K Number
K240430
Device Name
Otoport Pro
Manufacturer
Date Cleared
2024-03-15

(30 days)

Product Code
Regulation Number
874.1050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Otoport Pro device is indicated for use when there is a requirement to screen for hearing disorders by objective and non-invasive means. ABR, TEOAE and DPOAE screening test results are automatically interpreted and a clear "Pass' or 'Refer' result is presented to the device is indicated when the patient is unable to give reliable voluntary responses to sound, especially with infants. Use of the device facilitates the early detection of hearing loss and its characterization. Where the individual to be screened is healthy with no medical conditions related to the ear, as in the case of well-baby hearing screening, the user can be a trained screener. In all other cases the user should be an audiologist or medical professional. The TEOAE and DPOAE analytical functions of the device are indicated when objective non-invasive clinical investigations require the characterization and monitoring of the peripheral auditory function. For this purpose, the device is intended to be used by audiologists or other professionals skilled in audiology. These TEOAE and DPOAE tests are applicable to populations of any age to obtain objective evidence of peripheral auditory function.
Device Description
The Otodynamics Ltd ("Otodynamics") Otoport Pro device is a compact handheld device capable of high quality OAE measurements for clinical purposes and also automated ABR and OAE testing for fast infant screening. Responses to sound are recorded via an applied earphone and or adhesive surface electrode pad. Specifically, the device can record Otoacoustic Emissions (type DPOAEs or TEOAEs) and auditory brainstem responses (ABRs) to sound. These responses are especially useful in the hearing of infants for deafness. The more detailed analysis of DPOAE and TEOAE responses is additionally useful as a component of the audiological diagnostic test battery. The Otoport Pro is simple to use with customizable automation to make testing easy and the results clear. It has user access controls, graphical display panel, and extensive test database features. The Otoport Pro when configured for clinical use has advanced test features including extensive raw data capture for offline review and analysis if required. Otoport Pro device is a hardware/ software revision of the currently marketed Otoport OAE+ABR, having the same performance and intended uses as the Otoport OAE+ABR device.
More Information

Otoport/Otocheck OAE+ABR™, K143395

Not Found

No
The summary explicitly states "Mentions AI, DNN, or ML: Not Found" and the description focuses on standard signal processing and automated interpretation based on pre-defined criteria ("automatically interpreted and a clear 'Pass' or 'Refer' result is presented").

No
This device is designed to screen for hearing disorders and characterize and monitor peripheral auditory function, which falls under diagnostic and screening purposes rather than direct treatment or therapy.

Yes.

The device is explicitly stated to "facilitate the early detection of hearing loss and its characterization" and to be used for "clinical investigations [to] require the characterization and monitoring of the peripheral auditory function." The results are interpreted as "Pass" or "Refer," indicating it aids in identifying a condition.

No

The device description explicitly states it is a "compact handheld device" and records responses via an "applied earphone and or adhesive surface electrode pad," indicating it includes hardware components for data acquisition.

Based on the provided information, this device is not an In Vitro Diagnostic (IVD).

Here's why:

  • IVDs analyze biological samples: In Vitro Diagnostics are designed to examine specimens taken from the human body, such as blood, urine, tissue, etc., to provide information about a person's health.
  • This device measures physiological responses: The Otoport Pro device measures auditory responses (OAEs and ABRs) to sound stimuli. These are physiological responses from the patient's auditory system, not the analysis of a biological sample.

The device is a diagnostic tool used to assess hearing function, but it does so by interacting directly with the patient's body and recording their responses, not by analyzing a sample taken from them.

N/A

Intended Use / Indications for Use

The Otoport Pro device is indicated for use when there is a requirement to screen for hearing disorders by objective and non-invasive means. ABR, TEOAE and DPOAE screening test results are automatically interpreted and a clear "Pass' or 'Refer' result is presented to the device is indicated when the patient is unable to give reliable voluntary responses to sound, especially with infants. Use of the device facilitates the early detection of hearing loss and its characterization.

Where the individual to be screened is healthy with no medical conditions related to the ear, as in the case of well-baby hearing screening, the user can be a trained screener. In all other cases the user should be an audiologist or medical professional.

The TEOAE and DPOAE analytical functions of the device are indicated when objective non-invasive clinical investigations require the characterization and monitoring of the peripheral auditory function. For this purpose, the device is intended to be used by audiologists or other professionals skilled in audiology. These TEOAE and DPOAE tests are applicable to populations of any age to obtain objective evidence of peripheral auditory function.

Product codes (comma separated list FDA assigned to the subject device)

EWO, GWJ

Device Description

The Otodynamics Ltd ("Otodynamics") Otoport Pro device is a compact handheld device capable of high quality OAE measurements for clinical purposes and also automated ABR and OAE testing for fast infant screening.

Responses to sound are recorded via an applied earphone and or adhesive surface electrode pad. Specifically, the device can record Otoacoustic Emissions (type DPOAEs or TEOAEs) and auditory brainstem responses (ABRs) to sound. These responses are especially useful in the hearing of infants for deafness. The more detailed analysis of DPOAE and TEOAE responses is additionally useful as a component of the audiological diagnostic test battery.

The Otoport Pro is simple to use with customizable automation to make testing easy and the results clear. It has user access controls, graphical display panel, and extensive test database features. The Otoport Pro when configured for clinical use has advanced test features including extensive raw data capture for offline review and analysis if required.

Otoport Pro device is a hardware/ software revision of the currently marketed Otoport OAE+ABR, having the same performance and intended uses as the Otoport OAE+ABR device.

The previous version of the Otoport product (Otoport OAE+ABR) required an adaptor 'shell' to be fitted over the base unit if AABR functionality was required (Figure 1 below)

In the Otoport Pro design (Figure 2 below) all test modalities are included inside the base unit.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Scalp, forehead, nape, shoulder

Indicated Patient Age Range

population of any age, Infant and adult (Infant only for AABR)

Intended User / Care Setting

Where the individual to be screened is healthy with no medical conditions related to the ear, as in the case of well-baby hearing screening, the user can be a trained screener. In all other cases the user should be an audiologist or medical professional.

For clinical investigations, the device is intended to be used by audiologists or other professionals skilled in audiology.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Clinical measurement comparison: AABR, TEOAE and DPOAE measurements were made on a cohort of adult volunteers giving informed consent, using both the Otoport Pro and the predicate Otoport OAE+ABR, using the same probe which remained in position for each comparison. Tests were conducted in a sound booth by a qualified audiologist.

TEOAE screening: 17 subjects
DPOAE screening: 16 subjects
AABR screening: 14 subjects

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-Clinical Performance Testing:
Electrical safety and electromagnetic compatibility (EMC): The system complies with ANSI/AAMI/IEC ES60601-1:2005/(R)2012 and Al:2012. Electromagnetic compatibility of the system confirmed through third-party testing to collateral standard IEC 60601-1-2 Edition 4.0 2014-02.
Software Verification and Validation Testing: Conducted in compliance with IEC 62304 Edition 1.1 2015-06, ANSI/AAMI/IEC ES60601-1:2005//R/2012, and FDA guidance documents "General Principles of Software Validation; Final Guidance for Industry and FDA Staff" (11-Jan-2002) and "FDA Guidance for the Content of Premarket Submissions for Device Software Functions" (14-June-2023).
Cybersecurity: Issues regarding cybersecurity are included for resolution at all appropriate phases of the products development process. Cybersecurity addressed in the Software Risk mitigation measures against cybersecurity risks are incorporated into the software requirements specifications. The Otodynamics products/ services considered and followed the recommendations in "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, Guidance for Industry and Food and Drug Administration Staff (27-September- 2023)" and current medical devices cybersecurity regulations.

Clinical Performance Testing:
Measurements of the acoustic stimulus and recorded auditory response:

  • Measurement by a calibrated microphone of the amplitude and waveform of the acoustic stimulus delivered into 1. a calibrated ear simulator after level setting by the Otoport OAE+ABR, in TEOAE, DPOAE and AABR testing modes, using the same probe on each device.
  • Measurement by the device of the response recorded by both Otoport OAE+ABR in TEOAE, 2. DPOAE and AABR testing modes, when presented with an electronically generated from the device's acoustic stimulus (i.e., simulated) TEOAE, DPOAE or ABR response signal respectively.

Results:
Acoustic stimulus differences between devices (using the same probe):

  • TEOAE differences less than 0.4dB
  • DPOAE an average difference of less than 0.3dB from 1-8kHz, with a maximum deviation of 0.8dB
  • AABR an average difference of 0.2dB from 2-4.5kHz

Received response differences:

  • TEOAE average of 0.23 dBSPL
  • DPOAE average difference 0.4dB, with largest differences less than 2dB
  • AABR average difference 0.1dB

Clinical measurement comparison (on adult volunteers):

  • TEOAE screening (click stimulus at 84dBSPLpe, 6dBSNR pass criteria in 2 half octave bands): 17 subjects tested, 15 passed on both instruments, the same 2 failed on both.
  • DPOAE screening (stimulus levels L1 65 dBSPL for f1 and f2 tones at half octave intervals from 1.5 to 6kHz): 16 subjects tested, 13 subjects passed the default (newborn) screening criteria on both instruments, 3 failed on both instruments, and these subjects failed at the same frequency on each instrument.
  • AABR screening (chirp of 40dBHL audiologically calibrated on 25 healthy young adults): 14 subjects tested, 12 subjects passed on both instruments, the same 2 subjects failed on both instruments (with no clear response and high EEG). Test times were similar on both instruments at about 100 seconds. Average pass ABR amplitudes were 0.49 microV for Otoport Pro and 0.51 microV for the predicate. Both devices reported the same electrode impedances within 240 ohms.

Clinical (measurement) mode:

  • TEOAE: Difference between mean TEOAE levels reported by the two devices was less than 1dB at each half octave frequency (1k to 6kHz).
  • DPOAE: 13-point DPqrams recorded. Averaged across 15 subjects, the mean DPOAE level difference was less than 1dB at each frequency (1k-8kHz).

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Otoport/Otocheck OAE+ABR™, K143395

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 874.1050 Audiometer.

(a)
Identification. An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.(b)
Classification. Class II. Except for the otoacoustic emission device, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, if it is in compliance with American National Standard Institute S3.6-1996, “Specification for Audiometers,” and subject to the limitations in § 874.9.

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

March 15, 2024

Otodynamics Ltd % Chandler Thames Director of Quality Rook Quality Systems 1155 Mount Vernon Hwy #800 Dunwoody, Georgia 30338

Re: K240430

Trade/Device Name: Otoport Pro Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO, GWJ Dated: February 12, 2024 Received: February 14, 2024

Dear Chandler Thames:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

1

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shuchen Peng -S

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

2

Indications for Use

510(k) Number (if known) K240430

Device Name Otoport Pro

Indications for Use (Describe)

The Otoport Pro device is indicated for use when there is a requirement to screen for hearing disorders by objective and non-invasive means. ABR, TEOAE and DPOAE screening test results are automatically interpreted and a clear "Pass' or 'Refer' result is presented to the device is indicated when the patient is unable to give reliable voluntary responses to sound, especially with infants. Use of the device facilitates the early detection of hearing loss and its characterization.

Where the individual to be screened is healthy with no medical conditions related to the ear, as in the case of well-baby hearing screening, the user can be a trained screener. In all other cases the user should be an audiologist or medical professional.

The TEOAE and DPOAE analytical functions of the device are indicated when objective non-invasive clinical investigations require the characterization and monitoring of the peripheral auditory function. For this purpose, the device is intended to be used by audiologists or other professionals skilled in audiology. These TEOAE and DPOAE tests are applicable to populations of any age to obtain objective evidence of peripheral auditory function.

Type of Use ( Select one or both, as applicable )
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the

of information is estimated to average time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

SUBMITTER

Otodynamics Ltd 30-38 Beaconsfield Rd Hatfield, Hertfordshire ALI0 8BB, United Kingdom

Phone: +44 1707 267540 Email: daniel.budd@otodynamics.com

Contact Person: Daniel Budd, QA Manager Date Prepared: 15.03.2024

DEVICE

Trade Name:Otoport Pro
Common Name:Otoport Pro
Classification Name:Audiometer/ Evoked response auditory stimulator (874.1050 & 882.1900)
Regulatory Class:Class II
Product Codes:EWO &GWJ

PREDICATE DEVICE

The legally marketed device to which equivalence is being claimed [807.92(a)(3)] is the Otoport/Otocheck OAE+ABR™, K143395.

This predicate has not been the subject of any design-related recalls.

DEVICE DESCRIPTION

The Otodynamics Ltd ("Otodynamics") Otoport Pro device is a compact handheld device capable of high quality OAE measurements for clinical purposes and also automated ABR and OAE testing for fast infant screening.

Responses to sound are recorded via an applied earphone and or adhesive surface electrode pad. Specifically, the device can record Otoacoustic Emissions (type DPOAEs or TEOAEs) and auditory brainstem responses (ABRs) to sound. These responses are especially useful in the hearing of infants for deafness. The more detailed analysis of DPOAE and TEOAE responses is additionally useful as a component of the audiological diagnostic test battery.

The Otoport Pro is simple to use with customizable automation to make testing easy and the results clear. It has user access controls, graphical display panel, and extensive test database features. The Otoport Pro when configured for clinical use has advanced test features including extensive raw data capture for offline review and analysis if required.

Otoport Pro device is a hardware/ software revision of the currently marketed Otoport OAE+ABR, having the same performance and intended uses as the Otoport OAE+ABR device.

The previous version of the Otoport product (Otoport OAE+ABR) required an adaptor 'shell' to be fitted over the base unit if AABR functionality was required (Figure 1 below)

In the Otoport Pro design (Figure 2 below) all test modalities are included inside the base unit.

4

Image /page/4/Picture/0 description: The image shows a handheld Otodynamics device being held by two hands. The device has a screen at the top, followed by a series of buttons and a keypad. The keypad has numbers 1-9 and 0, along with other symbols. The device is white and has the brand name "Otodynamics" printed on it.

Image /page/4/Figure/1 description: The image shows a handheld device with a screen displaying a graph labeled "DPZ TEST". The graph shows a red line fluctuating between -10 and 20 dBSPL across a frequency range of 2 to 8 kHz. Below the screen, there are several buttons, including a numeric keypad from 1 to 9, a 0 button, and other function buttons labeled "CANCEL", "MANUAL", and "END".

Figure 2

INDICATIONS FOR USE

Figure 1

The Otoport Pro device is indicated for use when there is a requirement to screen for hearing disorders by objective and non-invasive means. ABR, TEOAE and DPOAE screening test results are automatically interpreted and a clear 'Pass' or 'Refer' result is presented to the user. Use of the device is indicated when the patient is unable to give reliable voluntary responses to sound, especially with infants. Use of the device facilitates the early detection of hearing loss and its characterization. Where the individual to be screened is healthy with no medical conditions related to the ear, as in the case of well-baby hearing screening, the user can be trained. In all other cases the user should be an audiologist or medical professional. The TEOAE and DPOAE analytical functions of the device are indicated when objective non-invasive clinical investigations require the characterization and monitoring of the functional status of the peripheral auditory function. For this purpose, the device is intended to be used by audiologists or other professionals skilled in audiology. These TEOAE and DPOAE tests are applicable to populations of any age to obtain objective evidence of peripheral auditory function.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The same Otoacoustic Emissions (OAEs) & Auditory Brainstem Response (ABR) techniques and principles of response analysis are utilized in both the subject and predicate devices.

5

TABLE 1. COMPARISON BETWEEN THE SUBJECT DEVICE AND THE PREDICATE OTOPORT/OTOCHECK 0AE+ABR (K143395)

TABLE 1. COMPARISON BETWEEN THE SUBJECT DEVICE AND THE PREDICATE OTOPORT/OTOCHECK 0AE+ABR (K143395)
Subject Device
Otoport Pro (K240430)Predicate Device
Otoport/Otocheck OAE+ABR
(K143395)Overview of similarities
and differences
Common NameAudiometer
Evoked response auditory stimulatorAudiometer
Evoked response auditory stimulatorSame
ClassIIIISame
Product CodeEWO&GWJEWO&GWJSame
Classification
Regulation874.1050
882.1900874.1050
882.1900Same
Intended UseFor the detection of peripheral
hearing loss and for use in the
characterization of and diagnosis of
auditory and hearing-related
disorders.For the detection of peripheral
hearing loss and for use in the
characterization of and diagnosis of
auditory and hearing-related
disorders.Same
Indications for UseThe Otoport Pro device is indicated for
use when there is a requirement to
screen for hearing disorders by
objective and non-invasive means.
ABR, TEOAE and DPOAE screening test
results are automatically interpreted
and a clear 'Pass' or 'Refer' result is
presented to the user.The Otoport/Otocheck OAE+ABR
device is indicated for use when there
is a requirement to screen for hearing
disorders by objective and non-
invasive means. Test results are
automatically interpreted and a clear
'Pass' or 'Refer' result is presented to
the user.Same
Use of the device is indicated when
the patient is unable to give reliable
voluntary responses to sound,
especially with infants.Use of the device is indicated when
the patient is unable to give reliable
voluntary responses to sound,
especially with infants.
Use of the device facilitates the early
detection of hearing loss and its
characterization.Use of the device facilitates the early
detection of hearing loss and its
characterization.
Where the individual to be screened is
healthy with no medical conditions
related to the ear, as in the case of
well-baby hearing screening, the user
can be a trained screener. In all other
cases the user should be an
audiologist or medical professional.Where the individual to be screened is
healthy with no medical conditions
related to the ear, as in the case of
well-baby hearing screening, the user
can be a trained screener. In all other
cases the user should be an
audiologist or medical professional.
The TEOAE and DPOAE analytical
functions of the device are indicated
when objective non-invasive clinical
investigations require the
characterization and monitoring of the
functional status of the peripheral
auditory function. For this purpose,
the device is intended to be used by
audiologists or other professionals
skilled in audiology. These TEOAE
and DPOAE tests are applicable to
populations of any age to obtain
objective evidence of peripheral
auditory functionThe TEOAE and DPOAE analytical
functions of the device are indicated
when objective non-invasive clinical
investigations require the
characterization and monitoring of the
functional status of the peripheral
auditory function. For this purpose,
the device is intended to be used by
audiologists or other professionals
skilled in audiology. These TEOAE
and DPOAE tests are applicable to
populations of any age to obtain
objective evidence of peripheral
auditory function
Overall Device DesignHandheld device (L195mm x W70mm x D2.8mm; 214g)
(7 3/4" x 2 3/4"x 11/8" 0.47pounds)
Stimulates the ear with quiet sounds.
Non-invasive, records physiological
responses to these sounds.
Automatically evaluates the evidence
for a response.
Reports a concise result output about
the presence or absence of normal
responses.
Saves a detailed record of the test.
Device alerts to the possibility of
hearing loss and documents the status
of the ear.Handheld device (L280mm x W83mm x D45mm; 446g)
(11" x 3 1/4" x 13/4" 1 pound)
Stimulates the ear with quiet sounds.
Non-invasive, records physiological
responses to these sounds.
Automatically evaluates the evidence
for a response.
Reports a concise result output about
the presence or absence of normal
responses.
Saves a detailed record of the test.
Device alerts to the possibility of
hearing loss and documents the status
of the ear.Otoport Pro is smaller and
lighter.
The predicate Otoport
OAE+ABR was heavier and
larger because it incorporated
an adaptor 'shell', fitted over the
basic Otoport OAE product, to
provide AABR functionality. In
the Otoport Pro subject device,
the AABR functionality has been
built into the basic handheld
device as the attachment of an
ABR adaptor shell is not
required.
Device
Components/
FeaturesStimulates the ear with either
transient or tonal sounds via an
inserted earpieceStimulates the ear with either
transient or tonal sounds via an
inserted earpieceSame
Registers interfering acoustic noise
and advises users when this is too
large for testing to take place.Registers interfering acoustic noise and
advises users when this is too large for
testing to take place.Same
Implements otoacoustic emission
technology to record TEOAE and
DPOAEs responses to sound applied
via an inserted earpiece (probe)Implements otoacoustic emission
technology to record TEOAE and
DPOAEs responses to sound applied
via an inserted earpiece (probe)Same
Provides an intensity and frequency
analysis of the OAE response,
numerically and graphicallyProvides an intensity and frequency
analysis of the OAE response,
numerically and graphicallySame
Automatically determines and displays
the presence or absence of a normal
TEOAE or DPOAE responseAutomatically determines and displays
the presence or absence of a normal
TEOAE or DPOAE responseSame
Implements auditory brainstem
response (ABR) technology to record
electro physiological responses via
surface electrodesImplements auditory brainstem
response (ABR) technology to record
electro physiological responses via
surface electrodesSame
Stimulates the ear for ABR screening
at 30,35,40, or 45dBnHL (60dBHL for
training)Stimulates the ear for ABR screening
at 30,35,40, or 45dBnHL (60dBHL for
training)Same
Registers interfering electrical noise
and advises users when this is too
large for testing to take place.Registers interfering electrical noise
and advises users when this is too
large for testing to take place.Same
Registers the electrical impedance of
electrode connections to the patient
and advises users when this is too
large for testing to take placeRegisters the electrical impedance of
electrode connections to the patient
and advises users when this is too
large for testing to take placeSame
Performs statistical analysis on the
ABR response to decide if it is present
and normalPerforms statistical analysis on the
ABR response to decide if it is present
and normalSame
Automatically determines and displays
the presence or absence of a normal
ABR responseAutomatically determines and displays
the presence or absence of a normal
ABR responseSame
Serves as a hearing based on the
presence or
absence of a normal OAE or ABR
responseServes as a hearing based on the
presence or
absence of a normal OAE or ABR
responseSame
PopulationInfant and adult (Infant only for
AABR)Infant and adult (Infant only for AABR)Same
Intended userTrained operators for screening, and
professionals for screening and clinical
useTrained operators for screening, and
professionals for screening and clinical
useSame
Safety
characteristicsElectrically isolated from any device
connected to its UCB-C charging and
data connector.
Acoustic stimulation levels physically
constrained and monitored with active
protection against single faults.Electrically isolated from any device
connector to its Hirose charging and
data connector.
Acoustic stimulation levels physically
constrained and monitored with active
protection against single faults.Same
Enclosure materialThe device enclosure materials are
Polylac PA-765 Flame retardant ABS.
Injection moulding.The device enclosure materials are
Polylac PA-765 Flame retardant ABS.
Injection moulding.Same materials. Comply with
the current biocompatibility
standards.
Materials of
ConstructionProbe: plastic encapsulated insert
earphone containing microphone and
receiver, fitted with disposable single
use plastic tipProbe: plastic encapsulated insert
earphone containing microphone and
receiver, fitted with disposable single
use plastic tipSame
For ABR screening - Surface, self-
adhesive skin electrodes.
Electrode cables: PVC (Biocompatible)For ABR screening - Surface, self-
adhesive skin electrodes
Electrode cables: PVC; Nylon
(Biocompatible)Same materials. Comply with
the biocompatibility standards.
Electronic circuitry providing
stimulation drive to the probe,
amplification for the signals, received
from the probe microphone and
electrodes, signal processing for
response enhancement and noise
rejection, microprocessor for signal
and statistical analysis, all
encapsulated in a plastic housing,
including graphic display unit and data
input facilities plus rechargeable
battery.Electronic circuitry providing stimulation
drive to the probe, amplification for the
signals, received from the probe
microphone and electrodes, signal
processing for response enhancement
and noise rejection, microprocessor for
signal and statistical analysis, all
encapsulated in a plastic housing,
including graphic display unit and data
input facilities plus rechargeable
battery.Similar. Individual circuit
components have been changed
to avoid obsolescence with
equivalent or superior
performance items.
SterileNoNoSame
Duration of
placementLess than 5 minutes per earLess than 5 minutes per earSame
Usage of patient
contact itemsAcoustic probe tip and electrodes are
single patient use onlyAcoustic probe tip and electrodes are
single patient use onlySame
Anatomical
placementScalp, forehead, nape, shoulderScalp, forehead, nape, shoulderSame
Number of
Electrodes33Same
Electrode SocketsYellow, Touchproof Connector, SSB,
4-pin, 1mm. PVC. BiocompatibleDIN 42-802 Type ST 'Touchproof'
terminated electrode cables. Electrode
cable connectors Nylon: ZYTEL
70G33L.
BiocompatibleSimilar
The same Plastics One
manufacturer of the connector.
Both types are for EEG cables/
electrodes.
Otoport Pro has all 3 electrode
cables in one single connector.
Otoport ABR predicate has
individual connectors for each of
the 3 electrode cables.
The connectors for both Otoport
Pro and predicate Otoport ABR
are biocompatible and
substantially equivalent in
functionality and performance.
BatteryLithium-ion BatteryLithium-ion Batterysame
Battery voltage
operating
range:3.2-4.2V3.2-4.2Vsame
SourceSource: 1000mAh lithium polymer
internal
rechargeable cellsSource: 1000mAh lithium polymer
internal
rechargeable cellssame
Electrode
CableGray, Tinsel Wire, 3 x 0.023 conductor,
Shielded,
0.120 Dia. PVC jacket. Biocompatible.Electrode Cable Insulations: PVC,
Mexichem
2235L-75. Biocompatible.Similar. The Otoport Pro cable
has all 3 electrode cables inside
one shield and also ground
shielding on most of the length.
The Otoport Pro electrode
cables with single connector to
device run to 3 cable ends-
electrode connections to patient
(same with Otoport ABR).
Otoport ABR electrode cables
are 3 individual electrode cables
running from device to patient.
The electrode cables for both
Otoport Pro and predicate
Otoport ABR are biocompatible
and substantially equivalent in
functionality and performance.
Probe ConnectorTwo (14 pin) 'Medi-Snap' connector
push pull connector'Triad' (8 pin) connector metal collar to
screw lock in place. Binaural option has
two such probe connectorsSimilar. The probe connector
for the subject device is a 14
pin "Medi-Snap" connector
while the predicate device
uses an 8 pin "Triad"
connector. The Otoport Pro
probes are substantially
equivalent with legacy probes
of predicate Otoport ABR,
which have been in service for
20+years. The only difference
is the connectors.
The 14 pin Medi-Snap
connector of the Otoport Pro
serves the same purpose as
the predicate, but it's insertion
and removal are mechanically
different. Instead of a screwed
collar locking method, the
Medi-Snap features a Push-
in/Pull-out mechanism with
adequate retention when in.
This provides an easier user
experience and reduces risk of
connector damage by
mishandling. Connector
damage would be detected by
recommended QA tests. The
Mini-Snap further improves
safety by not exposing metal
parts to user. The residual user
risks remain low/ are
acceptable.
ProbesUPS, UPD & XPD probes
TEOAE/ DPOAE Functionality probes.All models take UGS & UGD probes.
Can also take the XGD probe with a
firmware update.
TEOAE/ DPOAE Functionality probes.Similar. Substantially equivalent
TEOAE/ DPOAE probes. The
subject device is compatible
with all the probe types useable
with the predicate device when
supplied with a Medi-Snap
connector (UPS, UPD and
XPD).
Differences in the probe
connectors to the device:
Predicate has UGS/ UGD/ XGD
probes with metallic screw
connectors to the Otoport.
Otoport Pro has UPS/UPD &
XPD probes with plastic clip-on
connectors to the device
(easier to use).
Number of
Probes
ConnectedThe Otoport Pro can accept 2 probesThe Otoport OAE+ABR binaural option
can accept 2 probes, otherwise only
1 probeThe subject and predicate base
models (Otoport Pro and
Otoport) are similar in that both
can accept 2 probes for
binaural recording. With the
predicate 2 probes was only
available with the binaural
OAE+ABR variant. With the
subject device, 2 probes
capacity is standard.
Electrode
Socket
ConnectorOtoport Pro has one 4 Pin Plug &
Socket ConnectorThree Single Pin Plug and Socket
ConnectorsThe Otoport Pro and the
predicate device are similar that
they require 3 electrode
connections to perform the ABR
screening test.
The Otoport Pro uses a single 4
pin plug and socket connector,
while the predicate uses three
single pin plug and socket
connectors. The electrode
cabling is different and incudes
an EMC shield, which reduce
interference noise.
The 4-pin single electrode cable
socket of the port Pro provides a
Charging and data
transfer connectorUSB-C for charging and downloadHirose proprietary connector for
charging and downloadSimilar.
Improved useability and
reliability.
Device
Components /
FeaturesA handheld battery powered hearing
screening device with integral keypad
and display which shows processed
real time data. OAE/ ABR
Functionality.A handheld battery powered hearing
screening device with integral
keypad and display which shows
processed real time data. For the
ABR functionality the ABR module
attachment fitted on the Otoport is
required. OAE/ ABR Functionality.In the Otoport Pro ABR
functionality is integral to the
basic unit. No adaptor is
needed to enable ABR
screening functionality. This
enhances usability and
convenience- simplicity.
Residual risks associated with
ABR adaptor shell fitting are
removed. The residual risks
are found to be low/ are
acceptable.
User InterfaceThe Otoport Pro preinstalled software
provides complete user control over the
testing hardware inside the Otoport Pro
and displays test progress, test results.,
and tests saved in the in-built database.
Optionally, Otoport Pro database
contents can be securely transferred to
Otodynamics Otolink software for
review, printing, or archiving.
Otoport Pro firmware can be securely
updated from OtolinkThe Otoport preinstalled software
provides complete user control over
the testing hardware inside the
Otoport and displays test progress,
test results, and tests saved in the
in-built database. Optionally, Otoport
database contents can be securely
transferred to Otodynamics Otolink
software for review, printing, or
archiving.
Otoport firmware can be securely
updated from Otolink"Similar.
Products usability engineering
compliance to current
standards/ regulations/
acceptable.
Test data storageInternal, encrypted searchable
database with capacity for 1024
patients, 5000 test records, maximum
256 tests/patientsInternal, proprietary encoded
searchable database with capacity
1000 patients, 5000 test recordssimilar
Data TransferData transfer is via wired USBI.1 or 2.0
to third party computer running
Otodynamics Otolink software. Otoport
Pro has Bluetooth 5 functionality
which can also be used to transfer
data to Otolink. The USB connection is
made by a
USB C connector.Data transfer is via wired USBI.1 or 2.0
to third party computer running
Otodynamics Otolink software. Otoport
OAE+ABR has Bluetooth 2.0
functionality can also be used to
transfer data to Otolink.Similar, both the subject and
predicate have both Bluetooth
capability for completed test data
transfer; however, the Bluetooth
2 which is used on the predicate
device is now obsolete and
superseded by Bluetooth 5
which is used on the subject
device. Data can also be
transferred by wired USB cable
as with the predicate. The use of
a USB C connector instead of
the predicate proprietary HiRose
connector increases user
convenience. Device safety
benefits from total electrical
isolation (onto isolation) between
it and the attached device (PC or
charger).
Barcode
readerHas optional 2D barcode scanner for
patient ID and demographics inputs.Has optional 1D barcode scanner for
patent ID and demographics inputs.Similar. The provision of the
optional 2D barcode reader
allows more patient data to be
transferred to the device
database and is also compatible
with 2D barcoding which are still
in widespread use.
ChargingCharging/ Data connector - connects to
Otodynamics PSU (charging) or to PC
USB port (USB 1.lor 2.0) via Data Cable.Charging/Data connector - connects to
Otodynamics PSU (charging) or to PC
USB port (USB 1.lor 2.0) via Data Cable.Similar, but now the subject
device has total electrical
isolation from any device
connected to its USB-C port.
Power StatesIntelligent multi-level power control for
charging/testing/idle/sleep/shutdownIntelligent multi-level power control for
charging/testing/idle/sleep/shutdownThe same way of charging,
testing, Idle mode, sleep,
shutdown.
Charge Time3.5 Hrs. to 100% for supplied PSU charge,
5 hours for PC charge4 Hrs. to 100%Similar
Max
consumption when
testing:1W1W (Otoport) or 1.3W (Otoport ABR)Similar
Max
consumption
when
charging:2.65W when charging from USB
connection. 3.5W when charging from
supplied PSU.2.5WSimilar. Consumption of the unit is
2.5W when charging battery from
a USB to comply with USB
standards. When charged with
the supplied charger/PSU,
charging is faster, and the
consumption is 3.5W. This
improvement in charge speed is
possible because of the Otoport
Pro's improved 5V isolator circuit.
Subject Device
Otoport ProPredicate Device
Otoport/ Otocheck OAE+ABR
(K143395)Overview of similarities and
differences
Transient Evoked Otoacoustic Emissions (TEOAE)
Stimulus
LevelsIdle
  • 80µs positive broadband square
    wave pulse with an intensity of
    64dBpeSPL (peak equivalent) in a
    1cc cavity. Adjusted 80µs positive
    broadband square wave.

Testing

  • 300µs biphasic broadband
    triangular pulse;
  • Screening level: 84dBpeSPL-
    Clinical mode: adjustable in 2dB
    increments; Range 50-90dBpeSPL | Idle
  • 80µs positive broadband square
    wave pulse with an intensity of 64
    dBpeSPL (peak equivalent) in a 1cc
    cavity. Adjusted 80µs positive
    broadband square wave.

Testing

  • 300µs biphasic broadband
    triangular pulse;
  • Screening level: 84dBpeSPL-
    Clinical mode: adjustable in 2dB
    increments; Range: 50-90dBpeSPL | same |
    | Sample rate | 20kHz | 20kHz | same |
    | Stimulus pattern | Each sweep presents 8 stimuli
    responses with the stimulus
    presentation pattern:
    A A A B -A -A -A -B
    Where: B = -3A | Each sweep presents 8 stimuli
    responses with the stimulus
    presentation pattern:
    A A A B -A -A -A -B
    Where: B = -3A | same |
    | Noise level
    calculation | The noise level for noise reject is
    calculated from the difference between
    alternate averaged sweeps | The noise level for noise reject is
    calculated from the difference between
    alternate averaged sweeps | same |
    | Stimulus
    repetition rate | One stimulus every 13ms,
    approximately 80 stims per second | One stimulus every 13ms,
    approximately 80 stims per second | same |
    | Response window | 3-13ms or 3-9ms after start of stimulus
    presentation. Cosine filtered with rise
    and fall time of 2ms | 3-13ms or 3-9ms after start of stimulus
    presentation. Cosine filtered with rise
    and fall time of 2ms | same |
    | Response
    frequency bands | Half octave, centered at 1k, 1.4k, 2k,
    2.8k, 4k and 5.6kHz | Half octave, centered at 1k, 1.4k, 2k,
    2.8k, 4k and 5.6kHz | same |
    | | Distortion Product Otoacoustic Emissions (DPOAE) | | |
    | Stimulus level | Preset L1:L2
    60:50dBSPL; 65:55dBSPL;
    70:60dBSPL
    For clinical measurement
    L1 and L2 adjustable
    Range 50-75dBSPL | Preset L1:L2
    60:50dBSPL; 65:55dBSPL;
    70:60dBSPL
    For clinical measurement
    L1 and L2 adjustable
    Range 50-75dBSPL | same |
    | Sample rate | 25.6kHz | 25.6kHz | same |
    | FFT frequency Bin | 25Hz | 25Hz | same |
    | DP noise
    calculation | DP noise is calculated from the five
    spectral points above and the five
    points below the DP frequency. A
    margin of 1 or 2 standard deviations is
    applied for precise clinical
    measurements | DP noise is calculated from the five
    spectral points above and the five
    points below the DP frequency.
    A margin of 1 or 2 standard deviations
    is applied for precise clinical
    measurements | same |
    | Stimulus
    frequencies (F2) | 1-6kHz, 1.5-8kHz, 2-6kHz
    (Recommended)
    1.5-8k tests at six F2 frequencies: 1.5k,
    2k, 3k, 4k, 6k and 8kHz
    1-6k tests at six F2 frequencies: 1k,
    1.5k, 2k, 3k, 4k and 6kHz
    2-6k tests at four F2 frequencies: 2k,
    3k, 4k and 6kHz
    Screening: half or quarter octaves | 1-6kHz, 1.5-8kHz, 2-6kHz
    (Recommended)
    1.5-8k tests at six F2 frequencies: 1.5k,
    2k, 3k, 4k, 6k and 8kHz
    1-6k tests at six F2 frequencies: 1k,
    1.5k, 2k, 3k, 4k and 6kHz
    2-6k tests at four F2 frequencies: 2k,
    3k, 4k and 6kHz
    Screening: half or quarter octaves | same |
    | | between 1k and 8kHz
    Clinical measurement mode.
    Configurable start/stop frequencies up
    within device resolution. Up to 16
    frequency points per sweep, between
    1k and 8kHz | between 1k and 8kHz
    Clinical measurement mode.
    Configurable start/stop frequencies up
    within device resolution. Up to 16
    frequency points per sweep, between
    1k and 8kHz | |
    | Frequency ratio
    f2/f1 | 1.22 | 1.22 | same |
    | | Automated Auditory Brainstem Response (AABR) | | |
    | Stimulus levels | For infant screening only; 35 or
    40dBHL are recommended.
    (A higher level up to 60dBHL is
    provided for the training of adult
    screeners) | For infant screening only; 35 or
    40dBHL are recommended.
    (A higher level up to 60dBHL is
    provided for the training of adult
    screeners) | same |
    | Stimulus rate | 51.8-57.9/second, alternating polarity. | 51.8-57.9/second, alternating polarity. | same |
    | Stimulus type | Chirp (default recommended) or click | Chirp (default recommended) or click | same |
    | Sampling rate | 25.6kHz | 25.6kHz | same |
    | Capture time
    frame | 17.3 to 18.8ms | 17.3 to 18.8ms | same |
    | Capture
    bandwidth | 150-1000Hz | 150-1000Hz | same |
    | Evaluation method | Fsp with infant template matching | Fsp with infant template matching | same |
    | Electrode
    Impedance
    sensing | 400Hz. Less than 5microamps
  • to over 16kohms | 400Hz. Less than 5microamps
  • to over 16kohms | same |
    | Amplifier | Differential, 75dB CMR>60dB@100Hz,
    1Mohm. Dynamic powerline rejection. | Differential, 75dB CMR>60dB@100Hz,
    1Mohm. Dynamic powerline rejection. | same |

6

7

8

9

10

11

12

TABLE 2. COMPARISON BETWEEN THE SUBJECT DEVICE AND THE PREDICATE – Stimulus Parameters

13

Non-Clinical Performance Testing

Electrical safety and electromagnetic compatibility (EMC)

Both electrical safety and EMC testing have been conducted on the Otoport Pro Device. The system complies with ANSI/AAMI/IEC ES60601-1:2005/(R)2012 and Al:2012 for basic safety and essential performance of medical electrical equipment.

14

The electromagnetic compatibility of the system has been confirmed through third-party testing to collateral standard IEC 60601-1-2 Edition 4.0 2014-02.

Software Verification and Validation Testing

Software verification and validation testing has been conducted in compliance with IEC 62304 Edition 1.1 2015-06 (which is the principal normative standard applied), ANSI/AAMI/IEC ES60601-1:2005//R/2012 and the FDA guidance documents "General Principles of Software Validation; Final Guidance for Industry and FDA Staff" (11-Jan-2002) and "FDA Guidance for the Content of Premarket Submissions for Device Software Functions" (14-June-2023).

Cybersecurity

lssues regarding cybersecurity are included for resolution at all appropriate phases of the products development process. Cybersecurity is addressed in the Software Risk mitigation measures against cybersecurity risks are incorporated into the software requirements specifications. The Otodynamics products/ services considered and followed the recommendations in "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, Guidance for Industry and Food and Drug Administration Staff (27-September- 2023)" and current medical devices cybersecurity regulations.

The hardware and firmware changes do not negatively affect the cybersecurity, safety, reliability, and longevity of service of the device relative to the predicate.

Clinical Performance Testing

The indications of use of the Otoport Pro are the same as for the predicate device Otoport OAE+ABR.

The performance of each test modality of the Otoport Pro has been demonstrated to be similar to that of the predicate Otoport OAE+ABR in bench tests and in a clinical measurement trial.

Measurements of the acoustic stimulus and recorded auditory response

  • Measurement by a calibrated microphone of the amplitude and waveform of the acoustic stimulus delivered into 1. a calibrated ear simulator after level setting by the Otoport OAE+ABR, in TEOAE, DPOAE and AABR testing modes, using the same probe on each device.
  • Measurement by the device of the response recorded by both Otoport OAE+ABR in TEOAE, 2. DPOAE and AABR testing modes, when presented with an electronically generated from the device's acoustic stimulus (i.e., simulated) TEOAE, DPOAE or ABR response signal respectively.

The acoustic stimulus differences between the devices (using the same probe) were

TEOAE differences were less than 0.4dB,

DPOAE an average difference of less than 0.3dB form 1-8kHz, with a maximum deviation of 0.8dB, and AABR an average difference of 0.2dB from 2-4.5kHz

The received response differences TEOAE average of 0.23 dBSPL, DPOAE average difference 0.4dB, with largest differences less than 2dB, and AABR average difference 0.1dB

Clinical measurement comparison

15

AABR, TEOAE and DPOAE measurements were made on a cohort of adult volunteers giving informed consent, using both the Otoport Pro and the predicate Otoport OAE+ABR, using the same probe which remained in position for each comparison. Tests were conducted in a sound booth by a qualified audiologist.

Screening mode (default for newborns)

TEOAE screening was performed with a click stimulus at level 84dBSPLpe, with 6dBSNR pass criteria in 2 half octave bands;

TEOAE screening of 17 subjects tested for TEOAE, 15 passed on both instruments, the same 2 failed on both instruments.

DPOAE screening was performed at stimulus levels of L1 65 dBSPL for f1 and f2 tones at half octave intervals from 1.5 to 6kHz;

DPOAE screening of 16 subjects tested for DPOAE, 13 subjects passed the default (newborn) screening criteria on both instruments, 3 failed on both instruments, and these subjects failed at the same frequency on each instrument.

AABR screening was performance with a chirp of 40dBHL - audiologically calibrated on 25 healthy young adults; AABR of 14 subjects tested for ABR, 12 subjects passed on both instruments, the same 2 subjects failed on both instruments (with no clear response and high EEG).

Test times were similar on both instruments at about 100 seconds. Average pass ABR amplitudes were 0.49 µV for the Otoport Pro and 0.51 µV for the predicate. This difference is not significant. Both devices reported the same electrode impedances within 240 ohms.

Clinical (measurement) mode

TEOAE: the TEOAE level in half octave bands were measured from 1k to 6kHz. The difference between the mean TEOAE levels reported by the two devices was less than 1dB at each half octave frequency. DPOAE: 13-point DPqrams were recorded on both devices, using the same probe. Levels of L1=L2 of 70 dBSPL were chosen for this older subject cohort, as this might be used clinically to measure DPOAE strength. Averaged across 15 subjects, the mean DPOAE level difference was less than 1dB at each frequency (1k-8kHz). This difference in DPOAE levels between instrument is not audiologically significant.

AABR - no clinical /diagnostic ABR facilities are available on the Otoport Pro.

CONCLUSIONS

The Otoport Pro Device has the same intended use as the predicate device.

The Otoport Pro is substantially equivalent to the Otoport/Otocheck OAE+ABR predicate device with respect to overall device design elements and features. Functional bench testing supported the electronic and acoustic equivalence of the subject and predicate devices. The non-clinical testing has supported the safety and functionality of the device and the indications for use whilst the performance testing data demonstrates that the device is substantially equivalent to the predicate.