(70 days)
K221618 - Spine & Trauma Navigation
Not Found
No
The summary describes an image-guided surgery system using optical tracking and image registration. There is no mention of AI or ML in the intended use, device description, or performance studies. The focus is on mechanical accuracy, usability, and biocompatibility.
No.
This device is an image-guided navigation system and surgical instruments used to assist surgeons in performing procedures, rather than directly treating a medical condition itself.
No
Explanation: The device is described as an "intraoperative image-guided localization system" and "image guided surgery system" intended to assist surgeons in navigation during surgical procedures, not to diagnose a medical condition.
No
The device description explicitly states that the system "consists of software, Image Guided Surgery platforms and surgical instruments," including modified drill guides and drill bits. This indicates the presence of hardware components beyond just software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health.
- Device Function: The Spine & Trauma Navigation system is an image-guided surgery system. Its primary function is to assist surgeons during procedures by providing real-time localization and navigation based on medical images. It does not perform tests on biological samples.
- Intended Use: The intended use clearly states it's an "intraoperative image-guided localization system" for surgical procedures.
- Device Description: The description focuses on the software, hardware (platforms, instruments), and tracking technology used for surgical navigation. It does not mention any components or processes related to analyzing biological samples.
The device is a surgical navigation system, which falls under a different category of medical devices than In Vitro Diagnostics.
N/A
Intended Use / Indications for Use
Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments.
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy.
Product codes (comma separated list FDA assigned to the subject device)
OLO
Device Description
The Spine & Trauma Navigation is an image quided surgery system for navigated treatments in the fields of spine and trauma surgery, whereas the user may use image data based on CT, MR, 3D fluoroscopic image reconstruction (cone beam CT) or 2D fluoroscopic images. It offers different patient image registration methods and instrument calibrations to allow surgical navigation by using optical tracking technology. To fulfil this purpose, it consists of software, Image Guided Surgery platforms and surgical instruments.
Modified Drill Guides and Drill Bits have been introduced as part of the Subject Device. The Drill Guide instruments are navigated instruments which support the surgeon in guiding drill bits and K-wires during spinal procedures. They consist of a guide tube, a trocar insert (both available in five different diameters), a body with two available handles, an array and a depth control (available in two different sizes for various drilling depths). The Drill Guide Tubes and Drill Guide Trocar Inserts have patient contact. All instruments are delivered unsterile and require end user sterilization.
The Drill Bits are used for drilling of bone. They are made of stainless steel and are delivered non-sterile. They require steam sterilization onsite before use. There are several variants in terms of diameter, length, and presence of a depth stop feature.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
preoperative or intraoperative 2D or 3D image data.
CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image
Anatomical Site
skull, the pelvis, a long bone or vertebra
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
- Assembly stability: The Drill Guide was able to withstand the forces without losing accuracy or function.
- Accuracy: Accuracy testing was performed with acceptance criteria: tip position deviation must be equal or below 1.7 mm (95th percentile), and the angular deviation must be equal or below 1.7° (95th percentile). Testing showed the modified Drill Guide fulfills the same accuracy requirements as the predecessor one.
- Lifecyle assessment: Accuracy as well as label readability requirements were met.
- Skiving: Skiving testing showed the modified Drill Guide provides a more stable hold with the improved teeth design compared to the predicate Guide Tubes.
- Handling and interface analysis: The Drill Guide design changes in terms of depth control and array attachment concept were evaluated and acceptance criteria were met in all cases.
- Usability: Summative usability testing was performed in order to validate the changed design of the Drill Guides. The evaluation was conducted with 15 representative users which had to assemble and use the Drill Guide in a simulated scenario. The final design was proven safe and effective for use in the defined use scenarios.
- Mechanical failure testing: The new worst case drill bit (diameter 2 mm) was tested under different scenarios in order to ensure it is strong enough to withstand the expected torques and possible bending under worst case conditions.
- Biocompatibility: Was evaluated according to ISO 10993-1:2018 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
- Reprocessing validation: Was evaluated according to ISO 11737-1:2018, ANSI/AAMI ST98:2022 and ISO 15883-5:2021-07.
No clinical testing was required for the subject device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Mean Positional Error of the placed instrument's tip
§ 882.4560 Stereotaxic instrument.
(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).
0
May 23, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Brainlab AG Esther Moreno Garcia Manager Regulatory Affairs Olof-Palme-Str. 9 Munich, 81829 Germany
Re: K240701
Trade/Device Name: Drill Guide; Drill Bit; Spine & Trauma Navigation Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 14, 2024 Received: April 23, 2024
Dear Esther Moreno Garcia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
1
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tejen D. Soni -S
For
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
2
Indications for Use
510(k) Number (if known) K240701
Device Name Drill Guide; Drill Bit; Spine & Trauma Navigation
Indications for Use (Describe)
Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments.
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink brain icon on the left, followed by the word "BRAINLAB" in pink, uppercase letters. The brain icon appears to be made up of several curved lines, giving it a modern and abstract look.
510(k) Summary
March 14th, 2024
General Information | |
---|---|
Manufacturer | Brainlab AG; Olof-Palme-Str.9, 81829, Munich, Germany |
Establishment Registration | 8043933 |
Trade Name | Drill Guide |
Drill Bit | |
Spine & Trauma Navigation | |
Classification Name | Orthopedic Stereotaxic Instrument |
Product Code | OLO |
Regulation Number | 882.4560 |
Regulatory Class | II |
Panel | Orthopedic |
Predicate Device(s) | K221618 - Spine & Trauma Navigation |
Contact Information | |
Primary Contact | Alternate Contact |
Esther Moreno Garcia | |
Manager Regulatory Affairs | |
Phone: +49 89 99 15 68 0 | |
Email: regulatory.affairs@brainlab.com | Chiara Cunico |
Phone: +49 89 99 15 68 0 | |
Fax: +49 89 99 15 68 5033 | |
Email: chiara.cunico@brainlab.com |
1. Indications for Use
Spine & Trauma Navigation is intended as an intraoperative image-guided localization system to enable open and minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Spine & Trauma Navigation enables computer-assisted navigation of medical image data, which can either be acquired preoperatively or intraoperatively by an appropriate image acquisition system. The software offers screw and interbody device planning and navigation with surgical instruments.
The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to the acquired image (CT, MR, 3D fluoroscopic image reconstruction or 2D fluoroscopic image) and/or an image databased model of the anatomy.
2. Device Description
4
Image /page/4/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink symbol on the left, resembling a brain or a network, followed by the word "BRAINLAB" in pink, block letters. The overall design is clean and modern, suggesting a focus on technology and innovation in the field of neuroscience or medical technology.
The Spine & Trauma Navigation is an image quided surgery system for navigated treatments in the fields of spine and trauma surgery, whereas the user may use image data based on CT, MR, 3D fluoroscopic image reconstruction (cone beam CT) or 2D fluoroscopic images. It offers different patient image registration methods and instrument calibrations to allow surgical navigation by using optical tracking technology. To fulfil this purpose, it consists of software, Image Guided Surgery platforms and surgical instruments.
Modified Drill Guides and Drill Bits have been introduced as part of the Subject Device. The Drill Guide instruments are navigated instruments which support the surgeon in guiding drill bits and K-wires during spinal procedures. They consist of a guide tube, a trocar insert (both available in five different diameters), a body with two available handles, an array and a depth control (available in two different sizes for various drilling depths). The Drill Guide Tubes and Drill Guide Trocar Inserts have patient contact. All instruments are delivered unsterile and require end user sterilization.
The Drill Bits are used for drilling of bone. They are made of stainless steel and are delivered non-sterile. They require steam sterilization onsite before use. There are several variants in terms of diameter, length, and presence of a depth stop feature.
3. Substantial Equivalence
For the Substantial Equivalence determination, comparison of the Subject Device features with the following predicate device(s) was carried out:
K221618 - Spine & Trauma Navigation (product code: OLO)
The predicate device was chosen since it's the predecessor device and therefore very similar to the subject device w.r.t the indications for use, technological characteristics and use cases.
The main difference compared to the predicate device are modified instruments.
Feature | Predicate Device K221618 | Subject Device |
---|---|---|
Indications | ||
for use | Spine & Trauma Navigation is | |
intended as an intraoperative image- | ||
guided localization system to enable | ||
open and minimally invasive surgery. | ||
It links a freehand probe, tracked by | ||
a passive marker sensor system to | ||
virtual computer image space on a | ||
patient's preoperative or | ||
intraoperative 2D or 3D image data. |
Spine & Trauma Navigation enables
computer-assisted navigation of
medical image data, which can either | Spine & Trauma Navigation is
intended as an intraoperative image-
guided localization system to enable
open and minimally invasive surgery.
It links a freehand probe, tracked by
a passive marker sensor system to
virtual computer image space on a
patient's preoperative or
intraoperative 2D or 3D image data.
Spine & Trauma Navigation enables
computer-assisted navigation of
medical image data, which can either |
| Feature | Predicate Device K221618 | Subject Device |
| | be acquired preoperatively or
intraoperatively by an appropriate
image acquisition system. The
software offers screw and interbody
device planning and navigation with
surgical instruments. | be acquired preoperatively or
intraoperatively by an appropriate
image acquisition system. The
software offers screw and interbody
device planning and navigation with
surgical instruments. |
| | The system is indicated for any
medical condition in which the use of
stereotactic surgery may be
appropriate and where a reference to
a rigid anatomical structure, such as
the skull, the pelvis, a long bone or
vertebra can be identified relative to
the acquired image (CT, MR, 3D
fluoroscopic image reconstruction or
2D fluoroscopic image) and/or an
image databased model of the
anatomy. | The system is indicated for any
medical condition in which the use
of stereotactic surgery may be
appropriate and where a reference
to a rigid anatomical structure, such
as the skull, the pelvis, a long bone
or vertebra can be identified relative
to the acquired image (CT, MR, 3D
fluoroscopic image reconstruction or
2D fluoroscopic image) and/or an
image databased model of the
anatomy. |
| Navigation
accuracy | Mean Positional Error of the placed
instrument's tip ≤ 2 mm
Mean Angular Error of the placed
instrument's axis ≤ 2° | Mean Positional Error of the placed
instrument's tip ≤ 2 mm
Mean Angular Error of the placed
instrument's axis ≤ 2° |
| Tracking
technology | Optical tracking | Optical tracking |
| Drill guide
components | -Drill Guide Handles
-Drill Guide Compact Handle
-Drill Guide Tubes
-Drill Guide Trocars
-Tissue Protection Sleeve
-Drill Guide Depth Controls | -Drill Guide Handle
-Drill Guide Handle Compact
-Drill Guide Tubes
-Drill Guide Trocar Inserts
-Drill Guide Depth Controls
-Drill Guide Array |
| Adjustment
and
handling
features of
Drill Guides | • Adjusting position of handle
requires loosening of the
counter nut.
• Array is fixed to the handle body
• The guide tubes are attached to
the main drill guide body by
means of a left-handed thread.
• Depth control is attached to drill | • Rotatable handle allows to
adjust position relative to array
without loosening any fixation
elements.
• Detachable array with integrated
magnet to prevent it from
dropping. |
| Feature | Predicate Device K221618 | Subject Device |
| | depth setting is secured with a
counter nut. | The Guide Tube is attached to
the Handle by simple linear
insertion and held in place by a
click interface. Depth control is attached and
depth setting is secured with a
snapping mechanism. |
| Drill Bits | Drill bits with diameters between
2.4mm to 4.5mm. | Four new drill bits are introduced,
three within the same range of
diameters (3.2 and 4.5mm). A new
worst case drill bit with a diameter of
2mm. |
| Cleaning
and
sterilization | Automated cleaning and automated
thermal disinfection
Steam sterilization | Automated cleaning and automated
thermal disinfection
Steam sterilization |
5
Image /page/5/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized pink brain icon on the left, followed by the word "BRAINLAB" in pink, sans-serif capital letters. The brain icon is abstract and appears to be made up of several curved lines.
6
Image /page/6/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized brain icon on the left and the word "BRAINLAB" in capital letters on the right. The color of the logo is pink.
4. Performance Data
The following testing was conducted on the Subject Device to establish substantial equivalence with the predicate device:
- Assembly stability: In order to verify that the stiffness and fixation of the instruments can resist unintended loads. The Drill Guide was able to withstand the forces without losing accuracy or function.
- . Accuracy: Accuracy testing was performed with acceptance criteria: tip position deviation must be equal or below 1.7 mm (95th percentile), and the angular deviation must be equal or below 1.7° (95th percentile). Testing showed the modified Drill Guide fulfills the same accuracy requirements as the predecessor one.
- Lifecyle assessment: To evaluate the influence of wear, reprocessing and aging on . selected aspects of product performance throughout the product lifetime. Accuracy as well as label readability requirements were met.
- . Skiving: Skiving testing showed the modified Drill Guide provides a more stable hold with the improved teeth design compared to the predicate Guide Tubes.
- . Handling and interface analysis: The Drill Guide design changes in terms of depth control and array attachment concept were evaluated and acceptance criteria were met in all cases.
- . Usability: Summative usability testing was performed in order to validate the changed design of the Drill Guides. The evaluation was conducted with 15 representative users
7
Image /page/7/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized brain icon on the left and the word "BRAINLAB" in capital letters on the right. The color of the logo is pink.
which had to assemble and use the Drill Guide in a simulated scenario. The final design was proven safe and effective for use in the defined use scenarios.
- . Mechanical failure testing: The new worst case drill bit (diameter 2 mm) was tested under different scenarios in order to ensure it is strong enough to withstand the expected torques and possible bending under worst case conditions.
- . Biocompatibility: Was evaluated according to ISO 10993-1:2018 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
- . Reprocessing validation: Was evaluated according to ISO 11737-1:2018, ANSI/AAMI ST98:2022 and ISO 15883-5:2021-07.
No clinical testing was required for the subject device.
5. Conclusion
The comparison of the Subject Device with the predicate device shows that the modified Spine & Trauma Navigation including modified instruments has similar functionality, intended use and technological characteristics as the predicate device. Based on the comparison to the predicate and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.