K Number
K240045
Device Name
QStat Cartridge
Manufacturer
Date Cleared
2024-03-27

(82 days)

Product Code
Regulation Number
864.5430
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous or arterial whole blood sample using the Quantra Hemostasis Analyzer. The QStat Cartridge includes tests to assess coagulation via the intrinsic and extrinsic pathways and includes a test with tranexamic acid to evaluate clot lysis characteristics.

The OStat Cartridge is intended for in vitro diagnostic use by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood by means of the following functional parameters: Clot Time (CT), Clot Stiffness (CS), Fibrinogen Contribution to Clot Stiffness (FCS), Platelet Contribution to Clot Stiffness (PCS), and Clot Stability to Lysis (CSL).

The QStat Cartridge is indicated for the evaluation of blood coagulation and clot lysis in patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions in trauma and liver transplantation procedures.

Results obtained with the QStat Cartridge should not be the sole basis for patient diagnosis.

For prescription use only.

Device Description

The QStat Cartridge is a single-use, multi-channel disposable plastic cartridge used with the Quantra Hemostasis Analyzer for the evaluation of blood coagulation and clot lysis. The measurements are performed in four test channels of the disposable cartridge which enable differential testing with different sets of reagents without the need for any reagent preparation or controlled pipetting. The cartridge utilizes a citrated evacuated blood collection tube filled with a patient whole blood sample The proprietary technology SEER Sonorheometry measures the evolution of shear modulus (i.e., clot stiffness) in all four channels as a function of time. The QStat Cartridge is intended for use in patients 18 years or older by professionals in a hospital setting (point of care or laboratory) to assess possible hypocoagulable and hypercoagulable conditions in trauma and liver transplantation procedures.

Each QStat Cartridge is pre-filled with lyophilized reagent beads individually sealed in an airtight pouch. After a QStat Cartridge is removed from its primary packaging, it is inserted into the instrument dock. A whole blood sample, collected in a 3.2% sodium citrate anticoagulant blood collection tube (minimum volume 2.7 mL), is attached directly to the cartridge and the test is initiated using the touch screen interface on the Quantra Hemostasis Analyzer. The cartridge is the only component of the Quantra System that is in direct contact with blood. The fluidic system within the instrument draws the sample into the cartridge where it is warmed to 37°C, aliquoted, introduced and mixed with the lyophilized reagents, and analyzed. When the test is complete, the cartridge is released from the dock to be disposed of in an appropriate biosafety sharps container.

AI/ML Overview

The provided text is a 510(k) Premarket Notification from the FDA regarding the HemoSonics QStat Cartridge. The purpose of this submission is to expand the approved sample matrices to include arterial whole blood, in addition to the previously cleared venous whole blood.

The document does not detail a study proving the device meets acceptance criteria in the manner of an AI/ML device, as it concerns an in vitro diagnostic (IVD) device for measuring blood viscoelastic properties. Therefore, many of the requested items (e.g., number of experts, adjudication methods, MRMC study, training sets, ground truth establishment for AI/ML) are not applicable or described in this type of submission.

However, based on the information provided, we can infer some aspects relevant to device performance and the expansion of its indications for use.

Here's an analysis based on the provided text, addressing the applicable points and explaining why others are not relevant in this context:

1. A table of acceptance criteria and the reported device performance

The document defines the device's function and indications for use, but it does not provide a table of quantitative acceptance criteria and reported performance metrics for the device itself (e.g., accuracy, precision) in the context of an AI/ML algorithm evaluation. This is a 510(k) submission for a modification to an existing IVD, primarily focused on expanding the sample type. The "acceptance criteria" here refer to demonstrating substantial equivalence for the new claim (arterial blood).

What is reported is the intent to demonstrate equivalency between arterial and venous whole blood samples.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document states: "HemoSonics is submitting this Special 510(k) to demonstrate the equivalency of arterial and venous whole blood samples analyzed on the Quantra Hemostasis Analyzer with the QStat Cartridge in order to expand the sample matrices accepted on this system."

While it explicitly states the intent to perform a study for equivalency, the sample size and data provenance (country, retrospective/prospective) are not provided in this document. This information would typically be detailed in a separate study report submitted as part of the 510(k) supporting data, which is not included in this high-level summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This question is typically relevant for interpretative devices, especially those using AI/ML where human expert interpretation forms the ground truth. For an IVD like the QStat Cartridge, which measures objective viscoelastic properties of blood, the "ground truth" is established by the accuracy and precision of the measurement itself against a reference method or known values, not by expert consensus on image interpretation. Therefore, this concept of "experts" to establish ground truth does not directly apply here.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Similar to point 3, adjudication methods are used in AI/ML performance studies where multiple human readers might disagree, and a consensus process is needed to establish ground truth. This is not applicable to a laboratory device that performs quantitative measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC study was not done, nor would it be relevant for this type of device. MRMC studies are specific to evaluating the impact of AI assistance on human reader performance in tasks like image interpretation (e.g., radiology for diagnostic accuracy). The QStat Cartridge is an IVD that provides quantitative measurements of blood coagulation; it does not involve human "readers" interpreting cases in the sense of an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is also more applicable to AI/ML software. The QStat Cartridge is a physical device that performs measurements. Its performance is evaluated fundamentally in a standalone manner (the device's ability to accurately measure blood properties), but not in the context of an "algorithm only" as understood for AI/ML. The device generates quantitative parameters (Clot Time, Clot Stiffness, etc.). An "algorithm" within the device would process sensor data to yield these parameters, and the evaluation would focus on the accuracy of these output parameters.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For an IVD like the QStat Cartridge, the "ground truth" for its measurements would be established through:

  • Reference Methods: Comparison of the QStat Cartridge's measurements (Clot Time, Clot Stiffness, etc.) against established, validated reference methods for measuring blood viscoelastic properties or related coagulation parameters.
  • Known Samples: Testing with samples having known or characterized coagulation properties (e.g., calibrated controls, samples spiked with known substances).
  • Clinical Correlation: While not a direct "ground truth" for the measurement itself, the clinical utility of the device's outputs is validated by correlating its results with patient clinical status or outcomes.

The document states the device provides "semi-quantitative indications" and its results "should not be the sole basis for patient diagnosis," indicating its role as a supportive diagnostic tool. The specific ground truth used for the arterial vs. venous equivalency study is not detailed here, but it would involve comparing measurements from paired arterial and venous samples from the same patient.

8. The sample size for the training set

This question is applicable to AI/ML models. Since this is an IVD device measuring physiological parameters, the concept of a "training set" for an AI model as typically distinguished from a "test set" does not apply in the same way. The device's measurement principles are based on established biophysical principles (SEER Sonorheometry), not on training a machine learning algorithm on a dataset. Any internal "algorithms" would be deterministic or model-based, not learned from data in the AI sense.

9. How the ground truth for the training set was established

Not applicable, as explained in point 8.

§ 864.5430 Coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients.

(a)
Identification. A coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients is an in vitro diagnostic device used to evaluate blood coagulation, fibrinolysis, or both, in perioperative patients, as an aid in the assessment of coagulopathies when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include detailed documentation of, and results from, the following:
(i) A study assessing precision using protocols determined to be acceptable by FDA, to cover the measurement range for each reported parameter (test output). Testing must include native specimens with coagulation profiles representative of the intended use population. In order to cover the measuring range, testing may include a limited number of contrived specimens, not to exceed 10 to 20 percent, or as otherwise deemed appropriate by FDA. The contrived specimens must be prepared to resemble clinical specimens. This testing must evaluate repeatability and reproducibility and provide assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system;
(ii) Studies that demonstrate the performance of each parameter (test output) throughout the claimed measurement range, to include linearity studies or dose-response studies, as applicable to the parameter (test output);
(iii) Potential interferent study that includes evaluation of hemolyzed and lipemic samples as potential interferents; exogenous and endogenous interferents associated with each patient population intended for use with the device, and which might be expected to affect assay performance, must be evaluated; and potential interferents that are specific for, or related to, the technology or methodology of the device. Evaluation of all potential interferents must be performed using a protocol determined to be acceptable to the FDA (
e.g., an FDA-recognized standard) and include both normal and abnormal specimens covering coagulation profiles representative of the intended use population;(iv) A study that evaluates specimen stability under the intended conditions for specimen collection, handling, and storage, using samples that cover the coagulation profiles representative of the intended use population, and using protocols determined to be acceptable by FDA;
(v) A multisite clinical study, determined to be acceptable by FDA, demonstrating performance, relative to clinically relevant and clinically validated laboratory test(s) for each parameter (test output). Further, the study must meet all of the following criteria:
(A) The study must be performed in the intended use population and include representation from all patient populations for whom the device is intended to be used. Potential endogenous and exogenous interferents for each target patient population must be evaluated or known prior to the study;
(B) The study must be conducted at a minimum of three external sites representative of the intended use setting by the intended operators;
(C) Test samples must be collected at time intervals relevant to the device's use in the intended use population;
(D) Clinical specimens, which cover coagulation profiles representative of the intended use population, must be evaluated at each of the three clinical sites in the study;
(E) Analysis of the concordance of clinical interpretation of patient coagulation status made from individual test parameter (test output) results as compared to clinical interpretation of coagulation status from a clinically relevant laboratory test or tests (
e.g., a comparative viscoelastic device or standard laboratory tests) must be conducted; and(F) Expected (reference) values for each parameter (test output) must be demonstrated by testing a statistically appropriate number of samples from apparently healthy normal individuals;
(vi) For a device with a user interface that has information that needs to be interpreted by the user in correctly using the device to achieve the intended test results or a device that does not provide a final output that is a comprehensive interpretation of all parameter (test output) results, a study evaluating the ability of device users to correctly interpret results;
(vii) For any device indicated to guide blood product use, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding blood product use; and
(viii) For any device indicated to guide use of medication, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding use of medication.
(2) The labeling required under § 809.10(b) of this chapter must include the following:
(i) A summary of results from the study required by paragraph (b)(1)(i) of this section, including repeatability, reproducibility, and assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system.
(ii) The claimed measurement range of each parameter (test output), as supported by demonstrated performance of the parameter (test output) throughout the claimed measurement range, including studies required by paragraphs (b)(1)(i) through (iii) and (v) of this section, and, if applicable, paragraphs (b)(1)(vii) and (viii) of this section.
(iii) Identification of known interferents, including all endogenous, exogenous, technology-specific, and patient population-specific interferents, specific to each parameter (test output). The information must include the concentration(s) or level(s) at which interference was found to occur and the concentration range or levels at which interference was not found to occur.
(iv) Information regarding the multisite clinical study required by paragraph (b)(1)(v) of this section, including:
(A) Each patient population evaluated;
(B) Each intended use setting and the operators;
(C) A summary of the results, including the concordance analysis to clinically relevant laboratory test(s); and
(D) Demonstrated expected (reference) values for each parameter (test output).
(3) The labeling required under § 809.10 of this chapter must include the following:
(i) A limiting statement that the result(s) from the device is(are) not intended to be used as the sole basis for a patient diagnosis.
(ii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(vii) of this section that specifically validate an indication for the device's use in guiding blood product use, a limiting statement that the device has not been evaluated to guide blood product use.
(iii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(viii) of this section that specifically validate an indication for the device's use in guiding use of medication, a limiting statement that the device has not been evaluated to guide use of medication.