | Devices are capable of producing peak suction levels between -40 and -290 mmHg. |
| Cycles per Minute: Between 20 and 120 cycles per minute | Devices are capable of producing speeds between 20 and 120 cycles per minute. |
2. Sample size used for the test set and the data provenance
The document does not specify sample sizes for the performance tests (e.g., how many units were tested for vacuum levels, how many use-life cycles). Data provenance details (country of origin, retrospective/prospective) are not provided as these are primarily engineering performance tests rather than clinical studies on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The ground truth for these performance tests is based on engineering specifications and adherence to international standards, not expert clinical consensus.
4. Adjudication method for the test set
Not applicable. This is not a study requiring adjudication of clinical findings.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an electric breast pump, a mechanical medical device, and does not involve AI or human readers for diagnostic interpretation.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device does not use algorithms for diagnostic or interpretive performance separate from its mechanical function.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the performance testing is based on:
- Engineering specifications: Pre-defined ranges for suction levels, cycles per minute, and other functional parameters.
- International standards: Compliance with standards like ISO 10993 for biocompatibility, IEC 60601 for electrical safety, and IEC 60601-1-2 for EMC.
- Design requirements verification: Ensuring the device meets its intended design functions (e.g., backflow protection, battery function).
8. The sample size for the training set
Not applicable. There is no "training set" in the context of this 510(k) submission for a non-AI/ML device.
9. How the ground truth for the training set was established
Not applicable. (See #8)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 19, 2024
Joytech Healthcare Co., Ltd. Yanyan Zhang Regulation Affair No.365. Wuzhou Road Hangzhou, Zhejiang 311100 CHINA
Re: K241322
Trade/Device Name: Electric Breast Pump (LD-208L, LD-3010L, LD-2010L, LD-3010, LD-2010) Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: Class II Product Code: HGX Dated: November 21, 2024 Received: November 21, 2024
Dear Yanyan Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/uniquedevice-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific
{2}------------------------------------------------
regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K241322
Device Name
Electric Breast Pump (Model LD-208L, LD-3010L, LD-3010L, LD-3010, LD-2010)
Indications for Use (Describe)
The Electric Breast Pump (Model LD-208L, LD-3010L, LD-3010, LD-2010) is a powered breast pump to be used by lactating women to express and collect milk from their breast Pump is intended for a single user.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The letters are stylized and the "O" is a circle, the "Y" is a V shape, the "E" is a rectangle, and the "H" is a T shape.
510(k) Summary
1. Submitter Information
Applicant: JOYTECH Healthcare Co.,Ltd. Address: No.365, Wuzhou Road, Hangzhou, Zhejiang 311100,China
2. Correspondent Information
| Contact: | Yanyan Zhang |
|---|---|
| Regulation Affairs | |
| Email: | zhangyy@sejoy.com |
| Phone: | 86 (571) 8195-7767 |
- Date prepared: December 12, 2024
4. Device Information
| Device Name: | Electric Breast Pump (LD-208L, LD-3010L, LD-2010L, LD-3010, LD-2010) |
|---|---|
| Common Name: | Powered breast pump |
| Regulation Number: | 21 CFR 884.5160 |
| Regulation Name: | Powered Breast Pump |
| Product Code: | HGX (Pump, Breast, Powered) |
| Regulatory Class: | Class II |
5. Predicate Device Information
| Device Name: | Electric Breast Pump Model: YM-8166 |
|---|---|
| 510(k) Number: | K232466 |
| Manufacturer: | Guangdong Youmeng Electrical Technology Co., Ltd. |
The predicate device has not been subject to a design-related recall.
6. Device Description
The Electric Breast Pump (LD-208L, LD-3010L, LD-3010, LD-3010, LD-2010) is an electrically powered breast pump to be used in a home environment by a single user. The device is provided non-sterile and can be used on one breast (single pumping) or on both breasts at the same time (double pumping).
The subject devices are capable of expression, stimulation, and the hybrid 'bionio' and 'variable frequency' modes. Stimulation mode is associated with six suction levels, and expression, bionic, and variable frequency modes are associated with nine suction levels. The subject devices feature various combinations of on/off buttons, level up/down buttons, mode selection buttons, suction level/time indicators, battery visual indicator display, and mode displays. The LD-208L, LD-3010L, and LD-2010L pumps are powered by internal, nonreplaceable, rechargeable lithium-ion batteries which are charged using the included AC power supply and cable. The LD-2010 and LD-3010 are powered by alkaline batteries or AC power.
The breast pumps use cyclic negative pressure (suction) to mimic the suckling patterns of a feeding infant. A DC motor drives a membrane vacuum pump to generate the suction required to stimulate and express breast milk. The timing of this pattern is dependent upon the suction/speed settings selected by the user and is pre-
{5}------------------------------------------------
K241322 Page 2 of 5
JC
programmed in the devices. The devices are capable of producing peak suction levels between -40 and -290 mmHg at speeds between 20 and 120 cycles per minute. The subject device ensures backflow protection between the breast shield and the electronic components via a physical barrier (silicone diaphragm) mechanism.
All other components (i.e., motor unit) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 177.
7. Indications for Use
The Electric Breast Pump (Model LD-208L, LD-3010L, LD-3010, LD-2010) is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Electric Breast Pump is intended for a single user.
8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Electric Breast Pump(Model LD-208L, LD-3010L, LD-2010L, LD-3010, LD-2010)K241322Subject Device | Electric Breast PumpModel: YM-8166K232466Predicate Device | Comparison | |
|---|---|---|---|
| Product Code | HGX | HGX | Same |
| Regulation Number | 21 CFR 884.5160 | 21 CFR 884.5160 | Same |
| Regulatory Class | Class II | Class II | Same |
| Patient Population | Lactating Women | Lactating Women | Same |
| Indications for Use | The Electric Breast Pump(Model LD-208L, LD-3010L,LD-2010L, LD-3010, LD-2010)is a powered breast pump to beused by lactating women toexpress and collect milk fromtheir breasts. The Electric BreastPump is intended for a singleuser. | The Electric Breast PumpModel: YM-8166 is intended toexpress milk from lactatingwomen in order to collect milkfrom their breasts. The device isintended for a single user | Similar |
| Single/doublepump | Single or double | Single or double | Same |
| Media separation(backflowprotection) | Yes | Yes | Same |
| Cycling controlmechanism | Microcontroller | Microcontroller | Same |
| Expression pattern | 2-Phase | 2-Phase | Same |
| Power supply | LD-3010L,LD-2010L: | USB connector (Only | Different |
| In: 100-240 VAC 50/60 Hz0.25AMains Out: 5VDC, 1.0ABatt Out: 3.7VDC 2000mAhLD-3010,LD-2010:In:100-240 VAC 50/60 Hz0.25AMains Out: 5VDC, 1.0ABatt Out: 4 AA batteriesLD-208L:In: 100-240 VAC 50/60 Hz0.25AMains Out: 5VDC, 1.0A(Onlyfor charging)Batt Out: 3.7VDC 1200mAh | for charging) by DC 5V, 2 A | ||
| Suction levels(expression) | 50-290 ± 30 mmHg | 85-280 ± 20 mmHg | Different |
| Suction levels(stimulation) | 40-100 ± 30 mmHg | 60-165 ± 20 mmHg | Different |
| Cycles per minute(expression) | 20-65 ± 5 cpm | 24-84 ± 5 cpm | Different |
| Cycles per minute(stimulation) | 80-120 ± 5 cpm | 39-123 ± 5 cpm | Different |
| Suction levels | 9 expression, bionic, variablefrequency; 6 stimulation | 16 levels | Different |
| Available modes | Stimulate mode,Expressionmode, Bionic suckingmode, Variable frequencypumping mode | Stimulation Mode, ExpressionMode,Pseudo-true Expression Mode,Array Mode, and Single/ doublePumping Mode | Different |
| User Interface | On-Off switch, mode changebutton, vacuum adjustmentbuttons, LED | Power button:Mode adjustmentbutton :Vacuum level increase/decreasebutton;Night light button(optional) ;Single/ double pumpingchange button | Similar |
| Adjustable SuctionLevels | Yes | Yes | Same |
| Mobile Application | No | No | Same |
| Design | LD-3010L, LD-3010, LD-2010L, LD-2010: TabletopLD-208L: Wearable | Not specified | -- |
Table 1: Comparator Table for Subject and Predicate Devices
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "JOYTECH" in a stylized font. The letters "JOY" are in dark blue, with a red triangle above the "Y". The letters "TEC" are in dark blue and red, with the "H" being red. The logo is simple and modern.
{7}------------------------------------------------
K241322 Page 4 of 5
The indications for use of the subject and predicate device are identical and they have the same intended use - the expression and collection of breast milk.
The subject and predicate devices have similar technological features, user interface and single/double pumping operation. However, as shown in the table above, there are technological differences between the subject and predicate device, including different overall vacuum pressure specifications/cycle speeds, available pumping modes, etc. However, the different technological characteristics of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness.
9. Summary of Non-Clinical Performance Testing
Biocompatibility
Biocompatibility studies, including Skin Irritation Testing, and Skin Sensitization testing were performed in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" as follows:
- . Cytotoxicity (ISO 10993-5:2009)
- . Skin Sensitization (ISO 10993-10:2021)
- Skin Irritation (ISO 10993-23:2021) .
The testing supports the biocompatibility of each device model. The user-contacting materials were shown to be non-cytotoxic, non-irritating, and non-sensitizing.
Electrical Safety
Testing was conducted in accordance with IEC 60601-1 Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance), IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems, IEC 60086-5:2016 Primary batteries - Part 5: Safety of batteries with aqueous electrolyte, and IEC 60601-1-11 Edition 3.1 2020-07 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
Electromagnetic Compatibility
Testing was conducted in accordance with IEC 60601-1-2:2014+A1: 2020 Medical Electrical Equipment -Part 1-2: "General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests."
Software
Software was evaluated at the Basic Documentation level as recommended in the 2023 FDA guidance document "Content of Premarket Submissions for Device Software Functions."
Performance Testing
Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:
- . Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications.
- . Backflow protection testing was conducted to verify liquid does not backflow into the tubing.
{8}------------------------------------------------
K241322 Page 5 of 5
Image /page/8/Picture/1 description: The image shows the word "JOYTECH" in a stylized font. The letters "J", "O", "Y", and "T" are in blue, while the "E", "C", and "H" are in red. The "Y" has a red triangle above it, and the "T" has a blue bar above it.
- . Use life testing was conducted to demonstrate that the device maintains its specifications throughout its proposed use life.
- Battery performance testing was conducted to demonstrate that the battery remains functional during its stated battery use-life.
- . Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
10. Conclusion
The results of the performance testing described above demonstrate that the Electric Breast Pump (LD-208L, LD-3010L, LD-3010L, LD-3010, LD-2010) is as safe and effective as the predicate device and supports a determination of substantial equivalence.
§ 884.5160 Powered breast pump.
(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).