(97 days)
Resona I9, Resona I9 Exp. Resona I9T. Resona I9 Easi, Resona I9 Nasa, Resona IV. Imagyn I9, Imagyn I9S, Imagyn 19 Easi, Nuewa 19, Nuewa 19T, Nuewa 19 Exp, Nuewa 19 Easi, Anesus 19, Anesus 19, Anesus 19, Resona 19W, Recho 19, Recho 19 Exp, Recho 19S, Recho 19T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, Intra-operative (includes abdominal, thoracic, and vascular), Laparoscopic, pediatric, small organ (breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), Thoracic/Pleural, adult and pediatric cardiac, trans-esoph. (Cardiac), peripheral vessel, and urology exams.
Modes of operation include: B, M, PWD, CWD, Color Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, Smart3D, 4D(Reape View, TDI, Color M, Strain Elastography, Contrast agent for LVO), V Flow, STE, STO, Contrast imaging (Contrast agent for Liver).
This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healtheare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.
The Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus 19, Resona I9W, Recho 19, Recho 19 Pro, Recho 19 Exp, Recho 19S, Recho 19T Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images in Modes of operation include: B. M. PWD, CWD , Color Doppler, Amplitude Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B. Power+PW+B). Tissue Harmonic Imaging, Smart3D, 4D(Realtime 3D), iScape View, TDI, Color M, Strain Elastography, Contrast imaging (Contrast agent for LVO), V Flow, STE, STQ, Contrast imaging (Contrast agent for Liver).
The Resona 19, Resona 19 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System can also measure anatomical structures and offer analysis packages to provide information based on which the competent health care professionals can make the diagnosis.
I'm sorry, but this document does not contain the information required to answer your request. The document is an FDA 510(k) clearance letter for a diagnostic ultrasound system. It outlines the device's indications for use, its substantial equivalence to predicate devices based on non-clinical tests (acoustic output, biocompatibility, electrical/mechanical safety, etc.), and applicable safety standards it conforms to.
However, it explicitly states under "8. Clinical Studies": "Not applicable. The subject of this submission... does not require clinical studies to support substantial equivalence."
Therefore, there is no information in this document regarding:
- Acceptance criteria tables with reported device performance for an AI/algorithm-related study.
- Sample sizes for test sets or data provenance.
- Number of experts for ground truth establishment or their qualifications.
- Adjudication methods.
- MRMC studies or effect sizes of AI assistance on human readers.
- Standalone algorithm performance studies.
- Type of ground truth used.
- Sample size for training sets.
- How ground truth for training sets was established.
This document is a regulatory clearance for a general purpose diagnostic ultrasound system, not a document detailing the performance and validation of an AI/ML component within such a system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date June 17, 2024. The month is June, the day is the 17th, and the year is 2024. The text is written in a clear, legible font.
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the agency's name on the right. The symbol is a stylized representation of a staff entwined with a serpent, which is a symbol of medicine and healing. To the right of the symbol, the text reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in a clear, sans-serif font. The color scheme is primarily blue and white, giving the logo a clean and professional appearance.
Shenzhen Mindray Bio-medical Electronics Co., LTD. % Tang Jing Technical Manager Regulatory Affairs Technical Regulation Department Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan Shenzhen, Guangdong 518057 CHINA
Re: K240684
Trade/Device Name: Resona 19. Resona 19 Exp. Resona 19T. Resona 19T. Resona 19 Easi. Resona 19 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T. Nuewa I9 Exp. Nuewa I9 Easi. Anesus I9 Easi. Eagus I9. Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System
Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: May 17, 2024 Received: May 17, 2024
Dear Tang Jing:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K240684
Device Name
Resona I9, Resona I9 Exp. Resona I9T. Resona I9 Easi. Resona IV. Imagyn 19, Imagyn 19, Imagyn 19, Imagyn 19 Easi, Nuewa I9, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System
Indications for Use (Describe)
Resona I9, Resona I9 Exp. Resona I9T. Resona I9 Easi, Resona I9 Nasa, Resona IV. Imagyn I9, Imagyn I9S, Imagyn 19 Easi, Nuewa 19, Nuewa 19T, Nuewa 19 Exp, Nuewa 19 Easi, Anesus 19, Anesus 19, Anesus 19, Resona 19W, Recho 19, Recho 19 Exp, Recho 19S, Recho 19T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, Intra-operative (includes abdominal, thoracic, and vascular), Laparoscopic, pediatric, small organ (breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), Thoracic/Pleural, adult and pediatric cardiac, trans-esoph. (Cardiac), peripheral vessel, and urology exams.
Modes of operation include: B, M, PWD, CWD, Color Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, Smart3D, 4D(Reape View, TDI, Color M, Strain Elastography, Contrast agent for LVO), V Flow, STE, STO, Contrast imaging (Contrast agent for Liver).
This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healtheare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.
| Type of Use ( Select one or both, as applicable ) | |
|---|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete.
review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c).
The assigned 510(k) number: K240684
1. Submitter
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 8188 6129 Fax: +86 755 2658 2680
Contact Person
Tang Jing Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Date Prepared: March 10, 2024
2. Device Name
Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi. Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System
Classification
Regulatory Class: II Review Category: Tier II 21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN) 21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO) 21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)
3. Predicate and reference devices
Resona I9 series Diagnostic Ultrasound System is comparable with and
{5}------------------------------------------------
| Device | Manufacturer | Model | DeviceClass | Product Code | 510KNumber |
|---|---|---|---|---|---|
| 1. Main predicate device | Mindray | Resona 19 | II | IYN, IYO, ITX | K216099 |
| 2. Reference device | Mindray | Resona R9 | II | IYN, IYO, ITX | K222928 |
| 3. Reference device | Mindray | MX7 | II | IYN, IYO, ITX | K212900 |
| 4. Reference device | Mindray | TEX20 | II | IYN, IYO, ITX | K220242 |
| 5. Reference device | Mindray | TE7 | II | IYN, IYO, ITX | K203391 |
| 6. Reference device | Mindray | DC-80A | II | IYN, IYO, ITX | K201693 |
| 7. Reference device | Mindray | ZS3 | II | IYN, IYO, ITX | K192410 |
| 8. Reference device | Mindray | Consona N9 | II | IYN, IYO, ITX | K221300 |
| 9. Reference device | GE | Voluson E10 | II | IYN, IYO, ITX | K181985 |
substantially equivalent to the predicate and reference devices listed below. Resona I9(K216099) is the main predicate device.
The result shows the conformance of subject device to the predicate and reference devices.
Regulation name and code
21 CFR 892.1550 Ultrasonic Pulsed Dopple Imaging System
(IYN)
21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO)
21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)
4. Device Description:
The Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus 19, Resona I9W, Recho 19, Recho 19 Pro, Recho 19 Exp, Recho 19S, Recho 19T Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images in Modes of operation include: B. M. PWD, CWD , Color Doppler, Amplitude Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B. Power+PW+B). Tissue Harmonic Imaging, Smart3D, 4D(Realtime 3D), iScape View, TDI, Color M, Strain Elastography, Contrast imaging (Contrast agent for LVO), V Flow, STE, STQ, Contrast imaging (Contrast agent for Liver).
The Resona 19, Resona 19 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System can also measure anatomical structures and offer analysis packages to provide information based on which the competent health care professionals can make the diagnosis.
{6}------------------------------------------------
5. Intended Use:
Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, intra-operative (includes abdominal, thoracic, and vascular), laparoscopic, pediatric, small organ (breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), Thoracic/pleural, adult and pediatric cardiac, trans-esoph. (Cardiac), peripheral vessel, and urology exams.
Modes of operation include: B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, Smart3D, 4D(Real-time 3D), iScape View, TDI, Color M, Strain Elastography, Contrast imaging (Contrast agent for LVO), V Flow, STE, STQ, Contrast imaging (Contrast agent for Liver).
This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.
6. Comparison with predicate and reference devices:
Subject device Resona I9, Resona I9 Exp. Resona I9T, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa 19 Exp, Nuewa I9 Easi, Anesus 19, Anesus 19 Easi, Eagus 19, Resona I9W, Recho 19, Recho 19 Pro, Recho 19 Exp, Recho 19S, Recho 19T Diagnostic Ultrasound System is comparable with and substantially equivalent to the predicate and reference devices mentioned in 3. Predicate and reference devices with regards to intended use, imaging modes, features and technological characteristics.
- . All systems transmit ultrasonic energy into patients, perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, as well as calculations.
- Subject device Resona I9, Resona I9 Exp. Resona I9S. Resona I9T, Resona I9 . Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi,
{7}------------------------------------------------
Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System has the same intended uses as the predicate device Resona I9 (K210699) except laparoscopic and intra-operative (includes abdominal, thoracic, and vascular) application. At the same time, TE7(K203391) has the laparoscopic and intra-operative (includes abdominal, thoracic, and vascular) applications instead.
- The patient contact materials of the transducers and needle-guided brackets of subject device Resona 19, Resona 19 Exp, Resona I9S, Resona 19T, Resona 19 Easi, Resona I9 Nasa, Resona IV, Imagyn I9S, Imagyn I9S, Imagyn I9 Easi, Nuewa 19, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho 19 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System are the same to the predicate and reference devices.
- The acoustic power levels of Resona I9. Resona I9 Exp. Resona I9S. Resona . I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho 19 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System are below the limits of FDA, which are the same as the predicate device Resona I9 (K210699).
- Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System is designed in compliance with the FDA recognized electrical and physical safety standards, which are the same as the predicate device Resona I9 (K210699).
- . Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa 19 Easi, Anesus 19, Anesus 19 Easi, Eagus 19, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp. Recho I9S, Recho I9T Diagnostic Ultrasound System has the equivalent features and functions as the predicate and reference devices. Among these features, the USAT, HRI+, HiFR CEUS, CEUS Chrono-Parametric Mode, TCMR, 3D-Print Format, Ultra-Micro Angiography, DICOM Pediatric SR, DICOM Urology SR and Biopsy Grid supported in proposed Resona 19 series has been cleared in reference device Resona R9(K222928).
- Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 . Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System has the equivalent features and functions as the
{8}------------------------------------------------
predicate and reference devices. Among these features, the Smart Bladder and ClamAV supported in proposed Resona I9 series has been cleared in reference device Consona N9 (K221300).
- . Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System has the equivalent features and functions as the predicate and reference devices. Among these features, the IOTA supported in proposed Resona I9 series has been cleared in reference device Voluson E10 (K181985).
7. Non-clinical Tests:
Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho 19, Recho 19 Pro, Recho 19 Exp, Recho 19S, Recho 19T Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and this device has been designed to conform with applicable medical safety standards.
This device has been tested and evaluated under the following standards:
- . NEMA UD 2-2004 (R2009), acoustic output measurement standard for diagnostic ultrasound equipment revision 3.
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012. C1:2009/(R)2012 & . A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)].
- . IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
- IEC TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- . IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of
{9}------------------------------------------------
ultrasonic medical diagnostic and monitoring equipment.
- ISO 14971 Third Edition 2019-12, Medical devices - Application of risk management to medical devices.
- ISO 10993-1 Fifth edition 2018-08, biological evaluation of medical devices part 1: evaluation and testing within a risk management process.
- . IEC 62304 Edition 1.1 2015-06, medical device software - software life cycle processes.
- . IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, Medical devices -Part 1: Application of usability engineering to medical devices.
These non-clinical tests relied on in this premarket notification submission can support the determination of substantial equivalence of the subject device.
8. Clinical Studies
Not applicable. The subject of this submission, Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn 19, Imagyn I9S, Imagyn 19 Easi, Nuewa 19, Nuewa 19S, Nuewa 19T, Nuewa 19 Exp, Nuewa 19 Easi, Anesus I9. Anesus I9 Easi, Eagus 19. Resona I9W, Recho 19. Recho 19 Pro, Recho 19 Exp, Recho I9S, Recho I9T Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence.
9. Summary
Based on the performance data as documented in the study, Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa 19 Exp, Nuewa I9 Easi, Anesus I9, Anesus I9 Easi, Eagus I9, Resona I9W, Recho I9, Recho I9 Pro, Recho I9 Exp. Recho I9T Diagnostic Ultrasound System was found to have a safety and effectiveness profile that is similar to the predicate and reference devices.
10.Conclusion:
Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer confirms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the Resona I9, Resona I9 Exp, Resona I9S, Resona I9T, Resona I9 Easi, Resona I9 Nasa, Resona IV, Imagyn I9, Imagyn I9S, Imagyn I9 Easi, Nuewa I9, Nuewa I9S, Nuewa I9T, Nuewa I9 Exp, Nuewa I9 Easi, Anesus I9, Anesus 19 Easi. Eagus 19, Resona I9W, Recho I9, Recho I9 Pro. Recho I9 Exp. Recho
{10}------------------------------------------------
I9S, Recho I9T Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to the predicate and reference devices.
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.