K Number
K240926

Validate with FDA (Live)

Device Name
PeekMed web
Manufacturer
Date Cleared
2024-12-06

(246 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the planning.

The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.

This medical device consists of a decision support tool for qualified healthcare professionals to quickly and efficiently perform the pre-operative planning for several surgical procedures, using medical imaging with the additional capability of planning the 2D or 3D environment. The system is designed for the medical specialties within surgery and no specific use environment is mandatory, whereas the typical use environment is a room with a computer. The patient target group is adult patients who have an injury or disability diagnosed previously. There are no other considerations for the intended patient population.

Device Description

PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning.

The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.

As PeekMed web is capable of representing medical images in a 2D or 3D environment, performing relevant measurements on those images, and also capable of adding templates, it then can perform a total overview of the surgery. Being software it does not interact with any part of the body of the user and/or patient.

AI/ML Overview

The provided document describes the PeekMed web device and its performance testing to demonstrate substantial equivalence to a predicate device. Here's a breakdown of the acceptance criteria and the study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document explicitly states the acceptance criteria for each ML model but does not provide the reported device performance values against these criteria. It only states that the device met the acceptance criteria.

ML modelAcceptance CriteriaReported Device Performance
SegmentationDICE is no less than 90%HD-95 is no more than 8STD DICE is between +/- 10%Precision is more than 85%Recall is more than 90%Met acceptance criteria
LandmarkingMRE is no more than 7mmSTD MRE is between +/- 5mmMet acceptance criteria
ClassificationAccuracy is no less than 90%Precision is no less than 85%Recall is no less than 90%F1 score is no less than 90%Met acceptance criteria

2. Sample Sizes and Data Provenance

  • Test Set (External Validation Dataset):

    • Segmentation ML model: 367 unique datasets
    • Landmarking ML model: 367 unique datasets
    • Classification ML model: 367 unique datasets
  • Training and Development Datasets:

    • Total CR datasets: 2852
    • Total CT scans: 1903
    • Training: 80% of the combined CR/CT datasets
    • Development: 10% of the combined CR/CT datasets
    • Testing (internal): The remaining 10%
  • Data Provenance: The document states that ML models were developed with datasets from multiple sites. It does not specify the country of origin of the data. The data used for external validation was collected independently from the development data to prevent bias, implying a prospective collection or at least a carefully curated retrospective collection to ensure independence. It states the external validation dataset was "not a sampling of the development dataset" and "leakage between development and validation data sets did not occur."

3. Number of Experts and Qualifications

The document states that the external validation dataset was "labeled by a separate team" to establish ground truth. However, it does not specify the number of experts used or their qualifications (e.g., radiologist with 10 years of experience).

4. Adjudication Method

The document mentions that the external validation dataset was "labeled by a separate team," implying that this team established the ground truth. However, it does not describe the specific adjudication method used (e.g., 2+1, 3+1, none).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The study focuses on the standalone performance of the ML models and their equivalency to a predicate device, not on the improvement of human readers with AI assistance.

6. Standalone (Algorithm Only) Performance

Yes, the performance evaluation described focuses on the standalone performance of the ML models (segmentation, landmarking, classification) as incorporated into the PeekMed web system. The results against the predefined acceptance criteria evaluate the algorithm's performance without specific human-in-the-loop metrics, although the overall device is intended to assist human healthcare professionals.

7. Type of Ground Truth Used

The ground truth for the test set (external validation datasets) was established by predefined ground truth. The term "predefined ground truth" and "labeled by a separate team" suggests expert consensus or a similar meticulous labeling process, but it does not specify whether it was pathology, outcomes data, or directly observed clinical outcomes. Given the nature of medical imaging software, it is highly probable to be expert consensus.

8. Sample Size for the Training Set

The training set comprised 80% of a total of 2852 CR datasets and 1903 CT scans.
Total datasets = 2852 (CR) + 1903 (CT) = 4755 datasets.
Training set size = 80% of 4755 = 3804 datasets.

9. How the Ground Truth for the Training Set was Established

The document states that the ML models were developed with datasets from multiple sites. It implies that these datasets, used for training, development, and internal testing, also had established ground truth necessary for supervised learning. However, the document does not explicitly describe the process for establishing the ground truth specifically for the training set. It only states that the external validation data was "labeled by a separate team" for independence, suggesting a similar process might have been used for the training data but this is not confirmed.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Peek Health, S.A. Sara Silva Chief Quality Officer (CQO) Centro de Negócios Ideia Atlântico, Rua Padres Carmelitas Braga, 4719-005 Portugal

Re: K240926

December 6, 2024

Trade/Device Name: PeekMed web Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ, QIH Dated: November 1, 2024 Received: November 1, 2024

Dear Sara Silva:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pm.cfm identifies combination product submissions. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/unique-deviceidentification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-how-report-medicaldevice-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuingeducation/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240926

Device Name PeekMed web

Indications for Use (Describe)

PeekMed web is a system designed to help healthcare professionals carvy out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the planning.

The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the text 'K240926' at the top, followed by a logo in the center. The logo is a stylized, geometric shape in blue, resembling a stylized eye. Below the logo, the text 'PeekMed' is written in a simple, sans-serif font, also in blue. The logo and text likely represent the brand or company named PeekMed.

510(k) summary / K240926

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.

1. Submitter

Peek Health, S.A. Centro de Negócios Ideia Atlântico Rua Padres Carmelitas 4719-005 Braga Portugal

Contact Person:Sara Silva
Chief Quality Officer (CQO)
Email: sara.silva@peekmed.com
Office number: + 351 253 128 941

Date Summary Prepared: December 5, 2024

2. Device

2.1 PeekMed web

Trade Name:PeekMed web
Common or Usual Name:Medical image management and processing system
Classification Name:System, Image Processing, Radiological(21 C.F.R. § 892.2050)
Regulatory Class:Class II
Product Code:LLZ/QIH

3. Legally Marketed Predicate Device

3.1 PeekMed web

510(k)Product NameClearance Date
K222767PeekMed webDecember 2022

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image features the logo for PeekMed, a medical technology company. The logo consists of a stylized, three-dimensional shape in a teal color, resembling a cube with a section cut out to form a stylized eye. Below the graphic is the company name, "PeekMed," also in teal, presented in a clean, sans-serif typeface.

4. Device Description Summary

PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning.

The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.

As PeekMed web is capable of representing medical images in a 2D or 3D environment, performing relevant measurements on those images, and also capable of adding templates, it then can perform a total overview of the surgery. Being software it does not interact with any part of the body of the user and/or patient.

5. Intended Use/Indications for Use

PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surqical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment is necessary for the proper use of the system in the revision and approval of the output of the planning.

The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.

5.1. Contraindications

No contraindications specific to this device.

5.2 Indications for Use Comparison

There are NO differences between the indications for use of this device and its predicate.

6. Technological Comparison to Predicate

PeekMed was compared to its respective predicate device in intended use, indications for use, design, function, and technology and it was demonstrated that they are substantially equivalent. Any

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for PeekMed. The logo features a stylized, three-dimensional cube shape in a teal color. Within the cube, there is a white outline of an eye. Below the cube, the word "PeekMed" is written in a simple, sans-serif font, also in teal.

technological differences within this 510(k), between the subject device and the predicate device, do not impact substantial equivalence, or safety and effectiveness.

The subject device and its predicate are both medical software that allows healthcare professionals to perform orthopedic pre-surgical planning efficiently in the musculoskeletal system of adults in a healthcare environment, therefore sharing the same intended use, and intended patient population. To properly and fully use both devices, clinical judgment, and experience are mandatory.

Both devices have the same workflows, use requirements (e.g., internet connection, output validation), and planning features (e.g., model representation, digital overlap of prosthetic material, possible 2D and 3D environments). Both devices generate a final report of the planning which consists of the selected images with templates, measurements, and textual information describing the patient and/or the surgical procedure to be performed.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for PeekMed. The logo is a stylized, geometric shape in blue, resembling a cube or a stylized letter 'P'. Within the shape, there is a white eye symbol, suggesting a focus on vision or medical imaging. Below the shape, the word "PeekMed" is written in a simple, sans-serif font, also in blue.

Table 1: Summary of Predicate and Subject Device Characteristics to Demonstrate Substantial Equivalence

CharacteristicPeekMed webPredicate device K222767PeekMed webSubject deviceSubstantiallyEquivalent?Justification and rationale
Product CodeLLZ, QIHLLZ, QIHYes--
Regulation Number21 CFR 892.205021 CFR 892.2050Yes--
Regulation NameMedical Image Management AndProcessing SystemMedical Image Management AndProcessing SystemYes--
Intended usePeekMed web is a systemdesigned to help healthcareprofessionals carry outpre-operative planning for severalsurgical procedures, based ontheir imported patients' imagingstudies. Experience in usage anda clinical assessment is necessaryfor the proper use of the system inthe revision and approval of theoutput of the planning. Themulti-platform system works with adatabase of digital representationsrelated to surgical materialssupplied by their manufacturers.PeekMed web is a systemdesigned to help healthcareprofessionals carry outpre-operative planning for severalsurgical procedures, based ontheir imported patients' imagingstudies. Experience in usage anda clinical assessment is necessaryfor the proper use of the system inthe revision and approval of theoutput of the planning. Themulti-platform system works with adatabase of digital representationsrelated to surgical materialssupplied by their manufacturers.Yes--
Indications forThis medical device consists of aThis medical device consists of aYes--
CharacteristicPeekMed webPredicate device K222767PeekMed webSubject deviceSubstantiallyEquivalent?Justification and rationale
usedecision support tool for qualifiedhealthcare professionals to quicklyand efficiently perform thepre-operative planning for severalsurgical procedures, using medicalimaging with the additionalcapability of planning the 2D or 3Denvironment. The system isdesigned for the medicalspecialties within surgery and nospecific use environment ismandatory, whereas the typicaluse environment is a room with acomputer. The patient target groupis adult patients who have aninjury or disability diagnosedpreviously. There are no otherconsiderations for the intendedpatient population.decision support tool for qualifiedhealthcare professionals to quicklyand efficiently perform thepre-operative planning for severalsurgical procedures, using medicalimaging with the additionalcapability of planning the 2D or 3Denvironment. The system isdesigned for the medicalspecialties within surgery and nospecific use environment ismandatory, whereas the typicaluse environment is a room with acomputer. The patient target groupis adult patients who have aninjury or disability diagnosedpreviously. There are no otherconsiderations for the intendedpatient population.
ContraindicationsNo contraindications specific tothis device.No contraindications specific tothis device.Yes---
Clinical purposePeekMed web allows the surgeonto perform orthopedic pre-surgicalplanning efficiently in thePeekMed web allows the surgeonto perform orthopedic pre-surgicalplanning efficiently in theYes---
CharacteristicPeekMed webPredicate device K222767PeekMed webSubject deviceSubstantiallyEquivalent?Justification and rationale
musculoskeletal system (e.g., Hipprocedures, Knee procedures)musculoskeletal system (e.g., Hipprocedures, Knee procedures)
AnatomicalregionsPeekMed web allows the surgeonto perform the pre-surgicalplanning efficiently in the followinganatomical regions:- Hip- Knee- Upper limbPeekMed web allows the surgeonto perform the pre-surgicalplanning efficiently in the followinganatomical regions:- Hip- Knee- Upper limb- FootYesThe subjectdevice alsoallows theplanning oforthopedicsurgery in thefoot region.The subject device allows for the surgeryplanning on one more anatomical region(foot), this does not constitute a newintended purpose nor does it raisequestions of safety and performance,since the development, verification,validation, and release processes areexactly the same between the devices.The new variants of the ML model for footare considered in this Traditional 510(k)and were developed according to what isdefined for each ML model developmentand validation according to their intendedrequirements and performance.This change improves the medical devicein a way that the user can benefit from theexisting features on one more anatomicalregion. This change does not interferewith or create an effect on otheranatomical regions covered by the device.
Patient Population AdultsAdultsYes---
CharacteristicPeekMed webPredicate device K222767PeekMed webSubject deviceSubstantiallyEquivalent?Justification and rationale
End usersHealthcare ProfessionalsHealthcare ProfessionalsYes---
Device availabilitySoftware is cloud-based (notinstallable) and can be displayedon any personal device orworkstation that can run on a webbrowserSoftware is cloud-based (notinstallable) and can be displayedon any personal device orworkstation that can run on a webbrowserYes---
SoftwareArchitectureDistributed system (cloud-based).This distributed system is acombination of software modulesplaced on servers that are able tocommunicate with each other.Distributed system (cloud-based).This distributed system is acombination of software modulesplaced on servers that are able tocommunicate with each other.Yes---
WorkflowThe workflow is as follows: Importcase images, configure images,identify the case, pre-surgicalplanning, and export the case. Theworkflow chart is in Annex 2.The workflow is as follows: Importcase images, configure images,identify the case, pre-surgicalplanning, and export the case. Theworkflow chart is in Annex 2.Yes---
InternetconnectionRequiredRequiredYes---
Images sourceReceives medical images fromvarious sourcesReceives medical images fromvarious sourcesYes---
Data processingThe software processes data toprovide an overlap andThe software processes data toprovide an overlap andYes---
CharacteristicPeekMed webPredicate device K222767PeekMed webSubject deviceSubstantiallyEquivalent?Justification and rationale
dimensioning of digitalrepresentations of the prostheticmaterialdimensioning of digitalrepresentations of the prostheticmaterial
Digital overlap oftemplatesAllows the overlap of models andthe intersection of the modelsAllows the overlap of models andthe intersection of the modelsYes---
Interactive modelpositioningYesYesYes---
Interactive modeldimensioningYesYesYes---
Model rotationYesYesYes---
Support for digitalprostheticmaterialsprovided by themanufacturersYesYesYes---
Pre-surgicalplanningYesYesYes---
Type ofpre-surgicalplanningAutomatic or ManualAutomatic or ManualYes---
Contact with thepatientNoNoYes---
CharacteristicPeekMed webPredicate device K222767PeekMed webSubject deviceSubstantiallyEquivalent?Justification and rationale
Control of lifesupportingdevicesNoNoYes ----
Humanintervention forimageinterpretationYesYesYes ----
Ability to addadditionalmodules whenavailableYesYesYes ----
Automatic bonesegmentationYesYesYesThe updatedML modelvariants werecomparedagainst theacceptancecriteria and,thereforeconfirmed tobe animprovementThe subject device includes new andupdated ML model variants for thepurpose of bone segmentation. These MLmodel variants were developed andvalidated using the same methods andacceptance criteria and deemed asCompliant. The updated model variantswere also compared against theacceptance criteria, therefore confirmed tobe an improvement over the last MLmodel variants versions.
CharacteristicPeekMed webPredicate device K222767PeekMed webSubject deviceSubstantiallyEquivalent?Justification and rationale
over the lastML modelvariantsversions.
ConfigurationWindows forOblique PlanesfunctionalityYesYes---
Type oflandmarkingAutomatic or ManualAutomatic or ManualYesThe updatedML modelvariants werecomparedagainst theacceptancecriteria and,thereforeconfirmed tobe animprovementover the lastML modelvariantsversions.The subject device includes new andupdated ML model variants for thepurpose of landmarking. These ML modelvariants were developed and validatedusing the same methods and acceptancecriteria and deemed as Compliant. Theupdated model variants were alsocompared against the acceptance criteria,therefore confirmed to be an improvementover the last ML model variants versions.

{7}------------------------------------------------

Peek Health, S.A. 510(k) Summary Traditional 510(k) Premarket Notification PeekMed web

Image /page/7/Picture/1 description: The image shows the logo for PeekMed. The logo is a blue, stylized cube with a white eye in the center. Below the cube is the text "PeekMed" in a simple, sans-serif font. The logo is clean and modern, suggesting a focus on technology and vision.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for PeekMed. The logo is a blue, stylized cube with a white eye in the center. Below the cube is the text "PeekMed" in a sans-serif font. The logo is simple and modern, and the eye in the cube suggests that the company is involved in vision or medical technology.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for PeekMed. The logo is a blue, stylized cube with a white eye in the center. The word "PeekMed" is written in blue below the cube. The logo is simple and modern.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image contains the logo for PeekMed. The logo features a stylized, teal-colored geometric shape resembling a cube with a cutout in the form of the letter 'P'. Within the cutout, there is an eye symbol. Below the geometric shape, the word "PeekMed" is written in a simple, sans-serif font, also in teal.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for PeekMed. The logo is a blue, stylized eye inside of a geometric shape. The text "PeekMed" is written in a sans-serif font below the logo.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the logo for PeekMed. The logo is a stylized blue eye inside of a geometric shape. The text "PeekMed" is written in a sans-serif font below the logo.

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image shows the logo for PeekMed. The logo features a stylized, teal-colored geometric shape resembling a cube with a cutout in the form of a 'P'. Within the cutout, there is an eye symbol. Below the geometric shape, the word "PeekMed" is written in a simple, sans-serif font, also in teal.

PeekMed web shares the same intended use, indications for use, end has patient population, and overall technical and functional capabilities, and therefore is substant to the predicate device. The subject device has the same design and function as the predicate device for the modes of operation and use. The subject device includes newimproved ML models and variants.

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the logo for PeekMed. The logo features a stylized, teal-colored geometric shape resembling a cube with a cutout in the center. Within the cutout, there is a white eye symbol. Below the geometric shape, the text "PeekMed" is written in a simple, sans-serif font, also in teal.

7. Performance Data

Nonclinical performance testing performed on the subject device, PeekMed web, supports substantial equivalence to the predicate device. The following testing was performed on the subject device:

  • A. Verification activities to ensure that features were implemented following the requirements and covering the acceptance criteria.
  • B. ML models incorporated into PeekMed web were also developed, trained, tested, and externally validated for their performance according to the internal procedures.
    • o A dedicated validation dataset containing different data from the ML development dataset was used. Specifically, the validation dataset was not a sampling of the development dataset, has never been used for the algorithm training or for tunning the algorithm, and leakage between development and validation data sets did not occur.

Training, development, testing and external validation data information

ML models were developed with datasets from multiple sites in a total of 2852 CR datasets, and 1903 CT scans. We trained the ML models with 80% of the datasets, developed with 10%, and tested with the remaining 10%. External validation is performed by sample size with a total unique dataset, for segmentation ML model: 367; Landmarking ML model: 367; and Classification ML model: 367. This comprehensive dataset was designed to cover the intended use population while ensuring a variety of data maintaining diverse patient characteristics.

Subgroup definition (generalizability)

Datasets were divided according to the subgroups listed below:

  • Demographics .
    • o Patient Sex
    • o Patient Age
  • Equipment and Protocols for Image Collection .
    • Institution Name O
    • Manufacturer O
    • Manufacturer Model Name O

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the logo for PeekMed. The logo features a stylized eye within a geometric shape, rendered in a teal color. Below the graphic is the word "PeekMed" in a simple, sans-serif font, also in teal.

Acceptance criteria

The acceptance criteria for each ML model are shown in the following table:

ML modelAcceptance Criteria
SegmentationDICE is no less than 90%HD-95 is no more than 8STD DICE is between +/- 10%Precision is more than 85%Recall is more than 90%
LandmarkingMRE is no more than 7mmSTD MRE is between +/- 5mm
ClassificationAccuracy is no less than 90%.Precision is no less than 85%Recall is no less than 90%F1 score is no less than 90%

Reference Standard ("Truthing" Process)

Comparison of the efficacy results of the Bone Segmentation ML model using the testing and external validation datasets against the predefined ground truth met the acceptance criteria for ML model performance, demonstrating the substantial equivalence of the subject device to its predicate.

Comparison of the efficacy results of the Landmarking ML model using the testing and external validation datasets against the predefined ground truth met the acceptance criteria for ML model performance, demonstrating the substantial equivalence of the subject device to its predicate.

Comparison of the efficacy results of the Classification ML model using the testing and external validation datasets against the predefined ground truth met the acceptance criteria for ML model performance, demonstrating the substantial equivalence of the subject device to its predicate.

{16}------------------------------------------------

Image /page/16/Picture/1 description: The image shows the logo for PeekMed. The logo features a stylized eye within a geometric shape, rendered in a teal color. Below the graphic is the word "PeekMed" in a simple, sans-serif font, also in teal.

Independence of Training and External Validation Data

External validation datasets were collected independently of the development data to prevent bias, ensuring the reliability of the results. For the external validation, a fully independent dataset, labeled by a separate team, was employed to provide an accurate assessment of the model's performance, over the whole population and for each sub-group mentioned before in order to prove that it generalizes well to unseen, real-world data. All the testing and external validation performed indicate acceptable performances of the ML models for its intended population.

  • C. Validation tests were performed internally before the release to the market by qualified personnel, in an environment simulating the real end-user environment. This validation follows a pre-defined test script document, according to the tests defined in the TestRail.
    Nonclinical performance testing allowed us to understand that there are no related problems in the subject device. Furthermore, these tests will be repeated and updated when appropriate to ensure that the software is always properly validated, making it possible to understand in which version the problems arise and in which they are solved. Consequently, any problem that may appear in a given PeekMed web version will be identified and can be solved in subsequent versions, as all steps are traceable. All anatomical areas were tested, as well as other main areas of the software, such as the planning final report, and saved planning, ML models, among others.

After these successful validation tests, it is possible to deem the subject device, PeekMed web, as substantially equivalent to its predicate device.

9. Conclusion

Based on the information provided in this 510(k) submission, it was determined that the subject device, PeekMed web, is substantially equivalent to the legally marketed predicate device concerning indications for use, intended use, design, technology, and performance.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).