(28 days)
Axium™ Detachable Coil:
Axium™ Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. Axium™ Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
Axium™ Prime Detachable Coil:
(Models: APB-X-Y-3D-ES, APB-X-Y-3D-SS, APB-X-Y-HX-ES, APB-X-Y-HX-SS)
The Axium™ Prime Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. The Axium™ Prime Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
Axium™ Prime Detachable Coil:
(Models: FC-X-Y-3D)
The Axium™ Prime Detachable Col is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature.
The Axium™ Detachable Coil and Axium™ Prime Detachable Coil consist of a platinum embolization coil attached to a composite implant delivery pusher with a radiopaque positioning marker and a hand-held Instant Detacher (I.D.) which when activated detaches the coil from the delivery pusher tip. The I.D. is sold separately.
The document provided focuses on the substantial equivalence of the "Axium™ Detachable Coil" and "Axium™ Prime Detachable Coil" with the addition of fluorosafe markers. The primary purpose of the submission is to demonstrate that these changes do not raise new questions of safety and effectiveness compared to the predicate devices. Therefore, the "device" in question is not an AI/ML powered diagnostic device, but rather a neurovascular embolization device with a minor design change.
As such, many of the typical acceptance criteria and study details relevant to AI/ML powered devices, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details, are not applicable or provided in this document. The provided text describes bench testing and biocompatibility testing for the physical device itself.
However, I can extract the relevant "acceptance criteria" and "reported device performance" as presented for the specific tests conducted for this submission.
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Implicit from "Results" column) | Reported Device Performance |
|---|---|---|
| Biocompatibility | ||
| Cytotoxicity | Not induce cytotoxicity | Did not induce cytotoxicity. |
| Sensitization | Not be considered a sensitizer | Were not considered a sensitizer. |
| Irritation | Be considered non-irritant | Are considered non-irritant. |
| Acute Systemic Toxicity | Show no mortality or evidence of acute systemic toxicity | Showed no mortality or evidence of acute systemic toxicity. |
| Indirect (extract) Hemolysis | Be considered non-hemolytic | Are considered non-hemolytic. |
| Material-Mediated Pyrogenicity | Meet USP 151 requirements and be non-pyrogenic | Met the requirements and were found to be non-pyrogenic. |
| Bench Testing | ||
| Visual Inspection | Darkness of fluorosafe etch mark and 360° etch mark around delivery pusher meet specifications | Met the acceptance criteria for visual inspection. |
| Marker Dimensional (Marker Position and Total Marker Length) Inspection | Total length of fluorosafe markers and their position on the delivery pusher meet specifications | Met the acceptance criteria for marker dimensional inspection. |
| Corrosion Resistance | No corrosion on the delivery pusher after soaking in saline and immersion in boiling water per ISO 10555-1, Annex A | Met the acceptance criteria for corrosion resistance. |
| Fluorosafe Marker Visibility | Meet user needs for visibility under simulated use conditions | Met the user needs for which it was designed and tested. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated for each test, but standard for device verification and validation. For biocompatibility, animal models were used (guinea pig, rabbit, mice). For bench tests, "devices" (plural) were evaluated, implying a sample size greater than one but not specified numerically.
- Data Provenance: Not applicable in the context of AI/ML data sets. These are laboratory tests on physical devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable. The "ground truth" for this device modification is established through physical and objective measures against predefined specifications and recognized international standards (e.g., ISO 10993, USP 151) and user needs (for marker visibility). The "clinical users" who evaluated fluorosafe marker visibility are mentioned, but their number and specific qualifications are not detailed.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. Adjudication methods are typically for subjective assessments (e.g., image interpretation). These tests involve objective measurements and evaluations against specified criteria and standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This submission is for a physical medical device (embolization coil) with a minor design change (adding fluorosafe markers), not an AI-powered diagnostic tool. No MRMC study was conducted.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" here is based on objective measurements against established engineering specifications, chemical/biological compatibility standards (ISO 10993, USP 151), and documented user needs. For example, meeting the criteria for "non-cytotoxic" or demonstrating "no corrosion."
8. The sample size for the training set
- Not applicable. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established
- Not applicable. There is no "training set."
{0}------------------------------------------------
November 7, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Lekshmi Pillai, MS Senior Regulatory Affairs Specialist 5290 California Avenue Irvine, California 92617
Re: K233420
Trade/Device Name: Axium Detachable Coil; Axium Prime Detachable Coil Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: October 9, 2023 Received: October 10, 2023
Dear Lekshmi Pillai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K233420
Device Name Axium™ Detachable Coil; Axium™ Prime Detachable Coil
Indications for Use (Describe)
Axium™ Detachable Coil:
Axium™ Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. Axium™ Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
Axium™ Prime Detachable Coil:
(Models: APB-X-Y-3D-ES, APB-X-Y-3D-SS, APB-X-Y-HX-ES, APB-X-Y-HX-SS)
The Axium™ Prime Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. The Axium™ Prime Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
Axium™ Prime Detachable Coil:
(Models: FC-X-Y-3D)
The Axium™ Prime Detachable Col is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K233420 510(k) Summary
| 510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular9775 Toledo WayIrvine, CA 92618, USAEstablishment Registration: 2029214 |
|---|---|
| Contact Person: | Lekshmi PillaiSenior Regulatory Affairs SpecialistTelephone: (917) 291-3810Email : lekshmiajaykumar.pillai@medtronic.com |
| Date SummaryPrepared: | 06 November 2023 |
|---|---|
| Trade Name of Device: | Axium™ Detachable CoilAxium™ Prime Detachable Coil |
| Common Name ofDevice: | Neurovascular Embolization Device, Vascular Embolization Device |
| Review Panel: | Neurology, Cardiovascular |
| Product Code: | HCG, KRD |
| Regulation Number: | 21 CFR 882.5950; 21 CFR 870.3300 |
| Regulation Name: | Neurovascular Embolization Device, Vascular Embolization Device |
| Device Classification | Class II |
| Predicate Device: | K203432Axium™ Detachable CoilAxium™ Prime Detachable Coil |
Device Description:
The Axium™ Detachable Coil and Axium™ Prime Detachable Coil consist of a platinum embolization coil attached to a composite implant delivery pusher with a radiopaque positioning marker and a hand-held Instant Detacher (I.D.) which when activated detaches the coil from the delivery pusher tip. The I.D. is sold separately.
Indications for Use Statement:
| Axium™ Detachable Coil and Axium Prime Detachable Coil Indications for Use | ||
|---|---|---|
| Model Numbers | Trade Name | Indications for Use |
| QC-X-Y-HELIX | ||
| QC-X-Y-3D | Axium™DetachableCoil | Axium™ Detachable Coils are intended for the endovascular embolizationof intracranial aneurysms. Axium™ Detachable Coils are also intended forthe embolization of other neuro vascular abnormalities such asarteriovenous malformations and arteriovenous fistulae. |
| NC-X-Y-HELIX | ||
| PC-X-Y-HELIX | ||
| PC-X-Y-3D |
{4}------------------------------------------------
| Axium™ Detachable Coil and Axium Prime Detachable Coil Indications for Use | ||
|---|---|---|
| Model Numbers | Trade Name | Indications for Use |
| APB-X-Y-3D-ES | Axium™PrimeDetachableCoil | The Axium™ Prime Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. The Axium™ Prime Detachable Coils are also intended for the embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. |
| APB-X-Y-3D-SS | ||
| APB-X-Y-HX-ES | ||
| APB-X-Y-HX-SS | ||
| FC-X-Y-3D | Axium™PrimeDetachableCoil | The Axium™ Prime Detachable Coil is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Prime Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature. |
The model numbers are formatted to summarize sizes available, where X is the coil loop outer diameter in mm and Y is the implant length in cm.
Proposed Change:
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular requests clearance for the Axium™ Detachable Coil and Axium™ Prime Detachable Coil with addition of fluorosafe markers on the proximal end of the coil delivery pusher.
Device Comparison:
| Design Feature | Predicate Devices (K203432):Axium™ Detachable Coil;Axium™ Prime Detachable Coil | Subject Devices:Axium™ DetachableCoil;Axium™ PrimeDetachable Coil |
|---|---|---|
| Indications for Use | Axium™ Detachable Coil:Axium™ Detachable Coils are intended for the endovascularembolization of intracranial aneurysms. Axium™ Detachable Coilsare also intended for the embolization of other neuro vascularabnormalities such as arteriovenous malformations andarteriovenous fistulae.Axium™ Prime Detachable Coil (APB-X-Y-3D-ES, APB-X-Y-3D-SS,APB-X-Y-HX-ES, APB-X-Y-HX-SS):The Axium™ Prime Detachable Coils are intended for theendovascular embolization of intracranial aneurysms. The Axium™Prime Detachable Coils are also intended for the embolization ofother neuro vascular abnormalities such as arteriovenousmalformations and arteriovenous fistulae.Axium™ Prime Detachable Coil (FC-X-Y-3D):The Axium™ Prime Detachable Coil is intended for the endovascularembolization of intracranial aneurysms and other neurovascularabnormalities, such as arteriovenous malformations andarteriovenous fistulae. The Axium™ Prime Detachable Coils are also | Same |
| Design Feature | Predicate Devices (K203432):Axium™ Detachable Coil;Axium™ Prime Detachable Coil | Subject Devices:Axium™ DetachableCoil;Axium™ PrimeDetachable Coil |
| intended for arterial and venous embolizations in the peripheralvasculature. | ||
| Visual Markers | None | Fluorosafemarkers |
| Dimensions | ||
| Device Size Range | Axium™ Bare Helix:1.5-20 mm - Loop OD1-50 cm - LengthAxium™ Bare 3D:2-25 mm - Loop OD2-50 cm - LengthAxium™ Nylon 3D:2-4 mm – Loop OD1-10 cm – LengthAxium™ PGLA Helix:2-10 mm – Loop OD1-30 cm – LengthAxium™ PGLA 3D:2-18 mm – Loop OD2-40 cm – LengthAxium™ Prime 3D:1-6 mm – Loop OD2-20 cm – LengthAxium™ Prime Helix:1-6 mm – Loop OD1-20 cm – LengthAxium™ Prime 3D:3-25 mm – Loop OD6-50 cm – Length | Same |
| Coil Shape | Helical and 3D | Same |
| Compatible Accessories | ||
| CatheterCompatibility | Axium™ Detachable Coils (Bare) should be delivered throughmicrocatheters with minimum ID of 0.0165".Axium™ Detachable Coils (Nylon and PGLA) should be deliveredthrough microcatheters with minimum ID of 0.0165"-0.020". | SameSame |
| Axium™ Prime Detachable Coils should be delivered throughmicrocatheters with minimum ID of 0.0165"-0.017". | Axium™ PrimeDetachable Coilsshould be deliveredthrough | |
| Design Feature | Predicate Devices (K203432):Axium™ Detachable Coil;Axium™ Prime Detachable Coil | Subject Devices:Axium™ Detachable Coil;Axium™ PrimeDetachable Coilmicrocatheters withminimum ID of0.0165". |
| Guide CatheterCompatibility | 6-8F | Same |
| Method of CoilDetachment | Instant Detacher - standalone hand-held mechanical unit that, when connected to the proximal end of the pusher, pulls the release element inside of the pusher, resulting in release of the implant from the distal end of the delivery pusher. | Same |
| Material of Construction | ||
| Main Coil | Platinum/Tungsten alloy | Same |
| Tension Safe/Stretch ResistantFilament | Polypropylene | Same |
| Delivery System | Pusher: Composite hypotube stainless steel with PTFE outer jacket and liner and PET & Pt/w alignment marker | Same |
| Introducer Sheath | Polypropylene and HDPE | Same |
| Sterilization | ||
| Method of Supply | Sterile and single use | Same |
| SterilizationMethod | Ethylene Oxide (EO) | Same |
| Packaging | ||
| Pouch Material | For PC-X-Y-HELIX and PC-X-Y-3D:PLK235/Uncoated 2FS TyvekFor all other models:PET/Tyvek | SameFor all other models:Nylon/Tyvek |
| Pouch Dimensions | For PC-X-Y-HELIX and PC-X-Y-3D:17" X 9.8" (pre-sterilization)For all other models:10 1/8" X 10.838" | SameFor all other models:11.98" X 10.94" |
| Carton | Cardboard, Solid Bleach Sulfate | Same |
| MRI Card | None | 3 1/2" x 4" (+/-1/8"), 1/1 Black -Black and white,2 sided |
| Stability | ||
| Shelf Life | 36 months | Same |
| MRI Compatibility | Magnetic Resonance ImagingMR Conditional | Same |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Biocompatibility:
The fluorosafe marker (subject to this submission) added onto the proximal end of the delivery pusher is considered external communicating device having indirect blood contact via the fluid path, as it is only in contact with the delivery catheter and does not have direct exposure to circulating blood. The distal end of the delivery pusher is categorized as external communicating device with limited exposure duration (د 24 hrs.) with circulating blood. The embolization coil is considered an implant with long-term (> 30 days) direct contact with circulating blood. There are no material or contact duration changes to the components of the device that directly contact circulating blood in the current submission. Therefore, prior hemocompatibility (complement activation, direct hemolysis, and thrombogenicity) tests demonstrate the biocompatibility of the unmodified regions and components of the subject device that contact blood directly.
To assess the addition of fluorosafe markers on the proximal end of the delivery pusher, the following biocompatibility tests were performed on the final, finished delivery pusher subassemblies incorporating the fluorosafe marker.
| Test Description | Study Method | Results |
|---|---|---|
| Cytotoxicity | Axium™ Detachable Coil and Axium™ PrimeDetachable Coil were evaluated per ISO 10993-5: 2009, using the MEM elution method. | The Axium™ Detachable Coil andAxium™ Prime Detachable Coil didnot induce cytotoxicity. |
| Sensitization | Axium™ Detachable Coil and Axium™ PrimeDetachable Coil were evaluated per ISO 10993-10:2021, using a guinea pig maximization test. | The Axium™ Detachable Coil andAxium™ Prime Detachable Coilwere not considered a sensitizer. |
| Irritation | Axium™ Detachable Coil and Axium™ PrimeDetachable Coil were evaluated per ISO 10993-23:2021, using a rabbit intracutaneous reactivitytest. | The Axium™ Detachable Coil andAxium™ Prime Detachable Coil areconsidered non-irritant. |
| AcuteSystemicToxicity | Axium™ Detachable Coil and Axium™ PrimeDetachable Coil were evaluated per ISO 10993-11:2017, using an acute systemic injection testin mice. | The Axium™ Detachable Coil andAxium™ Prime Detachable Coilshowed no mortality or evidence ofacute systemic toxicity. |
| Indirect(extract)Hemolysis | Axium™ Detachable Coil and Axium™ PrimeDetachable Coil were evaluated per ISO 10993-4:2017, using ASTM hemolysis study. | The Axium™ Detachable Coil andAxium™ Prime Detachable Coil areconsidered non-hemolytic. |
| Material-MediatedPyrogenicity | Axium™ Detachable Coil and Axium™ PrimeDetachable Coil were evaluated per USP 151(2020), using USP rabbit pyrogen test -material-mediated. | The Axium™ Detachable Coil andAxium™ Prime Detachable Coil metthe requirements and were foundto be non-pyrogenic. |
The Axium™ Detachable Coil and Axium™ Prime Detachable Coil have been evaluated to meet requirements specified in FDA guidance, "Use of International Standard ISO 10993-1, "Biocompatibility Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process", " issued
{8}------------------------------------------------
on September 8, 2023.
Performance Data – Bench:
Non-clinical bench testing was conducted to evaluate the performance of Axium™ Detachable Coil and Axium™ Prime Detachable Coil in a clinically representative neurovascular access model. The successful results of the testing demonstrated that the changes do not raise new questions of safety and effectiveness, supporting the substantial equivalence to the predicate devices.
The following non-clinical bench tests were conducted:
| Design Verification | ||
|---|---|---|
| Test | Test Method Summary | Results |
| Visual Inspection | Visual inspection verified for thedarkness of the etch mark offluorosafe markers and the 360°etch mark around the deliverypusher. | Axium™ Detachable Coil andAxium™ Prime Detachable Coilmet the acceptance criteria forvisual inspection. |
| Marker Dimensional (MarkerPosition and Total MarkerLength) Inspection | Marker dimensional inspectionverified the total length of thefluorosafe markers and theirposition on the delivery pusher. | Axium™ Detachable Coil andAxium™ Prime Detachable Coilmet the acceptance criteria formarker dimensional inspection. |
| Corrosion Resistance | The test verified the corrosionresistance on the delivery pusherusing the method of soaking thedevices in saline followedbyimmersion in boiling water per themethod specified in Annex A of ISO10555-1. | Axium™ Detachable Coil andAxium™ Prime Detachable Coilmet the acceptance criteria forcorrosion resistance. |
| Design Validation | |||||
|---|---|---|---|---|---|
| Test | Test Method Summary | Results | |||
| Fluorosafe Marker Visibility | The clinical users evaluated thedevicesunder simulateduseconditions and rated fluorosafemarker visibility. | The devices met the user needsfor which it was designed andtested. |
Performance Data – Animal and Clinical:
The determination of substantial equivalence is based upon non-clinical bench testing as there is no change to the indications for use statement, fundamental scientific technology, principles of operation or materials of construction except for the addition of fluorosafe markers. No clinical data or animal study were deemed necessary to support the subject device submission.
{9}------------------------------------------------
Conclusion:
There is no change to the indications for use statement, fundamental scientific technology or principle of operation of the Axium™ Detachable Coils and Axium™ Prime Detachable Coils in comparison to the predicate devices. The addition of fluorosafe markers on the proximal end of the embolization coil delivery pusher does not raise new questions of safety and effectiveness and is supported by non-clinical bench testing using well-established acceptable scientific methods. The information provided in this submission supports a determination of substantial equivalence for Axium™ Detachable Coils and Axium™ Prime Detachable Coils to the predicate devices.
§ 882.5950 Neurovascular embolization device.
(a)
Identification. A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).