K Number
K240226
Device Name
Pixie Skin Tag
Manufacturer
Date Cleared
2024-10-18

(266 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pixie® Skin Tag is a cryogenic device for home use intended to treat skin tags in persons over the age of 21.

Device Description

Pixie Skin Taq is a pen-like cryogenic device that utilizes pressurized nitrous oxide (NoO, liquefied) to destroy skin tag tissue. The N2O is released into a single-use, disposable applicator, cooling the applicator as the result of rapid evaporation. The cooled applicator is briefly contacted with the skin tag while the surrounding tissue is protected by a disposable thermally insulating adhesive bandage. Each Pixie Skin Tag device is supplied in a carton with instructions for use and replacement applicator tips and insulating adhesive bandages.

AI/ML Overview

The provided text is a 510(k) summary for the Pixie Skin Tag device. This type of submission is for establishing substantial equivalence to a predicate device, not for proving a device meets specific performance acceptance criteria via a rigorous clinical study of AI performance, especially not as would be described for an AI/ML-driven medical device.

Therefore, the information required to answer your questions regarding acceptance criteria, sample sizes for test/training sets, expert qualifications, ground truth establishment, MRMC studies, or standalone algorithm performance, as typically found in submissions for AI/ML devices, is not present in this document.

This document describes a physical medical device (a cryosurgical unit) and its comparison to a predicate device. The performance evaluations cited are limited to:

  • Bench studies: Comparison of applicator temperatures, device robustness (activation, N2O evacuation, drop protection), and thermal protection of insulating bandages.
  • Animal studies: Application to healthy rabbit skin to determine epithelial damage and wound healing.
  • Usability studies: Novice laypersons' ability to set up and apply simulated treatments after reading instructions.
  • Post-market review of complaints: For CE-marked devices and a reference Pixie device for warts, to assess safety and reported adverse events.

In summary, this document does not contain the information needed to answer the specific questions about acceptance criteria for an AI/ML device.

{0}------------------------------------------------

October 18, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Oystershell NV % Phil Triolo Founder and Principal Consultant Phil Triolo and Associates LC 86 Skycrest Lane Salt Lake City, Utah 84108

Re: K240226

Trade/Device Name: Pixie Skin Tag Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: ORE, GEH Dated: September 19, 2024 Received: September 20, 2024

Dear Phil Triolo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

James H. Digitally signed by James H. Jang -S Date: 2024.10.18 Jang -S 17:30:13 -04'00'

For

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K240266

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Submission Number (if known)

K240266

Device Name

Pixie Skin Tag

Indications for Use (Describe)

Pixie® Skin Tag is a cryogenic device for home use intended to treat skin tags in persons over the age of 21.

Type of Use (Select one or both, as applicable)

[X] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Prescription Use (Part 21 CFR 801 Subpart D)

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for Oystershell. The logo consists of a stylized image of a person with their arms raised above their head, with a red dot above the head. To the right of the image is the word "Oystershell" in blue font. Below the word "Oystershell" is the phrase "Innovating Consumer Health" in a smaller font.

Traditional 510(k) Premarket Notification

Pixie Skin Taq

510(k) Summary Page 1

510(k) Summary

Date Prepared: 18 October 2024

1. SUBMITTER

  • 510(k) Sponsor: Oystershell NV
    Contact Person: Phil Triolo, PhD, RAC President and Senior Consultant Phil Triolo and Associates LC Phone: 801-699-9846 / Fax 801-328-2399 Email: philt@philt.com

  • Phil Triolo and Associates, LC Prepared By: 86 Skycrest Lane SLC UT 84108 www.philt.com

2. DEVICE

Trade/Device Name:Pixie Skin Tag
Common Name:Over-the-counter cryogenic skin tag remover
Classification Name:Cryosurgical unit and accessories
Regulation number:21 CFR 878.4350
Product Code:ORE (Cryogenic Skin Tag Removal, OTC)
Device Class:Class II

3. PREDICATE DEVICE

K211099 CryoConcepts Freeze'n Clear Skin Clinic Warts & Tags, Regulation 21 CFR 878.4350, Product Code GEH

REFERENCE DEVICE

K172373 Compound W Nitro Freeze, Regulation 21 CFR 878.4350, Product Code GEH

DEVICE DESCRIPTION 4.

Pixie Skin Taq is a pen-like cryogenic device that utilizes pressurized nitrous oxide (NoO, liquefied) to destroy skin tag tissue. The N2O is released into a single-use, disposable applicator, cooling the applicator as the result of rapid evaporation. The cooled applicator is briefly contacted with the skin tag while the surrounding tissue is protected by a disposable thermally insulating adhesive bandage. Each Pixie Skin Tag device is supplied in a carton with instructions for use and replacement applicator tips and insulating adhesive bandages.

5. INDICATIONS FOR USE

{5}------------------------------------------------

Pixie® Skin Tag is a cryogenic device for home use intended to treat skin tags in persons over the age of 21.

COMPARISON OF CHARACTERISTICS WITH THE PREDICATE DEVICE 6.

The main difference between the new Pixie Skin Tag and the predicate device is that the new Pixie Skin Tag uses compressed nitrous oxide as the cryogen instead of a combination of dimethyl ether and propane. The evaporation of nitrous oxide results in cooling of the applicator that is contacted with the skin tag during treatments to a lower temperature (-83°C) than the evaporation of the dimethyl ether/propane mixture (-55°C). Contact of the skin tag with the chilled applicator of the new and predicate devices results in cell death due to freezing and blood flow stasis that occurs during thawing of the affected tissue. To protect non-target skin surrounding the skin tag during treatment using the Pixie Skin Tag applicator, thermally insulating bandages are provided for the user to adhere to the skin surrounding the skin tag before treatments. In contrast, the predicate device is provided with tweezers that are used to isolate the skin tag from the surrounding skin to make it easier to contact the skin tag with the applicator without contacting non-target skin.

The Pixie Skin Tag device design and materials are the same as those of the reference Compound W Nitro Freeze Product cleared after FDA review of K172373. The devices differ in the indication for use of the new device to treat skin tags and the inclusion of the single-use, disposable thermally insulating bandages with the new devices.

The following table compares Pixie Skin Tag to the predicate devices with respect to intended use and technological characteristics. This comparison of the devices provides the detailed information that is the basis for the determination of substantial equivalence.

Substantial Equivalence Table: Comparison of New Device with Legally Marketed PredicateDeviceNone of the identified differences raise different questions of safety and effectiveness
ATTRIBUTENEW DEVICEOystershell Pixie Skin Tag (PST)K240226PREDICATE DEVICECryoConcepts Freeze'n Clear (FnC)Skin Clinic Warts & TagsK211099
DescriptionPixie Skin Tag is a pen-like cryogenicdevice that utilizes pressurized nitrousoxide (N₂O, liquefied) to destroy skin tagtissue. The N₂O is released into a single-use, disposable applicator, cooling theapplicator to approximately -83°C as theresult of rapid evaporation. The cooledapplicator tip is contacted with the skin tagfor 20s while the surrounding tissue isprotected by a disposable thermallyinsulating adhesive bandage. The frozenskin tag falls away over time and newepidermis grows in its place. One to threetreatments delivered on a 2-week schedulemay be delivered (maximum of 3applications for any one skin tag).The Freeze 'n Clear Skin Clinic for Wartsand Skin Tags is for OTC use and utilizesa combination of dimethyl ether, propaneand isobutane (DMEP) delivered from thecanister into a single-use, disposableapplicator which acts as a reservoir for thecryogenic gas. The gas rapidly evaporatesand cools the applicator to approximately-55°C. After a 15-second wait, theapplicator is placed against the wart or skintag for 40 seconds which freezes thetargeted tissue. The frozen skin tag or wartfalls away over time and new epidermisgrows in its place. One to four treatmentswith intervals of two weeks may berequired.
ContentsEach Pixie Skin Tag (PST) device issupplied in an outer carton with aninstructions for use leaflet and a blisterpack containing 4 replacement single-useapplicator tips for Pixie 10mL and 9replacement single-use tips for Pixie 21mL,and a sheet of 10 disposable, single-useinsulating adhesive bandages.Freeze 'n Clear Skin Clinic for Warts andSkin Tags is supplied in an outer cartonwith an instructions for use leaflet, multiple-use actuator, 8 single-use applicator tips,and a multiple-use tweezer.
Substantial Equivalence Table: Comparison of New Device with Legally Marketed PredicateDeviceNone of the identified differences raise different questions of safety and effectiveness
ATTRIBUTENEW DEVICEOystershell Pixie Skin Tag (PST)K240226PREDICATE DEVICECryoConcepts Freeze'n Clear (FnC)Skin Clinic Warts & TagsK211099
ClassificationClass IIClass II
Regulation21 CFR 878.435021 CFR 878.4350
NameUnit, Cryosurgical, and AccessoriesUnit, Cryosurgical, and Accessories
PanelGeneral and Plastic SurgeryGeneral and Plastic Surgery
Product CodeORECryogenic Skin Tag Removal, OTCGEHCryosurgical unit and accessories
Intended andIndicated UsePixie® Skin Tag is a cryogenic device forhome use intended to treat skin tags inpersons over the age of 21.The Freeze 'n Clear Skin Clinic for Wartsand Skin Tags product is intended for theOTC treatment of common warts, plantarwarts, and skin tags.
TargetPopulationPersons over the age of 21No age restriction
AnatomicalSite(s)Only treat skin tags if they are located onthe neck, breast, armpits, or underarms viaa narrow stalkInstructions do not restrict the use of thedevice to treat skin tags to any specificanatomical locations.
Where usedHome (over-the-counter) UseHome (over-the-counter) Use
Principles ofOperationThe application of extreme cold removesheat from the targeted skin tag, resulting infreezing and destruction of the tissue.The application of extreme cold removesheat from the targeted skin tag, resulting infreezing and destruction of the tissue.
TechnologyNitrous Oxide is used to cool an applicatorwhich is applied to the skin tag.The same technology is used by Pixiedevices cleared for use to treat warts afterFDA review of K172373.DMEP (mixture of dimethyl ether andpropane) is used to cool a foam applicatorwhich is applied to the lesion.
Mode ofApplicationUser adheres a disposable, thermallyinsulating bandage to the skin surroundingthe skin tag.The user dispenses Nitrous Oxide for 2sinto the disposable applicator tip attachedto the pen. The applicator is immediatelyapplied to the skin tag for a recommendedduration of 20s.The bandage is removed and discarded.Up to 3 treatments may be applied to eachskin tag, separated by 2 weeks.The applicator stick is attached to theactuator. The actuator is pressed into thecanister of DMEP (mixture of dimethylether and propane) for 3s to 5s to dispensethe DMEP and saturate the applicator. Theapplicator and actuator are removed fromthe cannister assembly, and the user waits15s before attempting treatment.To treat the skin tag, the provided tweezersare used to grasp the skin tag and pull itaway from its base, exposing the stalk (orstem) of the skin tag. The applicator tip isthen applied to the stalk of the skin tag forup to a recommended duration of 40s.Up to 4 treatments may be applied to eachskin tag, separated by 2 weeks.
Number oftreatmentsAvailable in 10mL and 21mL sizes,capable of delivering, respectively, 5 and10 treatments.Available in one size, capable of deliveringup to 8 treatments.
FlammabilityNon-flammable, but will support fireHighly Flammable

{6}------------------------------------------------

K240226

Oystershell

Traditional 510(k) Premarket Notification

Pixie Skin Tag

{7}------------------------------------------------

Image: logoOystershellINNOVATING CONSUMER HEALTH
--------------------------------------------------------

Traditional 510(k) Premarket Notification

Pixie Skin Tag

Substantial Equivalence Table: Comparison of New Device with Legally Marketed PredicateDeviceNone of the identified differences raise different questions of safety and effectiveness
ATTRIBUTENEW DEVICEOystershell Pixie Skin Tag (PST)K240226PREDICATE DEVICECryoConcepts Freeze'n Clear (FnC)Skin Clinic Warts & TagsK211099
StorageConditionsand SafetyPrecautionsContents under pressureNever pierce or burn product, even afteruse.Keep away from fire or flame.Do not smoke while using this product.Do not expose to heat or store attemperatures above 122°F (50°C).Store unused product at room temperatureaway from heat.Keep out of reach of children.Do not smoke while using the product.Do not puncture or incinerate canister.Do not expose to heat or store attemperatures above 120°F.Store at room temperature, away from thereach of children, and out of the way ofdirect sunlight or other heat sources.
DisposalDevice is disposable.Device is disposable after emptied ofcryogen.
Shelf Life3 years (unopened)3 years (unopened)
SterilityProduct is provided non-sterileProduct is provided non-sterile

7. PERFORMANCE EVALUATIONS

7.1 Bench Studies

A comparison of the temperatures reached by the new and predicate applicators was conducted and demonstrating that, after dispensing the cryogenic gases into the applicators, the new Pixie Skin Tag applicator reaches and maintains a temperature of approximately -83°C, whereas the predicate applicator reaches a temperature of approximately -33°C, over recommended application durations.

Results of numerous studies demonstrate the ability of PST, after exposure to elevated temperatures or anticipated device misuse, to be activated and dispense N2O to the applicators, and, independently, to safely evacuate the pressurized nitrous oxide; and to protect the N2O cartridge when the device is dropped.

An analysis of the ability of the thermally insulating bandages to protect the skin from thermal damage during Pixie Skin Tag treatments demonstrated adequate thermal protection.

7.2 Animal Studies

A study conducted in rabbits demonstrated that application of PST to healthy rabbit skin resulted in epithelial damage that did not extend deeper than the superficial fascia and evidence of wound healing occurring within 7 days after application.

7.3 Clinical Evidence

Results of three usability studies demonstrated that novice laypersons could set up and apply simulated treatments after reading the provided instructions for use.

CE-marked PST devices have been sold in the EU and other countries for several years. Post-market review of complaints reported for CE-marked PST devices sold through the end of 2022 in countries outside the US revealed no reports of serious incidents, and a total of only 5 reported incidents for non-serious events identified as pain, discoloration, blister formation, skin burns, and hypersensitivity reactions. These are known side effects/risks associated with use of

{8}------------------------------------------------

Traditional 510(k) Premarket Notification 510(k) Summary Oystershel Pixie Skin Tag Page 5

the device. Further, the low incidence of reported skin burns strongly implies that the insulating bandages are being used properly, and are effective.

Post-market review of complaints reported for the over 3 million devices that have been sold, including over 2 million sold in the US, for reference Pixie devices intended to treat warts, revealed only one serious event. The very low and acceptable rates of reported non-serious events, all of which are communicated to end users in the instructions for use, were 0.019% in 2021 and 0.012% in 2022 (based on an assumption that 50% of devices sold were used).

8. SOFTWARE

No software is employed in the Pixie Skin Tag device.

9. BIOCOMPATIBILITY

Pixie Skin Tag pen and applicator components and their processing are the same as those of the reference predicate device which is intended for the same user (patient) contact. Therefore, the biological and toxicological risks associated with the Pixie Skin Tag applicator assembly are the same as the reference predicate device and are low and acceptable.

The thermally insulating bandages make direct contact with intact skin. Results of in-vitro cytotoxicity (MEM Elution), sensitization, and intracutaneous injection studies, conducted in accordance with ISO 10993-5, ISO 10993-10, and ISO 10993-10, respectively, met all acceptance criteria, demonstrating low and acceptable biological risks.

10. STERILIZATION

Neither the new Pixie Skin Tag nor the predicate device is sterilized.

11. STANDARDS

The following standards were referenced in this 510k to support substantial equivalence:

  • ISO 10993-1

  • ISO 10993-12

  • ISO 14155 ●

  • ISO 10993-5

  • ISO 20417 ●

  • ISO 10993-10 ●

  • IEC 62366 ●

  • ISO 14971

12. CONCLUSION

The results of bench, pre-clinical, and usability studies and comparisons with the predicate device provide the evidence needed to find that Oystershell's new Pixie Skin Tag devices are substantially equivalent to the predicate CryoConcepts Freeze'n Clear Skin Clinic Warts & Tags.

N/A