K Number
K233347
Manufacturer
Date Cleared
2023-12-04

(66 days)

Product Code
Regulation Number
878.4350
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Verruca-Freeze® H cryosurgical system is indicated for use in the following skin lesions:

  • · Actinic Keratosis
  • · Genital Warts
  • Lentigo
  • Molluscum Contaqiosum
  • Seborrheic Keratosis
  • Skin Tags
  • · Verruca Plantaris
  • · Verruca Vulgaris • Verruca Plana
  • Verruca Plana
Device Description

The Verruca-Freeze® H cryosurgical system is used for treatment of benign skin lesions.
The device consists of:

  • Canister filled with a liquid mixture of the compressed gases dimethyl ether, propane, and isobutane
  • . Polyurethane foam swabs
  • . Detailed instructions including illustrated descriptions
    The main component of the device is the pressurized aerosol canister containing cryogen spray. The canister is used with foam swabs, which are non-sterile, single-use applicators that are disposed after each use. The foam swabs allow the cryogen to be localized to a target lesion.
    The one- to two-minute procedure begins by spraying cryogen onto a foam swab, which is then placed directly onto the lesion and held for 15-40 seconds. Upon removal of the swab from the skin, the frozen tissue appears white and begins thawing. The typical thaw time is 20-40 seconds.
    This treatment methodology is simple and has been an accepted practice used by physicians for decades. It utilizes a standard cryogen composition profile to freeze lesions and destroy the underlying tissue through evaporative cooling. On the cellular level, both extracellular and intracellular ice formation are occurring.
AI/ML Overview

The provided document is a 510(k) summary for the Verruca-Freeze® H cryosurgical system. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving that the device meets specific acceptance criteria through a machine learning or AI algorithm's performance.

Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving a device meets these criteria in the context of an AI/ML powered device. The "performance data" mentioned refers to non-clinical tests of the cryosurgical process itself, not an algorithm's performance.

Here's why the document cannot answer the prompt's specific questions about AI/ML device performance:

  • Device Type: The Verruca-Freeze® H is a cryosurgical system, a physical medical device that applies cold to treat skin lesions. It is not an AI-powered or software-based medical device.
  • "Study" Context: The "performance testing" described in Section 6.2 ("Non-Clinical Performance Data") is about measuring the output temperature and effectiveness of the cryogen and applicators, and visually confirming ice ball formation. This is physical hardware performance testing, not an algorithmic study.
  • Absence of AI/ML Specifics: There is no mention of algorithms, training sets, test sets, ground truth establishment by experts, adjudication methods, or MRMC studies, all of which are relevant to evaluating AI/ML-based medical devices.
  • "Clinical Performance Data": Section 6.3 explicitly states, "There was no clinical testing required to support the medical device as the indications for use are equivalent to those of the predicate device. These types of devices have been on the market for decades and their clinical safety and efficacy has been established." This further confirms that no new clinical study was conducted for this submission, let alone one focused on AI/ML performance.

In summary, the provided document is a regulatory submission for a traditional medical device demonstrating substantial equivalence, not an AI/ML device showing performance against specific acceptance criteria.

Therefore, I cannot populate the requested table or answer the specific questions about the study design (sample sizes, expert qualifications, adjudication, MRMC, standalone performance, ground truth types, training set details) because this information is not present in the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 4, 2023

CryoSurgery, Inc. % Rachel Paul Lead Consultant QA&RA Emergo Europe Consulting Prinsessegracht 20 The Hague, 2514AP Netherlands

Re: K233347

Trade/Device Name: Verruca-Freeze® H Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: September 29, 2023 Received: September 29, 2023

Dear Rachel Paul:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation reguires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark Trumbore -SDigitally signed by Mark Trumbore -SDate: 2023.12.04 08:24:35 -05'00'
---------------------------------------------------------------------------------------------
Mark Trumbore, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K233347

Device Name

Verruca-Freeze® H

Indications for Use (Describe)

The Verruca-Freeze® H cryosurgical system is indicated for use in the following skin lesions:

  • · Actinic Keratosis

  • · Genital Warts

  • Lentigo

  • Molluscum Contaqiosum

  • Seborrheic Keratosis

  • Skin Tags

  • · Verruca Plantaris

  • · Verruca Vulgaris • Verruca Plana

  • Verruca Plana

Type of Use (Select one or both, as applicable)

ر Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:

1. Submitter Information

Company:Ronald A. McDow, MD
Medical Director
CryoSurgery, Inc.
5829 Old Harding Pike
Nashville, TN 37205 USA
Telephone: (615) 354-0414
Fax: (615) 354-0466
rmcdow@comcast.net
Contact:Rachel Paul
Lead Consultant QA&RA
Emergo Europe Consulting
Prinsessegracht 20
The Hague, 2514AP The Netherlands
Telephone: + 31 (0) 70 345 85 70/Direct: + 31(0) 70 850 82 49
Fax:
LST.AUS.ProjectManagement@ul.com
Date Summary Prepared:November 27, 2023
2. Name of the Device
Trade Name:Verruca-Freeze® H
Common Name:Unit, Cryosurgical, Accessories
Classification Name:General & Plastic Surgery
Review Panel:General & Plastic Surgery (SU)
Regulation:878.4350
Class:Class II
Product Code:GEH

{4}------------------------------------------------

3. Equivalence Claimed to Predicate Device

The Verruca-Freeze® H is equivalent to the CryoDose H Or Similar (K161337), manufactured by Nuance Medical, LLC.

4. Device Description

The Verruca-Freeze® H cryosurgical system is used for treatment of benign skin lesions.

The device consists of:

  • Canister filled with a liquid mixture of the compressed gases dimethyl ether, propane, and isobutane
  • . Polyurethane foam swabs
  • . Detailed instructions including illustrated descriptions

The main component of the device is the pressurized aerosol canister containing cryogen spray. The canister is used with foam swabs, which are non-sterile, single-use applicators that are disposed after each use. The foam swabs allow the cryogen to be localized to a target lesion.

The one- to two-minute procedure begins by spraying cryogen onto a foam swab, which is then placed directly onto the lesion and held for 15-40 seconds. Upon removal of the swab from the skin, the frozen tissue appears white and begins thawing. The typical thaw time is 20-40 seconds.

This treatment methodology is simple and has been an accepted practice used by physicians for decades. It utilizes a standard cryogen composition profile to freeze lesions and destroy the underlying tissue through evaporative cooling. On the cellular level, both extracellular and intracellular ice formation are occurring.

Substantial Equivalence Table
CharacteristicProposed Device: Verruca-Freeze® HPrimary Predicate Device:CryoDose H Or Similar(K161337)
Intended UseThe Verruca-Freeze Hcryosurgical system is usedfor treatment of benign skinlesions.The CryoDose H is used fortreatment of benign skinlesions.

5. Substantial Equivalence Comparison

{5}------------------------------------------------

Indication for UseThe Verruca-Freeze Hcryosurgical system isindicated for use in thetreatment of the followingskin lesions: ActinicKeratosis, Genital Warts,Lentigo, MolluscumContagiosum, SeborrheicKeratosis, Skin Tags, VerrucaPlantaris, Verruca Vulgaris,and Verruca PlanaThe CryoDose H indicationsfor use as follows: CryoDoseH is indicated for use in thetreatment of the following:actinic keratosis, genitalwarts, lentigo, molluscumcontagiosum, seborrheickeratosis, skin tags, verrucaplantaris, verruca vulgaris,verruca plana
Product CodeGEHGEH
Regulation Number878.4350878.4350
Classification NameUnit, Cryosurgical,AccessoriesUnit, Cryosurgical,Accessories
Mechanism of ActionApplication of extreme coldcausing tissue destructionApplication of extreme coldcausing tissue destruction
Cryogen Gas MaterialDimethyl Ether (95%),Propane (2%) and Isobutane(3%)Dimethyl Ether (95%),Propane (2%) and Isobutane(3%)
Freeze Time15-40 seconds15-40 seconds
ApplicatorsFoam swabsFoam swabs
SterileNoNo
OTC or RxRxRx
DesignAerosol-filled canister andparts within the canister(valve, actuator, cover cap)Aerosol-filled canister andparts within the canister(valve, actuator, cover cap)
Freezing temperature-50°C-50°C
Shelf life2-year2-year

6. Additional Information

6.1 Substantial Equivalence Discussion

The technological characteristics of Verruca-Freeze® H, including materials, design, and energy source, are identical to the predicate device. The table as above compares Verruca-Freeze® H to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence.

Other elements of Verruca-Freeze® H not mentioned in the table that have the same technical characteristics as the predicate device include:

{6}------------------------------------------------

  • Canister ●
  • . Valve
  • Actuator ●
  • Cover Cap

Verruca-Freeze® H has the same technological characteristics as the predicate device. Both devices use the same design, energy source, materials, and other features. When the cryogen is dispensed, it saturates the foam swab is then applied directly to the desired area of treatment. Subsequently, the skin lesion is frozen and destroyed. Verruca-Freeze® H does not raise any new questions of safety or effectiveness as compared to the predicate device.

It is concluded that Verruca-Freeze® H is substantially equivalent to the predicate device.

6.2 Non-Clinical Performance Data

Performance testing was conducted to determine the output temperature and effectiveness of Verruca-Freeze® H as compared to the predicate device. The output temperature was measured in the cryogen-saturated foam-tipped applicators followed by a visual confirmation of an "ice ball" on a neoprene pad. The tests determined that Verruca-Freeze® H cryogen and applicators perform equivalently to the cryogen and applicators of the predicate device. Further, the tests demonstrated that the device is capable of reaching the minimum required temperature to destroy the underlying tissue of a skin lesion.

6.3 Clinical Performance Data

There was no clinical testing required to support the medical device as the indications for use are equivalent to those of the predicate device. These types of devices have been on the market for decades and their clinical safety and efficacy has been established. The nonclinical testing detailed in this submission supports the substantial equivalence of the device.

7. Substantial Equivalence Conclusion

Based on the device characteristics and testing, the Verruca-Freeze® H is substantially equivalent to the predicate device and raises no new questions related to safety or effectiveness.

§ 878.4350 Cryosurgical unit and accessories.

(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.