K Number
K234115
Device Name
APRO 55 Catheter
Manufacturer
Date Cleared
2024-03-15

(79 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices.

Device Description

The APRO 55 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 55 Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 55 Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 55 Catheter is provided with an Introducer Sheath.

AI/ML Overview

The provided document is a 510(k) summary for a medical device (APRO 55 Catheter) seeking FDA clearance, not a study report detailing acceptance criteria for an AI/ML device. Therefore, it does not contain information typically found in such a study, such as acceptance criteria for machine learning performance metrics (e.g., sensitivity, specificity, AUC), sample sizes for test sets in an AI context, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone algorithm performance, or training set details.

The document describes non-clinical bench testing for a physical medical catheter, including mechanical, material, and biocompatibility tests. The "acceptance criteria" mentioned are for these engineering and biological tests, not for an AI/ML diagnostic or predictive model.

Therefore, I cannot fulfill your request for information related to AI/ML acceptance criteria and study details based on the provided text.

{0}------------------------------------------------

March 15, 2024

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Alembic, LLC % Prithul Bom Most Responsible Person Regulatory Technology Services 1000 Westgate Drive, Suite #510k Saint Paul. Minnesota 55114

Re: K234115

Trade/Device Name: APRO 55 Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, QJP Dated: March 6, 2024 Received: March 6, 2024

Dear Prithul Bom:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K234115

Device Name APRO 55 Catheter

Indications for Use (Describe)

The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

510(k) Number: K234115

This 510(k) Summary is provided in accordance with the requirements of 21 CFR §807.92.

Submitter information 1)

Date Prepared:

Submitter:Alembic, LLC627 National Ave.Mountain View, CA 94043
Contact:Lisa YenDirector of Regulatory and QualityTelephone Number: (650) 388-5087Fax: (650) 390-0107Email: lyen@alembicllc.com

March 13, 2024

2) Device Name and Classification

Trade/Proprietary Name:APRO® 55 Catheter
Common Name:Percutaneous Catheter
Classification Name:Percutaneous Catheter, 21 CFR 870.1250Catheter, Percutaneous, Neurovasculature, 21 CFR 870.1250
Regulatory Class:Class II
Product Codes:DQY, QJP
Review Panel:Cardiovascular, Neurology

3) Legally Marketed Predicate Device and Reference Device

Predicate Device:K151667 AXS Catalyst Distal Access Catheter
Reference Device:K223545 APRO 70 Catheter and Alembic Aspiration Tubing

4) Device Description

The APRO 55 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 55 Catheter is designed to facilitate the insertion and guidance of interventional devices into peripheral and neuro vasculature. Using standard catheterization techniques under fluoroscopic guidance, the APRO 55 Catheter is introduced through a guide catheter or guide sheath and over a guidewire into the target vasculature. The distal segment of the catheter shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal tip of the catheter for visualization under fluoroscopy. The APRO 55 Catheter is provided with an Introducer Sheath.

{4}------------------------------------------------

5) Indications for Use

The APRO 55 Catheter is indicated for use in facilitating the insertion and guidance of appropriately sized interventional devices into a selected blood vessel in the peripheral and neurovascular systems. The APRO 55 Catheter is also indicated for use as a conduit for retrieval devices.

6) Technological Characteristics Comparison

Alembic has demonstrated that the APRO 55 Catheter is substantially equivalent to the predicate device based on the similarity in materials, similarity in design concept, and the same fundamental operating principles. A comparison of the APRO 55 Catheter with the predicate device and reference device is summarized in Table 1 below.

CategorySubject DeviceAPRO 55 CatheterPredicate DeviceAXS Catalyst Distal AccessCatheterReference DeviceAPRO 70 Catheter andAlembic Aspiration Tubing
510(k) NumberK234115K151667K223545
Regulatory ClassClass II, 21 CFR 870.1250,DQY, QJPClass II, 21 CFR 870.1250,DQYClass II, 21 CFR 870.1250,NRY
Indications forUseThe APRO 55 Catheter isindicated for use infacilitating the insertion andguidance of appropriatelysized interventional devicesinto a selected blood vessel inthe peripheral andneurovascular systems. TheAPRO 55 Catheter is alsoindicated for use as a conduitfor retrieval devices.The AXS Catalyst DistalAccess Catheter is indicated foruse in facilitating the insertionand guidance of appropriatelysized interventional devices intoa selected blood vessel in theperipheral and neurovascularsystems. The AXS CatalystDistal Access Catheter is alsoindicated for use as a conduitfor retrieval devices.The APRO 70 Catheter with anaspiration pump and the AlembicAspiration Tubing is intended foruse in the revascularization ofpatients with acute ischemic strokesecondary to intracranial largevessel occlusive disease (within theinternal carotid, middle cerebral -M1 and M2 segments, basilar, andvertebral arteries) within 8 hours ofsymptom onset. Patients who areineligible for intravenous tissueplasminogen activator (IV t-PA) orwho fail IV t-PA therapy arecandidates for treatment.The Alembic Aspiration Tubing isintended to connect the APRO 70Catheter to the aspiration pump.
Materials
HubNylonNylonPolycarbonate
AdhesiveCyanoacrylateNoneCyanoacrylate
Strain ReliefThermoplastic elastomerThermoplastic rubberThermoplastic elastomer
LinerPolytetrafluoroethylene(PTFE)/Tecoflex compositePTFEPTFE/Tecoflex composite
Shaft Coil orBraid304V stainless steel braid304V stainless steel coilStainless steel with nitinol wireand polymer fiber304V stainless steel braid304V stainless steel coil
ExtrusionsThermoplastic polyurethanes,thermoplastic elastomerPebax, NylonThermoplastic polyurethanes,thermoplastic elastomer
Marker BandPlatinum/ iridiumPlatinum/ iridiumPlatinum/ iridium
AdhesiveNoneCyanoacrylateNone
CoatingHydrophilic coating, identicalHydrophilic coatingHydrophilic coating

Table 1 - APRO 55 Catheter Comparison with the Predicate Device and Reference Device

{5}------------------------------------------------

CategorySubject DeviceAPRO 55 CatheterPredicate DeviceAXS Catalyst Distal AccessCatheterReference DeviceAPRO 70 Catheter andAlembic Aspiration Tubing
to reference device
Dimensions
Proximal OuterDiameter0.066 inch5.6 F (0.073 inch)6.0 F (0.078 inch)0.083 inch
Distal OuterDiameter0.066 inch5.3 F (0.069 inch)5.4 F (0.070 inch)0.083 inch
Inner Diameter0.055 inch0.058 inch, 0.060 inch0.070 inch
Effective Lengths125, 137 cm115, 132 cm125, 132, 135 cm
Coated Length90, 102 cm65, 82 cm90, 97, 100 cm
Tip ShapeStraightStraightStraight
Accessories
Introducer SheathYesYesYes
RotatingHemostasis ValveNoneYesNone
Packaging Materials
PouchNylon/polyethylene/ TyvekNylon/polyethylene/ TyvekNylon/polyethylene/ Tyvek
Packaging TubePolyethylenePolyethylenePolyethylene
Packaging CardYesYesYes
Shelf CartonSolid bleached sulfatepaperboardSolid bleached sulfatepaperboardSolid bleached sulfatepaperboard
Other
SterilizationEthylene oxideEthylene oxideEthylene oxide
Shelf Life6 months2 years6 months
Use ConditionsSterile, single use, disposableSterile, single use, disposableSterile, single use, disposable

Performance Data 7)

Alembic performed non-clinical bench, sterility, shelf-life, and biocompatibility testing. The results demonstrate substantial equivalence of the APRO 55 Catheter to the legally marketed predicate device.

A. Design Verification Testing - Non-Clinical Bench

Performance testing was conducted to support the APRO 55 Catheter submission. The results of the design verification and validation testing performed confirm that the APRO 55 Catheter conforms to the predefined specifications and met test acceptance criteria. A summary of the testing is shown in Table 2.

TestAcceptance CriteriaConclusion
Visual and DimensionalCharacteristicsCatheter meets the visual and dimensionalspecifications.The APRO 55 Catheter met theacceptance criteria.
ParticulateCatheter meets the acceptance criteria.Subject device was evaluated with apredicate device under the same testconditions.The APRO 55 Catheter met theacceptance criteria.
Kink ResistanceCatheter shaft shall not kink at clinicallyrelevant radii.The APRO 55 Catheter met theacceptance criteria.
Hub Air LeakageCatheter does not leak air into hubThe APRO 55 Catheter met the

Table 2 – Summary of Non-Clinical Bench Test Results

{6}------------------------------------------------

TestAcceptance CriteriaConclusion
Hub Compatibilityassembly with methods specified in ISO10555-1, Annex D.acceptance criteria.
Torque StrengthCatheter meets the requirements specifiedin ISO 80369-7.The APRO 55 Catheter met theacceptance criteria.
Dynamic Burst PressureCatheter must withstand the minimumrequired number of rotations withoutbreakage and without kinking compared tolegally marketed devices.The APRO 55 Catheter met theacceptance criteria.
Liquid LeakageNo damage to catheter with dynamicpressure.The APRO 55 Catheter met theacceptance criteria.
Static BurstCatheter must withstand pressurewith methods specified in ISO 10555-1, Annex C.The APRO 55 Catheter met theacceptance criteria.
Hub and Shaft TensileStrengthCatheter must withstand pressuresanticipated for clinical use.The APRO 55 Catheter met theacceptance criteria.
Tip Tensile StrengthCatheter hub and shaft must meet tensilestrength specification.The APRO 55 Catheter met theacceptance criteria.
Tip Buckling ForceCatheter tip must meet tip tensile strengthspecification.The APRO 55 Catheter met theacceptance criteria.
Delivery and RetrievalForceCatheter tip buckling force must beacceptable. Forces were compared to apredicate.The APRO 55 Catheter met theacceptance criteria.
Simulated UseCatheter delivery and retrieval force mustbe acceptable. Forces were compared to apredicate.The APRO 55 Catheter met theacceptance criteria.
When used per the Instructions for Usewith accessory devices in an anatomicalneurovascular model, the Catheter mustmeet functionality specifications.The APRO 55 Catheter met theacceptance criteria.

B. Design Verification Testing - Animal

Substantial equivalence was established based on non-clinical bench performance data. Animal testing data were not deemed necessary.

C. Sterilization and Shelf-Life

The APRO 55 Catheter is sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of 1 x 10-6 in accordance with ISO 11135.

Aging studies for the APRO 55 Catheter have established that the subject device and packaging remain functional for the labeled expiration date. Aging studies for packaging integrity, seal strength, and the device functionality were performed and met the acceptance criteria.

D. Biocompatibility

Biocompatibility testing has been completed for the APRO 55 Catheter in accordance with ISO 10993-1 and the device is deemed non-toxic (local or systemic), non-sensitizing, not locally irritating or otherwise harmful. Test results obtained were acceptable for the intended

{7}------------------------------------------------

use as shown in Table 3. The thrombogenicity evaluation of the APRO 55 Catheter was supplemented by the prior in vivo thrombogenicity evaluation of the reference device, APRO 70 Catheter, in a good laboratory practice animal study.

TestResultsConclusions
Sensitization(Guinea Pig Maximization)The APRO 55 Catheter did not elicit asensitization response.Non-sensitizing
Irritation/IntracutaneousReactivityThe APRO 55 Catheter demonstrated noevidence of irritation.Non-irritant
Cytotoxicity(MEM Elution, L929 cells)The APRO 55 Catheter did not elicit acytotoxic response at 24 hours and 48 hours.Non-cytotoxic
Hemolysis - IndirectThere were no significant differences betweenthe test article extract and negative controlarticle results.Non-hemolytic
Hemolysis - DirectThere were no differences between thehemolytic index of the test article and thenegative control.Non-hemolytic
Partial ThromboplastinTime (PTT)The average clotting time of the test article wasgreater than vehicle control and negativecontrol.Acceptable clottingtimes
SC5b9 ComplementActivationThe Sc5b9 concentration of the test article wasstatistically less than the positive control and wasnot statistically higher than the negative control.Acceptable
Acute Systemic ToxicityNo weight loss, mortality, or evidence ofsystemic toxicity from the extract exposure tothe mice.Non-toxic
Material-MediatedPyrogenicityAll individual rabbits for both the test articleand negative control showed a total rise intemperature of < 0.5 °C and were determined tobe nonpyrogenic.Non-pyrogenic

Table 3 - Biocompatibility Test Results

E. Clinical Testing

Substantial equivalence was established based on non-clinical performance data. Human clinical data were not deemed necessary.

8) Conclusion

Based on the comparison of the technological characteristics and the non-clinical testing, the subject device is found to be substantially equivalent to the predicate device. The differences in technological characteristics do not raise new questions of safety and effectiveness. Testing was conducted to demonstrate that the subject device meets the specifications and performs as intended.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).