(262 days)
Yes
The device description explicitly states "Machine learning skeletal and anatomic structure segmentation."
No.
The device provides quantitative analysis of medical images to assist in interpretation and reporting, but it is explicitly stated that it is "not intended to diagnose any disease, replace the diagnostic procedures for interpretation of CT or PET/CT images, recommend any specific treatment, nor is it intended to replace the skill and judgment of a qualified medical professional." It is an adjunct to image interpretation, not a device that directly applies or provides therapy.
No
The "Intended Use / Indications for Use" section explicitly states "TRAQinform IQ is not intended to diagnose any disease, replace the diagnostic procedures for interpretation of CT or PET/CT images, recommend any specific treatment, nor is it intended to replace the skill and judgment of a qualified medical professional." Instead, it provides quantitative analysis of lesions as an adjunct to locally reported image interpretation.
Yes
The device description explicitly states "TRAQinform IQ is a software only device".
Based on the provided information, TRAQinform IQ is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs are used to examine specimens derived from the human body. The intended use and device description clearly state that TRAQinform IQ analyzes imaging data (PET/CT DICOM compliant imaging data). This data is acquired from the patient, but the device is processing the images themselves, not biological specimens like blood, urine, or tissue.
- The device's function is image processing and quantitative analysis of imaging data. It performs tasks like ROI matching, segmentation, quantitative analysis of volume, density, and tracer uptake from the images. This is distinct from the analysis of biological markers or substances within a specimen, which is the core function of an IVD.
- The output is a report based on image analysis. The TRAQinform Report provides quantitative information derived from the images. It does not provide a diagnostic result based on the analysis of a biological specimen.
While the information derived from the images can be used by clinicians in the diagnostic process, the device itself is performing image analysis, not in vitro testing of a biological sample.
No
The letter does not state that the FDA has reviewed and approved or cleared a PCCP for this specific device.
Intended Use / Indications for Use
TRAQinform IQ is a software only device that provides a quantitative TRAQinform Report on lesions identified as Regions of Interest (ROI) in PET/CT DICOM compliant imaging data acquired, interpreted, and reported on per local practice prior to device use.
Clinicians responsible for patient care and for ordering TRAQinform Reports as an adjunct to locally reported image interpretation do not interact directly with the device. Clinicians responsible for local image interpretation do not interact with the device and generate their reporting before and independently of the TRAQinform Report.
The TRAQinform Report is generated by the device manufacturer and signed by a U.S. board certified physician responsible for supervising central report generation and qualified to practice nuclear radiology/medicine. The TRAQinform Report is for use by trained medical professionals including but not limited to oncologists, nuclear radiologists/physicians, medical imaging technologists, dosimetrists, and physicists.
TRAQinform IQ software contains the following functionalities:
- Automated matching of ROI between previously performed CT and PET/CT DICOM 3.0 volumetric medical images.
- In order to perform automated matching of ROI and quantitative analysis of previously performed CT and PET/CT DICOM 3.0 volumetric medical images, the software initially performs the following functions:
- Machine learning skeletal and anatomic structure segmentation.
- Threshold-based ROI identification and contouring.
- Automated quantitative analysis to assess previously performed CT and PET/CT DICOM 3.0 volumetric medical images, including: change in total volume and density of each identified ROI, and change in Fludeoxyglucose F18 (FDG) tracer uptake of each identified ROI among images.
- Generation of images of the anatomy combined with spatial and quantitative information, including computed classification of quantitative FDG ROI changes.
For multi-timepoint quantitative analysis, recommended use is in adult patients 22 years and older with partial or whole-body PET/CT acquired following administration of FDG per approved drug prescribing information and with the second FDG administration separated from the first by a period not to exceed 12 months.
For single-timepoint quantitative analysis, recommended use is in adult patients 22 years and older with partial or whole-body PET/CT following administration of FDG, a PSMA targeted PET drug, or a SSTR-targeted PET drug per approved drug prescribing information.
Discrepancies between TRAQinform IQ and local PET/CT reporting have been investigated and use of TRAQinform IQ has not been established for binary patient level progression or non-progression decisions without multidisciplinary review. Discrepancies between TRAQinform IQ and local PET/CT reporting that could impact patient care should therefore prompt consultation with subject matter experts (for example, in tumor board), with a patient-centered focus on discrepant imaging regions and with blinded or otherwise neutral adjudication regarding interpretation/classification source.
TRAQinform IQ is not intended to diagnose any disease, replace the diagnostic procedures for interpretation of CT or PET/CT images, recommend any specific treatment, nor is it intended to replace the skill and judgment of a qualified medical professional.
Product codes
LLZ
Device Description
TRAQinform IQ is a software only device that provides quantitative analysis of lesions identified as Regions of Interest (ROI) in PET/CT DICOM compliant imaging data acquired, interpreted, and reported on per local practice prior to device use.
The input to TRAQinform IQ is CT and PET/CT images as supported by ACR/NEMA DICOM 3.0.
The following steps are performed by the software:
- Automatic threshold-based ROI segmentation:
- ROI can also be imported from external sources (other validated tools or manual contouring or by qualified medical personnel).
- Automatic ROI registration between multiple images:
- Images can be from the same or different imaging modality.
- Images can be from the same or different PET tracer.
- Images can be from the same or different date.
- Automatic matching of ROI between multiple, previously performed images.
- Automatic quantification of dynamic changes among images including, but not limited to:
- Changes in ROI shape.
- Single ROI splitting into multiple ROI.
- Multiple ROI combining into a single ROI.
- ROI appearing, disappearing, and re-appearing across images.
- A comprehensive summary analysis.
TRAQinform IQ calculates spatial and quantitative metrics for each individual ROI. These metrics are provided as a TRAQinform Report. TRAQinform IQ uses computational algorithms to detect, fuse and analyze ROI and provides the following outputs:
- Identification of anatomic location of ROI in all areas of the body.
- A quantitative analysis of functional and anatomic data for CT and PET/CT scans, including:
- Volume of all identified ROI on each image;
- Change in volume of each identified ROI among images:
- Total volume of all identified ROI on each image;
- Change in total volume of all identified ROI on each image;
- Heterogeneity of change in volume of each identified ROI:
- For PET scans:
- Tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of each identified ROI on each image:
- Change in tracer uptake (SUVmax, SUVtotal, SUVmean) of each identified ROI among images:
- Total tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of all identified ROI on each image:
- Change in total tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of all identified ROI on each image:
- Heterogeneity of change in tracer uptake (SUVhetero) among identified ROI.
- For CT scans
- Radio density (HUmax, HUtotal, HUmean, HUhetero) for each identified ROI on each image:
- Changes in radio density (HUmax, HUtotal, HUmean) of all identified ROI on each image:
- Change in total radio density (HUmax, HUtotal, HUmean, HUhetero) of all identified ROI on each image;
- Heterogeneity of change in radio density (HUhetero) among identified ROI.
- 2D graphical renderings of medical images, including Maximum Intensity Projections of the PET and CT, with overlayed and labeled/color-coded ROI, for inclusion in TRAQinform Reports.
- 3D labeled contours for ROI, anatomic structures, and skeletal structures.
The TRAQinform IQ software operates in a secure cloud environment.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Yes
Input Imaging Modality
PET/CT, CT
Anatomical Site
Not Found
Indicated Patient Age Range
adult patients 22 years and older
Intended User / Care Setting
Trained medical professionals including but not limited to oncologists, nuclear radiologists/physicians, medical imaging technologists, dosimetrists, and physicists.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
A study was conducted to evaluate the clinical utility of TRAQinform IQ. In this study, three report evaluators (oncologists) were assigned to distinguish each of 103 patients as positive versus negative for progression based on presentation of the locally generated radiology report without and then with adjunctive presentation of the TRAQinform Report.
Each of the 103 patients had two sequential FDG PET/CT scans acquired between 2005 and 2022. Scans were acquired from patients scanned at 10 or more imaging centers in at least 3 U.S. states (exact image center information was redacted on scan transfer). A full breakdown is shown in Table 2. A breakdown of patients by disease type, patient sex, weight, and age are also shown in Table 2. Out of the 206 total scans (2 per patient), scanner information was available for 199 (99 patients). Scans were acquired on a variety of scanner manufacturers and models, with 84 patients receiving scans on the same scanner for baseline and follow-up.
A total of 1046 ROI were detected across the baseline scans (median per patient: 6, range: 0-80) and 935 were detected on the follow-up (median per patient: 4, range: 0-60).
Panel agreement was also evaluated at the ROI level. The panel examined only the subset of ROI that had been highlighted by the report evaluators. Of 111 total ROIs highlighted on the TRAQinform Report.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance Testing:
Results of software validation and verification (V&V) and early feasibility testing of device component functionality for CT segmentation, regional PET thresholding, and ROI matching supported further FDG test-retest and pivotal device output investigation. For detail on component testing, published reporting is available respectively in Weisman 2023 (CT anatomy segmentation, doi.org/10.1088/2057-1976/acfb06), Perk 2018 (ROI detection methodology, doi.org/10.1088/1361-6560/aaebba), and Huff 2023 (ROI matching, doi.org/10.1088/1361-6560/acef8f).
Test-Retest Study:
This study established test-retest limits of agreement used to categorize regions of interest based off changes between FDG PET/CT images. 31 patients with non-small lung cancer patients received two FDG PET/CT images within 1 week pretreatment. The limits of repeatability endpoint estimated the interval within which 95% of repeated measurements are expected to lie when clinically relevant change is lacking. For TRAQinform IQ generated features, these limits are shown in Table 1.
SUVmax: Lower limit (%) -27.0, Upper limit (%) 56.8
SUVmean: Lower limit (%) -20.2, Upper limit (%) 38.5
SUVtotal: Lower limit (%) -54.1, Upper limit (%) 144.5
Volume: Lower limit (%) -52.6, Upper limit (%) 113.9
Pivotal Reader Study:
A study was conducted to evaluate the clinical utility of TRAQinform IQ. In this study, three report evaluators (oncologists) were assigned to distinguish each of 103 patients as positive versus negative for progression based on presentation of the locally generated radiology report without and then with adjunctive presentation of the TRAQinform Report.
Of the 103 patients investigated, the TRAQInform Report prompted at least one oncologist interpreter to reevaluate 23 patients classified without the device as negative for progression to positive, and to reevaluate 9 patients classified without the device as positive for progression to negative. In these cases, the oncologist interpreter highlighted a subset of ROI for further review. Imaging and local reporting on these 23 + 9 = 32 patients was sent to a panel of two radiologists and one oncologist, together serving as a reference source against which to quantify positive percent agreement (PPA, representing patients classified as positive by both the panel and an oncologist), negative percent agreement (NPA; representing patients classified as negative by both the panel and an oncologist), and the 95% confidence interval (95%Cl) for overall percent agreement (OPA; prevalence-biased PPA and NPA aggregation). Compared to chance performance (50%), the 95% CI for OPA was 41% to 76%. PPA was 14/18=78% and NPA was 5/14=36%.
Because reevaluation at the patient level was derived from reporting at a lower ROI level, panel agreement was also evaluated at this level. The panel examined only the subset of ROI that had been highlighted by the report evaluators. Of 111 total ROIs highlighted on the TRAQInform Report, the panel agreed with the device classification (new, increasing, unchanged, decreasing, and disappearing) in 37/53=70%, 18/26=70%, 11/13=85%, 5/8 (63%), and 10/11 (90%) ROIs, respectively.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Test-Retest:
Limits of Repeatability (95% of repeated measurements):
SUVmax: -27.0% to 56.8%
SUVmean: -20.2% to 38.5%
SUVtotal: -54.1% to 144.5%
Volume: -52.6% to 113.9%
Pivotal Reader Study:
Overall Percent Agreement (OPA): 95% CI for OPA was 41% to 76%.
Positive Percent Agreement (PPA): 14/18 = 78% (for patients classified as positive by both the panel and an oncologist)
Negative Percent Agreement (NPA): 5/14 = 36% (for patients classified as negative by both the panel and an oncologist)
ROI-level agreement with panel classification:
New ROIs: 37/53 = 70%
Increasing ROIs: 18/26 = 70%
Unchanged ROIs: 11/13 = 85%
Decreasing ROIs: 5/8 = 63%
Disappearing ROIs: 10/11 = 90%
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
December 5, 2024
AIQ Global, Inc. % Kelliann Payne Partner Hogan Lovells US LLP 1735 Market Street Floor 23 Philadelphia, Pennsylvania 19103
Re: K233998
Trade/Device Name: TRAQinform IQ Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ
Dear Kelliann Payne:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated September 5, 2024. Specifically, FDA is updating this SE Letter to address a typo in the IFU form as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Daniel Krainak, OHT8: Office of Radiological Health, 301-796-0478, daniel.krainak@fda.hhs.gov.
Sincerely.
D. R. K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
1
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol of the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
September 5, 2024
AIQ Global, Inc. % Kelliann Payne Partner Hogan Lovells US LLP 1735 Market Street Floor 23 Philadelphia, Pennsylvania 19103
Re: K233998
Trade/Device Name: TRAQinform IQ Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: December 18, 2023 Received: December 18, 2023
Dear Kelliann Payne:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
3
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D. G. K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
4
Indications for Use
510(k) Number (if known) K233998
Device Name
TRAQinform IQ
Indications for Use (Describe)
TRAQinform IQ is a software only device that provides a quantitative TRAOinform Report on lesions identified as Regions of Interest (ROI) in PET/CT DICOM compliant imaging data acquired, interpreted, and reported on per local practice prior to device use.
Clinicians responsible for patient care and for ordering TRAQinform Reports as an adjunct to locally reported image interpretation do not interact directly with the device. Clinicians responsible for local image interpretation do not interact with the device and generate their reporting before and independently of the TRAQinform Report.
The TRAQinform Report is generated by the device manufacturer and signed by a U.S. board certified physician responsible for supervising central report generation and qualified to practice nuclear radiology/medicine. The TRAQinform Report is for use by trained medical professionals including but not limited to oncologists, nuclear radiologists/physicians, medical imaging technologists, dosimetrists, and physicists.
TRAQinform IQ software contains the following functionalities:
- · Automated matching of ROI between previously performed CT and PET/CT DICOM 3.0 volumetric medical images.
- · In order to perform automated matching of ROI and quantitative analysis of previously performed CT and PET/CT DICOM 3.0 volumetric medical images, the software initially performs the following functions:
- · Machine learning skeletal and anatomic structure segmentation.
- · Threshold-based ROI identification and contouring.
- · Automated quantitative analysis to assess previously performed CT and PET/CT DICOM 3.0 volumetric medical images, including: change in total volume and dentified ROI, and change in Fludeoxygluose F18 (FDG) tracer uptake of each identified ROI among images.
- · Generation of images of the anatomy combined with spatial and quantitative information, including computed classification of quantitative FDG ROI changes.
For multi-timepoint quantitative analysis, recommended use is in adult patients 22 years and older with partial or wholebody PET/CT acquired following administration of FDG per approved drug prescribing information and with the second FDG administration separated from the first by a period not to exceed 12 months.
For single-timepoint quantitative analysis, recommended use is in adult patients 22 years and older with partial or wholebody PET/CT following administration of FDG, a PSMA targeted PET drug, or a SSTR-targeted PET drug per approved drug prescribing information.
Discrepancies between TRAQinform IQ and local PET/CT reporting have been investigated and use of TRAQinform IQ has not been established for binary patient level progression decisions without multidisciplinary review. Discrepancies between TRAQinform IQ and local PET/CT reporting that could impact patient care should therefore prompt consultation with subject matter experts (for example, with a patient-centered focus on discrepant imaging regions and with blinded or otherwise neutral adjudication regarding interpretation/classification source.
TRAQinform IQ is not intended to diagnose any disease. replace the diagnostic procedures for interpretation of CT or PET/CT images, recommend any specific treatment, nor is it intended to replace the skill and judgment of a qualified medical professional.
5
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
6
K233998
510(k) SUMMARY AIQ Global, Inc's TRAQinform IQ
Submitter
AIQ Global, Inc. (dba AIQ Solutions) 8000 Excelsior Dr. Suite 400 Madison, WI 53717 (608) 268-9684
Contact Person
Alberti Dona, COO (608) 268-9684 dona.alberti@aiq-solutions.com
Date Prepared
Auqust 28, 2024
Device Identification
Proprietary Name: TRAQinform IQ Common Name: System, Image Processing, Radiological Regulation No: 21 CFR 892.2050 Picture Archiving and Communications System Product Code: LLZ Regulatory Class: Class II Classification Panel: Radiology
Predicate Device
Name of Device: QTxI 510(k) Number: K173444 Regulation No: 21 CFR 892.2050 Product Code: LLZ Regulatory Class: Class II Classification Panel: Radiology
Reference Device
Name of Device: MIM 4.1 (SEASTAR) 510(k) Number: K071964 Regulation No: 21 CFR 892.2050 Product Code: LLZ Regulatory Class: Class II Classification Panel: Radiology
7
Device Description and Technological Characteristics
TRAQinform IQ is a software only device that provides quantitative analysis of lesions identified as Regions of Interest (ROI) in PET/CT DICOM compliant imaging data acquired, interpreted, and reported on per local practice prior to device use.
The input to TRAQinform IQ is CT and PET/CT images as supported by ACR/NEMA DICOM 3.0.
The following steps are performed by the software:
- . Automatic threshold-based ROI segmentation:
- ROI can also be imported from external sources (other validated tools or manual contouring о by qualified medical personnel).
- Automatic ROI registration between multiple images:
- Images can be from the same or different imaging modality. O
- Images can be from the same or different PET tracer. O
- Images can be from the same or different date. O
- Automatic matching of ROI between multiple, previously performed images. ●
- Automatic quantification of dynamic changes among images including, but not limited to: ●
- Changes in ROI shape. o
- Single ROI splitting into multiple ROI. O
- Multiple ROI combining into a single ROI. O
- ROI appearing, disappearing, and re-appearing across images. O
- A comprehensive summary analysis. ●
TRAQinform IQ calculates spatial and quantitative metrics for each individual ROI. These metrics are provided as a TRAQinform Report. TRAQinform IQ uses computational algorithms to detect, fuse and analyze ROI and provides the following outputs:
- . Identification of anatomic location of ROI in all areas of the body.
- . A quantitative analysis of functional and anatomic data for CT and PET/CT scans, including:
- Volume of all identified ROI on each image; O
- Change in volume of each identified ROI among images: O
- Total volume of all identified ROI on each image; O
- Change in total volume of all identified ROI on each image; O
- Heterogeneity of change in volume of each identified ROI: o
- For PET scans: O
- Tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of each identified ROI on . each image:
- . Change in tracer uptake (SUVmax, SUVtotal, SUVmean) of each identified ROI among images:
- . Total tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of all identified ROI on each image:
- I Change in total tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of all identified ROI on each image:
- . Heterogeneity of change in tracer uptake (SUVhetero) among identified ROI.
- For CT scans O
- Radio density (HUmax, HUtotal, HUmean, HUhetero) for each identified ROI on each '
8
image:
- Changes in radio density (HUmax. HUtotal. HUmean) of all identified ROI on each image:
- . Change in total radio density (HUmax, HUtotal, HUmean, HUhetero) of all identified ROI on each image;
- 트 Heterogeneity of change in radio density (HUhetero) among identified ROI.
- . 2D graphical renderings of medical images, including Maximum Intensity Projections of the PET and CT, with overlayed and labeled/color-coded ROI, for inclusion in TRAQinform Reports.
- . 3D labeled contours for ROI, anatomic structures, and skeletal structures.
The TRAQinform IQ software operates in a secure cloud environment.
Intended Use / Indications for Use
.
TRAQinform IQ is a software only device that provides a quantitative TRAQinform Report on lesions identified as Regions of Interest (ROI) in PET/CT DICOM compliant imaging data acquired, interpreted, and reported on per local practice prior to device use.
Clinicians responsible for patient care and for ordering TRAQinform Reports as an adjunct to locally reported image interpretation do not interact directly with the device. Clinicians responsible for local image interpretation do not interact with the device and generate their reporting before and independently of the TRAQinform Report.
The TRAQinform Report is generated by the device manufacturer and signed by a U.S. board certified physician responsible for supervising central report generation and qualified to practice nuclear radiology/medicine. The TRAQinform Report is for use by trained medical professionals including but not limited to oncologists, nuclear radiologists/physicians, medical imaging technologists, dosimetrists, and physicists.
TRAQinform IQ software contains the following functionalities:
- Automated matching of ROI between previously performed CT and PET/CT DICOM 3.0 . volumetric medical images.
- . In order to perform automated matching of ROI and quantitative analysis of previously performed CT and PET/CT DICOM 3.0 volumetric medical images, the software initially performs the following functions:
- Machine learning skeletal and anatomic structure segmentation.
- Threshold-based ROI identification and contouring.
- Machine learning skeletal and anatomic structure segmentation.
- . Automated quantitative analysis to assess previously performed CT and PET/CT DICOM 3.0 volumetric medical images, including: change in total volume and density of each identified ROI, and change in Fludeoxyglucose F18 (FDG) tracer uptake of each identified ROI among images.
- . Generation of images of the anatomy combined with spatial and quantitative information. including computed classification of quantitative FDG ROI changes.
For multi-timepoint quantitative analysis, recommended use is in adult patients 22 years and older with partial or whole-body PET/CT acquired following administration of FDG per approved drug prescribing information and with the second FDG administration separated from the first by a period not to exceed 12 months.
9
For single-timepoint quantitative analysis, recommended use is in adult patients 22 years and older with partial or whole-body PET/CT following administration of FDG, a PSMA targeted PET drug, or a SSTR-targeted PET drug per approved drug prescribing information.
Discrepancies between TRAQinform IQ and local PET/CT reporting have been investigated and use of TRAQinform IQ has not been established for binary patient level progression or non-progression decisions without multidisciplinary review. Discrepancies between TRAQinform IQ and local PET/CT reporting that could impact patient care should therefore prompt consultation with subject matter experts (for example, in tumor board), with a patientcentered focus on discrepant imaging regions and with blinded or otherwise neutral adjudication regarding interpretation/classification source.
TRAQinform IQ is not intended to diagnose any disease, replace the diagnostic procedures for interpretation of CT or PET/CT images, recommend any specific treatment, nor is it intended to replace the skill and judgment of a qualified medical professional.
Performance Testing
Results of software validation and verification (V&V) and early feasibility testing of device component functionality for CT segmentation, regional PET thresholding, and ROI matching supported further FDG test-retest and pivotal device output investigation. For detail on component testing, published reporting is available respectively in Weisman 2023 (CT anatomy segmentation, doi.org/10.1088/2057-1976/acfb06), Perk 2018 (ROI detection methodology, doi.org/10.1088/1361-6560/aaebba), and Huff 2023 (ROI matching, doi.org/10.1088/1361-6560/acef8f).
Test-Retest
This study established test-retest limits of agreement used to categorize regions of interest based off changes between FDG PET/CT images. 31 patients with non-small lung cancer patients received two FDG PET/CT images within 1 week pretreatment. The limits of repeatability endpoint estimated the interval within which 95% of repeated measurements are expected to lie when clinically relevant change is lacking. For TRAQinform IQ generated features, these limits are shown in Table 1.
Feature | Lower limit (%) | Upper limit (%) |
---|---|---|
SUVmax | -27.0 | 56.8 |
SUVmean | -20.2 | 38.5 |
SUVtotal | -54.1 | 144.5 |
Volume | -52.6 | 113.9 |
Table 1: Lower and upper limits of repeatability for each imaging feature.
Pivotal Reader Study
A study was conducted to evaluate the clinical utility of TRAQinform IQ. In this study, three report evaluators (oncologists) were assigned to distinguish each of 103 patients as positive versus
10
neqative for progression based on presentation of the locally generated radiology report without and then with adjunctive presentation of the TRAQinform Report.
Each of the 103 patients had two sequential FDG PET/CT scans acquired between 2005 and 2022. Scans were acquired from patients scanned at 10 or more imaging centers in at least 3 U.S. states (exact image center information was redacted on scan transfer). A full breakdown is shown in Table 2. A breakdown of patients by disease type, patient sex, weight, and age are also shown in Table 2. Out of the 206 total scans (2 per patient), scanner information was available for 199 (99 patients). Scans were acquired on a variety of scanner manufacturers and models, with 84 patients receiving scans on the same scanner for baseline and follow-up.
A total of 1046 ROI were detected across the baseline scans (median per patient: 6, range: 0-80) and 935 were detected on the follow-up (median per patient: 4, range: 0-60).
Table 2: Patient and scan characteristics for all patients and scans where information was not redacted during the scan transfer process.
| Imaging Site /
Location | Site 1 (North Carolina, USA), n = 35
Site 2 (Unknown State, USA) , n = 31 Site 3
(North Carolina, USA), n = 13
Site 4 (North Carolina, USA), n = 12
Site 5 (Iowa, USA), n = 4
Site 6 (North Carolina, USA), n = 3
Site 7 (North Carolina, USA), n = 2
Site 8 (Unknown State, USA), n = 1
Site 9 (North Carolina, USA), n = 1
Site 10 (Missouri, USA), n = 1 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cancer Type, n | Breast Cancer, n = 23
Lung Cancer, n = 20
Prostate, n = 16
Melanoma, n = 15
Head & Neck Cancer, n = 8
Lymphoma, n = 7
Colorectal Cancer, n = 6
Other, n = 5
Gynecological Cancer, n = 3 |
| Patient sex, n
Female / Male | 57/46 |
| Patient age, years
Median (range) | 66 (26, 87) |
| Patient weight, kg
Median (range) | 78.5 (43.0, 132.9) |
| Patient Race | Unreported, n = 75
White, n = 25
Hispanic, n = 2
Black, n = 1 |
11
| Scanner model
(n=199) | Siemens Healthineers Biograph 20, n = 76
Siemens Healthineers BioGraph TruePoint, n = 46
Canon Medical Systems Celesteion, n = 41
Canon Medical Systems Cartesion Prime, n = 14
GE Healthcare Discovery ST, n = 12
Siemens Healthineers Biograph6, n = 7
Siemens Healthineers Biograph Horizon, n = 3 |
-------------------------- | -------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
---|
Of the 103 patients investigated, the TRAQInform Report prompted at least one oncologist interpreter to reevaluate 23 patients classified without the device as negative for progression to positive, and to reevaluate 9 patients classified without the device as positive for progression to negative. In these cases, the oncologist interpreter highlighted a subset of ROI for further review. Imaging and local reporting on these 23 + 9 = 32 patients was sent to a panel of two radiologists and one oncologist, together serving as a reference source against which to quantify positive percent agreement (PPA, representing patients classified as positive by both the panel and an oncologist), negative percent agreement (NPA; representing patients classified as negative by both the panel and an oncologist), and the 95% confidence interval (95%Cl) for overall percent agreement (OPA; prevalence-biased PPA and NPA aggregation). Compared to chance performance (50%), the 95% CI for OPA was 41% to 76%. PPA was 14/18=78% and NPA was 5/14=36%.
Because reevaluation at the patient level was derived from reporting at a lower ROI level, panel agreement was also evaluated at this level. The panel examined only the subset of ROI that had been highlighted by the report evaluators. Of 111 total ROIs highlighted on the TRAQInform Report, the panel agreed with the device classification (new, increasing, unchanged, decreasing, and disappearing) in 37/53=70%, 18/26=70%, 11/13=85%, 5/8 (63%), and 10/11 (90%) ROIs, respectively.
Comparison of Technological Characteristics to Predicate
TRAQinform IQ is a modification of the predicate device. Both the subject and predicate devices are intended to be used by trained medical professionals as an aid in evaluation and information management of digital medical images. Specifically, both devices are radiological image processing systems intended to assess patient radiological scans to provide supplemental information to physicians on changes to regions of interest (ROI).
Comparison of key technological features are provided in Table 3.
Conclusion
The TRAQinform IQ has the same intended use and similar indications, technological characteristics, and principles of operation as the predicate device. Differences do not alter the intended use of the device and do not raise different questions of safety and effectiveness when used as labeled. The performance data demonstrate that the TRAQinform IQ is as safe and effective as the QTxl. Thus, the TRAQinform IQ is substantially equivalent.
12
Table 3: Substantial Equivalence Table
| Specification /
Characteristic | TRAQinform IQ
Subject Device | Quantitative Total Extensible
Imaging (QTxl) (K173444)
Predicate Device |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | LLZ | LLZ |
| Regulation
Number | 892.2050 | 892.2050 |
| Regulatory Class | II | II |
| Review Panel | Radiology | Radiology |
| Users | Trained medical professionals including, but
not limited to, radiologists, oncologists,
nuclear medicine physicians, medical imaging
technologists, dosimetrists and physicists. | Trained medical professionals
including, but not limited to,
radiologists, oncologists, nuclear
medicine physicians, medical imaging
technologists, dosimetrists and
physicists. |
| Intended Use | A software tool to aid in evaluation and
information management of digital medical
images. | A software tool to aid in evaluation
and information management of digita
medical
images. |
| Indications for
Use | TRAQinform IQ is a software only device that
provides quantitative analysis of lesions
identified as Regions of Interest (ROI) in
PET/CT DICOM compliant imaging data
acquired, interpreted, and reported on per local
practice prior to device use.
Clinicians responsible for patient care and for
ordering TRAQinform Reports as an adjunct to
locally reported image interpretation do not
interact directly with the device. Clinicians
responsible for local image interpretation do
not interact with the device and generate their
reporting before and independently of the
TRAQinform Report.
The TRAQinform Report is generated by the
device manufacturer and signed by a U.S.
board certified physician responsible for
supervising central report generation and
qualified to practice nuclear
radiology/medicine. The TRAQinform Report is
for use by trained medical professionals
including but not limited to oncologists, nuclear | Quantitative Total Extensible Imaging
(QTxl) is a software tool used to aid in
evaluation and information management
of digital medical images by trained
medical professionals including, but not
limited to, radiologists, oncologists,
nuclear medicine physicians, medical
imaging technologists, dosimetrists and
physicists. The medical modalities of
these medical images include, but are
not limited to, DICOM CT and PET as
supported by ACR/NEMA DICOM 3.0. |
| Specification /
Characteristic | TRAQinform IQ
Subject Device | Quantitative Total Extensible Imaging
(QTxI) (K173444)
Predicate Device |
| | TRAQinform IQ software contains the
following functionalities:
Automated matching of ROI between
previously performed CT and PET/CT
DICOM 3.0 volumetric medical
images. In order to perform automated
matching of ROI and quantitative
analysis of previously performed CT
and PET/CT DICOM 3.0 volumetric
medical images, the software initially
performs the following functions: Machine learning skeletal and
anatomic structure
segmentation. Threshold-based ROI
identification and contouring. Automated quantitative analysis to
assess previously performed CT and
PET/CT DICOM 3.0 volumetric
medical images, including: change in
total volume and density of each
identified ROI, and change in
Fludeoxyglucose F18 (FDG) tracer
uptake of each identified ROI among
images. Generation of images of the anatomy
combined with spatial and quantitative
information, including computed
classification of quantitative FDG ROI
changes. | QTxI assists in the following indications:
Receive, store, retrieve, display
and process digital medical
images Create, display and print reports
from those images Provide medical professionals
with the ability to display,
register, and fuse medical
images Identify Regions of Interest (ROIs)
and perform ROI contouring
allowing quantitative/statistical
analysis of full or partial body scans Evaluate quantitative change in
ROIs (total or partial body;
individual ROI within individual) with
3D interactive rendering of images
with highlighted ROIs. |
| Specification /
Characteristic | TRAQinform IQ
Subject Device | Quantitative Total Extensible Imaging
(QTxI) (K173444)
Predicate Device |
| | For multi-timepoint quantitative analysis,
recommended use is in adult patients 22
years and older with partial or whole-body
PET/CT acquired following administration
of FDG per approved drug prescribing
information and with the second FDG
administration separated from the first by
a period not to exceed 12 months. | |
| | For single-timepoint quantitative analysis,
recommended use is in adult patients 22
years and older with partial or whole-body
PET/CT following administration of FDG,
a PSMA targeted PET drug, or a SSTR-
targeted PET drug per approved drug
prescribing information. | |
| | Discrepancies between TRAQinform IQ
and local PET/CT reporting have been
investigated and use of TRAQinform IQ
has not been established for binary
patient level progression or non-
progression decisions without
multidisciplinary review. Discrepancies
between TRAQinform IQ and local
PET/CT reporting that could impact
patient care should therefore prompt
consultation with subject matter experts
(for example, in tumor board), with a
patient-centered focus on discrepant
imaging regions and with blinded or
otherwise neutral adjudication regarding
interpretation/classification source. | |
| | TRAQinform IQ is not intended to
diagnose any disease, replace the
diagnostic procedures for interpretation of
CT or PET/CT images, recommend any
specific treatment, nor is it intended to
replace the skill and judgement of a
qualified medical professional. | |
| Specification /
Characteristic | TRAQinform IQ
Subject Device | Quantitative Total Extensible Imaging
(QTxI) (K173444)
Predicate Device |
| Medical Modalities | CT and PET/CT images as supported by
ACR/NEMA DICOM 3.0 | DICOM, CT, MRI, SPECT and PET as
supported by ACR/NEMA DICOM 3.0 |
| Medical Modalities | Radiotracers for multiple timepoints: FDG | Radiotracers: NaF PET |
| Medical Modalities | Radiotracers for single timepoint: FDG,
PSMA-targeted PET drug, or a SSTR-
targeted PET drug | |
| Operating System | Secure Cloud Environment with Web
Browser Interface | Windows 7 or Windows 10 |
| Identification of
ROI | Automatic Threshold based ROI
identification in bone and anatomical
structures | Threshold based ROI identification in bone |
| Importation of
ROI contours | TRAQinform IQ allows the importation of
physician established contours of ROI | No importation functionality |
| Segmentation | Segmentation of bone to identify skeletal
structures and segmentation of
anatomical structures to identify organs.
Segmentation is driven by machine
learning algorithms. | Atlas based segmentation of bone to identify
skeletal structures |
| Quantitative
Analysis | Scan-to-scan registration and ROI
matching allows automatic
quantification of ROIs and assessment
of changes/differences in ROI:
Changes in ROI shapeSingle ROI splitting into multiple ROIMultiple ROI combining into a single ROIROI appearing, disappearing, and re- appearing across images
Quantitative analysis of functional and
anatomical data for CT and PET/CT
scans, including:
Volume of each identified ROI on each image | Scan-to-scan registration and ROI matching
allows automatic quantification of ROIs and
assessment of changes/differences in ROI.
Quantitative/statistical analysis of full or
partial-body scans |
| | Change in volume of each identified ROI among imagesTotal volume of all identified ROI on each image | |
| Specification /
Characteristic | TRAQinform IQ
Subject Device | Quantitative Total Extensible Imaging
(QTxI) (K173444)
Predicate Device |
| | Change in total volume of all identified ROI on each image Heterogeneity of change in volume of each identified ROI For PET scans: Tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of each identified ROI on each image Change in tracer uptake (SUVmax, SUVtotal, SUVmean) of each identified ROI among images Total tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of all identified ROI on each image Change in total tracer uptake (SUVmax, SUVtotal, SUVmean, SUVhetero) of all identified ROI on each image Heterogeneity of change in tracer uptake (SUVhetero) among identified ROI For CT scans: Radio density (HUmax, HUtotal, HUmean, HUhetero) for each identified ROI on each image Changes in radio density (HUmax, HUtotal, HUmean) of all identified ROI on each image Change in total radio density (HUmax, HUtotal, HUmean, HUhetero) of all identified ROI on each image Heterogeneity of change in radio density (HUhetero) among identified ROI 2D Graphical renderings of medical images, including Maximum Intensity Projections of the PET and CT, with overlayed and labeled/color-coded ROI for inclusion in TRAQinform Reports. | |
| | 3D labeled contours for ROI, anatomic structures, and skeletal structures. | |
13
14
15
16