AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Model BB811 and Model BB811-NS: The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

Model BB841: Oxygenator with Integrated Arterial Filter: The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery. Cardiotomy/Venous Reservoir: The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures. The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

Device Description

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (BB811), Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (BB841), and Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (Non-Sterile) (B811-NS) are collectively referred to as the Affinity Fusion Oxygenator in the summary. The Affinity Fusion Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator contains both an integrated arterial filter and integrated heat exchanger. The Affinity Fusion Oxygenator is a microporous, hollow-fiber, gas-exchange devices available with Balance Biosurface bonded to the blood contacting surface of the device. The integrated arterial filter is designed to filter from the circuit microemboli larger than the specified micron size from the circuit for periods up to six hours during cardiopulmonary bypass surgery. Some models of the Affinity Fusion Oxygenator are packaged with an Affinity Fusion Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface which is designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Affinity Fusion CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Fusion CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery.

AI/ML Overview

The provided text discusses the Affinity Fusion Oxygenator and its acceptance criteria, specifically focusing on the addition of an alternate hollow fiber supplier.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance:

Test ItemAcceptance Criteria (Implied: PASS)Reported Device Performance
O2 TransferPASSPASS
CO2 TransferPASSPASS
Blood Side Pressure DropPASSPASS
6 Hour O2 TransferPASSPASS
6 Hour CO2 TransferPASSPASS
Pressure IntegrityPASSPASS*
BurstPASSPASS
Gas Pathway IntegrityPASSPASS
Plasma BreakthroughPASSPASS
Filtration EfficiencyPASSPASS
Hemolysis (Max Flow)PASSPASS
White Blood Cell RetentionPASSPASS
Platelet (PLT) retentionPASSPASS
Functional platelet retentionPASSNA (not specified as tested at this time point)
Platelet PLT Function (min flow)PASSNA (not specified as tested at this time point)
Prime VolumePASSPASS
CoveragePASSPASS
LeachingPASSPASS

Biocompatibility Testing:

Biocompatibility Test MethodTesting CategoryReported Device Performance
ISO MEM Elution Cytotoxicity, L929 Mouse Fibroblast CellsCytotoxicityPASS
ISO Maximization Sensitization Study, 0.9%Sodium Chloride & Sesame Oil, NF extractSensitizationPASS
ISO Intracutaneous Study – 0.9% Sodium Chloride & Sesame Oil, NF extractIntracutaneous ReactivityPASS
ISO Systemic Toxicity Study - 0.9% Sodium Chloride & Sesame Oil, NF extractAcute Systemic ToxicityPASS
Pyrogen Study – Material Mediated -0.9%Sodium ChlorideMaterial Mediated Pyrogen StudyPASS
ASTM In-vitro Hemolysis, Rabbit BloodHemocompatibilityPASS
Complement Activation, SC5b-9, Human SerumHemocompatibilityPASS
ASTM Partial Thromboplastin Time, Direct Contact, Human PlasmaHemocompatibilityPASS
Genotoxicity, Bacterial Reverse Mutation StudyGenotoxicityPASS
Genotoxicity: Mouse Lymphoma AssayGenotoxicityPASS
Platelet and Leukocyte Count- with comparison articleHemocompatibilityPASS

*Note on "Pressure Integrity": Devices conditioned only to factors significant for leaks (per Two Proportion statistical test).

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

The document does not explicitly state the sample sizes for the test sets (number of oxygenator units tested) for design verification or biocompatibility. It also does not specify the country of origin of the data or whether the studies were retrospective or prospective. The nature of the tests (engineering performance and biocompatibility) suggests these were primarily laboratory-based prospective tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

This information is not applicable to the type of device and testing described. The "ground truth" for a medical device like an oxygenator is established through objective performance measurements and standardized biocompatibility tests, not subjective expert assessment as would be the case for image interpretation or diagnosis.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or studies involving subjective human interpretation (e.g., radiology reads) to resolve discrepancies. The tests described here are objective performance and safety tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device is an oxygenator, not an AI-powered diagnostic tool requiring human reader studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This is not applicable. The device is an oxygenator, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

The ground truth for this device is based on:

  • Engineering specifications and standards: For performance characteristics like O2 Transfer, CO2 Transfer, Pressure Drop, Burst, Filtration Efficiency, etc. "PASS" implies meeting predefined, objective measurable criteria.
  • International standards for biocompatibility (ISO 10993-1:2018): For tests like cytotoxicity, sensitization, genotoxicity, hemocompatibility, etc. "PASS" means meeting the established safety thresholds for these biological endpoints.

8. The sample size for the training set:

This is not applicable. This is not an AI/machine learning device that requires a training set. The "study" described is a series of laboratory-based design verification and biocompatibility tests.

9. How the ground truth for the training set was established:

This is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".

July 10, 2024

Medtronic Perfusion Systems Megan Schlichting Regulatory Affairs Manager 7611 Northland Drive Minneapolis, Minnesota 55428

Re: K240666

Trade/Device Name: Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (BB811); Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (BB841); Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface NON STERILE (BB811-NS)

Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary bypass oxygenator Regulatory Class: Class II Product Code: DTZ Dated: June 7, 2024 Received: June 7, 2024

Dear Megan Schlichting:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Nicole M. Gillette -S

Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

Device Name

Affinity FusionTM Oxygenator with Integrated Arterial Filter and BalanceTM Biosurface (BB811); Affinity FusionTM Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with BalanceTM Biosurface (BB841);

Affinity FusionTM Oxygenator with Integrated Arterial Filter and BalanceTM Biosurface NON STERILE (BB811-NS)

Indications for Use (Describe)

Model BB811 and Model BB811-NS

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

Model BB841

Oxygenator with Integrated Arterial Filter

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

Cardiotomy/Venous Reservoir

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures.

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use affer open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

12.0 510(k) Summary

Date Prepared: June 3, 2024

Submitter:Medtronic, Inc.
Medtronic Perfusion Systems
7611 Northland Drive
Minneapolis, MN 55428
Establishment Registration Number: 2184009

Contact Person: Megan Schlichting Regulatory Affairs Manager Medtronic Perfusion Systems Phone: (763) 526-3087 Fax: (763) 514-9521 Email: megan.schlichting@medtronic.com

Alternate Contact: Kimberly Peterson

Regulatory Affairs Director Medtronic Perfusion Systems Phone: (763) 526-6751 Email: kimberly.m.peterson@medtronic.com

{5}------------------------------------------------

12.1 Device Name and Class

ModelTrade nameCommonNameClassification NameClassClassClassificationPanelRegulation(21 CFR)ProductCode
BB811AffinityTM FusionOxygenator withIntegrated Arterial Filterand BalanceTM BiosurfaceOxygenatorCardiopulmonarybypass oxygenatorClassIICardiovascular870.4350DTZ
BB841AffinityTM FusionOxygenator withIntegrated Arterial Filterand Cardiotomy/VenousReservoir with BalanceTMBiosurfaceOxygenatorCardiopulmonarybypass oxygenatorandCardiopulmonarybypass bloodreservoirClassIICardiovascular870.4350870.4260870.4400DTZDTMDTN
BB811-NSAffinity FusionTMOxygenator withIntegrated Arterial Filterand BalanceTM Biosurface(Non-Sterile)OxygenatorCardiopulmonarybypass oxygenatorClassIICardiovascular870.4350DTZ

{6}------------------------------------------------

12.2 Predicate Device Information

Trade NameAffinity FusionTM Oxygenator with Integrated Arterial Filter and BalanceTMBiosurface (BB811)
Affinity FusionTM Oxygenator with Integrated Arterial Filter andCardiotomy/Venous Reservoir with BalanceTM Biosurface (BB841)
Affinity FusionTM Oxygenator with Integrated Arterial Filter and BalanceTMBiosurface (Non-Sterile) (B811-NS)
510(k)K230640
Concurrence DateApril 6, 2023

12.3 Device Description

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (BB811), Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (BB841), and Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (Non-Sterile) (B811-NS) are collectively referred to as the Affinity Fusion Oxygenator in the summary. The Affinity Fusion Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator contains both an integrated arterial filter and integrated heat exchanger. The Affinity Fusion Oxygenator is a microporous, hollow-fiber, gas-exchange devices available with Balance Biosurface bonded to the blood contacting surface of the device. The integrated arterial filter is designed to filter from the circuit microemboli larger than the specified micron size from the circuit for periods up to six hours during cardiopulmonary bypass surgery. Some models of the Affinity Fusion Oxygenator are packaged with an Affinity Fusion Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface which is designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Affinity Fusion CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Fusion CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery.

Principles of Operation 12.4

Affinity Oxygenators are designed to be an integral part of the cardiopulmonary heart bypass circuit for use during cardiac surgery. The Affinity Oxygenators are intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

{7}------------------------------------------------

The oxygenator is composed of the "fiber bundle" which is the gas exchange portion of an oxygenator. The "fiber bundle" is comprised of tiny hollow tubes, or "hollow fibers" that consist of a microporous membrane that allows for gas exchange. Blood flows over the outside surfaces of the hollow fibers while fresh gas is delivered to the inside of the fibers allowing for gas exchange.

Blood that comes from the patient is delivered through a pump to the oxygenator and other auxiliary devices, and back to the patient. The oxygenator can be connected to a heater/cooler device, recirculation circuit, cardioplegia circuit, and the main blood path. The oxygenator is under constant fluid pressure.

There is pressure exerted on the blood-side of the device from the blood pump and patient, the water side of the device due to the flow of the heater-cooler for water, and the gas-side of the device due to the flow of gases through the device. The water-side of the oxygenator is connected to a heater/cooler device to enable temperature control of the blood.

12.5 Indications for Use

Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance™ Biosurface (Model BB811 and BB811-NS)

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance™ Biosurface (Model BB841)

Oxygenator with Integrated Arterial Filter

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

Cardiotomy/Venous Reservoir

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures.

{8}------------------------------------------------

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

12.6 Comparison to Predicate Device

A comparison of the modified product to the currently marketed predicate product Affinity Fusion Oxygenator (K230640) indicates the following similarities:

  • Intended Use / Indications for use ●
  • Contraindications
  • Operating Principle
  • Mechanism of action
  • Performance ●
  • . Shelf Life
  • Packaging configuration and materials
  • . Sterilization process and requirements

When compared to the predicate device. the Affinity Fusion Oxygenator devices presented in this submission have the following differences:

  • addition of an alternate supplier of the hollow fiber for use in the oxygenator fiber ● bundle
    The alternate hollow fiber was previously cleared for use in Medtronic Affinity NT Oxygenator (K191077).

Design verification and biocompatibility testing were completed, and they have led to the conclusion that no newly emerging hazards or risks were identified.

12.6.1 Design Verification testing

The proposed Affinity Fusion Oxygenator has been evaluated through testing to demonstrate that it meets the pressure integrity, burst, and other Special Controls requirements. Test results are presented in Table 12-1. Testing demonstrated results within specification, and therefore the Affinity Fusion Oxygenator manufactured with the hollow fiber supplied by the new supplier is deemed to be substantially equivalent to the predicate Affinity Fusion Oxygenator manufactured with the current hollow fiber supplier. Testing results demonstrate that the performance is comparable to the predicate Affinity Fusion Oxygenator devices (BB811, BB841, and BB811-NS) currently in market.

Test ItemT=0T=2yrAAT=20 months RTA
O2 TransferPASSPASSNA
CO2 TransferPASSPASSNA
Blood Side Pressure DropPASSPASSNA
6 Hour O2 TransferPASSPASSNA

Table 12-1: Summary of Device Testing

{9}------------------------------------------------

6 Hour CO2 TransferPASSPASSNA
Pressure IntegrityPASS*PASS*NA
BurstPASSPASSNA
Gas Pathway IntegrityNANAPASS
Plasma BreakthroughPASSPASSNA
Filtration EfficiencyPASSPASSNA
Hemolysis (Max Flow)PASSPASSNA
White Blood Cell RetentionPASSPASSNA
Platelet (PLT) retentionPASSPASSNA
Functional platelet retentionPASSPASSNA
Platelet PLT Function (min flow)PASSPASSNA
Prime VolumePASSPASSNA
CoveragePASSPASSNA
LeachingPASSPASSNA

*Devices conditioned only to factors significant for leaks (per Two Proportion statistical test)

12.6.2 Biocompatibility

Biocompatibility testing was performed in accordance with ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The Affinity Fusion Oxygenator is categorized as an external communicating device in contact with circulating blood for less than twenty-four (24) hours per ISO 10993-1:2018 and appropriate tests were selected. Test results are presented in Table 12-2. The results of the biological endpoint testing concluded that the Affinity Fusion Oxygenator with the alternate hollow fiber is non-cytotoxic, a non-sensitizer, a non-irritant, non-pyrogenic, non-hemolytic, non-genotoxic, and hemocompatible in regard to complement activation, partial thromboplastin time and platelet & leukocyte count.

{10}------------------------------------------------

Biocompatibility Test MethodTesting CategoryResults
ISO MEM Elution Cytotoxicity, L929 Mouse FibroblastCellsCytotoxicityPASS
ISO Maximization Sensitization Study, 0.9%SodiumChloride & Sesame Oil, NF extractSensitizationPASS
ISO Intracutaneous Study – 0.9% Sodium Chloride &Sesame Oil, NF extractIntracutaneous ReactivityPASS
ISO Systemic Toxicity Study - 0.9% Sodium Chloride &Sesame Oil, NF extractAcute Systemic ToxicityPASS
Pyrogen Study – Material Mediated -0.9%SodiumChlorideMaterial Mediated Pyrogen Study– 0.9% Sodium ChloridePASS
ASTM In-vitro Hemolysis, Rabbit BloodHemocompatibilityPASS
Complement Activation,SC5b-9, Human SerumHemocompatibilityPASS
ASTM Partial Thromboplastin Time, Direct Contact,Human PlasmaHemocompatibilityPASS
Genotoxicity, Bacterial Reverse Mutation StudyGenotoxicityPASS
Genotoxicity: Mouse Lymphoma AssayGenotoxicityPASS
Platelet and Leukocyte Count- with comparison articleHemocompatibilityPASS

Table 12-2: Summary of Biocompatibility Testing

12.7 Conclusion

Medtronic has demonstrated through testing completed that the addition of a second hollow fiber supplier for the Affinity Fusion Oxygenator described in this summary results in a substantially equivalent device because fundamental scientific principle, operating principle, design features and intended use are unchanged.

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”