(83 days)
The Portrait Mobile Monitoring Solution is intended to acquire, store, calculate, display and export patient monitoring data as well as provide real time alarming for monitoring adult and pediatric patients (3 years of age and older, and weighing more than 10 kg).
Physiological parameters and waveforms supported are:
- Pulse oximetry (SpO2/pulse rate)
- Respiration rate (RR)
Continuous pulse oximetry and respiration rate monitoring may be used for patients at risk of cardiorespiratory and infectious complications.
The Portrait Mobile Monitoring Solution is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
This device is not an Apnea monitor (i.e., do not rely on the device for detection of alarm for the cessation of breathing).
This device should not be used for life sustaining/supporting purposes.
The Portrait Mobile Monitoring Solution is not intended for use in a controlled Magnetic Resonance (MR) environment.
The Portrait Mobile Monitoring Solution V1.1 is a wireless monitoring system for monitoring SpO2, pulse rate and respiration rate of adult and pediatric patients. The system can be used for monitoring adult and pediatric patients (3 years of age and older and weighing more than 10 kg) within a hospital or healthcare facility. The system acquires, stores, calculates, displays, and exports patient physiological parameter data, alarms, and information. It supports pulse oximetry (SpO2/pulse rate) and respiration rate parameters. Measurement values are displayed as graphic or numeric values, like waveforms and numbers, and when applicable, also as alarm messages.
The Portrait Mobile Monitoring Solution V1.1 consists of the following general categories of medical devices:
Central Monitoring Devices:
- Portrait Core Services hosted on the GE HealthCare nonmedical device Edison Health Link platform. Portrait Core Services is a set of software services that enable the communication and interaction of the system components and are capable of integrating into existing healthcare facility infrastructure and clinical information systems.
- Portrait Central Viewer Application software hosted on a Windows off-the-shelf computer. Portrait Central Viewer Application provides the ability to view patient real-time and historical data, capable of displaying data from multiple patients.
- Portrait Clinical Alarming Unit provides audible alarms at each Central Viewer.
Mobile Monitoring Devices:
- Portrait Mobile Patient Monitor Hardware, a battery-powered, wireless, hand-held patient monitor. The Portrait Mobile Patient Monitor is a completely wireless, hand-held device that is capable of acting as a standalone patient monitor including alarms, with a 3.7-inch capacitive touchscreen capable of displaying numeric data and waveforms for SpO2, Pulse Rate (PR), and Respiration Rate (RR). Compared to the predicate (K230626), the Portrait Mobile Patient Monitor has been split into separate hardware and software components for serviceability. The Portrait Mobile Patient Monitor Hardware is shipped with no clinical software installed. The clinical software (i.e., Portrait Mobile Patient Monitor Software) is installed at the customer site before the device is put into use. The Portrait Mobile Patient Monitor Hardware when the Portrait Mobile Patient Monitor Software is installed (prior to patient use) is then what we will refer to as the Portrait Mobile Patient Monitor. Portrait Wearable SpO2 sensors for acquiring SpO2 and pulse rate data from a patient wirelessly.
- Portrait Mobile Patient Monitor Software is the clinical software which gets installed at the customer site on the Portrait Mobile Patient Monitor Hardware. The Portrait Mobile Patient Monitor Software is this submission is equivalent to the software installed on the predicate Portrait Mobile Patient Monitor cleared in K230626 with minor changes as discussed in this submission.
- Portrait Wearable Respiration Rate sensor and Portrait RR electrode patch for acquiring impedance respiration data from a patient wirelessly.
- Portrait Sensor battery used for powering the wearable sensors and provide wireless communication to the Portrait Mobile Patient Monitor.
- Portrait Bedside Charger for charging the Portrait Sensor Batteries and Portrait Mobile Patient Monitor (including when the Portrait Mobile Patient Monitor is in clinical use).
- Portrait SpO2 Attachment accessory band which provides means to secure the SpO2 sensors to the patient's wrist
- Portrait Mobile Patient Monitor Pouch, which allows the Portrait Mobile Patient Monitor to be carried while the patient is ambulatory
The provided text does not contain specific acceptance criteria or a detailed study proving the device meets said criteria in a table format. It focuses on demonstrating substantial equivalence to a predicate device, K230626 Portrait Mobile Monitoring Solution v1.0, through non-clinical testing and updates to software and standards compliance.
Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is not:
1. Table of Acceptance Criteria and Reported Device Performance
This information is not explicitly provided in a table format in the given document. The document lists "Specifications" for the proposed device and compares them to the predicate device, stating "Identical" or "Equivalent" for most technical specifications and performance characteristics, rather than specific acceptance criteria and detailed performance results from a study.
A table comparing the proposed device specifications to the predicate device is provided on pages 19-27, with "Identical" or "Equivalent" noted for each entry. The key takeaway is that the proposed device maintains the performance characteristics of the predicate.
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "Bench testing related to software, hardware and performance" but does not specify sample sizes for test sets.
- Sample Size for Test Set: Not specified for individual performance tests.
- Data Provenance: The document does not explicitly state the country of origin or whether the data is retrospective or prospective for the non-clinical tests. It focuses on compliance with FDA guidance and recognized standards.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document for the non-clinical performance and validation tests mentioned. The document describes compliance with standards and guidance documents but does not detail the process of establishing ground truth for testing data, nor the number or qualifications of experts involved in such a process.
4. Adjudication Method for the Test Set
This information is not provided in the document.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
This type of study is not mentioned in the document. The submission explicitly states: "The Portrait Mobile Monitoring Solution v1.1 does not contain clinical testing data."
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done
The document describes "SpO2 measurement," "Pulse Rate (PR) measurement," and "Impedance Respiration" for which "no changes have been made to the algorithms since K230626, only regression testing performed." This implies that the algorithms for these measurements are operating in a standalone capacity (device only), and their performance compared to the predicate device has been affirmed through regression testing. However, no specific standalone performance metrics or studies are detailed beyond this.
7. Type of Ground Truth Used
For the SpO2 and Pulse Rate measurements, the document mentions:
- "low perfusion condition (this testing covers the full accuracy range in challenging low-perfusion conditions using a simulator, proving that there has been no system regression)"
This indicates that simulators were used to establish reference data for certain conditions (low perfusion) for SpO2 and PR. For other measurements, the ground truth methodology is not explicitly specified beyond compliance with recognized standards.
8. Sample Size for the Training Set
This information is not provided. The document states that "no changes have been made to the SpO2 algorithms since K230626, only regression testing performed" and similarly for Respiration Rate algorithms. This suggests the underlying algorithms were developed and potentially trained earlier, but details of training data are absent.
9. How the Ground Truth for the Training Set was Established
This information is not provided, as details about training sets are omitted.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, the text "FDA U.S. FOOD & DRUG ADMINISTRATION" is displayed in blue. The word "FDA" is in a larger font size than the rest of the text.
GE Medical Systems Information Technologies, Inc. Joel Kent Director, Regulatory Affairs, Strategy 9900 Innovation Drive Wauwatosa, Wisconsin 53226
March 20, 2024
Re: K234130
Trade/Device Name: Portrait™ Mobile Monitoring Solution consists of:
- Portrait™ Central Viewer Application (Portrait CVAXB) ●
- Portrait™ Core Services (Portrait CSSXB) ●
- Portrait™ Clinical Alarming Unit (Portrait CAU01)
- Portrait™ Mobile Patient Monitor Hardware (Portrait HUBXB)
- Portrait™ Mobile Patient Monitor Software (Portrait HSWXB)
- Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SA01)
- Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SP01) ●
- . Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-W01, Portrait SpO2 P-SE01)
- Portrait™ SpO2 Attachment Accessory Band (Portrait AAB01)
- Portrait™ Wearable Respiration Rate Sensor (Portrait RR P-RR01)
- . Portrait™ RR Electrode Patch (Portrait RRP01)
- Portrait™ Sensor Battery (Portrait SBT01) ●
- . Portrait™ Bedside Charger (Portrait BCH01)
- Portrait™ Mobile Patient Monitor Pouch (Portrait MMP01) ●
Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, MSX, DRG, BZQ, DQA Dated: December 28, 2023 Received: December 28, 2023
Dear Joel Kent:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls
{1}------------------------------------------------
provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K234130
Device Name
Portrait™ Mobile Monitoring Solution
Indications for Use (Describe) Portrait™ Mobile Monitoring Solution:
The Portrait Mobile Monitoring Solution is intended to acquire, store, calculate, display and export patient monitoring data as well as provide real time alarming for monitoring adult and pediatic patients (3 years of age and older, and weighing more than 10 kg).
Physiological parameters and waveforms supported are:
- · Pulse oximetry (SpO2/pulse rate)
· Respiration rate (RR)
Continuous pulse oximetry and respiration rate monitoring may be used for patients at risk of cardiorespiratory and infectious complications.
The Portrait Mobile Monitoring Solution is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
This device is not an Apnea monitor (i.e., do not rely on the device for the cessation of breatling), This device should not be used for life sustaining/supporting purposes.
The Portrait Mobile Monitoring Solution is not intended for use in a controlled Magnetic Resonance (MR) environment.
The Portrait™ Mobile Monitoring Solution consists of:
Portrait™ Central Viewer Application (Portrait CVAXB):
The Portrait Central Viewer Application (Portrait CVAXB) provides monitoring station capability running as an application for the Portrait Mobile Monitoring Solution that meets minimum system requirements. It provides the ability to view real-time data for multiple patients and historical data for a single patient including configurable visual and audible alarm notifications.
The Portrait Central Viewer Application is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Portrait™ Core Services (Portrait CSSXB):
The Portrait Core Services (Portrait CSSXB) are a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems. The Portrait Core Services can transmit patient physiological trends and numerics (IHE PCD DEC), waveforms (IHE PCD WCM), and alarm events (IHE PCD ACM) outbound. The Portrait Core Services can also receive HL7 ADT information to admit patients to the Portrait Mobile Monitoring Solution.
Portrait™ Clinical Alarming Unit (Portrait CAU01)
The Portrait Clinical Alarming Unit (Portrait CAU01) is a required accessory to the Portrait Central Viewer Apolication that provides audio alarming capability.
{4}------------------------------------------------
The Portrait Clinical Alarming Unit is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Portrait™ Mobile Patient Monitor Hardware (Portrait HUBXB):
The Portrait Mobile Patient Monitor Hardware (Portrait HUBXB) is intended for use with adult and pediatric patients as a handheld, mobile device which supports running the Portrait Monitor Software to monitor patients, and to provide and visual alarms. The Portrait Mobile Patient Monitor Hardware enables non-invasive continuous vital sign montoring of patients by acquiring signals wirelessly from Portrait wearable sensors, as well as wireless clinical and service data communication.
The Portrait Mobile Patient Monitor Hardware is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Mobile Patient Monitor Hardware is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ Mobile Patient Monitor Software (Portrait HSWXB):
The Portrait Mobile Patient Monitor Software (Portrait HSWXB) is intended to run on Portrait Monitor Hardware for the continuous monitoring of oxygen saturation (SpO2), pulse rate (PR) and respiration rate (RR) parameters. Portrait Mobile Patient Monitor Software is intended for use with adult and pediatic patients (3 years of age and older, and weighing more than 10 kg). It enables non-invasive continuous monitoring of patients by acquiring signals from Portrait wearable sensors, as well as displaying trends, events, and a QR code containing patient demographics and parameter data. When installed on the Portrait Monitor Hardware, the Portrait Mobile Patient Monitor Software provides real-time, trend and event data to Portrait Mobile Patient Mobile Patient Monitor Software, when configured to do so, also enables the Portrait Monitor Hardware to provide audible and visual alarms locally.
The Portrait Mobile Patient Monitor Software is intended for use under the direct supervision of a licensed healtheare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SA01):
The Portrait Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SA01) is intended for use with adult and pediatric patients (3 years of age and older, and weighing more than 30 kg) for continuous physiologic monitoring of oxygen saturation (SpO2) and pulse rate (PR) parameters. The Portrait Wearable Pulse Oximetry Sensor acquires parameter data from the patient and transmits it to the Portrait Sensor Battery for communication to a host device through the Medical Body Area Network (MBAN) connection.
The Portrait Wearable Pulse Oximetry Sensor is intended for use under the direct supervision of a licensed healtheare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Wearable Pulse Oximetry Sensor is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SP01):
The Portrait Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SP01) is intended for use with pediatric patients (3 years of age and older, and weighing 15 kg to 30 kg) for continuous physiologic monitoring of oxygen saturation (SpO2) and pulse rate (PR) parameters. The Portrait Wearable Pulse Oximetry Sensor acquires parameter data from the patient and transmits it to the Portrait Sensor Battery for communication to a host device through the Medical Body Area Network (MBAN) connection.
The Portrait Wearable Pulse Oximetry Sensor is intended for use under the direct supervision of a licensed healtheare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
{5}------------------------------------------------
The Portrait Wearable Pulse Oximetry Sensor is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-W01, Portrait SpO2 P-SE01):
The Portrait Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SE01, Portrait SpO2 P-W01) is intended for use with adult and pediatric patients (3 years of age and older, and weighing more than 10 kg) for continuous physiologic monitoring of oxygen saturation (SpO2) and pulse rate (PR) parameters. The Portrait Wearable Pulse Oximetry Sensor acquires parameter data from the patient and transmits it to the Portrait Sensor Battery for communication to a host device through the Medical Body Area Network (MBAN) connection.
The Portrait Wearable Pulse Oximetry Sensor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Wearable Pulse Oximetry Sensor is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait SpO2 Attachment Accessory Band (Portrait AAB01):
The Portrait SpO2 Attachment Accessory Band (Portrait AAB01) is intended to provide a means to secure the Portrait Wearable Pulse Oximetry Sensor with Portrait Sensor Battery to the patient's wrist.
The Portrait SpO2 Attachment Accessory Band is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Portrait™ Wearable Respiration Rate Sensor (Portrait RR P-RR01):
The Portrait Wearable Respiration Rate Sensor (Portrait P-RR01) is intended for use with adult and pediatric patients (3 years of age and older, and weighing more than 10 kg) for continuous physiologic monitoring of the respiration rate (RR) parameter. The Portrait Wearable Respiration Rate Sensor acquires parameter data from the patient it to the Portrait Sensor Battery for communication to a host devical Body Area Network (MBAN) connection.
The Portrait Wearable Respiration Rate Sensor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Wearable Respiration Rate Sensor is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ RR Electrode Patch (Portrait RRP01):
The Portrait RR Electrode Patch (Portrait RRP01) is intended for use with adult and pediatic patients (3 years of age and older, and weighing more than 10 kg) for continuous physiologic monitoring of the respiration rate (RR) parameter. The Portrait RR Electrode Patch transfers impedance and biopotential signals from the patient and transmits them to the Portrait Wearable Respiration Rate Sensor.
The Portrait RR Electrode Patch is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait RR Electrode Patch is not in a controlled Magnetic Resonance (MR) environment.
Portrait™ Sensor Battery (Portrait SBT01):
The Portrait Sensor Battery (Portrait SBT01) is intended for use as a power supply for the Portrait Wearable Sensors and to provide wireless communication to a host device.
The Portrait Sensor Battery is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Sensor Battery is not intended for use in a controlled Magnetic Resonance (MR) environment.
{6}------------------------------------------------
Portrait™ Bedside Charger (Portrait BCH01):
The Portrait Bedside Charger (Portrait BCH01) is intended for charging the Portrait Sensor Batteries and the Portrait Mobile Patient Monitor (including while the Portrait Mobile Patient Monitor is in use).
The Portrait Bedside Charger is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Bedside Charger is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ Mobile Patient Monitor Pouch (Portrait MMP01):
The Portrait Mobile Patient Monitor Pouch (Portrait MMP01) is an optional accessory intended to enable the Portrait Mobile Patient Monitor to be carried while the patient is ambulatory.
The Portrait Mobile Patient Monitor pouch is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. The letters are stylized and appear in white against a black background, with the circle also outlined in white.
GE Medical Systems Information Technologies, Inc. 9900 Innovation Drive Wauwatosa, WI 53226, USA
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Owner/Contact/Date (807.92(a)(1)):
Date: February 26, 2024
GE Medical Systems Information Technologies, Inc. Owner/Submitter: 9900 Innovation Drive Wauwatosa, WI 53226, USA
Joel Kent Primary Contact Person: Director, Regulatory Affairs, Strategy GE HealthCare, Patient Care Solutions Phone: 617-851-0943 E-mail: joel.kent(@gehealthcare.com
- Secondary Contact Person: William Jung Regulatory Affairs Director GE HealthCare, Monitoring Solutions Phone: 571-396-1558 E-mail: william.jung@gehealthcare.com
{8}------------------------------------------------
| Device names (807.92(a)(2)): | |
|---|---|
| Trade Name: | Portrait Mobile Monitoring Solution V1.1 consists of the following:PortraitTM Central Viewer Application (Portrait CVAXB)PortraitTM Core Services (Portrait CSSXB)PortraitTM Clinical Alarming Unit (Portrait CAU01)PortraitTM Mobile Patient Monitor Hardware (Portrait HUBXB)PortraitTM Mobile Patient Monitor Software (Portrait HSWXB)PortraitTM Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SA01)PortraitTM Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SP01)PortraitTM Wearable Pulse Oximetry Sensor (Portrait SpO2 P-W01, Portrait SpO2 P-SE01)PortraitTM SpO2 Attachment Accessory Band (Portrait AAB01)PortraitTM Wearable Respiration Rate Sensor (Portrait RR P-RR01)PortraitTM RR Electrode Patch (Portrait RRP01)PortraitTM Sensor Battery (Portrait SBT01)PortraitTM Bedside Charger (Portrait BCH01)PortraitTM Mobile Patient Monitor Pouch (Portrait MMP01) |
| Common/Usual Name: | Multiparameter patient monitor (Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)) |
| Classification Names: | 21 CFR 870.2300 Cardiac monitor (including cardiotachometer and rate alarm21 CFR 870.2910 Radiofrequency physiological signal transmitter and receiver21 CFR 868.2375 monitor, breathing frequency21 CFR 870.2700 oximeter |
| Product Code: | MWI |
| Subsequent Product Codes: | MSXDRGBZQDQA |
| Predicate Device(s) (807.92(a)(3)): | The primary predicate for this submission is K230626 Portrait Mobile Monitoring Solution v1.0 |
| Device Description (807.92(a)(4)): | The Portrait Mobile Monitoring Solution V1.1 is a wireless monitoring system for monitoring SpO2, pulse rate and respiration rate of adult and pediatric patients. The system can be used for monitoring adult and pediatric patients (3 years of age and older and weighing more than 10 kg) within a hospital or healthcare facility. The system acquires, stores, calculates, displays, and exports patient physiological parameter data, alarms, and information. It supports pulse oximetry (SpO2/pulse rate) and respiration rate parameters. Measurement values are displayed as graphic or numeric values, like waveforms and |
510(k) Summary GE HealthCare Portrait Mobile Monitoring Solution v1.1 Page 2 of 27
{9}------------------------------------------------
numbers, and when applicable, also as alarm messages.
The Portrait Mobile Monitoring Solution V1.1 consists of the following general categories of medical devices:
Central Monitoring Devices:
· Portrait Core Services hosted on the GE HealthCare nonmedical device Edison Health Link platform. Portrait Core Services is a set of software services that enable the communication and interaction of the system components and are capable of integrating into existing healthcare facility infrastructure and clinical information systems.
• Portrait Central Viewer Application software hosted on a Windows off-the-shelf computer. Portrait Central Viewer Application provides the ability to view patient real-time and historical data, capable of displaying data from multiple patients.
• Portrait Clinical Alarming Unit provides audible alarms at each Central Viewer.
Mobile Monitoring Devices:
· Portrait Mobile Patient Monitor Hardware, a battery-powered, wireless, hand-held patient monitor. The Portrait Mobile Patient Monitor is a completely wireless, hand-held device that is capable of acting as a standalone patient monitor including alarms, with a 3.7-inch capacitive touchscreen capable of displaying numeric data and waveforms for SpO2, Pulse Rate (PR), and Respiration Rate (RR). Compared to the predicate (K230626), the Portrait Mobile Patient Monitor has been split into separate hardware and software components for serviceability. The Portrait Mobile Patient Monitor Hardware is shipped with no clinical software installed. The clinical software (i.e., Portrait Mobile Patient Monitor Software) is installed at the customer site before the device is put into use. The Portrait Mobile Patient Monitor Hardware when the Portrait Mobile Patient Monitor Software is installed (prior to patient use) is then what we will refer to as the Portrait Mobile Patient Monitor. Portrait Wearable SpO2 sensors for acquiring SpO2 and pulse rate data from a patient wirelessly.
· Portrait Mobile Patient Monitor Software is the clinical software which gets installed at the customer site on the Portrait Mobile Patient Monitor Hardware. The Portrait Mobile Patient Monitor Software is this submission is equivalent to the software installed on the predicate Portrait Mobile Patient Monitor cleared in K230626 with minor changes as discussed in this
{10}------------------------------------------------
submission.
· Portrait Wearable Respiration Rate sensor and Portrait RR electrode patch for acquiring impedance respiration data from a patient wirelessly.
· Portrait Sensor battery used for powering the wearable sensors and provide wireless communication to the Portrait Mobile Patient Monitor.
• Portrait Bedside Charger for charging the Portrait Sensor Batteries and Portrait Mobile Patient Monitor (including when the Portrait Mobile Patient Monitor is in clinical use).
· Portrait SpO2 Attachment accessory band which provides means to secure the SpO2 sensors to the patient's wrist
• Portrait Mobile Patient Monitor Pouch, which allows the Portrait Mobile Patient Monitor to be carried while the patient is ambulatory
Intended Use: (807.92(a)(5)):
The Intended Use of the Portrait™ Mobile Monitoring Solution v1.1 as a vital sign monitor is identical to the primary predicate K230626 Portrait Mobile Monitoring Solution v1.0.
The indications for use are described separately for each of the individual subsystems/components of the Portrait Mobile Monitoring Solution V1.1 and for the system as a whole.
Portrait™ Mobile Monitoring Solution
The Portrait Mobile Monitoring Solution is intended to acquire, store, calculate, display and export patient monitoring data as well as provide real time alarming for monitoring adult and pediatric patients (3 years of age and older, and weighing more than 10 kg).
Physiological parameters and waveforms supported are:
- . Pulse oximetry (SpO2/pulse rate)
- Respiration rate (RR)
Continuous pulse oximetry and respiration rate monitoring may be used for patients at risk of cardiorespiratory and infectious complications.
The Portrait Mobile Monitoring Solution is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a
{11}------------------------------------------------
professional healthcare facility.
This device is not an Apnea monitor (i.e., do not rely on the device for detection of alarm for the cessation of breathing).
This device should not be used for life sustaining/supporting purposes.
The Portrait Mobile Monitoring Solution is not intended for use in a controlled Magnetic Resonance (MR) environment.
The Portrait Mobile Monitoring Solution consists of:
Portrait™ Central Viewer Application (Portrait CVAXB)
The Portrait Central Viewer Application (Portrait CVAXB) provides monitoring station capability running as an application for the Portrait Mobile Monitoring Solution on a PC platform that meets minimum system requirements. It provides the ability to view real-time data for multiple patients and historical data for a single patient including configurable visual and audible alarm notifications.
The Portrait Central Viewer Application is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Portrait™ Core Services (Portrait CSSXB)
The Portrait Core Services (Portrait CSSXB) are a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems. The Portrait Core Services can transmit patient physiological trends and numerics (IHE PCD DEC), waveforms (IHE PCD WCM), and alarm events (IHE PCD ACM) outbound. The Portrait Core Services can also receive HL7 ADT information to admit patients to the Portrait Mobile Monitoring Solution.
Portrait™ Clinical Alarming Unit (Portrait CAU01)
The Portrait Clinical Alarming Unit (Portrait CAU01) is a required accessory to the Portrait Central Viewer Application that provides audio alarming capability.
The Portrait Clinical Alarming Unit is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
{12}------------------------------------------------
Portrait™ Mobile Patient Monitor Hardware (Portrait HUBXB)
The Portrait Mobile Patient Monitor Hardware (Portrait HUBXB) is intended for use with adult and pediatric patients as a handheld, mobile device which supports running the Portrait Mobile Patient Monitor Software to monitor patients, and to provide audible and visual alarms. The Portrait Mobile Patient Monitor Hardware enables non-invasive continuous vital sign monitoring of patients by acquiring signals wirelessly from Portrait wearable sensors, as well as wireless clinical and service data communication.
The Portrait Mobile Patient Monitor Hardware is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Mobile Patient Monitor Hardware is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ Mobile Patient Monitor Software (Portrait HSWXB)
The Portrait Mobile Patient Monitor Software (Portrait HSWXB) is intended to run on Portrait Mobile Patient Monitor Hardware for the continuous monitoring of oxygen saturation (SpO2), pulse rate (PR) and respiration rate (RR) parameters. Portrait Mobile Patient Monitor Software is intended for use with adult and pediatric patients (3 years of age and older, and weighing more than 10 kg). It enables non-invasive continuous monitoring of patients by acquiring signals from Portrait wearable sensors, as well as displaying trends, events, and a QR code containing patient demographics and parameter data. When installed on the Portrait Mobile Patient Monitor Hardware, the Portrait Mobile Patient Monitor Software provides real-time, trend and event data to Portrait Core Services. The Portrait Mobile Patient Monitor Software, when configured to do so, also enables the Portrait Mobile Patient Monitor Hardware to provide audible and visual alarms locally.
The Portrait Mobile Patient Monitor Software is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
{13}------------------------------------------------
Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SA01)
The Portrait Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SA01) is intended for use with adult and pediatric patients (3 years of age and older, and weighing more than 30 kg) for continuous physiologic monitoring of oxygen saturation (SpO2) and pulse rate (PR) parameters. The Portrait Wearable Pulse Oximetry Sensor acquires parameter data from the patient and transmits it to the Portrait Sensor Battery for communication to a host device through the Medical Body Area Network (MBAN) connection.
The Portrait Wearable Pulse Oximetry Sensor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Wearable Pulse Oximetry Sensor is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SP01)
The Portrait Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SP01) is intended for use with pediatric patients (3 years of age and older, and weighing 15 kg to 30 kg) for continuous physiologic monitoring of oxygen saturation (SpO2) and pulse rate (PR) parameters. The Portrait Wearable Pulse Oximetry Sensor acquires parameter data from the patient and transmits it to the Portrait Sensor Battery for communication to a host device through the Medical Body Area Network (MBAN) connection.
The Portrait Wearable Pulse Oximetry Sensor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Wearable Pulse Oximetry Sensor is not intended for use in a controlled Magnetic Resonance (MR) environment.
{14}------------------------------------------------
Portrait™ Wearable Pulse Oximetry Sensor (Portrait SpO2 P-W01, Portrait SpO2 P-SE01)
The Portrait Wearable Pulse Oximetry Sensor (Portrait SpO2 P-SE01, Portrait SpO2 P-W01) is intended for use with adult and pediatric patients (3 years of age and older, and weighing more than 10 kg) for continuous physiologic monitoring of oxygen saturation (SpO2) and pulse rate (PR) parameters. The Portrait Wearable Pulse Oximetry Sensor acquires parameter data from the patient and transmits it to the Portrait Sensor Battery for communication to a host device through the Medical Body Area Network (MBAN) connection.
The Portrait Wearable Pulse Oximetry Sensor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Wearable Pulse Oximetry Sensor is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ SpO2 Attachment Accessory Band (Portrait AAB01)
The Portrait SpO2 Attachment Accessory Band (Portrait AAB01) is intended to provide a means to secure the Portrait Wearable Pulse Oximetry Sensor with Portrait Sensor Battery to the patient's wrist.
The Portrait SpO2 Attachment Accessory Band is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Portrait™ Wearable Respiration Rate Sensor (Portrait RR P-RR01)
The Portrait Wearable Respiration Rate Sensor (Portrait P-RR01) is intended for use with adult and pediatric patients (3 years of age and older, and weighing more than 10 kg) for continuous physiologic monitoring of the respiration rate (RR) parameter. The Portrait Wearable Respiration Rate Sensor acquires parameter data from the patient and transmits it to the Portrait Sensor Battery for communication to a host device through the Medical Body Area Network (MBAN) connection.
The Portrait Wearable Respiration Rate Sensor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
{15}------------------------------------------------
The Portrait Wearable Respiration Rate Sensor is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ RR Electrode Patch (Portrait RRP01)
The Portrait RR Electrode Patch (Portrait RRP01) is intended for use with adult and pediatric patients (3 years of age and older, and weighing more than 10 kg) for continuous physiologic monitoring of the respiration rate (RR) parameter. The Portrait RR Electrode Patch transfers impedance and biopotential signals from the patient and transmits them to the Portrait Wearable Respiration Rate Sensor.
The Portrait RR Electrode Patch is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait RR Electrode Patch is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ Sensor Battery (Portrait SBT01)
The Portrait Sensor Batterv (Portrait SBT01) is intended for use as a power supply for the Portrait Wearable Sensors and to provide wireless communication to a host device.
The Portrait Sensor Battery is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Sensor Battery is not intended for use in a controlled Magnetic Resonance (MR) environment.
Portrait™ Bedside Charger (Portrait BCH01)
The Portrait Bedside Charger (Portrait BCH01) is intended for charging the Portrait Sensor Batteries and the Portrait Mobile Patient Monitor (including while the Portrait Mobile Patient Monitor is in use).
The Portrait Bedside Charger is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
The Portrait Bedside Charger is not intended for use in a controlled Magnetic Resonance (MR) environment.
{16}------------------------------------------------
Portrait™ Mobile Patient Monitor Pouch (Portrait MMP01)
The Portrait Mobile Patient Monitor Pouch (Portrait MMP01) is
an optional accessory intended to enable the Portrait Mobile
Patient Monitor to be carried while the patient is ambulatory.
The Portrait Mobile Patient Monitor pouch is intended for use
under the direct supervision of a licensed healthcare practitioner,
or by personnel trained in proper use of the equipment in a
professional healthcare facility.
Technology (807.92(a)(6)): The Portrait Mobile Monitoring Solution uses the same
fundamental technology and functionality as the primary
predicate primary predicate K230626 Portrait Mobile
Monitoring Solution v1.0.
The basis for this TRADITIONAL 510(k) submission is that the
Portrait Mobile Monitoring Solution with software v1.1 is a
revised version of the predicate Portrait™ Mobile Monitoring
Solution K230626 with changes as described below.
The Intended Use of the Portrait™ Mobile Monitoring Solution
with software v1.1 as a vital sign monitor is identical to the
predicate primary predicate Portrait™ Mobile Monitoring
Solution K230626 with software version 1.0.
The Portrait Mobile Monitoring Solution with software v1.1 uses
the same fundamental technology and functionality as the
primary predicate primary predicate Portrait™ Mobile
Monitoring Solution K230626.
Changes since the previous submission (K230626)
The main change is a new software version 1.1 replacing
software 1.0 and revised labeling. There are no changes to the
hardware of the devices that comprise Portrait Mobile
Monitoring Solution.
The main differences with respect to the predicate discussed
throughout this submission are summarized below:
Service enhancements to support ease of deployment
- Installation enhancements for Portrait Core Services to
make for a more streamlined software installation
process. - Ability to do over-the-air updates to sensor software from
the Portrait Mobile Patient Monitor and to update
510(k) Summary GE HealthCare Portrait Mobile Monitoring Solution v1.1 Page 10 of 27
{17}------------------------------------------------
batteries from the Portrait Mobile Patient Monitor when both are docked on the charger
- Service enhancements to allow the ability for a reset of ● all configured parameters to factory defaults when performing a software update of the Portrait Mobile Patient Monitor.
- Electronic delivery to a field engineer laptop of software for the Portrait Central Viewer and Portrait Mobile Patient Monitor, as well as electronic delivery of software updates for sensors, sensor battery, and charger to a field engineer laptop. This feature eliminates the need to ship physical software kits (i.e., USB sticks) to customer sites.
Cybersecurity improvements
- . Updates to off-the-shelf software to address open cybersecurity vulnerabilities and software end-of-life issues.
- Software signing key updates.
- Improvements to device authentication of a Sensor Battery and Sensor (from the Portrait Mobile Patient Monitor).
- Adding support for WPA3 wireless security to the ● Portrait Mobile Patient Monitor.
Minor Software functional enhancements
- . Alarm improvements to support a 2 stage check probe alarm and adding a 60s delay option for Resp High, Resp Low, Pulse Rate High and Pulse Rate Low.
- OR code display in the Portrait Mobile Patient Monitor ● to allow patient vital signs to be read by an external vital signs monitor.
- Adding IHE WCM outbound feeds (i.e., HL7) for ● Pleth/Resp waveforms.
Minor Changes to the Indications for Use
We have revised the Indications for Use statement from the predicate for clarity and to reflect the following changes that are new for this submission:
- . Separating the Portrait Mobile Patient Monitor hardware and software into 2 separate medical devices with corresponding indication statements (Portrait HUBXB and Portrait HSWXB).
- Updates to device model numbers for the Portrait Central . Viewer Application and Portrait Core Services.
{18}------------------------------------------------
- . Addition of IHE WCM outbound feeds (i.e., HL7) to the indications for Portrait Core Services (licensed feature).
- . Addition of a QR code containing patient demographics and parameter data (used to allow patient vital signs to be read by external vital signs monitor) to the indications for the Portrait Mobile Patient Monitor Software.
The scope of modifications does not affect the patient population or environment of use or parameters monitored compared to the predicate Portrait Mobile Monitoring Solution (K230626).
Miscellaneous updates
- . Bluetooth beaconing to enable RTLS (real-time location system) support. Used for device tracking only. This added feature is not used for clinical use.
- Localization support of Portrait Mobile Monitoring Solution user interface and manuals to support new languages required for new target countries.
- Adding a compatibility check to Sensor, Sensor Battery, ● and Portrait Mobile Patient Monitor Software to ensure that only verified software configurations can be used for patient monitoring.
- . Updates to shared communication libraries to support functional enhancements referenced above.
- Separating the Portrait Mobile Patient Monitor hardware . and software into 2 separate medical devices, which resulted in changes to device labels.
- Miscellaneous updates to packaging labels including the ● following:
- Updates to the minimum hardware specifications ● required for the Portrait Central Viewer Application software (to support expanded functionality in the future).
- Updates to the humidity specifications for the Portrait Mobile Patient Monitor to 10-90% from 5-95%.
- Deliver new IOU and software upgrade parts (i.e., new ● part numbers for service parts).
- . Updates to user and service documentation needed to explain the changes, as well as continuous improvement clarifications
A summary of the main changes compared to the predicate are listed below in the comparison table.
Product Comparison versus Predicate Main features chart follows below.
{19}------------------------------------------------
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution -K230626 | Proposed devicePortrait Mobile Monitoring Solution - | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
|---|---|---|---|
| Intended Use | Vital Sign Monitor | Vital Sign Monitor | Identical |
| FDA Primary Product Code | MWI | MWI | Identical |
| FDA Classification Regulation | 21 CFR 870.2300 | 21 CFR 870.2300 | Identical |
| Indications for Use (entire system) | The Portrait Mobile MonitoringSolution is intended to acquire, store,calculate, display and export patientmonitoring data as well as provide realtime alarming for monitoring adultand pediatric patients (3 years of ageand older, and weighing more than 10kg).Physiological parameters andwaveforms supported are:• Pulse oximetry (SpO2/pulserate)• Respiration rate (RR)Continuous pulse oximetry andrespiration rate monitoring may beused for patients at risk ofcardiorespiratory and infectiouscomplications.The Portrait Mobile MonitoringSolution is intended for use under thedirect supervision of a licensedpractitioner, or by personnel trained inproper use of the equipment in aprofessional healthcare facility.This device is not an Apnea monitor(i.e., do not rely on the device fordetection of alarm for the cessation ofbreathing).This device should not be used for lifesustaining/supporting purposes.The PORTRAIT Mobile MonitoringSolution is not intended for use in a | The Portrait Mobile MonitoringSolution is intended to acquire, store,calculate, display and export patientmonitoring data as well as provide realtime alarming for monitoring adult andpediatric patients (3 years of age andolder, and weighing more than 10 kg).Physiological parameters andwaveforms supported are:• Pulse oximetry (SpO2/pulserate)• Respiration rate (RR)Continuous pulse oximetry andrespiration rate monitoring may beused for patients at risk ofcardiorespiratory and infectiouscomplications.The Portrait Mobile MonitoringSolution is intended for use under thedirect supervision of a licensedhealthcare practitioner, or bypersonnel trained in proper use of theequipment in a professional healthcarefacility.This device is not an Apnea monitor(i.e., do not rely on the device fordetection of alarm for the cessation ofbreathing).This device should not be used for lifesustaining/supporting purposes.The PORTRAIT Mobile MonitoringSolution is not intended for use in a | EquivalentA typo was corrected in the indicationsstatement to add the word"healthcare" so that "…licensedprofessional, " changes to "…licensedhealthcare professional, "This difference does not significantlyaffect safety and/or effectiveness. |
| controlled Magnetic Resonance (MR)environment. | controlled Magnetic Resonance (MR)environment. | ||
| Patient Population | Adult and pediatric patients (3 years ofage and older, and weighing morethan 10 kg). | Adult and pediatric patients (3 years ofage and older, and weighing morethan 10 kg). | Identical |
| Environments of Use | A professional healthcare facility | A professional healthcare facility | Identical |
| Location of devices in the patientvicinity | In the Portrait Mobile MonitoringSolution, only the sensors must belocated on the patient. The monitor isconnected wirelessly to the sensors,and as such, must only be located inthe vicinity of the patient. However,the monitor may be worn by thepatient using Portrait Mobile PatientMonitor Pouch. The monitor is alsocapable of being carried in one hand. | In the Portrait Mobile MonitoringSolution, only the sensors must belocated on the patient. The monitor isconnected wirelessly to the sensors,and as such, must only be located inthe vicinity of the patient. However,the monitor may be worn by thepatient using Portrait Mobile PatientMonitor Pouch. The monitor is alsocapable of being carried in one hand. | Identical |
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution –K230626 | Proposed devicePortrait Mobile Monitoring Solution – | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
| Remote Viewing and MonitoringModes | Once a patient is admitted, thePortrait Mobile Patient Monitorcomponent of the Portrait MobileMonitoring Solution can be usedstandalone (viewing of data and alarmannunciation) if properly configured orif it becomes disconnected from therest of the system. In addition, patientdata and alarms are always remotelydisplayed on the Portrait CentralViewer unless there is a system ornetwork failure. | Once a patient is admitted, thePortrait Mobile Patient Monitorcomponent of the Portrait MobileMonitoring Solution can be usedstandalone (viewing of data and alarmannunciation) if properly configured orif it becomes disconnected from therest of the system. In addition, patientdata and alarms are always remotelydisplayed on the Portrait CentralViewer unless there is a system ornetwork failure. | Identical |
| STANDARDS COMPLIANCE | |||
| Medical Standards | IEC 60601-1:2005 + A1:2012IEC 60601-1-2:2014IEC 60601-1-8:2012 / EN 60601-1-8:2007/A1:2013IEC 80601-2-49:2018ISO 80601-2-61:2017/EN ISO 80601-2-61:2017IEC 60601-1-6:2013/EN 60601-1-6:2010/A1:2015IEC 62366-1:2015 / EN 62366-1:2015ISO 10993-1:2018IEC 62304:2015 / EN 62304:2006 /A1:2016ISO 14971 Third Edition 2019-12LabelingISO 20417:2021ISO 15223-1:2021ISO 17664:2017ISO 17664-2:2021Battery:UL 2054:2004IEC 62133-2:2017UL 1642:2012RFID:AIM 7351731:2017Wireless:ANSI IEEE C63.27:2017 | ANSI AAMI ES60601-1:2005/®2012 &A1:2012, C1:2009/®2012 &A2:2010/®2012 (Cons. Text) [Incl.AMD2:2021]60601-1-2 Edition 4.1 2020-09IEC 60601-1-8 Edition 2.2 2020-07IEC 80601-2-49 Edition 1.0 2018-03ISO 80601-2-61 Second edition 2017-12IEC 60601-1-6 Edition 3.2 2020-07IEC 62366-1 Edition 1.1 2020-06ISO 10993-1 Fifth editionIEC 62304 Edition 1.1 2015-06ISO 14971 Third Edition 2019-12LabelingISO 20417 First edition 2021-04Corrected version 2021-1215223-1 Fourth edition 2021-0717664-2 First edition 2021-02Battery:UL 2054 2nd EditionIEC 62133-2:2017/AMD1:2021UL 1642:2012Wireless:IEEE ANSI C63.27-2021 | EquivalentNewer updated versions of standardswere used for the following standards:ANSI AAMI ES60601-1:2005/®2012 & A1:2012,C1:2009/®2012 &A2:2010/®2012 (Cons.Text) [Incl. AMD2:2021]60601-1-2 Edition 4.12020-0960601-1-8 Edition 2.22020-0760601-1-6 Edition 3.22020-07IEC 62366-1 Edition 1.12020-06IEEE ANSI C63.27-2021The AIM 7351731 standard wasremoved since the new IEC 60601-1-2Edition 4.1 now takes the same testinginto account. Similarly, the ISO17664:2017 standard was droppedsince relevant content for Portrait V1.1devices are covered by the newer ISO17664-2: 2021.Updating to newer standard versionsdoes not significantly affect safetyand/or effectiveness. |
| System Components | |||
| Patient Device | PORTRAIT Mobile Patient Monitor | Portrait HUBXB with software PortraitHSWXB | EquivalentThe Portrait Mobile Patient Monitorhad been split into separate hardwareand software components. ThePortrait Mobile Patient Monitor isshipped with no clinical softwareinstalled. The clinical software (i.e.,Portrait Mobile Patient MonitorSoftware) is installed at the customersite before the device is put into use.This difference does not significantlyaffect safety and/or effectiveness. |
| Sensors | PORTRAIT SpO2 Wearable PulseOximetry SensorsPORTRAIT Wearable Respiration RateSensor(wireless connection to the Mobile | PORTRAIT SpO2 Wearable PulseOximetry SensorsPORTRAIT Wearable Respiration RateSensor(wireless connection to the Mobile | Identical |
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution –K230626 | Proposed devicePortrait Mobile Monitoring Solution – | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
| Sensor Battery | Detachable sensor battery.Interchangeable. May be used withany sensor. | Detachable sensor battery.Interchangeable. May be used withany sensor. | Identical |
| SpO2 Probes | Integrated in sensors | Integrated in sensors | Identical |
| Respiration Electrode Patches | Portrait RR Electrode Patch | Portrait RR Electrode Patch | Identical |
| Battery Charger | Portrait Bedside Charger | Portrait Bedside Charger | Identical |
| Accessories for Portrait MobilePatient Monitor | Portrait Mobile Patient Monitor Pouch | Portrait Mobile Patient Monitor Pouch | Identical |
| Central Viewing | Portrait Central Viewer SoftwareApplication withPortrait Clinical Alarming Unit | Portrait Central Viewer SoftwareApplication withPortrait Clinical Alarming Unit | Identical |
| Storage and Centralized Services | Portrait Core Services | Portrait Core Services | Identical |
| Accessories for SpO2 | Portrait SpO2 Attachment AccessoryBand | Portrait SpO2 Attachment AccessoryBand | Identical |
| Patient Device – General HardwareSpecifications | |||
| Device Name | Portrait Mobile Patient Monitor | Portrait HUBXB with software PortraitHSWXB | EquivalentThe Portrait Mobile Patient Monitorhad been split into separate hardwareand software components. ThePortrait Mobile Patient Monitor isshipped with no clinical softwareinstalled. The clinical software (i.e.,Portrait Mobile Patient MonitorSoftware) is installed at the customersite before the device is put into use.This difference does not significantlyaffect safety and/or effectiveness. |
| Size (H x W x D) | 14.1 cm x 6.3 cm x 2.1 cm | 14.1 cm x 6.3 cm x 2.1 cm | Identical |
| Weight | 223 g | 223 g | Identical |
| Power source | Battery (integrated into device) | Battery (integrated into device) | Identical |
| Battery Type | Lithium-ion | Lithium-ion | Identical |
| Run-time on battery | 16 Hours with full capacity (display off) | 16 Hours with full capacity (display off) | Identical |
| Charge time | 8 Hours from empty to full capacity. | 8 Hours from empty to full capacity. | Identical |
| Patient Device – User Interface | |||
| Device Name | Portrait Mobile Patient Monitor | ||
| Touch display interface | Yes | Yes | Identical |
| Visual and Audible Alarm capability | Available | Available | Identical |
| Patient Device – DisplaySpecifications | |||
| Display type | SFT LCD (Super Fine TFT) | SFT LCD (Super Fine TFT) | Identical |
| Display Size and resolution | 480 x 800 pixels, 3.7 inches in 16:9format | 480 x 800 pixels, 3.7 inches in 16:9format | Identical |
| Parameter Blocks DisplayedSimultaneously | 3 Parameters (all availableparameters) | 3 Parameters (all availableparameters) | Identical |
| Number of traces (waveforms) | One waveform per page in detailedparameter view | One waveform per page in detailedparameter view | Identical |
| Trends | Displays graphical trend of connectedparameters for the last 4 hours.Movable cursor displays details for thegiven point in time. | Displays graphical trend of connectedparameters for the last 4 hours.Movable cursor displays details for thegiven point in time. | Identical |
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution -K230626 | Proposed devicePortrait Mobile Monitoring Solution - | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
| Patient Device - Connectivity | |||
| Wireless connection to network | Requires a wireless networkconnection to begin monitoring. IEEE802.11 a/b/g/n supported (dual band).After patient admission, the PortraitMobile Patient Monitor will continueto monitor locally in the event of anetwork failure. | Requires a wireless networkconnection to begin monitoring. IEEE802.11 a/b/g/n supported (dual band).After patient admission, the PortraitMobile Patient Monitor will continueto monitor locally in the event of anetwork failure. | Identical |
| Connection to Sensors | Communicates with the sensors over awireless Medical Body Area Network(MBAN) using a proprietary protocol.The MBAN connection to the sensorsincludes a GE-proprietary protocol anda low-power radio interface that canoperate on both the unlicensed 2.4GHz ISM band and certain otherprotected frequency bands whereMBAN traffic is allowed. In the USA,the MBAN communication occurs inthe 2390-2400 MHz band, which hasbeen reserved by the FCC for use byMedical Body Area Networks. | Communicates with the sensors over awireless Medical Body Area Network(MBAN) using a proprietary protocol.The MBAN connection to the sensorsincludes a GE-proprietary protocol anda low-power radio interface that canoperate on both the unlicensed 2.4GHz ISM band and certain otherprotected frequency bands whereMBAN traffic is allowed. In the USA,the MBAN communication occurs inthe 2390-2400 MHz band, which hasbeen reserved by the FCC for use byMedical Body Area Networks. | Identical |
| RTLS | None | Portrait HUBXB with software PortraitHSWXB transmits Bluetooth beaconsto support RTLS (real time locationservices). | EquivalentBluetooth beacons sent in theproposed device are intended forasset tracking only and are notintended for clinical use.This difference does not significantlyaffect safety and/or effectiveness. |
| QR Code | None | Portrait HUBXB with software PortraitHSWXB supports QR code for readingof patient demographics and vital signsby external device. | EquivalentIn V1.1, Portrait HUBXB with softwarePortrait HSWXB adds a QR code thatcan be read by an external device totransfer patient demographics andcurrent vital signs information.This difference does not significantlyaffect safety and/or effectiveness. |
| Accessories for Portrait Mobile Patient | |||
| Device Name | MonitorPortrait Mobile Patient Monitor Pouch | ||
| Single Use / Multiple Use | Single Use | Single Use | Identical |
| Respiration Sensors - Hardware | |||
| Device Name | Portrait Wearable Respiration RateSensor | ||
| Dimensions | 6.9 x 5.3 x 1.9 cm | 6.9 x 5.3 x 1.9 cm | Identical |
| Weight | 34 g | 34 g | Identical |
| Accessories for Respiration Rate | |||
| Device Name | sensorPortrait RR Electrode Patch | ||
| Operating Temperature | 0 to 40 °C | 0 to 40 °C | Identical |
| Single Use / Multiple Use | Single Use | Single Use | Identical |
| Shelf Life | 18 months | 18 months | Identical |
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution -K230626 | Proposed devicePortrait Mobile Monitoring Solution - | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
| Pulse Oximetry Sensor - Hardware | |||
| Device Name | Portrait Wearable Pulse OximetrySensor | ||
| Sensor Application Site | Finger | Finger | Identical |
| Dimensions | SpO2 P-SA01: 27.2 x 5.3 x 1.9 cmSpO2 P-SP01: 23.6 x 5.3 x 1.9 cmSpO2 P-W01: 26.2 x 5.3 x 1.9 cmSpO2 P-SE01: 27.1 x 5.3 x 1.9 cm | SpO2 P-SA01: 27.2 x 5.3 x 1.9 cmSpO2 P-SP01: 23.6 x 5.3 x 1.9 cmSpO2 P-W01: 26.2 x 5.3 x 1.9 cmSpO2 P-SE01: 27.1 x 5.3 x 1.9 cm | Identical |
| Weight | SpO2 P-SA01: 52 gSpO2 P-SP01: 43 gSpO2 P-W01: 38 gSpO2 P-SE01: 38 g | SpO2 P-SA01: 52 gSpO2 P-SP01: 43 gSpO2 P-W01: 38 gSpO2 P-SE01: 38 g | Identical |
| Fluid Ingress Rating | SpO2 P-SA01, SpO2 P-SP01: IP67SpO2 P-W01 and SpO2 P-SE01: IP44 | SpO2 P-SA01, SpO2 P-SP01: IP67SpO2 P-W01 and SpO2 P-SE01: IP44 | Identical |
| Operating Temperature | 0 to 40 °C | 0 to 40 °C | Identical |
| Non-Operating Temperature | -30 to 70 °C | -30 to 70 °C | Identical |
| Operating Humidity | 5 to 95% non-condensing | 5 to 95% non-condensing | Identical |
| Non-Operating Humidity | 5 to 95% non-condensing | 5 to 95% non-condensing | Identical |
| Operating Pressure | 620 to 1060 hPa | 620 to 1060 hPa | Identical |
| Note: For comparison purposes this is62 kPa to 106 kPa | Note: For comparison purposes this is62 kPa to 106 kPa | ||
| Non-Operating Pressure | 500 to 1060 hPa | 500 to 1060 hPa | Identical |
| Note: For comparison purposes this is50 kPa to 106 kPa | Note: For comparison purposes this is50 kPa to 106 kPa | ||
| Accessories for SpO2 sensors | |||
| Device Name | Portrait SpO2 Attachment AccessoryBand | ||
| Indications for Use | The Portrait SpO2 AttachmentAccessory Band (Portrait AAB01) isintended to provide a means to securethe Portrait Wearable Pulse OximetrySensor with Portrait Sensor Battery tothe patient's wrist.The Portrait SpO2 AttachmentAccessory Band is intended for useunder the direct supervision of alicensed practitioner, or by personneltrained in proper use of theequipment in a professionalhealthcare facility | The Portrait SpO2 AttachmentAccessory Band (Portrait AAB01) isintended to provide a means to securethe Portrait Wearable Pulse OximetrySensor with Portrait Sensor Battery tothe patient's wrist.The Portrait SpO2 AttachmentAccessory Band is intended for useunder the direct supervision of alicensed healthcare practitioner, or bypersonnel trained in proper use of theequipment in a professional healthcarefacility. | EquivalentA typo was corrected in the indicationsstatement to add the word"healthcare" so that "…licensedprofessional, …" changes to "…licensedhealthcare professional, …".This difference does not significantlyaffect safety and/or effectiveness. |
| Operating Temperature | 0 to 40 °C | 0 to 40 °C | Identical |
| Non-Operating Temperature | -30 °C to -70 °C | -30 °C to -70 °C | Identical |
| Non-Operating Humidity | 5 to 95% non-condensing | 5 to 95% non-condensing | Identical |
| Non-Operating pressure | 500 to 1060 hPaNote: For comparison purposes this is50 kPa to 106 kPa | 500 to 1060 hPaNote: For comparison purposes this is50 kPa to 106 kPa | Identical |
| Sterility | The SpO2 Attachment is supplied asnon-sterile and is not intended to besterilized. | The SpO2 Attachment is supplied asnon-sterile and is not intended to besterilized. | Identical |
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution –K230626 | Proposed devicePortrait Mobile Monitoring Solution – | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
| Sensor Batteries - Hardware | |||
| Device Name | Portrait Sensor Battery | ||
| Dimensions | 3.6 x 5.3 x 1.7 cm | 3.6 x 5.3 x 1.7 cm | Identical |
| Weight | 31 g | 31 g | Identical |
| Charger - Hardware | |||
| Device Name | Portrait Bedside Charger | ||
| Input voltage and frequency | 100-240 V, 50-60 Hz (using externalpower supply) | 100-240 V, 50-60 Hz (using externalpower supply) | Identical |
| Power | 36 W (Output power - 3.0 A at 12 Vdc) | 36 W (Output power - 3.0 A at 12 Vdc) | Identical |
| Dimensions (exclusive of powersupply and cable) | 11.5 x 23.8 x 4.3 cm | 11.5 x 23.8 x 4.3 cm | Identical |
| Weight(exclusive of power supply andcable) | 412 gWall mount: 603 gTable mount: 908 g | 412 gWall mount: 603 gTable mount: 908 g | Identical |
| Central/Remote Viewer | |||
| Device Name | Portrait Central Viewer Applicationwith Portrait Clinical Alarming Unit | ||
| Display | Required Specification for customersupplied display:20 in (minimum),1920 x 1080 resolution (minimum) | Required Specification for customersupplied display:20 in (minimum),1920 x 1080 resolution (minimum) | Identical |
| Operating System | Required Specification for customersupplied PC:Windows 10, build 10.0.17763(minimum) | Required Specification for customersupplied PC:Windows 10, build 10.0.17763(minimum) | Identical |
| Number of patients per viewer | Maximum 24 | Maximum 24 | Identical |
| Simultaneous view of numerics forall patients on viewer | Supported on for all measuredparameters on Multiple Patient View –SpO2%, Pulse Rate, Respiration Rate | Supported on for all measuredparameters on Multiple Patient View –SpO2%, Pulse Rate, Respiration Rate | Identical |
| Viewing of patient waveforms | One patient at time - Supported onsingle patient view | One patient at time - Supported onsingle patient view | Identical |
| Viewing of patient trends | Graphical trend view supported forone patient at time – single patientview | Graphical trend view supported forone patient at time – single patientview | Identical |
| Visual/Audible alarms supported | Supported | Supported | Identical |
| Storage and Centralized Services –General | |||
| Device Name | Portrait Core Services | ||
| Description | The Portrait Core Services are a set ofsoftware services that enable thecommunication and interaction of thePortrait Monitoring Solutioncomponents and will integrate intoexisting healthcare facilityinfrastructure and clinical informationsystems. The Portrait Core Servicesprovide system configuration,administration, data storage, andtransmission of patient physiologicaltrends and events. The Portrait CoreServices are installed on the Portraitnon-medical device EHL (Edison HealthLink) Server. | The Portrait Core Services are a set ofsoftware services that enable thecommunication and interaction of thePortrait Monitoring Solutioncomponents and will integrate intoexisting healthcare facilityinfrastructure and clinical informationsystems. The Portrait Core Servicesprovide system configuration,administration, data storage, andtransmission of patient physiologicaltrends and events. The Portrait CoreServices are installed on the Portraitnon-medical device EHL (Edison HealthLink) Server. | Identical |
| HL7 outbound support | Supported - The Portrait Core Servicescan transmit patient physiologicaltrends and numerics (IHE PCD DEC)and alarm events (IHE PCD ACM)outbound. | Supported - The Portrait Core Servicescan transmit patient physiologicaltrends and numerics (IHE PCD DEC)and alarm events (IHE PCD ACM)outbound and waveforms (IHE PCDWCM). | EquivalentThe addition of transmitting IHE PCDWCM waveforms in addition tophysiological trends and numerics inthe predicate does not significantlyaffect safety and/or effectiveness. |
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution -K230626 | Proposed devicePortrait Mobile Monitoring Solution - | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
| HL7 Inbound support | Supported - The Portrait Core Servicescan also receive HL7 ADT informationto admit patients to the PortraitMonitoring Solution. | Supported - The Portrait Core Servicescan also receive HL7 ADT informationto admit patients to the PortraitMonitoring Solution. | Identical |
| Storage and Centralized Services -General | |||
| Server Model | HP ProLiant DL360 Gen10 | HP ProLiant DL360 Gen10 | Identical |
| Monitored Parameters | |||
| Monitored Parameters | Oxygen saturation (SpO2)Pulse rate (PR)Respiration Rate (RR) | Oxygen saturation (SpO2)Pulse rate (PR)Respiration Rate (RR) | Identical |
| Parameters Acquisition Method | The Portrait Mobile Monitor acquiresparameter data from wireless patientsensors over the MBAN wireless link.The parameter electronics areencapsulated into the respectivewireless sensors.The sensors include:Portrait SpO2 P-SA01, SpO2 WearablePulse Oximetry Sensor (SpO2, PulseRate)Portrait SpO2 P-SP01, SpO2 WearablePulse Oximetry Sensor (SpO2, PulseRate)Portrait SpO2 P-W01, SpO2 WearablePulse Oximetry Sensor (SpO2, PulseRate)Portrait SpO2 P-SE01, SpO2 WearablePulse Oximetry Sensor (SpO2, PulseRate)Portrait RR P-RR01, WearableRespiration Rate Sensor (RespirationRate) | The Portrait HUBXB with softwarePortrait HSWXB acquires parameterdata from wireless patient sensorsover the MBAN wireless link. Theparameter electronics areencapsulated into the respectivewireless sensors.The sensors include:Portrait SpO2 P-SA01, SpO2 WearablePulse Oximetry Sensor (SpO2, PulseRate)Portrait SpO2 P-SP01, SpO2 WearablePulse Oximetry Sensor (SpO2, PulseRate)Portrait SpO2 P-W01, SpO2 WearablePulse Oximetry Sensor (SpO2, PulseRate)Portrait SpO2 P-SE01, SpO2 WearablePulse Oximetry Sensor (SpO2, PulseRate)Portrait RR P-RR01, WearableRespiration Rate Sensor (RespirationRate) | Identical (except for name change ofthe device) |
| Alarms | |||
| Classification / Alarm Levels | Three levels - High (red), Medium(yellow), Low (cyan) in compliancewith IEC 60601-1-8Informational messages (gray) alsoincluded | Three levels - High (red), Medium(yellow), Low (cyan) in compliancewith IEC 60601-1-8Informational messages (gray) alsoincluded | Identical |
| Notification | Audible and visual | Audible and visual | Identical |
| List of Physiological Alarmssupported (for parameters includedin the proposed device) | PR HighPR LowRR HighRR LowSpO2 HighSpO2 LowSpO2 Critically lowApnea | PR HighPR LowRR HighRR LowSpO2 HighSpO2 LowSpO2 Critically lowApnea | Identical |
| Technical / Equipment Alarms | Supported | Supported | Equivalent |
| The proposed device supports thesame Technical/Equipment alarmswith a modified SpO2 check probetechnical alarm from a single levelalarm to a 2-stage (i.e., escalating)alarm. | This change allows for the responsibleorganization flexibility in configuringalarm priority defaults. Theconfigurations options in the predicatedevice remain available in theproposed device. The proposed deviceprovides the responsible organizationto escalate the alarm priority after aconfigurable delay. The configurationof alarm priorities and time delay arepassword protected. | ||
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution –K230626 | Proposed devicePortrait Mobile Monitoring Solution – | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
| This difference does not significantlyaffect safety and/or effectiveness. | |||
| Trending | |||
| Trend Visualization | Graphical Trends only. Individualvalues for each trended value arevisible at the position of the cursor. | Graphical Trends only. Individualvalues for each trended value arevisible at the position of the cursor. | Identical |
| Trend Visualization Duration | Up to 4 Hours on the Portrait MobilePatient MonitorUp to 24 Hours on the Portrait CentralViewer Application | Up to 4 Hours on the Portrait MobilePatient MonitorUp to 24 Hours on the Portrait CentralViewer Application | Identical |
| Trend Smoothing | On the Portrait Central ViewerApplication, the trend data undergoesan additional smoothing process. Thepurpose of this smoothing is toremove brief changes and normalvariations in the physiological trenddata acquired by the Portrait MobilePatient Monitor. | On the Portrait Central ViewerApplication, the trend data undergoesan additional smoothing process. Thepurpose of this smoothing is toremove brief changes and normalvariations in the physiological trenddata acquired by the Portrait MobilePatient Monitor. | Identical |
| Event management | Physiological Events (i.e., alarms) ofmedium or high priority are stored inthe Portrait Mobile MonitoringSolution and are viewable in the Trenddisplay on the Portrait Central ViewerApplication. | Physiological Events (i.e., alarms) ofmedium or high priority are stored inthe Portrait Mobile MonitoringSolution and are viewable in the Trenddisplay on the Portrait Central ViewerApplication. | Identical |
| Trend resolution | 1-minute resolution | 1-minute resolution | Identical |
| RESPIRATION | |||
| Measurement Method | Impedance Respiration (i.e.,Impedance Pneumography) | Impedance Respiration (i.e.,Impedance Pneumography) | Identical |
| Display Range | 0 to 99 breaths/min | 0 to 99 breaths/min | Identical |
| Units | Respiration Rate (RR) in breaths/min | Respiration Rate (RR) in breaths/min | Identical |
| Resolution | 1 breath/min | 1 breath/min | Identical |
| Accuracy Range | 4 to 60 breaths/min | 4 to 60 breaths/min | Identical |
| Accuracy | +/- 3 breaths/min | +/- 3 breaths/min | Identical |
| Waveforms | Impedance Respiration waveform withautomatic scaling, 6.25 mm/s sweepspeed (Portrait Mobile PatientMonitor), 25 mm/s sweep speed(Portrait Central Viewer Application) | Impedance Respiration waveform withautomatic scaling, 6.25 mm/s sweepspeed (Portrait HUBXB with softwarePortrait HSWXB, 25 mm/s sweepspeed (Portrait Central ViewerApplication) | Identical (except for name change ofthe device) |
| PULSE OXIMETRY - SpO2 | |||
| Measurement | Arterial oxygen saturation (SpO2) | Arterial oxygen saturation (SpO2) | Identical |
| Units | Percent | Percent | Identical |
| Display Range | 0 to 100% | 0 to 100% | Identical |
| Resolution | 1% | 1% | Identical |
| Accuracy Range | 70 to 100% | 70 to 100% | Identical |
| Accuracy | SpO2 (70 to 100%) <= 2%Low perfusion SpO2: <= 3%With motion: <= 3%SpO2 (<70%) - Unspecified | SpO2 (70 to 100%) <= 2%Low perfusion SpO2: <= 3%With motion: <= 3%SpO2 (<70%) - Unspecified | Identical |
| Averaging | Adjustable from 0 to 60 seconds(default 10 seconds) | Adjustable from 0 to 60 seconds(default 10 seconds) | Identical |
| Optical Wavelengths / power | Pulse oximetry sensor LED peakwavelengths are within 600 to 1000nm and the maximum optical outputpower for each LED is less than 15mW. | Pulse oximetry sensor LED peakwavelengths are within 600 to 1000nm and the maximum optical outputpower for each LED is less than 15mW. | Identical |
| Specification | Reference Predicate DevicePortrait Mobile Monitoring Solution -K230626 | Proposed devicePortrait Mobile Monitoring Solution - | Discussion of differences betweenPortrait Mobile Monitoring Solutionand Predicate |
| Waveforms | Pleth Waveform, Not normalized(Amplitude of the displayedplethysmographic waveform reflectsthe strength of the arterial bloodpulsation at the measurement site.)25mm/s sweep speed (Portrait MobilePatient Monitors and Portrait CentralViewer Application)PULSE OXIMETRY - Pulse Rate | Pleth Waveform, Not normalized(Amplitude of the displayedplethysmographic waveform reflectsthe strength of the arterial bloodpulsation at the measurement site.)25mm/s sweep speed (Portrait MobilePatient Monitors and Portrait CentralViewer Application) | Identical |
| Measurement | Peripheral Pulse rate (PR) | Peripheral Pulse rate (PR) | Identical |
| Units | Beats per minute (bpm) | Beats per minute (bpm) | Identical |
| Display Range | 30 to 300 bpm | 30 to 300 bpm | Identical |
| Resolution | 1 bpm | 1 bpm | Identical |
| Accuracy Range | 30 to 250 bpm | 30 to 250 bpm | Identical |
| Accuracy | <= 2 bpm (30 to 250 bpm)Low perfusion: <= 2 bpm (30 to 250bpm)With motion: <=5 bpm (30 to 250bpm) | <= 2 bpm (30 to 250 bpm)Low perfusion: <= 2 bpm (30 to 250bpm)With motion: <=5 bpm (30 to 250bpm) | Identical |
{20}------------------------------------------------
{21}------------------------------------------------
{22}------------------------------------------------
{23}------------------------------------------------
{24}------------------------------------------------
{25}------------------------------------------------
{26}------------------------------------------------
{27}------------------------------------------------
{28}------------------------------------------------
Determination of Substantial Equivalence (807.92(b)(1)):
Summary of Non-Clinical Tests:
Bench testing related to software, hardware and performance including applicable consensus standards was conducted on the Portrait Mobile Monitoring Solution v1.1, demonstrating the design meets the specifications.
This section addresses the Non-Clinical testing for Portrait Mobile Monitoring Solution v1.1 relied on for a determination of substantial equivalence to the predicate K230626 Portrait Mobile Monitoring Solution v1.0.
Per the FDA guidance titled "Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions; Guidance for Industry and Food and Drug Administration Staff, Document issued on December 20, 2019", the following was verified:
- . SpO2 measurement (no changes have been made to the SpO2 algorithms since K230626, only regression testing performed)
- low perfusion condition (this testing covers the full ● accuracy range in challenging low-perfusion conditions using a simulator, proving that there has been no system regression)
- Pulse Rate (PR) measurement, non-motion condition ●
- Pulse Rate (PR) measurement, motion condition
- Pulse Rate (PR) measurement, low perfusion condition
- Impedance Respiration (no changes have been made to the Respiration Rate algorithms since K230626, only regression testing performed)
- . Respiration Rate (RR)
- Apnea
- Alarms Bench Testing ●
- Manuals Bench Testing ●
- Compliance to FDA Recognized Standards
- Compliance to FDA Human Factors Guidance
- Compliance to FDA ECG Electrodes Guidance ●
- Compliance to FDA Pulse Oximeters Guidance .
{29}------------------------------------------------
The Portrait Mobile Monitoring Solution v1.1 meets the EMC requirements described in 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. In addition, Portrait Mobile Monitoring Solution v1.1 shows compliance according to the "Electromagnetic Compatibility (EMC) of Medical Devices Guidance for Industry and Food and Drug Administration Staff, issued on June 6, 2022"
The Portrait Mobile Monitoring Solution meets the electrical safety requirements of ANSI AAMI ES60601-1:2005/2012 & A1:2012, C1:2009/2012 & A2:2010/2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]".This testing was performed by a recognized independent and Certified Body Testing Laboratory (CBTL) under the IECEE CB Scheme. The Portrait Mobile Monitoring Solution was designed and tested for compliance to the FDA 21 CFR Part 898, § 898.12 (Performance standard for electrode lead wires and cables). The performance standard is fulfilled because of compliance with IEC 60601-1:2005 + A1:2012+A2:2020, Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance - Edition 3.2, clause 8.5.2.3, which is equivalent with clause 56.3c in IEC 60601-1:1988.
Additional data is provided for compliance to:
- 60601-1-8 Edition 2.2 2020-07 CONSOLIDATED . VERSION Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 80601-2-49 Edition 1.0 2018-03 Medical electrical . equipment - Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitors - Edition 1.0
- ISO 80601-2-61 Second edition 2017-12 (Corrected . version 2018-02) Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
{30}------------------------------------------------
- . Guidance for Industry - Pulse Oximeters-Premarket Notification submission 510(k)s: Document issued on March 4, 2013
- Guidance for Industry and Food and Drug Administration . Staff; Class II Special Controls Guidance Document: Electrocardiograph Electrodes, Document issued on: July 21, 2011, NOTE: Included because the Portrait Mobile Monitoring Solution includes electrodes used to measure changes in thoracic impedance in order to measure a patient's respiratory rate.
The Portrait Mobile Monitoring Solution follows the FDA Biocompatibility guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" Guidance for Industry and Food and Drug Administration Staff- Document issued on: September 8, 2023" and ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Testing data for showing biocompatibility of patient contacting devices is provided in the submission.
The Portrait Mobile Monitoring Solution follows the guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, Guidance for Industry and Food and Drug Administration Staff. Document issued on: March 17, 2015" and the following standards:
-
· 17664-2 First edition 2021-02 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices.
The Portrait Mobile Monitoring Solution follows the Applying Human Factors and Usability Engineering to Medical Devices: Guidance for Industry and Food and Drug Administration Staff, Document issued on: February 3, 2016 and the following standards: -
IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
{31}------------------------------------------------
- IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED ● VERSION Medical devices - Part 1: Application of usability engineering to medical devices
Summative Usability testing has been concluded with 15 US Clinical and 15 US Technical users. The usability testing of the Portrait Mobile Monitoring Solution follows the FDA Guidance for Industry and Food and Drug Administration Staff "Applying Human Factors and Usability Engineering to Medical Devices" (Feb 3, 2016).
Batteries performance data is provided related to:
- . IEC 62133-2:2017/AMD1:2021 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- UL 2054 2nd Edition Household and Commercial ● Batteries
Wireless performance data was provided related to:
- . Radio Frequency Wireless Technology in Medical Devices; Guidance for Industry and Food and Drug Administration Staff, Document issued on: August 14, 2013
- IEEE ANSI C63.27-2021 American National Standard for . Evaluation of Wireless Coexistence
Additional Labeling standards followed:
- . ISO 20417 First edition 2021-04 Corrected version 2021-12 Medical devices — Information to be supplied by the manufacturer
- 15223-1 Fourth edition 2021-07 Medical devices -● Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
The Portrait Mobile Monitoring Solution follows the FDA software guidance documents as outlined in this submission.
- . Content of Premarket Submissions for Device Software Functions, Guidance for Industry and Food and Drug Administration Staff Document issued on June 14, 2023
- General Principles of Software Validation: Final ●
{32}------------------------------------------------
Guidance for Industry and FDA Staff, January 11, 2002
- . Off-The-Shelf Software Use in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, August 11, 2023
- Design Considerations and Premarket Submission ● Recommendations for Interoperable Medical Devices, Guidance for Industry and Food and Drug Administration Staff. Document issued on: September 6, 2017
Software testing was conducted, and documentation was provided as recommended by Content of Premarket Submissions for Device Software Functions, Guidance for Industry and Food and Drug Administration Staff: Document issued on June 14, 2023" The software for this device is considered as a "Enhanced" documentation level. Software standards IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes and the risk management standard ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices were also applied to the design.
Patient safety, security, and privacy risks have been addressed in the design and development of Portrait Mobile Monitoring Solution v1.1 including a Security Risk Assessment, Threat model and Penetration testing. This includes system integrity controls, access controls, audit controls, network controls, and remote service controls which address the General Principles and Security Capabilities outlined in the "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions Guidance for Industry and Food and Drug Administration Staff Document issued on September 27, 2023".
In addition, the Portrait Mobile Monitoring Solution v1.1 complies with section 524B "Ensuring Cybersecurity of Medical Devices" of the FD&C Act. The Portrait Mobile Monitoring Solution v1.1 meets the definition of a "cyber device" per Section 524B(c) of the FD&C Act . To meet the requirements under section "3305 of the Consolidated Appropriations Act, 2023, enacted on December 29, 2022, section 524B "Ensuring Cybersecurity of Medical Devices" to the FD&C Act" GE HealthCare has also specifically attached a complete cybersecurity management plan and SBOMs.
{33}------------------------------------------------
Clinical (807.92(b)(2)): Summary of Clinical Tests:
The Portrait Mobile Monitoring Solution v1.1 does not contain clinical testing data.
- Conclusion (807.92(b)(3)): GE HealthCare considers the Portrait Mobile Monitoring Solution v1.1 to be substantially equivalent to the predicate device.
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).