K Number
K240320
Date Cleared
2024-06-28

(147 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RespArray™ patient monitor is intended to be used for monitoring, storing, reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitor is for prescription use only.

The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and carbon dioxide (CO2).

The arrhythmia detection and ST Segment analysis are intended for adult patients.

The SpO2 (Nellcor™) module is intended to be used for spot-check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR), in motion and no motion conditions, and in patients who are well or poorly perfused.

The Microstream™ capnography module is intended for continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR). The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for ages 22 and up.

The monitors are not intended for MRI environments.

Device Description

The RespArray patient monitor (hereinafter called RespArray) can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time.

AI/ML Overview

The provided text is a 510(k) summary for the Edan Instruments Patient Monitor (RespArray). It focuses on establishing substantial equivalence to a predicate device, primarily through non-clinical performance and software verification/validation.

Crucially, the document explicitly states: "Clinical data: Not applicable." This means there was no clinical study conducted to prove the device meets specific acceptance criteria in a human-use setting, particularly relating to diagnostic accuracy where AI assistance or expert consensus would be relevant.

Therefore, many of the requested items, such as multi-reader multi-case studies, ground truth establishment for a test set, and sample sizes for clinical test sets, are not applicable to this submission as per the document's contents.

Here's a breakdown based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criteria discussed are primarily related to electrical safety, electromagnetic compatibility (EMC), and general performance of the physiological parameter measurements (ECG, RESP, NIBP, TEMP, SpO2, CO2). The document states that the device was found to comply with relevant standards and that bench testing shows it meets its accuracy specification and relevant consensus standards.

Table of Acceptance Criteria and Reported Device Performance (as inferred from the text):

Acceptance Criteria CategorySpecific Acceptance Criteria (Standards Adhered To)Reported Device Performance (Compliance)
Electrical SafetyANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]Found to comply.
EMCANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021]Found to comply.
Alarm SystemsIEC 60601-1-8:2020Bench testing results show compliance.
ECG MonitoringIEC 60601-2-27:2011Bench testing results show compliance, meets accuracy specification.
NIBPIEC 80601-2-30:2018Bench testing results show compliance, meets accuracy specification.
Multifunction Patient MonitoringIEC 60601-2-49:2018Bench testing results show compliance, meets accuracy specification.
Respiratory Gas MonitorsISO 80601-2-55:2018Bench testing results show compliance, meets accuracy specification.
Clinical ThermometersISO 80601-2-56:2017+A1:2018Bench testing results show compliance, meets accuracy specification.
Pulse OximeterISO 80601-2-61:2017Bench testing results show compliance, meets accuracy specification.
Wireless CoexistenceIEEE ANSI C63.27:2017Bench testing results show compliance.
Software FunctionalityFDA Guidance: "Content of Premarket Submissions for Device Software Functions"Software verification and validation testing were conducted and documentation was provided as recommended by FDA guidance.

Study Proving Device Meets Acceptance Criteria:

The study proving the device meets the acceptance criteria is described as "Non-clinical data" including:

  • Electrical safety and electromagnetic compatibility (EMC) assessments.
  • Performance testing-Bench
  • Software Verification and Validation Testing.

Detailed Information on the Study:

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly quantified in terms of number of patients or physiological measurements. The testing was "bench" testing, meaning laboratory-based tests on the device's functionality and accuracy against reference standards, rather than patient-derived data.
    • Data Provenance: Not applicable in the context of clinical data. For testing against standards, it implies standardized test setups and simulated physiological signals/conditions.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No experts were used for ground truth establishment as it was non-clinical bench testing against established engineering and medical device performance standards. There's no "ground truth" of a diagnostic nature being established by human experts in this context.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication is relevant for human-interpreted diagnostic data. This study relies on objective measurements against engineering specifications and international standards.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, not applicable. The device is a patient monitor, not an AI-assisted diagnostic tool that interprets medical images/signals for improved human reader performance. The submission explicitly states "Clinical data: Not applicable."
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in essence. The "Performance testing-Bench" and "Software Verification and Validation Testing" evaluate the device's inherent algorithms and functionality in a standalone manner (without a human in the loop for diagnostic interpretation, but rather for operational functionality and accuracy of physiological measurements). However, this is not in the context of a diagnostic AI algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Reference standards and engineering specifications. The "ground truth" for the non-clinical tests is derived from established international and national standards for medical electrical equipment, specific performance criteria for various physiological measurement modules, and the device's own accuracy specifications validated through bench testing.
  7. The sample size for the training set:

    • Not applicable for a clinical training set. This is not an AI/ML device in the sense of learning from a large dataset of patient cases. The device's algorithms are designed and verified, not "trained" on patient data in a machine learning sense.
  8. How the ground truth for the training set was established:

    • Not applicable. As above, no clinical training set or associated ground truth establishment process is described or relevant for this type of device submission.

In summary, the provided 510(k) focuses on demonstrating substantial equivalence through adherence to recognized performance and safety standards via non-clinical bench testing and software validation, rather than clinical studies involving human patients or complex AI diagnostic algorithms requiring expert review and adjudication.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 28, 2024

Edan Instruments, Inc Tracy Yue Regulatory Affairs Engineer 15 Jinhui Road, Jinsha Community, Kengzi Sub-district, Pingshan District, Shenzhen, Guangdong 518122 China

Re: K240320

Trade/Device Name: Patient Monitor (RespArray) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, DSB Dated: February 2, 2024 Received: May 24, 2024

Dear Tracy Yue:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kimberly N. Crowley -S

For: Jennifer Shih Kozen Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices

{2}------------------------------------------------

OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240320

Device Name

Patient Monitor (RespArray)

Indications for Use (Describe)

The RespArray™ patient monitor is intended to be used for monitoring, storing, reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitor is for prescription use only.

The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and carbon dioxide (CO2).

The arrhythmia detection and ST Segment analysis are intended for adult patients.

The SpO2 (Nellcor™) module is intended to be used for spot-check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR), in motion and no motion conditions, and in patients who are well or poorly perfused.

The Microstream™ capnography module is intended for continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR). The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for ages 22 and up.

The monitors are not intended for MRI environments.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary

Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92

1. Submitter:

Applicant:Edan Instruments, Inc.#15 Jinhui Road, Jinsha Community,Kengzi Sub-District, Pingshan District,Shenzhen, 518122 P.R.China.Tel: +86(0755) 26858736Fax: +86(0755) 26882223
Contact person:Tracy Yue
Contact person:Tracy Yue
Preparing date:February 02, 2024

2. Device name and classification:

Trade name:Patient Monitor, Model: RespArray
Common/Usual Name:Patient Monitor
Classification Name:21 CFR 870.1025
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Regulatory class:Class II
Primary product code:MHX-Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)
Subsequent productcode:Regulation number/DeviceProduct Code
Regulation number/DeviceProduct Code
21 CFR 870.2340Detector and Alarm, ArrhythmiaDSI
21 CFR 870.1025Monitor, ST Segment with AlarmMLD
21 CFR 870.2300Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm)DRT
21 CFR 870.1130System, Measurement, Blood-Pressure, Non-InvasiveDXN
21 CFR 870.1110Computer, Blood-PressureDSK
21 CFR 880.2910Thermometer, Electronic, ClinicalFLL
21 CFR 870.2700OximeterDQA
21 CFR 870.1400Analyzer, Gas, Carbon-Dioxide, Gaseous-PhaseCCK

{5}------------------------------------------------

Traditional 510(k) of RespArray Patient Monitor

21 CFR 870.2770DSB
Impedance plethysmograph
  1. Predicate Device(s): 1) Edan Instruments, Inc., Patient Monitor, Model: RespArray, K220308 (Primary predicate device)
    1. Edan Instruments, Inc., Vital Signs Monitor, Model: iM3s, iM3As, iM3Bs, iHM3s K202892 (Reference device)
    1. Covidien LLC., MicrostreamTM CO2 NanoPod, K213911 (Reference device)

The RespArray patient monitor (hereinafter called RespArray) can perform long-time 4. Device Description: continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time.

  • The RespArray™ patient monitor is intended to be used for monitoring, storing, reviewing 5. Indication for Use of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.
    The monitor is for prescription use only.

The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and carbon dioxide (CO2).

The arrhythmia detection and ST Segment analysis are intended for adult patients.

The SpO2 (Nellcor™) module is intended to be used for spot-check or continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR), in motion and no motion conditions, and in patients who are well or poorly perfused.

The Microstream™ capnography module is intended for continuous non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath and carbon dioxide concentration of the exhaled breath and inspired (etCO2) and respiration rate (RR). The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for age 22 and up. The monitors are not intended for MRI environments.

6. Predicate Device Comparison

The table below compares the indication for use and key technological feature of the subject devices to the predicate device (Patient Monitor Model RespArray, K220308).

{6}------------------------------------------------

ItemRespArrayRespArrayComparisonResult
Manufacturer/K#Current SubmissionEdan Instrument.Inc/K220308
Intended UseThe RespArray™ patient monitor isintended to be used for monitoring,storing, reviewing of, and to generatealarms for, multiple physiologicalparameters of adults, pediatrics andneonates. The monitors are intendedfor use by trained healthcareprofessionals in hospital environments.The monitor is for prescription useonly.The monitored physiologicalparameters include: ECG, respiration(RESP), temperature (TEMP), oxygensaturation of arterial blood (SpO2),pulse rate (PR), non-invasive bloodpressure (NIBP), and carbon dioxide(CO2).The arrhythmia detection and STSegment analysis are intended foradult patients.The SpO2 (Nellcor™) module isintended to be used for spot-check orcontinuous non-invasive monitoring offunctional oxygen saturation of arterialhemoglobin (SpO2) and pulse rateThe RespArray™ patient monitor isintended to be used for monitoring,storing, reviewing of, and to generatealarms for, multiple physiologicalparameters of adults, pediatrics andneonates. The monitors are intendedfor use by trained healthcareprofessionals in hospital environments.The monitor is for prescription useonly.The monitored physiologicalparameters include: ECG, respiration(RESP), temperature (TEMP), oxygensaturation of arterial blood (SpO2),pulse rate (PR), non-invasive bloodpressure (NIBP), and carbon dioxide(CO2).The arrhythmia detection and STSegment analysis are intended foradult patients.The SpO2 (Nellcor™) module isintended to be used for spot-check orcontinuous non-invasive monitoring offunctional oxygen saturation of arterialhemoglobin (SpO2) and pulse rateSame
(PR), in motion and no motionconditions, and in patients who arewell or poorly perfused.The Microstream™ capnographymodule is intended for continuous(PR), in motion and no motionconditions, and in patients who arewell or poorly perfused.The Microstream™ capnographymodule is intended for continuous
non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR).The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for ages 22 and up.The monitors are not intended for MRI environments.non-invasive monitoring of carbon dioxide concentration of the expired and inspired breath (etCO2) and respiration rate (RR).The monitor also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr) and oxygen desaturation index (ODI) values. IPI is not intended for patients up to the age of one year. A/hr and ODI are intended for ages 22 and up.The monitors are not intended for MRI environments.
ECG monitor
Lead Mode3 Electrodes; 5 Electrodes;3 Electrodes; 5 Electrodes;Same
Arrhythmia AnalysisASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED BEATS, IRR, VBRADY, PNC, PNPASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED BEATS, IRR, VBRADY, PNC, PNPSame
RESP monitor from ECG
Principle of OperationThoracic impedanceThoracic impedanceSame
Measurement Range0 rpm to 200 rpm0 rpm to 200 rpmSame
NIBP monitor
Principle of OperationoscillationoscillationSame
Measurement RangeMeasurement range:Adult Pediatric NeonateSystolic 25-290 25-240 25-140Diastolic 10-250 10-200 10-115Mean 15-260 15-215 15-125Measurement range:Adult Pediatric NeonateSystolic 25-290 25-240 25-140Diastolic 10-250 10-200 10-115Mean 15-260 15-215 15-125Same
PR from NIBP
Measurement range40 bpm to 240 bpmSame
Continuous TEMP
Measurement Range0 °C to 50 °C (32 °F to 122 °F)Same
SpO2 monitor
ModuleNell-1Same
Spot Temp
Measuring Range16 °C ~ 43 °C (61 °F ~ 110 °F)The added spot temp is substantial equivalent to the Spot Temp Module cleared in K202892
Power supply
AC power100 V to 240 V~50 Hz/60 HzSame
Rechargeable BatteryYesSame
Battery Capacity6800 mAhThe battery comply with IEC 62133-2.
CO2 Monitor
ModuleMicrostream™ NanoMediCO2 EtCO2The changed CO2 module is substantial equivalent to the CO2 module cleared in

{7}------------------------------------------------

{8}------------------------------------------------

As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.

7. Performance Data:

Non-clinical data:

Electrical safety and electromagnetic compatibility (EMC)

The Resparray Patient Monitors were assessed for conformity with the relevant requirements of the following 510(k)_Summary 5 / 7

{9}------------------------------------------------

standards and found to comply:

  • ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
  • ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: electromagnetic disturbances - Requirements and tests.

Performance testing-Bench

Edan has conducted functional and system level testing to validate the performance of the results of the bench testing show that the subject device meets its accuracy specification and meet relevant consensus standards.

  • IEC 60601-1-8:2020 Medical electrical equipment - part 1-8: general requirements for basic safety and essential performance - collateral standard: general requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
  • IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
  • IEC 80601-2-30:2018 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
  • IEC 60601-2-49:2018 Medical electrical equipment –Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
  • ISO 80601-2-55:2018 Medical electrical equipment - part 2-55: particular requirements for the basic safety and essential performance of respiratory gas monitors
  • ISO 80601-2-56:2017+A1:2018 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
  • ISO 80601-2-61:2017 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment
  • IEEE ANSI C63.27:2017 American National Standard for Evaluation of Wireless Coexistence

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and Food and Drug Administration Staff, "Content of Premarket Submissions for Device Software Functions".

Clinical data: Not applicable.

Summary

510(k)_Summary

{10}------------------------------------------------

The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate device.

8. Conclusion

The bench testing data and software verification and validation demonstrate that RespArray Patient Monitor is substantially equivalent to the predicate device.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.