AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Green Nitrile Examination Gloves, Powder Free Pink Nitrile Examination Gloves and Powder Free Purple Nitrile Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

Powder Free Green Nitrile Examination Gloves, Powder Free Pink Nitrile Examination Gloves and Powder Free Purple Nitrile Examination Gloves are manufactured to meet the all current specifications listed under the ASTM Specification D6319 - 19, Standard Specification for Nitrile Examination Gloves for Medical Application. Powder Free Green Nitrile Examination Gloves, Powder Free Pink Nitrile Examination Gloves and Powder Free Purple Nitrile Examination Gloves are mainly composed of nitrile latex. It is mainly suitable for hand protection during medical examination. These gloves are powder free. The difference between the devices is the color of powder free green nitrile examination gloves is green, the color of powder free pink nitrile examination gloves is pink, the color of powder free purple nitrile examination gloves is purple. The proposed device(s) are sold non-sterile and are intended to be single-use, disposable devices.

AI/ML Overview

The medical device in question is the "Powder Free Green/Pink/Purple Nitrile Examination Gloves." The documentation provided describes the acceptance criteria and the non-clinical studies performed to demonstrate its performance.

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodAcceptance CriteriaReported Device Performance
ASTM D6319-19Physical Dimensions: - Length (mm): XS/S/M/L/XL: ≥230 - Width(mm): XS: 70±10mm; S: 85±10mm; M: 95±10mm; L: 110±10 mm; XL: 120±10 mm - Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Physical Dimensions: - Length(mm): XS/S/M/L/XL: 236239 (Pass) - Width(mm): XS 70; S 8586; M 9596; L 110112; XL 120 (Pass) - Finger: 0.230.47 (Pass) - Palm: 0.260.46 (Pass)
ASTM D5151-19Watertightness Test for Detection of Holes: - Be free from holes when tested in accordance with ASTM D5151Pass
ASTM D412Physical Properties: - Before Aging: Tensile Strength: 14MPa, min; Ultimate Elongation: 500% min - After Aging: Tensile Strength: 14MPa, min; Ultimate Elongation: 400% minPhysical Properties: - Before Aging: Tensile Strength: 20.1628.78MPa (Pass); Ultimate Elongation: 522558% (Pass) - After Aging: Tensile Strength: 18.2520.39MPa (Pass); Ultimate Elongation: 525555% (Pass)
ASTM D6124Powder Content: - Meet the requirements of Less than 2mg per glove≤0.46mg (Pass)
ISO 10993-10Irritation: - Non-irritatingUnder the conditions of the study, not an irritant.
ISO 10993-11Acute systemic toxicity: - Non-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo.
ISO 10993-23Sensitization: - Non-sensitizingUnder the conditions of the study, not a sensitizer.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes for each specific test within the non-clinical studies. However, the tests were conducted according to recognized international and national standards (ASTM and ISO). The provenance of the data is not specified beyond indicating that the studies were performed as part of the regulatory submission process for the manufacturer, Hebei Titans Hongsen Medical Technology Co., Ltd. in China. These studies are typically prospective tests conducted on representative samples of the manufactured devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

Not applicable. The ground truth for these types of devices (examination gloves) is established by performance against pre-defined, objective physical, chemical, and biological criteria outlined in the referenced standards (ASTM, ISO). Expert consensus is not typically used to establish ground truth for these quantitative, standardized tests.

4. Adjudication Method for the Test Set:

Not applicable. As the tests are based on objective, standardized measurements against specified criteria, an adjudication method for test results is not typically required. The results are quantitative and compared directly to the predetermined acceptance limits.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. This type of study is not applicable for examination gloves, which are physical barriers designed for medical purposes. MRMC studies are typically used for diagnostic devices that involve human interpretation or decision-making aided by AI.

6. Standalone Performance Study:

Yes, a standalone study was performed. The non-clinical tests described (physical dimensions, watertightness, physical properties, powder content, biocompatibility) assess the performance of the device itself (the gloves) against established standards. This represents the standalone performance of the algorithm (in this context, the gloves as a product) without human-in-the-loop performance measurement.

7. Type of Ground Truth Used:

The ground truth used for the non-clinical tests is based on pre-established, objective criteria and limits defined within recognized international and national standards (e.g., ASTM D6319-19, ASTM D5151-19, ISO 10993 series). These standards outline the acceptable ranges for physical properties and the expected biological responses.

8. Sample Size for the Training Set:

Not applicable. This is not an AI/ML device that requires a training set. The "device" is a physical product (gloves) whose performance is evaluated through non-clinical testing against established standards.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for a physical device like examination gloves.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

April 4, 2024

Hebei Titans Hongsen Medical Technology Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd. FangShan District Beijing, 102401 China

Re: K240361

Trade/Device Name: Power Free Green Nitrile Examination Gloves (XS/S/M/L/XL); Power Free Pink Nitrile Examination Gloves (XS/S/M/L/XL); Power Free Purple Nitrile Examination Gloves (XS/S/M/L/XL) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: February 6, 2024 Received: February 6, 2024

Dear Ray Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240361

Device Name

Powder Free Green Nitrile Examination Gloves (XS/S/M/L/XL); Powder Free Pink Nitrile Examination Gloves (XS/S/M/L/XL); Powder Free Purple Nitrile Examination Gloves (XS/S/M/L/XL)

Indications for Use (Describe)

Powder Free Green Nitrile Examination Gloves, Powder Free Pink Nitrile Examination Gloves and Powder Free Purple Nitrile Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

{4}------------------------------------------------

510(k) Summary

The assigned 510(k) Number: K240361

    1. Date of Preparation: 03/13/2024
    1. Submitter

Hebei Titans Hongsen Medical Technology Co., Ltd. Eastern Industrial Zone, Nangong City, Xingtai City, 051800, Hebei, China Contact Person: Xiaojing Zhao

Position: Office clerk Tel: +86-13383292209 Email: sara@titans-cn.com

    1. Submission Correspondent

Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401 Contact Person: Ray Wang Position: General Manager Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com

    1. Subject Device Identification
      Trade Name: Powder Free Green Nitrile Examination Gloves

Powder Free Pink Nitrile Examination Gloves

Powder Free Purple Nitrile Examination Gloves

Common Name: Medical Examination Glove

Regulatory Information: Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

{5}------------------------------------------------

  • ડ. Predicate Device Identification
    510(k) Number: K232008

Product Name: Disposable Medical Examination Nitrile Gloves

Manufacturer: Longgang City Ailiya Arts Crafts Co., Ltd.

Regulatory Information: Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital Common Name: Medical examination gloves

6. Device Description

Powder Free Green Nitrile Examination Gloves, Powder Free Pink Nitrile Examination Gloves and Powder Free Purple Nitrile Examination Gloves are manufactured to meet the all current specifications listed under the ASTM Specification D6319 - 19, Standard Specification for Nitrile Examination Gloves for Medical Application. Powder Free Green Nitrile Examination Gloves, Powder Free Pink Nitrile Examination Gloves and Powder Free Purple Nitrile Examination Gloves are mainly composed of nitrile latex. It is mainly suitable for hand protection during medical examination. These gloves are powder free. The difference between the devices is the color of powder free green nitrile examination gloves is green, the color of powder free pink nitrile examination gloves is pink, the color of powder free purple nitrile examination gloves is purple.

The proposed device(s) are sold non-sterile and are intended to be single-use, disposable devices.

    1. Indication For Use Statement
      Powder Free Green Nitrile Examination Gloves, Powder Free Pink Nitrile Examination Gloves and Powder Free Purple Nitrile Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
    1. Technological Characteristic Comparison Summary

{6}------------------------------------------------

ITEMProposed Device (K240361)Predicate Device (K232008)Remark
Powder Free Green/Pink/PurpleNitrile Examination GlovesDisposable Medical ExaminationNitrile Gloves
Product CodeLZALZASAME
Regulation No.21 CFR 880.625021 CFR 880.6250SAME
ClassIISAME
Intended Use /Indications for UsePowder Free Green NitrileExamination Gloves, Powder FreePink Nitrile Examination Glovesand Powder Free Purple NitrileExamination Gloves is adisposable device intended formedical purposes that is worn onthe examiner 's hands to preventcontamination between patient andexaminer.The Disposable MedicalExamination Nitrile Gloves is adisposable device intended formedical purposes that is worn onthe examiner's hands to preventcontamination between patient andexaminer.SAME
Powder freeYesYesSAME
Design featureAmbidextrousAmbidextrousSAME
MaterialNitrileNitrileSAME
ColorGreen, Pink, PurpleBlackDifferent1
OTC useYesYesSAME
SterilityNon-sterileNon-sterileSAME
UseSinge useSinge useSAME
LabelingInformationSingle-use indication, powder free,device color, device name, glovesize and quantity, non-sterile.Single-use indication, powder free,device color, device name, glovesize and quantity, non-sterile.SAME
Size andDimensions (mm)Length (mm):XS/S/M/L/XL: ≥230Width(mm):XS: 70±10mm; S: 85±10mm; M:95±10mm; L: 110±10 mm; XL:120±10 mmLength (mm):XS/S: ≥220; M/L/XL/XXL: ≥230;Width(mm):XS: 70±10mm; S: 80±10mm; M:95±10mm; L: 110±10 mm; XL:120±10 mm; XXL: 130±10 mmDifferent2
Thickness (mm)Finger: ≥0.05; Palm: ≥0.05Finger: ≥0.05; Palm: ≥0.05SAME
PhysicalPropertiesBeforeAgingTensile Strength: 14MPa, minUltimate Elongation:500% minTensile Strength: 14MPa, minUltimate Elongation:500% minSAME
AfterAgingTensile Strength: 14MPa, minUltimate Elongation:400% minTensile Strength: 14MPa, minUltimate Elongation:400% minSAME
Freedom fromHolesBe free from holes when tested inaccordance with ASTMD5151Be free from holes when tested inaccordance with ASTMD5151SAME
Powder ContentMeet the requirements of ASTMD6124Less than 2mg per gloveMeet the requirements of ASTMD6124Less than 2mg per gloveSAME
BiocompatibilityISO 10993-10, under theconditions of the study, not asensitizerISO 10993-23, under theconditions of the study, not anirritantISO 10993-11, under the conditionof acute systemic toxicity test, theISO 10993-10, under theconditions of the study, not anirritant or a sensitizerISO 10993-11, under the conditionof acute systemic toxicity test, theSAME
systemic toxicity in vivo.systemic toxicity in vivo.

Table 1 Comparison of Technology Characteristics

{7}------------------------------------------------

Different 1-Color

The color of the proposed device is different from the predicate device. However, the biocompatibility and specification performance tests have been performed on each of the proposed devices and the test results show no significant concerns. Therefore, the difference will not affect the safety and effectiveness of the proposed device.

Different 2-Size and dimension

The size and dimension of the proposed device is not exactly same as the predicate device. However, the size and dimension of the proposed device has been covered by ASTM D6319-19. The user can select appropriate model depended on size of user's hand. In addition, its dimension has been tested and met the requirement of ASTM D6319-19. Therefore, the difference will not affect the safety and effectiveness of the proposed device.

9. Summary of Non-Clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10: 2021 Biological Evaluation Of Medical Devices - Part 10: Tests For Skin Sensitization ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ASTM D412-16(2021) Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers -Tension

Table 2 Summary of non clinical performance testing

Test MethodPurposeAcceptance CriteriaResults
ASTM D6319-19Physical DimensionsTestLength (mm):XS/S/M/L/XL: ≥230Length(mm):XS/S/M/L/XL:236~239 Pass
Width(mm):XS: 70±10mm; S: 85±10mm;Width(mm):XS 70; S 8586;M 9596;L110~112;XL 120Pass
M: 95±10mm; L: 110±10 mm;
XL: 120±10 mm
Thickness (mm):Finger: ≥0.05; Palm: ≥0.05Finger:0.230.47Palm: 0.260.46

{8}------------------------------------------------

Pass
ASTM D5151-19Watertightness Test for Detection of HolesBe free from holes when tested in accordance with ASTM D5151Pass
ASTM D412Physical PropertiesBefore AgingTensile Strength:14MPa, min20.16~28.78MPa/Pass
Ultimate Elongation:500% min522~558MPa/Pass
After AgingTensile Strength:14MPa, min18.25~20.39MPa/Pass
Ultimate Elongation:400% min525~555MPa/Pass
ASTM D6124Powder ContentMeet the requirements of Less than 2mg per glove≤0.46mg/Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant.
ISO 10993-11Acute systemic toxicityNon- acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo.
ISO 10993-23SensitizationNon-sensitizingUnder the conditions of the study, not a sensitizer.
    1. Summary of Clinical Testing Not applicable

11. Conclusions

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Disposable Medical Examination Nitrile Gloves, cleared under K232008.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.