K Number
K240185
Manufacturer
Date Cleared
2024-05-03

(101 days)

Product Code
Regulation Number
878.4800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cypris eXact Suture Placement Device is indicated for use in general soft tissue approximation when used with sutures.

Device Description

The Cypris eXact Suture Placement Device is for use in general soft tissue approximation by surgeons, and it is single-use and provided sterile. It is a hand-held device used for the placement of a surgical suture, and the device consists of a handle, needle driver, and tissue capture port. The handle is designed for manual manipulation of the suture placement device, allowing the surgeon to advance the device to the desired deployment site. The needle driver functions to pass the suture when it is advanced and retracted. The tissue capture port consists of a capture chamber, needle, and hook assembly, and holds the tissue to be sutured during either manual palpation or vacuum-assisted palpation of tissue into the tissue capture port. Suture placement is achieved by advancing and retracting the needle driver.

AI/ML Overview

{
"1. A table of acceptance criteria and the reported device performance": {
"Acceptance Criteria": [
"Biocompatibility endpoints evaluated (cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity)",
"Integrity of the sterile barrier and functionality of the device after 3 years of real-time aging",
"Performance as expected",
"Use of vacuum-assisted palpation equivalent to manual palpation by the predicate device",
"Peak suction force measurement at 25.0 inHg"
],
"Reported Device Performance": [
"Device underwent testing for cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity.",
"Testing demonstrated integrity of the sterile barrier and functionality after 3 years of real-time aging (shelf-life is 3 years).",
"Testing demonstrated that the subject device performs as expected.",
"Testing demonstrated that the use of vacuum-assisted palpation is equivalent to the use of manual palpation by the predicate device.",
"At a vacuum level of 25.0 inHg, the minimum, maximum, and average measured peak suction force achieved was 0.753 lbf, 1.052 lbf, and 0.904 lbf, respectively."
]
},
"2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)": "The provided text describes non-clinical testing. It does not mention specific sample sizes for a test set in the context of clinical data or human subjects. The data provenance is not specified, but the testing appears to be laboratory-based (non-clinical).",
"3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/imaging device requiring expert ground truth establishment for a test set.",
"4. Adjudication method (e.g. 2+1, 3+1, none) for the test set": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/imaging device requiring an adjudication method.",
"5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/imaging device. No MRMC study was conducted or mentioned.",
"6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done": "Not applicable. This is a non-clinical study for a manual surgical instrument. No algorithm or standalone performance testing was mentioned.",
"7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)": "Not applicable. For this non-clinical study, ground truth was established by adherence to performance specifications and established biocompatibility standards.",
"8. The sample size for the training set": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/machine learning device requiring a training set.",
"9. How the ground truth for the training set was established": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/machine learning device. No training set was used or ground truth established for one."
}

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 3, 2024

Cypris Medical % Erin Gontang Senior Consultant Ram+ 2251 San Diego Avenue, B-257 San Diego, California 92110

Re: K240185

Trade/Device Name: Cypris eXact Suture Placement Device Regulation Number: 21 CFR 878.4800 Regulation Name: Manual Surgical Instrument For General Use Regulatory Class: Class I Product Code: GEJ Dated: January 23, 2024 Received: January 23, 2024

Dear Erin Gontang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Julie A. Morabito -S

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices

{2}------------------------------------------------

OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240185

Device Name Cypris eXact Suture Placement Device

Indications for Use (Describe)

The Cypris eXact Suture Placement Device is indicated for use in general soft tissue approximation when used with sutures.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY K240185

DATE PREPARED

May 1, 2024

MANUFACTURER AND 510(k) OWNER

Cypris Medical, Inc. 4541 N. Ravenswood Ave. #202 Chicago, IL 60640, USA Phone: +1 888-433-5002

OFFICIAL CONTACT Dan Holton, President & CEO

REPRESENTATIVE/CONSULTANT

Erin A. Gontang, Ph.D. Allison C. Komiyama, Ph.D., RAC ROM+ 2251 San Diego Ave, Suite B-257 San Diego, CA 92110, USA Phone: +1 412-816-8290 Email: egontang@rqmplus.com akomiyama@rqmplus.com

DEVICE INFORMATION

Trade/Proprietary NameCypris eXact Suture Placement Device
Common NameSuture Placement Device
Regulatory Class:I
Regulation Number/Name:21 CFR 878.4800/Manual Surgical Instrument for General Use
Product Code:GEJ
Classification PanelGeneral & Plastic Surgery
Premarket ReviewOffice of Health Technology 4(Surgical and Infection Control Devices)Division of Health Technology 4B(Infection Control and Plastic and Reconstructive Surgery)

{5}------------------------------------------------

PREDICATE DEVICE IDENTIFICATION

The Cypris eXact Suture Placement Device is substantially equivalent to the following:

510(k) NumberPredicate Device Name/ ManufacturerPredicate/Reference
Class I, 510(k) ExemptCypris eXact 5.1 Suture Placement Device/ Cypris Medical, Inc.Primary Predicate
K233355Cypris eXact Suturing System/ Cypris Medical, Inc.Reference Device
K994290Bard Endoscopic Suturing System/ C.R. Bard, Inc.Reference Device

The predicate and reference devices have not been subject to a design related recall.

DEVICE DESCRIPTION

The Cypris eXact Suture Placement Device is for use in general soft tissue approximation by surgeons, and it is single-use and provided sterile. It is a hand-held device used for the placement of a surgical suture, and the device consists of a handle, needle driver, and tissue capture port. The handle is designed for manual manipulation of the suture placement device, allowing the surgeon to advance the device to the desired deployment site. The needle driver functions to pass the suture when it is advanced and retracted. The tissue capture port consists of a capture chamber, needle, and hook assembly, and holds the tissue to be sutured during either manual palpation or vacuum-assisted palpation of tissue into the tissue capture port. Suture placement is achieved by advancing and retracting the needle driver.

INDICATIONS FOR USE

The Cypris eXact Suture Placement Device is indicated for use in general soft tissue approximation when used with sutures.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Cypris Medical believes the Cypris eXact Suture Placement Device is substantially equivalent to the primary predicate device based on the information summarized here:

The Cypris eXact Suture Placement Device has the same intended use as the primary predicate device, and they are identical with respect to formulation, processing, and sterilization. The difference between the subject and predicate devices is that the Cypris eXact Suture Placement Device is packaged with a vacuum connector, which gives the option of connecting the suture placement device to a vacuum source.

{6}------------------------------------------------

SUBSTANTIAL EQUIVALENCE

The Cypris eXact Suture Placement Device is substantially equivalent to the primary predicate device in its intended use, as well as formulation, processing, and sterilization. The substantial equivalence table for the comparison of the subject and predicate devices is provided below.

Subject DevicePredicate Device
Cypris Medical, Inc.Cypris Medical, Inc.
Cypris eXact Suture Placement DeviceCypris eXact 5.1 Suture Placement Device
K240185Class I (510(k) Exempt)
Indications for UseThe Cypris eXact Suture Placement Deviceis indicated for use in general soft tissueapproximation when used with sutures.The Cypris eXact 5.1 Suture PlacementDevice is indicated for use in general softtissue approximation.
Regulation Number21 CFR 878.480021 CFR 878.4800
Regulation DescriptionManual Surgical Instrument for General UseManual Surgical Instrument for General Use
Product Code/NameGEJ / Carrier, LigatureGEJ / Carrier, Ligature
ClassificationClass IClass I
510(k) ExemptNoYes
Device DescriptionThe Cypris eXact Suture Placement Deviceis for use in general soft tissue approximationby surgeons, is single-use, and is providedsterile. It is a hand-held device used for theplacement of a surgical suture, and the deviceconsists of a handle, needle driver, and tissuecapture port.The Cypris eXact Suture Placement Deviceis for use in general soft tissue approximationby surgeons, is single-use, and is providedsterile. It is a hand-held device used for theplacement of a surgical suture, and the deviceconsists of a handle, needle driver, and tissuecapture port.
Intended UseGeneral Soft Tissue ApproximationGeneral Soft Tissue Approximation
Device MaterialsPolycarbonateStainless SteelPolycarbonateStainless Steel
Single Use DeviceYesYes
Tissue CaptureManual PalpationManual Palpation
MethodVacuum-Assisted Palpation
Suture DeploymentThe suture is threaded into needle proximalof tissue capture port. When deployed, theneedle passes through tissue, pushing thesuture through the tissue. The suture iscaptured in the distal end of the device, andthen the needle is retractedThe suture is threaded into needle proximalof tissue capture port. When deployed, theneedle passes through tissue, pushing thesuture through the tissue. The suture iscaptured in the distal end of the device, andthen the needle is retracted
Sterilization MethodEthylene Oxide (EtO)Ethylene Oxide (EtO)
BiocompatibilityCytotoxicitySensitizationIrritationAcute Systemic ToxicityMaterial-Mediated PyrogenicityCytotoxicitySensitizationIrritationAcute Systemic ToxicityMaterial-Mediated Pyrogenicity
Reference DeviceReference Device
Cypris Medical, Inc.C.R. Bard, Inc.
Cypris eXact Suturing SystemBard Endoscopic Suturing System
K233355K994290
Indications for UseCypris eXact Suture PP (Polypropylene)The Cypris eXact Suture (PP) is indicated foruse in general soft tissue approximationand/or ligation, including use incardiovascular, ophthalmic and neurologicalprocedures.Cypris eXact Suture PTFEFor endoscopic placement of suture(s) in thesoft tissue of the esophagus and stomach andfor approximation of tissue for the treatmentof symptomatic Gastroesophageal RefluxDisease.
(Polytetraflouroethylene)The Cypris eXact Suture (PTFE) is indicatedfor use in all types of soft tissueapproximation and/or ligation, includingdental and general surgeries.
Cypris eXact Suture PDO (Polydioxanone)The Cypris eXact Suture (PDO) is indicatedfor use in general soft tissue approximationand/or ligation, but not for use incardiovascular or neurological tissues,microsurgery or ophthalmic surgery.
The Cypris eXact 5.1 Suture PlacementDevice can be used in conjunction witheXact Sutures, which are indicated for use ingeneral soft tissue approximation and/orligation.
Regulation Number21 CFR 878.484021 CFR 878.501021 CFR 878.503521 CFR 878.480021 CFR 876.1500
RegulationDescriptionAbsorbable polydioxanone surgical sutureNonabsorbable polypropylene surgical sutureNonabsorbable expandedpolytetrafluoroethylene surgical sutureManual Surgical Instrument for General UseEndoscope and Accessories
Product Code/NameNEW / Suture, Surgical, Absorbable,PolydioxanoneGAW / Suture, Nonabsorbable, Synthetic,PolypropyleneNBY / Suture, Surgical, Nonabsorbable,Expanded, PTFEGEJ / Carrier, LigatureODE / Endoscopic Suture/Plication System,Gastroesophageal Reflux Disease (GERD)
ClassificationClass IIClass II
510(k) ExemptNoNo
Device DescriptionThe Cypris eXact Suturing System iscomprised of the Cypris eXact Suture and theCypris eXact 5.1 Suture Placement Device.The Cypris eXact Suture and the CyprisThe Bard Endoscopic Suturing Systemconsists of a capsule assembly with suctiontubing and a fixed head knot pusher thatattaches to the distal end of a flexible
eXact 5.1 Suture Placement Device are forendoscope or a through-the-scope knot
Intended Useuse in general soft tissue approximation by surgeons, are single-use, and are provided sterile.pusher, plus a needle assembly, pusher wire, guidewire and suture cutter that all pass through an endoscope's biopsy channel.
General Soft Tissue ApproximationPlace Stitches and Tie Suture Material to Approximate Soft Tissue Under Endoscopic Visualization
Device MaterialsPolycarbonateStainless SteelPlasticStainless Steel
Single Use DeviceYesYes
Tissue Capture MethodManual PalpationVacuum-Assisted Palpation
Suture DeploymentThe suture is threaded into needle proximal of tissue capture port. When deployed, the needle passes through tissue, pushing the suture through the tissue. The suture is captured in the distal end of the device, and then the needle is retractedThe handle assembly is manually activated, advancing the needle through the tissue, at which time the suture tag is pushed through the needle assembly by the pusher wire until the suture tag exits the needle and is captured within the end cap of the capsule assembly. The handle and needle are then retracted.
Sterilization MethodEthylene Oxide (EtO)Ethylene Oxide (EtO)
BiocompatibilityCytotoxicitySensitizationIrritationAcute Systemic ToxicityMaterial-Mediated PyrogenicityCytotoxicitySensitizationIrritationAcute Systemic ToxicityMaterial-Mediated Pyrogenicity

{7}------------------------------------------------

The named reference devices provided additional information to help substantiate the safety and effectiveness of the subject device and the use of vacuum-assisted palpation. Information about both reference devices is provided below.

{8}------------------------------------------------

SUMMARY OF NON-CLINICAL TESTING

The Cypris eXact Suture Placement Device underwent biocompatibility testing as an external communicating device with limited contact (≤24 hours) with tissue. The following biocompatibility endpoints were evaluated: cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity.

The Cypris eXact Suture Placement Device was tested to evaluate the integrity of the sterile barrier and the functionality of the device following 3 years of real-time aging. The shelf-life of the Cypris eXact Suture Placement Device is 3 years.

Performance testing of the Cypris eXact Suture Placement Device and a comprehensive vacuum safety evaluation were completed. Testing demonstrated that the subject device performs as expected and that the use of vacuum-assisted palpation is equivalent to the use of manual palpation by the predicate device. Peak suction force was measured at a vacuum level of 25.0 inHg, a likely vacuum level used in clinical settings. At this vacuum level, the minimum, maximum, and average measured peak suction force achieved by the Cypris eXact Suture Placement Device was 0.753 1bf, 1.052 1bf, and 0.904 1bf, respectively.

CONCLUSION

Based on the totality of evidence, including performance testing and a comprehensive vacuum safety evaluation. Cypris Medical concluded that use of vacuum-assisted palpation with the subject device does not raise different questions of safety and effectiveness relative to the predicate device. The Cypris eXact Suture Placement Device is determined to be substantially equivalent to the predicate device.

§ 878.4800 Manual surgical instrument for general use.

(a)
Identification. A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.