K Number
K241125

Validate with FDA (Live)

Device Name
VIVIX-S 1751S
Manufacturer
Date Cleared
2024-11-15

(206 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VIVIX-S 1751S series is used for the general-purpose diagnostic procedures, and as well as intended to replace radiographic film/ screen systems. The VIVIX-S 1751S series is not intended for mammography applications.

Device Description

The VIVIX-S 1751S, available in models FXRD-1751SA and FXRD-1751SB, features a 17x51 inch imaging area. This device intercepts x-ray photons and uses a scintillator to emit visible spectrum photons. The FXRD-1751SA model uses a Csl:Tl (Thallium doped Caesium lodide) scintillator, while the FXRD-1751SB model uses a Gadox (Gadolinium Oxysulfide) scintillator. These photons illuminate an array of photo (a-SI) detectors, creating electrical signals are then converted to digital values, which are processed by software to produce digital images displayed on monitors. The VIVIX-S 1751S must be integrated with an operating PC and an X-ray generator, and it can communicate with the generator via cable. It is designed for capturing and transferring digital x-ray images for radiography diagnosis. Note that the X-ray generator and imaging software are not included with the VIVIX-S 1751S.

AI/ML Overview

The document describes a 510(k) submission for the VIVIX-S 1751S digital X-ray detector. The acceptance criteria and the study proving the device meets these criteria are primarily demonstrated through a comparison to a predicate device (K190611) and performance testing based on established standards.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by demonstrating substantial equivalence to the predicate device in terms of technological characteristics and performance metrics, as well as compliance with recognized standards. The "performance" column shows the subject device's reported values relative to the predicate device.

ParameterAcceptance Criteria (Predicate Device K190611, FXRD-1751SB)Reported Device Performance (Subject Device K241125, FXRD-1751SA)Equivalence
Technological CharacteristicsSame as Predicate deviceSame as Predicate deviceSubstantially Equivalent
Intended UseVIVIX-S 1751S series is used for general-purpose diagnostic procedures, and to replace radiographic film/screen systems. Not for mammography.VIVIX-S 1751S series is used for general-purpose diagnostic procedures, and to replace radiographic film/screen systems. Not for mammography.Equivalent
Operating PrincipleSame as Predicate deviceSame as Predicate deviceEquivalent
Design FeaturesSame as Predicate deviceSame as Predicate deviceEquivalent
Communication MethodSame as Predicate deviceSame as Predicate deviceEquivalent
ResolutionSame as Predicate deviceSame as Predicate deviceEquivalent
Scintillator TypeGadoxCsI (FXRD-1751SA), Gadox (FXRD-1751SB)Different scintillator for FXRD-1751SA model, but performance shown to be comparable. FXRD-1751SB is identical.
Performance (Optical / Imaging)
MTF (0.5 lp/mm)≥ 81≥ 83Similar
MTF (1 lp/mm)≥ 56≥ 63Similar
MTF (2 lp/mm)≥ 22≥ 30Similar
MTF (3 lp/mm)≥ 9≥ 14Similar
DQE (0.5 lp/mm)≥ 29≥ 38Similar
DQE (1 lp/mm)≥ 22≥ 33Similar
DQE (2 lp/mm)≥ 11≥ 23Similar
DQE (3 lp/mm)≥ 4≥ 14Similar
Compliance with StandardsCompliance with 21CFR1020.30, 21CFR1020.31, IEC 60601-1, CAN/CSA-C22.2 No. 60601-1, ANSI/AAMI ES60601-1, IEC 60601-1-2Complies with all listed standardsCompliant
Diagnostic CapabilityEquivalent to predicate deviceDemonstrated equivalent diagnostic capabilityEquivalent

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states, "A-Qualified Expert Evaluation study according to CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted..." However, it does not specify the sample size of cases/images used in this clinical evaluation study.
The data provenance (country of origin, retrospective/prospective) is not explicitly stated in the provided text.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document mentions an "Expert Evaluation study" but does not specify the number of experts or their exact qualifications (e.g., "radiologist with 10 years of experience"). It only indicates that it was a "Qualified Expert Evaluation study."

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe the adjudication method used for the expert evaluation study.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader multi-case (MRMC) study was conducted as part of the "Qualified Expert Evaluation study." The primary goal of this study was to confirm that the subject device (VIVIX-S 1751S - FXRD-1751SA) provides images of equivalent diagnostic capability to the predicate device (VIVIX-S 1751S - FXRD-1751SB).

The document does not mention the involvement of AI in this study, nor does it quantify any improvement of human readers with AI assistance. The study described is a comparison of two different X-ray detector technologies, not an AI-assisted reading study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This device is an X-ray detector, not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" performance without a human-in-the-loop is not applicable in the same way it would be for an AI diagnostic software. The performance metrics reported (MTF, DQE) are physical image quality parameters of the detector itself, which could be considered "standalone" in this context as they characterize the device's inherent imaging capability.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the "clinical test" (Expert Evaluation study) was established by "Qualified Expert Evaluation" to assess "equivalent diagnostic capability" of the images. This suggests a form of expert consensus or individual expert readings to determine the diagnostic quality of the images produced by the subject device compared to the predicate. It does not mention pathology or outcomes data as the primary ground truth.

8. The sample size for the training set

The document does not mention a training set in the context of this device because it is a hardware device (X-ray detector), not an AI algorithm that requires a training phase.

9. How the ground truth for the training set was established

Since there is no mention of a training set for an AI algorithm, the question of how its ground truth was established is not applicable to this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

November 15, 2024

Vieworks Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 115 Roosevelt. STE 200 IRVINE, CA 92620

Re: K241125

Trade/Device Name: Vivix-S 1751s Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: April 23, 2024 Received: April 23, 2024

Dear Priscilla Chung:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241125

Device Name VIVIX-S 1751S

Indications for Use (Describe)

VIVIX-S 1751S series is used for the general-purpose diagnostic procedures, and as well as intended to replace radiographic film/ screen systems. The VIVIX-S 1751S series is not intended for mammography applications.

Type of Use (Select one or both, as applicable)

✓ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

(K241125)

This 510(k) summary information is prepared in accordance with 21 CFR807.92

1. Date Prepared [21 CFR 807.92(a) (1)]

11/13/2024

2. Submitter's Information [21 CFR 807.92(a) (1)]

Name of Sponsor: Vieworks Co., Ltd. (Gwanyang-dong) 41-3, Burim-ro 170beon-gil,Dongan-gu, Address: Anyang-si, Gyeonggi-do, 431-060 Republic of Korea Contact Name: Oh, Kevin / Regulatory Affairs Associate Registration Number: 3006013411 Name of Manufacturer: Same as Sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]

Trade Name:VIVIX-S 1751S
Model Name:FXRD-1751SAFXRD-1751SB
Common Name:Digital Flat Panel X-ray Detector
Classification Name:Stationary X-ray system
Classification Panel:Radiology
Classification Regulation:21 CFR 892.1680
Product Code:MQB
Device Class:2

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

510(k) Number:K190611
Trade Name:VIVIX-S 1751S
Model Name:FXRD-1751SB
Common Name:Digital Flat Panel X-ray Detector
Classification Name:Stationary X-ray system
Classification Panel:Radiology
Classification Regulation:21 CFR 892.1680
Product Code:MQB
Device Class:2

{5}------------------------------------------------

5. Description of the Device [21 CFR 807.92(a) (4)]

o General Description

The VIVIX-S 1751S, available in models FXRD-1751SA and FXRD-1751SB, features a 17x51 inch imaging area. This device intercepts x-ray photons and uses a scintillator to emit visible spectrum photons. The FXRD-1751SA model uses a Csl:Tl (Thallium doped Caesium lodide) scintillator, while the FXRD-1751SB model uses a Gadox (Gadolinium Oxysulfide) scintillator. These photons illuminate an array of photo (a-SI) detectors, creating electrical signals are then converted to digital values, which are processed by software to produce digital images displayed on monitors. The VIVIX-S 1751S must be integrated with an operating PC and an X-ray generator, and it can communicate with the generator via cable. It is designed for capturing and transferring digital x-ray images for radiography diagnosis. Note that the X-ray generator and imaging software are not included with the VIVIX-S 1751S.

6. Intended Use [21 CFR 807.92(a)(5)]

VIVIX-S 1751S series is used for the general-purpose diagnostic procedures, and as well as intended to replace radiographic film/ screen systems. The VIVIX-S 1751S series is not intended for mammography applications.

7. Technological Characteristics [21 CFR 807.92(a) (6)]

Comparisons with the predicate, the subject device shows the technological characteristics substantially equivalent to the predicate device. The proposed device is functionally identical to the predicate device.

8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

When compared to the predicate device (K190611), the VIVIX-S 1751S (FXRD-1751SA and FXRD-1751SB) presented in this submission has the same:

  • Intended Use
  • Technological characteristics
  • · Operating principle
  • · Design features
  • Communication Method
  • Resolution

There is similar performance as follows:

  • Performance (MTF)
  • Performance (DQE)

There are no significant differences between the subject device and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles, and intended use.

{6}------------------------------------------------

ParameterPredicate DeviceSubject DeviceEquivalence
510(k) NumberK190611K241125-
ManufacturerVieworks Co., Ltd.--
Device NameVIVIX-S 1751SVIVIX-S 1751S-
DetectorFXRD-1751SBFXRD-1751SAFXRD-1751SB
ScintillatorGadoxCsIGadox
Power Supply UnitFXRP-01AFXRP-01AFXRP-01APower Supply
IO Box---Power Supply
Software---Software
MTF(Min)0.5 lp/mm: 811lp/mm: 562lp/mm: 223lp/mm: 90.5 lp/mm: 831lp/mm: 632lp/mm: 303lp/mm: 140.5 lp/mm: 811lp/mm: 562lp/mm: 223lp/mm: 9
DQE(Min)0.5 lp/mm: 291lp/mm: 222lp/mm: 113lp/mm: 40.5 lp/mm: 381lp/mm: 332lp/mm: 233lp/mm: 140.5 lp/mm: 291lp/mm: 222lp/mm: 113lp/mm: 4
Common NameDigital Flat Panel X-ray DetectorEquivalent
Classification NameStationary X-ray systemEquivalent
Classification PanelRadiologyEquivalent
Classification Regulation21 CFR 892.1680Equivalent
Product CodeMQBEquivalent
Device Class2Equivalent
Intended UseVIVIX-S 1751S series is used for the general-purpose diagnostic procedures, and as well as intended to replace radiographic film/ screen systems. The VIVIX-S 1751S series is not intended for mammography applications.VIVIX-S 1751S series is used for the general-purpose diagnostic procedures, and as well as intended to replace radiographic film/ screen systems. The VIVIX-S 1751S series is not intended for mammography applications.Equivalent

9. Summary of Non-Clinical Data

A comparison test was conducted between the subject device (VIVIX-S 1751S - FXRD-1751SA) and the predicate device (K190611) on the items such as DQE, MTF and spatial resolution. Both devices show similar performance results.

{7}------------------------------------------------

These detectors comply with the following international and FDA-recognized consensus standards:

  • · 21CFR1020.30, Diagnostic X-ray Systems and their major components
  • · 21CFR1020.31, Radiographic equipment
    • IEC 60601-1 Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety and Essential Performance.
  • CAN/CSA-C22.2 No. 60601-1 (2008) (Medical Equipment Part 1 : General Requirements for Basic Safety and Essential Performance) (includes National Differences for Canada)
  • ANSI/AAMI ES60601-1 (2005+ C1:09+A2:10) (Medical Electrical Equipment Part 1
  • IEC 60601-1-2 Medical Electrical Equipment Part 1-2 : General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests

10. Summary of Clinical Data

A-Qualified Expert Evaluation study according to CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices was conducted, and the study confirmed that the VIVIX-S 1751S (FXRD-1751SA) provide images of equivalent diagnostic capability to the predicate device, the VIVIX-S 1751S (FXRD-1751SB), and its results demonstrate substantial equivalence.

11. Conclusion [21 CFR 807.92(b) (3)]

The subject device is substantially equivalent to the currently marketed and predicate device (K190611) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

Additionally, Substantial equivalence was demonstrated through the non-clinical performance, which complied with the requirements specified in the international and FDA-recognized consensus standards, IEC60601-1, IEC 60601-1-2, 3 and the clinical test, which complied with the requirements specified in the CDRH's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.

The results of these tests demonstrate that VIVIX-S 1751S Digital X-ray detectors meet the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, non-clinical performance data, safety testing, and clinical image concurrence data demonstrates that the device is as safe, as effective, and performs as well the predicate devices.

{8}------------------------------------------------

12. FDA Guidance Documents Referenced

Technical documents for FXRD-1751SA comply with the following FDA guidance documents.

  • FDA Guidance on Submission of 510(k)s for Solid State X-ray Imaging Devices
  • FDA Guidance on Pediatric Information for X-ray Imaging Device Premarket Notifications
  • FDA Guidance on Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions
  • FDA Guidance on Off-The-Shelf Software Use in Medical Devices

N/A