AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Swan-Ganz pacing-TD catheters are indicated for use for temporary, atrial, ventricular, or A-V sequential cardiac pacing of adult surgical and/or critically ill patients. This includes, but is not limited to, patients with arrythmias, acute myocardial infarction, cardiac surgery, and coronary angiography. In addition, these catheters are indicated for use in hemodynamic monitoring of patients experiencing medical conditions such as, but not limited to, post-major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, stroke, and cardiac disease including heart failure.

The Swan-Ganz bipolar pacing catheters are indicated for use in surgical and/or critically ill cardiac patients who may benefit from temporary, transvenous right ventricular pacing.

The Chandler transluminal V-pacing probe is a transluminal bipolar pacing probe, which must be used with a Swan-Ganz Paceport catheter is indicated for use in adult surgical and/or critically ill cardiac patients. This includes, but is not limited to, patients with arrhythmias, acute myocardial infarction, cardiac surgery and coronary angiography.

Device Description

The Swan-Ganz Pacing Catheters and Pacing Probes are a series of specialized catheters and pacing probes, which are temporarily inserted into the central circulatory system vasculature to facilitate temporary atrial, ventricular, or A-V sequential pacing and to serve as diagnostic and/or therapeutic tools. The catheters may be used to overdrive suppression of arrhythmias and diagnosis of complex arrhythmias. The pacing catheters and probes may be used for ECG detection during placement but are not intended to be used for ECG monitoring.

AI/ML Overview

I am sorry, but based on the provided text, there is no information about the acceptance criteria and study proving the device meets those criteria in the context of an AI/ML powered medical device. The document is a 510(k) summary for a set of physical medical devices (catheters and pacing probes) and does not mention any AI or machine learning components.

Therefore, I cannot provide the requested information, including:

  1. A table of acceptance criteria and the reported device performance
  2. Sample size used for the test set and data provenance
  3. Number of experts used to establish ground truth and their qualifications
  4. Adjudication method for the test set
  5. MRMC comparative effectiveness study results or effect size
  6. Standalone (algorithm only) performance
  7. Type of ground truth used
  8. Sample size for the training set
  9. How the ground truth for the training set was established

The document focuses on the substantial equivalence of the Swan-Ganz Pacing Probe and Catheters to predicate devices based on design, material, chemical composition, principle of operation, and compliance with biocompatibility and bench testing standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 27, 2024

Edwards Lifesciences Anna Hwang Manager, Regulatory Affairs One Edwards Way Irvine, California 92614

Re: K233983

Trade/Device Name: Swanz Pacing Probe and Catheters; Swan-Ganz Bipolar Pacing Catheter (D97120F5); Swan-Ganz Bipolar Pacing Catheter (D97130F5); Swan-Ganz Pacing Thermodilution TD Catheter (D200F7); Chandler Transluminal V-Pacing Probe (D98100) Regulation Number: 21 CFR 870.1240 Regulation Name: Flow-directed catheter Regulatory Class: Class II Product Code: DYG, LDF Dated: May 24, 2024 Received: May 28, 2024

Dear Anna Hwang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/7 description: The image shows a digital signature. The signature is for Sara M. Royce. The date of the signature is 2024.06.27, and the time is 14:46:29 -04'00'.

Sara Royce Assistant Director DHT2A: Division of Cardiac

{2}------------------------------------------------

Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K233983

Device Name

Swan-Ganz Pacing Probe and Catheters; Swan-Ganz Bipolar Pacing Catheter (D97120F5); Swan-Ganz Bipolar Pacing Catheter (D97130F5); Swan-Ganz Pacing Thermodilution TD Catheter (D200F7); Chandler Transluminal V-Pacing Probe (D98100)

Indications for Use (Describe)

The Swan-Ganz pacing-TD catheters are indicated for use for temporary, atrial, ventricular, or A-V sequential cardiac pacing of adult surgical and/or critically ill patients. This includes, but is not limited to, patients with arrythmias, acute myocardial infarction, cardiac surgery, and coronary angiography. In addition, these catheters are indicated for use in hemodynamic monitoring of patients experiencing medical conditions such as, but not limited to, post-major surgical recovery, trauma, sepsis, burns, pulmonary disease, pulmonary failure, stroke, and cardiac disease including heart failure.

The Swan-Ganz bipolar pacing catheters are indicated for use in surgical and/or critically ill cardiac patients who may benefit from temporary, transvenous right ventricular pacing.

The Chandler transluminal V-pacing probe is a transluminal bipolar pacing probe, which must be used with a Swan-Ganz Paceport catheter is indicated for use in adult surgical and/or critically ill cardiac patients. This includes, but is not limited to, patients with arrhythmias, acute myocardial infarction, cardiac surgery and coronary angiography.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

Swan-Ganz Pacing Probe and Catheters
510(k) SubmitterEdwards Lifesciences LLCOne Edwards WayIrvine, CA 92614(949) 250-111
Contact PersonAnna HwangManager, Regulatory AffairsEdwards LifesciencesOne Edwards WayIrvine, CA 92614Telephone: (949) 250 - 0902Fax: (949) 809 - 7303Email:Anna_Hwang@edwards.comKaren O'LearySr. Director, Regulatory AffairsEdwards LifesciencesOne Edwards WayIrvine, CA 92614Telephone: (949) 610-9179Fax: (949) 809 - 7214Email:Karen_Oleary@edwards.com
Date PreparedDecember 15, 2023
Trade NameSwan-Ganz Pacing probe and catheters
Regulation Number/21 CFR §870.1240
Regulation Name21 CFR §870.3680
Product CodeDYG - Flow-directed catheterLDF - Cardiovascular permanent or temporary pacemaker electrode
Regulation ClassClass II
Predicate DeviceSwan-Ganz Flow Directed Catheter(K791183, cleared on July 12, 1979)Swan-Ganz Bipolar Pacing Catheter(K822723, cleared on Oct 15, 1982)Chandler Transluminal Bipolar Pacing Probe(K152633, cleared Oct 15, 2015)
Device DescriptionThe Swan-Ganz Pacing Catheters and Pacing Probes are a series ofspecialized catheters and pacing probes, which are temporarily insertedinto the central circulatory system vasculature to facilitate temporaryatrial, ventricular, or A-V sequential pacing and to serve as diagnosticand/or therapeutic tools. The catheters may be used to overdrivesuppression of arrhythmias and diagnosis of complex arrhythmias. Thepacing catheters and probes may be used for ECG detection duringplacement but are not intended to be used for ECG monitoring.
Swan-Ganz Pacing Probe and Catheters
Indications for UseThe Swan-Ganz pacing-TD catheters are indicated for use for temporary,atrial, ventricular, or A-V sequential cardiac pacing of adult surgicaland/or critically ill patients. This includes, but is not limited to, patientswith arrythmias, acute myocardial infarction, cardiac surgery, andcoronary angiography. In addition, these catheters are indicated for usein hemodynamic monitoring of patients experiencing medical conditionssuch as, but not limited to, post-major surgical recovery, trauma, sepsis,burns, pulmonary disease, pulmonary failure, stroke, and cardiac diseaseincluding heart failure.The Swan-Ganz bipolar pacing catheters are indicated for use in surgicaland/or critically ill cardiac patients who may benefit from temporary,transvenous right ventricular pacing.The Chandler transluminal V-pacing probe is a transluminal bipolarpacing probe, which must be used with a Swan-Ganz Paceport catheteris indicated for use in adult surgical and/or critically ill cardiac patients.This includes, but is not limited to, patients with arrhythmias, acutemyocardial infarction, cardiac surgery and coronary angiography.
ComparativeAnalysisThe subject device is identical to the predicate device in terms of thetechnological characteristics including design, material, chemicalcomposition, and principle of operation.The subject devices indications for use were updated for clarificationpurposes, with inclusion of medical and clinical conditions of use, whilealso removing redundancy. The indications for use and changes to thelabeling were updated in compliance with the European Union MedicalDevice Regulations (EU MDR), and in better alignment with 21 CFR841.20 requirements.
Device TestingBiocompatibility testing was performed in accordance with ISO 10993-1: 2018 – Biological evaluation of medical devices – Part 1: Evaluationand testing within a risk management process, and FDA guidancedocument, Use of International Standard ISO 10993-1, "Biologicalevaluation of medical devices – Part 1: Evaluation and testing within arisk management process", issued on September 4, 2020.Bench testing was performed in accordance with Edwards' designrequirements. In addition, shelf-life, packaging, and sterilizationvalidations have been performed.
ConclusionAll testing met the existing predetermined acceptance criteriaThe Swan-Ganz Pacing probe and catheters remain substantiallyequivalent to the predicate devices.

{5}------------------------------------------------

§ 870.1240 Flow-directed catheter.

(a)
Identification. A flow-directed catheter is a device that incorporates a gas-filled balloon to help direct the catheter to the desired position.(b)
Classification. Class II (performance standards).