(162 days)
The Sleep Apnea Notification Feature (SANF) is a software-only mobile medical application that analyzes Apple Watch sensor data to identify patterns of breathing disturbances suggestive of moderate-to-severe sleep apnea and provides a notification to the user. This feature is intended for over-the-counter (OTC) use by adults age 18 and over who have not previously received a sleep apnea diagnosis and is not intended to diagnose, treat, or aid in the management of sleep apnea. The absence of a notification is not intended to indicate the absence of sleep apnea.
The Sleep Apnea Notification Feature (SANF) is an over-the-counter mobile medical application (MMA) intended to identify patterns of breathing disturbances suggestive of moderate-to-severe sleep apnea and provide a notification to the user. SANF is intended to run on compatible iOS (e.g. iPhone, iPad) and Apple Watch platforms. Users set up SANF and view their health data on the iOS platform. Prior to use, users must undergo educational onboarding. SANF uses accelerometer sensor data collected by the Apple Watch to calculate breathing disturbance values while a user is asleep. Breathing disturbances describe transient changes in breathing patterns, such as temporary breathing interruptions.
Breathing disturbance data is analyzed in discrete, consecutive 30-day evaluation windows, If patterns consistent with moderate-to-severe sleep apnea are identified within the 30-day evaluation window, the user is notified. SANF provides visualizations depicting the user's breathing disturbance data over various time scales. SANF is not intended to provide instantaneous measurements. Instead, once activated, SANF runs opportunistically in the background receiving signals from Apple Watch sensors for processing.
Here's a summary of the acceptance criteria and study details for the Sleep Apnea Notification Feature (SANF), based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
| Metric | Acceptance Criteria (Stated Goal) | Reported Device Performance (95% CI) |
|---|---|---|
| Sensitivity | Optimized for high specificity given SANF is designed as an opportunistic detection feature. | 66.3% [62.2%, 70.3%] for moderate-to-severe sleep apnea (AHI ≥ 15) |
| Specificity | Optimized for high specificity given SANF is designed as an opportunistic detection feature. | 98.5% [98.0%, 99.0%] for normal-to-mild sleep apnea (AHI < 15) |
| False Positives | SANF did not falsely notify any subjects with normal AHI (AHI < 5). | 0% (implicitly, based on the statement above) |
| Breathing Disturbance Estimates (Proportion within pre-specified performance zone) | Not explicitly stated as a numerical acceptance criterion, but implicitly that it demonstrates effectiveness. | 91.4% (1,193 out of 1,305 subjects) |
Note: The document emphasizes that performance was "optimized for high specificity" given the opportunistic detection nature of the device. This implies that while a specific numerical sensitivity might not have been a hard "acceptance criterion" per se, the reported sensitivity alongside high specificity demonstrated sufficient effectiveness for clearance.
2. Sample Size Used for the Test Set and Data Provenance
-
Sample Size for Test Set (Clinical Study):
- Notification Performance Analysis: 1,278 subjects
- Breathing Disturbance Performance Analysis: 1,305 subjects
- Total Subjects Enrolled: 1,499 subjects (some had insufficient data for analysis)
-
Data Provenance:
- Country of Origin: United States (from "several sites across the United States").
- Retrospective or Prospective: Prospective. The study "enrolling 1,499 subjects" suggests a prospective collection of data specifically for this validation study.
3. Number of Experts Used to Establish the Ground Truth and Qualifications
The document refers to the "Nox T3s home sleep apnea testing (HSAT) device (K192469) as a reference device" for ground truth. The HSAT device itself provides the AHI (Apnea-Hypopnea Index) which is the clinical standard for sleep apnea diagnosis.
Thus, the ground truth was established by the HSAT device, not by human experts directly adjudicating each case. The output of the HSAT device is the ground truth measure (AHI).
4. Adjudication Method for the Test Set
The ground truth was established by the Nox T3s HSAT device, which is an objective measurement device. Therefore, a human expert adjudication method (like 2+1 or 3+1) was not explicitly mentioned or performed for the primary clinical endpoint, as the HSAT device is considered the reference standard. The AHI values derived from the HSAT device served as the diagnostic ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
There is no mention of an MRMC comparative effectiveness study involving human readers with or without AI assistance. The study focuses on the standalone performance of the SANF device against a reference standard (HSAT).
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)
Yes, a standalone performance study was done. The reported sensitivity and specificity values are for the algorithm's performance in identifying patterns suggestive of moderate-to-severe sleep apnea and providing a notification, without human intervention in the interpretation or decision-making process based on the device's output. The device itself "provides a notification to the user," implying direct algorithm output.
7. Type of Ground Truth Used
The ground truth used was objective diagnostic data derived from a medical device: The Nox T3s home sleep apnea testing (HSAT) device, which provides the Apnea-Hypopnea Index (AHI). This is considered a gold standard for diagnosing and classifying the severity of sleep apnea in a home setting.
8. Sample Size for the Training Set
The algorithm development dataset included over 11,000 nights of concurrent reference and watch sensor data.
9. How the Ground Truth for the Training Set Was Established
The ground truth for the training set was established using concurrent in-lab polysomnography (PSG) and Home Sleep Apnea Test (HSAT) reference recordings. These are the gold standard diagnostic tests for sleep apnea, providing objective measures like the Apnea-Hypopnea Index (AHI). The document also mentions that the distribution of sleep apnea classifications (normal, mild, moderate, severe) was broad in this dataset.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 13, 2024
Apple Inc. Lynda Ikejimba Principal Regulatory Affairs Associate One Apple Park Way Cupertino, California 95014
Re: K240929
Trade/Device Name: Sleep Apnea Notification Feature (SANF) Regulation Number: 21 CFR 868.2378 Regulation Name: Over-the-counter device to assess risk of sleep apnea Regulatory Class: Class II Product Code: QZW Dated: April 4, 2024 Received: April 4, 2024
Dear Lynda Ikejimba:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP) titled "SLEEP APNEA NOTIFICATION FEATURE (SANF) PREDETERMINED CHANGE
{1}------------------------------------------------
CONTROL PLAN" version 1.0. Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia. Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
Submission Number (if known)
K240929 Device Name
Sleep Apnea Notification Feature (SANF)
Indications for Use (Describe)
The Sleep Apnea Notification Feature (SANF) is a software-only mobile medical application that analyzes Apple Watch sensor data to identify patterns of breathing disturbances suggestive of moderate-to-severe sleep apnea and provides a notification to the user. This feature is intended for over-the-counter (OTC) use by adults age 18 and over who have not previously received a sleep apnea diagnosis and is not intended to diagnose, treat, or aid in the management of sleep apnea. The absence of a notification is not intended to indicate the absence of sleep apnea.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
This summary of 5 l 0(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
1. Submitter
| Applicant | Apple Inc.One Apple Park WayCupertino, CA 95014 |
|---|---|
| SubmissionCorrespondent | Lynda Ikejimba, PhDRegulatory AffairsPhone: (669) 227-8858Email: lc_ikejimba@apple.com |
| SecondaryCorrespondent | Kevin GoRegulatory AffairsPhone: (669) 225-1032Email: kevin_f_go@apple.com |
| Date Prepared | Sept 13, 2024 |
2. Device Names and Classifications
Subject Device:
| Name of Device | Sleep Apnea Notification Feature (SANF) |
|---|---|
| Classification Name | Over-the-counter device to assess risk of sleep apnea, 21 CFR868.2378 |
| Regulatory Class | Class II |
| Product Code | QZW |
| 510(k) Review Panel | Anesthesiology |
3. Predicate Device
| Predicate | Value |
|---|---|
| Manufacturer | Samsung Electronics Co., Ltd |
| Trade Name | Sleep Apnea Feature |
| 510(k) | DEN230041 |
{5}------------------------------------------------
4. Device Description
The Sleep Apnea Notification Feature (SANF) is an over-the-counter mobile medical application (MMA) intended to identify patterns of breathing disturbances suggestive of moderate-to-severe sleep apnea and provide a notification to the user. SANF is intended to run on compatible iOS (e.g. iPhone, iPad) and Apple Watch platforms. Users set up SANF and view their health data on the iOS platform. Prior to use, users must undergo educational onboarding. SANF uses accelerometer sensor data collected by the Apple Watch to calculate breathing disturbance values while a user is asleep. Breathing disturbances describe transient changes in breathing patterns, such as temporary breathing interruptions.
Breathing disturbance data is analyzed in discrete, consecutive 30-day evaluation windows, If patterns consistent with moderate-to-severe sleep apnea are identified within the 30-day evaluation window, the user is notified. SANF provides visualizations depicting the user's breathing disturbance data over various time scales. SANF is not intended to provide instantaneous measurements. Instead, once activated, SANF runs opportunistically in the background receiving signals from Apple Watch sensors for processing.
5. Indications for Use
The Sleep Apnea Notification Feature (SANF) is a software-only mobile medical application that analyzes Apple Watch sensor data to identify patterns of breathing disturbances suggestive of moderate-to-severe sleep apnea and provides a notification to the user. This feature is intended for over-the-counter (OTC) use by adults age 18 and over who have not previously received a sleep apnea diagnosis and is not intended to diagnose, treat, or aid in the management of sleep apnea. The absence of a notification is not intended to indicate the absence of sleep apnea.
6. Comparison with the Predicate Device
SANF and the predicate device (DEN230041) have the same intended use, technological characteristics, and principles of operation, and the difference in indications does not represent a new intended use. Both the subject and predicate devices are software-only mobile medical applications intended to detect signs of moderate-to-severe sleep apnea for individuals who have not been previously diagnosed with sleep apnea and are not intended to provide a standalone diagnosis.
The subject device contains some differences in technological characteristics:
- The subject device is compatible with Apple products (i.e., iOS device, Apple Watch), while ● the predicate is compatible with Samsung products (i.e., Galaxy Watch and Phone).
- . The subject device utilizes passive, opportunistic detection to monitor the user over a 30day period, and only alerts the user if it detects signs of sleep apnea. The predicate device provides an on-demand two-day assessment, and returns either a positive or negative finding to the user.
- . The subject device utilizes accelerometer sensor data while the predicate device utilizes blood oxygen sensor data.
The differences in technological characteristics described above do not raise new questions of safety or effectiveness. The differences can properly be evaluated through the special controls
{6}------------------------------------------------
established in 21 CFR 868.2378. The subject device has been appropriately verified and validated through non-clinical and clinical testing to ensure that the device is substantially equivalent to the predicate. A complete comparison of the subject and predicate device can be found in Table 1 below.
| Item | Subject DeviceSleep Apnea Notification Feature | Predicate Device(DEN230041) |
|---|---|---|
| Device Name | Sleep Apnea Notification Feature(SANF) | Sleep Apnea Feature |
| Manufacturer | Apple Inc. | Samsung Electronics Co., Ltd |
| Regulation Number | 21 CFR 868.2378 | 21 CFR 868.2378 |
| Product Code | QZW | QZW |
| Regulation Name | Over-the-counter device to assess risk of sleep apnea | Over-the-counter device to assess risk of sleep apnea. |
| DeviceClassification | Class II | Class II |
| OTC/Prescription | OTC | OTC |
| Intended Use | An over-the-counter device to assessrisk of sleep apnea intended to providea notification of the risk of sleep apneain users who have not been previouslydiagnosed with sleep apnea. Thisdevice uses software algorithms toanalyze input sensor signals andprovide a risk assessment for sleepapnea. It is not intended to provide astandalone diagnosis, replace traditionalmethods of diagnosis (e.g.,polysomnography), assist clinicians indiagnosing sleep disorders, or be usedas an apnea monitor. | An over-the-counter device to assessrisk of sleep apnea is intended toprovide a notification of the risk of sleepapnea in users who have not beenpreviously diagnosed with sleep apnea.This device uses software algorithms toanalyze input sensor signals andprovide a risk assessment for sleepapnea. It is not intended to provide astandalone diagnosis, replace traditionalmethods of diagnosis (e.g.,polysomnography), assist clinicians indiagnosing sleep disorders, or be usedas an apnea monitor. |
| Table 1 : SANF Comparison with the Predicate | ||
|---|---|---|
{7}------------------------------------------------
| Item | Subject DeviceSleep Apnea Notification Feature | Predicate Device(DEN230041) |
|---|---|---|
| Indications for Use | The Sleep Apnea Notification Feature(SANF) is a software-only mobilemedical application that analyzes AppleWatch sensor data to identify patternsof breathing disturbances suggestive ofmoderate-to-severe sleep apnea andprovides a notification to the user. Thisfeature is intended for over-the-counter(OTC) use by adults age 18 and overwho have not previously received asleep apnea diagnosis and is notintended to diagnose, treat, or aid in themanagement of sleep apnea. Theabsence of a notification is not intendedto indicate the absence of sleep apnea. | The Sleep Apnea Feature is an over-the-counter (OTC) software-only,mobile medical application operating ona compatible Samsung Galaxy Watchand Phone. This feature is intended todetect signs of moderate to severeobstructive sleep apnea in the form ofsignificant breathing disruptions in adultusers 22 years and older, over a two-night monitoring period. It is intendedfor on demand use. This feature is notintended for users who have previouslybeen diagnosed with sleep apnea.Users should not use this feature toreplace traditional methods of diagnosisand treatment by a qualified clinician.The data provided by this device is alsonot intended to assist clinicians indiagnosing sleep disorders. |
| Principle ofOperation | SANF uses software algorithms toanalyze input accelerometer sensorsignals and provide a risk assessmentfor sleep apnea. | The Sleep Apnea Feature uses softwarealgorithms to analyze input bloodoxygen sensor signals and provide arisk assessment for sleep apnea. |
| Overall DeviceDesign | A software-only device, and usessoftware algorithms to analyze inputsensor signals from a general purposecomputing platform and provide a riskassessment for sleep apnea.Assessments are based on sensor datacollected over 30-day periods. Thedevice is intended to provideopportunistic detection of sleep apnea,such that after initial enrollment no userinteraction is required for the device toperform as intended. | A software-only device, and usessoftware algorithms to analyze inputsensor signals from a general purposecomputing platform and provide a riskassessment for sleep apnea.Assessments are based on sensor datacollected over a 2-day period. Thedevice is intended to provide ondemand assessments to detect signs ofsleep apnea, such that a user mustactively choose to initiate a monitoringperiod. |
| Use Environment | Over-the-counter | Over-the-counter |
| Device Components | Software-only | Software-only |
| Device Input | Accelerometer data | Blood oxygen level (SpO2) data |
| Item | Subject DeviceSleep Apnea Notification Feature | Predicate Device(DEN230041) |
| ClinicalPerformance | The performance was optimized forhigh specificity given SANF is designedas an opportunistic detection feature(i.e., passive, recurring).Sensitivity: 66.3%95% CI [62.2%, 70.3%]Specificity: 98.5%95% CI [98.0%, 99.0%] | Sensitivity: 82.7%95% CI [76.7%, 87.6%]Specificity: 87.7%95% CI [83.1%, 91.4%] |
{8}------------------------------------------------
7. Summary of Non-Clinical Testing
Algorithm Development
SANF includes a deep learning algorithm to identify breathing disturbances using accelerometer sensor data from Apple Watch. The model was trained on Apple Watch accelerometer signals collected during sleep sessions with concurrent in-lab polysomnography (PSG) and Home Sleep Apnea Test (HSAT) reference recordings. The algorithm development dataset included over II,000 nights of concurrent reference and watch sensor data. The distribution of sleep apnea classifications in this dataset was broad and spanned all four clinically defined categories of sleep apnea: normal (AHI 0 to <5), mild (AHI 5 to <15), moderate (AHI 15 to <30), and severe (AHI >30). For the purposes of algorithm development, data from the studies was pooled and split into four sets: Training, Validation, Test, and Sequestration. The model was trained on the Training set, with the Validation set used for early stopping and threshold selection. The model was then evaluated on the Test set at regular intervals during model development was complete and the model was locked, it was evaluated on the Sequestration set as a last test to ensure it had not been over-fit to the training data. This process ensured no subject overlap and matching distributions of sex, age, BMI, and disease severity. The development data included a diverse group of subjects with respect to demographic factors (e.g., age, AHI, race, ethnicity, and BMI) representative of the intended use population.
Non-clinical Testing Summary
Apple conducted the necessary non-clinical testing on SANF with passing results supporting a determination of substantial equivalence. Non-clinical testing conducted included the following:
Software Verification and Validation
Software verification and validation was conducted in accordance with Apple's robust Quality Management System and documented to address the recommendations in FDA's 2023 Guidance, "Content of Premarket Submissions for Device Software Functions." SANF was determined to require a Basic Documentation Level. Apple's good software engineering practices, as demonstrated through the submission's documentation, supports a conclusion that SANF was appropriately designed, verified, and validated.
{9}------------------------------------------------
Cybersecurity
Apple approach to cybersecurity aligns with FDA's 2023 Guidance, " Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions." The device also conforms to the cybersecurity requirements identified in Section 524B to the FD&C Act.
Human Factors Validation
The Sleep Apnea Notification Feature was found to be safe and effective as compared to the predicate for the intended users, uses, and use environments. This conclusion is supported by iterative human factors analyses and evaluations on the device, resulting design modifications and the analysis of the summative validation testing results as recommended by 2016 FDA Guidance , "Applying Human Factors and Usability Engineering to Medical Devices".
General Purpose Computing Platform Assessment
SANF is a software-only device available on compatible general purpose computing platforms (e.g. Apple Watch); therefore, medical device hardware testing is not applicable. However, as a multiple function device product, the impact of the general purpose computing platform on SANF was assessed per FDA's 2020 Guidance, "Multiple Function Device Products: Policy and Considerations" and determined to be acceptable. This is consistent with the impact assessment of other Apple medical device features made available on Apple Watch, such as the Irregular Rhythm Notification Feature (K231173) and the Atrial Fibrillation History Feature (K213971).
8. Summary of Clinical Testing
The performance of the Sleep Apnea Notification Feature was validated in a prospective, nonsignificant risk study enrolling 1,499 subjects from several sites across the United States. The purpose of the study was to evaluate the performance of SANF using the Nox T3s home sleep apnea testing (HSAT) device (K192469) as a reference device. The study enrolled subjects across the spectrum of sleep apnea severity classifications, with a broad distribution across each of the following AHI categories using the "4%" hypopnea scoring rule: 559 normal subjects (AHI < 5), 362 mild subjects (5 ≤ AHI ≤ 15), 216 moderate subjects (15 ≤ AHI < 30), and 201 severe subjects (AHI ≥ 30), plus 161 subjects with missing HSAT reference. Subjects were also enrolled based across a broad range of demographic factors, including enrollment targets for age, sex, BMI, skin tone, race, and ethnicity subgroups to ensure the study population was representative of the intended user population. Study demographic characteristics are summarized in Table 2 below.
| N = 1,499 | |
|---|---|
| Age Group (years) | |
| 18-49 | 855 (57.0%) |
| 50-64 | 491 (32.8%) |
| ≥65 | 153 (10.2%) |
| Sex |
Table 2: SANF Clinical Study Subject Demographics
{10}------------------------------------------------
| Female | 847 (56.5%) |
|---|---|
| Male | 652 (43.5%) |
| Ethnicity | |
| Hispanic or Latino | 181 (12.1%) |
| Non-Hispanic or Latino | 1,318 (87.9%) |
{11}------------------------------------------------
| Race | |
|---|---|
| American Indian or AlaskaNative | 25 (1.7%) |
| Asian | 103 (6.9%) |
| Black or African American | 347 (23.1%) |
| Native Hawaiian or OtherPacific Islander | 3 (0.2%) |
| White | 1,021 (68.1%) |
Of the 1,499 enrolled subjects, 1,278 contributed to the notification performance analysis and 1,305 contributed to the breathing disturbance performance analysis. Those not included in the performance had insufficient Apple Watch data and/or reference data.
The sensitivity of notifications for subjects with moderate-to-severe sleep apnea (AHI ≥ 15) was 66.3%; 95% Cl [62.2%, 70.3%]. The specificity of the notifications for those with normal-to-mild sleep apnea (AHI
< 15) was 98.5%; 95% Cl [98.0%, 99.0%]. SANF did not falsely notify any subjects with normal AHI (AHI < 5). The performance was similar for identified sub-groups.
To assess performance of Breathing Disturbance estimates, Apple evaluated the proportion of paired (Breathing Disturbance, reference AHI). Of the total 1,305 subjects who had at least one paired measurement, 1,193 (91.4%) were within the pre-specified performance zone.
These results demonstrate that the Sleep Apnea Notification is effective in generating accurate notifications for moderate-to-severe sleep apnea and Breathing Disturbance values.
9. Predetermined Change Control Plan
The SANF contains a Predetermined Change Control Plan (PCCP), which complies with Section 3308 of the Food and Drug Omnibus Reform Act (FDORA) of 2022, enacted on December 29, 2022. The PCCP does not include provisions for implementation of adaptive algorithms that will continuously learn in the field. All algorithm modifications will be trained, and locked prior to release of the software to the field. A procedure has also been established for updating the Instructions for Use in order to inform users about algorithm changes implemented under this FDAauthorized PCCP, including a summary of the changes, a characterization of algorithm performance, and the availability and compatibility of the feature. Apple will publish updated Instructions for Use on its website and make them accessible within the Health App.
The PCCP specifies possible modifications to the device software as well as verification and validation activities in place to implement the changes in a controlled manner such that the modified device remains as safe and effective as the predicate device. The PCCP includes a specific list of potential software modifications defining the region of potential changes that can be made to the algorithms in the device. Details of the potential changes are summarized in Table
{12}------------------------------------------------
3 below. The modification protocol incorporates impact assessment considerations and specifies requirements for data management, including data sources, collection, storage, and sequestration, as well as documentation and data re-use practices. Specific test methods are specified in the PCCP to establish substantial equivalence the Sleep Apnea Notification Feature and include sample size determination, analysis methods, and acceptance criteria. To help ensure validation test datasets are representative of the intended use population, each will meet minimum demographic requirements for age, sex, race, BMI and ethnicity.
| Detailed List of Changes | Requirements | Test Method | |
|---|---|---|---|
| Modifications toBreathingDisturbances (BD)computation | • Adjust the operating point• Re-train algorithm withadditional datasets whilemaintaining the same algorithmarchitecture and number ofparameters.• Revise signal input module• Add additional classifier outputs• Modifications to signal qualityand post-processing modules | • No change to input type• No change to output type• No concurrent change toother modules• No change to intended useof device• Can be fully verified and/orvalidated by requirements ofthe modification protocol | Verification of BDaccuracy andsubstantialequivalence innotification-levelsensitivity andspecificity whencompared to theperformance of SANF1.0 |
| Modifications to sleepapnea estimation | • Modify the number of nightly BDreadings required to surface anotification• Reduce the interval of thenotification window• Modify logic for surfacing anotification based on BDs | Substantialequivalence insensitivity andspecificity whencompared to theperformance of SANF1.0 |
Table 3: Proposed modifications to the SANF under the PCCP
10. Conclusion
The Sleep Apnea Notification Feature is substantially equivalent to the predicate device as they are identical with respect to intended use and there are no differences in technological or performance characteristics that raise different questions of safety and effectiveness.
N/A