(104 days)
Use the 3M TM Attest TM Rapid Readout Biological Indicator 1295 in conjunction with the 3M TM Attest TM Auto-reader 490H or 490 Auto-reader having software version 4.0.0 or greater or 490M Auto-reader as a standard method of routine monitoring of vaporized hydrogen peroxide sterilization processes in the following systems: STERRAD 100S® Sterilization System STERRAD NX® Sterilization System (Standard and Advanced cycles) STERRAD 100NX® Sterilization System (Standard, Flex, Express and Duo cycles) STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles) STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express and Duo cycles) V-PRO® 1 Low Temperature Sterilization System (Lumen cycle) V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles) V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles) V-PRO® 60 Low Temperature Sterilization System (Lumen and Flexible cycles) V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast Non Lumen cvcles) V-PRO® &2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) STERIZONE® VP4 Sterilizer (Cycle 1) SteroScope® Sterilization System
Use the 3M IM Attest TM Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E for pack control monitoring of the following hydrogen peroxide sterilization sterilizers and cycles: STERRAD 100S® Sterilization System STERRAD NX® Sterilization System (Standard and Advanced cycles) STERRAD 100NX® Sterilization System (Standard, Flex, Express and Duo cycles) STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles) STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express and Duo cycles) V-PRO® 1 Low Temperature Sterilization System (Lumen cycle) V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles) V-PRO® maXLow Temperature Sterilization System (Lumen, and Flexible cycles) V-PRO® 60 Low Temperature Sterilization System (Lumen and Flexible cycles) V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles) V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) STERIZONE® VP4 Sterilizer (Cycle 1) SteroScope® Sterilization System
3M™ Attest™ Rapid Readout Biological Indicator 1295: The 3M™ Attest™ Rapid Readout Biological Indicator 1295 is a self-contained biological indicator specifically designed for rapid and reliable routine monitoring of vaporized hydrogen peroxide sterilization processes when used in conjunction with the 3MTM Attest™ Auto-reader 490H or the 3MTM Attest™ Auto-reader 490 having software version 4.0.0 or greater or a 3MTM Attest™ Mini Auto-reader 490M. The 1295 BI is a single-use device composed of a plastic sleeve containing a spore carrier and media ampoule, enclosed with a color-coded cap. A chemical process indicator printed with stripes which change from blue toward pink upon exposure to vaporized hydrogen peroxide is located on the top of the cap. The detection of fluorescence upon incubation of the 1295 BI in one of the designated Attest™ Auto-readers indicates a sterilization failure.
3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E: The 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E is a chemical indicator consisting of a non-cellulose based coated indicator strip sensitive to vaporized hydrogen peroxide, contained in a film laminate. There has been no change to the product construction compared to the predicate submission, K212022. The 3M™ Attest™ CI 1348/1348E verifies that the stated values for the three critical parameters of exposure time, temperature, and concentration of vaporized hydrogen peroxide have been achieved within a package or containment device (i.e. wrapped trays, rigid containers, sterilization pouches, and other types of packs) and/or at a specific location within the load or empty chamber. Upon exposure to vaporized hydrogen peroxide, the color of the coated indicator strip progressively changes from blue toward pink along the strip. The progression of the blue to pink color change along the strip is visible through a window with marked "REJECT" and "ACCEPT" zones. The extent of the progression depends on exposure time, temperature, and concentration of vaporized hydrogen peroxide.
The provided text describes premarket notification for two devices: 3M™ Rapid Readout Biological Indicator 1295 and 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E. The key new information in this submission is the addition of the "SteroScope® Sterilization System" to the indications for use for both devices. The acceptance criteria and the study performance for this new indication are detailed.
Here's a breakdown of the requested information based on the provided text:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
Device: 3M™ Attest™ Rapid Readout Biological Indicator 1295 (for SteroScope® Sterilization System)
Test Name | Purpose | Acceptance Criteria | Result |
---|---|---|---|
Full Cycle Performance | Verify performance in each of the full cycles in the SteroScope® Sterilization System. | All biological indicators display a negative fluorescent and negative growth response. | Pass |
Fractional Cycle Performance | Verify performance in fractional cycles for each of the cycles within the SteroScope® Sterilization System. | All biological indicators display a negative fluorescent and negative growth response. | Pass |
Chemical Indicator (CI) Color Change | Demonstrate the color change of the CI when exposed to the SteroScope® Sterilization System. | Color change from blue toward pink. | Pass (implicitly, as part of the BI unit) |
Device: 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E (for SteroScope® Sterilization System)
Test Name | Purpose | Acceptance Criteria | Result |
---|---|---|---|
Complete (Full) Cycle Performance | Verify performance in the complete (full) cycle of a loaded SteroScope® Sterilization System. | Visual assessment of the chemical indicator yields an ACCEPT result with ≥ 95% accuracy. | |
At least 80% of the population of indicators tested has a measured color change value corresponding to an ACCEPT result. | Pass, 100% accuracy | ||
Pass, >99.9% population | |||
Fractional Cycle Performance | Verify performance in an incomplete cycle of a loaded SteroScope® Sterilization System. | Visual assessment of the chemical indicator yields a REJECT result with ≥ 95% accuracy. | |
At least 80% of the population of indicators tested has a measured color change value corresponding to a REJECT result. | Pass, 100% accuracy | ||
Pass, >99.9% population |
Study Details
2. Sample size used for the test set and the data provenance
The document does not explicitly state the specific numerical sample sizes for the test sets (e.g., number of BIs or CIs tested in each condition). Instead, it uses qualitative statements like "All biological indicators" and quantitative statements related to percentages of the population of indicators.
- Data Provenance: The testing was "performed in a health care facility" (for the Chemical Indicator) and implicitly in a laboratory/testing environment for the Biological Indicator. The data is prospective, as it was generated specifically to demonstrate performance for the expanded indications. The country of origin is not specified, but given the FDA submission, it's likely U.S.-based or compliant with U.S. regulatory standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not mention the use of experts or their qualifications for establishing ground truth. The ground truth for these types of devices (Biological and Chemical Indicators) is typically established through precisely controlled sterilization cycles using calibrated equipment and defined parameters (e.g., successful sterilization leading to no growth for BIs, or specific chemical changes for CIs). The "Visual assessment" mentioned for the CI implies human observation, but it doesn't specify if these were expert readers or how many.
4. Adjudication method for the test set
The document does not specify any formal adjudication method (e.g., 2+1, 3+1). For the chemical indicator, it mentions "Visual assessment," but no details on multiple readers or adjudication are provided.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This type of study is more common for diagnostic imaging AI devices, where human perception and interpretation are key variables. For sterilization indicators, the evaluation focuses on the device's intrinsic performance against defined physical/chemical/biological criteria rather than how human readers' diagnostic accuracy is improved with AI assistance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This question is not directly applicable in the context of these specific devices. The Biological Indicator's readout is automated by an "Auto-reader," which itself performs the "algorithm" (fluorescence detection). The Chemical Indicator's performance is determined by a visual color change. There isn't an "AI algorithm" in the typical sense (e.g., for image analysis) being evaluated here. The "standalone" performance is essentially the performance of the indicators themselves, which is what the non-clinical tests evaluate.
7. The type of ground truth used
- For the 3M™ Attest™ Rapid Readout Biological Indicator 1295:
- Biological Ground Truth: Absence or presence of fluorescent and/or microbial growth, indicating kill or survival of Geobacillus stearothermophilus spores after exposure to sterilization cycles. This is established by standard microbiological methods and incubation in the auto-reader.
- For the 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E:
- Physical/Chemical Ground Truth: The successful achievement of specified exposure time, temperature, and vaporized hydrogen peroxide concentration (ACCEPT result) or failure to achieve these conditions (REJECT result). This is confirmed by the color change progression of the indicator and validated against the known conditions of the sterilization cycles (i.e., "complete/full cycle" vs. "fractional/incomplete cycle"). The "Visual assessment" is against this physical/chemical ground truth.
8. The sample size for the training set
The document does not provide information on a "training set" or its sample size. These devices are physical/chemical/biological indicators, not AI/machine learning models that typically require training data. The testing described is verification and validation, not model training.
9. How the ground truth for the training set was established
As there is no mention of a "training set" or AI/ML model for these devices, this question is not applicable to the provided information.
§ 880.2800 Sterilization process indicator.
(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).