K Number
K240882
Date Cleared
2024-05-01

(30 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

Device Description

BruxZir® Esthetic blanks are zirconia milling blanks used for the production of highly esthetic zirconia dental restorations. The high esthetics and high strength of BruxZir® Esthetic restorations make them viable for use in all regions of the mouth. The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. BruxZir® Esthetic blanks are available in white, bleach, and pre-shaded varieties that can be easily adjusted to match the final shade.

AI/ML Overview

The provided text describes the submission for a dental device, BruxZir® Esthetic (K240882), and compares it to a predicate device (K213425). The focus of the submission is to demonstrate substantial equivalence through non-clinical testing. This is not a submission for an AI/ML powered device, and therefore, many of the requested elements related to AI/ML specific studies (e.g., MRMC studies, standalone AI performance, training set details, ground truth establishment for AI models) are not applicable.

However, I can extract the relevant information regarding acceptance criteria and the study that proves the device meets those criteria, specifically concerning the physical and chemical properties of the dental material.

Here's a breakdown of the requested information based on the provided text, while noting the inapplicability of AI/ML specific questions:


Device: BruxZir® Esthetic (K240882)
Device Type: Zirconia milling blanks for dental restorations. This is a material, not an AI/ML powered device.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Flexural Strength>800 MPa (per ISO 6872:2015/Amd 1:2018, Type II Class 5)Met the criteria (>800 MPa)
Visual Shade EvaluationGrade of "Pass" for Visual Shade Match for all shades.Met the criteria
Color ConsistencyΔE calculation needs to be less than 2.00 between samples measured.Met the criteria
TranslucencyWithin 1% of the equivalent BruxZir Anterior-16 shade at 700nm.Met the criteria
Coefficient of Thermal Expansion (CTE)Between 10-11 x 10^-6 /°C, and standard deviation of CTE no greater than 0.5 x 10^-6 /°C (per ISO 6872:2015/Amd 1:2018).Met the criteria (Reported range: 10-11 x 10^-6 /°C; implies standard deviation also met based on compliance statement)
Chemical Solubility<100 µg/cm² (per ISO 6872:2015/Amd 1:2018)Met the criteria (<100 µg/cm²)
Radioactivity (Uranium-238)Not exceed 1.0 Bq/g (per ISO 6872:2015/Amd 1:2018)Met the criteria (No more than 1.0 Bq/g)
BiocompatibilityBiocompatible, safe, and non-toxic (per ISO 10993-1:2018)Met the criteria (Biocompatible and non-toxic)

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes for each non-clinical test. It states that "the test samples were made using the standard manufacturing process and QC process for BruxZir® Esthetic."

  • Data Provenance: The tests were conducted by Prismatik Dentalcraft, Inc. directly, as part of their design verification "according to BruxZir® Esthetic Design Verification Protocol." This indicates the data is prospective, generated specifically for validating the device's properties based on established industry standards (ISO). The country of origin is not explicitly stated for the testing location, but the applicant (Prismatik Dentalcraft, Inc.) is based in Irvine, California, USA.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This question is not applicable as the studies conducted are non-clinical, laboratory-based physical and chemical property tests, not clinical evaluations requiring expert interpretation of medical images or patient outcomes. The "ground truth" for these tests is the quantitative measurement against an established ISO standard.

4. Adjudication Method for the Test Set

Not applicable. The tests are quantitative measurements against predefined ISO standards. There is no human subjective adjudication process for these material property tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. This is not an AI-powered device, nor is it a diagnostic imaging device that would typically involve human readers. The submission focuses on the material properties of dental zirconia.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

No. This is not an algorithm or software device.

7. The Type of Ground Truth Used

The ground truth for these non-clinical tests is established by industry-standardized test methods (e.g., ISO 6872:2015/Amd 1:2018, ISO 10993-1:2018). The "ground truth" is adherence to the specified quantitative values and qualitative interpretations (e.g., biocompatible).

8. The Sample Size for the Training Set

Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

May 1, 2024

Prismatik Dentalcraft, Inc. So Hyun Park Regulatory Affairs Manager 2144 Michelson Drive Irvine, California 92612

Re: K240882

Trade/Device Name: BruxZir® Esthetic Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: March 29, 2024 Received: April 1, 2024

Dear So Hyun Park:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bobak Shirmohammadi -S

For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240882

Device Name

BruxZir® Esthetic

Indications for Use (Describe)

The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

Type of Use (Select one or both, as applicable)

< Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a black border. The inside of the triangle is filled with a gradient of colors, including pink, red, orange, yellow, green, blue, and purple. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC." in a blue sans-serif font.

K240882 - 510(k) Summary

Date Prepared: March 29, 2024

CONTACT DETAILS (21 CFR 807.92(a)(1))

Applicant Name: Prismatik Dentalcraft, Inc.

Applicant Address: 2144 Michelson Drive, Irvine, CA 92612, USA

Applicant Contact Telephone: 949-863-5479

Applicant Contact: So Hyun Park, Regulatory Affairs Manager, MS

Applicant Contact Email: so.park(@glidewelldental.com

DEVICE NAME (21 CFR 807.92(a)(2))

Device Trade Name: BruxZir® Esthetic Common Name: Zirconia milling block or dental CAD/CAM block Classification Name: Porcelain powder for clinical use Regulation Number: 872.6660 Regulatory Class: Class II Product Code: EIH

LEGALLY MARKETED PREDICATE DEVICE (21 CFR 807.92(a)(3))

Predicate 510(k) #: K231425

Predicate Trade Name: BruxZir® Esthetic

Product Code: EIH

DEVICE DESCRIPTION SUMMARY (21 CFR 807.92(a)(4))

BruxZir® Esthetic blanks are zirconia milling blanks used for the production of highly esthetic zirconia dental restorations. The high esthetics and high strength of BruxZir® Esthetic restorations make them viable for use in all regions of the mouth. The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. BruxZir® Esthetic blanks are available in white, bleach, and pre-shaded varieties that can be easily adjusted to match the final shade.

INTENDED USE/INDICATIONS FOR USE (21 CFR 807.92(a)(5))

The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a triangle on the left side that is divided into sections of different colors, including pink, red, yellow, green, blue, and purple. To the right of the triangle is the company name, "PRISMATIK DENTALCRAFT, INC.", in a sans-serif font.

INDICATIONS FOR USE COMPARISON (21 CFR 807.92(a)(5))

The subject device, BruxZir® Esthetic, has the same indications for use as the predicate device, BruxZir® Esthetic (K213425). Both devices are indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

TECHNOLOGICAL COMPARISON (21 CFR 807.92(a)(6))

The subject device, BruxZir® Esthetic, is substantially equivalent in technical characteristics to the predicate device, BruxZir® Esthetic (K213425) in terms of overall design principles, material, and performance.

The subject device, BruxZir® Esthetic, is similar in terms of overall design to the predicate device, BruxZir® Esthetic (K213425). The subject device, BruxZir® Esthetic, is offered in a milling blank and GL Single Unit design to be used for milling single-unit restorations. The predicate device, BruxZir® Esthetic (K213425), is offered in the same designs with the addition of the GL Bridge Block, which is designed to be used for milling bridges (restorations of three or more units). The fundamental principle of operation of the subject device, BruxZir® Esthetic, and the predicate device, BruxZir® Esthetic (K213425), is the same. The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. Although no color needs to be added additionally, pre-coloring before sintering and/or basic staining and glazing techniques may be used after sintering to achieve the desired shade. Also, the change in the manufacturing process from the press process to the colloidal process (slip casting) does not affect the mechanical properties of the final finished device.

The subject device, BruxZir® Esthetic, is similar in terms of material composition to the predicate device, BruxZir® Esthetic (K213425). The main ceramic component is composed of yttriastabilized zirconia with varying trace amount of colorants. As the generic type of material has not changed, it does not alter the performance specifications of the subject device. The slight differences in overall chemical composition do not affect the biocompatibility and performance characteristics.

The technological comparison table below outlines and provides the similarities between the subject device, BruxZir® Esthetic, and the predicate device, BruxZir® Esthetic (K213425). Both the subject device and the predicate device have similar physical/mechanical and biocompatibility properties that met the requirements of ISO 6875:2015/Amd 1:2018 (Type II, Class 5) and ISO 10993. Any differences between the subject device and the predicate device do not raise any new concerns of safety and effectiveness.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

AttributesSubject Device(K240882)Predicate Device(K213425)Comparison
Design CharacteristicsDevice NameBruxZir® EstheticBruxZir® EstheticSame
ManufacturerPrismatik Dentalcraft, Inc.Prismatik Dentalcraft, Inc.Same
Product CodeEIHEIHSame
Prescription DeviceYesYesSame
Indications for UseThe device is indicatedfor use by dentaltechnicians in theconstruction of custommade all ceramicrestorations for anteriorand posterior location.The device is indicatedfor use by dentaltechnicians in theconstruction of custommade all ceramicrestorations for anteriorand posterior location.Same
Chemical CompositionThe main ceramiccomponent iscomposed of yttria-stabilized zirconia. Thedifferent shades havevarying trace colorantsas required to matchthe desired shades.The main ceramiccomponent iscomposed of yttria-stabilized zirconia. Thedifferent shades havevarying trace colorantsas required to matchthe desired shades.Similar; Differentyttria proportion ofthe main ceramiccomponents anddifferent oxidecolorants for thesubject device.
DesignVarious size andthicknessesVarious size andthicknessesSimilar; Differentsizes andthicknesses offeredfor the subjectdevice.
Principle of OperationThe manufactureddental restorations aremade utilizing aCAD/CAM system fordesign andmanufacture. Thedesigned andmanufactured dentalrestorations are thensintered at a hightemperature. No colorneeds to be addedbefore sintering. Thefinal target shades areachievable withoutadditional coloringsteps.The manufactureddental restorations aremade utilizing aCAD/CAM system fordesign andmanufacture. Thedesigned andmanufactured dentalrestorations are thensintered at a hightemperature. No colorneeds to be addedbefore sintering. Thefinal target shades areachievable withoutadditional coloringsteps.Same
AttributesSubject Device(K240882)Predicate Device(K213425)Comparison
ShadeA1, A2, A3, A3.5, A4,B1, B2, B3, B4, C1,C2, C3, C4, D2, D3,D4, White, BleachA1, A2, A3, A3.5, A4,B1, B2, B3, B4, C1,C2, C3, C4, D2, D3,D4, White, BL1, BL3Similar; Fewerbleach shades areoffered for thesubject device.
Final VITAShadesA1, A2, A3, A3.5, A4,B1, B2, B3, B4, C1,C2, C3, C4, D2, D3,D4, White, BleachA1, A2, A3, A3.5, A4,B1, B2, B3, B4, C1,C2, C3, C4, D2, D3,D4, White, BL1, BL3Similar; Fewerfinal target shadesare offered for thesubject device.
FlexuralStrength>800 MPa>800 MPaSame
Coefficient ofThermalExpansion(CTE/25-500°C)10-11 x 10-6 /°C10-11 x 10-6 /°CSame
ChemicalSolubility<100 µg/cm²<100 µg/cm²Same
RadioactivityThe activityconcentration ofUranium-238 is nomore than 1.0 Bq/g-1.The activityconcentration ofUranium-238 is nomore than 1.0 Bq/g-1.Same
BiocompatibilityBiocompatible andnon-toxicBiocompatible andnon-toxicSame
SterilityNon-sterileNon-sterileSame
ManufacturingProcessColloidal processBi-axial pressingprocessDifferentmanufacturingprocess

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo consists of a triangle on the left side that is filled with a gradient of colors, including pink, red, yellow, green, blue, and purple. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC." in a blue sans-serif font.

NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS (21 CFR 807.92(b))

Design verification according to BruxZir® Esthetic Design Verification Protocol was conducted by Prismatik Dentalcraft, Inc. The tests conducted are listed below.

  • Flexural Strength according to ISO 6872:2015/Amd 1:2018
  • Visual Shade Evaluation
  • Color Consistency .
  • Translucency ●
  • Coefficient of Thermal Expansion according to ISO 6872:2015/Amd 1:2018 ●
  • Chemical Solubility according to ISO 6872:2015/Amd 1:2018 ●
  • Radioactivity according to ISO 6872:2015/Amd 1:2018 ●
  • Biocompatibility according to ISO 10993-1:2018 ●

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo features a colorful triangle on the left side, with the company name in blue text on the right. The triangle is divided into sections of different colors, including pink, red, yellow, green, and blue. The text is in a simple, sans-serif font.

No clinical data is included in this submission.

The test samples were made using the standard manufacturing process and OC process for BruxZir® Esthetic to ensure that the test samples meet the following criteria for the finished device:

  • Flexural Strength must be greater than 800 MPa after sintering, as measured per ISO . 6872:2015/Amd 1:2018, Type II Class 5.
  • Grade of "Pass" for Visual Shade Match for all shades. ●
  • . Color consistency ΔE calculation needs to be less than 2.00 between samples measured.
  • Translucency of each formulation must be within 1 % of the equivalent BruxZir Anterior-● 16 shade at 700nm.
  • The Coefficient of Thermal Expansion must be between 10-11 x 10-9℃ and the standard . deviation of CTE must be no greater than 0.5 x 10 %/℃ per ISO 6872:2015/Amd 1:2018.
  • The solubility must be lower than 100 ug/cm2 per ISO 6872:2015/Amd 1:2018. ●
  • . The activity concentration of Uranium-238 must not exceed 1.0 Bq g-1 per ISO 6872:2015/Amd 1:2018.
  • . BruxZir® Esthetic must be biocompatible, safe, and non-toxic.

The results of the verification activities confirmed that the design of the subject device met all the predetermined criteria according to the design specifications. The results of the testing were used to address questions related to substantial equivalence based on differences in technological features between the subject device, BruxZir® Esthetic, and the predicate device, BruxZir® Esthetic (K213425), by verifying that the modified device met all the design specification requirements, and the design changes were implemented successfully without causing any unexpected issues.

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.