K Number
K240351
Manufacturer
Date Cleared
2024-03-06

(30 days)

Product Code
Regulation Number
862.3650
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FaStep™ Fentanyl Rapid Test Device (Urine) is a rapid visual immunoassay for the qualitative, presumptive detection of Fentanyl in human urine specimens at a cut-off concentration of 1.0 ng/mL. It is for in vitro diagnostic use only.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The FaStep™ Rapid Fentanyl Urine Test (Urine) is a rapid visual immunoassay for the qualitative, presumptive detection of Fentanyl in human urine specimens at a cut-off concentration of 1.0 ng/mL. It is for in vitro diagnostic use only.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Device Description

The FaStep™ Fentanyl Tests are immunoassays intended for the qualitative detection of fentanyl in human urine. Each FaStep™ Fentanyl Test device consists of a Test Cassette, a Dropper and a package insert. Each Test Cassette is sealed with sachets of desiccant in an aluminum pouch.

AI/ML Overview

The provided text is a 510(k) summary for the FaStep™ Fentanyl Rapid Test Device (Urine), an in vitro diagnostic device. It details the device's technical specifications and performance characteristics through analytical and lay-user studies. It does not describe a clinical study comparing human readers with and without AI assistance (MRMC study), nor does it involve the use of AI. Therefore, the information typically associated with acceptance criteria and studies for AI-powered devices, such as specific performance metrics like AUC, sensitivity, specificity for diagnostic accuracy in a clinical setting, or expert consensus and adjudication methods for ground truth, is not directly applicable or present in this document.

However, I can extract the relevant information from the document regarding the acceptance criteria and the study that proves the device meets the acceptance criteria, adapting to the context of a rapid diagnostic test.

Here’s the breakdown based on the provided text, focusing on the analytical and lay-user performance of the rapid immunoassay:

Acceptance Criteria and Device Performance (Adapted for a Rapid Diagnostic Test)

The primary performance evaluation in this document centers around the device's ability to accurately detect fentanyl in urine at a specific cut-off concentration (1.0 ng/mL) under various conditions. The "acceptance criteria" are implied by the desired accuracy rates at and around this cut-off, as demonstrated in the precision and lay-user studies.

1. Table of Acceptance Criteria (Implied) and Reported Device Performance:

Study/ParameterImplied Acceptance Criteria (e.g., High Agreement, Low False Results)Reported Device Performance (from Precision Study) - Example Lot 1Reported Device Performance (from Lay-User Study)
Precision (Analytical Performance)Minimal false positives/negatives at concentrations far from cutoff; Balanced detection around cutoff.
-100% cut off (0 ng/mL)Ideal: 100% Negative60-/0+ (100% Negative)100% Correct (Negative)
-75% cut off (0.25 ng/mL)Ideal: 100% Negative60-/0+ (100% Negative)100% Correct (Negative)
-50% cut off (0.5 ng/mL)Ideal: 100% Negative60-/0+ (100% Negative)100% Correct (Negative)
-25% cut off (0.75 ng/mL)Ideal: Predominantly Negative, some positive acceptable around cutoff58-/2+ (96.7% Negative; 3.3% False Positive)95% Correct (Predominantly Negative, 1 false positive out of 20)
Cut off (1.0 ng/mL)Ideal: Mix of Positive/Negative, indicating performance at threshold35+/25- (58.3% Positive; 41.7% False Negative)N/A (not directly presented for lay-user study as a 'cutoff' row)
+25% cut off (1.25 ng/mL)Ideal: Predominantly Positive, some negative acceptable around cutoff60+/0- (100% Positive)100% Correct (Positive)
+50% cut off (1.5 ng/mL)Ideal: 100% Positive60+/0- (100% Positive)100% Correct (Positive)
+75% cut off (1.75 ng/mL)Ideal: 100% Positive60+/0- (100% Positive)100% Correct (Positive)
+100% cut off (2.0 ng/mL)Ideal: 100% Positive60+/0- (100% Positive)N/A (not directly presented for lay-user study)
InterferenceNo interference from common substances at specified concentrations."Compounds that showed no interference at a concentration of 100µg/mL or specified concentrations are summarized..." (List provided, indicating no interference).Not applicable.
SpecificityMinimal cross-reactivity with structurally similar compounds; no cross-reactivity with unrelated opioids.Minimum concentrations causing positive results for related compounds vary; "Negative results were obtained for all these compounds [unrelated opioids]. There is no cross-reactivity..."Not applicable.
StabilityDevice remains stable over stated shelf life."The devices are stable at 36-86F for 24 months based on the real-time stability study."Not applicable.
Lay-User InstructionsInstructions are easy to understand and follow for lay users.Flesch-Kincaid reading grade level < 7.All lay users indicated instructions were easily followed.

2. Sample Sizes and Data Provenance:

  • Test Set (Precision Study): For each concentration level, 60 tests were performed (2 tests/day for 10 days per device lot, across 3 lots). Thus, for each concentration, 180 total tests (60x3) were conducted across the three lots combined. The document mentions urine samples were "prepared by spiking fentanyl in negative samples." This suggests a controlled laboratory setting.
  • Test Set (Lay-User Study): 140 lay persons participated, each testing one blind-labeled sample. A total of 140 samples were tested (20 per concentration level for 7 levels).
  • Test Set (Method Comparison Study): 82 unaltered clinical samples (42 negative and 40 positive) were used.
  • Data Provenance: The document does not explicitly state the country of origin for the samples or if they were retrospective or prospective patient samples. However, the use of "spiked" samples and "unaltered clinical samples" suggests a mix, likely from a laboratory and potentially local clinical collection. The "Method Comparison Study" description implies retrospective use of clinical samples ("unaltered clinical samples").

3. Number of Experts and Qualifications for Ground Truth:

  • For Precision, Interference, Specificity Studies: The ground truth for these analytical studies was established by preparing samples with known concentrations of fentanyl or interfering substances, verified by LC/MS (Liquid Chromatography/Mass Spectrometry). This is a highly accurate and accepted laboratory method for quantification, rather than relying on human experts for reading.
  • For Method Comparison Study: Ground truth was established by LC/MS results.
  • For Lay-User Study: Ground truth for the spiked samples was established by LC/MS.

4. Adjudication Method for the Test Set:

  • For the analytical studies (precision, interference, specificity), the ground truth was concentrations confirmed by LC/MS, so an 'adjudication' in the human-reader sense is not applicable. The device's result (positive/negative) was compared directly to the LC/MS confirmed concentration.
  • For the method comparison study, the results of the FaStep device were compared against LC/MS results. The "Discordant Results" table shows where the device result did not match the LC/MS result, demonstrating the instances where reconciliation (if any) might happen in a real-world scenario, but no formal adjudication process between human readers is described for this specific device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No, an MRMC comparative effectiveness study was NOT done. This document describes the performance of a rapid diagnostic test (an immunoassay), not an AI algorithm assisting human readers in interpreting medical images or data. Therefore, the concept of human readers improving with AI assistance is not applicable here.

6. Standalone Performance:

  • Yes, a standalone performance study was done. The entire analytical performance section (Precision, Interference, Specificity, Effect of Urine Specific Gravity and pH) and the lay-user study demonstrate the performance of the device itself without human intervention in terms of interpretation beyond visual reading of the test line. The device autonomously produces a "positive" or "negative" line.

7. Type of Ground Truth Used:

  • The primary ground truth used for performance validation was LC/MS (Liquid Chromatography/Mass Spectrometry) confirmed concentrations. This is a highly accurate chemical analytical method considered the gold standard for drug detection and quantification in biological samples.

8. Sample Size for the Training Set:

  • Not applicable in the context of an immunoassay device. This device is a chemical immunoassay, not a machine learning or AI algorithm that requires a "training set" to learn. The device's components and design are established during manufacturing, not through machine learning.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As explained above, this device does not utilize a training set in the AI/machine learning sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION".

Assure Tech LLC % Joe Shia Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877

Re: K240351

Trade/Device Name: FaStep™ Fentanyl Rapid Test Device (Urine); FaStep™ Rapid Fentanyl Urine Test (Urine) Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: NGL Dated: February 2, 2024 Received: February 5, 2024

Dear Joe Shia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A. Digitally signed by Kotarek -S Date: 2024.03.06
19-05'00'

Joseph Kotarek Branch Chief for Toxicology Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240351

Device Name

FaStep™ Fentanyl Rapid Test Device (Urine) FaStep™ Rapid Fentanyl Urine Test (Urine)

Indications for Use (Describe)

The FaStep 111 Fentanyl Rapid Test Device (Urine) is a rapid visual immunoassay for the qualitative, presumptive detection of Fentanyl in human urine specimens at a cut-off concentration of 1.0 ng/mL.

It is for in vitro diagnostic use only.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The FaStep™ Rapid Fentanyl Urine Test (Urine) is a rapid visual immunoassay for the qualitative, presumptive detection of Fentanyl in human urine specimens at a cut-off concentration of 1.0 ng/mL. It is for in vitro diagnostic use only.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY K240351

    1. Date: March 5, 2024 2. Submitter: Assure Tech. LLC. 1521 Concord Pike, Suite 201 Wilmington, DE 19803 3. Contact person: Joe Shia LSI International Inc. 504E Diamond Ave., Suite H Gaithersburg, MD 20877 Telephone: 240-505-7880 Email: shiajl@yahoo.com
  • FaStepTM Fentanyl Rapid Test Device (Urine) 4. Device Names: FaStepTM Rapid Fentanyl Urine Test (Urine)
Product CodeClassificationRegulation SectionPanel
NGLII21 CFR § 862.3650Opiate Test SystemToxicology (91)
    1. Predicate Devices:
      AllTest Fentanyl Urine Test Cassette (K233417)
    1. Indications for Use
      The FaStep™ Fentanyl Rapid Test Device (Urine) is a rapid visual immunoassay for the qualitative, presumptive detection of Fentany] in human urine specimens at a cut-off concentration of 1.0 ng/mL. It is for in vitro diagnostic use only.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The FaStep™ Rapid Fentanyl Urine Test (Urine) is a rapid visual immunoassay for the qualitative, presumptive detection of Fentanyl in human urine specimens at a cut-off concentration of 1.0 ng/mL. It is for in vitro diagnostic use only.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

    1. Device Description
      The FaStep™ Fentanyl Tests are immunoassays intended for the qualitative detection of fentanyl in human urine. Each FaStep™ Fentanyl Test device consists of a Test Cassette, a Dropper and a package insert. Each Test Cassette is sealed with sachets of desiccant in an aluminum pouch.
    1. Substantial Equivalence Information

{4}------------------------------------------------

A summary comparison of features of the FaStep™ Fentanyl Test and the predicate devices is provided in following table.

ItemDevicePredicate - K233417
Indication(s)for UseFor the qualitative determination offentanyl in human urine.Same
Calibrator and Cut-OffValuesFentanyl (FTY)1 ng/mlSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor OTC useSame
ConfigurationsCassetteSame
Storage4-30°CSame

Table 1: Features Comparison of FaStep™ Fentanyl Test and the Predicate Device

    1. Test Principle
      The FaStepTM Fentanyl Tests are immunoassays based on the principle of competitive binding. During testing, a urine specimen migrates upward by capillary action. Fentanyl, if present in the urine specimen below 1 ng/mL, will not saturate the binding sites of antibody-coated particles in the test device. The antibody-coated particles will then be captured by immobilized fentanyl conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the fentanyl level exceeds 1 ng/mL because it will saturate all the binding sites of anti-fentanyl antibodies.

To serve as a procedural control, a colored line will always appear at the control line region indicating that proper volume of specimen has been added and membrane wicking has occurred.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking fentanyl in negative samples. Each fentanyl concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two tests per day for 10 days per device lot in a randomized order.

{5}------------------------------------------------

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 160-/0+60-/0+60-/0+58-/2+35+/25-60+/0-60+/0-60+/0-60+/0-
Lot 260-/0+60-/0+60-/0+58-/2+36+/24-60+/0-60+/0-60+/0-60+/0-
Lot 360-/0+60-/0+60-/0+57-/3+35+/25-60+/0-60+/0-60+/0-60+/0-

c. Stability

The devices are stable at 36-86F for 24 months based on the real-time stability study.

d. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drug fentanyl urine with concentrations at 50% below and 50% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following tables.

AcetaminophenDoxepinNortriptyline
Acetone (1000 mg/dL)Ecgonine methyl esterNoscapine
AcetophenetidinEphedrineO-Hydroxyhippuric acid
Acetylsalicylic acidErythromycinOctopamine
Albumin (100 mg/dL)Ethanol (1%)Oxalic acid (100 mg/dL)
AlbuterolFenoprofenOxazepam
AminopyrineFluphenazineOxolinic acid
AmitriptylineFurosemideOxymetazoline
AmobarbitalGalactose (10 mg/dL)Papaverine
AmoxicillinGamma globulin (500 mg/dL)Penicillin G
AmpicillinGentisic acidPerphenazine
ApomorphineGlucose (3000 mg/dL)Phencyclidine
Ascorbic AcidHemoglobinPhenelzine
AspartameHydralazinePhenobarbital
AtropineHydrochlorothiazidePrednisone
Benzilic acidHydrocortisonePropoxyphene
Benzoic acidHydroxytyraminePropranolol
BenzoylecgonineIbuprofenPseudoephedrine
BilirubinImipramineQuinine
Boric acid (1% w/v)IsoproterenolRanitidine
BupropionIsoxsuprineRiboflavin (10 mg/dL)
CaffeineKetamineSalicylic acid
CarbamazepineKetoprofenSecobarbital
Chloral hydrateLabetalolSerotonin (5-hydroxytyramine)
ChloramphenicolLidocaineSulfamethazine
ChlorothiazideLoperamideSulindac
ChlorpromazineMaprotilineTetrahydrocortisone 3-(β-Dglucuronide)
CholesterolMeperidineTetrahydrocortisone 3-acetate
ClomipramineMeprobamateTetrahydrozoline
ClonidineMethapyrileneThiamine
CortisoneMethaqualoneThioridazine
CotinineMethoxyphenamineTriamterene
CreatinineMetronidazole (300 µg/mL)Trifluoperazine

{6}------------------------------------------------

CyclobenzaprineN-AcetylprocainamideTrimethoprim
DeoxycorticosteroneNaCl (4000 mg/dL)Tyramine
DesipramineNalidixic acidUrea (2000 mg/dL)
DextromethorphanNaloxoneUric acid
DiclofenacNaltrexoneValproic acid (250 ug/mL)
DiflunisalNaproxenVenlafaxine
DigoxinNiacinamideVerapamil
DiphenhydramineNicotineZomepirac
DL-TryptophanNifedipineB-Estradiol
DL-TyrosineNorethindrone

e. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of device. The lowest concentration that caused a positive result for each compound are listed below.

Fentanyl (Cutoff=1ng/mL)Minimumconcentration requiredto obtain a positiveresult (ng/mL)% Cross-Reactivity
Acetyl fentanyl520.00
Acrylfentanyl520.00
ω-1-Hydroxyfentanyl500000.002
Isobutyryl fentanyl520.00
Ocfentanil1001.00
Butyryl fentanyl254.00
Furanyl fentanyl1010.00
Valeryl fentanyl502.00
(±) β-hydroxythiofentanyl520.00
4-Fluoro-isobutyrylfentanyl502.00
Para-fluorobutyryl fentanyl254.00
Para-fluoro fentanyl254.00
(±)-3-cis-methyl fentanyl502.00
Carfentanil100000.01
Despropionyl fentanyl (4-ANPP)500000.002
Sufentanil>10000<0.01
Alfentanil>100000<0.001
Remifentanil>10000<0.01
Norfentanyl>100000<0.001
Acetyl norfentanyl>100000<0.001
Norcarfentanil>10000<0.01

The following opioids compounds were tested at a concentration of 100ug/mL. Negative results were obtained for all these compounds. There is no cross-reactivity for these compounds using the FaStep™ Fentanyl Test.

6-Acetyl morphineNaltrexone
AmphetamineNorbuprenorphine
BuprenorphineNorcodeine
BuprenorphineglucuronideNorketamine

{7}------------------------------------------------

CodeineNormeperidine
DextromethorphanNormorphine
DihydrocodeineNoroxycodone
EDDPOxycodone
EMDPOxymorphone
FluoxetinePentazocine (Talwin)
HeroinPipamperone
HydrocodoneRisperidone
HydromorphoneTapentadol
KetamineThioridazine
LevorphanolTilidine
MeperidineTramadol
MethadoneTramadol-O-Desmethyl
MorphineTramadol-N-Desmethyl
Morphine-3-glucuronideTrazodone
Naloxone

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target fentanyl at 50% below and 50% above Cut-Off levels. These samples were tested using three lots of device. Results were all positive for samples at and above +50% Cut-Off and all negative for samples at and below -50% Cut-Off.

2. Comparison Studies

Method comparison studies for the FaStep™ Fentanyl Test were performed using three different lots of the device. Operators ran 82 (42 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to LC/MS results are presented in the tables below.

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
OperatorPositive0021622
1Negative5161920
OperatorPositive0021622
2Negative5161920
OperatorPositive0031722
3Negative5161810

Discordant Results

OperatorSample IDLC/MS Result (ng/mL)Rapid Test Result
Operator 1F00310.89+
Operator 1F00940.94+
Operator 1F00011-

{8}------------------------------------------------

Operator 1F00841.1-
Operator 2F00480.93+
Operator 2F00940.94+
Operator 2F00011-
Operator 2F00841.1-
Operator 3F00310.89+
Operator 3F00480.93+
Operator 3F00940.94+
Operator 3F00011-

3. Lay-user study

A lay user study was performed at three testing sites with 140 lay persons. They had diverse educational and professional backgrounds and ranged in age from 20 to >50 years. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cut-off by spiking fentanyl into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers, blind-labeled and randomized. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below:

% of CutoffNumberofsamplesFentanyl Concentrationby LC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff200.28020100
-50% Cutoff200.51020100
-25% Cutoff200.7311995
+25% Cutoff201.23200100
+50% Cutoff201.48200100
+75% Cutoff201.75200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on the package insert and the score revealed a reading grade level of less than 7.

4. Clinical Studies

Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison and Lay-user studies of the devices, it's concluded a substantial equivalence decision.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).