(135 days)
No
The document does not mention AI, ML, deep learning, or any related terms in the device description, intended use, or performance studies. While it mentions image processing and algorithms (like myNeedle Guide), these are not explicitly stated to be AI/ML-based.
No
Explanation: The device is a diagnostic imaging system (Computed Tomography X-ray system) intended to generate images for aid in diagnosis and treatment, not to directly provide therapy. While the images can be used for "therapeutic interventions," the device itself does not perform or deliver therapeutic action.
Yes
Explanation: The "Intended Use / Indications for Use" section explicitly states that "The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions." This clearly indicates its role in the diagnostic process.
No
The device description explicitly states it is a "Computed Tomography X-ray system" and features "two continuously rotating tube-detector systems". This indicates it is a hardware device with integrated software, not a software-only medical device.
Based on the provided information, this device is not an In Vitro Diagnostic (IVD).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. The tests are performed outside the body (in vitro).
- Device Function: The description clearly states that this device is a computed tomography (CT) system that generates and processes cross-sectional images of patients by computer reconstruction of x-ray transmission data. This is an in vivo imaging technique, meaning it is performed inside the living body.
- Intended Use: The intended use is to provide images as an aid in diagnosis and treatment, as well as for diagnostic and therapeutic interventions. This aligns with the function of an imaging device, not a device that analyzes biological samples.
While the device produces images that are used for diagnosis, the diagnostic process itself is based on the interpretation of these images by trained medical professionals, not on the analysis of biological samples by the device.
N/A
Intended Use / Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Product codes
JAK
Device Description
Siemens intends to market a new software version, SOMARIS/10 syngo CT VB10, for the Dual Source CT system NAEOTOM Alpha based on the SOMARIS/10 platform.
The subject device NAEOTOM Alpha with software version SOMARIS/10 syngo CT VB10 is a Computed Tomography X-ray system which features two continuously rotating tube-detector systems, denominated as A- and B-systems respectively (dual source CT scanner system). The detectors' function is based on photon-counting technology. The NAEOTOM Alpha with SOMARIS/10 syngo CT VB10 produces CT images in DICOM format, which can be used by trained staff for postprocessing applications commercially distributed by Siemens and other vendors as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions. The computer system delivered with the CT scanner is able to run optional post-processing applications.
Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.
The platform software for the NAEOTOM Alpha is syngo CT VB10 (SOMARIS/10 syngo CT VB10). It is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation.
The software platform provides plugin software interfaces that allow for the use of specific commercially available post-processing software algorithms in an unmodified form from the cleared stand-alone post-processing version.
Software version syngo CT VB10 (SOMARIS/10 syngo CT VB10) is a modified software version of the primary predicate device, syngo CT VA50 (SOMARIS/10 syngo CT VA50) cleared in K220814.
Software version SOMARIS/10 syngo CT VB10 will be offered ex-factory and as optional upgrade for the existing NAEOTOM Alpha systems.
The subject device NAEOTOM Alpha will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use and the indications for use remain unchanged compared to the predicate devices.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Computed Tomography X-ray
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Trained staff (physicians, radiologists, or technologists) in a Professional Healthcare Facility.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Non-clinical testing, (integration and functional) including phantom tests were conducted for the NAEOTOM Alpha during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.
The general purpose of each test is to verify and validate the functionality of the subject device modifications.
Testing will cover all related subsystems that contribute to the device modifications. Test levels are defined. For each test level several test activities are performed. The test specification and acceptance criteria are related to the corresponding requirements. Various test activities are performed to specific modifications on different test levels to ensure safe and effective integration in the system. Three test levels are defined:
System Validation test:
- Acceptance test (workflow and user manual test)
- Legal and Regulatory test
System Verification test:
- System Integration Test (functional)
- Functionality verification
- Image Quality (IQ) Evaluation
Tests are conducted for all software components developed in product development and for the complete product itself. Several activities are considered for this process, including creation of test specifications that relate to software/hardware requirements including tests to address risk mitigations that are identified, documented, and traced by hazard keys.
Additional evaluation tests are performed as bench tests to support the new device or device modification on Non-Clinical Performance Testing as listed in Table 6 below.
Table 6: Non-clinical performance testing (bench testing).
FAST 3D Camera / FAST Integrated Workflow Bench Testing performed: The bench test evaluates and compares the accuracy of the three subfeatures FAST Isocentering, FAST Range, and FAST Direction to the accuracy of the predicate device with syngo CT VA50 using the old camera hardware and the then only available ceiling mount. The objectives of the bench test are to demonstrate that the FAST 3D Camera feature of the subject device with SOMARIS/10 syngo CT VB10, where the algorithms have been optimized for a new camera hardware in two mounting positions, achieves comparable results as the predicate device with syngo CT VA50.
Multi-Purpose Table Bench Testing performed: The range of possible movement for the mobile C-arm in different positions between CT gantry and patient table was tested and documented by measurement of angles.
ZeeFree Bench Testing performed: The bench tests evaluate the performance of ZeeFree reconstruction. The objectives of the tests are to demonstrate:
- that the number of artefacts which can be attributed to a stack misalignment (e.g. discontinuities in vessel structures, anatomical steps at air-soft-tissue interfaces, doubling of vessel or other anatomy) and which are often caused by incomplete patient breath-hold can be reduced in a "Cardiac Stack Artefact Correction" (SAC) reconstruction compared to the standard reconstruction with otherwise matching reconstruction parameters.
- that no artefacts are introduced by a SAC reconstruction.
myNeedle Guide (with myNeedle Detection) Bench Testing performed: Tests were performed to ensure clinical usability of the myNeedle Guide needle detection algorithm. Two individual tests were performed. The accuracy of the automatic needle detection algorithm was tested. The reduction of necessary user interactions for navigating to a needle-oriented view with and without the support of the automatic needle detection algorithm was analyzed.
Quantum Spectral Imaging Bench Testing performed: Tests were performed to demonstrate that:
T3D reconstructions in Quantumpeak mode are possible with the sharpest available kernels up to Qr89, Br98, etc. Quantumpeak scan mode allows reconstructions of monoenergetic images at energy levels from 40 to 190 keV. The evaluation has been performed based on phantom studies.
Quantum HD Cardiac Bench Testing performed: Quantitative assessment in terms of image noise and a visual assessment of the image quality in ECG gated acquisitions between different scan modes is performed: standard acquisition mode with spectral and nonspectral reconstruction, ultra-high resolution acquisition mode with non-spectral reconstruction (120x0.2 mm and 96x0.2 mm) and spectral image reconstruction (limited to 96x0.2 mm).
HD FoV Bench Testing performed: Tests were performed to evaluate the performance of HD FoV on NAEOTOM Alpha. For this purpose, the HU accuracy in the extended field of view region was measured based on phantom studies. The phantom diameter accuracy has been also evaluated.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
FAST 3D Camera / FAST Integrated Workflow:
The FAST Isocentering accuracy of the subject device with syngo CT VB10 is comparable to the predicate device with syngo CT VA50, regardless of the camera mounting position. For the FAST Range feature, the detection accuracy of all body region boundaries is comparable between the subject device with syngo CT VB10 and predicate device with syngo CT VA50. The FAST Direction pose detection results are of comparable accuracy for subject and predicate device, regardless of the camera mounting position. Overall, the SOMARIS/10 syngo CT VB10 delivers comparable accuracy to the SOMARIS/10 syngo CT VA50 predicate for the new FAST 3D Camera hardware.
Multi-Purpose Table:
Based on the test results it can be concluded that a CT scanner, equipped with a Multi-Purpose (Vitus) Patient Table, which is installed with enhanced distance (674 mm) to the CT gantry and offers the iCT mode functionality, provides sufficient freedom of movement for a mobile C-arm X-ray system to be used for clinical routine without any significant limitations for my needle Laser or 3D Camera.
ZeeFree:
The test results show: If misalignment artefacts are identified in non-corrected standard ECG-gated reconstructed sequence or spiral images, the feature "Cardiac Stack Artefact Correction" (SAC, marketing name: ZeeFree) enables optional stack artefact corrected images, which reduce the number of alignment artefacts. The SAC reconstruction does not introduce new artefacts, which were previously not present in the non-corrected standard reconstruction. The SAC reconstruction does realize equivalent image quality in quantitative standard physics phantom-based measurements (ACR, Gammex phantom) in terms of noise, homogeneity, high-contrast resolution, slice thickness and CNR compared to a non-corrected standard reconstruction. The SAC reconstruction does realize equivalent image quality in quantitative and qualitative phantom-based measurements with respect to metal objects compared to a non-corrected standard reconstruction. The SAC algorithm can be successfully applied to phantom data if derived from a suitable motion phantom demonstrating its correct technical function on the tested device. The SAC algorithm is independent from the physical detector width of the acquired data.
myNeedle Guide (with myNeedle Detection):
It has been shown that the algorithm was able to consistently detect needle-tips over a wide variety of scans in 90.76% of cases. Further, the results of this bench test clearly shows that the auto needle detection functionality reduces the number of interactions steps needed to generate a needle-aligned view in the CT Intervention SW. Zero user interactions are required and a needle-aligned view is displayed right away after a new scan, if auto needle detection is switched on in the SW configuration. Therefore, the test is already passed if only a single user interaction is necessary to achieve a needle-oriented view in the manual workflow.
Quantum Spectral Imaging:
The results showed that: with T3D reconstructions from Quantumpeak scan modes, high resolution images with sharp kernel up to Br98 are obtained. The resolution is comparable to other Highresultra scan modes of the NAEOTOM Alpha. Monoenergetic reconstructions from Quantumpeak scan modes are free of artifacts. Measured CT values precisely match the reference values. The accuracy of monoenergetic reconstructions in iodine and calcium inserts at the NAEOTOM Alpha is comparable or better than on the secondary predicate device SOMATOM Force.
Quantum HD Cardiac:
Based on the results it can be concluded that substantial equivalence in image quality is achieved by the images derived from the spectral capable cardiac acquisition mode 96x0.2mm for both, the highresolution UHR and the standard resolution spectral image cases, compared to the single source spectral capable 120x0.2mm UHR scan mode.
HD FoV:
Based on the results it can be concluded that HD FoV enables the reconstruction of images while significantly improving the visualization of anatomy in the regions outside the scan field of view of 50 cm. In the phantom study, an HU value accuracy of about +/- 40 HU was achieved with skin-line accuracy of about +/- 3 mm.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
myNeedle Guide (with myNeedle Detection): accuracy of 90.76% for detecting needle-tips.
HD FoV: HU value accuracy of about +/- 40 HU, skin-line accuracy of about +/- 3 mm.
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1750 Computed tomography x-ray system.
(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 28, 2024
Siemens Medical Solutions USA, Inc. % Tabitha Estes Regulatory Affairs Proffecional 810 Innovation Drive KNOXVILLE, TN 37932
Re: K233657
Trade/Device Name: NAEOTOM Alpha Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: February 22, 2024 Received: February 23, 2024
Dear Tabitha Estes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
1
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
2
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
Device Name
NAEOTOM Alpha
Indications for Use (Describe)
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a graphic of orange dots arranged in a circular pattern.
510(k) Summary for
NAEOTOM Alpha CT Scanner System
with software version SOMARIS/10 syngo CT VB10
Date Prepared: March 22, 2024
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
l. Submitter
Siemens Medical Solutions USA, Inc. 810 Innovation Drive Knoxville, TN 37932 Establishment Registration Number: 1034973
Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number: 2240869
Location of Manufacturing Site Siemens Healthcare GmbH Siemensstr. 1 -OR- Rittigfeld 1 D-91301 Forchheim, Germany Establishment Registration Number: 3004977335
Note: Descriptions in this submission use the short company name Siemens. Brand name on all products is Siemens Healthineers.
Submitter Contact Person: Tabitha Estes Regulatory Affairs Siemens Medical Solutions USA, Inc. (865) 804-4553 (work cell) tabitha.estes@Siemens-healthineers.com
4
Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are several orange dots arranged in a circular pattern.
II. Device Name and Classification
Product Name: | NAEOTOM Alpha |
---|---|
Trade Name: | NAEOTOM Alpha |
Classification Name: | Computed Tomography X-ray System |
Classification Panel: | Radiology |
Regulation Number: | 21 CFR §892.1750 |
Device Class: | Class II |
Product Code: | JAK |
III. Predicate Device
Primary Predicate Device: | |
---|---|
Trade Name: | NAEOTOM Alpha, Scan&GO |
510(k) Number: | K220814 |
Clearance Date: | July 12, 2022 |
Classification Name: | Computed Tomography X-ray System |
Classification Panel: | Radiology |
Regulation Number: | 21 CFR §892.1750 |
Device Class: | Class II |
Product Code: | JAK |
Recall Information: | All predicate device recalls have been considered in the subject |
device design. | |
Secondary Predicate Device: | |
Trade Name: | SOMATOM Force |
510(k) Number: | K230421 |
Clearance Date: | June 16, 2023 |
Classification Name: | Computed Tomography X-ray System |
Classification Panel: | Radiology |
Regulation Number: | 21 CFR §892.1750 |
Device Class: | Class II |
Product Code: | JAK |
Recall Information: | All predicate device recalls have been considered in the subject |
device design. |
Note: K230421 was a bundle submission with various Siemens CT Scanner Systems, including the dual source CT scanner systems SOMATOM Drive, SOMATOM Force, SOMATOM Definition Flash and the single source CT scanner CT systems SOMATOM Edge Plus, SOMATOM Confidence, SOMATOM Definition Edge, SOMATOM Definition AS/AS+, SOMATOM Definition AS Open.
5
Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are a series of orange dots arranged in a circular pattern.
IV. Device Description
Siemens intends to market a new software version, SOMARIS/10 syngo CT VB10, for the Dual Source CT system NAEOTOM Alpha based on the SOMARIS/10 platform.
The subject device NAEOTOM Alpha with software version SOMARIS/10 syngo CT VB10 is a Computed Tomography X-ray system which features two continuously rotating tube-detector systems, denominated as A- and B-systems respectively (dual source CT scanner system). The detectors' function is based on photon-counting technology. The NAEOTOM Alpha with SOMARIS/10 syngo CT VB10 produces CT images in DICOM format, which can be used by trained staff for postprocessing applications commercially distributed by Siemens and other vendors as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions. The computer system delivered with the CT scanner is able to run optional post-processing applications.
Only trained and qualified users, certified in accordance with country-specific regulations, are authorized to operate the system. For example, physicians, radiologists, or technologists. The user must have the necessary U.S. qualifications in order to diagnose or treat the patient with the use of the images delivered by the system.
The platform software for the NAEOTOM Alpha is syngo CT VB10 (SOMARIS/10 syngo CT VB10). It is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation.
The software platform provides plugin software interfaces that allow for the use of specific commercially available post-processing software algorithms in an unmodified form from the cleared stand-alone post-processing version.
Software version syngo CT VB10 (SOMARIS/10 syngo CT VB10) is a modified software version of the primary predicate device, syngo CT VA50 (SOMARIS/10 syngo CT VA50) cleared in K220814.
Software version SOMARIS/10 syngo CT VB10 will be offered ex-factory and as optional upgrade for the existing NAEOTOM Alpha systems.
The subject device NAEOTOM Alpha will support previously cleared software and hardware features in addition to the applicable modifications as described within this submission. The intended use and the indications for use remain unchanged compared to the predicate devices.
V. Indications for Use
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
6
Image /page/6/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
VI. Indications for Use Comparison
Subject Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Primary Predicate Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations*.
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Secondary Device Indications for Use:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment and radiation therapy planning as well as for diagnostic and therapeutic interventions.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Comparison:
-
The subject device Indications for Use is the exact same as the primary predicate's Indications for Use (K220814), however it is slightly different than the secondary predicate's Indications for Use.
-
Compared to the secondary predicate device, the subject device does not contain the phrase "radiation therapy planning" in the sentence "The images delivered by the system can be used by a trained staff as an aid in diagnosis and treatment as well as for diagnostic and therapeutic interventions." because NAEOTOM Alpha does not support radiation therapy planning.
7
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
None of the intended uses include computed tomography as the principal means of guidance in invasive procedures (involving the introduction of a device, such as a needle or a catheter into the body of the patient).
The NAEOTOM Alpha is not the principal means of guidance, because the CT System does not guide the invasive procedures, the needle orientation and the needle advance, and handling is always done under the physicians control.
VII. Comparison of Technological Characteristics with the Predicate Device
Supported by the subject device, SOMARIS/10 syngo CT VB10 software version is a further development of the SOMARIS/10 syngo CT VA50 software version which is cleared in K220814.
The NAEOTOM Alpha with SOMARIS/10 syngo CT VB10 software version provides the same technological characteristics in terms of materials, energy source, and control mechanisms when compared to the predicate devices. The software features of NAEOTOM Alpha have been modified or improved in comparison to the predicate devices to support enhanced device functionality compared to the predicate devices.
The new syngo CT VB10 software reuses all unmodified software features of the legacy software syngo CT VA50 cleared in K220814. Additionally, no features present in the predicate device are descoped.
Software version SOMARIS/10 syngo CT VB10 is designed to reuse hardware independent extended functionalities and GO technologies provided by Siemens cleared software applications.
The intended use and fundamental scientific technology for the NAEOTOM Alpha remain unchanged from the predicate devices.
At a high level, the subject and predicate devices are based on the same subset of technological elements:
- . Scanner Principle – Whole body X-Ray Computed Tomography Scanner
- System Acquisition Continuously rotating tube detector system ●
- Iterative Reconstruction – Support of various iterative reconstruction principles
- Workplaces – Support of workplaces that include reconstruction and image evaluation software
- Patient table
- Patient table foot switch for movement
- Tin filtration technology
- Vectron X-ray Tube
- Power Generator
- Scan&GO
- Mobile workflow (Tablet) ●
- Optional injector arm
- Optional support of CT guided intervention workflow (myNeedle Guide)
- Optional support of FAST 3D Camera operation for fast patient positioning workflow
- Scanner display and control functionality
- Remote Scan Control
- Long scan range
8
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots.
The subject device NAEOTOM Alpha with SOMARIS/10 syngo CT VB10 will support the modifications/further developments in comparison to the predicate devices as listed in the tables under subsections 1) Modified Hardware and 2) Modified Software below.
The configuration table and comparison table use the following terms to describe various technological characteristics in comparison to the primary and secondary predicate devices information:
Term | Definition |
---|---|
New | The feature is newly supported for Siemens CT Scanners and the subject device |
Modified | This feature is a modified form of a feature cleared within the predicate devices |
Enabled | This feature is currently supported by other cleared Siemens CT systems or |
cleared Siemens stand-alone software applications. |
1) Modified Hardware
Table 2: Overview of hardware modifications of NAEOTOM Alpha supported by software version SOMARIS/10 syngo CT VB10 compared to the predicate/secondary devices.
| | Hardware properties | NAEOTOM Alpha
SOMARIS/10 syngo CT VB10
(subject device) |
|----|-----------------------------|---------------------------------------------------------------|
| 1. | Tin Filtration | modified |
| 2. | FAST 3D Camera | modified |
| 3. | Multi-Purpose Table (Vitus) | modified |
2) Modified Software (syngo CT VB10)
Table 3: Overview of software modifications of NAEOTOM Alpha supported by software version SOMARIS/10 syngo CT VB10 compared to the predicate/secondary devices.
| | Software properties | NAEOTOM Alpha
SOMARIS/10 syngo CT VB10
(subject device) |
|----|-----------------------------------------------------|---------------------------------------------------------------|
| 1. | ZeeFree | new |
| 2. | FAST Integrated Workflow | modified |
| 3. | myNeedle Guide (with MyNeedle Detection) | modified |
| 4. | myExam Companion – myExam Cockpit/myExam
Compass | modified |
| 5. | Recon&GO | modified |
9
Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
| | Software properties | NAEOTOM Alpha
SOMARIS/10 syngo CT VB10
(subject device) |
|-----|---------------------------------|---------------------------------------------------------------|
| 6. | CT View&GO | modified |
| 7. | Scan&GO | enabled |
| 8. | Quantum Spectral Imaging | modified |
| 9. | Quantum HD Cardiac | modified |
| 10. | HD FoV | enabled |
| 11. | Multi-Threshold Acquisition | modified |
| 12. | myExam Satellite (Remote Recon) | modified |
A tabular summary of the comparable hardware and software properties between the subject device NAEOTOM Alpha with software version syngo CT VB10 and primary/secondary predicate device are listed in Table 4 and Table 5 below (modifications are in gray shaded sections).
Table 4: Technical hardware characteristics for the subject device NAEOTOM Alpha (software version SOMARIS/10 syngo CT | |
---|---|
VB10) compared to the predicate devices. |
| Hardware
property | Subject device | Primary predicate
device | Secondary predicate
device |
|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | NAEOTOM Alpha
SOMARIS/10 syngo CT
VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT
VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT VB30
(K230421) |
| Scanner | whole body X-ray
computed
tomography scanner | whole body X-ray
computed
tomography scanner | whole body X-ray
computed tomography
scanner |
| System
configuration | Dual Source | Dual Source | Dual Source |
| Environment of
Use | Professional
Healthcare Facility | Professional
Healthcare Facility | Professional Healthcare
Facility |
| Generator
max. power (kW) | 2x 120 | 2x 120 | 2x 120 |
| Detector
technology | QuantaMax | QuantaMax | UFC
(Ultra Fast Ceramic) |
| Hardware
property | Subject device | Primary predicate
device | Secondary predicate
device |
| | NAEOTOM Alpha | NAEOTOM Alpha | SOMATOM Force |
| | SOMARIS/10 syngo CT
VB10 | SOMARIS/10 syngo CT
VA50
(K220814) | SOMARIS/7 syngo CT VB30
(K230421) |
| | Direct Conversion
with "Quantum
Technology" | Direct Conversion
with "Quantum
Technology" | |
| Detector
volume coverage
(mm) | 2x 57.6 | 2x 57.6 | 2x 57.6 |
| Detector
physical rows | 2x 288 | 2x 288 | 2x 96 |
| Detector
slice width (mm) | 0.2 | 0.2 | 0.6
(optional: 0.4, 0.5) |
| Detector | 2752 (A system) | 2752 (A system) | 920 (A system) |
| DAS channel no. | 1984 (B system) | 1984 (B system) | 640 (B system) |
| Tube
technology | VECTRON | VECTRON | VECTRON |
| Tube
kV steps | 70, 90, 100, 120, 140,
150
(150 kV only available
on the smaller tube-
detector system (B
system) and only in
combination with the
additional Sn filter, 0.7
mm) | 70, 90, 100, 120, 140 | 70, 80, 90, 100, 110, 120,
130, 140, 150 |
| Tube
max. current
(mA) | 2x 1300 | 2x 1300 | 2x 1300 |
| Tube | 0.4 x 0.5/8° | 0.4 x 0.5/8° | 0.4 x 0.5/8° |
| tube focus (mm) | 0.6 x 0.7/8° | 0.6 x 0.7/8° | 0.6 x 0.7/8° |
| | 0.8 x 1.1/8° | 0.8 x 1.1/8° | 0.8 x 1.1/8° |
| Hardware
property | Subject device | Primary predicate
device | Secondary predicate
device |
| | NAEOTOM Alpha | NAEOTOM Alpha | SOMATOM Force |
| | SOMARIS/10 syngo CT
VB10 | SOMARIS/10 syngo CT
VA50
(K220814) | SOMARIS/7 syngo CT VB30
(K230421) |
| | (for both tubes) | (for both tubes) | (for both tubes) |
| Tube
heat capacity | higher than 30 MHU | higher than 30 MHU | Higher than 30 MHU |
| Gantry
bore size (cm) | 82 | 82 | 78 |
| Gantry
Scan FoV (cm) | 50 | 50 | 50 |
| Gantry
rotation time
(sec) | 0.25, 0.5, 1.0 | 0.25, 0.5, 1.0 | 0.25*, 0.285, 0.325*, 0.5,
1.0
(*optional) |
| Gantry
Tilt (degree) | N/A | N/A | N/A |
| Maximum
temporal
resolution in ECG
gated or
triggered
examination
(ms) | mono-segment: 66
bi-segment: 33 | mono-segment: 66
bi-segment: 33 | mono-segment, standard:
75
mono-segment, optional:
66
bi-segment, standard: 38
bi-segment, optional: 33 |
| Maximum scan
speed at pitch
(mm/s at pitch x) | 737 mm/s at pitch 3.2 | 737 mm/s at pitch 3.2 | 737 mm/s at pitch 3.2 |
| Patient Table
Type | Vario 2.D
Vitus | Vario 2.D
Vitus | PHS5
MPT4 |
| Max. Scan length
Topogram (mm) | Vario 2.D: 2080
Vitus: 2080 | Vario 2.D: 2080
Vitus: 2080 | PHS5: 1970
MPT4: 1970 |
| Max. Scan length | Vario 2.D: 2000 | Vario 2.D: 2000 | PHS5: 1953 |
| Hardware
property | Subject device | Primary predicate
device | Secondary predicate
device |
| | NAEOTOM Alpha
SOMARIS/10 syngo CT
VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT
VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT VB30
(K230421) |
| Image acquisition
(mm) | Vitus: 2000 | Vitus: 2000 | MPT4: 1953 |
| Patient table | Vario 2.D: 307 | Vario 2.D: 307 | PHS5: 227 |
| Max. weight
capacity (kg) | Vitus: 307 or 340 | Vitus: 307 | MPT4: 227 or 307 (with
bariatric/trauma table top) |
| Patient table
Installation
option | Regular installation
(Vario 2.D and Vitus):
474 mm | Regular installation
(Vario 2.D and Vitus):
474 mm | Regular installation (PHS5
and MPT4):
400 mm |
| | Installation option
with extended
distance (Vitus):
674 mm (474 mm +
200 mm) | | |
| Spectral
filtration | Tin Filter for both
tubes:
0.4 mm
additional Tin Filter
for the smaller tube-
detector system (B
system) only:
0.7 mm | Tin Filter for both
tubes:
0.4 mm | Tin Filter for both tubes:
0.6 mm |
| FAST 3D Camera
for patient
positioning | option for patient
positioning with 3D
Camera
ceiling mounted,
modified design | option for patient
positioning with 3D
Camera
ceiling mounted | option for patient
positioning with 3D Camera
ceiling mounted |
| X-ray foot switch | Option to trigger
hands-free scanning | Option to trigger
hands-free scanning | Option to trigger hands-
free scanning |
| Hardware
property | Subject device | Primary predicate
device | Secondary predicate
device |
| | NAEOTOM Alpha
SOMARIS/10 syngo CT
VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT
VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT VB30
(K230421) |
| Table foot switch | Option for table
patient movement | Option for table
patient movement | N/A |
| Tablet dock for
patient table | Option for mounting
of the tablet on the
patient table. | Option for mounting
of the tablet on the
patient table. | N/A |
| Interventional
Joystick (IVJ) | Option to move the
table during
myNeedle Guide
procedures in the
examination room.
electrical connection
for the tablet dock
which allows charging
the tablet when
mounted. | Option to move the
table during
myNeedle Guide
procedures in the
examination room.
electrical connection
for the tablet dock
which allows charging
the tablet when
mounted. | i-control:
option to operate some of
the CT functions including
patient table movement as
an alternative to the gantry
operating panel and the
input units at the console. |
| Laser supported
workflow | Laser in combination
with FAST
Isocentering visualize
coordinates for
patient isocenter
position;
myNeedle Laser
visualizes a planned
needle path for
interventions | Laser in combination
with FAST
lsocentering visualize
coordinates for
patient isocenter
position;
myNeedle Laser
visualizes a planned
needle path for
interventions | Laser in combination with
FAST Isocentering visualize
coordinates for patient
isocenter position; |
| Software property | Subject device | Primary predicate
device | Secondary predicate
device |
| | NAEOTOM Alpha
SOMARIS/10 syngo CT
VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT
VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT
VB30
(K230421) |
| Operating System | Windows based
SOMARIS/10 syngo CT
VB10
Note: the short version
syngo CT VB10 is also
used as labeling
information | Windows based
SOMARIS/10 syngo CT
VA50
Note: the short version
syngo CT VA50 is also
used as labeling
information | Windows based
SOMARIS/7 syngo CT
VB30
Note: the short version
syngo CT VB30 is also
used as labeling
information |
| Workplace | syngo Acquisition
Workplace (ICS) named
as "myExam Console"
Image Reconstruction
for Quantum
Technology (IRS)
2nd workplace option
named as "myExam
Satellite" with Remote
Recon function | syngo Acquisition
Workplace (ICS) named
as "myExam Console"
Image Reconstruction
for Quantum
Technology (IRS)
2nd workplace option
named as "myExam
Satellite" | syngo Acquisition
Workplace (AWP)
Optional second
operating system
(Remote Recon
Workplace, RRWP) |
| Standard system
software | • syngo Examination
• syngo Viewing
• syngo Filming
• syngo Archiving &
Network | • syngo Examination
• syngo Viewing
• syngo Filming
• syngo Archiving &
Network | • syngo Examination
• syngo Viewing
• syngo Filming
• syngo Archiving &
Network |
| Detector
firmware | QuantaMax detector
firmware supported | QuantaMax detector
firmware supported | Stellar detector
firmware supported |
| Teamplay | Support teamplay
Protocols | Support teamplay
Protocols | Support teamplay
Protocols |
| Protocols | Support of:
• Protocol supporting
contrast bolus-
triggered data
acquisition | Support of:
• Protocol supporting
contrast bolus-
triggered data
acquisition | Support of:
• Protocols
supporting contrast
bolus-triggered data
acquisition |
| Software property | Subject device | Primary predicate device | Secondary predicate device |
| | NAEOTOM Alpha
SOMARIS/10 syngo CT
VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT
VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT
VB30
(K230421) |
| | Contrast media protocols (including coronary CTA) Pediatric Protocols Flex Dose Profile Turbo Flash Spiral Dual Energy acquisition Dynamic imaging (Flex 4D Spiral) Protocols supporting CT Intervention, Cardiac Scanning, Spectral imaging for child examination, Spectral imaging with high resolution Protocols for Quantum Imaging modes: Quantum Quantumplus Quantum HD (previously: Quantum High resolution) Quantum HD Cardiac (previously High resolution Dual Source Cardiac modes). In addition, a spectral image results from dual source 96x0.2mm ultra high-resolution mode can optionally be obtained | Contrast media protocols (including coronary CTA) Pediatric Protocols Flex Dose Profile Turbo Flash Spiral Dual Energy acquisition Dynamic imaging (Flex 4D Spiral) Protocols supporting CT Intervention, Cardiac Scanning, Spectral imaging for child examination, Spectral imaging with high resolution Protocols for Quantum Imaging modes: Quantum Quantumplus Quantum High resolution High resolution Dual Source Cardiac modes | Contrast media protocols (including coronary CTA) Pediatric Protocols Turbo Flash Spiral Dual Source Dual Energy protocols Adaptive 4D Spiral Protocols for Radiation Therapy Planning Dual Source Dual Energy protocols for Radiation Therapy Planning Protocols for CT intervention, Cardiac Scanning |
| Software property | Subject device | Primary predicate device | Secondary predicate device |
| | NAEOTOM Alpha | NAEOTOM Alpha | SOMATOM Force |
| | SOMARIS/10 syngo CT VB10 | SOMARIS/10 syngo CT VA50 | SOMARIS/7 syngo CT VB30 |
| | | (K220814) | (K230421) |
| | - Quantumpeak
(Quantumpeak mode is
identical to the Dual Source
Dual Energy modes at
SOMATOM Dual Source Scanners
in K230421 | | |
| Advanced Reconstruction | Recon&GO:
- Spectral Recon (Dual
Energy Reconstruction
from photon-counting
data) / including Virtual
Unenhanced,
Monoenergetic plus - Inline Results DE SPP
(Spectral Post-
Processing with photon-
counting image data) - Inline Anatomical
ranges (Parallel/Radial)
incl. Virtual
Unenhanced,
Monoenergetic plus
(already cleared with
the stand-alone medical
device syngo.via
(K191040) - Inline Spine and Rib
Ranges (already cleared
with the stand-alone
medical device syngo.CT
Applications (syngo.CT
Bone Reading)
K220450)) | Recon&GO: - Spectral Recon (Dual
Energy Reconstruction
from photon-counting
data) / including Virtual
Unenhanced,
Monoenergetic plus - Inline Results DE SPP
(Spectral Post-
Processing with
photon-counting image
data) | Advanced
reconstruction tools
supported:
The syngo acquisition
workplace provides
image reconstruction
and routine
postprocessing. |
| Software property | Subject device | Primary predicate
device | Secondary predicate
device |
| | NAEOTOM Alpha | NAEOTOM Alpha | SOMATOM Force |
| | SOMARIS/10 syngo CT
VB10 | SOMARIS/10 syngo CT
VA50 | SOMARIS/7 syngo CT
VB30 |
| | | (K220814) | (K230421) |
| | - Inline table and bone
removal (already
cleared with the stand-alone medical device
syngo.CT Extended
Functionality (K221727)) | | |
| Image viewing | CT View&GO offers: | CT View&GO offers: | syngo Viewing offers: |
| | - basic post-processing
viewer (CT View&GO) | - basic post-processing
viewer (CT View&GO) | - 2D and 3D (MPR, VRT,
MIP and minIP) |
| | - 2D and 3D (MPR, VRT,
MIP and minIP) | - 2D and 3D (MPR, VRT,
MIP and minIP) | - Evaluation tools,
Filming, Printing |
| | - Evaluation tools,
Filming, Printing | - Evaluation tools,
Filming, Printing | |
| | - Interactive Spectral
Imaging (ISI) | | |
| | - Basic visualization
tools: Endoscopic View | | |
| | - Basic manipulation
tools: DE ROI, ROI HU
Threshold, Average | | |
| | - Automated table and
bone removal
(already cleared with
the stand-alone medical
device syngo.CT
Extended Functionality
(K221727)) | | |
| Post-Processing
interface | • Recon&GO Inline
Results:
Software interface to
post-processing
algorithms which are
unmodified when
loaded onto the CT | • Recon&GO Inline
Results:
Software interface to
post-processing
algorithms which are
unmodified when
loaded onto the CT | syngo.via - Wide Range
of individual
applications, syngo.via
is a software solution
intended to be used for
viewing, manipulation,
communication, and |
| Software property | Subject device | Primary predicate
device | Secondary predicate
device |
| | NAEOTOM Alpha | NAEOTOM Alpha | SOMATOM Force |
| | SOMARIS/10 syngo CT
VB10 | SOMARIS/10 syngo CT
VA50 | SOMARIS/7 syngo CT
VB30 |
| | | (K220814) | (K230421) |
| | cleared as medical
devices in their own
right.
• software interfaces
for post-processing
functionalities to
provide advanced
visualization tools to
prepare and process
medical images for
diagnostic purpose.
Note: The clearance of
standalone Advanced
Visualization Application
software is mandatory
precondition.
These advanced
visualization tools are
designed to support the
technician & physician
in the qualitative and
quantitative
measurement & analysis
of clinical data acquired
and reconstructed by
Computed Tomography
scanners. Additional
information regarding
the points of interface
and inputs for this
feature is provided in
Section 16. | cleared as medical
devices in their own
right.
• software interfaces
for post-processing
functionalities to
provide advanced
visualization tools to
prepare and process
medical images for
diagnostic purpose.
Note: The clearance of
standalone Advanced
Visualization
Application software is
mandatory
precondition.
These advanced
visualization tools are
designed to support
the technician &
physician in the
qualitative and
quantitative
measurement &
analysis of clinical data
acquired and
reconstructed by
Computed Tomography
scanners. Additional
information regarding
the points of interface
and inputs for this
feature is provided in
Section 16. | images. It can be used
as a standalone device
or together with a
variety of cleared and
unmodified syngo
based software
options. |
| Cybersecurity | IT Hardening | IT Hardening | IT Hardening |
| HD FoV | supported | N/A | supported |
| Software property | Subject device | Primary predicate device | Secondary predicate device |
| | NAEOTOM Alpha
SOMARIS/10 syngo CT VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT VB30
(K230421) |
| Standard
technologies | FAST technologies CARE technologies GO technologies CARE keV | FAST technologies CARE technologies GO technologies CARE keV | FAST technologies CARE technologies |
| myExam
Companion –
myExam
Compass/myExam
Cockpit | myExam Compass collects information about the current patient to dynamically adapt the scan parameters or exchange recon jobs according to the patient's characteristics myExam Cockpit option of displaying, modifying, creating, and deleting Clinical Decision Trees (CDTs). myExam Compass functionality offers the possibility to activate/deactivate diagnostic scan ranges, Bolus Tracking and Test Bolus ranges. myExam Cockpit allows to define these new settings. | myExam Compass collects information about the current patient to dynamically adapt the scan parameters or exchange recon jobs according to the patient's characteristics myExam Cockpit option of displaying, modifying, creating, and deleting Clinical Decision Trees (CDTs). | N/A |
| Scan&GO | Scan&GO
With software version syngo CT VB10, the previously stand-alone Scan&GO software functionality is fully incorporated into the subject device CT scanner system. The | Scan&GO | N/A |
| Software property | Subject device | Primary predicate device | Secondary predicate device |
| | NAEOTOM Alpha
SOMARIS/10 syngo CT VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT VB30
(K230421) |
| | functionality remains unchanged. | | |
| Reconstruction Options for Cardiac Imaging | • Standard (renamed from “TrueStack ‘off’” on the primary predicate device)
• TrueStack (renamed from “TrueStack ‘on’” on the primary predicate device)
• ZeeFree allows the reconstruction of ECG-gated spiral or ECG-triggered sequence data in a cardiac cycle-to-cycle border aligned fashion | • TrueStack “off”
• TrueStack “on” | • TrueStack “off”
• TrueStack “on” |
| Iterative Reconstruction Methods | Quantum Iterative Reconstruction
iMAR | Quantum Iterative Reconstruction
iMAR | ADMIRE
SAFIRE
iMAR |
| Precision Matrix | Precision Matrix resolution
support image matrix sizes of:
512 x 512 pixels
768 x 768 pixels
1024 x 1024 pixels | Precision Matrix resolution
support image matrix sizes of:
512 x 512 pixels
768 x 768 pixels
1024 x 1024 pixels | Precision Matrix resolution
support image matrix sizes of:
512 x 512 pixels
768 x 768 pixels
1024 x 1024 pixels |
| Multi-Threshold Acquisition | | | N/A |
| Software property | Subject device | Primary predicate device | Secondary predicate device |
| | NAEOTOM Alpha
SOMARIS/10 syngo CT VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT VB30
(K230421) |
| | acquisition, storage, and reconstruction of projection raw data of different energy thresholds (T1, T2, T3 and T4) projection raw data of the three higher energy thresholds (T2, T3, and T3) can be reconstructed by offline reconstruction tools or on the CT system | acquisition, storage, and reconstruction of projection raw data of different energy thresholds (T1, T2, T3 and T4) projection raw data of the three higher energy thresholds (T2, T3, and T3) are reconstructed by offline reconstruction tools. | |
| myNeedle Guide | plan the needle path and perform control scans (i-Sequence, i-Spiral, i-Fluoro) myNeedle Detection algorithm (modification of the myNeedle Guide 3D software) | plan the needle path and perform control scans (i-Sequence, i-Spiral, i-Fluoro) | N/A |
| FAST 3D Camera with FAST Integrated Workflow | FAST Integrated Workflow including the sub-features FAST Range, FAST Isocentering and FAST Direction
FAST Range, FAST Isocentering and FAST Direction algorithms | FAST Integrated Workflow including the sub-features FAST Range, FAST Isocentering and FAST Direction | FAST Integrated Workflow including the sub-features FAST Range, FAST Isocentering and FAST Direction |
| Software property | Subject device | Primary predicate
device | Secondary predicate
device |
| | NAEOTOM Alpha
SOMARIS/10 syngo CT
VB10 | NAEOTOM Alpha
SOMARIS/10 syngo CT
VA50
(K220814) | SOMATOM Force
SOMARIS/7 syngo CT
VB30
(K230421) |
| | from the redesigned
camera. | | |
10
Image /page/10/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a series of orange dots that are arranged in a pattern.
11
Image /page/11/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are several orange dots that are arranged in a triangular shape.
12
Image /page/12/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" are several orange dots of varying sizes.
13
Image /page/13/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.
14
Image /page/14/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
Table 5: Software characteristics for the subject device NAEOTOM Alpha (software version SOMARIS/10 syngo CT VB10) compared to the predicate devices.
15
Image /page/15/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
16
Image /page/16/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
17
Image /page/17/Picture/0 description: The image contains the logos for Siemens and Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.
18
Image /page/18/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots.
19
Image /page/19/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
20
Image /page/20/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
21
Image /page/21/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
22
Image /page/22/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a cluster of orange dots.
Any differences in technological characteristics do not raise different questions of safety and effectiveness. Testing and validation is completed. Test results show that the subject device, the NAEOTOM Alpha with syngo CT VB10, is comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore are substantially equivalent to the predicate devices.
VIII. Performance Data
Non-Clinical Testing
Non-clinical testing, (integration and functional) including phantom tests were conducted for the NAEOTOM Alpha during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.
The general purpose of each test is to verify and validate the functionality of the subject device modifications.
Testing will cover all related subsystems that contribute to the device modifications. Test levels are defined. For each test level several test activities are performed. The test specification and acceptance criteria are related to the corresponding requirements. Various test activities are performed to specific modifications on different test levels to ensure safe and effective integration in the system. Three test levels are defined:
System Validation test:
- Acceptance test (workflow and user manual test)
- . Legal and Regulatory test
System Verification test:
- System Integration Test (functional)
- . Functionality verification
- . Image Quality (IQ) Evaluation
Tests are conducted for all software components developed in product development and for the complete product itself. Several activities are considered for this process, including creation of test specifications that relate to software/hardware requirements including tests to address risk mitigations that are identified, documented, and traced by hazard keys.
23
Image /page/23/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
Additional evaluation tests are performed as bench tests to support the new device or device modification on Non-Clinical Performance Testing as listed in Table 6 below.
Table 6: Non-clinical performance testing (bench testing).
| Feature/Non-clinical
supportive testing | Bench Testing performed |
---|---|
FAST 3D Camera / | |
FAST Integrated | |
Workflow | The bench test evaluates and compares the accuracy of the three sub- |
features FAST Isocentering, FAST Range, and FAST Direction to the | |
accuracy of the predicate device with syngo CT VA50 using the old | |
camera hardware and the then only available ceiling mount. | |
The objectives of the bench test are to demonstrate that the FAST 3D | |
Camera feature of the subject device with SOMARIS/10 syngo CT VB10, | |
where the algorithms have been optimized for a new camera hardware | |
in two mounting positions, achieves comparable results as the predicate | |
device with syngo CT VA50. | |
The FAST Isocentering accuracy of the subject device with syngo CT VB10 | |
is comparable to the predicate device with syngo CT VA50, regardless of | |
the camera mounting position. | |
For the FAST Range feature, the detection accuracy of all body region | |
boundaries is comparable between the subject device with syngo CT | |
VB10 and predicate device with syngo CT VA50. | |
The FAST Direction pose detection results are of comparable accuracy for | |
subject and predicate device, regardless of the camera mounting | |
position. | |
Overall, the SOMARIS/10 syngo CT VB10 delivers comparable accuracy to | |
the SOMARIS/10 syngo CT VA50 predicate for the new FAST 3D Camera | |
hardware. | |
Multi-Purpose Table | |
The range of possible movement for the mobile C-arm in different | |
positions between CT gantry and patient table was tested and | |
documented by measurement of angles. | |
Based on the test results it can be concluded that a CT scanner, equipped | |
with a Multi-Purpose (Vitus) Patient Table, which is installed with | |
enhanced distance (674 mm) to the CT gantry and offers the iCT mode | |
functionality, provides sufficient freedom of movement for a mobile C- | |
arm X-ray system to be used for clinical routine without any significant | |
limitations for my needle Laser or 3D Camera. | |
ZeeFree | The bench tests evaluate the performance of ZeeFree reconstruction. |
The objectives of the tests are to demonstrate: | |
• | |
that the number of artefacts which can be attributed to a stack | |
misalignment (e.g. discontinuities in vessel structures, | |
anatomical steps at air-soft-tissue interfaces, doubling of vessel | |
Feature/Non-clinical | |
supportive testing | Bench Testing performed |
or other anatomy) and which are often caused by incomplete patient breath-hold can be reduced in a "Cardiac Stack Artefact Correction" (SAC) reconstruction compared to the standard reconstruction with otherwise matching reconstruction parameters. | |
that no artefacts are introduced by a SAC reconstruction. The test results show: If misalignment artefacts are identified in non-corrected standard ECG-gated reconstructed sequence or spiral images, the feature "Cardiac Stack Artefact Correction" (SAC, marketing name: ZeeFree) enables optional stack artefact corrected images, which reduce the number of alignment artefacts. The SAC reconstruction does not introduce new artefacts, which were previously not present in the non-corrected standard reconstruction. The SAC reconstruction does realize equivalent image quality in quantitative standard physics phantom-based measurements (ACR, Gammex phantom) in terms of noise, homogeneity, high-contrast resolution, slice thickness and CNR compared to a non-corrected standard reconstruction. The SAC reconstruction does realize equivalent image quality in quantitative and qualitative phantom-based measurements with respect to metal objects compared to a non-corrected standard reconstruction. The SAC algorithm can be successfully applied to phantom data if derived from a suitable motion phantom demonstrating its correct technical function on the tested device. The SAC algorithm is independent from the physical detector width of the acquired data | |
myNeedle Guide | |
(with myNeedle | |
Detection) | Tests were performed to ensure clinical usability of the myNeedle Guide needle detection algorithm. Two individual tests were performed. The accuracy of the automatic needle detection algorithm was tested. The reduction of necessary user interactions for navigating to a needle-oriented view with and without the support of the automatic needle detection algorithm was analyzed. |
It has been shown that the algorithm was able to consistently detect needle-tips over a wide variety of scans in 90.76% of cases.
Further, the results of this bench test clearly shows that the auto needle detection functionality reduces the number of interactions steps needed to generate a needle-aligned view in the CT Intervention SW. Zero user interactions are required and a needle-aligned view is displayed right away after a new scan, if auto needle detection is switched on in the SW |
| Feature/Non-clinical
supportive testing | Bench Testing performed |
| | configuration. Therefore, the test is already passed if only a single user
interaction is necessary to achieve a needle-oriented view in the manual
workflow. |
| Quantum Spectral
Imaging | Tests were performed to demonstrate that:
T3D reconstructions in Quantumpeak mode are possible with the sharpest available kernels up to Qr89, Br98, etc. Quantumpeak scan mode allows reconstructions of monoenergetic images at energy levels from 40 to 190 keV. The evaluation has been performed based on phantom studies.
The results showed that: with T3D reconstructions from Quantumpeak scan modes, high resolution images with sharp kernel up to Br98 are obtained. The resolution is comparable to other Highresultra scan modes of the NAEOTOM Alpha. Monoenergetic reconstructions from Quantumpeak scan modes are free of artifacts. Measured CT values precisely match the reference values. The accuracy of monoenergetic reconstructions in iodine and calcium inserts at the NAEOTOM Alpha is comparable or better than on the secondary predicate device SOMATOM Force. |
| Quantum HD Cardiac | Quantitative assessment in terms of image noise and a visual assessment
of the image quality in ECG gated acquisitions between different scan
modes is performed: standard acquisition mode with spectral and non-
spectral reconstruction, ultra-high resolution acquisition mode with non-
spectral reconstruction (120x0.2 mm and 96x0.2 mm) and spectral image
reconstruction (limited to 96x0.2 mm).
Based on the results it can be concluded that substantial equivalence in
image quality is achieved by the images derived from the spectral
capable cardiac acquisition mode 96x0.2mm for both, the high-
resolution UHR and the standard resolution spectral image cases,
compared to the single source spectral capable 120x0.2mm UHR scan mode. |
| HD FoV | Tests were performed to evaluate the performance of HD FoV on
NAEOTOM Alpha.
For this purpose, the HU accuracy in the extended field of view region
was measured based on phantom studies. The phantom diameter
accuracy has been also evaluated.
Based on the results it can be concluded that HD FoV enables the
reconstruction of images while significantly improving the visualization |
| Feature/Non-clinical
supportive testing | Bench Testing performed |
| | of anatomy in the regions outside the scan field of view of 50 cm. In the
phantom study, an HU value accuracy of about $\pm$ 40 HU was achieved
with skin-line accuracy of about $\pm$ 3 mm. |
24
Image /page/24/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.
25
Image /page/25/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.
26
Image /page/26/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots that are arranged in a pattern that resembles an arrow.
A list of recognized and general consensus standards considered for the subject devices is provided as Table 7 and Table 8 below.
Table 7: Recognized Consensus Standards.
| Date of Entry | Recognition
Number | Standard
Developing
Organization | Standard
Designation Number
and Date | Title of Standard |
|---------------|-----------------------|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 12/19/2022 | 12-349 | NEMA | PS 3.1 - 3.20 2022d | Digital Imaging and
Communications in
Medicine (DICOM) Set |
| 07/06/2020 | 12-325 | NEMA | XR 25-2019 | Computed
Tomography Dose
Check |
| 07/06/2020 | 12-330 | NEMA | XR 28-2018 | Supplemental
Requirements for
User Information and
System Function
Related to Dose in CT |
| 12/23/2019 | 12-328 | IEC | 61223-3-5 Edition
2.0 2019-09 | Evaluation and
routine testing in
medical imaging
departments - Part 3-
5: Acceptance tests
and constancy tests -
Imaging performance
of computed
tomography X-ray
equipment [Including:
Technical
Corrigendum 1
(2006)] |
| 03/14/2011 | 12-226 | IEC | 61223-2-6 Second
Edition 2006-11 | Evaluation and
routine testing in
medical imaging
departments - Part 2-
6: Constancy tests -
Imaging performance
of computed
tomography X-ray
equipment |
| Date of Entry | Recognition
Number | Standard
Developing
Organization | Standard
Designation Number
and Date | Title of Standard |
| 01/14/2014 | 12-269 | IEC | 60601-1-3 Edition
2.1 2013-04 | Medical electrical
equipment - Part 1-3:
General requirements
for basic safety and
essential performance
- Collateral Standard:
Radiation protection
in diagnostic X-ray
equipment |
| 06/27/2016 | 12-302 | IEC | 60601-2-44 Edition
3.2: 2016 | Medical electrical
equipment - Part 2-
44: Particular
requirements for the
basic safety and
essential performance
of x-ray equipment
for computed
tomography |
| 12/23/2019 | 5-125 | ANSI AAMI
ISO | 14971: 2019 | Medical devices -
Applications of risk
management to
medical devices |
| | | ISO | 14971 Third Edition
2019-12 | Medical devices -
Application of risk
management to
medical devices |
| 01/14/2019 | 13-79 | ANSI AAMI
IEC | 62304:2006/A1:2016 | Medical device
software - Software
life cycle processes
[Including
Amendment 1 (2016)] |
| | | IEC | 62304 Edition 1.1
2015-06
CONSOLIDATED
VERSION | Medical device
software - Software
life cycle processes |
| 07/09/2014 | 19-46 | ANSI AAMI | ES60601-
1:2005/(R)2012 &
A1:2012,
C1:2009/(R)2012 &
A2:2010/(R)2012
(Cons. Text) [Incl.
AMD2:2021] | Medical electrical
equipment - Part 1:
General requirements
for basic safety and
essential performance
(IEC 60601-1:2005,
MOD) [Including
Amendment 2 (2021)] |
| Date of Entry | Recognition
Number | Standard
Developing
Organization | Standard
Designation Number
and Date | Title of Standard |
| 09/17/2018 | 19-36 | ANSI AAMI
IEC | 60601-1-2:2014
[Including AMD
1:2021] | Medical electrical
equipment - Part 1-2:
General requirements
for basic safety and
essential performance - Collateral Standard:
Electromagnetic
disturbances -
Requirements and
tests |
| | | IEC | 60601-1-2 Edition
4.1 2020-09
CONSOLIDATED
VERSION | Medical electrical
equipment - Part 1-2:
General requirements
for basic safety and
essential performance - Collateral Standard:
Electromagnetic
disturbances -
Requirements and
tests |
| 12/23/2016 | 5-129 | ANSI AAMI
IEC | 62366-
1:2015+AMD1:2020
(Consolidated Text) | Medical devices Part
1: Application of
usability engineering
to medical devices,
including Amendment
1 |
| | | IEC | 62366-1 Edition 1.1
2020-06
CONSOLIDATED
VERSION | Medical devices - Part
1: Application of
usability engineering
to medical devices |
| 07/09/2014 | 12-273 | IEC | 60825-1 Edition 2.0
2007-03 | Safety of laser
products - Part 1:
Equipment
classification, and
requirements |
| 12/21/2020 | 5-132 | IEC | 60601-1-6 Edition
3.2 2020-07
CONSOLIDATED
VERSION | Medical electrical
equipment - Part 1-6:
General requirements
for basic safety and
essential performance - Collateral standard:
Usability |
| Date of Entry | Recognition
Number | Standard
Developing
Organization | Standard
Designation Number
and Date | Title of Standard |
| 12/23/2019 | 12-309 | IEC | 60601-2-28 Edition
3.0 2017-06 | Medical electrical
equipment - Part 2-
28: Particular
requirements for the
basic safety and
essential performance
of X-ray tube
assemblies for
medical diagnosis |
| 12/20/2021 | 12-341 | IEC | 62563-1 Edition 1.2
2021-07
CONSOLIDATED
VERSION | Medical electrical
equipment - Medical
image display systems - Part 1: Evaluation
methods |
27
Image /page/27/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots in a cluster.
28
Image /page/28/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are several orange dots that are arranged in a triangular shape.
29
Image /page/29/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots that are arranged in a pattern that resembles an arrow.
Table 8: General Use Consensus Standards.
| Standard
Developing
Organization | Standard Designation
Number and Date | Title of Standard | How was Standard Used |
|----------------------------------------|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| IEC | 60601-
1:2005+A1:2012+A2:2020 | Medical electrical
equipment - part 1:
general requirements
for basic safety and
essential performance | ANSI AAMI ES60601-
1:2005/(R)2012 &
A1:2012,
C1:2009/(R)2012 &
A2:2010/(R)2012 (Cons.
Text) [Incl. AMD2:2021] |
| IEC/ISO | 17050-1 | Conformity Assessment
– Supplier's declaration
of conformity – Part 1:
General requirements | Declaration of
conformance to FDA
recognized consensus
standards. |
| IEC/ISO | 17050-2 | Conformity assessment
– Supplier's declaration
of conformity - Part 2:
Supporting
documentation. | General consensus
standards not currently
recognized by FDA. |
A list of applicable guidance documents considered for this submission is provided as Table 9 below. Table 9: FDA Guidance Document and Effective Date
FDA Guidance Document | Issue date |
---|---|
User Fees and Refunds for Premarket Notification Submissions (510(k)s | 10/05/2022 |
Refuse to Accept Policy for 510(k)s | 04/21/2022 |
30
Image /page/30/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern that resembles a comet tail.
FDA Guidance Document | Issue date |
---|---|
Electronic Submission Template for Medical Device 510(k) Submissions | 10/2/2023 |
Deciding When to Submit a 510(k) for a Change to an Existing Device | 10/25/2017 |
The 510(k) Program: Evaluating Substantial Equivalence in Premarket | |
Notifications [510(k)] | 07/28/2014 |
Content of Premarket Submissions for Device Software Functions | 06/14/2023 |
Off-The-Shelf Software Use in Medical Devices | 09/27/2019 |
Applying Human Factors and Usability Engineering to Medical Devices | 02/03/2016 |
Pediatric Information for X-ray Imaging Device Premarket Notifications | 11/28/2017 |
Cybersecurity in Medical Devices: Quality System Considerations and Content | |
of Premarket Submissions | 09/27/2023 |
Electromagnetic Compatibility (EMC) of Medical Devices | 06/06/2022 |
Design Considerations and Pre-market Submission Recommendations for | |
Interoperable Medical Devices | 09/06/2017 |
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions | |
for Medical Devices | 09/14/2018 |
Verification and Validation
The Risk Analysis was completed and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claims of substantial equivalence.
Cybersecurity
Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.
Wireless Coexistence Testing
Additionally, Siemens conforms to the requirements for Radio Frequency Wireless Technology as defined in FDA guidance document "Radio Frequency Wireless Technology in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, issued on August 14, 2013" by adhering to the EMC and risk based verification and validation requirements in design, testing, and labeling of the wireless remote control components of the subject devices.
The Radio Frequency Wireless Technology of the optional Remote Scan Control and supporting Control Device iPad for Scan&GO complies to 47 CFR part 15 subpart c – Intentional Radiators. All Radio device labels will show an FCC ID code to show compliance. Shielding requirement applicable to the NAEOTOM Alpha and respective Scatter Radiation diagrams for typical room installations are provided in the User Documentation and Planning Guide of the intended Scanners in accordance with IEC60601-2-44.
31
Image /page/31/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
Siemens has considered several measures to address wireless coexistence by design to ensure the safe operation of the wireless components in combination with the applicable system supported functionality. Wireless technology in the system setup to perform a task in a given shared environment where other systems have an ability to perform their tasks and may or may not be using the same set of rules has been considered. According to FDA guidance 'Radio Frequency Wireless Technology in Medical Devices" Siemens has addressed the safety, effectiveness, and high likelihood of coexistence with other devices of this technology in our product design by our Risk Management Process, Failure Mode and Effects Analysis (FMEA) Process, and Requirement Engineering Process. As part of the risk management process, hazardous situations associated with the Scan&GO and its connection to the host system via Wi-Fi were addressed as part of the Risk Management process.
Testing for co-existence considered for following scenarios:
- Co-Channel Testing
- . Adjacent Channel Testing
- . RF Interference Testing
- Separation Distance/Location Testing
Scan&GO is designed to allow dynamic frequency selection and transmission power control by default in accordance with IEEE 802.11h. Adjacent channel testing is addressed by the fact that Scan&GO does not support shared medium access to Siemens Wi-Fi network. RF interference was tested by successfully ensuring that wireless communications were actively transmitting in situations where possible interference may exist. Recommended distance and router locations requirements are documented in the user documentation.
Summary
The features described in this premarket notification are supported with verification and validation testing, dosimetry and imaging performance, and analysis of phantom images to assess device and feature performance during product development. The risk analysis was completed, and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria. Verification and validation testing of the device was found acceptable to support the claim of substantial equivalence.
General Safety and Effectiveness Concerns
The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a system related risk analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing according to the risk management process. In order to minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
IX. Conclusions
The non-clinical data supports the safety of the device and the hardware and software verification and validation demonstrates that the NAEOTOM Alpha performs as intended in the specified use conditions. Verification and validation, clinical/patient and phantom testing were performed. The
32
Image /page/32/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
data included in this submission demonstrates that the NAEOTOM Alpha with described modifications performs comparably to the predicate devices currently marketed for the same intended use. The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate devices.