K Number
K241129
Device Name
Upper Arm Electronic Blood Pressure Monitor
Date Cleared
2024-07-03

(70 days)

Product Code
Regulation Number
870.1130
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Upper Arm Electronic Blood Pressure Monitor is intended for used by a person older than twelve (12) years to measure the systolic and diastolic blood pressure and pulse rate.
Device Description
The Upper Arm Electronic Blood Pressure Monitor uses the Oscillometric Measuring method to detect blood pressure. Before every measurement, the unit establishes a " zero pressure" equivalent to the atmospheric pressure. Then it starts inflating the arm cuff, meanwhile, the unit detects pressure oscillations generated by beat-to-beat pulsation , which is used to determine the systolic and diastolic pressure, and also pulse rate.
More Information

Not Found

No
The description details a standard oscillometric blood pressure measurement method and does not mention any AI or ML components.

No.
The device measures blood pressure and pulse rate, which are diagnostic measurements for monitoring, rather than actively treating a condition.

Yes
Explanation: The device is intended to "measure the systolic and diastolic blood pressure and pulse rate," which are measurements used to assess a patient's health status.

No

The device description explicitly states it uses an "Upper Arm Electronic Blood Pressure Monitor" and "starts inflating the arm cuff," indicating the presence of hardware components beyond just software.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. These tests are performed outside of the body (in vitro).
  • Device Function: The description clearly states that this device measures blood pressure and pulse rate by applying a cuff to the upper arm and using the oscillometric method. This is a direct measurement taken on the body (in vivo), not a test performed on a sample taken from the body.
  • Intended Use: The intended use is to measure blood pressure and pulse rate, which are physiological parameters, not to diagnose a disease or condition based on analysis of a biological sample.

The information provided describes a standard non-invasive blood pressure monitor, which is a medical device but not an IVD.

N/A

Intended Use / Indications for Use

Upper Arm Electronic Blood Pressure Monitor is intended for used by a person older than twelve (12) years to measure the systolic and diastolic blood pressure and pulse rate.

Product codes

DXN

Device Description

The Upper Arm Electronic Blood Pressure Monitor uses the Oscillometric Measuring method to detect blood pressure. Before every measurement, the unit establishes a " zero pressure" equivalent to the atmospheric pressure. Then it starts inflating the arm cuff, meanwhile, the unit detects pressure oscillations generated by beat-to-beat pulsation , which is used to determine the systolic and diastolic pressure, and also pulse rate.

Upper Arm Electronic Blood Pressure Monitor is intended for used by a person older than twelve (12) years to measure the systolic and diastolic blood pressure and pulse rate.

There are two models of cuffs that can be used together: 22-42cm and 32-52cm

This device has two series: A01(including A01-SLB, A01-BL, A01-SL, A01) and A02 (including A02-SE4、A02-SEB、A02-BE、A02-SE、A02-E)

They have same Indications for use and similar technological characteristics, the electrical circuit design, critical components used and internal wiring are also similar.We chose model A02-SE4 with the most complex functions as the representative model.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Upper Arm

Indicated Patient Age Range

older than twelve (12) years

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The device was tested according to ISO 81060-2:2018+A1:2020 Non-invasive sphygmomanometers - Part 2: Clinical validation of intermittent automated measurement type. Two clinical studies were performed on the device matching with different models of cuff. The study population consisted of 87, and 86 qualified subjects respectively. all subjects were adults. All data's mean error and standard deviation of differences in systolic, diastolic pressure is not beyond the limits set as per ISO 81060-2:2020. No adverse effect and/or complications are found in this study.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K160019

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

K241129

ShenZhen GoodlyMed Technology Co., Ltd. % Kyra Kang, Director Landlink Healthcare Technology (Shanghai) Co., Ltd. Room 1308, Baohua International Plaza, West Guangzhong Road Jingan District, Shanghai China

Re: K241129

Trade/Device Name: Upper Arm Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 23, 2024 Received: April 24, 2024

Dear Kyra Kang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

2

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

Submission Number (if known)

K241129

Device Name

Upper Arm Electronic Blood Pressure Monitor

Indications for Use (Describe)

Upper Arm Electronic Blood Pressure Monitor is intended for used by a person older than twelve (12) years to measure the systolic and diastolic blood pressure and pulse rate.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

510(k) Summary

l. Submitter

ShenZhen GoodlyMed Technology Co.,Ltd.

701, Building C, Area C, Datianyang Industrial Zone, Shiwei Community, Matian Street, Guangming District, 518107 Shenzhen, PEOPLE'S REPUBLIC OF CHINA

Contact person: Ms. Maria Position: Regulatory manager Tel.: +86 -18002580887 E-mail: manager@goodlymed.com

Submission Correspondent

Ms. Kyra Kang Landlink Healthcare Technology (Shanghai) Co., Ltd. E-mail: kyra.kang@landlink-health.com

II. Proposed Device

Device Trade Name:Upper Arm Electronic Blood Pressure Monitor
Common name:Electronic Blood Pressure Monitor
Regulation Number:21 CFR 870.1130
Regulatory Class:Class II
Product code:DXN
Review Panel:Cardiovascular

III. Predicate Devices

510(k) Number:K160019
Trade name:U80 Series Electronic Blood Pressure Monitor
Common name:Electronic Blood Pressure Monitor
Classification:Class II
Product Code:DXN
Manufacturer:Shenzhen Urion Technology Co., Ltd.

5

IV. Device description

The Upper Arm Electronic Blood Pressure Monitor uses the Oscillometric Measuring method to detect blood pressure. Before every measurement, the unit establishes a " zero pressure" equivalent to the atmospheric pressure. Then it starts inflating the arm cuff, meanwhile, the unit detects pressure oscillations generated by beat-to-beat pulsation , which is used to determine the systolic and diastolic pressure, and also pulse rate.

Upper Arm Electronic Blood Pressure Monitor is intended for used by a person older than twelve (12) years to measure the systolic and diastolic blood pressure and pulse rate.

There are two models of cuffs that can be used together: 22-42cm and 32-52cm

This device has two series: A01(including A01-SLB, A01-BL, A01-SL, A01) and A02 (including A02-SE4、A02-SEB、A02-BE、A02-SE、A02-E)

They have same Indications for use and similar technological characteristics, the electrical circuit design, critical components used and internal wiring are also similar.We chose model A02-SE4 with the most complex functions as the representative model.

V. Indication for use

Upper Arm Electronic Blood Pressure Monitor is intended for used by a person older than twelve (12) years to measure the systolic and diastolic blood pressure and pulse rate.

| Character
istics | Proposed device | Predicate device
(K160019) | Discussion
n |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Product
Code | DXN | DXN | Same |
| Indications
For Use | Upper Arm Electronic Blood
Pressure Monitor is intended
for used by a person older
than twelve (12) years to
measure the systolic and
diastolic blood pressure and
pulse rate. | U80 Series Electronic Blood
Pressure Monitor is intended
to measure the systolic and
diastolic blood pressure as
well as the pulse rate of adult
person
via non-invasive oscillometric
technique in which an
inflatable cuff is wrapped
around the upper arm. It can
be used at medical facilities
or at home. The intended | Same |
| | | upper arm circumference is
22-36 cm. | |
| Measurem
ent
Type | Upper arm | Upper arm | Same |
| Patient
Population | people over 12 years old | Adults
Person
over
12 | Same |
| Measurem
ent
Item | SYS, DYS, Pulse Rate | SYS, DYS, Pulse Rate | Same |
| Principle | Oscillometric | Oscillometric | Same |
| BP Range | 0 ~ 299 mmHg | 0 ~ 299 mmHg | Same |
| BP
Accuracy | $\pm$ 3mmHg | $\pm$ 3 mmHg | Same |
| PR Range | 40-199 bpm | 40-199 bpm | Same |
| Cuff Size | Arm Circumference: 22-
42cm, 32-52cm | Arm Circumference:
22-36 cm | Similar |
| Power
Supply | A01 Series : Four AAA
batteries or AC adapter
A02 Series : Four AA batteries
or AC adapter | Four AA batteries or AC
adapter | Similar |
| Software
Level
Concern | Moderate | Moderate | Same |

Table 1 General Comparison of Unner Arm Electronic Blood Pressure Monitor

VI. Comparison of technological characteristics with the predicate devices

6

7

WirelessA01: bluetooth/Different 1
A02:4G

Analysis 1

The subject device has Wireless module while the predicate device does not. However, it is to transfer measurement results only. As for the wireless technology of the subject device, it has also been validated according to Relevant standards. As demonstrated in relevant test reports, the difference here does not raise any issues concerning safety and effectiveness.

VII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"

The biocompatibility testing includes the following tests:

  • Cytotoxicity
  • Sensitization
  • Irritation

Electrical Safety, EMC, Performance testing

● IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance

● IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC 60601-1-6: Medical Electrical Equipment - Part 1-6: General requirements ● for safety - Collateral Standard: Usability.

● IEC 62366-1: Medical devices – Application of usability engineering to medical devices.

8

● IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests

● IEC 80601-2-30: Medical electrical equipment - Particular requirements for basic safety and essential performance of automated non-invasive the sphygmomanometers.

● IEC 80369-5 Small-bore connectors for liguids and gases in healthcare applications - Part 5: Connectors for limb cuff inflation applications.

Wireless testing

● ANSI C63.27-2021 American National standard for evaluation of wireless coexistence

● AAMI TIR69:2017/(R2020) Risk management of radio-frequency wireless coexistence for medica devices and systems

● Radio frequency wireless technology in medical devices Guidance for industry and food and drug administration staff (August 14,2013)

● FCC Part 15 Subpart C 10-1-2022 Edition Part 15-Radio frequency devices Subpart C-Intentional Radiators

Cybersecurity

● Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions

Shelf-life

● 5-year shelf-life of the device has been evaluated by Shelf-life test.

Reprocessing

● ISO 17664 Processing of health care products -Information to be provided by the medical device manufacturer for the processing of medical devices

VIII. Clinical Testing

The device was tested according to ISO 81060-2:2018+A1:2020 Non-invasive sphygmomanometers - Part 2: Clinical validation of intermittent automated measurement type. Two clinical studies were performed on the device matching with different models of cuff. The study population consisted of 87, and 86 qualified subjects

9

respectively. all subjects were adults. All data's mean error and standard deviation of differences in systolic, diastolic pressure is not beyond the limits set as per ISO 81060-2:2020. No adverse effect and/or complications are found in this study.

IX. Conclusion

The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is as safe and effective as the predicate device. Accordingly, the proposed device is substantially equivalent to the predicate device.