AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Nitrile Examination Glove 3.5, Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

Powder Free Nitrile Examination Gloves 3.5, Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large, Extra Large. Gloves meet the specification of ASTM D6319- 19 and have been tested for resistance to permeation by chemotherapy drugs and Fentanyl Citrate as per ASTMD6978- 05(2019). The gloves are single use, disposable, and provided non-sterile.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for a medical device: "Powder Free Nitrile Examination Glove 3.5, Blue Color, Non-Sterile, and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate." This is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive study of an AI algorithm. Therefore, many of the requested points related to AI/algorithm performance are not applicable.

Here's the information extracted from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria and reported device performance are primarily based on established ASTM and ISO standards for medical gloves.

TestStandard (Acceptance Criteria)Reported Device Performance (Result)
Physical Dimensions
LengthASTM D6319-19: Minimum 220mm for size XS and S, 230mm for size M, L, XLPass
Palm WidthASTM D6319-19: XS: 70±10mm, S: 80±10mm, M: 95±10mm, L:110±10mm, XL: 120±10mm, XXL: 130±10mmPass
ThicknessASTM D6319-19: Finger: 0.05mm (min), Palm: 0.05mm (min)Pass
Physical Properties
Before AgingASTM D6319-19, ASTM D412-16 (2021): Tensile strength: 14MPa, min; Ultimate elongation: 500%, minPass
After AgingASTM D6319-19, ASTM D412-16 (2021): Tensile strength: 14MPa, min; Ultimate elongation: 400%, minPass
Freedom from HolesASTM D6319-19, ASTM D5151-19 (Water leak test): G-I, AQL 2.5 (ISO 2859- 1)Pass
Residual PowderASTM D6319-19, ASTM D6124-06 (2017): Max 2mg/glovePass
Biocompatibility
Skin SensitizationISO 10993-10:2021: No Skin sensitizationThe sample extract elicits no sensitization reactions under the test conditions.
IrritationISO 10993-23:2021: No Skin irritationUnder the conditions of this study, the irritation response category of the test article is classified as Negligible for polar extract and Negligible for non-polar extract.
Acute Systemic ToxicityISO 10993-11:2017: Subject showed no adverse biological reactionNo death or acute systemic toxicity was observed in the mice which received injection of extract.
Chemotherapy Drugs Permeation (ASTM D6978-05 (2019))Minimum Breakthrough Detection Time (Minutes)Minimum Breakthrough Detection Time (Minutes)
Carmustine (BCNU) (3.3mg/ml)Refer to table for specific times (N/A – no general acceptance criteria provided, specific data shown)1.3 (Note: "Warning: Do not use with Carmustine and Thiotepa" due to extremely low permeation times)
Cisplatin (1mg/ml)>240>240
Cyclophosphamide (20mg/ml)>240>240
Dacarbazine (10.0 mg/ml)>240>240
Doxorubicin HCl (2 mg/ml)>240>240
Etoposide (20mg/ml)>240>240
Fluorouracil (50mg/ml)>240>240
Paclitaxel (6mg/ml)>240>240
Thiotepa (10mg/ml)Refer to table for specific times (N/A – no general acceptance criteria provided, specific data shown)1.1 (Note: "Warning: Do not use with Carmustine and Thiotepa" due to extremely low permeation times)
Fentanyl Citrate Permeation (ASTM D6978-05 (2019))Minimum Breakthrough Detection Time (Minutes)Minimum Breakthrough Detection Time (Minutes)
Fentanyl Citrate Injection, 100mcg/2ml>240>240

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes used for each individual test (e.g., number of gloves for physical dimensions, number of test replicates for permeation). However, the standard ASTM D6319-19 for physical dimensions and freedom from holes, and ASTM D6978-05 for permeation, typically define the sampling plans.

The data provenance is from "Yue Kang Anhui Medical Products Co., Ltd" in China. The studies are non-clinical performance tests, conducted to verify the device meets design specifications and relevant standards. They are retrospective in the sense that they are tests performed on manufactured products, not prospective clinical trials.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This document is for a medical glove, and the "ground truth" for the test set is established by the specifications of industrial standards (ASTM, ISO) and laboratory testing methodologies, not by expert medical review or consensus on images/data.

4. Adjudication Method for the Test Set

Not applicable. As described above, the compliance is determined by adherence to standardized test methods and their defined acceptance criteria. There is no human adjudication process involved in these performance tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

Not applicable. This is a submission for a physical medical device (glove) and does not involve an AI algorithm or human-in-the-loop performance evaluation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This submission is for a medical glove, not an AI algorithm.

7. The Type of Ground Truth Used

The "ground truth" here is defined by the technical specifications and performance requirements outlined in the referenced industry standards (ASTM D6319-19, ASTM D5151-19, ASTM D6124-06, ASTM D412-16, ASTM D6978-05, ISO 10993-10, ISO 10993-23, ISO 10993-11). These standards represent established, objective, and reproducible test methods for evaluating medical glove performance and biocompatibility.

8. The Sample Size for the Training Set

Not applicable. This is a physical device, not an AI model, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. There is no training set for this type of device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.