(443 days)
InferRead Lung CT.AI is comprised of computer assisted reading tools designed to aid the radiologist in the detection of pulmonary nodules ≥ 4mm during the review of CT examinations of the chest on an asymptomatic population ≥ 55 years old. InferRead Lung CT.AI requires that both lungs be in the field of view. InferRead Lung CT.AI provides adjunctive information and is not intended to be used without the original CT series.
InferRead Lung CT.AI uses the deep learning (DL) technology to perform nodule detection. It is a dedicated post-processing application that generates CADe marks as an overlay on original CT scans. The software can be installed in a healthcare facility or a cloud-based platform and is comprised of computer-assisted reading tools designed to aid radiologists in detecting, segmenting, measuring and localizing actionable pulmonary nodules that are 4mm or above during the review of chest CT examinations of asymptomatic populations, with enhanced capabilities for pulmonary nodule follow-up comparison and lung analysis. InferRead Lung CT.AI provides auxiliary information and is not intended to be used if the original CT series is not available.
The provided 510(k) clearance letter and summary discuss the InferRead Lung CT.AI device, its indications for use, and a comparison to predicate devices. It also details some standalone performance tests conducted to assess the newly introduced features of the device.
Here's an analysis of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document details performance for newly added features rather than explicitly defined "acceptance criteria" for the overall device's primary function of nodule detection. However, it states that "predetermined testing criteria" were passed and that "validation tests indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met."
For the newly introduced functions, specific performance metrics are reported:
| Feature Tested | Acceptance Criteria (Implied/Expected) | Reported Device Performance |
|---|---|---|
| Nodule Registration | High accuracy in matching nodule pairs between current and prior scans. | Overall Nodule Match Rate: 0.970 (95%CI: 0.947-0.994) Scan Interval Subgroup: - 0-6 months: 0.976 (95%CI: 0.911-1.0) - 6-12 months: 1.000 (95%CI: N/A) - 12-24 months: 0.938 (95%CI: 0.880-0.997) |
| Nodule Lobe Localization | High accuracy in identifying the correct lung lobe for detected nodules. | Overall Lobe Localization Accuracy Rate: 0.957 (95%CI: 0.929-0.986) |
| Lung Lobe Segmentation | High geometric similarity between automated segmentation and ground truth. | Average Dice Coefficient: 0.966 (95%CI: 0.962 to 0.969) |
2. Sample Sizes Used for the Test Set and Data Provenance
- Nodule Registration Standalone Test: 98 lung cancer screening cases with 206 nodule pairs.
- Nodule Lobe Localization Standalone Test: 94 lung cancer screening scans with 188 nodules.
- Lung Lobe Segmentation Standalone Test: 22 lung cancer screening cases with 110 lung lobes.
Data Provenance: The document does not explicitly state the country of origin for the data used in these tests, nor does it specify if the data was retrospective or prospective. It refers to "lung cancer screening cases/scans," suggesting these are clinical datasets.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not explicitly state the number of experts or their qualifications who established the ground truth for the standalone performance tests.
4. Adjudication Method for the Test Set
The document does not specify the adjudication method used for establishing the ground truth for the test sets in these standalone performance evaluations.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done to assess the improvement of human readers with AI assistance versus without AI assistance. The performance tests described are standalone evaluations of specific AI functions.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the document explicitly describes standalone performance testing for the newly added functions: "For the newly added functions, including nodule registration, nodule localization and lung lobe segmentation, we conducted standalone performance testing." The reported results (Nodule Match Rate, Lobe Localization Accuracy Rate, Dice Coefficient) are all metrics of the algorithm's performance without human interaction.
The document also states: "Regarding the performance of the AI outputs, the nodule detection and segmentation functions were consistent with the predicate product (K192880), as verified through consistency testing." This implies that the primary nodule detection and segmentation capabilities were also assessed in a standalone manner, likely by comparing the AI's output to a ground truth.
7. The Type of Ground Truth Used
The document does not explicitly state the type of ground truth used for the standalone tests (e.g., expert consensus, pathology, outcomes data). However, for metrics like "Nodule Match Rate," "Lobe Localization Accuracy Rate," and "Dice Coefficient," the ground truth would typically be established by expert radiologists or reference standards. For Dice Coefficient in segmentation, it would likely be expert-drawn segmentations.
8. The Sample Size for the Training Set
The document does not provide any information regarding the sample size of the training set used to develop the InferRead Lung CT.AI device.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for the training set was established.
510(k) Clearance Letter for InferRead Lung CT.AI
Page 1
May 16, 2025
lnfervision Medical Technology Co., Ltd.
℅ Matt Deng
Official Correspondent
Room B401, 4th Floor, Building 1,
No.12 Shangdi Information Road, Haidian District
BEIJING, CHINA 100085
Re: K240554
Trade/Device Name: InferRead Lung CT.AI
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: OEB, QIH
Dated: April 3, 2025
Received: April 15, 2025
Dear Matt Deng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K240554 - Matt Deng Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K240554 - Matt Deng Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K240554
Device Name
InferRead Lung CT.AI
Indications for Use (Describe)
InferRead Lung CT.AI is comprised of computer assisted reading tools designed to aid the radiologist in the detection of pulmonary nodules ≥ 4mm during the review of CT examinations of the chest on an asymptomatic population ≥ 55 years old. InferRead Lung CT.AI requires that both lungs be in the field of view. InferRead Lung CT.AI provides adjunctive information and is not intended to be used without the original CT series.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(k) Summary K240554
Infervision Medical Technology Co., Ltd.
This 510(k) Summary is in conformance with 21CFR 807.92
Submitter: Infervision Medical Technology Co., Ltd.
Room B401, 4th Floor, Building 1,
No.12 Shangdi Information Road,
Haidian District, Beijing, 100085
Phone: +86 10-86462323
Primary Contact: Mr. Matt Deng
Email: matt.deng@infervision.ai
Phone: 929-335-4879
Company Contact: Na Li
Regulatory Manager
Date Prepared: May 16, 2025
Device Name and Classification
Trade Name: InferRead Lung CT.AI
Classification: Class II
510(k) Number: K240554
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Common Name: Lung Computed Tomography System, Computer-Aided Detection
Classification Panel: Radiology
Product Code: OEB, QIH
Primary Predicate Device:
Trade Name: InferRead Lung CT.AI
Classification: Class II
510(k) Number: K192880
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Common Name: Lung Computed Tomography System, Computer-Aided Detection
Classification Panel: Radiology
Product Code: OEB, LLZ
Secondary Predicate Device:
Trade Name: Lung Nodule Assessment and Comparison Option (LNA)
Page 6
Classification: Class II
510(k) Number: K162484
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Common Name: System, Image Processing, Radiological
Classification Panel: Radiology
Product Code: LLZ, JAK
Reference device is:
Trade Name: ClearRead CT
Classification: Class II
510(k) Number: K221612
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Common Name: Lung Computed Tomography System, Computer-Aided Detection
Classification Panel: Radiology
Product Code: OEB, LLZ
Device Description
InferRead Lung CT.AI uses the deep learning (DL) technology to perform nodule detection. It is a dedicated post-processing application that generates CADe marks as an overlay on original CT scans. The software can be installed in a healthcare facility or a cloud-based platform and is comprised of computer-assisted reading tools designed to aid radiologists in detecting, segmenting, measuring and localizing actionable pulmonary nodules that are 4mm or above during the review of chest CT examinations of asymptomatic populations, with enhanced capabilities for pulmonary nodule follow-up comparison and lung analysis. InferRead Lung CT.AI provides auxiliary information and is not intended to be used if the original CT series is not available.
Indications for Use
InferRead Lung CT.AI is comprised of computer assisted reading tools designed to aid the radiologist in the detection of pulmonary nodules ≥ 4mm during the review of CT examinations of the chest on an asymptomatic population ≥ 55 years old. InferRead Lung CT.AI requires that both lungs be in the field of view. InferRead Lung CT.AI provides adjunctive information and is not intended to be used without the original CT series.
Substantial Equivalence
Infervision Medical Technology Co., Ltd. asserts that InferRead Lung CT.AI is substantially equivalent in both intended use and technical characteristics to the cited predicate devices. The variations in design and performance compared to the predicate devices do not impact the safety or effectiveness of InferRead Lung CT.AI for its intended use. Table below lists the predicate devices alongside the subject device, including their respective Product Codes and Indications for Use.
Page 7
Detailed Comparison of the Subject and Predicate/Reference Devices
| Item | Subject Device: InferRead Lung CT.AI | Primary Predicate: InferRead Lung CT.AI (K192880) | Secondary Predicate: Lung Nodule Assessment and Comparison Option (LNA)(K162484) | Reference Device: ClearRead CT(K221612) |
|---|---|---|---|---|
| Indications for Use | InferRead Lung CT.AI is comprised of computer assisted reading tools designed to aid the radiologist in the detection of pulmonary nodules ≥ 4mm during the review of CT examinations of the chest on an asymptomatic population ≥ 55 years old. InferRead Lung CT.AI requires that both lungs be in the field of view. InferRead Lung CT.AI provides adjunctive information and is not intended to be used without the original CT series. | InferRead Lung CT.AI is comprised of computer assisted reading tools designed to aid the radiologist in the detection of pulmonary nodules during the review of CT examinations of the chest on an asymptomatic population. InferRead Lung CT.AI requires that both lungs be in the field of view. InferRead Lung CT.AI provides adjunctive information and is not intended to be used without the original CT series. | The Lung Nodule Assessment and Comparison Option is intended for use as a diagnostic patient-imaging tool. It is intended for the review and analysis of thoracic CT images, providing quantitative and characterizing information about nodules in the lung in a single study, or over the time course of several thoracic studies. Characterizations include diameter, volume and volume over time. The system automatically performs the measurements, allowing lung nodules and measurements to be displayed. | ClearRead CT is comprised of computer-assisted reading tools designed to aid the radiologist in the detection and characterization of pulmonary nodules during the review of screening and surveillance (low-dose) CT examinations of the chest on a non-oncological patient population. ClearRead CT requires both lungs be in the field of view and is not intended for monitoring patients undergoing therapy for lung cancer or limited field of view CT scans. ClearRead CT provides adjunctive information and is not intended to be used without the original CT series. |
| Product Code | OEB, QIH | OEB, LLZ | LLZ, JAK | OEB, LLZ |
| User Access Point | Post Processing Application | Post Processing Application | Post Processing Application | Post Processing Application |
| Image Input | DICOM | DICOM | DICOM | DICOM |
| Type of Scans | CT | CT | CT | CT |
| Anatomical Region | Chest | Chest | Chest | Chest |
| Automatically Locate and Identify Lung Nodules | Yes | Yes | Yes | Yes |
| Modifies the Original CT scan | No | No | No | No |
| Automatic calculation of measurements for each segmented nodule | The maximum axial plane longest diameter and shortest diameter, the mean diameter, volume, mean densities information are provided. | The maximum and axial plane longest diameter, mean diameter and volume information are provided. | Short axis- longest diameter perpendicular to the long axis on the slice(mm) Long Axis- Longest diameter on an axial slice (mm) Average\ Max 3D \ Effective diameter (mm) Volume(mm3) Mean densities (HU) | Volume, maximum, minimum, average axial plane diameters |
Page 8
| Temporal Comparison (Nodule Matching) | Yes, automatic | No | Yes, Semi-Automatic | Yes, Fully Automatic |
|---|---|---|---|---|
| Segmentation of lungs and lung lobes | Yes | No | Yes | Yes |
| Nodule location | Yes | No | No | Yes |
| Reporting results | Yes | No | Yes | Yes |
Key features identified that differ from the predicate are discussed below. The newly introduced features have been validated and does not compromise the device's safety and effectiveness.
Nodule location:
This feature displays the lobar location of each detected pulmonary nodule. It relies on the same nodule-detection algorithm used in predicate device K192880 and serves to enhance physicians' ability to accurately localize detected nodules.
Nodule density:
Compared with predicate device K192880, this product additionally provides nodule density information for detected nodules. Nodule density is objectively quantified based on CT-derived Hounsfield Unit (HU) values.
Nodule registration between prior and current scan:
If prior scans are available for a case, the software will automatically match identified nodules between the current series and the prior series, and display each nodule's change status as one of the following: "Increased, Decreased New Disappeared, No Change". Physicians may manually adjust the matching results if needed.
Performance testing:
Infervision Medical Technology Co., Ltd. has assessed and tested the product and has passed all predetermined testing criteria. The plans for validation tests were designed to evaluate outputs by InferRead CT Lung.AI and followed the procedures documented in the validation test plan. Validation tests indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
Page 9
Regarding the performance of the AI outputs, the nodule detection and segmentation functions were consistent with the predicate product (K192880), as verified through consistency testing. For the newly added functions, including nodule registration, nodule localization and lung lobe segmentation, we conducted standalone performance testing. The detailed results are as follows:
Nodule Registration Standalone Performance Test
InferRead Lung CT.AI was evaluated in a Nodule Registration standalone test that contained 98 lung cancers screening cases with 206 nodule pairs. InferRead Lung CT.AI achieved an overall nodule Match Rate of 0.970 (95%CI: 0.947-0.994). In the scan interval subgroup analysis, the nodule match rates were as follows: 0.976 (95%CI: 0.911-1.0) for the 0-6 month interval, 1.000 (95%CI: N/A) for the 6-12 month interval, and 0.938 (95%CI: 0.880-0.997) for the 12-24 month interval.
Nodule Lobe Localization Standalone Performance Test
InferRead Lung CT.AI was evaluated in a Nodule Lobe Localization standalone test that contained 94 lung cancer screening scans with 188 nodules. InferRead Lung CT.AI achieved an overall Lobe Localization Accuracy Rate of 0.957 (95%CI: 0.929-0.986).
Lung Lobe Segmentation Standalone Performance Test
InferRead Lung CT.AI was evaluated in a Lung Lobe Segmentation standalone test that contained 22 lung cancer screening cases with 110 lung lobes. The test results for pulmonary lobe segmentation showed that the average Dice Coefficient was 0.966 (95%CI: 0.962 to 0.969).
Other Non-Clinical Testing Summary
InferRead Lung CT.AI was designed and developed by Infervision Medical Technology Co., Ltd. in accordance with applicable requirements, design controls, and relevant standards. It has demonstrated substantial equivalence to the predicate device. Non-clinical testing has demonstrated that InferRead Lung CT.AI complies with the following FDA-recognized consensus standards:
- IEC 62304:2006+A1:2015
- ISO 14971:2019
- AAMI TIR57:2016
The substantial equivalence of InferRead Lung CT.AI has been demonstrated through verification and validation testing in accordance with applicable specifications, acceptance criteria, and performance standards. The traceability analysis provides traceability between the requirement specifications, design specifications, risks, and verification testing of the subject device. All requirements and risk controls have been successfully verified and traced. A comprehensive risk analysis was performed for the subject device and appropriate risk controls have been implemented to mitigate hazards.
Page 10
Software verification and validation activities were conducted in accordance with IEC 62304:2006+A1:2015 – Medical device software – Software lifecycle processes and ISO 14971:2019 Medical devices – Application of risk management to medical devices, and in accordance with relevant FDA guidance documents, Guidance for the Content of Premarket Submissions for Device Software Functions (issued June 14, 2023), and Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (issued September 27, 2023).
Cybersecurity and vulnerability analyses were conducted, and it has been determined that InferRead Lung CT.AI conforms to the cybersecurity requirements.
The following processes were followed and applied during the design and development of InferRead Lung CT.AI:
- Risk Analysis
- Unit Testing
- Integration Testing
- System Testing
- Performance Testing
- Software Verification & Validation
- Cybersecurity Testing and Analysis
InferRead Lung CT.AI was tested and was found to be substantially equivalent for the intended use, intended users, intended patient population, and use environments, as demonstrated through verification and validation testing evaluating its clinical usage and performance. Validation testing was performed to ensure that the final product meets the requirements for the specified clinical application and performs as intended to meet users' needs, while demonstrating substantial equivalence to the predicate device.
Conclusion
In preparing this 510(k) submission, Infervision Medical Technology Co., Ltd. has meticulously considered the relevant statutory and regulatory requirements and believes that the information provided herein meets the criteria for demonstrating substantial equivalence in terms of design features, fundamental technology, indications for use, and the safety and effectiveness of the device. Furthermore, verification and validation testing have been conducted to confirm that the device meets its intended use and specifications when operated as intended.
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).