K Number
K241242
Device Name
SIGNA MAGNUS
Date Cleared
2024-10-17

(167 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SIGNA MAGNUS system is a head-only magnetic resonance scanner designed to support high resolution, high signal-tonoise ratio, diffusion-weighted imaging, and short scan times. SIGNA MAGNUS is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the head, neck, TMJ, and limited cervical spine on patients 6 years of age and older. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by SIGNA MAGNUS reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.

Device Description

SIGNA MAGNUS is a 3.0T high-performance magnetic resonance imaging system designed to support imaging of the head, neck, TMJ and limited cervical spine. The system supports scanning in axial, coronal, sagittal, oblique, and double oblique planes using a variety of pulse sequences, imaging techniques, acceleration methods, and reconstruction algorithms. The system can be delivered as a new system installation, or as an upgrade to existing compatible whole-body 3.0T MR

Key aspects of the system design:
• An asymmetrically designed, head-only gradient coil that achieves up to 300 mT/m peak gradient amplitude and 750 T/m/s peak slew rate.
• A graduated patient bore size, starting at 74 cm at the entry down to 37 cm at isocenter.
• Uses the same magnet as a conventional whole-body 3.0T system, with integral active shielding and a zero boil-off cryostat.
• Can be installed as a new system or upgraded from an existing compatible whole-body 3.0T MR system.
• A dockable mobile patient table.
• Oscillating Diffusion Encoding (ODEN) - a spectral diffusion technique that uses a sinusoidal diffusion gradient waveform.

AI/ML Overview

The provided text is a 510(k) Summary for the GE Healthcare SIGNA MAGNUS, a magnetic resonance diagnostic device. The summary indicates that this device did not require clinical studies to support substantial equivalence. Therefore, there are no acceptance criteria, performance metrics, sample sizes for test/training sets, expert qualifications, or other details related to clinical performance studies to report for this device, as these were not performed or deemed necessary by the manufacturer for this submission.

The document states:
"The subject of this premarket submission, SIGNA MAGNUS did not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission. The sample clinical images demonstrate acceptable diagnostic image performance of SIGNA MAGNUS in accordance with the FDA Guidance 'Submission of Premarket Notifications fo October 10, 2023. The image quality of SIGNA MAGNUS is substantially equivalent to that of the predicate device."

This means that the substantial equivalence was primarily based on non-clinical tests and a comparison of technological characteristics and indications for use with a predicate device (SIGNA Premier (K193282)). The non-clinical tests focused on safety and performance in compliance with various voluntary standards (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, IEC 62304, IEC 60601-1-6, IEC 62366-1, IEC62464-1, ISO 10993-1, NEMA MS, NEMA PS3 DICOM).

Due to the absence of a clinical study, the requested information cannot be extracted from the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

October 17, 2024

GE Medical Systems, LLC Brian Zielski Sr. Lead Specialist, Regulatory Affairs - MR 3200 N Grandview Blvd Waukesha, Wisconsin 53188

Re: K241242

Trade/Device Name: SIGNA MAGNUS Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LNI Dated: September 10, 2024 Received: September 10, 2024

Dear Brian Zielski:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Justina Justina Tam -Tam -S Date: 2024.10.17 14:38:06 S -04'00' for Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241242

Device Name SIGNA MAGNUS

Indications for Use (Describe)

SIGNA MAGNUS system is a head-only magnetic resonance scanner designed to support high resolution, high signal-tonoise ratio, diffusion-weighted imaging, and short scan times. SIGNA MAGNUS is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the head, neck, TMJ, and limited cervical spine on patients 6 years of age and older. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by SIGNA MAGNUS reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram inside of a circle on the left, and the words "GE HealthCare" in a sans-serif font on the right. The logo is purple.

SIGNA MAGNUS

Traditional 510(k) Premarket Notification Attachment 15 – 510(k) Summary

510(k) Summary

K241242

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:May 1, 2024
Submitter:GE Medical Systems, LLC3200 N. Grandview Blvd.Waukesha, WI 53188
Primary Contact:Brian R. ZielskiSr. Lead Specialist, Regulatory Affairs - MRPhone: 262-227-3596Email: brian.zielski@gehealthcare.com
Secondary Contact:Andrew MendenDirector, Regulatory Affairs - MRPhone: 262-308-5719Email: andrew.menden@gehealthcare.com
Device Trade Name:SIGNA MAGNUS
Common / Usual Name:Magnetic Resonance Diagnostic Device
Classification Name:Magnetic Resonance Diagnostic Device per 21 CFR892.1000
Product Code:LNH, LNI
Predicate Device(s):SIGNA Premier (K193282)
Reference Device(s):SIGNA Champion (K233728)
Device Description:SIGNA MAGNUS is a 3.0T high-performance magneticresonance imaging system designed to support imagingof the head, neck, TMJ and limited cervical spine. Thesystem supports scanning in axial, coronal, sagittal,oblique, and double oblique planes using a variety ofpulse sequences, imaging techniques, accelerationmethods, and reconstruction algorithms. The system canbe delivered as a new system installation, or as anupgrade to existing compatible whole-body 3.0T MR
Key aspects of the system design:• An asymmetrically designed, head-only gradient coilthat achieves up to 300 mT/m peak gradientamplitude and 750 T/m/s peak slew rate.• A graduated patient bore size, starting at 74 cm atthe entry down to 37 cm at isocenter.• Uses the same magnet as a conventional whole-body 3.0T system, with integral active shielding and azero boil-off cryostat.• Can be installed as a new system or upgraded froman existing compatible whole-body 3.0T MR system.• A dockable mobile patient table.• Oscillating Diffusion Encoding (ODEN) - a spectraldiffusion technique that uses a sinusoidal diffusiongradient waveform.
Indications for Use:SIGNA MAGNUS system is a head-only magneticresonance scanner designed to support high resolution,high signal-to-noise ratio, diffusion-weighted imaging,and short scan times. SIGNA MAGNUS is indicated for useas a diagnostic imaging device to produce axial, sagittal,coronal, and oblique images, spectroscopic images,parametric maps, and/or spectra, dynamic images of thestructures and/or functions of the head, neck, TMJ, andlimited cervical spine on patients 6 years of age andolder. Depending on the region of interest being imaged,contrast agents may be used.
The images produced by SIGNA MAGNUS reflect thespatial distribution or molecular environment of nucleiexhibiting magnetic resonance. These images and/orspectra, when interpreted by a trained physician, yieldinformation that may assist in diagnosis.
Technology:SIGNA MAGNUS employs the same functional scientifictechnology as its predicate device.
Comparison of Indicationsfor Use:SIGNA MAGNUS and the predicate device's Indicationsfor Use are substantially equivalent. While the predicateis intended for all anatomies, SIGNA MAGNUS is intendedfor only the head, neck, TMJ, and cervical spine.
As such, GE HealthCare considers that SIGNA MAGNUShas the same intended use as the predicate device inaccordance with the FDA's guidance document "The510(k) Program: Evaluating Substantial Equivalence inPremarket Notifications [510(k)]", dated 28 July 2014.
Comparison ofTechnologicalCharacteristics:Overall, SIGNA MAGNUS employs the same fundamentalscientific technology as the predicate device.System Design: Both SIGNA MAGNUS and the predicatedevice include the 3.0 T magnet, RF transmitarchitecture, RF receive chain and software applicationsuite. However, the most notable technologicaldifference between the SIGNA™ MAGNUS and thepredicate device are:
Gradient System: The SIGNA MAGNUS can delivera peak gradient amplitude of 300 mT/m and peakgradient slew rate of 750 T/m/s compared to thepredicate's peak gradient amplitude of 80 mT/mand peak gradient slew rate of 200 T/m/s. PNSand CS thresholds are also significantly differentbetween the two systems, allowing moreperformance to be used on the SIGNA MAGNUSsystem.
Applications: The SIGNA MAGNUS uses the SIGNAWorks suite of applications with the addition ofOscillating Diffusion ENcoding (ODEN) - a spectraldiffusion technique that uses a sinusoidaldiffusion gradient waveform. ODEN is not offeredon the predicate device.
Field-of-View (FOV): The SIGNA MAGNUS FOV(26cm x 26 cm x 26cm) is intended for head, neck,TMJ, and limited cervical spine scanning. As such,a smaller maximum imaging FOV is specifiedcompared to the predicate device with an FOV of50cm x 50cm x 50cm.
Bore Size: Unlike the straight cylindrical patientbore of the predicate device, the SIGNA MAGNUS
system has a stepped progressively narrowingpatient bore. The widest part at the patient boreentry accommodates the arms and torso. It has a74 cm inner diameter. The patient bore diameterat the shoulders is narrowed to 54 cm. At thehead, the SIGNA MAGNUS patient bore diameteris 37 cm compared to 70 cm for the predicatedevice.
• Receive Channels: The SIGNA MAGNUS has 64independent RF channels to account for theimaged anatomies of head, neck, TMJ, and limitedcervical spine, compared to 146 for the predicatedevice which is used for whole-body imaging.
• Max RF Power Output: The SIGNA MAGNUS has7.5 kW peak total maximum RF power outputcompared to 30 kW for the predicate device.
Operating Principles: SIGNA MAGNUS functions usingthe same operating principles as the predicate device.
Materials: SIGNA MAGNUS and the predicate deviceboth use the same materials.
Safety and Performance Testing: Both SIGNA MAGNUSand the predicate device comply with the same safetyand performance testing (see Determination ofSubstantial Equivalence, below).
These technological differences do not raise anyquestions regarding safety and effectiveness. Bothdevices must address questions of whether they providean adequate level of image quality appropriate fordiagnostic use. The performance data described in thissubmission include results of both bench testing andclinical testing that show the image quality performanceof SIGNA MAGNUS compared to the predicate device.
Determination ofSubstantial Equivalence:Summary of Non-Clinical Tests:SIGNA MAGNUS and the predicate device were subject tosimilar risk management testing to demonstratesubstantial equivalence of safety and performance.
Testing to the following voluntary standards included:• ANSI AAMI ES60601-1• IEC 60601-1-2• IEC 60601-2-33• IEC 62304• IEC 60601-1-6• IEC 62366-1• IEC62464-1• ISO 10993-1
In addition, SIGNA MAGNUS complies with applicableNEMA MS standards for MRI and NEMA PS3 standard forDICOM, as does the predicate device.
Both SIGNA MAGNUS and the predicate device havedemonstrated successful biocompatibility, asdemonstrated by ISO 10993 testing and by their historyof use in previously cleared devices.
The following quality assurance measures were appliedto the development of the subject device, as they werefor the predicate device:• Risk Analysis• Requirements Reviews• Design Reviews• Verification• Validation
Verification documents, validation documents, and testreports have been provided for more details. No newquestions of safety and effectiveness were raised duringnonclinical testing.
Summary of Clinical Tests:The subject of this premarket submission, SIGNAMAGNUS did not require clinical studies to supportsubstantial equivalence. Sample clinical images havebeen included in this submission. The sample clinicalimages demonstrate acceptable diagnostic imageperformance of SIGNA MAGNUS in accordance with theFDA Guidance “Submission of Premarket Notifications fo
October 10, 2023. The image quality of SIGNA MAGNUSis substantially equivalent to that of the predicate device.
Substantial Equivalence Conclusion:
The indications for use of the proposed device arecomparable to the claimed predicate device. SIGNAMAGNUS employs equivalent technology to the claimedpredicate device.
Additionally, the results from the above non-clinical testsdemonstrate that the device performs as intended.
Therefore, SIGNA MAGNUS is substantially equivalent tothe predicate device to which it has been compared.
Conclusion:In conclusion, GE Healthcare considers SIGNA MAGNUSto be at least as safe and effective, and its performance issubstantially equivalent to the predicate device.

{5}------------------------------------------------

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is purple and features a stylized "GE" monogram surrounded by a circular design. The text "GE HealthCare" is also purple and is written in a clean, sans-serif font.

.

Traditional 510(k) Premarket Notification Attachment 15 – 510(k) Summary

{7}------------------------------------------------

SIGNA MAGNUS

Image /page/7/Picture/1 description: The image shows the logo for GE HealthCare. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is purple and features a stylized "GE" design within a circle. The text "GE HealthCare" is also purple and is written in a clean, sans-serif font.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram in a circle on the left, followed by the words "GE HealthCare" on the right. The logo is purple in color. The text is sans-serif.

SIGNA MAGNUS

Traditional 510(k) Premarket Notification Attachment 15 – 510(k) Summary

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for GE HealthCare. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is purple and contains a stylized "GE" in white. The text "GE HealthCare" is also purple.

Traditional 510(k) Premarket Notification Attachment 15 – 510(k) Summary

Image /page/9/Picture/5 description: The image is a blurry, out-of-focus shot with a predominantly white background. There is a small, dark red or maroon blob in the center-right of the frame. The blob is slightly pixelated, suggesting it may be a small object or a point of light.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.