K Number
K240763

Validate with FDA (Live)

Date Cleared
2024-12-18

(273 days)

Product Code
Regulation Number
874.3730
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Blom-Singer® FirstFit™ Surgical Kit (FFSK) is indicated for use during surgical creation of primary or secondary tracheoesophageal puncture, dilation of the tracheoesophageal wall, and to guide the placement of the voice prosthesis for tracheoesophageal voice restoration following total laryngectomy.

The FFSK contains a sterile single use indwelling Blom-Singer® Indwelling Voice Prosthesis device intended for voice rehabilitation after surgical removal of the larynx (laryngectomy). Cleaning of the voice prosthesis is performed by the patient while it remains in situ.

Device Description

The FirstFit Surgical Kit (FFSK) is a kit for creating a primary or secondary TE puncture, with subsequent dilatation of that puncture to a width that facilitates placement of the included Blom-Singer® Indwelling Voice Prosthesis, cleared in K082206.

The FFSK is for transient use as an aid to a surgical operation to puncture and stent the tracheoesophageal septum. The FFSK is provided sterile (by ethylene oxide) in a sealed tray, ready to be introduced into the surgical field. Additionally, it is used to guide voice prosthesis placement in a total laryngectomized patient as a primary or secondary procedure. Puncture and placement should take place in a sterile field with the patient being under anesthesia.

The FFSK is designed for primary or secondary tracheoesophageal puncture (TEP) with retrograde voice prosthesis insertion. This kit includes: one (1) 13-gauge curved core puncture needle, one (1) 36-inch guidewire with rounded tips, one (1) pharynx protector tool with handle and notched cylinder, one (1) voice prosthesis inserter pre-loaded with a Blom-Singer® Indwelling Voice Prosthesis of the diameter and length specified on the package label.

The Blom-Singer® Indwelling Voice Prosthesis devices (VP) are indwelling devices primarily made of silicone, designed to provide voicing after total laryngectomy. The voice prostheses are placed in a surgically created fistula between the trachea and esophagus in order to divert air through the prosthesis valve to create voicing. The use of an indwelling device means that routine removal for cleaning by the patient is not necessary, thus making it easier for the patient to maintain the device the risk of accidental dislodgment of the device.

AI/ML Overview

The provided FDA 510(k) summary does not contain information on rigorous acceptance criteria and a detailed study proving the device meets those criteria in the context of an AI/software medical device, as the device is a physical surgical kit (FirstFit Surgical Kit). Therefore, the requested information regarding AI model performance (e.g., sample sizes for training/test sets, ground truth establishment, MRMC studies, standalone performance, expert qualifications) is not applicable to this submission.

However, I can extract the relevant performance testing and findings for the physical device as described in the summary:

Device Name: FirstFit Surgical Kit
Regulation Name: Laryngeal Prosthesis (Taub Design)
Regulatory Class: Class II
Product Code: EWL

Here's a breakdown of the performance testing that serves as the "study" for this physical device:


1. Table of Acceptance Criteria and Reported Device Performance (as inferred from the submission)

For a physical surgical kit, the "acceptance criteria" are generally related to its ability to perform its intended surgical functions safely and effectively. The submission describes a cadaver study as the primary performance evaluation.

Acceptance Criteria (Inferred)Reported Device Performance
Usability/Functionality for Tracheoesophageal Puncture (TEP)Met: Study concluded users can effectively place the FirstFit Surgical Kit.
Usability/Functionality for TEP DilationMet with Caveat: Study concluded users can effectively place the FirstFit Surgical Kit. However, in 3 out of 37 cadavers, additional instrumentation was needed for dilation to complete voice prosthesis placement. This was attributed to post-mortem changes and increased tissue stiffness in fresh-frozen cadaver tissue compared to live tissue (Hohmanna et al, 2018 cited: 55.3MPa modulus for fresh-frozen vs 25.6MPa for fresh long bicep tendons). As a precaution, labeling was updated with a warning: "proceed with great care and abort the procedure if dilation of the TE puncture requires too much force." An additional recommendation for abandoning the procedure mid-way was also added to labeling.
Usability/Functionality for Retrograde Voice Prosthesis InsertionMet: Study concluded users can effectively place the FirstFit Surgical Kit.
Absence of Tissue Damage (Ripping, Tearing, etc.) during puncture and dilationMet: No tissue damage (ripping, tearing, etc.) reported during puncture and dilation of the tracheoesophageal wall and placement of the voice prosthesis.

2. Sample Size and Data Provenance for the Test Set

  • Sample Size: 15 users performing 2 punctures & placements each on fresh cadaver tissue. The exact number of cadavers is stated as "37 cadavers" needing additional instrumentation for dilation out of an unstated total number of cadavers used. However, the primary performance data seems to be derived from the "15 users performing 2 punctures & placements each".
  • Data Provenance: Fresh cadaver tissue. The country of origin is not specified. This was a prospective "usability validation testing methodology."

3. Number of Experts and Qualifications for Ground Truth

  • Ground Truth Establishment: For a physical device, "ground truth" usually refers to the successful and safe execution of the surgical procedure. In this case, the "usability engineering" study was conducted by "users" (presumably surgeons or medical professionals who would use the device). The document does not specify if these users were considered "experts" for establishing ground truth, nor does it explicitly state their qualifications (e.g., number of years of experience). The assessment of "no tissue damage" would likely have been made by the study observers or the users themselves.

4. Adjudication Method for the Test Set

  • The document does not describe a formal "adjudication method" in the sense of multiple reviewers arbitrating discrepancies. The study observed the performance of the 15 users.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

  • No. This is a physical surgical kit, not an AI/software device. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable.

6. Standalone Performance (Algorithm Only)

  • N/A. This is a physical surgical kit. There is no algorithm to test in standalone mode.

7. Type of Ground Truth Used

  • Observed Performance on Cadaveric Tissue: The "ground truth" was the observed successful (or, in 3 cases, difficult) completion of the surgical steps (puncture, dilation, prosthesis placement) and the absence of tissue damage in the simulated-use cadaver study. It's a pragmatic assessment of procedural success and safety.

8. Sample Size for the Training Set

  • N/A. This is not an AI/machine learning device that requires a "training set."

9. How Ground Truth for the Training Set Was Established

  • N/A. Not applicable for a physical device.

In summary, the "study" for this device was a usability engineering study conducted on cadavers to demonstrate the safe and effective performance of the FirstFit Surgical Kit in creating tracheoesophageal punctures, dilating them, and guiding voice prosthesis placement. The primary finding was that the device met its functional goals, with an important caveat regarding potential tissue stiffness in cadavers leading to a labeling update for real-world use.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 18, 2024

Trey Thorsen Regulatory Consultant O'Connell & Myers, LLC 1110 Mark Ave Carpinteria, California 93013

Re: K240763

Trade/Device Name: FirstFit Surgical Kit Regulation Number: 21 CFR 874.3730 Regulation Name: Laryngeal Prosthesis (Taub Design) Regulatory Class: Class II Product Code: EWL Dated: November 18, 2024 Received: November 18, 2024

Dear Trey Thorsen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Joyce C. Lin -S

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240763

Device Name

FirstFit Surgical Kit

Indications for Use (Describe)

The Blom-Singer® FirstFit™ Surgical Kit (FFSK) is indicated for use during surgical creation of primary or secondary tracheoesophageal puncture, dilation of the tracheoesophageal wall, and to guide the placement of the voice prosthesis for tracheoesophageal voice restoration following total laryngectomy.

The FFSK contains a sterile single use indwelling Blom-Singer® Indwelling Voice Prosthesis device intended for voice rehabilitation after surgical removal of the larynx (laryngectomy). Cleaning of the voice prosthesis is performed by the patient while it remains in situ.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours be response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary Date: December 17, 2024

SUBMITTER

Freudenberg Medical, LLC 1110 Mark Ave. Carpinteria CA 93013

Contact Person: Trey Thorsen Email: trey@oconnellmyers.com Phone: 850-450-3932

DEVICE

Name of Device: FirstFit™ Surgical Kit Common or Usual Name: Prosthesis, Laryngeal (Taub) Classification Name: Prosthesis, Laryngeal (Taub) Regulation: 21 CFR 874.3730 Regulatory Class: II Primary Product Code: EWL

PREDICATE DEVICES

Predicate: Provox® Vega Puncture Set, (K131947)

DEVICE DESCRIPTION

The FirstFit Surgical Kit (FFSK) is a kit for creating a primary or secondary TE puncture, with subsequent dilatation of that puncture to a width that facilitates placement of the included Blom-Singer® Indwelling Voice Prosthesis, cleared in K082206.

The FFSK is for transient use as an aid to a surgical operation to puncture and stent the tracheoesophageal septum. The FFSK is provided sterile (by ethylene oxide) in a sealed tray, ready to be introduced into the surgical field. Additionally, it is used to guide voice prosthesis placement in a total laryngectomized patient as a primary or secondary procedure. Puncture and placement should take place in a sterile field with the patient being under anesthesia.

The FFSK is designed for primary or secondary tracheoesophageal puncture (TEP) with retrograde voice prosthesis insertion. This kit includes: one (1) 13-gauge curved core puncture needle, one (1) 36-inch guidewire with rounded tips, one (1) pharynx protector tool with handle and notched cylinder, one (1) voice prosthesis inserter pre-loaded with a Blom-Singer® Indwelling Voice Prosthesis of the diameter and length specified on the package label.

The Blom-Singer® Indwelling Voice Prosthesis devices (VP) are indwelling devices primarily made of silicone, designed to provide voicing after total laryngectomy. The voice prostheses are placed in a surgically created fistula between the trachea and esophagus in order to divert air through the prosthesis valve to create voicing. The use of an indwelling device means that routine removal for cleaning by the

{5}------------------------------------------------

patient is not necessary, thus making it easier for the patient to maintain the device the risk of accidental dislodgment of the device.

INDICATIONS FOR USE / INTENDED USE

The Blom-Singer FirstFit Surgical Kit (FFSK) is indicated for use during surgical creation of primary or secondary tracheoesophageal puncture, dilation of the tracheoesophageal wall, and to guide the placement of the voice prosthesis for tracheoesophageal voice restoration following total laryngectomy. The FFSK contains a sterile single use indwelling Blom-Singer® Indwelling Voice Prosthesis device intended for voice rehabilitation after surgical removal of the larynx (laryngectomy). Cleaning of the voice prosthesis is performed by the patient while it remains in situ.

PATIENT POPULATION

For patients who had their larynx surgically removed.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The FirstFit Surgical Kit (FFSK) is a convenience kit for creating a primary or secondary TE puncture, with subsequent dilatation of that puncture to a width that facilitates placement of the included Blom-Singer® Indwelling Voice Prosthesis, cleared in K082206.

Biocompatibility testing

The biocompatibility evaluation for the FFSK was conducted in accordance with the FDA guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, May 1, 1995, and International Standard ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA.

Performance Testing

The following testing was performed according to:

  • . Sterilization
    • Sterilization method: Ethylene Oxide O
    • Validation method: overkill approach (half-cycle method) per ISO 11135, ISO 10993-7, о ISO 14937
  • . Shelf Life
    • Shelf life by accelerated aging- per ASTM F1980-21 (Accelerated Aging of Sterile Barrier O Systems for Medical Devices)
    • O Packaging Validation - Visual Inspection (per ASTM F1886), Burst Test (per ASTM F1140), Dye Penetration (per ASTM F1929), Peel Strength Testing (per TM-061)
  • Biocompatibility - EN ISO 10993-1
  • Bench testing - per internal protocols
    • o Usability engineering IEC 62366-1: Cadaver Study was performed under simulated-use conditions with 15 users performing 2 punctures & placements each on fresh cadaver tissue as a usability validation testing methodology for the FirstFit Surgical Kit to ensure the subject device can accommodate:
    • an anterograde tracheoesophageal puncture (TEP) (primary & secondary), O

{6}------------------------------------------------

  • a TEP dilation to allow for placement of 16Fr, 20Fr, & 22.5Fr voice prostheses O
  • a retrograde voice prosthesis insertion in the TEP (primary & secondary) During the study O 3 out of 37 cadavers needed additional instrumentation for dilation to complete the placement of the voice prostheses. This result was indicative of post-mortem changes in cadaveric anatomy where it is no longer alive and has undergone significant changes due to the death and preservation by freezing process, including potential tissue stiffening, which could lead to difficulty dilating the tissue. This is supported in Hohmanna et al, 2018 that suggests fresh-frozen cadaver tissue exhibits a higher modulus (55.3MPa) of elasticity/stiffness compared to fresh tissue (25.6MPa) as observed in cadaveric long bicep tendons, which may suggest frozen cadaveric tissue used in our cadaver study exhibited a higher stiffness than what would be observed in a live patient.

The study concluded users can effectively place the FirstFit Surgical Kit with no tissue damage (ripping, tearing, etc.) during puncture and dilation of the tracheoesophageal wall and placement of the voice prosthesis.

Animal Studies

No animal data submitted.

Clinical Studies

No clinical data submitted.

CONCLUSIONS

Based on the study result and as a precaution, the FFSK's labelling has been updated to include a warning to "proceed with great care and abort the procedure if dilation of the TE puncture requires too much force". The addition of this waring aligns with the predicate's labeling.

The FFSK labeling also includes an additional recommendation for how to proceed if the procedure needs to be abandoned mid-way after the puncture has been created and dilated.

The physician must determine if the patient's tissue or anatomy is eligible for safe use of this kit as suggested by the contraindications, warnings, and precautions of our Instructions for Use either prior to procedure or during.

Freudenberg Medical believes the additional warning and recommendation for how to proceed certain cases, combined with the performance testing demonstrate that the subject device is substantially equivalent to the predicate.

§ 874.3730 Laryngeal prosthesis (Taub design).

(a)
Identification. A laryngeal prosthesis (Taub design) is a device intended to direct pulmonary air flow to the pharynx in the absence of the larynx, thereby permitting esophageal speech. The device is interposed between openings in the trachea and the esophagus and may be removed and replaced each day by the patient. During phonation, air from the lungs is directed to flow through the device and over the esophageal mucosa to provide a sound source that is articulated as speech.(b)
Classification. Class II (special controls). The device, when it is a voice amplification device, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.